The 14th IFToMM World Congress, Taipei, Taiwan, October 25-30, 2015 DOI Number: 10.6567/IFToMM.14TH.WC.OS19.017 Sensitivity Calibration of a Lead-Zirconate-Titanate (PZT) Thin-Film Intracochlear Microphone Chuan Luo1 Tsinghua University Beijing, China M. Staley2 Siemens Wind Service Orlando, United States Abstract: An intracochlear microphone is a critical component to enable totally implantable cochlear prostheses. In this study, the intracochlear microphone takes the form of a piezoelectric diaphragm with an area . A of 0.8 mm by 0.8 mm and a thickness of roughly 2 tuning fork generates a pure-tone sound at 4096 Hz exciting the piezoelectric diaphragm. In the meantime, a laser Doppler vibrometer measures the vibration amplitude at the center of the diaphragm. Also, a charge amplifier measures the charge generated by the piezoelectric diaphragm. The measured sensitivity, defined as the charge-to-displacement ratio, is around 53 pC/ . For a purpose of comparison, a finite element analysis is conducted to estimate the sensitivity. The simulation shows that the piezoelectric diaphragm has a sensitivity of 30 pC/ . The close agreement between the numerical and experimental results assures the reliability of the measured sensitivity. Keywords: Intracochlear microphones, Lead-zirconate-titanate, thin films, diaphragms, sensitivity calibration. I. Introduction Normal ears rely on hair cells in cochlea to transduce pressure waves into neural signals that generate the sense of hearing. For patients with severe hearing loss, the hair cells are so badly damaged that the mechano-electric transduction is largely lost. In this case, a cochlear implant, also known as a neural prosthesis, is introduced to treat patients with severe hearing loss. Basically, a cochlear implant is an electrode array surgically inserted in scala tympani of cochlea. The electrode array stimulates the auditory nerve electrically, bypassing the mechano-electrical transduction mechanism (i.e., the hair cells) in cochlea. To perform these functions, cochlear implants have a behind-the-ear unit that houses a microphone and a speech processor. A latest development in cochlear implants is to make the implants totally implantable [1-5]. In other words, there will be no behind-the-ear unit. Therefore, the microphone must be buried inside the body. Totally implantable cochlear prostheses have the advantage of no bulky external components, more natural hearing via auditory pathways, more versatile speech processing algorithms, and reduced surgical time and complexity [1-5]. Nevertheless, recent clinical trials using invisible microphones implanted under the skin reveal increased body noise interference and reduced microphone sensitivity [4]. 1 luochuan@mail.tsinghua.edu.cn mckenzie.staley@siemens.com 3 gzcao@u.washington.edu 4 ishen@u.washington.edu 2 G. Z. Cao3 University of Washington Seattle, United States I. Y. Shen4 University of Washington Seattle, United States One way to bypass the current difficulty encountered in the invisible microphones is to use an intracochlear microphone, i.e., a microphone housed inside the cochlear [6]. Since a good percentage of patients qualified for cochlear implants do not have conduction hearing loss, a microphone inside the inner ear will not only be a viable approach but also provide easy integration with the electrode array. Piezoelectric microphones are good candidates for intracochlear applications because of their high bandwidth and sensitivity especially in aqueous environments. In fact, many hydrophones are piezoelectric based. Piezoelectric microphones usually take the form of a piezoelectric diaphragm [7-11], which is relatively easy to deflect thus capable of producing more charge and a higher sensitivity. There are, however, many challenges. One major challenge is size constraints. Most micromachined piezoelectric microphones have a size of 2-4 mm [7-11]. These microphones are all too big to fit into cochlea, where cross section of scala tympani is roughly 1 mm by 0.5 mm. Another challenge is to maintain a high sensitivity while keeping the size of the microphones small enough to fit in the cochlea. As a piezoelectric diaphragm is scaled down in size, its sensitive is reduced accordingly [12]. Yet another challenge is calibration of the microphones given their small size. Ideally, a direct calibration is preferred. For example, an acoustic wave