2141-375 Measurement and Instrumentation Analog Electrical Devices and Measurements Analog Devices: Current Measurements Force on a conductor I A conductor is placed in a uniform magnetic field B T, at an angle of θ. The current flow in the conductor is I A. Force exerted on the conductor can be calculated from r r r F = IL × B θ F F = ILB sinθ B Analog Devices: Current Measurements Force on a conductor With no current flowing through the conductor, the spring will be at its unstretched length. As current flows through the conductor, the spring will stretch and developed force required to balance the electromagnetic force. kx = ILB Fs = kx I I k = spring constant, x = the total distance moved by the spring and θ is 90o B F = IBL k I= x BL D’Arsonval or PMMC Instrument Important parts of PMMC Instrument •Permanent magnet with two softiron poles • Moving coil • Controlling or restoring spring PMMC = Permanent Magnet Moving Coil Torque Equation and Scale When a current I flows through a one-turn coil in a magnetic field , a force exerted on each side of the coil F = IBL F = IBL F = IBL D L = the length of coil perpendicular to the paper Since the force acts on each side of the coil, the total force for a coil of N turns is F = NIBL The force on each side acts at a coil diameter D, producing a deflecting torque TD = NIBLD Torque Equation and Scale The controlling torque exerted by the spiral springs is proportional to the angle of deflection of the pointer: TC = Kθ Where K = the spring constant. For a given deflection, the controlling and deflecting torques are equal BLIND = Kθ Since all quantities except θ and I are constant for any given instrument, the deflection angle is θ = CI Therefore the pointer deflection is always proportional to the coil current. Consequently, the scale of the instrument is linear. Galvanometer • Galvanometer is essentially a PMMC instrument designed to be sensitive to extremely current levels. • The simplest galvanometer is a very sensitive instrument with the type of center-zero scale, therefore the pointer can be deflected to either right or left of the zero position. • The current sensitivity is stated in µA/mm • Galvanometers are often used to detect zero current or voltage in a circuit rather than to measure the actual level of current or voltage. In this situation, the instrument is referred as a null detector. DC Ammeter • An ammeter is always connected in series with a circuit. PMMC instrument • The internal resistance should be very low Ammeter shunt • The pointer can be deflected by a very small current • Extension of ranges of ammeter can be achieved by connecting a very low shunt resistor Rs Rm Coil resistance Im I = Is + Im Is Vm Rs I Shunt resistance Vm = Vs I m Rm = I s Rs I m Rm Rs = Is I m Rm Rs = I − Im DC Ammeter Example: An ammeter has a PMMC instrument with a coil resistance of Rm = 99 Ω and FSD current of 0.1 mA. Shunt resistance Rs = 1 Ω. Determine the total current passing through the ammeter at (a) FSD, (b) 0.5 FSD, and (c) 0.25 FSD. Known: FSD of Im Rm and Rs Solution: Im (mA) Is (mA) I = Im + Is (mA) FSD 0.5FSD 0.25FSD 0.1 0.05 0.025 9.9 4.95 2.475 10 5 2.5 DC Ammeter Example: A PMMC instrument has FSD of 100 µA and a coil resistance of 1kΩ. Calculate the required shunt resistance value to convert