Digital Multimeters - Agilent Technologies

advertisement
Discover the next generation of bench/system
DMMs from Agilent
DISPLAY
DMM results in ways
you never have before
MEASURE
with unquestioned
Truevolt confidence
MOVE
to the next generation 34401A
DMM with 100% assurance
SERIES
Digital
Multimeters
► Agilent 34461A
a direct replacement for the
34401A 6½ digit DMM
► Agilent 34460A
a basic entry point into
the 6½ digit class
of Agilent DMMs
Display DMM results in ways you never have before
Easily display, save and document
your measurement results
Easily set up and see your results and get actionable
information faster
•Seeyourresultsclearlyonabright,4.3-inch(109mm),high-resolution,color
graphicaldisplay.Getanswersfasterusingtheintuitive,menu-driveninterface.
Ifyouhavequestionsaboutakeyorfunctionality,pressandholdthekeyto
displaybuilt-inhelponthesubject.
•Customizedisplayoperationtomeetyourneeds.Conigureyourpreferences
andpullthemupautomaticallyatthenextinstrumentstart-up.
•Addcustomizedlabelstodescribetheinstrument’smeasurementyourway.
•Getquickinsightwithgraphicalviewsofyourmeasurementresults:View
readings,long-termtrends(34461Aonly)andmeasurementhistograms
forastatisticalview.
The bright, 4.3” high-resolution monitor is a prominent
feature of Agilent’s new Truevolt multimeter family.
Bar meter mode provides the
number display along with an
analog meter to provide a visual
view of your measurements.
Number mode provides
the traditional “digits” view
of measurements.
Histogram mode gives you a statistical
view of your measurements.
Truevolt
D I S P L A Y
2
Display DMM results in ways you never have before
Save and document your DMM’s data to your PC
or mobile device using your preferred I/O:
•Control,captureandviewyourDMM’sdataonyourPCwithasingleclick
usingtheDigitalMultimeterConnectivityUtilityviaUSB,GPIB,LAN,
orRS-232(foroldergenerationAgilentDMMs).
•Easilyaccesstheilesonyour
DMMusingdrag-and-drop
totransferilesviaUSB–
nosoftwarerequired.
The free DMM Connectivity
Utility helps you get your
job done faster
Capture and export data to your PC with
drag and drop ease
Captureadigitizedrecord,logmeasurementsfor
longerperiodsunderPCcontroloruploaddata
captureddirectlyfromtheinstrument.Exportdata
fromasingleDMM,ormultipleDMMswithtime
alignment,topopulartoolssuchasMicrosoft
Excel,MicrosoftWordandMATLAB.Simplifytest
documentationanddataanalysiswithoutthehassles
ofprogramming.
•ChoosetheI/Othatworksbest
foryou:USB,LAN/LXICore
(optionalon34460A),GPIB
(optionalon34460A&34461A).
•Easilyaccessthefront-panelUSBmemoryconnector.
Test result screens allow you to show
measurement statistics on the display.
Visualize multiple DMM outputs at once
Displaysinglemeasurements,charts,orhistograms
fromasingleinstrumentoruptofourDMMs
simultaneously.Seewhat’shappeningonyour
bench,allononedisplay—tospotcorrelatedtrends
youmightotherwisemiss.
Simplify instrument coniguration
Instrumentcontrolsprovideeasymeasurementsetup
andinstrumentstatemanagement.Reduceset-up
timesandquicklyreconigurefornewtests.
Speed up instrument discovery and connection for a broad range of Agilent DMMs
Withone-clickaccesstoAgilentIOLibrariesSuite,
connecttoDMMsviaUSB,GPIB,LANorRS-232.
Spendyourvaluabletimetestingyourdesignsrather
thansettingup,connecting,andtroubleshooting
software.
Access and control tests on your DMM
remotely on your mobile device
Withacompanionappavailable,DMMMobile
Utility,asafreedownloadfromtheAppleAppStore,
viewandcontrolyourLAN-enabledinstrument
whereveryougoandreceiveemailalertswhen
problemsoccur.Remotelyremedyproblemsonlong
runningteststominimizeprojectdelays.
Download the DMM Connectivity Utility:
Trend chart mode (34461A only)
displays your measurements
over time.
www.agilent.com/find/DMMutilitysoftware
3
Measure with unquestioned Truevolt conidence
Worry about the quality of your design, not the quality
of your measurements
Inarackoronabenchreal-worldsignalsareneverlat.Theyhavesomelevel
ofACsignalridingontopfrompowerlinenoise,otherenvironmentalnoise,or
injectedcurrentfromthemeteritself.Howwellyourmeterdealswiththese
extraneousfactorsandeliminatesthemfromthetruemeasurementmakesabig
differencetoyouraccuracy.Behindthescenes,Agilent’sTruevolttechnology
accountsformeasurementerrorscreatedbythesereal-worldfactorssoyoucan
beconidentinyourmeasurementsanditisonlyavailableonAgilentDMMs.
