OIML/TC11/SC3/N5 ORGANISATION INTERNATIONAL DE METROLOGIE LEGALE International Recommendation Fifth Commitee Draft of Recommendation Blackbody Radiators for Calibration of Radiation Thermometers Calibration and Verification Procedures Subcommittee: TC11/SC3 “Radiation Thermometers” P-Members: Czech Republic, Germany, Japan, Korea (R.), Netherlands, Russian Federation, Slovakia, United Kingdom, United States O-Members: Austria, Bulgaria, Egypt, Finland, Hungary, Iran, Korea (D.P.R.), Norway, Poland, Serbia, Spain, Sri Lanka Liaisons: CIE, International Commission on Illumination ISO, International Standardization Organization 2013 OIML/TC11/SC3/N5 Contents Explanatory Note .................................................................................................................3 1 Scope ..........................................................................................................................3 2 Terms, Definitions, Units and References ...................................................................3 3 Technical Requirements for the Blackbody Radiators (BBR) ......................................5 4 Metrological requirements and tested characteristics of the BBR ...............................6 5 Calibration and verification conditions .........................................................................7 6 Methods for calibration and verification of the BBR. Form of report. ...........................7 7 Annex A.....................................................................................................................24 Recommended forms of the calibration and verification certificates ..................................24 1 Results of the evaluation of the blackbody radiator cavity geometry...........................24 2 Results of the evaluation of the warm-up time, temperature drift and transition time for the blackbody radiator to pass from a stationary mode to another.................................24 3 Results of the evaluation of the blackbody radiator temperature control uncertainty ..24 4 Results of the correction evaluation during sighting along the blackbody radiator axis.........24 8 Annex B.....................................................................................................................26 References ........................................................................................................................26 2 OIML/TC11/SC3/N5 Explanatory Note This Draft Recommendation deals with the metrological control in manufacturing and operating of blackbody radiator (further named BBR) with controlled temperature in the OIML member-states. The Recommendation can be a basis for calibration, verification, and certification of these instruments. Currently the majority of manufactured radiation thermometers are verified and calibrated using blackbody radiators. With the growth of the number of radiation thermometers, the number of manufactured blackbody radiators has increased. However, no international documents describing calibration and verification of such blackbody radiators are available. Therefore the present Recommendation has utmost importance. 1 Scope The present Recommendation applies to BBR intended for calibration, verification and engineering work in production, maintenance and adjustment of reference and working radiation thermometers, thermal imaging instruments and radiometers in the temperature range from minus 50 °C to plus 2500 °C; it sets up methods and procedures for their calibration and verification. 2 Terms, Definitions, Units and References 2.1 The Recommendation uses the following general terms and definitions: 2.1.1 Calibration (OIML VIM-2010; 2.39 [1]1) operation that, under specified conditions, in a first step, establishes a relation between the quantity values with measurement uncertainties provided by measurement standards and corresponding indications with associated measurement uncertainties and, in a second step, uses this information to establish a relation for obtaining a measurement result from an indication. 2.1.2 Verification (OIML VIM-2010; 2.44 [1]) provision of objective evidence that a given item fulfils specified requirement. Confirmation that performance properties or legal requirements of a measuring system are achieved. 2.1.3 Certification (OIML VIML-2000; 3.2 – 3.4 [2]) operation leading to the issue of a document certifying that the verification of the measuring instrument was carried out with a satisfactory result 2.1.4 Maximum permissible (measurement) error (OIML VIM-2010; 4.26 [1]) extreme value of measurement error, with respect to a known reference quantity value, permitted by specifications or regulations for a given measurement, measuring instrument, or measuring system 1 See bibliography in Annex B 3 OIML/TC11/SC3/N5 2.2 The Recommendation uses the following specific terms and definitions: 2.2.1 Blackbody radiator (BBR) – a source of thermal radiation with an effective emissivity ε close to 1 (as a rule, ε ≥ 0,95 for the radiators with radiating cavity, and ε ≥ 0,9, for the radiators with extended flat surface.). 2.2.2 Permissible uncertainty UpBB – expanded uncertainty at a specified confidence level (p=0,95 or p=0,99) declared in the technical documentation, at which the BBR is considered fit for its intended use. The standard uncertainty upBB = UpBB/k(p) is calculated from either k(p=0.95) = 2 or k(p=0.99)=3. 2.2.3 Temperature instability, Tki – instability of the BBR temperature maintained [or ‘controlled’]in a specified stationery temperature mode. 2.2.4 Temperature drift, Td – temperature drift of the BBR during its operation in a specified stationary temperature mode. 2.2.5 Transition time, tt – the required time for the BBR to pass from one stationary temperature mode to another. 2.2.6 Warm-up time, tw – time elapsed from the moment of turning on the BBR untill it reaches the specified working stationary temperature mode when it is allowed to determine the metrological characteristics of the BBR. 2.2.7 Demountable contact sensor – the contact thermometric sensor, which can be removed from the BBR without its dismantling for the purpose of a separate calibration and/or verification. 2.2.8 Permanent jointed contact sensor – the contact sensor of an internal or external thermometer that cannot be removed from the BBR without the dismantling of the latter. 2.2.9 BBR own thermometer – built-in sensor connected to an internal or external device having an output signal (showing device, or interface, or transmitter transforming a signal of the sensor to a normal electric signal) correlated with temperature of the BBR radiation. 2.2.10 Pyrometer – a thermometer using the optical radiation of a source and indicating temperature values which are calibrated traceable to (inter)national standards2 2.2.11 Pyrometer-comparator – a device using the optical radiation of sources and indicating their temperature differences and no traceability needed (see 2.1.10). 2.2.12 Emissivity – the ratio of the radiance of a substance to the radiance of a blackbody at the same temperature as that of the substance. 2.2.13 Effective emissivity – an apparent emissivity of a blackbody cavity or a surface of a planar blackbody radiator. That should be taking into account of an intrinsic emissivity of surface, a geometrical factor, a temperature distribution, and an ambient thermal radiation. 2.2.14 Spectral selectivity - is the wavelength range over which the BBR specifications are valid 2.3 Units Celsius degrees (ºC) or kelvins (K) are used in this Recommendation as units for temperature or temperature difference (drift, instability etc.). 