the instrument into an ammeter with (a) FSD = 100 mA and (b) FSD = 1 A. Known: FSD of Im Rm Solution: (a) FSD = 100 mA: Rs = 1.001 Ω (b) FSD = 1 A: Rs = 0.10001 Ω DC Ammeter: Multirange Rm C Rs1 B D Rs2 E Rs3 Rs4 C B A D E A Make-before break switch Multirange ammeter using switch shunts • A make-before-break must be used so that instrument is not left without a shunt in parallel to prevent a large current flow through ammeter. DC Ammeter: Ayrton shunt Rm Im VS Im R1 + R2 I R3 B I Is C Is - A D R1 + R2 + R3 in parallel with Rm Rm Im VS Im R1 + R2 R3 B I Is Is C A D R1 + R2 in parallel with Rm + R3 I An Ayrton shunt used with an ammeter consists of several seriesconnected resistors all connected in parallel with the PMMC instrument. Range change is effected by switch between resistor junctions DC Ammeter Example: A PMMC instrument has a three-resistor Ayrton shunt connected across it to make an ammeter. The resistance values are R1 = 0.05 Ω, R2 = 0.45 Ω, and R3 = 4.5 Ω . The meter has Rm = 1 kΩ and FSD = 50 µA. Calculate the three ranges of the ammeter. Known: FSD of Im Rm R1 R2 and R3 Solution: Rm Im Im R1 + R2 R3 B I Position B C D IFSD (mA) 10.05 100.05 1000.05 VS Is Is C A I D DC Voltmeter • An ammeter is always connected across or parallel with the points in a circuit at which the voltage is to be measured. • The internal resistance should be very high PMMC instrument V = I m Rs + I m Rm Series resistance or “multiplier” Given V = Range V Multiplier resistance Rs V Rs = − Rm Im Coil resistance Rm Im V Range Rs = − Rm Im The reciprocal of full scale current is the voltmeter sensitivity (kΩ Ω/V) The total voltmeter resistance = Sensitivity X Range DC Voltmeter: Multirange Multiplier resistors • Multirange voltmeter using switched multiplier resistors R1 Meter resistance R2 V = I m (Rm + R) R3 Rm Where R can be R1, R2, or R3 V Rm R1 R2 R3 • Multirange voltmeter using seriesconnected multiplier resistor V = I m (Rm + R) Where R can be R1, R1 + R2, or R1 + R2 + R3 V DC Ammeter Example: A PMMC instrument with FSD of 50 µA and a coil resistance of 1700 Ω is to be used as a voltmeter with ranges of 10 V, 50 V, and 100 V. Calculate the required values of multiplier resistor for the circuit (a) and (b) Known: FSD of Im Rm Solution: Multiplier resistors R1 Meter resistance Rm R2 R1 R2 R3 R3 Rm V V (b) (a) R1 R2 R3 R1 R2 R3 198.3 kΩ Ω 998.3 kΩ Ω 1.9983 MΩ Ω 198.3 kΩ Ω 800 kΩ Ω 1 MΩ Ω Ohmmeter: Voltmeter-ammeter method Pro and con: •Simple and theoretical oriented •Requires two meter and calculations •Subject to error: Voltage drop in ammeter (Fig. (a)) Current in voltmeter (Fig. (b)) + + VA A I A + IV + + Vx VS V V - - - Ix I VS - V V - Rx Fig. (b) Fig. (a) V V + VA V = Rx + A Measured Rx: Rmeas = = x I I I Rmeas ≈ Rx if Vx>>VA Therefore this circuit is suitable for measure large resistance Rx Measured Rx: Rmeas = if Ix>>IV V V Rx = = I I x + IV 1 + IV / I x Rmeas ≈ Rx Therefore this circuit is suitable for measure small resistance Ohmmeter: Series Connection •Voltmeter-ammeter method is rarely used in practical applications (mostly used in Laboratory) •Ohmmeter uses only one meter by keeping one parameter constant Example: series ohmmeter Resistance to be measured Nonlinear scale Standard resistance 15k Rx k 45 Vs − R1 − Rm I Basic series ohmmeter 75 A 