Physics 18L: Modern Physics Lab Lab 1: Malus’s Law Jacob Schwartz Partner: Samuel Jacobsen University of California, Los Angeles Professor Regan Lab Section 1 April 7, 2010 Abstract In this lab, we investigate the polarization of light and Malus’s Law. We project light through a polarizer and analyzer and measure the resulting levels of light with an analog light meter. By rotating the analyzer we can measure how polarized light is transmitted or blocked by a second sheet of polaroid at an angle to the first in accordance with Malus’s Law. After fitting our data to theory, I found that I had underestimated our error bars (χ2 /DOF ≈ 1.8) and so I attempt to give an explanation and to suggest improvements to our experiment. In the conclusion I propose that we did indeed verify Malus’s Law. 1 Introduction and Theory Polarization is an important optical effect which exploits light’s nature as a transverse wave. Polarization is important to science, for example as a way to view material stress, but also in our daily lives in applications from LCD screens on digital watches to sunglasses, photography, and 3D movies. When an unpolarized beam of light is passed through a first polarizer sheet, only the electric field vector’s component in the direction of polarization can pass through. This leads to a reduction in intensity of the initial beam by one half, but now all the light is polarized in one plane. When this plane-polarized light is passed through a second polarizer at an angle α to the first, only part of the electric field vector E = E0 cos(α) can get through. Because the intensity of light I is proportional to the square of the electric field vector E: I(α) = I0 cos2 (α) This equation is called Malus’s Law. In this experiment we will attempt to demonstrate Malus’s Law by shining light through a pair of polarizers and recording the level of light which is transmitted based on the angle between them. We will also investigate and attempt to consider the effects of non-ideal polarizers, ie those which besides polarizing filter out some additional light or which do not polarize the light completely. 1 2 2.1 Experimental Arrangement and Procedure Malus’s Law setup To investigate Malus’s Law we used a setup with four main components: a projector, two polaroid filters called the polarizer and the analyzer, and a light meter. Light shown from the projector passes through the polarizer (our filter marked ”#2”), set at 0 degrees, to become polarized. Then it passes through the analyzer (marked ”#8”), set at a variable angle α where the light is reduced according to Malus’s law. Finally we measure the remaining light using a fiber optic cable attached to a light meter. 0° Projector Polarizer “#2” α° Analyzer “#8” Light Meter Figure 1: Setup for Malus’s Law experiment 2.2 Measurement Procedures Before starting we zeroed the light meter by covering up the receiving end of the meter with a finger and adjusting the knob until the meter read zero. Removing the two polaroid filters for now, we turned on the projector and took readings of the light meter every minute to check its stability over time. Our second calibration measurement was to determine whether the light coming from the projector was already somewhat polarized. With each of our filters (called #2 and #8) in turn, we placed it in the path of the beam and rotated it until we found the maxima and minima of light meter readings. We also took note of the reduction in intensity due to the filter so that we could determine how much it blocked light besides polarizing it. Now we took data for Malus’s law. We placed both filters in the path of the beam and set the polarizer to 0 degrees. With the analyzer also initially at 0 degrees, we took a reading from the meter and turned the analyzer by 10 degrees until we covered the whole circle. 2 3 3.1 Data, Analysis, and Results Malus’s Law Data As specified in the procedure we took light meter readings for each α from 0 to 360 degrees. From those values this plot was made: Malus's Law Data and Best Fit 7 6 Fit Function Measured Values Relative Intensity 5 4 3 2 1 0 0 20 40 60 80 100 120 140 160 180 200 220 240 260 280 300 320 340 360 Polarizer Angle (degrees) Figure 2: Data from measurements and plot of best fit I attempted to find the best fit of the graph with the function I(α) = I0 + I1 cos2 (α − φ), which describes a cos2 function with a raised floor I0 , amplitude I1 and phase shift φ. With those three variables, I attempted to minimize the sum of the squares of the residuals. Incorporating the errors present in the data, my lowest value for χ2 /DOF was about 1.8, which tells that I did not take into account all the errors or different effects which were present. The best value found for I0 was 0.11 ± 0.07, for I1 was 6.1 ± 0.1, and for φ was −0.02 ± 0.04. The errors on those values were found by finding the difference between the best value for a parameter and one which increased the χ2 by one. In order to confirm that this best fit was reasonable, I made a plot to show the difference between the data points and the points from the theoretical equation. 3 Difference in Relative Intensity Residuals 0.20 Data minus Theory 0.10 0.00 -0.10 -0.20 -0.30 -0.40 -0.50 0 20 40 60 80 100 120 140 160 180 200 220 240 260 280 300 320 340 360 Polarizer Angle (Degrees) Figure 3: Residual differences between data and theory It is true that there is more area below the axis than above, but the way that I weighted the errors (discussed in the next section) causes this to not be an issue. The errors in absolute terms are optimized to be smaller around the minima and maxima of cos2 (0, 90, 180, 270 degrees) and larger on the regions with high slope. 