HANDBOOK FOR MAGNAFLUX Y8 ELECTROMAGNETIC YOKE PART NUMBER: 001Y024 Handbook for Magnaflux Y8 Electromagnetic Yoke Y6/Y8 YOKE KITS Certificate of Conformity Y8 Yoke Serial number: Certified that the above item conforms to and meets the requirements of the following: In kit form, the Y6/Y8 Yokes are supplied complete with black magnetic ink, white contrast paint and a wire brush and duster, all housed in a strong attractive carrying case. A light source, attached to one leg of the Y6 Yoke, is activated by the magnetic field and switches on automatically during testing. Ideal for darker corners. EC Directives Specifications Light source for Y6 Yoke – part number 002L115. 73/23/EEC 89/392/EEC 91/368/EEC 89/336/EEC (Emissions) 92/31/EEC (Emissions) ASME V ART7 ASTM E-709 ASTM E-1444 L-10 MAGNETISING COIL Certificate is issued under the auspices of the Equipment Product Manager (2011): Developed for testing shafts, spindles and similar components. Features: Magnaflux (A Division of ITW Ltd), Faraday Road, South Dorcan Industrial Estate, Swindon, Wiltshire, SN3 5HE, UK. Ideal for inspecting for transverse cracks Components are magnetised and demagnetised with the same coil 110V operation Wet or dry Magnetic Particle Inspection (MPI) method can be used Footswitch provides continuous control Magnetic particle colours available: Tel: +44 (0)1793 524566 Fax: +44 (0)1793 490459 Email: sales@magnaflux.co.uk www.magnaflux.com Dry method – grey, black, red Wet method – black, red, fluorescent USING THE Y8 YOKE PREPARED INKS Flash point (°C min) (PMCC) Viscosity @ 21°C (cS) Specific gravity Particle size (µm) Settlement volume (%V/V) (as supplied) 93 3.45 0.81 7 to 25 0.12 Magnaglo 410HF* A prepared ink consisting of Magnaglo MG410 in a high flash kerosene of low odour Magnaglo 14HF* A prepared ink consisting of Magnaglo 14A in a high flash kerosene of low odour Magnavis 7HF* A prepared ink consisting of Magnaflux 7C in a high flash kerosene of low odour 93 3.45 0.81 2 to 25 0.25 93 3.45 0.81 0.6 to 2.5 2.5 Note 1: This Yoke can be supplemented by a magnetising coil for inspecting mounted spindles, rear axles and other light parts for transverse cracks only. Note 2: Fluorescent magnetic ink with high intensity black light improves the sensitivity of both the Y8 Yoke and L10 Magnetising Coil 1. 2. 3. *Available in both aerosol and bulk format DRY POWDERS Place the Yoke on the workpiece perpendicular to the direction of the cracks. Double jointed articulated legs can be moved in two directions at each joint. You can then adjust the spacing between the legs from about 25 to 250mm and contour the feet to produce a good flat contact point on parts with irregular shapes. Trigger the switch, apply the inspection medium, either wet or dry, and surface crack indications form immediately. You can proceed rapidly in a series of steps each finding cracks in a 150 x 150mm area. WARNINGS 1 Grey Maximum working temperature 315°C 3A Black Maximum working temperature 230°C 8A Red Maximum working temperature 175°C Magnaflux dry powders offer a range of colour to give good contrast on as wide a variety of finishes as possible. In the event of contrast not being ideal for inspection, a thin layer of Magnaflux WCP-2 White Contrast Paint can be applied prior to testing. All powders are of closely controlled particle size and shape and have desirable magnetic properties. For best results application should be as a cloud applied near the surface while the current is flowing. Excess powder settling out can be blown off. 4. WCP-2 White Contrast Paint Magnaflux WCP-2 is a quick-drying white contrast paint which can be applied as a thin coating prior to testing where enhanced contrast is required. SPECIFICATIONS Magnaflux, Magnaglo and Magnavis concentrates, prepared inks and dry powders meet the requirements of ASTM E-1444 with oil suspended inks and concentrates meeting AMS specifications, plus appropriate industrial and Government specifications, Certifications are available on request. 