sns medium-beta cryomodule performance

advertisement
SNS MEDIUM-BETA CRYOMODULE PERFORMANCE*
M. Drury†, I. E. Campisi, E. Daly, G. K. Davis, J. R. Delayen, C. Grenoble, J. Hogan, L. King, T.
Powers, J. Preble, M. Stirbet, H. Wang, M. Wiseman.
Thomas Jefferson National Accelerator Facility Newport News, Va, 23606, USA
Abstract
Thomas Jefferson National Accelerating Facility
(Jefferson Lab) is producing 24 Superconducting Radio
Frequency (SRF) cryomodules for the Spallation Neutron
Source (SNS) cold linac. This includes one medium-beta
(0.61) prototype, 11 medium-beta production, and 12
high-beta (0.81) production cryomodules. Each of the
medium-beta cryomodules is scheduled to undergo
complete operational performance testing at Jefferson
Laboratory before shipment to ORNL. To date, the
prototype and three production models of the mediumbeta cryomodule have been tested. The performance
results of the tested production cryomodules will be
discussed.
performed using a low power (1 W) setup. To date, the
prototype and three production models of the mediumbeta cryomodule have been tested. The performance
results of the tested production cryomodules will be
discussed.
TUNERS
The SNS tuner design has been adapted from a Saclay
tuner design for the TESLA cavities. For SNS application,
the tuner requirements are listed in table 1. In order to
meet the SNS requirements a piezo element has been
incorporated into the “dead leg” and provides the required
fine tuning adjustment [2].
Mechanical Tuner Tests
Jefferson Lab has completed construction of six
medium-beta cryomodules, and testing is currently
underway in the recently upgraded Cryomodule Test
Facility (CMTF) [1].
While installed in the CMTF, each cryomodule will be
subjected to a series of performance tests based on SNS
specifications.
The testing process begins once the cryomodule has
been cooled to 2K. Table 1 has a list of tests performed
as a part of the production testing process.
Table 1: List of Production Tests
Test or Procedure
Tuner Range,
Hysteresis
Tuner Resolution
High Power
Processing
Extended Run
Field Emission
Qext’s
Q0
Lorentz Force Pulse
Response
Purpose
Determine proper operation of
piezo and mechanical tuners.
Determine frequency resolution
of piezo and mechanical tuners.
Condition Coupler Vacuum,
Determine Emax.
Determine Maximum Stable
Gradient, Emaxop.
Measure Field Emission vs.
Gradient.
Measure Qext’s of Fundamental
Power Coupler, Field Probe and
Higher Order Mode Couplers.
Measure Q0 vs. Gradient.
Measure Dynamic Lorentz Force
Coefficient
Other measurements may be added as needed. The
high power production tests require the use of a 1 MW
pulsed klystron. Other tests, such as tuner tests are
The mechanical tuners are tested to determine
frequency range, hysteresis and resolution. The tuner is
cycled through its range from the clockwise limit switch
to the counter-clockwise limit switch. This is done in
steps of ~25 kHz (60000 half-steps of the stepper motor).
The tuner is then cycled through a smaller, 4 kHz loop
using increments of 600 half-steps (~200 Hz) for
determination of hysteresis. See Figure 1 for a typical
plot of frequency vs. motor position.
Cavity #1 (M02) Tuner Hysteresis
805.003
Frequency (MHz)
INTRODUCTION
805.002
805.001
805.000
804.999
804.998
804.997
-576000 -574000 -572000 -570000 -568000 -566000 -564000 -562000
Step Position
Figure 1: Mechanical Tuner Range and Hysteresis.
The SNS requirement for the mechanical tuner range is
200 kHz (805 MHZ ± 100 kHz). The average range for
cryomodules M1-M3 is 486 kHz ± 28 kHz. The average
hysteresis for this group of cavities is 90 Hz ± 33 Hz.
In order to determine the resolution of the mechanical
tuner, the stepper motor is cycled back and forth by a set
number of steps while using a cavity resonance monitor
(CRM) to look at the signal from the cavity field probe.
A VCO-PLL system and a 1 W amplifier are used to drive
the cavity. The number of motor steps is reduced until the
frequency shift measured by the CRM is less than 60 Hz
(based on the SNS requirement).
result of a typical test.