2 Term ‘pyrometer’ is synonym of term ‘radiation thermometer’ in this document 4 OIML/TC11/SC3/N5 3 Technical Requirements for the Blackbody Radiators (BBR) 3.1 Types of BBRs 3.1.1 The BBRs have a radiating area that is composed of a cavity or a flat surface. 3.1.2 The BBRs can be intended for working with temperature fixed points (Fixed- PointsBlackbody) or for the variable temperature range (Variable-Temperature-Blackbody) and shall have a temperature adjustment system. 3.1.3 The BBRs are subdivided with respect to the method of the radiance temperature measurement into the following types: - BBRs with demountable contact sensors; - BBRs with permanent mounted contact sensors that are either included into the automatic temperature adjustment system, or that are operating off-line; - BBRs with non-contact sensors that are either included into the automatic temperature adjustment system, or that are operating off-line. 3.1.4 The BBRs can be portable or stationary installed. 3.1.5 Requirements to the design of BBRs. The BBR should be equipped with a handle for temperature control, an automatic temperature adjustment system and with a temperature display or/and with a socket or plugs of an output signal (analogue or digital) correlated with the BBR temperature value. 3.1.6 It is desirable (for verification - required), that characteristics of the BBR, such as: - the working temperature range (or fixed values), - the radiating area, - the temperature instability, - the drift, - the effective emissivity, - the spectral selectivity, - the permissible uncertainty of the BBR temperature, - the warm-up time, - the transition time, - the correction factor or value to the readings of the BBR own thermometer – should be given in its technical documentation (hereinafter referred to as TD). All these values are results of measurements and have to be stated either - the associated expanded k 2 uncertainty or mostly - as the interval of minimum and maximum values specifying a rectangular probability distribution – and the one chosen has to be stated, clearly. If any of the characteristics are not resulted in the TD, its actual value together with the associated expanded k 2 uncertainty can be defined at calibration, and it is not checked at verification. 5 OIML/TC11/SC3/N5 The main task during calibration of BBR temperature is the determination of the correction factor or value to the readings of a BBR own thermometer, associated with the appropriate expanded uncertainty and the clear statement which coverage factor or coverage probability p 0,95 or p 0,99 was used for calculation. Notes: 1. Among the above mentioned characteristics, the effective emissivity is the hardest to be separately evaluated (calibrated, verified), because it considerably depends on the BBR design, on the form of the radiating area and on the materials used. It is therefore impossible within the framework of this Recommendation, which does not deal with technological and constructive differences of BBRs, to regulate specific requirements and operations which allows to calculations or measurement of this quantity. The effective emissivity should be taken into account in the correction value applied to the readings of the BBR’s own thermometer and determined by the comparison with a standard BBR traceable to the International Temperature Scale. 2. The spectral selectivity of the BBR is also considerably dependent on the BBR design, the shape of the radiating area and the materials used. This dependence is well-defined for the BBRs with cavity-type radiating element and can be significant for the BBRs with radiating surface with special coating to ensure a high spectral emissivity. In the latter case it is necessary to know the spectral curve of the dependence of emissivity (or reflectance) on the wavelength and this curve shall be given in the specifications. In the case of precision BBRs (devices developed for minimum fluctuation of all their characteristics), if a separate evaluation of the effect of these characteristics is required in the process of their subsequent use, then it is necessary to follow the main principles and methods of their evaluation that are specified in documents [3-5] from bibliography (subsection 8 Annex B) in this Recommendation. 3.1.7 Generally the result of the calibration of BBR temperature has to be stated for a coverage probability p 0,95 using a coverage factor k p 0,95 2 . For some applications to use the BBR the regulations define a higher coverage probability p 0,99 with a corresponding coverage factor k p 0,99 3 . Therefore, the appropriate expanded uncertainty is determined from the standard uncertainty by multiplication with the related coverage factor k p . 3.1.8 Depending on the intended use of a BBR as standard for temperature measurements the TD will also state the permissible uncertainty of the BBR temperature. 4 Metrological requirements and tested characteristics of the BBR In the process of testing for calibration and verification the following metrological characteristics of the BBR shall be determined: 4.1 4.2 4.3 Temperature range (or fixed values). Size of the radiating area of the BBR. Warm-up time required for the BBR to reach the specified stationary mode at the lower and upper levels of the working temperature range of the BBR. 6 OIML/TC11/SC3/N5 4.4 4.5 4.6 4.7 4.8 Transition time required for the BBR to pass from one stationary mode to another. Temperature drift during the operation in specified stationary modes. Instability of temperature control at specified levels. Corrections to the readings of the BBR own thermometer (or output signal). Expanded uncertainty of the BBR temperature at a specified confidence level. 5 Calibration and verification conditions 5.1 Process of verification and calibration The process of verification and calibration should be carried out in a stable indoor environment within the measurement conditions (a temperature range and the relative humidity range) accepted in each country, unless the other conditions are specified by customer; the conditions should be written down in the calibration certificate. A BBR should not be affected by shocks, vibrations, external electromagnetic fields, extraneous radiation sources influencing the readings of measuring instruments. 5.2 Accreditation The calibration shall be carried out by an accredited calibration laboratory. 6 Methods for calibration and verification of the BBR. Form of report. 6.1 Methods 6.1.1 Methods for the calibration and verification The actual temperature T X c Y X Y X measured by a pyrometer is calculated from the reading Y X and a correction value Y X . The factor c 1 with a value defined as unity converts the reading Y X from arbitrary units Y X in the unit T X wanted for the temperature. Thus, the unit of the factor c T X Y X is just the ratio of the units of the temperature and the reading. The symbols “ T X ” and “ Y X ” stand for actual temperatures and for readings of pyrometers specified for the device by index “ X ”. The indices “ b ” and “ sb ” replacing index “ X ” refer to the BBR to be calibrated and a standard BBR used as reference, while the indices “ p ” and “ sp ” refer to an external pyrometer-comparator and a standard pyrometer used for reference. The calibration of a BBR assigns a value to the correction value Yb to the BBR own thermometer with associated combined standard uncertainty u Yb . Finally, the standard uncertainty is converted in an expanded uncertainty U Yb k uYb by multiplication with the coverage factor k with values either k 2 or k 3 depending on the fraction p of probability defined for the application. There are two fundamental methods to determine this correction value: 7 OIML/TC11/SC3/N5 - Direct measurement with an external standard pyrometer as reference - Comparison with a standard BBR as reference and using an external pyrometercomparator. 