5. DO NOT smoke while performing NDT DO NOT operate the Y8 Yoke for longer than 3 seconds ON followed by 10 seconds OFF, that is a 23% duty cycle If the Yoke is too hot to hold in the bare hand it is a sign that the duty cycle has been exceeded. Wait for the Yoke to cool before continuing. DO NOT use any means to permanently operate the Yoke switch DO NOT use the supply cable to pull, lift or carry the equipment When the Yoke is switched OFF the magnetic attraction to the tested component will be weakened and either the part or the Yoke could fall and cause injury Laminar magnetic iron provides maximum efficiency with light weight. The trigger switch is easily operated without changing hand positions. As magnetic contact exists only between the Yoke and the test piece, the switch, completely enclosed by a rubber gasket, is the only break in the electrical circuit. Thus arcing on the part itself is impossible. SAFETY DATA RECHARGING DATA FLAMMABILITY Date The Y8 Yoke is intended to be used in conjunction with appropriate chemicals as a Non Destructive means of Testing (NDT) for defects, such as cracks, on a wide range of manufactured components. Some chemicals may produce a flammable atmosphere at the point of use and it is important that the testing is carried out in a well ventilated place and that all sources of ignition are excluded. Good quality NDT chemicals such as Magnaflux magnetic inks and powders are formulated to minimise the risk of a flammable atmosphere when correctly used. MAINTENANCE Before using the Yoke, battery or battery charger: Ensure that they are physically undamaged Ensure the cables are free from cuts which expose the wiring Tighten the articulated joints if they are excessively loose If the switch area of the Yoke becomes wet with either kerosene or water it should be disconnected from its supply until the area has dried If the Y8 shows signs of malfunctioning or if cracks appear in the casing, the Yoke should be removed from service immediately and examined by a qualified electrician. Cracking is usually caused by dropping the Yoke or by twisting the articulated legs. It is a legal requirement in many countries that the electrical safety of equipment is checked periodically, commensurate with use, but at least once a year. The basic tests required are to check the condition and effectiveness of the electrical insulation and the continuity of the protective conductor (earth). These tests can be performed using as Portable Appliance Tester (PAT) or Portable Appliance Checker (PAC). Test lift Remarks Signed HANDLING AND STORAGE OF BATTERIES MAGNETIC FIELDS Electric and magnetic fields (EMF’s) arise from the generation, transmission and use of electricity. The result is that everyone today is exposed to a complex mix of EMF’s both at home and work. Store batteries in a cool and dry area with an impervious surface Store batteries under a roof and protect against adverse weather conditions Protect against physical damage and exposure to organic solvents Do not allow metal objects to contact both terminals at the same time, as this will cause sparking and possible injury Never install batteries in a gas tight enclosure as gases may be generated during use Batteries must be charged on a voltage regulated charging system and adequate ventilation provided to avoid build up of ignitable gases Under extreme conditions of charging equipment malfunction and/or battery failure, high voltage and high temperature conditions may occur causing evolution of hydrogen sulphide (H2S) gas, which is toxic. If detected by its odour of rotting eggs (at extremely low concentrations) switch of the charging equipment, evacuate all personnel from the area, and ventilate the area well. Seek appropriate advice before attempting to re-starting charging. ACCIDENTAL RELEASE MEASURES The batteries are designed not to leak under normal conditions. If, however, electrolyte does leak out of the battery for any reason, it should be absorbed onto dry sand, earth or other inert material and must not be allowed to enter the drains. If possible, neutralise any leaked electrolyte using soda ash, sodium bicarbonate, sodium carbonate or calcium carbonate, and then wash thoroughly with water. Collect the absorbed material and place in an inert sealed container for appropriate disposal. Magnetic fields are created when an electric current flows. Magnetic and electric fields then exist together. The greater the current the stronger the