Figure 2 shows the
The first task is to process the coupler vacuum and
define the maximum gradient (Emax).
Cavity 2 M03 Piezo Resolution,
5V piezo amp output
2.5
2.0
1.5
1.0
0.5
0.0
-0.5
-1.0
-1.5
-2.0
-2.5
15
Cavity Frequency Shift
(Hz)
Cavity detuing (Hz)
M03 Cav 1 Mech Tuner Resolution
20 half-steps
10
5
0
-5
-10
-15
-10
-6
-4
-2
0
2
4
-8
-6
-4
-2
0
2
4
6
8
10
Time (sec)
6
Time (sec)
Figure 4: Piezo Tuner Resolution.
Figure 2: Mechanical tuner Resolution.
The average measured resolution for the tested cavities is
9.9 Hz ± 8.9 Hz.
Piezo Tuners
The same types of tests are performed on the piezo
tuners: range, hysteresis and resolution.
During the range and hysteresis tests, the piezo drive
voltage is increased from zero to 850 Volts in steps of 100
Volts (~ 250 Hz), and then decreased in the same fashion
to zero Volts. A typical plot is shown in Figure 3.
This is accomplished by gradually raising the gradient
while allowing time for the coupler vacuum to improve.
Eventually, the gradient is raised to a level that cannot be
sustained due to cavity quenches or an interlock fault is
triggered and the RF permit is removed. This condition
occurs as a result of arcing or vacuum degradation.
Figure 5 shows the progress over time as the gradient is
raised over a nine-hour period. It depicts the slow increase
in gradient and the behavior of the coupler vacuum. It is a
fairly typical session, although the times required vary
from as little as four hours to several days.
Processing Cavity 2-M02 6-24-03
Piezo Tuner Range Cavity M02-2
40
Eacc (MV/m)
Freq (MHz)
1.0E-07
45
805.0035
805.0030
805.0025
805.0020
35
30
25
1.0E-08
20
15
10
805.0015
5
805.0010
0
200
400
600
800
1000
Piezo Voltage (V)
0
8:00
Cplr 2 Vacuum (mbar)
50
805.0040
1.0E-09
10:00
12:00
14:00
16:00
18:00
20:00
time
Emax (MV/m) Ptran
CPLR TWO
Figure 3: Piezo Range and Hysteresis.
Figure 5: Processing Session.
The average measured frequency range for the piezo
tuners is 2.9 kHz ± 0.4 kHz. The average hysteresis is
643 Hz ± 349 Hz.
Finally, we measure the piezo tuner’s resolution. A
pulse generator is used to drive the piezo tuner with a 0.2
Hz, 50 % duty cycle pulse train. The cavity under test is
driven with a VCO-PLL and a 1 W amplifier. The CRM
is used to measure the frequency shift. The amplitude of
the pulse train is set initially at 0.05 V and reduced until
the measured frequency change is less than 60 Hz (peak
to peak). Figure 4 shows the result of a typical test.
Once Emax has been determined, we look for the
maximum stable gradient (Emaxop). We lower the
gradient by about 1 MV/m in order to avoid unstable
conditions such as quenches. The cavity must operate at
Emaxop in a stable fashion for at least an hour.
HIGH POWER MEASUREMENTS
After processing has been completed and the maximum
gradient has been defined, it is time to measure the
unloaded Q (Q0).
A calorimetric method is used. The cryomodule under
test is isolated from the cryogenic system by closing the
During high power measurements, a 1 MW klystron is
connected to the cavity under test. It is set up to deliver a
1.2 ms pulse with a 60 Hz repetition rate.
Table 2: Emax for Measured Cavities
SNS Requirement
•10.1 MV/m
Average for Emax
Average for Emaxop
18.8 MV/m ± 1 MV/m
17.4 MV/m ± 1.4 MV/m
helium input and output valves. The rate of change in
pressure (¨3¨W DV D UHVXOW RI WKH VWDWLF KHDW ORDG the static heat load plus a known input from a cryogenic
heater and 3) the static heat load plus the RF heat load.
From these measurements we are able to determine both
Q0 (and the primary (2 K) static heat load. Figure 6
shows the results of Q0 measurements on the first three
production cryomodules.