6.1.2 Direct measurement: The BBR temperature Tb c Yb Y b indicated by readings Yb of the BBR-own-thermometer is measured by an external standard pyrometer T sp Tb c Y sp Tb Y sp , which is certified with the correction value Y sp . Then the correction value Yb of the BBR-thermometer reads: Yb Tsp Tb c Yb where with T T c Y T Y sp b sp b (1) sp Yb , Yb - reading, correction value for BBR-own-thermometer; Tb , T sp Tb - BBR temperature, indicated by standard pyrometer; c1 - factor to match the units; Y sp Tb , Y sp - reading, correction value of standard pyrometer. indicated by its own T c Y T Y . 6.1.3 Substitution method: The BBR temperature Tb c Yb Y b thermometer is measured by the pyrometer-comparator T p b p b p The temperature T sb c Y sb Y sb of a standard BBR indicated by its own thermometer is also measured by the pyrometer-comparator T p Tsb c Y p T sb Y p . A combination of the two equations cancels out the Y p correction value and the correction value Yb of the BBR-thermometer reads: Tsb c Y sb Y sb Yb Tsb c Yb Y p Tb with Y p Tb Y p Tb Y p Tsb where Yb , Yb (2) - reading, correction value of BBR-own-thermometer; Tsp - temperature of standard BBR; c1 - factor to match the units; Ysb , Ysb Y p Tb - reading, correction value of standard BBR-own-thermometer; - difference of indications of external pyrometer–comparator; Y p Tb , Y p Tsb - indication of pyrometer-comparator for BBR, standard BBR. 6.2 Operations and means for calibration and verification of BBRs 6.2.1 The operations and measuring instruments to be used for calibration and verification are listed in Table 1. 8 OIML/TC11/SC3/N5 Table 1 Operations and Measuring Instruments No. Operation Item in the text Verification instruments and their characteristics 1 2 3 4 1 External examination 2 Testing Evaluation of the BBR 3 radiating geometry Evaluation of the warm-up time, temperature drift and transition time of BBR from one stationary mode to another 4 Evaluation of the BBR 5 temperature keeping instability Evaluation of the correction to the 6 readings of the sensor of the BBR to be calibrated (verified) Calculation of the 7 expanded uncertainty of the BBR temperature Obligation of a verification procedure Initial Periodical 5 6 - Linear (length) measuring instrument with the 6.5 scale factor, provided measurement of size of the radiating area with uncertainty less than 5%. - chronometer; - pyrometer-comparator with a temperature measuring range corresponding to that of the BBR and with the combined uncertainty up1 from contributions of instability and limited resolution of the thermometer has to be valid up1 upBB 3 (less than1/3 of the permissible standard uncertainty of the tested BBR), 6.6 - the field of view has to be smaller than the output apertures of both, the standard BBR and the tested BBRs; - a device to measure the BBR thermometer output signal (if necessary) the uncertainty contribution up2 of limited resolution of the thermometer has to be valid up2 upBB 3 (less than1/3 of the permissible standard uncertainty of the tested BBR) Yes Yes Yes Yes Yes No Yes Yes 6.7 Equipment from item 4 in this Table Yes Yes - standard BBR with variable temperature in a corresponding measurement temperature range or with temperatures of the fixed points (depends on tested BBR), the combined uncertainty of all characteristics specified in lines 3 – 7 of this Table shall be valid up3 upBB 3 (less than 1/3 of the permissible standard uncertainty of the tested BBR), - emissivity has to be 0,99 , - pyrometer-comparator with a corresponding temperature measurement range; its combined uncertainty up4 from contributions of instability and Yes Yes Yes Yes 6.3 6.4 6.8 temperature resolution has to be valid u u p4 pBB 3 (less than 1/3 of the permissible standard uncertainty of the tested BBR); its field of view has to be smaller than the output apertures of both, the standard BBR and the tested BBR; the size-ofsource effect of the pyrometer-comparator should be sufficiently small so that either the effect of different aperture sizes can either be neglected - a device to measure or decode an output signal (if necessary) with resolution Y upBB 3 (in a temperature equivalent) less than 1/3 of the permissible standard uncertainty of the tested BBR) 6.9 Calculation of the uncertainty should carried out according to [3-5] 9 OIML/TC11/SC3/N5 6.2.2 6.2.3 6.2.4 6.2.5 Measuring instruments specified in Table 1 should be calibrated traceable to national standards and provided with the corresponding legal documents about their verification or calibration. Standard BBRs shall be calibrated traceable to national standards and the International Temperature Scale (the ITS-90). Measuring instruments are prepared for operation in accordance with their valid documentation. Demountable BBR own thermometers shall have valid calibration or verification certificates. 6.3 External examination During external examination the following points shall be checked: 6.3.1 6.3.2 6.3.3 6.3.4 6.4 6.4.1 6.4.2 6.5 6.5.1 6.5.2 6.5.3 6.5.4 3 Correspondence of the completeness of BBR and associated equipment to the requirements of its valid documentation; Correspondence of the BBR to the safety requirements specified in the TD; Absence of external damage of the calibrated (verified) BBR and associated equipment that may adversely affect its metrological performance and main functions. BBR that does not comply with the requirements of item 6.3.3 is not subject to calibration or verification. Testing for serviceability The BBR is connected to a power supply and its serviceability is tested in compliance with the valid documentation. A BBR in which a defect was found during the testing (for example: inability to look some of the display readout, apparent instability, etc.) is rejected for a further calibration (verification). Evaluation of the BBR radiating geometry The outlet diameter of the BBR radiating size and the cavity depth (in case of a cavity BBR)3 shall be measured once. The difference between the measured values and the values specified in the TD shall not exceed the limits ± 5 % with reference to the declared values. If relative differences calculated according to 6.5.1 exceed the limits ± 5 %, the calibration (verification) certificate shall specify the actual dimensions and the recommendation to the customer to amend the TD and to update (verify) the emissivity value. Evaluation of the warm-up time, temperature drift and transition time for the BBR to pass from one stationary mode to another The warm-up time of the BBR is interrelated with its temperature drift. Therefore these parameters shall be determined simultaneously. It must be kept in mind that it is impossible to insert any instruments into the blackbody cavity which could damage the coating 10 OIML/TC11/SC3/N5 6.5.5 6.5.6 6.5.7 6.5.8 6.5.9 6.5.10 6.5.11 6.5.12 6.5.13 6.5.14 The warm-up time of the BBR at the lower temperature limit is determined by setting the value corresponding to the lower temperature limit on the temperature control device of the BBR control unit. The BBR is turned on and entered in the specified stationary mode in compliance with TD. When the BBR reaches the stationary mode after the time tW , its radiance temperature is determined every (10 to 15) seconds during fifteen minutes by the readings of the radiance temperature measuring device having a sufficient resolution (see Table. 