INSTRUCTION MANUAL Digital Multimeter Model 178 SPECIFICATIONS DC VOLTAGE RANGE 2OV 200 v 1200v ACCURACY (12 months) 18°-28°C ±(% rdg + digits) MAXIMUM READING 1200V momentary 1200v 1200v 1200v 0.04% + 1d 1.9999 19.999 0.04% +1Id 0.04% +1Id 0.04% +1Id 199.99 1200.0 Temperature Coefficient (0°-18° and 28°-55°C): Normal Mode Rejection Ratio: G r e a t e r t h a n 6 0 d B a t 50Hz and 60Hz. +(0.006% + 0.2 digit)/°C Common Mode Rejection Ratio (lkfi unbalance): Input Resistance: 10 M Q ± 0 . 1% Settling Time: 1 s e c o n d t o w i t h i n 1 d i g i t o f final MAXIMUM ALLOWABLE INPUT G r e a t e r t h a n 1 2 0 d B a t D C , 50Hz and 60Hz. reading. AC VOLTAGE RANGE 2 v 20 v 200 v 1000 v MAXIMUM READING ACCURACY (12 months) (above 2000 counts) 1 8 ° - 2 8 ° C ; 1 00Hz-1 Ok Hz ±(% rdg + digits) TEMPERATURE COEFFICIENT 0°-I 8° and 28°-55°C ±(% rdg + digits)/°C 45Hz-10kHz I0kHz-20kHz 0.4% + 15d 1.9999 19.999 199.99 1000.0 0.04% 0.01% 0.01% 0.01% 0.3% + 15d 0.3% + 15d 0.3% + 15d Extended Frequency Accuracy: Input (45Hz-100Hz) ±(0.5% + 1 5 d i g i t s ) (10kHz-20kHz) ±(1.0% + 1 5 d i g i t s ) Response: A v e r a g e r e s p o n d i n g c a l i b r a t e d i n r m s of a sinewave. Settling Time: 2 . 5 s e c o n d s t o w i t h i n 1 0 d i g i t s of final reading. + + + + 0.09% 0.03% 0.03% 0.03% 0.5d 0.5d 0.5d 0.5d + 0.5d + 0.5d + 0.5d + 0.5d Impedance: IMfi + 1% shunted by less than 75pF. Maximum Allowable Input Voltage: 1000V rms, 1400V p e a k , 107V·Hz maximum. Common Mode Rejection Ratio (lkfi unbalance): 6 0 d B a t D C , 50Hz a n d 60Hz. RESISTANCE RANGE MAXIMUM READING 2kQ 1.9999 20 kQ 19.999 200 kQ 2000 kQ 20Ma 199.99 1999.9 19.999 ACCURACY (12 months) TEMPERATURE COEFFICIENT 18°-28°C 0 ° - 1 8 ° and 2 8 ° - 5 5 ° C ±(% rdg + digits) ±(% rdg + digits)/°C 0.04% + 2d 0.04% + I d 0.04% + I d 0.04% + I d 0.10% + 1 d M a x i m u m A l l o w a b l e I n p u t : 250V rms sine, 350Vpeak. Maximum Voltage Across Unknown: 2V within range, 0.003% 0.003% 0.003% 0.003% 0.02 % + + + + + 0.2d 0.2d 0.2d 0.2d 0.2d NOMINAL APPLIED CURRENT 1mA 100~ A 10~ A 1P A 0.1~ A Settling Time: 1 s e c o n d t o w i t h i n 1 d i g i t o f f i n a l r e a d i n g e x c e p t 2 s e c o n d s o n t h e 20Mfi range. 5V open circuit. GENERAL POWER: DISPLAY: Five 0.5” LED digits, appropriate position and polarity indication. CONVERSION PERIOD: 400 milliseconds. decimal ENVIRONMENT: Operating: 0 ° C Storage: AA to 55°C; 0% to 80% relative humidity up to 40°C. - 2 5 ° C t o +65°C. 105-125 or 210-250 volts (switch select e d ) , 9 0 - I 10V a v a i l a b l e . 50-60Hz, 7 watts. Optional 6 hour battery pack, Model 1788. DIMENSIONS, WEIGHT: 85mm high x 235mm w i d e x 2 7 5 m m d e e p (3-1/2 in. x 9-1/4 in. x 10-3/4 i n . ) . N e t w e i g h t : 1.7kg (3lbs., 13 oz). OVERRANGE INDICATION: Display blinks all zeros above 19999 counts. MAXIMUM COMMON MODE VOLTAGE: 1400V peak. OPERATING INSTRUCTION MANUAL INSTRUCTIONS Digital Multimeter Model 178 2-7 7. OPERATING Refer to Figure 2-3 and operate the INSTRUCTIONS. a. Turn on the power by depressing the ON/OFF b Select l DMM as follows: pushbutton. the function with the DCV, ACV or 52 pushbutton. For ac and dc voltage Select the range by depressing the appropriate pushbutton. c. For resistance measurements there are measurements there are four ranges available. The pushbuttons are interlocked to avoid improper settings. five ranges. d. Connect the source to the INPUT terminals. POLARlTY (NEGATIVE IS INDICATED, POSITIVE IS IMPLIED WHEN MINUS (-) DISPLAY IS OFF) DCV ACV ELECT FUNCTION * A D J U S T ZERO AS REQUIRED FOR MEASUREMENT 2-3. FIGURE SELECT RANGE OF DC VOLTAGE Operating AND CONNECT-SOURCE RESISTANCE. SEE SEC. 2-11. Controls. CAUTION MAXIMUM DCV: RATINGS: n (2V): 45OV rms continuous; (20-1200V): 12OOV ACV: (All a: (All Ranges): Ranges): 12OOV peak, for 8 seconds peak. 1OOOV rms; 10 7V·Hz. 25OV rms sine wave or 35OV peak. per minute. 