Truevolttechnologystartswithananalog-to-digitalconverterthatenablesa
patentedmetrology-gradearchitecture.Usingthisarchitecture,Agilentdelivers
agoodbalanceofmeasurementresolution,linearity,accuracy,andspeedata
valueprice,allderivedandguaranteedperISO/IEC17025industrystandards.
Truevolt
MEASUREMENT
4
Measure with unquestioned Truevolt conidence
What Truevolt technology means to you:
You can measure your real-world signals,
not instrument error
Noise and injected current: AgilentTruevoltDMMscontributelessthan
30%oftheinjectedcurrentthanalternatives.Comparedtosomelowercost
alternatives,TruevoltDMMsofferalmost100%lessnoise.
800
You can measure your real-world signals with confidence
AllTruevoltDMMspeciicationsaretestedandguaranteedforcompliance
withISO/IEC17025standardssoyoucanprovetheeffectiveness
ofyourlaborproductionline’squalitymanagementsystem.
Manylower-costDMMsinthisclassdonotcarryaguarantee
oftheirmeasurementspeciications.
700
600
nA
500
400
300
You can take advantage of expanded
measurement functionality
200
100
0
Truevolt
Brand A
Brand B
Brand C
Input bias current:Ideally,nocurrentlowsintothemeasurementterminals
ofyourDMM.Inrealmeasurementsituations,therearealwaysinputcurrents
creatingadditionalmeasurementerrors.TruevoltDMMstakecareofinputbias
current.SomealternativeDMMsoffer20%toininitelypoorerperformance
(somearetoonoisytomeasure).
Compared to the 34401A DMM, Truevolt DMMs offer expanded
current ranges from 100 µA to 10 A. We have also added a temperature measurement function (RTD/PT100, 5 kΩ thermistor).
Additionally, diode measurement capability has been expanded
to allow a larger full-scale voltage to be measured (5 V) to
enable the measurement of more diode types such as LEDs.
60
50
Input
bias
current
too
noisy
to
measure
pA
40
30
20
10
0
Truevolt
Brand A
Brand B
Brand C
Digital AC rms measurements:Formetersinthisclass,onlyAgilentuses
digitaldirectsamplingtechniquestomakeACrmsmeasurements.Thisresults
inatruermscalculationtechniquethatavoidstheslowerresponseofanalog
RMSconvertersusedinallothervendor’s6½digitDMMs.Thisallowsfor
crestfactorsupto10withoutadditionalerrorterms.Thisisaunique,patented
technique–onlyusedbyAgilent.
Crest factor: the ratio of the
peak value to the RMS value
of a waveform
Peak value
RMS value
AC measurement
accuracy is
degraded when
signals have energy
contained in higher
frequencies than a
typical sine wave.
5
Move to the next-generation
34401A DMM with 100% assurance
Migrate with confidence: Everything you
depend on with the 34401A and more
Likemost34401ADMMowners,yourelyonyour
DMMandyoutrusttheanswersitgivesyou.Now,
withtheAgilentTruevolt34461ADMM,youcan
getalloftheadvantagesofthe34401Aandmore.
Nowyoucangetfasteranswersandhaveeven
moreconidenceinyourresults.
Thebestnewsofall?Youcanmigratefromthe
34401Atothe34461Awithoutahassle.Noneed
torewriteyoursoftwareprogramsorspendhours
learningabrand-new,complicatedinterface.
Use your existing programs:The34461A
DMMistheindustry’sonly100%drop-in,SCPIcompatiblereplacementforthe34401ADMM.
OtherDMMsmayclaim34401ASCPIcompatibility,butonlyasubsetofSCPIcommandsare
implemented.
MIGRATION Q&A
QUESTION
ANSWER
Program
compatibility
Willmyexistingprogramsstillwork
ifIswitchtothe34461A?
YES
Measurements
WillIhavethesameperformancesoit
doesn’taffecttheresultsonmyline?
YES
Cost
Willitcostthesametobuy,use,
maintain,andrepair?
YES (and potentially less since
the DMMs now include a 3-year
standard warranty)
Reliability
My34401Aneverbreaks.Arethe
TruevoltDMMsgoingtobeasgood?
YES
Use
Willwebeabletouseiteasily?Quickly?
YES
No long learning curve: TheTruevoltDMMs
weredesignedbythesameteamthatcreatedthe
34401A.Theteamkept34401Ameasurements,
reliabilityandfamiliarityinmindastheycreatedthe
TruevoltfamilyofDMMs.Soyoucanuseitwithout
spendinghourslearninghow.