1). Simultaneously the indicated reading of the BBR own thermometer is also recorded. The average temperature values are determined in accordance with the measurement results during the first five minutes, second five minutes and third five minutes. The difference between the average temperature values shall not exceed the temperature drift value specified in the TD. The calculation is carried out for the radiance temperature values and for the indicated temperatures of the BBR own thermometer. If in the process of verification the maximum difference between the average temperature values of the BBR is bigger than the temperature drift value, the BBR is rejected as defective. In the process of calibration the warm-up time of the BBR is determined with more precision through additional measurements. With this purpose, the operations of items 6.6.3 – 6.6.4 are repeated until the measured temperature drift becomes equal to the value specified in the TD. If the measured temperature drift value agrees with the value specified in the TD, the BBR warm-up time being less than 2tW , a new value of the BBR warm-up time is indicated in the TD. If the measured drift value exceeds the value specified in the TD at the BBR warm-up time being equal to 2 tW , the BBR is rejected as defective. The BBR transition time from one stationary mode to another (tt) is determined by setting, on the temperature control device of the BBR unit, the value corresponding to the next temperature mode of the BBR, and, after the time specified in the TD as the transition time from one stationary mode to another expires, the operations mentioned in items 6.6.2 – 6.6.6 are repeated. The transition time of the BBR to the stationary mode at the upper temperature limit is determined after disconnecting the BBR from the power supply and cooling down to the room temperature. Then the value corresponding to the upper temperature limit is set on the temperature control device of the BBR unit. The BBR is turned on again and, when the transition time of the BBR to the stationary mode at the upper temperature limit (tW2) expires, the operations of items 6.6.2 – 6.6.6 are repeated, and the values tW2 and 2tW2 substitute the values tW and 2tW . 6.5.15 Evaluation of the temperature control instability of the BBR 6.5.16 The value corresponding to the lower temperature limit is set on the temperature handle of the BBR control unit and then the BBR is adjusted to this temperature value according to its operation manual. 6.5.17 When the BBR reaches the stationary mode, the radiance temperature value Ti is measured every (10 to 15) seconds during (15 to 20) minutes by the readings of the temperature measuring device with a resolution no worse than 0,1 °C ( T 0 ,1 C ). 11 OIML/TC11/SC3/N5 The average temperature value Ta during the period 15 t min 20 , the empirical standard deviation Ti of the current temperature value Ti and standard uncertainty type B u(T) are calculated using the formulae: n n Ti Ta = i=1 uTa ; n where Ta Ti n Ti uTa - Ti with Ti T i Ta 2 i=1 , (3) n1 n average temperature value; the i-th temperature measurement result; number of independent readings; empirical standard deviation of readings; standard uncertainty type B equals the empirical standard deviation of the mean of readings. 6.5.18 Similar procedure is carried out for values of temperature which are indicated by the BBR own thermometer. 6.5.19 For calculations of an expanded uncertainty largest of the average temperature values obtained in item 6.7.1 - 6.7.3 is used. 6.5.20 The expanded uncertainty of the temperature keeping instability depends on the confidence probability p, and is equal to the standard deviation multiplied by the coverage factor k (for example, k = 2 when p = 0.95). 6.5.21 If the standard deviation exceeds half of the temperature keeping instability specified in the TD, the BBR shall be rejected as defective. 6.5.22 Operations in items 6.7.1 – 6.7.4 are repeated for the BBR with regulated temperature at the mid and maximum temperature values in the working range. 6.6 Evaluation of the correction to apply to the readings of the BBR own thermometer 6.6.1 The correction to the readings of the BBR own thermometer is determined by comparing it with a standard BBR by a pyrometer-comparator, or by direct measurement of its radiance temperature with a standard radiation thermometer. A standard BBR and a standard radiation thermometer shall be calibrated with radiance temperature traceable to national standards and the International Temperature Scale (the ITS-90) by the calibration laboratory4. The correction to the readings of the BBR own thermometer to be calibrated (verified) in the low temperature range 50 T C 300 ) is determined by comparing it with a standard BBR by a pyrometer-comparator working in a full radiation or in a spectral interval 8 μm 14 , or by direct measurement of its radiance temperature with a standard radiation thermometer working in such a spectral interval.5 The BBR to be calibrated (verified) is placed on a test bench connected to the power supply and adjusted to the specified lower stationary temperature mode. Comparisons are performed by means of a comparator using the equal signals method 6. With this method, the comparator is placed in such a way that its optical 6.6.2 6.6.3 6.6.4 4 calibration laboratories must be accredited according to the state legislation The numerical ranges of temperature and wavelength in clauses 6.8.2, 6.8.12 and 6.8.14 are given as an informative example. 6 Method, in which the comparator signals received from the test and standard BBR are equal 5 12 OIML/TC11/SC3/N5 6.6.5 6.6.6 6.6.7 6.6.8 6.6.9 6.6.10 6.6.11 6.6.12 6.6.13 axis lies in the axis of the standard BBR and passes through the centre of its radiating aperture. Then the comparator is directed at the tested BBR. The temperature of the tested BBR is selected in such a way that the comparator signal is equal to the signal of the standard BBR. The standard BBR temperature and the readings of the calibrated (verified) BBR own thermometer are recorded. The measurements shall be repeated 10 times. The average-values of temperatures of the standard BBR and of the BBR own thermometer to be calibrated (verified) are calculated. The correction to the readings of the BBR own thermometer to be calibrated (verified) is determined as a difference of the average temperature values of the standard and calibrated (verified) BBRs. In the calibration of the variable temperature BBR, the standard BBR and the BBR to be calibrated (verified) are entered in the next stationary temperature mode and operations are carried out according to items 6.8.3 – 6.8.6. These operations are repeated at all temperature modes of the BBR specified in theTD (or required by customer). In the verification of the variable temperature BBR the number of specified temperature modes may be reduced to three (including minimal and maximal temperatures). The correction values obtained during the calibration are specified in the certificate of calibration, if they exceed half the value of the expanded (permissible) uncertainty of the BBR temperature. If the correction value obtained during the verification differs from the correction value given in the TD by more than half the value of the permissible expanded uncertainty of the BBR temperature at one temperature mode, the correction shall be revised at