Warning Indicators and Fault Codes On the control panel there are a series of ‘hidden until active’ warning indicators. Some of these indicators will appear in conjunction with a fault code in the temperature display. The error message will appear in the red Temp Indication area. The meaning of these warnings, why they appear, and what to do when they appear, is as follows: F000 - Measuring System Error This is an internal temperature monitoring system via Analogue Input 06 (PRT). It highlights that the Tactrol Program software has recognised a temperature issue. Note: This fault will abort the cycle and NOT complete the sterilization process . F001 - Load Sense Probe Failure This warning is activated in the event of the failure of the load sensing thermocouple. If the autoclave is fitted with load sensed process timing, this should be deselected to enable the autoclave to run without this feature. The thermocouple should be replaced as soon as possible. Great care should be taken to ensure that loads which would ordinarily be autoclaved with load sensed process timing are adequately sterilised. Note: This fault will abort the cycle and NOT complete the sterilization process. F002 - Thermocouple Input Fault This highlights a temperature probe problem, and pinpoints the following probes specifically: F102 Control Probe Fault F202 Display Probe Fault F302 Load Sense Probe Fault (if fitted) F402 Load Simulator Probe Fault F502 Jacket Probe Fault (if fitted) F702 Heater-Over Temp 1 Probe Fault (if fitted) F802 Heater-Over Temp 2 Probe Fault (if fitted) Note: This fault will abort the cycle and NOT complete the sterilization process. F003 - Overheat on Process Dwell This indicates that one of the sensing probes has exceeded the Overheat Abort Temperature. If fitted, the heater over-temperature protection may have sensed that the heating element became too hot. This is probably due to a low water condition, which was not sensed by the low water probe. The water level and the condition of the probe should be checked before attempting to use the autoclave again. If heater over temperature protection is not fitted then the overheat cut out will only operate under extreme conditions, such as a failure of the temperature control system or a probe becoming open circuit. The next attempt to run the autoclave should be closely observed and as the fault(s) are cancelled all the code(s) displayed should be noted. F004 - Low Water Level The water level has fallen below the minimum level and must be topped up before the autoclave can be run. The warning will automatically cancel when the door is opened and the water level is topped up. The low water condition may have caused a running cycle to abort, and the load may need to be autoclaved again. Note: This fault code is not relevant on a steam heated machine. Should it occur on a steam heated machine then the configuration is incorrect or the main-board has been corrupted. (On dual heated machines the code should only occur when electric heating is selected.) Note: This fault will abort the cycle and NOT complete the sterilization process. F005 - Temperature Fall during Process Cycle This indicates a temperature drop of 3 degrees (default) from the cycle dwell set-point during the dwell time period. The temperature can be selected via the Dwell abort offset temperature on the GENERAL Tab Note: This fault will abort the cycle and NOT complete the sterilization process. F006 - Power Fail This indicates that a power failure of some kind has occurred during the process cycle. Note: This fault will abort the cycle and NOT complete the sterilization process. F007 - Vacuum Pre-cycle Pressure Loop Break The autoclave has not achieved the pre-set level of vacuum during the Precycle vacuum stage during the pre-set time. Note: This fault will abort the cycle and NOT complete the sterilization process. F008 - Heating stage timeout. The autoclave has not reached process temperature within the Pre-set time. Note: This fault will abort the cycle and NOT complete the sterilization process. F009 - Vacuum cooling. Set-point not achieved. The autoclave has not achieved a low enough level of vacuum during the post cycle vacuum stage (Vacuum Cooling or Drying Cycle). Note: This fault will abort the cycle and NOT complete the sterilization process. F010 - Air detector If fitted the air detector system has detected an over pressure condition symptomatic of excess air remaining in the load. Note: This fault will abort the cycle and NOT complete the sterilization process. F011 - Printer Timeout / Malfunction The control system has not received confirmation from the printer within its pre-set timeout. Note: This fault code is not relevant unless the machine is fitted with a printer. Should it occur on a machine without a printer then the configuration is incorrect or the main board has been corrupted. Note: This fault will abort the cycle and NOT complete the sterilization process F012 - Door micro-switch fault If a door micro-switch opens during a cycle this fault code is displayed. F013 Jacket timeout If a jacket is fitted it has not reached the required temperature within the pre-set time. This would indicate a problem with steam supply, inlet or drain valve operation. F014 - Jacket over temperature If a jacket is fitted the temperature has exceeded the pre-set alarm temperature. F015 - Jacket under temperature If a jacket is fitted the temperature has fallen below the pre-set operating temperature band. F016 - Water-fill timeout The upper level water probe level has not been reached within the allowed time for filling and the filling operation has been stopped. This function prevents continuous unsupervised operation of the water fill, which could lead to flooding. F017 – NOT USED F018 - Vessel Over temperature This shows that either the HOT 1 or HOT 2 probe has recorded a reading in excess of the Overheat Abort Temperature, indicating a probable low water level. Note – US Spec machines only. In addition to the fault codes displayed, there are hidden message indicators located below the main Temp / Time indicators as follows: SERVICE This means that 500 cycles, or six months have passed since the autoclave was last serviced. The engineer will cancel the message when the autoclave is serviced. This indicator does not signify a machine fault, nor does it affect the Autoclave function. WATER Indicated along with a Low Water fault, F004 O/HEAT Indicates a vessel Over-heat fault along with F003 FAULT Indicates a function fault along with FAULT CODES F000, F002, F005, F006, F007, F008, F009, F010 & F011 LOCK This warning will light when the thermal lock key-switch is in the override position. LOAD Indicates a Load Sense Probe fault along with F001 CANCELLING FAULT MESSAGES Firstly, correcting the source of the original fault Single Program machines: Locate the manual reset button to the rear of the machine / control panel. Press and hold for 3-5 seconds. The Autoclave will reboot and re-set any error messages displayed. Multi-program machines: Insert the SETTING key into the Autoclave. Turn the key-switch to position 3, back to position 1, then back to position 3. If 2 or more faults occur at the same time, the one with the highest priority is displayed. (F000 is the highest priority and F012 is the lowest.) If a higher priority fault is cleared it will be replaced by the next active fault, unless this too is cleared by the same action.