Qo vs Eacc
SNS Medium Beta Production Cryomodules
Qo
5x10
9
10.1 MV/m
1.0E+09
1.0E+08
7.0E+06 9.0E+06 1.1E+07 1.3E+07 1.5E+07 1.7E+07 1.9E+07 2.1E+07
Eacc (V/m)
Figure 6: Q0 vs. Eacc
Table 3 summarizes the results. The SNS requirement
is Q0 • 5× 10 9 at Eacc = 10.1 MV/m. The static heat load
requirement is 25 Watts or less.
Table 3 Average Q0 Values
Q0 at 10.1 MV/m
9
Eacc at 5× 10
Static Heat Load
10.5 W ± 1.1 W
-0 .0 6
C R M out
P tra n s
400
-0 .1 2
200
-0 .1 5
100
-0 .1 8
0
-0 .2 1
-1 0 0
-0 .02 5
50
40
30
20
10
0
15
20
Figure 7: Field Emission.
-0 .0 1 5
-0 .0 0 5
0 .0 05
T im e (s e c )
0 .0 1 5
-0 .2 4
0 .0 2 5
Figure 8: Frequency Response to Pulsed RF.
Finally, as part of the high power measurements, the
external Q’s (Qext) of the FPC, the field probes and the
HOM ports are measured. Table 3 summarizes the results
of those measurements.
SNS requirement
Measured value
7.3 × 10 5 ± 20%
1.052×1012 ”4H[W”
2.642×1012
7.1× 10 5 ± 13%
1.9 ×1012 ± 6.5 ×1011
Qext
HOM1
> 3×1010
1.2 ×1014 ± 1.9 × 1014
Qext
HOM2
> 3×1010
1.9 ×1010 ± 9.1× 1011
Qext field
probe
So far, two of the cavities tested have failed to meet the
requirement for the HOM 2 port.
CONCLUSION
M1-1
M1-2
M1-3
M2-1
M2-2
M2-3
M3-1
M3-2
M3-3
Eacc (MV/m)
-0 .0 9
300
Qext FPC
60
Radiation (mR/hr)
-0 .0 3
500
Table 4: Qext
Field Emission vs. Eacc
SNS Medium Beta Production Cryomodules
10
0 .0 0
600
1.4 × 1010 ± 1.9 × 10 9
15.6 MV/m ± 1.1 MV/m
At this time, we also look at field emission. A set of
five Geiger-Mueller (GM) tubes is placed around the
cryomodule and the response at different gradient levels is
recorded. Figure 7 shows the response of the GM tube
located at the beamline on the helium supply end of the
first three production cryomodules. This is the location
where the highest level of field emission has been
recorded so far.
5
700
(V)
1.0E+10
M1-1
M1-2
M1-3
M2-1
M2-2
M2-3
M3-1
M3-2
Ptrans Xtal Det Output
S N S M 0 3 C a v 1 R F Pu ls e R e s p o n s e a t 1 0 M V/m , 6 0 H z,
1 .3 m s e c
Cavity Freq Shift (Hz)
1.0E+11
The average gradient at which field emission starts for
these cavities is 10.3 MV/m ± 2.8 MV/m.
The Qo measurements are followed by the dynamic
Lorentz force measurement, which measures the
frequency response of the cavity to an RF pulse. Figure
89 shows the frequency shift to an RF pulse for a typical
cavity operating at 10 MV/m.
The average frequency shift for the medium-beta
production cavities tested so far is 283 Hz ± 29 Hz.
25
The first three production models of the SNS mediumbeta cryomodule have completed testing. The three
cryomodules have exceeded all requirements other than
the two exceptions noted above.
REFERENCES
[1] Powers, T. Allison, G. K. Davis, M. Drury, C.
Grenoble, L. King, T. Plawski, and J. Preble,
“Upgrade To Cryomodule Test Facility at Jefferson
Lab”, SRF Workshop 2003
[2] J. Preble, I. E. Campisi, E. Daly, G. K. Davis, J. R.
Delayen, M. Drury, C. Grenoble, J. Hogan, L. King,
P. Kneisel, J. Mammosser, T. Powers, M. Stirbet, H.
Wang, T. Whitlatch, M. Wiseman, “SNS Cryomodule
Performance”, PAC 2003
Download