excites the piezoelectric diaphragm to generate a voltage output of the microphone. At the same time, the pressure of the acoustic wave is measured via a reference microphone. By comparing the voltage output and the pressure from the reference microphone, one can obtain the sensitivity of the microphone in V/Pa [7-11]. The direct calibration method described above is very costly. It often requires capital investment, such as an anechoric chamber [11] or a plane wave tube [8]. Also, the piezoelectric diaphragm has a very tiny surface area. Technically, the reference microphone does not exactly measure the pressure acting on the diaphragm. In this paper, we investigate the feasibility of using a piezoelectric diaphragm as an intracochlear microphone. Specifically, we develop strategies to address the three challenges described above. Figure 1 shows a piezoelectric diaphragm studied in this paper. The diaphragm is 0.8 mm by 0.8 mm in size, . The diaphragm is and its thickness is roughly 2 located at the tip of a cantilever probe. The diaphragm can serve as an acoustic actuator when voltage is applied. Conversely, the diaphragm can serve as a microphone. When an acoustic load is applied to the diaphragm, a charge amplifier connected Lead-Zirconiumto the electrodes can measure novel small-scale piezoelectric the generated charge. As an uators to be implanted in the inneradded ear. convenience, The PZT such a piezoelectric diaphragm has been successful fabricated re wave directly stimulating in the and thoroughly studied as perilymph an acoustic actuator [13,14]. It Togetherhas with a shortened electrode, the PZTfor animal also been implanted in guinea pig cochlea testsacoustic [15]. Therefore, the piezoelectric electric and stimulation (CEAS)diaphragm of the in Fig. meets the size small enough to fit in s research,1 we have twoconstraint, goals to i.e., achieve. cochlea. Therefore, it can readily be tested as an intracochlear microphone for its sensitivity in the feasibility ate a piezoelectric acoustic actuator to fit inside study. lop an actuator probe with 1 mm wide, 10 mm the tip of the Driving Voltage Piezoelectric from Amplifier gm serving as Diaphragm 0.8 mm x 0.8 mm is applied, the erate acoustic The designed .8 mm, and is 8.5-9 mm tire actuator is 0.4 mm (outside cochlea) together with 1-1.5 mm 1 mm (inside cochlea) trodes of the and analyzed. Finally, we conduct a finite element analysis to estimate the sensitivity to assure the measured sensitivity. II. Specimen Preparation The design and fabrication of the specimen is published in [13,14] with great detail. Only key features are summarized here. The diaphragm consists of a layer of silicon base materials (silicon oxide and nitride) of about thick and a layer of lead-zirconate-titanate (PZT) of 1 thick; see Figs. 2 and 3. There are three roughly 1 electrodes. One is a bottom electrode providing the electric ground (not visible in Fig. 3). There are also a center electrode and an outer electrode (separated by a yellow polygon in Fig. 3). Fig. 1 Proposed PZT probe Fig. 1. Piezoelectric diaphragm as an Pressure acoustic actuator and a microphone Guinea Pig Wave Cochlea , the tip of the mm) will To be bypass the challenge encountered in direct calibration, we adopt an indirect calibration procedure as cochlea before follows. An acoustic wave is first generated to excite the The remaining diaphragm. At the same time, the diaphragm displacement side cochlea to Auditory to give the and the charge generated are measured Brainstem e piezoelectric sensitivity in the form of a charge-to-displacement ratio Response e, the actuator ). Then one can convert the displacement (i.e., Piezoelectric sensitivity ing the basilar Probe Tip from to the pressure sensitivity V/Pa by Round Window to Pa, one can simple calculations. To convert from piral of cochlea use the stiffness of the diaphragm obtained from a finite Fig. 2 Feasibility element analysis, for example.study To convert from pC to V, one can use the capacitance of the piezoelectric nction of the intracochlear acoustic diaphragm. With thispiezoelectric approach, we can use no anechoric chamber. Moreover, the displacement measurement is eve this goal, we will acutely implant the actuator directly from the piezoelectric diaphragm. ea of guinea pigs through the round window; see It is, however, diaphragm important to generate stable acoustic vibrate the piezoelectric