0µ 10 Rx = 25 0 VS Meter Infinity resistance 0 Rm ∞ R1 Battery 5k 50 Meter Ohmmeter scale Basic series ohmmeter consisting of a PMMC and a series-connected standard resistor (R1). When the ohmmeter terminals are shorted (Rx = 0) meter full scale defection occurs. At half scale defection Rx = R1 + Rm, and at zero defection the terminals are open-circuited. Loading Effect: Voltage Measurement Rth a Linear Circuit Vab V Vth Rm Vab b Undisturbed condition: Rm = ∞ Measured condition: Rm ≠ ∞ General equation: Measurement error: a V Rm b Vab = Vu = Vth Vab = Vm = Vm = Rm Vth Rm + Rth 1 Vu 1 + Rth / Rm Vm − Vu × 100 Vu Rth 1 =− ×100% = − ×100% Rm + Rth 1 + Rm / Rth error = Therefore, in practice, to get the acceptable results, we must have Rm ≥ 10 Rth (error ~ 9%) Loading Effect R1 100kΩ Ω 100kΩ Ω 5V 6.7 V 10 V 10 V R2 100kΩ Ω 5V 3.3 V V 100kΩ Ω Vmeas = 100 // 100 10 V = 3.3 V 100 + 100 // 100 Circuit under measurement Circuit before measurement 100kΩ Ω 100kΩ Ω 100kΩ Ω 6V 5.2 V 10 V 10 V 100kΩ Ω Vmeas = 4V 200 // 100 10 V = 4.0 V 100 + 200 // 100 V 100kΩ Ω 200kΩ Ω Vmeas = 4.8 V V 1000 // 100 10 V = 4.8 V 100 + 1000 // 100 1000kΩ Ω Loading Effect Example Find the voltage reading and % error of each reading obtained with a voltmeter on (i) 5 V range, (ii) 10 V range and (iii) 30 V range, if the instrument has a 20 kΩ/V sensitivity, an accuracy 1% of full scale deflection and the meter is connected across Rb SOLUTION The voltage drop across Rb without the voltmeter connection Rb 5k V= × 50 = 5 V Ra + Rb 45 k + 5 k Vb = Ra 45kΩ Ω On the 5 V range 45 V 50 V Rb 5kΩ Ω 5V Rm = S × range = 20kΩ / V × 5V = 100 kΩ Req = Rm Rb 100 k × 5 k = = 4.76 kΩ Rm + Rb 100 k + 5 k The voltmeter reading is Vb = Req Ra + Req V= 4.76 k 50 = 4.782 V 45 k + 4.76 k Loading Effect Error of the measurement is the combination of the loading effect and the meter error The loading error = 4.782 - 5 = -0. 218 V The meter error = ± 5 x 1 = ± 0.05 V 100 ∴% of error on the 5 V range: = − 0.218 V ± 0.05 V ×100 = ±5.36% 5V Range (V) Vb . (V) Loading error (V) Meter error (V) Total error (V) % error 5 4.78 -0.22 ± 0.05 ± 0.27 ± 5.36 10 4.88 -0.12 ± 0.1 ± 0.22 ± 4.40 30 4.95 -0.05 ± 0.3 ± 0.35 ± 6.10 Loading Effect: Current Measurement a Linear Circuit Vab Rth I A Rm Vth b Undisturbed condition: Rm = 0 Measured condition: Rm ≠ 0 General equation: Measurement error: I = I u = Vth / Rth a Vab I A Rm b I = I m = Vth / (Rth + Rm ) I m = I u / (1 + Rm / Rth ) I m − Iu ×100 Iu Rm 1 =− ×100% = − × 100% Rm + Rth 1 + Rth / Rm error = Therefore, in practice, to get the acceptable results, we must have Rm ≤ Rth /10 (error ~ 9%) AC Voltmeter: PMMC Based Waveform Amplitude Average RMS A 0 A 2 A A A 2 A D W π 2A π A 2 A 0 A 3 A 0 A A D A D +W D A D +W AC Voltmeter: PMMC Based • Basic PMMC instrument is polarized, therefore its terminals must be identified as + and -. • PMMC instrument can not response quite well with the frequency 50 Hz or higher, So the pointer will settle at the average value of the current flowing through the moving coil: average-responding meter. •Using 4 diodes Full-wave Rectifier Voltmeter Multiplier resistors Rs D1 D3 Rm Vrms Vp Vav D2 •On positive cycle, D1 and D4 are forward-biased, while D2 and D3 are reverse-biased •On negative cycle, D2 and D3 are forward-biased, while D1 and D4 are