all specified temperature modes of the BBR. The obtained new correction values are to be specified in the TD of the BBR in the same way as during the calibration. If the corrections are determined by means of a standard pyrometer, they are calculated as a difference of the average readings of the standard pyrometer and the BBR own thermometer to be calibrated (verified). The correction to the readings of the BBR own thermometer to be calibrated (verified) in the mid temperature range ( 300 T C 1000 ) is determined by comparing it with a standard BBR by means of pyrometer-comparators with the partial spectral range, or standard pyrometers with two or three partial spectral ranges, e.g., with the ranges 2 μm 5 and 8 μm 14 . The operations described in items 6.8.3 – 6.8.11 are carried out for each spectral intervals7. If the correction obtained with different spectral intervals (within one temperature mode) do not agree with each other within a half value of the permissible expanded uncertainty U pBB , they shall be averaged over all spectral intervals and the temperature uncertainty component, usi (one standard uncertainty), shall be taken to be the maximum deviation of the correction from its average value.. Otherwise yields usi U pBB . 6.6.14 The corrections to the readings of the BBR own thermometer to be calibrated (verified) at the temperature T 800 C can be determined by comparing it with a standard BBR by means of a spectrocomparator with a narrow spectral band 7 See footnote to Clause 6.8.2 13 OIML/TC11/SC3/N5 depending on the purpose of the BBR to be calibrated (verified). During the verification it is allowed to use the pyrometer-comparator, or the standard radiation thermometer with the partial spectral interval instead of the spectrocomparator8. 6.6.15 When the BBR to be calibrated (verified) is intended for calibration of radiation thermometers of a special type, it is possible to calibrate (verify) the BBR by means of the same calibrated radiation thermometer type, if it meets the requirements of the Table 1, irrespective of the temperature range. 6.6.16 If the BBR to be calibrated (verified) is intended to be applicable for calibration of radiation thermometers with wide-angle objectives, the dependence of correction on viewing angle9 shall be determined. With this purpose, the operations from items 6.8.2 - 6.8.15 are carried out for each viewing angle depending on the temperature mode. The average correction value is determined for all viewing angles. If the correction obtained is higher than half the value of the permissible expanded uncertainty, the maximum deviation of the corrections from the average value by all viewing angles is taken into account as the uncertainty component uva . Otherwise yields uva U pBB 6.6.17 The dependence of correction on the radiating surface non-uniformity of the radiance temperature Ti shall be determined only for BBRs with the extended radiating surface. With this purpose, the average correction value is determined as the maximum difference between radiance temperature values Ti from 5 points of the surface (in the center and on periphery) and their average value. It is taken 1 5 into account as the uncertainty component urs max Ti T j . 1 i 5 5 j 1 These measurements are made according to items 6.8.2 – 6.8.13. In this case the dependence of correction on the viewing angle is not determined. Such correction, as a rule in practice, is most powerful at a temperature T 300 C . 6.7 Evaluation of the uncertainty of the BBR temperature The basic components of the uncertainty budget in calibration (verification) of a BBR temperature are listed below and grouped for the two methods. The uncertainty contribution from the reading of the BBR own thermometer is independent of the method of calibration (verification) and reads: Uncertainty contribution due to reading of the BBR-own-thermometer The uncertainty contribution from the readings Yb of the BBR-own-thermometer is determined as the maximum from either resolution Yb or empirical standard deviation Yb of a number n of independent readings. u Yb max Yb where 3 , Yb n (4) - standard uncertainty of BBR-own-thermometer reading; u Yb Yb - resolution of BBR-own-thermometer; Yb - empirical standard deviation of reading; n - number of independent readings. 8 9 See footnote to Clause 6.8.2 viewing angle is the angle between the line of observation and the normal line to the radiating area 14 OIML/TC11/SC3/N5 The readings are modified by a possible difference Ta , b Ta , b Ta , b, 0 between the actual value Ta , b of the ambient temperature and the value Ta , b, 0 stated in the certificate. The small correction b Ta , b 1 is the product of ambient temperature difference Ta , b and a relative temperature coefficient b . In the TD of the BBR the value b is given with associated expanded uncertainty U b and coverage factor k p . Thus, the standard uncertainty u b U b k p can be calculated. The corrected readings Y b 1 b Ta , b correction factor. are the product with a u b c u Yb 1 b Ta , b c u 2 Yb Yb2 u 2 Yb Ta2, b b2 u 2 b u 2 Ta , b where (5) ub - standard uncertainty of BBR temperature by own thermometer; c 1 - factor to match the units; Yb - reading of BBR-own-thermometer; b - relative temperature coefficient; Ta , b - ambient temperature difference. Provided the ambient temperature difference is less than the associated uncertainty Ta , b u Ta , b , than the higher order term in brackets has to be used. 6.7.1 Uncertainty contribution of the standard measuring instruments a) Calibration (verification) option by comparison with a standard BBR as reference using a pyrometer-comparator and described by the following model of evaluation: Yb Tsb c Yb Y p Tb 1 2 u Tsb u 2 Yb u 2 Y p Tb 2 c uYb where Yb , Yb Tsb c1 u Yb Y p Tb - (6) reading, correction value of BBR-own-thermometer; temperature of standard BBR; factor to match the units; standard uncertainty associated with Yb ; difference of indications from external pyrometer-comparator. Uncertainty of temperature relating to the standard BBR as reference usb = u(Tsb) – includes the uncertainty of its calibration; its instability; uncertainty due to its positioning against the optical axis of the comparator; uncertainty of measuring instruments used to maintain the conditions of its operation; uncertainty dues to the effect of the ambient conditions. It is evaluated using the following models: The temperature Tsb of the standard BBR is evaluated from the reading Ysb and 15 OIML/TC11/SC3/N5 the correction value Ysb modified for small effects of ambient temperature 1 sb Ta , sb and ageing 1 sb t sb , respectively. sb Ta , sb 1 with sb t sb 1 (7) sb2 u 2 t sb Tsb c Ysb 1 sb Ta , sb Y sb 1 sb t sb u sb uTsb c where Y Y T u 2 Ysb Y sb2 t sb2 u 2 sb u2 2 sb sb Tsb Ysb с≡1 Ysb sb Ta,sb sb tsb - 2 a , sb u2 sb 2 sb u u T 2 2 sb a , sb temperature of standard BBR; reading of standard BBR; factor to match the units; correction value standard BBR; relative temperature coefficient standard BBR; ambient temperature difference; relative ageing coefficient standard BBR; time since calibration, standard uncertainty. u Ysb U Ysb k p is the uncertainty associated with the correction value of the standard BBR from calibration, certified as expanded uncertainty U Ysb with coverage factor k p . For long term stability the uncertainty contribution Ysb2 tsb2 u 2 sb sb2 u 2 tsb is evaluated from the relative ageing coefficient sb with associated uncertainty u sb U sb k p and coverage factor k p given in the TD. The period of time tsb since last calibration is determined from