to generate excitations in the indirect calibration process. Ideally, the m response is measured to prove the feasibility acoustic excitations should contain all frequencies (e.g., a random noise), so that the measured response can be averaged to enhance the signal-to-noise ratio and the measured sensitivity will be valid for a wide range of frequencies. In our preliminary studies, however, we find that the PZT diaphragm is very sensitive to electroy in agingmagnetic seniorsinterference. and people work longmeans to Anywho electromagnetic ority of persons hearing loss have generatewith acoustic excitations (e.g., sensorispeakers) severely interferes the piezoelectric measurements. As a result, the calibration process is limited only to mechanical excitations. Three means that mechanically generate acoustic excitations are tested: an air jet via a 1nozzle, an air horn, and a tuning fork. Only the tuning fork gives repeatable results. For the rest of the paper, we first briefly explain the design and fabrication of the specimen that contains the piezoelectric diaphragm. Then we show the experimental setup that adopts the indirect calibration process to measure the sensitivity. Experimental results are shown Fig. 2. A microphone probe with a PZT diaphragm at the tip Fig. 3. PZT diaphragm viewed under a magnifying glass The PZT diaphragm is fabricated via a sol-gel process and a deep reactive etch (DRIE) as follows. The silicon wafer is first treated with oxide and nitride. A Pt-Ti bottom electrode is deposited. Then PZT sol is coated and sintered three times to form the PZT thin film. Next, an Au-Cr top electrode layer is deposited, patterned, and etched to form the center and outer electrodes. Finally, the silicon wafer with the PZT thin film is etched via DRIE from the backside to form the diaphragm and to release the probes simultaneously. The probes are further coated with Parylene to prevent the probes from short circuits in aqueous environments. The first natural frequency of the PZT diaphragm is generally between 60 and 90 kHz. The large variation of Displacement[nm/V] the natural frequency results from thermal residual stresses developed in the PZT diaphragm. During the C. As fabrication, the PZT needs to be sintered at the PZT, the electrode, and the silicon wafer cool down to room temperature, significant thermal stresses develop. ) compared Since the diaphragm is very thin (cf. 2 ), membrane residual with its large width (cf. 800 stresses will significantly affect the natural frequency of the diaphragm. Figure 4 shows a measured frequency response function when the PZT diaphragm is excited electrically as an acoustic actuator [14]. A distinct resonance at the first natural frequency (about 80 kHz) is observed. Below the first natural frequency, the frequency response of the PZT diaphragm is virtually flat. Since the audible frequency range (e.g., less than 15 kHz) is much lower than the natural frequency, the PZT diaphragm can be considered as a static structure in the audible frequency range. This implies that calibration of the diaphragm at a single audible frequency will suffice. To O-Scope PZT Thin Film LDV Top Electrode To Charge Amplifier Bottom Electrode Silicon Substrate Diaphragm Suspension Sound Source Fig. 5. Experimental setup to measure microphone sensitivity + 800 Resonance peak 600 Low frequency flat zone 400 200 0 0 20 40 60 80 100 Frequency [kHz] Phase [deg] 200 Fig. 6. Tuning fork used to generate the acoustic excitations in the experiment 100 0 -100 -200 0 20 40 60 80 100 Frequency [kHz] Fig. 4. A typical frequency response function of the PZT diaphragm when driven electrically [14] III. Experimental Setup and Measured Results Figure 5 illustrates the experimental setup, which includes the following elements. The first element is a tuning fork as the sound source (Fig. 6). A plastic mallet is struck on the tuning force to produce a pure-tone acoustic excitation at 4.096 kHz. Once the PZT diaphragm is excited, a laser Doppler vibrometer measures the velocity at the center of the diaphragm. Since the excitation frequency is known, the velocity can be converted to a displacement. At the same time, a double-end charge amplifier measures the generated charge from the PZT diaphragm via the center electrode. The charge amplifier needs to be carefully grounded and shielded to reduce electromagnetic interference [7]. The velocity