reverse-biased D4 •Actually voltage to be indicated in ac measurements is normally the rms quantity •The scale is calibrated for pure sine with the scale factor of 1.11 (A/√ √2 / 2A/π π) AC Voltmeter: PMMC Based Example: A PMMC instrument has FSD of 100 µA and a coil resistance of 1kΩ is to be employed as an ac voltmeter with FSD = 100 V (rms). Silicon diodes are used in the full-bridge rectifier circuit (a) calculate the multiplier resistance value required, (b) the position of the pointer when the rms input is 75 V and (c) the sensitivity of the voltmeter Known: FSD of Im Rm Solution: Multiplier resistors Rs D1 D3 Rm Vrms Vp Vav D2 D4 (a) Rs = 890.7 kΩ (b) 0.75 of FSD (c) 9 kΩ/V AC Voltmeter: PMMC Based Half-wave Rectifier Voltmeter Rs D1 Vp Rm D2 Vrms RSH Vav •On positive cycle, D1 is forward-biased, while D2 is reverse-biased •On negative cycle, D2 is forward-biased, while D1 is reverse-biased •The shunt resistor RSH is connected to be able to measure the relative large current. •The scale is calibrated for pure sine with the scale factor of 2.22 (A/√ √2 / A/π π) AC Voltmeter: PMMC Based Example: A PMMC instrument has FSD of 50 µA and a coil resistance of 1700 Ω is used in the half-wave rectifier voltmeter. The silicon diode (D1) must have a minimum (peak) forward current of 100 µA. When the measured voltage is 20% of FSD. The voltmeter is to indicate 50 Vrms at full scale Calculate the values of RS and RSH. Known: FSD of Im Rm RS = 139.5 kΩ RSH = 778 Ω Solution: Rs D1 Vp Rm D2 Vrms RSH Vav AC Voltmeter: PMMC Based Example The symmetrical square-wave voltage is applied to an average-responding ac voltmeter with a scale calibrated in terms of the rms value of a sine wave. If the voltmeter is the full-wave rectified configuration. Calculate the error in the meter indication. Neglect all voltage drop in all diodes. E Solution 11% Em t T Bridge Circuit Bridge Circuit is a null method, operates on the principle of comparison. That is a known (standard) value is adjusted until it is equal to the unknown value. Bridge Circuit AC Bridge DC Bridge (Resistance) Wheatstone Bridge Kelvin Bridge Megaohm Bridge Inductance Capacitance Maxwell Bridge Hay Bridge Owen Bridge Etc. Schering Bridge Frequency Wien Bridge Wheatstone Bridge and Balance Condition The standard resistor R3 can be adjusted to null or balance the circuit. Balance condition: A R2 R1 I1 V No potential difference across the galvanometer (there is no current through the galvanometer) I2 D B I4 I3 R3 R4 C Under this condition: VAD = VAB I1R1 = I 2 R2 And also VDC = VBC I3 R3 = I 4 R4 where I1, I2, I3, and I4 are current in resistance arms respectively, since I1 = I3 and I2 = I4 R1 R2 or = R3 R4 R2 Rx = R4 = R3 R1 Example 1Ω 1Ω 1Ω 1Ω 12 V 12 V 1Ω 2Ω 1Ω (a) Equal resistance 1Ω 2Ω (b) Proportional resistance 1Ω 10 Ω 10 Ω 12 V 12 V 2Ω 20 Ω (c) Proportional resistance 2Ω 10 Ω (d) 2-Volt unbalance Sensitivity of Galvanometer A galvanometer is use to detect an unbalance condition in Wheatstone bridge. Its sensitivity is governed by: Current sensitivity (currents per unit defection) and internal resistance. consider a bridge circuit under a small unbalance condition, and apply circuit analysis to solve the current through galvanometer Thévenin Equivalent Circuit Thévenin Voltage (VTH) A I1 VS VCD = VAC − VAD = I1 R1 − I 2 R2 I2 R1 C R2 G R3 where I1 = D R4 B Therefore V V and I 2 = R1 + R3 R2 + R4 R1 R2 VTH = VCD = V − R + R R + R 3 2 4 1 Sensitivity of Galvanometer (continued) Thévenin Resistance (RTH) R1 C R2 A R3 D R4 RTH = R1 // R3 + R2 // R4 B Completed Circuit RTH C Ig= G VTH VTH RTH+Rg Ig = VTH RTH + Rg D where Ig = the galvanometer current Rg = the galvanometer resistance Example 1 Figure below show the schematic diagram of a Wheatstone bridge with values of the bridge elements. The battery voltage is 5 V and its internal resistance negligible. The galvanometer has a current sensitivity of 10 mm/µA and an internal resistance of 100 Ω. Calculate the deflection of the galvanometer caused by the 5-Ω unbalance in arm BC SOLUTION The bridge circuit is in the small unbalance condition since the value of resistance in arm BC is 2,005 Ω. Thévenin Voltage (VTH) A 1000 Ω 100 Ω R1 5V G D R3 1000 100 VTH = V AD − VAC = 5 V × − 100 + 200 1000 + 2005 ≈ 2.77 mV R2 C R4 2005 Ω 200 Ω B Thévenin Resistance (RTH) (a) 100 Ω C RTH = 100 // 200 + 1000 // 2005 = 734 Ω 1000 Ω A 200 Ω D 2005 Ω B Ig = (b) RTH= 734 Ω The galvanometer current C VTH 2.77 mV = = 3.32 µ A RTH + Rg 734 Ω + 100 Ω Ig=3.34 µA VTH 2.77 mV G D (c) Rg= 100 Ω Galvanometer deflection d = 3.32 µ A × 10 mm = 33.2 mm µA Example 2 The galvanometer in the previous example is replaced by one with an internal resistance of 500 Ω and a current sensitivity of 1mm/µA. Assuming that a deflection of 1 mm can be observed on the galvanometer scale, determine if this new galvanometer is capable of detecting the 5-Ω unbalance in arm BC Example 3 If all resistances in the Example 1 increase by 10 times, and we use the galvanometer in the Example 2. Assuming that a deflection of 1 mm can be observed on the galvanometer scale, determine if this new setting can be detected (the 50-Ω unbalance in arm BC) Deflection Method Consider a bridge circuit which have identical resistors, R in three arms, and the last arm has the resistance of R +∆R. if ∆R/R <<1 Please correct A R R V Thévenin Voltage (VTH) C V D VTH = VCD = V R+∆ ∆R R B ∆R / R 4 + 2 ∆R / R Thévenin Resistance (RTH) Small unbalance occur by the external RTH ≈ R environment In an unbalanced condition, the magnitude of the current or voltage drop for the meter or galvanometer portion of a bridge circuit is a direct indication of the change in resistance in one arm. This kind of bridge circuit can be found in sensor applications, where the resistance in one arm is sensitive to a physical quantity such as pressure, temperature, strain etc. 5 kΩ Ω 5 kΩ Ω 6V Rv Output signal 5 kΩ Ω R v (kΩ ) Example Circuit in Figure (a) below consists of a resistor Rv which is sensitive to the temperature change. The plot of R VS Temp. is also shown in Figure (b). Find (a) the temperature at which the bridge is balance and (b) The output signal at Temperature of 60oC. 6 5 4 3 2 1 0 4.5 kΩ 0 20 40 80 Temp (oC) (b) (a) 60 100 120 AC Bridge: Balance Condition B Z2 Z1 I1 V I2 C D A all four arms are considered as impedance (frequency dependent components) The detector is an ac responding device: headphone, ac meter Source: an ac voltage at desired frequency Z1, Z2, Z3 and Z4 are the impedance of bridge arms Z3 Z4 At balance point: D I1 = General Form of the ac Bridge Complex Form: Polar Form: Z1Z4 ( ∠θ1 + ∠θ 4 ) =Z2 Z3 ( ∠θ 2 + ∠θ 3 ) EBA = EBC or I1Z1 = I 2 Z 2 V V and I 2 = Z1 + Z 3 Z2 + Z4 Z1 Z 4 = Z 2 Z 3 Magnitude balance: Phase balance: Z1Z4 =Z2 Z3 ∠θ1 + ∠θ 4 =∠θ 2 + ∠θ 3 Example The impedance of the basic ac bridge are given as follows: Z1 = 100 Ω ∠80o (inductive impedance) Z3 = 400 ∠30o Ω (inductive impedance) Z 2 = 250 Ω (pure resistance) Z 4 = unknown Determine the constants of the unknown arm. Example an ac bridge is in balance with the following constants: arm AB, R = 200 Ω in series with L = 15.9 mH R; arm BC, R = 300 Ω in series with C = 0.265 µF; arm CD, unknown; arm DA, = 450 Ω. The oscillator frequency is 1 kHz. Find the constants of arm CD. B Z1 I1 V Z1 = R + jω L = 200 + j100 Ω I2 C D A Z3 SOLUTION Z2 Z4 Z 2 = R + 1/ jω C = 300 − j 600 Ω Z3 = R = 450 Ω Z 4 = unknown D The general equation for bridge balance states that Z1 Z 4 = Z 2 Z 3 Operational Amplifier: Op Amp VCC(+) Inverting Input _ Non-inverting Input + I1 _ I2 Output + VV+ VEE(-) (a) Electrical Symbol for the op amp (b) Minimum connections to an op amp Ideal Op Amp Rules: 1. No current flows in to either input terminal 2. There is no voltage difference between the two input terminals Rule 1: I1 = I2 = 0; R+/- = ∞ Rule 2: V+ = V-; Virtually shorted Vout Inverting Amplifier Rf KCL R1 Use KCL at point A and apply Rule 1: (no current flows into the inverting input) v A − vin v A − vout + =0 R1 Rf _ A + + vout - vin Rearrange 1 1 vin vout vA + − + =0 R 1 R f R1 R f Apply Rule 2: (no voltage difference between inverting and non-inverting inputs) Since V+ at zero volts, therefore V- is also at zero volts too. vin vout + =0 R1 R f Rf vout =− vin R1 vA = 0 Inverting Amplifier: another approach No current flows into op amp vin Rf i From Rule 2: we know that V- = V+ = 0, and therefore 0 vin i= R1 − R1 _ i + −vin + iR1 − V = 0 Since there is no current into op amp (Rule 1) + vout - vout = −iR f −V − + iR f + vout = 0 Combine the results, we get mV Rf vout =− vin R1 vout 60 40 20 Given vin = 5sin3t, R1=4.7 kΩ and Rf =47 kΩ vout = -10vin = -50 sin 3t 0 mV -20 -40 -60 vin 1 2 3 4 5 6 time Non-inverting Amplifier Rf KCL R1 Use KCL at point A and apply Rule 1: v A v A − vout + =0 R1 Rf _ A + + vin vin = v A Apply Rule 2: vout - Rf vout = 1+ vin R1 mV vout 60 Given vin = 5sin3t, R1=4.7 kΩ and Rf =47 kΩ 40 20 vout = 11vin = 55 sin 3t mV -20 -40 -60 vin 1 2 3 4 5 6 time Summing Amplifier: Mathematic Operation i = i1 + i2 + i3 Use KCL and apply Rule 1: i i1 R i2 R i3 v1 v2 R vA v A − v1 v A − v 2 v A − v3 v A − vout + + + =0 R R R Rf Rf Since vA = 0 (Rule 2) _ vB + + vout - vout = − Rf R ( v1 + v2 + v3 ) Sum of v1, v2 and v3 v3 Difference Amplifier: Mathematic Operation Use KCL and apply Rule 1: R4 R1 R2 v1 v2 vA v A − v1 v A − vout + =0 R1 R4 _ vB + R3 Substitute eq. (2) into eq. (1), we get If R1 = R2 = R and R3 = R4 = Rf (1) Since vA = vB (Rule 2) and + vout - R3 v A = vB = v2 R2 + R3 vout R1 + R4 R3 v1 = v − 2 R4 R1 R4 R2 + R3 R1 vout = Rf R ( v2 − v1 ) Difference of v1and v2 (2) Differentiator and Integrator: Mathematic Operation R i vout = − iR C i _ + vin But i=C dvC and dt vout dvin = − RC dt + vout - Differentiator i C R vout = − vC + vc - t 1 But vC (t ) = ∫ idt + vC (0) and C 0 _ i + + vout - vin Integrator vin = vC t vout vin = iR 1 =− vin dt + vC (0) ∫ RC 0