The34461Arepresentseverythingyouhaveknown
andtrustedwithyourAgilentDMMmeasurements
fordecades–itjustkeepsgettingstronger.
34461A: The industry’s only 100% drop-in,
SCPI-compatible replacement for the 34401A DMM
6
That’s why we can offer
a 3-year standard warranty
Truevolt
M I G R A T I O N
Move to the next-generation
34401A DMM with 100% assurance
WHICH MODEL IS RIGHT FOR YOU?
34460A
34461A
34401A
Resolution
6½digits
6½digits
6½digits
Input terminals
Front
Frontandrear
Frontandrear
1-year DCV accuracy
±(% of reading + % of range)
0.0075+0.0005
0.0035+0.0005
0.0035+0.0005
Measurement speed – 4½ digits
300readings/s
1000readings/s
1000readings/s
DCV, ACV, resistance,
frequency, period, continuity
Sameas34401Abaseline
Sameas34401Abaseline
34401Abaseline
Diode
5V
5V
1V
Current
100µA–3A
100µA–10A
10mA–3A
Temperature
RTD/PT100,thermistor
RTD/PT100,thermistor
N/A
Internal memory
1,000readings
10,000readings
512readings
Graphical display
Number,histogram,barmeter
Number,histogram,barmeter,trend
Number
I/O connectivity
USB(LAN,GPIBoptional)
USB,LAN(GPIBoptional)
GPIB,RS-232
OPTIONS
34460A
34461A
34401A
3446ACCU: Accessory kit for 34460A
Documentation CDs, test leads, USB cable
OPTIONAL
INCLUDED STANDARD
INCLUDED STANDARD
3446GPBU
GPIB user installable interface module
OPTIONAL
OPTIONAL
INCLUDED STANDARD
3446LANU
Enable rear panel LAN/LXI web interface,
external triggering for 34460A
OPTIONAL
INCLUDED STANDARD
N/A
3446SECU
Enable NISPOM and file security
OPTIONAL
OPTIONAL
N/A
Measurements
7
S P E C I F I C AT I O N S
34460A
34460Aaccuracyspeciications:±(%ofreading+%ofrange)1
ThesespeciicationarecomplianttoISO/IEC17025forK=2 Range 2/frequency
24 hour 3
TCAL±1°C
90 day
TCAL±5°C
1 year
TCAL±5°C
2 year
TCAL±5°C
Temperature
coefficient/°C 4
DC voltage
100mV
1V
10V
100V
1000V
0.0040+0.0060
0.0030+0.0009
0.0025+0.0004
0.0030+0.0006
0.0030+0.0006
0.0070+0.0065
0.0060+0.0010
0.0050+0.0005
0.0065+0.0006
0.0065+0.0010
0.0090+0.0065
0.0080+0.0010
0.0075+0.0005
0.0085+0.0006
0.0085+0.0010
0.0115+0.0065
0.0105+0.0010
0.0100+0.0005
0.0110+0.0006
0.0110+0.0010
0.0005+0.0005
0.0005+0.0001
0.0005+0.0001
0.0005+0.0001
0.0005+0.0001
True RMS AC voltage 2, 5, 6
100mV,1V,10V,100V,and750Vranges
3–5Hz
5–10Hz
10Hz–20kHz
20–50kHz
50–100kHz
100–300kHz
1.00+0.02
0.38+0.02
0.07+0.02
0.13+0.04
0.58+0.08
4.00+0.50
1.00+0.03
0.38+0.03
0.08+0.03
0.14+0.05
0.63+0.08
4.00+0.50
1.00+0.03
0.38+0.03
0.09+0.03
0.15+0.05
0.63+0.08
4.00+0.50
1.00+0.03
0.38+0.03
0.10+0.03
0.16+0.05
0.63+0.08
4.00+0.50
0.100+0.003
0.035+0.003
0.005+0.003
0.011+0.005
0.060+0.008
0.200+0.020
Resistance 7
100Ω
1kΩ
10kΩ
100kΩ
1MΩ
10MΩ
100MΩ
Test current
1mA
1mA
100µA
10µA
5µA
500nA
500nA||10MΩ
0.0040+0.0060
0.0030+0.0008
0.0030+0.0005
0.0030+0.0005
0.0030+0.0010
0.015+0.001
0.300+0.010
0.011+0.007
0.011+0.001
0.011+0.001
0.011+0.001
0.011+0.001
0.020+0.001
0.800+0.010
0.014+0.007
0.014+0.001
0.014+0.001
0.014+0.001
0.014+0.001
0.040+0.001
0.800+0.010
0.017+0.007
0.017+0.001