and charge measurements are both fed into a digital oscilloscope for comparison. Figure 7 shows the measured response from the digital oscilloscope. The top trace is the velocity measurement from the laser Doppler vibrometer, and the bottom trace is the charge measurement from the charge amplifier. Both measurements have the same frequency. The velocity measurement is relatively clean with a good sinusoidal response. Although the charge measurement presents some noise, it does follow the velocity measurement in an out-of-phase manner very consistently at the same frequency. Fig. 7. Velocity and charge measurements from the digital oscilloscope Based on the measurements above, we take 23 peak values to calculate the sensitivity. The average sensitivity with a standard deviation of 6.7 . is 53.1 IV. Finite Element Simulations and Discussions To assure that the measured sensitivity is reasonable, we further conduct a finite element analysis to see if the sensitivity falls within the same ballpark range. Figure 8 shows the finite element model that we used to estimate the sensitivity. Since the finite element model has been published with great detail in [13,14], it is summarized only briefly for reference here. The finite element model is one-eighth of the PZT diaphragm. The diaphragm is fixed at the outside boundary simulated by the bulky silicon substrate. For the other two sides, symmetric boundary conditions are imposed. The diaphragm has multiple layers, such as silicon and PZT layers. The PZT layer is modeled via piezoelectric elements, while the rest is modeled via solid elements. At the juncture of the diaphragm and the silicon substrate, there is a transitional area in the form of trapezoid, which simulates residual silicon that is not removed by the DRIE process. Moreover, the residual silicon has a unique geometry. It is circular inside at the interface with the diaphragm, but it is square outside at the interface with the silicon substrate. It turns out that the residual silicon is important in determining mechanical response and consequently the sensitivity of the PZT diaphragm [16]. Therefore, it must be included in the model. potential can be calculated from the finite element analysis. The finite element analysis shows that the center electrode presents a voltage of 0.176 mV and the outer electrode presents a voltage of 0.0599 mV. Table 1 shows the area, capacitance, and the sensitivity of the center and outer electrodes. The capacitance is calculated from the parallel plate formula given the dielectric constant of PZT-7A (i.e., 274.15). With the voltage and capacitance, we can calculate the charge generated from each electrode. With a displacement of 1.921 nm, the sensitivity for the center and outer and 31.56 electrodes are found to be 29.84 , respectively. Fig. 8. Finite element model to estimate sensitivity of the PZT diaphragm V. Conclusions In this paper, we experimentally calibrate the sensitivity of a PZT thin-film diaphragm of a size 0.8 mm by 0.8 mm. A tuning fork is a reliable source to generate acoustic excitations driving the PZT diagram. Via a laser Doppler vibrometer and a charge amplifier, the measured . sensitivity of the PZT diaphragm is around 53 An independent finite element analysis shows that the . sensitivity of the PZT diaphragm is around 30 Given uncertainties associated with the PZT diaphragm, the agreement between the measured and calculated sensitivities is consider close thus assuring the accuracy of the calibration. The one-eighth model is sufficient for the purpose of this simulation for several reasons. First, a static analysis is performed on the finite element model, because the diaphragm behaves like a static structure in the audible frequency range. Second, a unit pressure of 1 Pa is applied to the diaphragm. Under these conditions and geometry, the deflection of the diaphragm with present symmetry with respect to the centerlines and diagonals. Therefore, one-eighth model will suffice. Under the static load of 1 Pa, the corresponding displacement at the center of the diaphragm is calculated as 1.921 nm. As shown in Fig. 8, the deflection is not uniform over the diaphragm. It reaches the maximal value of 1.921 nm at the center and gradually transition to zero somewhere in the residual silicon. For the rest of the paper, we will only refer to the displacement at the center of the diaphragm. For the PZT