CONDUCTIVITY AND RESISTIVITY

-

-

-

+

-

Current

Source

+

+

+

+

+

+

+

+

Conductivity measures the ability of a solution to conduct an electric current between two electrodes. In solution, the current flows by ion transport.

Therefore, with an increasing amount of ions present in the liquid, the liquid will have a higher conductivity. If the number of ions in the liquid is very small, the solution will be “resistive” to current flow. AC current is used to prevent complete ion migration to the two electrodes.

CONDUCTIVITY/RESISTIVITY

MEASUREMENT

Conductance = 1/Resistance

Conductivity: mho = Siemen

Normal unit of measurement is:

1 micromho (

µ mho) = 1 microSiemen (

µ

S)

1 millimho (mmho) = 1 milliSiemen (mS)

= 1,000

µ

S

Resistivity : ohm (

)

Normal unit of measurement is: megohm (M

) = 1,000,000

Example - Conductivity to Resistivity

Conversion:

Conductivity = 20

µ

S

= 20 x 10 -6 S

= 2 x 10

= 2 x 10

=

[

-5

-5

S

1 ohm mho

______

-5 ]

Resistivity = 1/conductivity

-1

=

[ ______

-5

1 ohm

]

= 1 ohm

2 x 10 -5

= 1 ⁄

2 x 10

= .5 x 10

= 5 x 10 4

5

5 ohm ohm ohm

= 50,000 ohm

CONDUCTIVITY/RESISTIVITY

CONVERSIONS

Conductivity Resistivity Dissolved

(Micromhos-cm) (Ohms-cm) Solids (ppm)

.

056

.084

.167

1.00

2.50

20.0

200

2000

20,000

18,000,000

12,000,000

6,000,000

1,000,000

400,000

50,000

5,000

500

50

.0277

.0417

.0833

.500

1.25

10.0

100

1,000

10,000

Note: ppm x 2 = Conductivity

Current

Source

If d = 1 cm, probe constant = 1.0

PROBE CONSTANTS

Probe constant defines the volume between electrodes.

Solutions with an extremely high conductivity require a sensor with a probe constant greater than 1.0.

d d d

Solutions with extremely low conductivity require a sensor with a probe constant less than 1.0.

The greater the distance between the electrodes, the smaller the current signal.

CONDUCTIVITY AND RESISTIVITY

(NaCI AND CaCO

3

SOLUTIONS AT 25°C (77°F)) ppm as CaCO

3

1700

1275

850

425

170

127.5

85.0

42.5

17.0

12.7

8.5

4.25

1.70

1.27

0.85

0.42

0.17

0.13

0.085

0.042

0.017

0.012

0.008

0.004

0.002

0.001

None ppm

NaCI

2000

1500

1000

500

200

150

100

50

20

15

10

5.0

2.0

1.5

1.00

0.50

0.20

0.15

0.10

0.05

0.02

0.015

0.010

0.005

0.002

0.001

None

Conductivity micromho/cm

3860

2930

1990

1020

415

315

210

105

42.7

32.1

21.4

10.8

4.35

3.28

2.21

1.13

0.49

0.38

0.27

0.16

0.098

0.087

0.076

0.066

0.059

0.057

0.005

Instrumentation Technical Services - ITS

20 Hagerty Bld. - Suite 1 - West Chester PA, 19380 - (610)436-9703

Resistivity megohm/cm

0.00026

2.65

3.70

6.15

10.2

11.5

13.1

15.2

16.9

17.6

18.3

0.00034

0.00050

0.00099

0.0024

0.0032

0.0048

0.0095

0.023

0.031

0.047

0.093

0.23

0.30

0.45

0.88

2.05