OL 770-DMS
Display Measurement System
The OL 770-DMS offers a complete solution for modern display measurement
requirements, giving accurate color, luminance and spectral information instantly at the
click of a button.
The software works with you, transferring the results or graphs to your own customized
report or spreadsheet.
The unique design ensures results are obtained in milliseconds, saving time and money.
The OL 610 CCD Imaging Telescope is available in 0.5° and 1° fields of view. Custom
FOVs are available upon request.
A clear picture of the measurement area and environment is provided and may be stored
with the data if desired.
Compatible with Westar’s family of FPM display measurement systems.
Bulletin 94 / Rev. 2-04
OL 770-DMS
Display Measurement System
Software Features:
•
•
•
•
•
Real-time graphics utility
Value Monitor with pass/fail, visual/audible alarms
ActiveX™ Control SDK for software development
Excel and Word compatible
Display, log, and store resultant data
•
•
•
•
•
Chromaticity calculations including x,y,z, u,v, u’, v’, etc.
Comparison cursors
High resolution CIE plots
Accumulation graphs for time studies
Intensity profiles
Measurement utilizing:
Value Monitor, which indicates Pass/Fail
of key measurement parameters with
visible and/or audible alarms
and
CIE Chromaticity Diagram indicating all
measurements over time.
OL 770 SPECIFICATIONS
Wavelength Range
OL 770-VIS-DMS
OL 770-VIS/UV-DMS
OL 770-VIS/NIR-DMS
380 to 780 nm
200 to 780 nm
380 to 1100 nm
± 0.75 nm
± 1.0 nm
± 0.5 nm
2 nm
3.5 nm
5.0 nm
0.4 nm
0.6 nm
0.75 nm
100 micron (standard), 50 & 200 micron (optional)
140 mm
fiber optic
f/2
Wavelength Accuracy
Optical Bandwidth (with 100 micron slit)
Spectral Resolution
Slits (user interchangeable)
Optical Focal Length
Optical Input
Optical Aperture
Operating Temperature
0 to 40°C
0 to 90% non-condensing
TE-cooled back-thinned 2-dimensional CCD array
Operating Humidity
Detector Technology
Detector Cooling Temperature
-10°C
> 90% at 650 nm
Quantum Efficiency
± 0.002 x, y
Chromaticity Accuracy
Chromaticity Repeatability (temp. stabilized blue LED)
Stray Light (tungsten source)
Integration Time
A/D Resolution
A/D Rate
Power Input
Interface
Dimensions
Weight
± 0.00015 x, ± 0.0002 y
-4
-4
-4
< 2.5E
< 2.5E at > 300 nm
<2.5E
20 ms – 60 s
16 bits
250 kHz
115/230 Vac
USB, RS232
W 7.25” (18.4 cm), H 13.25” (33.5 cm), D 13.0” (33.0 cm)
22.5 lbs. (10.2 kg)
OL 610 SPECIFICATIONS
FOV
Imaging
PC Interface
Optical Interface
*Wavelength Range
Dimensions
Weight
0.5° or 1.0° (user specified)
Internal CCD camera
USB
Single core fiber with SMA connector
360 to 1100 nm
W 3.00” (7.6 cm), H 4.00” (10.2 cm), D 5.25” (13.3 cm)
2.5 lbs. (1.1 kg)
Bulletin 94 / Rev. 2-04
th
Optronic Laboratories, Inc. 4632 36 Street, Orlando, FL 32811
Tel: 1 407 422 3171 Fax: 1 407 648 5412 Email: info@olinet.com