2-8 . a. DC VOLTAGE Turn on Use the Model 178 MEASUREMENT. power and depress the DCV DMM to measure dc volts as follows: pushbutton. Select the desired range from the four ranqes available. The maximum reading Over range is indicated by a flashing 0000 except on the 1OOO-volt range. 19999. b. is CAUTION 1. Do n not exceed Neqative polarity C. t h e m i n u s (-) d i s p l a y i s the maximum is displayed off. Instrument ratings. Positive automatically. Zero the instrument as described in Paragraph 2-11 d the’instrument is used outside the temperature range of weekly ,during normal use. 2-9 . a. AC VOLTAGE MEASUREMENT. damage may polarity occur. is before the first to 28°C, and 18" implied when use, whenever approximately Use the Model 178 DMM to measure ac volts as follows: Turn on power and depress the ACV pushbutton. CAUTION I. A Do not exceed the maximum ratings. Instrument damage may occur. Select the desired range from the four ranges available. The maximum reading is The Overrange is indicated by a flashing 0000 except on the lOOO-volt range. 19999 instrument reads the root mean square value of a sine wave with a frequency of 45 to 20 kHz. b The Model 1682 RF Probe (see Paragraph c. with a frequency of 20kHz to 100MHz. 2-10. RESISTANCE Turn on (R) power MEASUREMENT. and 2-12e) should . Use the 178 DMM depress the Sz to be used measure to measure resistance as ac voltages follows: pushbutton. CAUTION 1. Do n not exceed the maximum ratings. instrument damage may occur. b Select the desired range- from the five ranges available. The maximum reading is Overrange is indicated by a flashing 0000. The letter k refers to kilohms, and 19999 M refers to meqohms. c. The HI junctions. d . input terminal is positive and causes Two volts is applied at full range with 5 forward volts conduction maximum under of open semiconductor circuit conditions. Zero the i n s t r u m e n t as d e s c r i b e d in Paragraph 2-11 before the first use whenever theainstrument i s u s e d o u t s i d e t h e t e m p e r a t u r e r a n g e o f 18° to 28°C, and approximately weekly during normal use. ZERO ADJUSTMENT. 2-11. The zero adjustment nulls input offset on the 20, 200 and 1200 volt ranges and on all resistance ranges. Typically, this adjustment need not be performed more often than once a week unless the instrument is operated at ambient t e m p e r a t u r e s o u t s i d e t h e r a n g e o f 18° to 28°C. Zero adjustment may also be used for lead compensation on the Q function. Zero the instrument as follows: a. Turn on the power and select DCV and the 20 range. b. Plug in test leads and short them. A d j u s t t h e z e r o a d j u s t ( p o t R132) f r o m front panel with a small screwdriver to obtain a reading of 0000 or - 0 0 0 0 . the 2-12. ACCESSORIES. A wide range of accessories is available to facilitate DMM, extend its range and adapt it for additional uses. M o d e l 178 of use the