0.017+0.001
0.017+0.001
0.017+0.001
0.060+0.001
0.800+0.010
0.0006+0.0005
0.0006+0.0001
0.0006+0.0001
0.0006+0.0001
0.0010+0.0002
0.0030+0.0004
0.1500+0.0002
DC current
100µA
1mA
10mA
100mA
1A
3A
Burden voltage
<0.011V
<0.11V
<0.05V
<0.5V
<0.7V
<2.0V
0.010+0.020
0.007+0.006
0.007+0.020
0.010+0.004
0.050+0.006
0.180+0.020
0.040+0.025
0.030+0.006
0.030+0.020
0.030+0.005
0.080+0.010
0.200+0.020
0.050+0.025
0.050+0.006
0.050+0.020
0.050+0.005
0.100+0.010
0.200+0.020
0.060+0.025
0.060+0.006
0.060+0.020
0.060+0.005
0.120+0.010
0.230+0.020
0.0020+0.0030
0.0020+0.0005
0.0020+0.0020
0.0020+0.0005
0.0050+0.0010
0.0050+0.0020
8
S P E C I F I C AT I O N S
34460A
Range 2/frequency
24 hour 3
TCAL±1°C
90 day
TCAL±5°C
1 year
TCAL±5°C
2 year
TCAL±5°C
Temperature
coefficient/°C 4
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.015+0.006
0.030+0.006
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.015+0.006
0.030+0.006
0.23+0.04
0.23+0.04
0.23+0.04
0.23+0.04
0.23+0.04
0.23+0.04
0.23+0.04
0.23+0.04
0.015+0.006
0.030+0.006
Continuity
1kΩ
0.002+0.030
0.008+0.030
0.010+0.030
0.012+0.030
0.0010+0.0020
Diode test 10
5V
0.002+0.030
0.008+0.030
0.010+0.030
0.012+0.030
0.0010+0.0020
True RMS AC current 2, 6, 8
100µA,1mA,10mA,and100mAranges
3Hz–5kHz
5–10kHz9
1Arange
3Hz–5kHz
5–10kHz9
3Arange
3Hz–5kHz
5–10kHz9
Burden voltage
<0.011,<0.11,<0.05,<0.5V
<0.7V
<2.0V
DC ratio 11
(normalizedinputaccuracy)+(normalizedreferenceaccuracy)
12
Temperature
PT100(DIN/IEC751)
5kΩthermistor
Probeaccuracy+0.05°C
Probeaccuracy+0.1°C
Frequency: specification ± (% of reading) 13, 14
100mV,1V,10V,100V,and750Vranges15
3–10Hz
10–100Hz
100Hz–1kHz
1–300kHz
Squarewave 16
Additional gate time errors ± ( % of reading )
Frequency
3–40Hz
40–100Hz
100Hz–1kHz
1–300kHz
Squarewave16
0.100
0.030
0.003
0.002
0.001
0.100
0.030
0.010
0.008
0.008
0.100
0.030
0.012
0.012
0.012
1 second
0
0
0
0
0
0.1 second
0.200
0.060
0.020
0.004
0
0.01 second
0.200
0.200
0.200
0.030
0
0.100
0.035
0.017
0.017
0.017
0.0002
0.0002
0.0002
0.0002
0.0002
14
1. ForDC:Speciicationsarefor60-minutewarm-up,apertureof10or100NPLC,
andautozeroon.
ForAC:Speciicationsarefor60-minutewarm-up,slowACilter,sinewave.
2. 20%overrangeonallranges,except1000DCV,750ACV,3AAC,anddiodetest.
3. Relativetocalibrationstandards.
4. Addthisforeach°CoutsideTCAL±5°C.
5. Speciicationsareforsinewaveinput>0.3%ofrangeand>1mVrms.
750ACVrangelimitedto8x10 7Volt–Hz.
6. Low-frequencyperformance:threeiltersettingsareavailable:3Hz,20Hz,200Hz.
Frequenciesgreaterthantheseiltersettingsarespeciiedwithnoadditionalerrors.
7. Speciicationsarefor4–wireohmsfunctionor2–wireohmsusingmathnullforoffset.Without
mathnull,add0.2Ωadditionalerrorin2-wireohmsfunction.
8. Speciicationsareforsinewaveinput>1%ofrangeand>10µAAC.
9. ACcurrentspeciications>5kHzaretypical.
10. Speciicationsareforthevoltagemeasuredattheinputterminals.The1mAtestcurrent
istypical.Variationinthecurrentsourcewillcreatesomevariationinthevoltagedropacross
adiodejunction.
