LETTERS TO THE EDITOR reports showing changes in PEP and HR the directional results for PEPI are generally the same as for PEP itself, probably owing to disproportionate effects of the intervention on PEP and HR. Despite these criticisms, which I hope Dr. Lewis and associates will address in detail, this is an excellent paper that should be welcomed by everyone interested in systolic time intervals. I have already made it required reading among our noninvasive staff (as are most of Dr. Lewis's previous contributions). DAVID H. SPODICK, M.D., D.Sc. St. Vincent Hospital Worcester, Massachusetts 01604 References Downloaded from http://circ.ahajournals.org/ by guest on September 29, 2016 1. Lewis RP, Rittgers SE, Forester WR, Boudoulas H: A critical review of systolic time intervals. Circulation 56: 146, 1977 2. Spodick DH, Kumar S, Flessas A, Sriratanaban AD, Sosler BS, Klints RV, Muench HP: The relationship of paper speed to precision of pulse wave measurements: A multiple observer study of left ventricular ejection time. Aerospace Med 40: 707, 1969 3. Pigott VM, Ryan-Platt R, Elber E, Spodick DH: The influence of recording speed on apexcardiographic timing: a multi-observer study of precision and performance utilizing randomized traces in multiple subjects. Cardiology 57: 232, 1972 4. Spodick DH, Ball HG, Pigott VM: Effects of registration speeds on the precision of time-based polycardiographic measurements. (abstr) Proc 25th ACEMB, Oct 1972, p 453 5. Cokkinos DV, Heimonas ET, Demopoulos, Haralambakis A, Tsartsalis G, Gardikas CD: Influence of heart rate increase on uncorrected preejection period/left ventricular ejection time (PEP/LVET) ratio in normal individuals. Br Heart J 38: 683, 1976 6. Talley RC, Meyer JF, McNay JL: Evaluation of the pre-ejection period as an estimate of myocardial contractility in dogs. Am J Cardiol 27: 384, 1971 7. Knapp E, Aigner A, Baumgartl P, Raas E: Zur frage der frequenzabhangigkeit systolischer kreislaufzeiten und daraus berechneter indizes. Kreislaufforsch 61: 492, 1972 8. Higgins CB, Vatner SF, Franklin D, Braunwald E: Extent of regulation of the heart's contractile state in the conscious dog by alteration in the frequency of contraction. J Clin Invest 52: 1187, 1973 9. Harris WS: In Cardiac Mechanics. London, Wiley, 1974, p 238 10. Lance VQ, Spodick DH: Constant load versus rate targeted exercise: Responses of systolic intervals. J Appl Physiol 38: 794, 1975 The author replies: To the Editor: We wish to thank Dr. Spodick for his comments on our manuscript. Apparently he is in agreement with our results which indicate 100 mm/sec is the optimum paper speed. He raised the question of observer bias. We indeed conducted our study as he suggested ... "completely coded and randomized to reduce observer bias so that observers could never read data from the same subject in sequence except by chance." All observers received individual copies of each taperecorded pulse arbitrarily coded and in random order according to patients and paper recording speeds. Apparently in Dr. Spodick's as yet unpublished study, he found no statistically significant difference for the LVET while we noted such a difference. It should be noted that we used a computer analysis to achieve a standard against which to judge the LVET measurements at each paper speed. This may account for our different results. However, even if the mean values were not statistically different, the greater observer variability at slower speeds dictates that 100 mm/sec is the optimal recording speed. I feel that this debate over paper speed has become somewhat academic since modern technology has resulted in a number of different recording devices with 100 mm/sec capability. The argument over whether to correct the PEP for heart rate is essentially an "empirical vs functional" argument. Until it can be proved that PEP is functionally independent of heart rate, we must rely on carefully collected empirical data. From such studies it is clear that a significant, albeit weak, correlation exists. Naturally, only studies with large numbers of subjects will reveal this relationship. Atrial pacing data cannot be applied to this situation, 201 as this is an unnatural mode of increasing the heart rate (i.e., no increase in adrenergic tone). RICHARD P. LEWIs, M.D. Ohio State University Hospitals Columbus, Ohio 43210 Small vs Large Surface Area Electrodes To the Editor: We were interested in the article by Hughes et al. (Circulation 54: 128, 1976) because