the certificate and usually the associated uncertainty utsb is negligible. u Ysb max Ysb 3 , Ysb n is the uncertainty from the readings of standard BBR- own-thermometer. It is determined as the maximum from either resolution Ysb or the empirical standard deviation Ysb of a number n of independent readings. The uncertainty contribution from ambient conditions e.g. the ambient temperature Ysb2 Ta2, sb u 2 sb sb2 u 2 sb u 2 Ta , sb is determined from the difference Ta , sb Ta , sb Ta , sb0 between the actual ambient temperature Ta , sb near the standard BBR and the value Ta , sb0 defined for its operation. Provided the temperature difference Ta , sb is less than the uncertainty uTa , sb , then the higher order term (in brackets) have to be regarded. The relative temperature coefficient sb with associated uncertainty u sb U sb k and coverage factor k is given in the TD. The nominal effective geometric extend Gsb sb Asb Щsb of the standard BBR has to be stated in the certificate of the standard BBR with individual values of emittance sb of the radiating area Asb , the size and location of this area and the solid angle Щsb to be collected. During the measurement with a radiation thermometer a geometric extend Gp, sb p, sb Ap, sb Щp, sb is used and the reading has to be corrected by a factor 1 p Gp, sb . It is combined from the imperfect match Gp, sb Gp, sb Gsb between the effective geometric extends and a relative weighting coefficient p . The value p and associated standard uncertainty u p U p k p are found in the TD or the latter is determined from the expanded uncertainty U p and the related coverage factor k p . 16 OIML/TC11/SC3/N5 usb u Tsb the uncertainty associated with the temperature of the standard BBR and modeled in equ. (7) shall be specified in the calibration certificate. It (uncertainty) shall be expressed in terms of expanded uncertainty with the confidence level of p 0,95 . If it is expressed as standard uncertainty, its value should be adjusted to correspond to the expanded value by multiplying by the coverage factor k p 0,95 2 determined by this probability10. Uncertainty due to pyrometer-comparator. For the pyrometer-comparator the model of evaluation with the sources of uncertainty shall be known. It contains the readings Yp Tb , Yp Tsb for the temperatures Tb , Tsb of the BBR and the standard BBR, respectively. In case of fine adjustment their difference Yp Tb Yp Tsb 1 will be negligible and the associated uncertainty can be evaluated in a simplified formula. The readings have to be corrected for fluctuations in ambient temperature Ta , p Ta , p, b Ta , p, sb , possible effects of ageing t p t p, b t p, sb (usually, negligible), and differences in the effective geometric extend Gp, b, sb Gp, b Gp, sb . All these corrections are small and determined as products with relative coefficients for temperature p , ageing p and geometry p . Yp Tb Yp Tb 1 p Ta , p, b p t p, b p Gp, b Yp Tsb 1 p Ta , p, sb p t p, sb p Gp, sb uY T u Y T u Y T u 2 p b 2 p 2 b p sb p 2 with Y T Y T p b 2 p sb 1 Ta,p u p t p u p Gp, b, sb (8) where Yp Tb Tb , Tsb - difference indications of pyrometer-comparator for BBR temperature; - BBR temperature, standard BBR temperature; Yp Tb - indication pyrometer-comparator for BBR temperature; Yp Tsb - indication pyrometer-comparator for standard BBR temperature; p , p , p - relative coefficients for ambient temperature, ageing, geometry; Ta,p , tp , Gp, b, sb - fluctuation ambient temperature, time since calibration, difference in geometry; - uncertainty component caused by the effect of ambient temperature fluctuations during the calibration. The tested BBR and the standard BBR are measured sequentially by the radiation thermometer. In between the radiation thermometer ambient temperature fluctuates from Ta , p, b to Ta , p, sb . u1 u p Ta,p is the related uncertainty and a part of equ. (7) with differences Ta , p,b Ta , p,b Ta, p0 and Ta , p,sb Ta , sp,b Ta , p0 . Provided, the adjustment of the BBR is such, that differences between the comparator readings Yp Tb Yp Tsb 1 are negligibly, then the simplification u1 Yp2 Tb u 2 p Ta, p, b Yp2 Tsb u 2 p Ta , p, sb 10 Sources of uncertainties that were not considered in the calibration certificate shall be estimated separately by the user of the reference standard BBR. Especially it concerns of uncertainty due to difference in dimensions of the compared BBRs 17 OIML/TC11/SC3/N5 u Y T is valid and with the higher order term in brackets yields: u u T T (9) u T T u1 Yp Tb u p Ta , p, b Ta , p, sb 1 p 2 2 b p where Yp Tb Ta , p, b , Ta , p, sb 2 p 2 2 p a , p, sb a , p, b - indication of pyrometer-comparator for BBR temperature; - relative temperature coefficient pyrometer; - ambient temperature for pyrometer when measuring the tested BBR, the standard BBR; - standard uncertainty of the ambient temperature difference. p u Ta , p, b Ta , p, sb a , p, sb a , p, b The values p of the relative temperature coefficient with associated uncertainty u p U p k p and coverage factor k p are given in the TD of the pyrometercomparator. The ambient temperatures Ta , p, b and Ta , p, sb are measured with a thermometer not necessarily certified. The uncertainty of the ambient temperature readings is found from either the resolution Ta, p or the empirical standard deviation Ta, p . The latter is not affected by the type of BBR and one gets: u Ta , p, b Ta , p, sb 2 max Ta, p 3 , Ta, p n . u2 u p t p – uncertainty component due to instability of the measurement transducer efficiency. The stability of the pyrometer between the sequential measurements is also part of equ. (7) and with comparator readings Yp Tb Yp Tsb 1 one gets: 2 u2 u p t p Yp Tb u 2 p t p, b tp, sb p2 u 2 tp, b t p, sb where Yp Tb p t p, b , t p, sb BBR, u t p, b t p, sb (10) - indication pyrometer-comparator for BBR temperature; - relative ageing coefficient of pyrometer; - time since calibration of pyrometer when measuring the tested the standard BBR; - standard uncertainty of difference of ambient times. u3 u Yp Tb Yp Tsb – uncertainty component due to resolution Yp of the instrument measuring the comparator output signal and the empirical standard deviations Yp Tb and Yp Tsb for a number of n readings for the BBR Yp Tb and for the standard BBR Yp Tsb , respectively: n maxY u3 u Yp Tb Yp Tsb max Yp2 3 , 2 Yp Tb where Yp Tb , Yp Tsb Tb , Tsb temperature; Yp Yp Tb , Yp Tsb 2 p n 3 , 2 Yp Tsb (11) - indications of pyrometer-comparator for BBR temperature; - the tested BBR temperature, the standard BBR - resolution of indication of pyrometer-comparator; - empirical standard deviation of pyrometer indication. – uncertainty component due to the difference in the dimensions of the reference radiation source Asb and the radiation source under u4 u p Gp, b, sb 18 OIML/TC11/SC3/N5 calibration Ab (size-of-source effect) 11. Their emittance sb , b and solid angles Ω sb ,Ω b forming the differences in the effective geometric extends measured by the pyrometer-comparator have to be regarded, additionally u p Gp, b, sb Yp Tb 1 p Gp, b Yp Tsb 1 p Gp, sb . If the adjustment of the tested BBR is such, that Yp Tb Yp Tsb 1 the comparator readings are (nearly) equal, then simplification Y p Tb Y p Tb 1 p b Ab Ω b sb Asb Ω sb is valid and one gets: u4 u p G p, b, sb Y p Tb u 2 Y p Tb u 2 p 2p u 2 b Ab Ω b u 2 sb Asb Ω sb (12) where Yp Tb - indication of pyrometer-comparator for BBR temperature; p relative geometry coefficient of pyrometer; Tb , Tsb the tested BBR temperature, the standard BBR temperature; Gp, b, sb effective geometric extents, the tested BBR minus the standard BBR; Yp Tb , Yp Tsb - the empirical standard deviation of the pyrometer indication. The combined uncertainty ust for the temperature is compiled from the contributions of equ. (4) to equ. (11) of the standard measuring instruments and shall be calculated by the formula: ust = usb2 ub2 c 2 u12 + u22 + u32 + u42 (13) b) Calibration (verification) option by direct measurement of the tested BBR temperature using the reference pyrometer is described by the following model: Yb Ysp Tb Yb Ysp uYb u 2 Ysp Tb u 2 Yb u 2 Ysp where Yb Ysp Tb Ysp Yb (14) reading of BBR-own-thermometer; correction value for the tested BBR; reading of the standard pyrometer; correction value of the standard pyrometer. 