3.2 Results of calibration measurements, sources of error In our first check, of the stability of the meter over time, we found that it varied by roughly ±0.1 from the initial relative intensity value of 42.0 over the span of five minutes. To find the error from meter fluctuations for the Malus’s Law data, we cannot simply extrapolate downward by saying the error was 1/400 of the measured value because the meter itself has a certain resolution, and we saw that at lower reading values it had a much higher jitter proportionally. For the values that we read on the meter (which were in the neighborhood of 5) I would estimate that the error from the meter’s instability was around δmeter = ±0.02. I will use that number as a constant source of error over the whole 0 to 360 degree range. The error in our ability to read the meter accurately varied between δread = ±0.02 and ±0.04 depending on where precisely the needle was in relation to the meter’s tick marks. If the needle was directly over a tick mark it was generally easier to read, for example. I recorded this error for each measurement and they are incorporated individually into the χ2 value. Another source of error was our ability to accurately measure the rotation of the analyzer. The analyzer could slip back and forth in its holder such that even when a certain value’s tick mark was lined up with the reference mark the actual tilt could be several degrees different. I estimate that 2 degrees was the uncertainty. Since the measured intensity 4 value’s rate of change depends on the angle, the error in measured value from this constant 2 degrees error changes with angle as well. It is larger for the angles where measured intensity is changing rapidly and goes to zero for the extrema. The error from this source is calculated by δ δα = (I0 cos(α)2 ) = 2δα I0 cos(α) sin(α) δα The sum of each of these three errors was added in quadrature to find the total error. q 2 2 + δα2 δ = δmeter + δread In our best fit model and in our data, the minimum value was not zero as it should have been. This is because some ambient light was able to make it into the meter even when the polarizer and analyzer were at right angles to each other. Perhaps we should have zeroed the meter with ambient conditions instead of with a condition of total darkness. Another possible source of error which we investigated was the polarization inherent in the light from the projector. Using one of the polarizers (#8) we found that the maximum light level measured by the detector was 11.6 ± 0.1 and the minimum was 10.5 ± 0.1. This min gives a degree of polarization (1 − max ) of 0.095 ± 0.012. It is likely that the polarization from the projector slightly affected our data and caused a larger error, but in precisely what way I do not know. Finally, we measured the non-ideal nature of our polarizers, defined by the transmissioncorrection-factor F , defined as the observed maximum intensity divided by the expected max intensity. Without any polarizers, our light meter read 41.8 ± 0.1. When we placed the first polarizer in the beam, it should have read half of that, because Z 2π 1 cos2 (α) dα = 1/2 2π 0 However the first polarizer, our #8, read a maximum of 11.6 ± 0.1 and the second, #2, read a maximum of 11.2 ± 0.1. These correspond to F of 0.555 ± 0.005 and 0.533 ± 0.05, respectively (an ideal polarizer would have F = 1). Multiplying 41.8 ∗ 0.5 ∗ .555 ∗ .533 gives a number very close to our observed maximum value when both polarizers were in the beam of 6.2. 5 4 Conclusion When I first plotted the data for Malus’s law, I was quite worried to see that χ2 /DOF was roughly 28. That was when I realized I had to incorporate the error from not knowing α and from the meter’s instability. After optimizing again, I obtained a χ2 /DOF of 1.78, which is still a bit too large to call the experiment a total success, but I believe this is a very good value because it doesn’t yet take into account sources of errors and effects that were not modeled, such as the ambient light in the room (which may have changed), and polarized light from the projector. If I were to do this experiment again, I would have been much more careful about the error in α because that was by far our largest contribution to errors. It would have helped to only turn the analyzer in one direction to lead up to a measurement rather than turning it back and forth, so that this way it would always be in the same position inside its housing when we took a measurement. It would have also been helpful to take more readings more closely spaced, so that we make sure we hit the actual peak and also to have a better resolution on the slopes of the graph, where data points are more sparse. However, that would take much more time so I suppose every 10 degrees is a good compromise. Looking at the graph, the plots of data and fit are almost on top of each other, which leads me to believe that the problem with our data right now is that we need to incorporate a few more sources of error, rather than Malus’s law actually being of a form having dependancies other than cos2 (α). So, while our errors were not quite spot-on, we did show that the intensity of light passing through a set of polarizers varies as cos2 (α), confirming Malus’s law. 6