11. Thesespeciicationsarefortypicalperformance.
12. Actualmeasurementrangeandprobeerrorswillbelimitedbytheselectedprobe.
ProbeaccuracyadderincludesallmeasurementandITS-90temperatureconversionerrors
PT100Rosettableto100Ω±5Ωtoremovetheinitialprobeerror.
13. Speciicationsarefor60-minutewarm-upandsinewaveinputunlessstatedotherwise.
Speciicationsarefor1-secondgatetime(7digits).
14. Appliestosineandsquareinputs≥100mV.
For10mVto<100mVinputs,multiply%ofreadingerrorx10.
15. Amplitude10%–120%ofrangeandlessthan750ACV.
16. Squarewaveinputspeciiedfor10Hz–300kHz.
9
S P E C I F I C AT I O N S
34461A
34461Aaccuracyspeciications:±(%ofreading+%ofrange)1
ThesespeciicationarecomplianttoISO/IEC17025forK=2
Range 2/frequency
24 hour 3
TCAL±1°C
90 day
TCAL±5°C
1 year
TCAL±5°C
2 year
TCAL±5°C
Temperature
coefficient/°C 4
DC voltage
100mV
1V
10V
100V
1000V
0.0030+0.0030
0.0020+0.0006
0.0015+0.0004
0.0020+0.0006
0.0020+0.0006
0.0040+0.0035
0.0030+0.0007
0.0020+0.0005
0.0035+0.0006
0.0035+0.0010
0.0050+0.0035
0.0040+0.0007
0.0035+0.0005
0.0045+0.0006
0.0045+0.0010
0.0065+0.0035
0.0055+0.0007
0.0050+0.0005
0.0060+0.0006
0.0060+0.0010
0.0005+0.0005
0.0005+0.0001
0.0005+0.0001
0.0005+0.0001
0.0005+0.0001
True RMS AC voltage 2, 5, 6
100mV,1V,10V,100V,and750Vranges
3–5Hz
5–10Hz
10Hz–20kHz
20–50kHz
50–100kHz
100–300kHz
1.00+0.02
0.35+0.02
0.04+0.02
0.10+0.04
0.55+0.08
4.00+0.50
1.00+0.03
0.35+0.03
0.05+0.03
0.11+0.05
0.60+0.08
4.00+0.50
1.00+0.03
0.35+0.03
0.06+0.03
0.12+0.05
0.60+0.08
4.00+0.50
1.00+0.03
0.35+0.03
0.07+0.03
0.13+0.05
0.60+0.08
4.00+0.50
0.100+0.003
0.035+0.003
0.005+0.003
0.011+0.005
0.060+0.008
0.200+0.020
Resistance 7
100Ω
1kΩ
10kΩ
100kΩ
1MΩ
10MΩ
100MΩ
Test current
1mA
1mA
100µA
10µA
5µA
500nA
500nA||10MΩ
0.0030+0.0030
0.0020+0.0005
0.0020+0.0005
0.0020+0.0005
0.002+0.001
0.015+0.001
0.300+0.010
0.008+0.004
0.008+0.001
0.008+0.001
0.008+0.001
0.008+0.001
0.020+0.001
0.800+0.010
0.010+0.004
0.010+0.001
0.010+0.001
0.010+0.001
0.010+0.001
0.040+0.001
0.800+0.010
0.012+0.004
0.012+0.001
0.012+0.001
0.012+0.001
0.012+0.001
0.060+0.001
0.800+0.010
0.0006+0.0005
0.0006+0.0001
0.0006+0.0001
0.0006+0.0001
0.0010+0.0002
0.0030+0.0004
0.1500+0.0002
DC current
100µA
1mA
10mA
100mA
1A
3A
10A
Burden voltage
<0.011V
<0.11V
<0.05V
<0.5V
<0.7V
<2.0V
<0.5V
0.010+0.020
0.007+0.006
0.007+0.020
0.010+0.004
0.050+0.006
0.180+0.020
0.050+0.010
0.040+0.025
0.030+0.006
0.030+0.020
0.030+0.005
0.080+0.010
0.200+0.020
0.120+0.010
0.050+0.025
0.050+0.006
0.050+0.020
0.050+0.005
0.100+0.010
0.200+0.020
0.120+0.010
0.060+0.025
0.060+0.006
0.060+0.020
0.060+0.005
0.120+0.010
0.230+0.020
0.150+0.010
0.0020+0.0030
0.0020+0.0005
0.0020+0.0020
0.0020+0.0005
0.0050+0.0010
0.0050+0.0020
0.0050+0.0010
10
S P E C I F I C AT I O N S
34461A
Range 2/frequency
24 hour 3
TCAL±1°C
90 day
TCAL±5°C
1 year
TCAL±5°C
2 year
TCAL±5°C
Temperature
coefficient/°C 4
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.015+0.006
0.030+0.006
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.10+0.04
0.015+0.006
0.030+0.006
0.23+0.04
0.23+0.04
0.23+0.04
0.23+0.04
0.23+0.04
0.23+0.04
0.23+0.04
0.23+0.04
0.015+0.006
0.030+0.006
0.15+0.04
0.15+0.04
0.15+0.04
0.15+0.04
