MASH 210, 311 T.Elsarnagawy A pH meter is an electronic instrument used to measure the pH (acidity or alkalinity) of a liquid A typical pH meter consists of a special measuring probe (a glass electrode) connected to an electronic meter that measures and displays the pH reading The pH term translates the values of the hydrogen ion concentrationinto numbers between 0 and 14. The lower-case letter "p" in pH stands for the negative common (base ten) logarithm, while the upper-case letter "H" stands for the element hydrogen. consequently, pH is a logarithmic measurement of the number of moles of hydrogen ions (H+) per liter of solution pH = log10 1 + + (H ) = −log10 (H ) Specially-prepared electrode are designed to allow hydrogen ions in the solution to migrate through a selective barrier producing a measurable potential (voltage) difference proportional to the solution‘s pH a measuring electrode must be constructed of special glass to screen out hydrogen ions from all the other ions floating around in the solution The measuring electrode, develops a potential (voltage) directly related to the hydrogen ion concentration of the solution ▪ Note the thin lithium-doped glass membrane across which the pH voltage is generated a reference electrode The reference electrode provides a stable potential against which the measuring electrode can be compared The porous junction shown at the bottom of the electrode is where the potassium chloride buffer and process liquid interface with each other a battery where the positive terminal is the measuring electrode and the negative terminal is the reference electrode. a temperature sensor Output of the measuring electrode changes with temperature (even though the process remains at a constant pH), so a temperature sensor is necessary to correct for this change in output. This is done in the analyser or transmitter software The preamplifier is a signal-conditioning device It takes the high-impedance pH electrode signal and changes it into a low impedance signal which the analyzer or transmitter can accept. The preamplifier also strengthens and stabilizes the signal, making it less susceptible to electrical noise. The sensor's electrical signal is then displayed The analyzer or transmitter has a man machine interface for calibrating the sensor and configuring outputs and alarms Two electrodes generate a voltage directly proportional to the pH of the solution At a pH of 7 (neutral) the electrodes will produce 0 volts between them At a low pH (acid) voltage will be developed of one polarity At a high pH (alkaline) a voltage will be developed of the opposite polarity Most often used pH electrodes are glass electrodes. Typical model is made of glass tube ended with small glass bubble. Inside of the electrode is usually filled with buffered solution of chlorides in which silver wire covered with silver chloride is immersed. pH of internal solution varies - for example it can be 1.0 (0.1M HCl) or 7.0 (different buffers used by different producers). A typical modern pH probe is a combination electrode, which combines both the glass and reference electrodes into one body. It consists of 1. 2. 3. 4. 5. 6. 7. a sensing part of electrode, a bulb made from specific glass sometimes electrode contain small amount of AgCl precipitate inside the glass electrode internal solution, usually 0.1M HCl for pH electrodes internal electrode, usually silver chloride electrode or calomel electrode body of electrode, made from non-conductive glass or plastics. reference electrode, usually the same type as 4 Junction , because we need to measure difference of potentials between sides of glass in the glass electrode. the junction closes the circuit through the solutions - internal and external - and the pH meter. It is usually made from ceramics or capillary with asbestos or quartz fiber The reference electrode potential does not change with the changing hydrogen ion concentration. A solution in the reference electrode also makes contact with the sample solution and the measuring electrode through a junction, completing the circuit The circuit of a simple pH meter usually consists of operational amplifiers in an inverting configuration, with a total voltage gain of about -17. The inverting amplifier converts the small voltage produced by the probe (+0.059 volt/pH) into pH units, which are then offset by seven volts to give a reading on the pH scale. For example: At neutral pH (pH 7) the voltage at the probe's output is 0 volts. 0 * 17 + 7 = 7. At basic pH, the voltage at the probe's output ranges from +0 to +0.41 volts (7 * 0.059 = 0.41). ▪ So for a sample of pH 10 (3 pH units above neutral), 3 * 0.059 = 0.18 volts), the output of the meter's amplifier is 0.18 * 17 + 7 = 10. At acid pH, the voltage at the probe's output ranges from - 0.41 volts to -0. ▪ So for a sample of pH 4 (3 pH units below neutral), -3 * 0.059 = -0.18 volts, the output of the meter's amplifier is -0.18 * 17 + 7 = 4. The two basic adjustments performed at calibration set the gain and offset of the inverting amplifier The calibration process correlates the voltage produced by the probe (approximately 0.06 volts / pH unit) with the pH scale. Calibration should be done before each measurement For normal use: calibration at the beginning of each day The pH meter has three controls 1st control (calibrate): to set the meter reading equal to the value of the first standard buffer 2nd control (slope): which is used to adjust the meter reading to the value of the second buffer. 3rd control allows the temperature to be set Calibration should be performed with at least two standard buffer solutions that span the range of pH values to be measured. Place the probe in a sample of pH 7.00 calibration buffer. Adjust the meter for a pH of 7.00. Rinse the probe with distilled water and then gently blot the tip with a tissue to prevent solution carryover. Immerse the probe in pH 4.00 buffer and adjust the pH meter so it gives a reading of 4.00. Rinse the probe and blot, repeating steps 1 to 3 until readings are stable and no further adjustments are necessary. Usually a couple of repetitions are sufficient. After each single measurement, the probe is washed with distilled water or deionized water to remove any traces of the solution being measured, blotted with a clean tissue to absorb any remaining water which could dilute the sample and thus alter the reading, and then quickly immersed in another solution About once a month, the probe may be cleaned using pH- electrode cleaning solution; generally a 0.1 M solution of hydrochloric acid (HCl) is used, having a pH of about one. Storing: The probe tip must be kept wet at all times and should be stored in buffer solution when not in use pH is a representation of hydrogen ion activity in a liquid. It is the negative logarithm of the amount of hydrogen ions (in moles) per liter of liquid. The basic pH scale extends from 0 (strong acid) to 7 (neutral, pure water) to 14 (strong alkaline). pH can be measured by measuring the voltage produced between two special electrodes immersed in the liquid solution. One electrode, made of a special glass, is called the measurement electrode. It's job it to generate a small voltage proportional to pH (ideally 59.16 mV per pH unit). The other electrode (called the reference electrode) uses a porous junction between the measured liquid and a stable, neutral pH buffer solution (usually potassium chloride) to create a zerovoltage electrical connection to the liquid http://www.ph-meter.info/pH-electrodeconstruction