magnetic field. Magnetic particle inspection involves the generation of magnetic fields to locate defects in ferrous materials. It has been used for many years with no reported health effects. Although and extremely useful and effective Non Destructive Testing (NDT) method, there are growing concerns that exposure to EMF’s could have detrimental health effects. There is very little conclusive information on the effects of exposure to intense fields. However, research by various authorities has shown that under certain circumstances acute exposure to intense fields of around 60 mT can result in headaches and alterations in visual response. The National Radiological Protection Board (NRPB) has developed guidelines to protect people from these potential effects. No health risks have been established for weak fields. Typical magnetic field strengths for domestic appliances Vacuum cleaners – 2-20 microtesla @ 30 cm from the appliance Electric drills – 2-20 microtesla @ 30 cm from the appliance Typical magnetic field strengths for some Magnaflux MPI equipment MAG 50 coil – 1.3 millitesla @ 30 cm MAG 50 head/tailstock – 1.5 millitesla @ 45 cm Y6 Yoke < 1 millitesla @ 15 cm Units of measurement for Magnetic Field Strength – tesla (T) 1 tesla = 1,000 millitesla (mT); 1 millitesla = 1,000 microtesla (µT) MAGNETIC FIELD RISK Persons susceptible to strong magnetic fields, including those with pacemakers, are advised not to use or approach this equipment without seeking professional advice. TECHNICAL SPECIFICATION BATTERY HEALTH AND SAFETY YOKE AND BATTERY WARNING: DC Voltage 6 Supply Current (A) in air Weight (Kg) – lifted at 140 mm pole spacing BATTERY CHARGER 1.5 22.5 The gases given off when charging are explosive. No smoking or naked flames Ensure that cables and electrical equipment are handled in such a way to avoid sparks Before fitting or removing the batteries, make sure that all equipment connected to it is switched off to avoid accidental sparking Never short circuit the battery terminals as the resulting sparks and arcs can injure personnel and are a fire hazard Do not charge batteries above above +50°C, or disc harge or store above +60°C When charging, connect the battery to the charger before switching the charger on. Reverse the process at the end of charging. When charging indoors, ensure there is adequate ventilation WARNING: Input: 230 volts 50/60Hz Output: 6.85/7.25 volts DC @ 2 amps Treat sulphuric acid with respect. It is a highly corrosive liquid and will cause severe injury if splashed on the skin or in the eyes, or if taken internally. It rapidly destroys most clothing. Sulphuric acid is scheduled as a poison in Part 2 of the Poisons List. CERTIFIED TEST LIFT WEIGHTS Always: Available for both AC and DC. AC Test Weight (4.5 Kg) – part number 026T018B DC Test Weight (18 Kg) – part number 026T018A Wear acid-resistant clothing, goggles, PVC gloves and rubber boots Have a copious supply of water available Keep an eyewash bottle ready, containing 1% of saline solution BATTERY CHARGER PARTS LIST OPERATION Description Y8 Yoke cable 3 core First stage – constant current mode Visual indication – RED LED ON, GREEN LED OFF Second stage – constant voltage mode Visual indication – AS ABOVE Part Number 001C065 Y8 cable bush 014B027 Cable cover 17771A3 Leg/foot assembly 075C004 Third stage – float charge mode. The battery will be maintained 100% charged. Visual indication – GREEN LED ON, RED LED OFF Switch kit 005K026 LED PROTOCOL Y8 plug (PLG 1) 016P085 Battery charger plug (PLG 2) 016P085 Battery charger (BC 1) 039B002 Battery socket (SKT 1) 015S135 Battery (BAT 1) 003B014 Fuse (FS 1) 001F115 Charger status Bulk charge mode Float charge mode (charge complete) Battery reversal detected High temperature detected Short circuit Open circuit LED Status Red – static ON Green – static ON Red – flashing Red + Green flashing simultaneously Red + Green flashing alternatively Red + Green static simultaneously Note: On power up the system will enter a ‘soft start’ mode. This facility checks for possible faults, for example, reverse battery connection etc, before operating maximum charge current. Battery pouch 520797 WARNING DO NOT REPLACE ANY PARTS RECOMMENDED BY MAGNAFLUX. WITH OTHER THAN THOSE