0.15+0.04
0.15+0.04
0.15+0.04
0.15+0.04
0.015+0.006
0.030+0.006
Continuity
1kΩ
0.002+0.030
0.008+0.030
0.010+0.030
0.012+0.030
0.0010+0.0020
Diode test 10
5V
0.002+0.030
0.008+0.030
0.010+0.030
0.012+0.030
0.0010+0.0020
True RMS AC current 2, 6, 8
100µA,1mA,10mA,and100mAranges
3Hz–5kHz
5–10kHz9
1Arange
3Hz–5kHz
5–10kHz9
3Arange
3Hz–5kHz
5–10kHz9
10Arange
3Hz–5kHz
5–10kHz9
DC ratio
Burden voltage
<0.011,<0.11,<0.05,<0.5V
<0.7V
<2.0V
<0.5V
11
(normalizedinputaccuracy)+(normalizedreferenceaccuracy)
12
Temperature
PT100(DIN/IEC751)
5kΩthermistor
Probeaccuracy+0.05°C
Probeaccuracy+0.1°C
Frequency: specification ± (% of reading) 13, 14
100mV,1V,10V,100V,and750Vranges15
3–10Hz
10–100Hz
100Hz–1kHz
1–300kHz
Squarewave 16
0.100
0.030
0.003
0.002
0.001
0.100
0.030
0.008
0.006
0.006
0.100
0.030
0.010
0.010
0.010
Additional gate time errors ± ( % of reading )14
Frequency
3–40Hz
40–100Hz
100Hz–1kHz
1–300kHz
Squarewave16
1 second
0
0
0
0
0
0.1 second
0.200
0.060
0.020
0.004
0
0.01 second
0.200
0.200
0.200
0.030
0
1. ForDC:Speciicationsarefor60-minutewarm-up,apertureof10or100NPLC,andautozeroon.
ForAC:Speciicationsarefor60-minutewarm-up,slowACilter,sinewave.
2. 20%overrangeonallranges,except1000DCV,750ACV,10ADC,3AAC,10AAC,
anddiodetest.
3. Relativetocalibrationstandards. 4. Addthisforeach°CoutsideTCAL±5°C.
5. Speciicationsareforsinewaveinput>0.3%ofrangeand>1mVrms.
750ACVrangelimitedto8x107Volt–Hz.
6. Low-frequencyperformance:threeiltersettingsareavailable:3Hz,20Hz,200Hz.
Frequenciesgreaterthantheseiltersettingsarespeciiedwithnoadditionalerrors.
7. Speciicationsarefor4-wireohmsfunctionor2-wireohmsusingmathnullforoffset.Without
mathnull,add0.2Ωadditionalerrorin2-wireohmsfunction. 8. Speciicationsareforsinewaveinput>1%ofrangeand>10µAAC.The10Arangeisonly
availableonaseparatefront-panelconnector. 0.100
0.035
0.015
0.015
0.015
0.0002
0.0002
0.0002
0.0002
0.0002
9. ACcurrentspeciications>5kHzaretypical.
10. Speciicationsareforthevoltagemeasuredattheinputterminals.The1mAtestcurrent
istypical.Variationinthecurrentsourcewillcreatesomevariationinthevoltagedrop
acrossadiodejunction.
11. Thesespeciicationsarefortypicalperformance. 12.Actualmeasurementrangeandprobeerrorswillbelimitedbytheselectedprobe.
ProbeaccuracyadderincludesallmeasurementandITS-90temperatureconversionerrors.
PT100R osettableto100Ω±5Ωtoremovetheinitialprobeerror.
13. Speciicationsarefor60-minutewarm-upandsinewaveinputunlessstatedotherwise.
Speciicationsarefor1-secondgatetime(7-digits).
14. Appliestosineandsquareinputs≥100mV.
For10mVto<100mVinputs,multiply%ofreadingerrorx10.
15. Amplitude10%–120%ofrangeandlessthan750ACV.
16. Squarewaveinputspeciiedfor10Hz–300kHz.
11
MEASUREMENT CHARACTERISTICS
34460A/34461A
Measurement characteristics
DC voltage
Measurementmethod:
A/Dlinearity:
Inputresistance:
0.1V,1V,10Vrange
100V,1000Vrange
Inputbiascurrent:
Inputterminals:
Inputprotection:
True RMS AC voltage
Measurementtype:
Measurementmethod:
Maximuminput:
Inputimpedance:
Inputprotection:
Agilentpatentedcontinuouslyintegrating
multi-slopeIVA/Dconverter
0.0002%ofreading+0.0001%ofrange
Selectable10MΩor>10GΩ
10MΩ±1%
<30pAat25°C
Copperalloy
1000Vonallranges
AC–coupledTrueRMS.