layer, its bottom side is grounded with zero electric potential. The electrical boundary condition at the topside of the PZT layer is free, so that an electric TABLE I. Parameters of the center and outer electrodes There are a couple of issues worthy of further discussions. First, the voltage of 0.176 mV is definitely small at 1 Pa. Nevertheless, it is better than other intracochlear microphone that has a response in the range of at 1 Pa [5]. Second, the calculated sensitivity is in the same order of the measured sensitivity, basically 30 vs. 53 . These two sensitivities agree very well with each other given many uncertainties in the finite element models. For example, the exact thickness of the diaphragm and the size of the residual silicon are often not known. Even under a scanning electronic microscope, dimension measurements can easily have 10% variations. Residual stresses in the diaphragm are also not known thus affecting stiffness of the diaphragm. VI. Acknowledgment This material is based upon work supported by the National Science Foundation under Grant Nos. CMMI-1030047 and CBET-1159623. Any opinions, findings, and conclusions or recommendations expressed in this material are those of the authors and do not necessarily reflect the views of the National Science Foundation. References [1] P. G. Berrang, H. V. Bluger, S. D. Jarvin, and A. J. Lupin, "Totally implantable cochlear prosthesis," US Patent 6648914. [2] A. J. Maniglia and W. H. Ko, "Totally implantable cochlear implant for improvement of partial and total sensorineural hearing loss," US Patent 6161046. [3] N. Cohen, "The totally implantable cochlear implant," Ear and Hearing, vol. 28, pp. 100S-101S, 2007. [4] R. J. S. Briggs, H. C. Eder, P. M. Seligman, R. S. C. Cowan, K. L. Plant, J. Dalton, D. K. Money, and J. F. Patrick, "Initial clinical experience with a totally implantable cochlear implant research device," Otology & Neurotology, vol. 29, pp. 114-119, 2008. [5] M. L. Carlson, C. L. W. Driscoll, R. H. Gifford,and S. O. McMenomey, 2012, “Cochlear Implantation: Current and Future Device Options,” Otolaryngologic Clinics of North America, Vol. 45, pp. 221-248. [6] L. A. Zhang, "Acoustic Sensors for Totally Implantable Cochlear Implant," in PhD Thesis, Melbourne, Australia: Department of Otolaryngology, University of Melbourne, 2008. [7] W. S. Lee and S. S. Lee, 2008, Piezoelectric Microphone Built on Circular Diaphragm, Sensors and Actuators A—Physical, Vol 144, pp. 367-373. [8] M. D. Williams, B. A Griffin, T. N. Reagan, J. R. nderbrink, and M. Sheplak, 2012, “An AlN MEMS Piezoelectric Microphone for Aeroacoustic Applications,” Journal of Microelectromechanical Systems, Vol. 21, pp. 270-283. [9] D. Kim, N. N. Hewa-Kasakarage, M. L. Kuntzman, K. D. Kirk, S H. Yoon, and N. A. Hall, 2013, “Piezoelectreic Micromachined Microphones with Out-of-plane directivity,” Applied Physics Letters, Vol. 103, paper 013502 [10] L. Baumgartel, A. Vafanejad, S.-J. Chen, and E. S. Kim, 2013, “Resonance-Enhanced Piezoelectric Microphone Array for Broadband or Prefiltered Acoustic Sensing,” Journal of Microelectromechani al Systems, Vol .22, pp. 107-114. [11] M. L. Kuntzman, N. N. Hewa-Kasakarage, A. Rocha, D. Kim and N A. Hall, 2015, “Micromachined In-Plane Pressure-Gradient Piezoelectric Microphones,” IEEE Sensors Journal, Vol 15, pp.1347-1357. [12] S. B. Horowitz, A. D. Mathias, J. R. Fox, J. P. Cortes, M. Sanghadasa, and P. Ashley, 2012, “Effects of Scalling and Geometry on the Performance of Piezoelectric Microphones,” Sensors and Actuators A—Physical, Vil. 185, pp.24-32. [13] Chuan Luo, G. Z. Cao, and I. Y. Shen, 2013: Development of a Lead-Zirconate-Titanate (PZT) Thin-Film Microactuator Probe for Intracochlear Applications. Sensors and Actuators A—Physical, Vol. 201, pp. 1-9. [14] Chuan Luo, 2012, PZT Thin Film Micro Probe Device with Dual Electrodes, PhD Dissertation, University of Washington. [15] Luo Chuan, Irina Omelchenko, Robert Manson, Carol Robbins, Elizabeth C. Oesterle, G. Z. Cao, I. Y. Shen, Clifford R. Hume, 2015, “Direct Intracochlear Acoustic Stimulation using a PZT Microactuator,” Abstracts of 2015 Conference on Implantable Auditory Prostheses, July 12-17, Lake Tahoe, CA. (podium talk). [16] Cheng-Chun Lee, Qing Guo, G. Z. Cao, and I. Y. Shen, 2008: Effect of Electrode Size and Silicon Residue on Piezoelectric Membrane Actuators. Sensors and Actuators A—Physical, Vol. 147, pp. 279-285.