11 Usually, in the process of calibration in the conditions of laboratory the fluctuations of the ambient temperature are negligibly small; the requirements for instability and resolution of the instrument measuring the output signal of the comparator given in Table 1 also allow to ignore these uncertainty components. What is important to be taken into account is the uncertainty due to the difference in the dimensions of the compared BBRs. According to the estimates of [4] and [5] the normal standard value of this uncertainty in the case of different dimensions of the compared BBRs varies between 0.1 and 0.2 %. If it is necessary to calibrate BBRs with an uncertainty below 0.5 %, one shall be guided by the rules and estimates given in [3-5]. 19 OIML/TC11/SC3/N5 Uncertainty of temperature related with the reference pyrometer (the standard radiation thermometer). usp – is the uncertainty of a BBR temperature Tb indicated by a standard pyrometer used as reference. It includes: the uncertainty of its calibration; its instability; uncertainty due to its positioning against the optical axis of the BBR; uncertainty due to the effect of ambient conditions. The temperature Tb of a BBR is evaluated from the reading Ysp(Tb) of the standard pyrometer corrected by a factor 1 sp Ta , sp sp G sp, b for effects of ambient temperature Ta , sp and effective geometric extend Gsp, b and from the correction value Ysp modified by a correction factor 1 sp tsp for ageing during the period of time tsp since calibration. The small corrections are products of variations in ambient temperature Ta , sp , differences in effective geometric extents Gsp, b and due to ageing time Ta , sp with the related relative coefficients sp , sp , sp . Tb c Y sp Tb 1 sp Ta , sp sp G sp, b Y sp 1 sp t sp u u T u u G u 2 Y sp Y sp2 t sp2 u 2 sp sp2 u 2 t sp u sp uTb c u 2 Y sp Tb Ta2, sp u 2 2 Y sp Tb G 2 u 2 sp, b where Tb Ysp Tb Ysp sp , sp , sp - sp 2 sp sp 2 sp 2 (15) 2 sp 2 a , sp 2 sp sp, b indication of the tested BBR temperature; reading of the standard pyrometer; correction value of the standard pyrometer; relative coefficients of the standard pyrometer for temperature, ageing and geometry. u Y sp U Y sp k p is the uncertainty of the pyrometer correction value assigned by the calibration of the pyrometer, and certified as expanded uncertainty U Ysp with coverage factor k p . For long term stability the uncertainty contribution 2 2 2 2 2 Y sp t sp u sp sp u t sp is evaluated from the relative ageing coefficient k p and a coverage factor k p sp with the associated uncertainty u sp U sp given in the TD for the standard pyrometer. The period of time tsp since last calibration is determined from the certificate and usually the uncertainty utsp is negligible. u Y sp Tb max Y sp 3 , Y sp n from the readings is determined as the maximum from either resolution Y sp Tb or the empirical standard deviation Y sp Tb of a number n of independent readings. The uncertainty contribution Y sp2 Tb Ta2, sp u 2 sp sp2 u 2 sp u 2 Ta , sp from ambient conditions e.g. the variation of the ambient temperature is determined from the difference Ta , sp Ta , sp Ta , sp0 between the actual ambient 20 OIML/TC11/SC3/N5 temperature Ta , sp near the standard pyrometer and the value Ta , sp0 defined for its operation. Provided the temperature difference Ta , sp is negligible, then the higher order term (in brackets) have to be regarded. The relative temperature coefficient sp with associated uncertainty u sp U sp k p and coverage factor k p is given in the TD of the standard pyrometer. The effective geometric extend G b, 0 b,0 Ab,0 Ω b,0 for the tested BBR is stated in the certificate with nominal values of emittance b, 0 of the radiating area Ab, 0 , the size and location of this area and the collected solid angle Ω b, 0 . It is compared with the measured geometric extend G sp, b sp, b Asp, b Ω sp, b of the standard pyrometer with individual values of emittance sp, b of the radiating area Asp, b , the size and location of this area and the collected solid angle Ω sp, b . The correction factor 1 sp G sp, b is product of the relative coefficient sp and the difference G sp, b G sp, 0 G b, 0 . 2 2 2 2 2 The uncertainty contribution G sp, during b u sp sp u sp u G sp, b the calibration of the tested BBR due to an imperfect match between these effective geometric extends has to be accounted for. The uncertainty usp of the standard pyrometer shall be specified in its calibration certificate. In the same way as mentioned in the previous section, this uncertainty shall be expressed in terms of the expanded uncertainty with the probability level of p = 0,95. If it is expressed as standard uncertainty (k = 1) , its value shall be adjusted to correspond to the expanded value with multiplying by the coverage factor k p 0 ,95 2 determined by this probability. The uncertainty due to the difference between (dimensions) effective geometric extent of the BBR used in calibration of the reference pyrometer and the dimensions of the BBR to be calibrated is explained in Note to Section 6.9.1 and taken into account in equ. (15). In this case the uncertainty of the standard measuring instruments u st from equ. (13) in this case shall be equal to usp. The standard uncertainty estimated repeated readings are obtained in accordance of 6.7.2. 21 OIML/TC11/SC3/N5 Uncertainty in the calibration of the tested BBR 6.7.2 The tested BBR temperature is determined from the average Yb of repeated readings of the tested BBR own thermometer obtained in accordance of 6.7.2 and the correction value Yb evaluated either from equ. (7) or equ. (15) and stated in the certificate as expanded uncertainty U Yb k p u Y b with specified coverage factor k p . The related standard uncertainties are determined from equ. (4) and either equ. (7) or equ. (15). Tb c Y b Y b where Tb c 1 Yb Yb - (16) the tested BBR temperature; factor to match the units; reading of the tested BBR-own-thermometer; correction value for the tested BBR-own-thermometer. 6.7.3 The maximum standard uncertainty of the tested BBR temperature for verification is calculated by the following formula using the correction value form equ. (7) or equ. (15) and the limiting intervals (rectangular probability distribution) in the TD estimated by Type B – u (Tb) – is calculated by the formula: u( Tb ) = u st2 + ( uci2 + u si2 + uva2 + u rs2 ) / 3 where (17) Tb - the tested BBR temperature; c 1 ust - factor to match the units; standard uncertainty form equ. (7) or equ. (15); uci - usi - uva - urs - standard interval from instability of the tested BBR-ownthermometer given as limits in the TD; standard interval from calibration (verification) for spectral intervals, limits of uncertainty according to 6.7.15; standard interval from calibration (verification) for viewing angle limits of uncertainty according to 6.7.16; standard interval from calibration (verification) for inhomogen radiating surface, limits of uncertainty according to 6.7.16. All components of the budget have to be expressed as standard uncertainties determined from the limits of permissible values taking into account the recommended coverage factor. 