MeasurestheACcomponentoftheinput.
Digitalsamplingwithanti-aliasilter
400DCV,1100Vpeak
1MΩ±1%,inparallelwith<100pF
750Vrmsallranges
DC and True RMS AC current
ACmeasurementtype:
Directlycoupledtothefuseandshunt.
ACTrueRMSmeasurement(measurestheACcomponentonly).
ACmeasurementmethod:
Digitalsamplingwithanti-aliasilter
Inputprotection3A:
Externallyaccessible3.15A,500Vfuse
(Replacementpartnumber2110-15473.15Aexternalfuse)
Internal11A,1000Vfuse
(Replacementpartnumber2110-140211Aexternalfuse)
Inputprotection10A:
Internal11A,1000Vfuse
(34461Aonly)
(Replacementpartnumber2110-140211Aexternalfuse)
Maximumleadresistance
(4-wireohms):
Inputprotection:
Continuity/diode test
Responsetime:
Continuitythreshold:
DC ratio
Measurementmethod:
InputHI-LO:
Reference(sense)HI-InputLO:
Inputtoreference(sense):
Reciprocal-countingtechnique.
MeasurementisAC-coupledusingACmeasurementfunctions.
Voltageranges:
100mVrmsfullscaleto750Vrms.Autoormanualranging.
Gatetime:
10ms,100ms,or1s
Measurementconsiderations Allfrequencycountersaresusceptibletoerrorwhen
measuringlow-voltage,low-frequencysignals.Shielding
inputsfromexternalnoisepickupiscriticalforminimizing
measurementerrors.
Autozero OFF operation
Followinginstrumentwarm–upatastableambienttemperature±1°Cand<10minutes.
Add0.0002%ofrange+5µVforDCVor+5mΩforresistance.
Selectable4-wireor2-wireohms.
CurrentsourcereferencedtoLOinput.
10%ofrangeperleadfor100Ω,1kΩranges.
1kΩperleadonallotherranges.
1000Vonallranges
300samples/swithaudibletone
Fixedat10Ω
1.Forpower-linefrequency±0.1% 2.Forpower-linefrequency±1%,theNMRis40dB
For±3%,use30dB
InputHI-LO/reference(sense)HI-LO
100mVto1000Vranges
100mVto10Vranges(autoranged)
HIandLOreference(sense)terminalsreference
toLOinput<12V
Temperature
PT100platinumRTDsensor,α=0.00385Ω/Ω/°C;DIN/IEC751.
Measurementconversionslimitedto-200to600°C.
5kΩthermistorβ=3891;YSI44007orequivalent.
Measurementconversionslimitedto-80to150°C.
12
Frequency and period
Measurementmethod:
Measurement settling considerations
ACV, ACI, Frequency, Period
Defaultdelaysareselectedtogiveirstreadingrightformostmeasurements.Errorswill
occurwhenattemptingtomeasurethefrequencyorperiodofaninputfollowingaDC
offsetvoltagechange.TheinputblockingRCtimeconstantmustbeallowedtofullysettle
(upto1second)beforethemostaccuratemeasurementsarepossible.
Applying>300VACrmsor>1Armswillcauseself-heatinginsignal-conditioningcomponents.Theseerrorsareincludedintheinstrumentspeciications.Internaltemperature
changesduetoself-heatingmaycauseadditionalerroronotherfunctionsorranges.
Theadditionalerrorwillgenerallydissipatewithinafewminutes.
DCV, DCI, Resistance
Applying>1ADCwillcauseself-heatinginsignal-conditioningcomponents.Theseerrors
areincludedintheinstrumentspeciications.Internaltemperaturechangesduetoselfheatingmaycauseadditionalerroronotherfunctionsorranges.Theadditionalerrorwill
generallydissipatewithinafewminutes.
Readingsettlingtimesareaffectedbysourceimpedance,cabledielectriccharacteristics,
andinputsignalchanges.AgilentrecommendstheuseofPTFEorotherhigh-impedance,
low-dielectricabsorptionwireinsulationforthesemeasurements.
AC crest factor and peak input
Crestfactor:
10:1maximumcrestfactor,(3:1atfull-scale).Measurement
bandwidthlimitedto300kHzforsignalplusharmonics.