the promise of lower long-term thresholds with small-area electrodes has brought them into widespread use. If a hazard exists in their failure to sense, it would seriously jeopardize their use, as competitive pacing is still a highly unattractive and potentially harmful long-term complication. We would agree with the observations of the authors that the combination of a small-area electrode and a low input impedance of the detection device may lead to failure of sensing. However, we would like to assert that small-area electrodes coupled with high impedance detection circuits such as are used in most modern pacemakers do not constitute such a hazard. In other words, the signal provided by a small-area electrode is determined by the combination of the interface impedance of the electrode and the input impedance of the detection device, or pacemaker. Also, a small-area electrode should produce a larger and sharper inherent signal than a very large area (within limits) since the large area will average the signal more than a smaller one. The failure to clearly point out the unique and almost artificial circumstances of their experiments is illustrated in their table 1. The results demonstrate how the 1 KU load, in combination with the electrode impedance, attenuates the signal strength. As pointed out by the authors, the 1 KQ load is much lower than commercially available pulse generators. The effects with actual pacemakers having impedances of 20 Kg will be about twenty times less severe. The true "sensed" electrogram would be represented by an unloaded measurement, but these values are unfortunately not provided. Some other points are: 1) bipolar electrodes do not inherently sense smaller signals than unipolar electrodes. 2) Fibrous tissue does not greatly affect the electrode-tissue impedance - the impedance is mainly at the metal-electrolyte interface. The fibrous tissue does, however, separate the electrode from viable tissue and so reduces the sensed signal. 3) In figure 5, the electrode tip design for optimal sensing is surely offered as an example of principle rather than as an actual electrode. The geometry of the large band sensing electrode is such that it would be very difficult to have it in intimate contact with excitable tissue if used as an endocardial electrode. In summary, small-area electrodes as they are currently used do not constitute a hazard for sensing when the higher interface impedance associated with these electrodes is appropriately used with a higher input impedance of the detecting pacemaker. T. E. CUDDY, M.D. M. B. RABER, P.ENG. University of Manitoba Winnipeg, Manitoba The authors reply: To the Editor: We agree completely with Dr. Cuddy that bipolar electrodes are not inherently inferior to unipolar electrodes. The ability of any type electrode to sense R-wave potentials is based primarily on its surface area, not the mode in which sensing is done.' I hope we did not leave that impression with anyone in our paper. Secondly, we have only limited information of the chronically implanted electrodes' ability to sense. We would, however, expect the sensing ability to be reduced by fibrous tissue ingrowth. Small vs large surface area electrodes. T E Cuddy and M B Raber Downloaded from http://circ.ahajournals.org/ by guest on September 29, 2016 Circulation. 1978;57:201-202 doi: 10.1161/01.CIR.57.1.201 Circulation is published by the American Heart Association, 7272 Greenville Avenue, Dallas, TX 75231 Copyright © 1978 American Heart Association, Inc. All rights reserved. Print ISSN: 0009-7322. Online ISSN: 1524-4539 The online version of this article, along with updated information and services, is located on the World Wide Web at: http://circ.ahajournals.org/content/57/1/201.citation Permissions: Requests for permissions to reproduce figures, tables, or portions of articles originally published in Circulation can be obtained via RightsLink, a service of the Copyright Clearance Center, not the Editorial Office. Once the online version of the published article for which permission is being requested is located, click Request Permissions in the middle column of the Web page under Services. Further information about this process is available in the Permissions and Rights Question and Answer document. Reprints: Information about reprints can be found online at: http://www.lww.com/reprints Subscriptions: Information about subscribing to Circulation is online at: http://circ.ahajournals.org//subscriptions/