6.7.4 The combined standard uncertainty of a temperature measurement of a BBR to be calibrated (verified) – uΣ ( T ) – is calculated by the formula: uΣ ( T ) = c 2 u 2 Yb + u 2 ( Tb ) where uΣ ( T ) - the measured BBR temperature; c1 - factor to match the units; u Yb - standard uncertainty form equ. (4); u( Tb ) - standard uncertainty from equ. (17). (18) 22 OIML/TC11/SC3/N5 6.7.5 The expanded uncertainty of the temperature value of the BBR to be calibrated (verified) is determined by the coverage factor k p depending on the confidence probability p ( k p 0 ,95 2 , k p 0 ,99 3 ) and is calculated by the formula: U = k p u Σ T (19) 6.7.6 The obtained expanded uncertainty of the temperature value of the BBR to be verified shall not exceed the corresponding uncertainty specified in the TD. 6.7.7 The calibration interval lasts usually for one or two years in the case of BBRs with permanent jointed thermometers or in the case of BBRs with demountable thermometers, respectively, unless otherwise specified in the TD. 6.8 Drawing up the results 6.8.1 The calibration and verification results are entered into the protocols, the forms for which are given in the Annex. 6.8.2 When verification or calibration results are favorable, a verification or calibration document (certificate, report) is issued. When verification or calibration results are unfavorable, a verification or calibration document (certificate, report) that clearly states the unserviceability of the instrument and the reasons being identified is issued. 6.8.3 Legal metrological control system on BBR is significantly different in each member state. Accordingly, the practical procedure, which is used to inform a result of verification or calibration and to grant a permission to use the instrument by issuing a certificate or a verification mark, is specified by the national authority in each member state 6.8.4 If appropriate, the following data and parameters shall be specified in calibration or verification certificates: 1) 2) 3) 4) 5) 6) 7) 8) 9) 10) 11) radiating area dimensions of the BBR forming the effective geometric extend (emissivity, area and solid angle, cavity dimensions in case of the cavity BBR); warm-up time of the BBR; transition time for the BBR to pass from one stationary mode to another; temperature drift of the BBR; temperature control instability of the BBR at the specified stationary mode; correction to the BBR own thermometer indication; expanded uncertainty of the BBR temperature value; positions of the temperature control device (if it given) of the BBR unit depending on temperature (in the table form); indication about using the BBR as a reference one; period of validity of the BBR calibration/verification. If appropriate, the following data and parameters shall be specified in calibration or verification certificates 23 OIML/TC11/SC3/N5 7 Annex A Recommended forms of the calibration and verification certificates 1 Results of the evaluation of the blackbody radiator cavity geometry Aperture diameter, mm Permissible value 2 Results of the evaluation of the warm-up time, temperature drift and transition time for the blackbody radiator to pass from a stationary mode to another Temperature T, oC* 3. Measured value Distance from the aperture to the back wall of the BB model, mm Permissible value Measured value Readings of the BBR own thermometer to be calibrated (verified), oC Average temperature values at the time intervals t1, t2, t3, oC t1 (0–5 min) t2 (5-10 min) t3 (10-15 min) Maximum difference of the average temperature values at the time intervals t, t2, t3, oC Value calculated Permissible by the value measurement data Results of the evaluation of the blackbody radiator temperature control uncertainty Temperature T, oC Readings of the thermometer of a BBR to be calibrated (verified), oC Average temperature value, o C Maximum deviation from the average temperature value, oC Value calculated by the Permissible value measurement data 4 Results of the correction evaluation during sighting along the blackbody radiator axis 4.1 Results of the correction evaluation using full radiation comparators Temperature T, oC Thermometer readings of a BBR to be calibrated (verified), oC Radiance temperature of a standard BBR, oC Difference between the radiance temperature of a standard BBR and thermometer readings of a BBR to be calibrated (verified), oC 24 OIML/TC11/SC3/N5 4.2 Results of the correction evaluation using pyrometer-comparators with a partial spectral range Temperature o T, C Thermometer readings of a BBR to be calibrated o (verified), C Spectral range, μm Radiance temperature of a standard BBR, oC Difference between the radiance temperature of a standard BBR and thermometer readings of a o BBR to be calibrated (verified), C The average temperature difference for all spectral ranges, oC… The maximum deviation from the average temperature difference, oC… 4.3 Results of the correction evaluation depending on the viewing angle of a BBR to be calibrated (verified) The table is filled for each viewing angle according to 4.2. Then the results are summarized in the following table: Temperature T, oC Viewing angle, i , 1 i m , grad Differences between the standard BBR temperature and temperature indicated by the thermometer of the calibrated (verified) BBR thermometer for 1 j n spectral bands, oC Average by all The j -th band angles Uncertainty of correction for each band and angle, o C Average value* of the temperature correction for all spectral ranges and angles, oC The maximum deviation from the average temperature correction for all spectral ranges and angles, oC… 4.4 Results of the correction evaluation depending on the location of sighting at a BBR to be calibrated (verified) Temperature T, o C Thermometer readings of a standard BBR, o C Coordinates of the location of view, mm The BBR own thermometer readings, o C Difference between the thermometer readings of a standard BBR and a BBR to be calibrated (verified), oC NB. T is the temperature set measured by the own thermometer of the BBR to be calibrated (verified) The average temperature difference for all sighting locations, oC… The maximum deviation from the average temperature difference, oC… 25 OIML/TC11/SC3/N5 8 Annex B References [1] OIML VIM-2010 – International Vocabulary of Metrology – Basic and General Concepts and Associated Terms (VIM). 3rd Edition (Bilingual E/F), 2010, minor corrections 2012 [2] OIML VIML-2000 – International vocabulary of terms in legal metrology (VIML) (bilingual French-English) / Vocabulaire international des termes de métrologie légale (VIML) (bilingue franзais-anglais) , edition 2000 [3] Guide to the Expression of Uncertainty in Measurement: First edition. – ISO, Switzerland, 1993 [4] J. Fischer et al., «CCT-WG5 on radiation thermometry, Uncertainty budgets for realization of scales by radiation thermometry», 2003, CIPM, CCT/03-03. Summary in Temperature, Its Measurement and Control in Science and Industry, 2003 vol.7, D.C.Ripple ed., Melville, New York, pp.631-638 [5] J. Fischer et al., «CCT-WG5 on radiation thermometry, Uncertainty budgets for calibration of radiation thermometers below the silver point», Ver. 1.71, CIPM, CCT-WG5/docs-03-2008, (http://www.bipm.org/wg/AllowedDocuments.jsp?wg=CCT-WG5); Int J Thermophys (2008) 29:1066–1083; DOI 10.1007/s10765-008-0385-1 26