Ohaus Corporation 29 Hanover Road - - Florh~m P~rk N.! , .-- DHAUS" 07932-0900 ~ ,/ ~~=~~~~~a Ohaus CompactScales ModelsCS-200,CS-2000and CS-5000 -OPERATING MANUALThe CS-200, CS-2000 and CS-5000 only differ in weighing capacity. Unless otherwise specified, the content of this manual applies to both models. Calibration weights are not provided with the scale. For best results, use the product in the specified optimum operating temperature. For best accuracy, perform your mass measurements in clean environments, free of electrostatic and magnetic fields, dust and dirt, moisture, vibration and air currents. Use your scale with care, gently loading items to be weighed onto the center of the pan. Avoiding rough treatment will aid in the life of your compact scale. Battery Installation Three AA size alkaline batteries are provided. . . . Remove battery cover on the bottom of the scale and place them into the compartment as indicated. Do not use excessive force or press the weighing pan. Re-insert the battery cover. AC Adapter Operation AC power adapters are available as optional accessories. . . Connect the AC adapter to the power input jack at the rear of the scale. Plug the AC adapter into a properly grounded wall outlet. General Product Features . Power Up Test: When the scale is turned on, all display segments will appear for approximately 3 seconds before resetting to zero Stable Reading Indication: On the display, the'" indicator will appear in the lower left corner of the display when a stable reading has been reached. Overload: If the applied load exceeds the capacity of the scale, an "E" will appear on the display and the load should be removed immediately. The scale will return to normal operation. Negative Value: When a load is removed from the scale, any zeroed value will be displayed as a negative number. To return to normal operation, the zeroed value can be canceled by pressing "On/Off/Zero". Zero Function: Values can progressively sample. By pressing the "On/Off/Zero" be added to a key, the scale display returns to zero. After adding additional mass, press "On/Off/Zero" to zero the scale again. Additional mass may be added up to the capacity of the scale. Low Battery Indication: The display will show "Lo" when the batteries are weak and need to be replaced. Auto Shut-Off: To extend battery life, the scale will automatically turn off after approximately five minutes if no active weighing is occurring. This feature is active with battery operation only. Two Button Keypad On/Off/Zero: Pressing this key turns on the scale. This same key operates the zero feature when the scale in on. Press and hold this key for three seconds to turn the scale off. Gal/Unit: Press this key briefly to change the weighing unit. The following weight units may be selected: grams (g), or pound-ounces (Ib-oz). Press and hold this key to begin the calibration procedure. Calibration For best results, calibrate the scale at regular intervals. This is especially important if the scale is in use for prolonged periods. 1. Allow the scale to warm up and stabilize for two minutes before commencing the calibration procedure. 2. Press and hold the "Cal/Unit" key until the calibration weight flashes on the display. 3. With the calibration weight value flashing, press the "On/Off/Zero" key, and the display will read "CAL 0." After pausing for at least two seconds, place the correct calibration weight gently on the center of the weighing pan. 4. The display will momentarily show "CAL F" and then return to active weighing. Calibration is complete when the display correctly shows the calibration weight, and the display has stabilized. 5. Remove the calibration weight and turn the scale off. Now turn on the calibrated scale. Note: If "CAL E" appears on the display instead of "CAL F", it indicates that a calibration error has occurred and the process should be repeated. The message "CAL E" will appear if the calibration steps are not followed in the correct order or if an incorrect calibration weight was applied. Accessories Supplied . Three AA size alkaline batteries Optional Accessories AC Power Adapters: 120VUSPlug 230V Euro Plug 240V UK Plug 240V Australian Plug Ohaus Part No. 80500457 80500458 80500459 80500460 Calibration Masses: 200g for CS-200 1000g for CS-2000 2000g for CS-5000 51025-16 51016-16 51026-16 1050 ml Bowl 80850075 Stainless steel weighing pan cover 80850076 Specifications Capacityx Readability Linearity Weighing units Calibration Stabilization time Maximum overload CS-200 CS-2000 2000X 19 200XO.1g 0 Ibs: 4lbs: 7.050zXO.O1oz 6.55ozX0,05oz :to. 19 :t1g :t2g g, Ib,oz Digital calibration from keypad 3 seconds 150% of capacity Overload indication IE] Underload indication Negative value Autoshut-off Optimum operating temperature WeighpansizeLxW (in/cm) Powerrequirements Dimensions L x Wx H (in/cm) NetWeight(Ib/kg) CS-5000 5000X29 111bs: O.4ozXO.1oz 5 minutes no activity 64.-7TF/18.-25.C 5.75 x 5.25/14.5 x 13.3 3-AA alkaline batteries 8.00 x 5.50 x 1.50/20.5 x 14.0 x 3.8 1.0/0.47