Peakinput:
300%ofrangeormaximuminput
Overloadranging:
Willselecthigherrangeifpeakinputoverloadisdetected
duringautorange.Overloadisreportedinmanualranging.
Resistance
Measurementmethod:
Measurement noise rejection
60Hz(50Hz)for1kΩLOleadunbalance(±500Vpeakmaximum)
DVCCMRR:140dB
AVCCMRR:70dB
Integration time
Normal mode rejection1
100PLC/1.67s(2s)
60dB2
10PLC/167ms(200ms)
60dB2
1PLC/16.7ms(20ms)
60dB2
0.2PLC/3ms(3ms)
0dB
0.02PLC/400µs(400µs)
0dB
OPERATING CHARACTERISTICS
34460A/34461A
Performance versus measurement speed
For DC voltage, DC current, and resistance 1
34460A
Readings/s
0.6(0.5)
6(5)
60(50)
100
300
Integration time
100PLC/1.67s(2s)
10PLC/167ms(200ms)
1PLC/16.7ms(20ms)
0.2PLC/3ms(3ms)
0.02PLC/400µs(400µs)
Digits
6½
6½
5½
5½
3½
AC voltage, AC current 3, 4
Digits
6½
6½
6½
6½
ACV
.4/s
1.6/s
40/s
50/s5
ACI
.6/s
4/s
40/s
50/s5
Aperture
1Second
0.1Second
0.01Second
Digits
7
6
5
Readings
1
10
80
Frequency, period
1.Readingspeedsfor60Hz(and50Hz)operation,autozerooff,ixedrange. 2.Add20µVforDCVand20mΩforresistance.
Add0.2µAforDCcurrent+10xtheaboverangeerrorforthe10mArange. 3.Maximumreadingratesfor0.01%ofACstepadditionalerror.
AdditionalsettlingdelayrequiredwheninputDClevelvaries.
4.Forexternaltriggerorremoteoperationusingdefaultsettlingdelay(DelayAuto).
5.Maximumusefullimitwithdefaultsettlingdelaysdefeated.
Digits
6½
6½
5½
5½
4½
34461A
Readings/s
0.6(0.5)
6(5)
60(50)
300
1000
AC filter
Slow
Medium
Fast
Fast
System Speeds (average)
DC voltage, DC current, resistance 1, 2
Autorangetime3
Maximuminternaltriggerrate
Maximumexternaltriggerrate
ASCIIreadingstobus
Singlereadingtransactionrate4
Additional noise error
0%ofrange
0%ofrange
0.001%ofrange
0.001%ofrange 2
0.01%ofrange2
34460A DMM rear panel with GPIB option installed.
34460A
<30ms
300/s
300/s
300/s
50/s
34461A
<30ms
1000/s
1000/s
1000/s
150/s
AC voltage, AC current 5
Autorangetime3
Maximuminternaltriggerrate
Maximumexternaltriggerrate
ASCIIreadingstobus
Singlereadingtransactionrate4
10/s
50/s
50/s
50/s
50/s
10/s
50/s
50/s
50/s
50/s5
Frequency, period 6
Autorangetime3
Maximuminternaltriggerrate
Maximumexternaltriggerrate
ASCIIreadingstobus
Singlereadingtransactionrate4
10/s
80/s
80/s
80/s
50/s
10/s
80/s
80/s
80/s
50/s
34461A DMM rear panel with GPIB option installed.
1.0.02NPLC,delay0,autozerooff,mathoff,anddisplayoff.
2.TheseratesapplytoallI/Ointerfaces.
3.Timetoautomaticallychangeonerangeandbereadyfornewmeasurement,≤10V,≤10MΩ.
4.IncludesmeasurementandIOtime(assumesconnectionviaSOCKETS.VXI-11connectionsmaybeslower).
5.FastACilter,delay0,mathoff,anddisplayoff.
6.10msaperture,fastACilter,delay0,mathoff,anddisplayoff.
13
GENERAL CHARACTERISTICS
34460A/34461A
General characteristics
Line power
Powersupply:
Powerlinefrequency:
Powerconsumption:
100/120(127)/220(230)/240VAC±10%,CATII
50/60/400Hz±10%
25VA
Environment
Operatingenvironment: Fullaccuracyfor0to55°C
Fullaccuracyto80%R.H.at40°Cnon–condensing
Operatingaltitude:
Upto3000m
Storagetemperature: -40to70°C
Mechanical
Rackdimensions:
Benchdimensions:
Weight:
(WxHxD):212.8mmx88.3mmx272.3mm
(WxHxD):261.2mmx103.8mmx303.2mm
34460A:3.68kg(8.1lb)
34461A:3.76kg(8.3lb)
Regulatory
Safety
EN61010-1:2010(3rdEdition)
ANSI/ISA-61010-1(82.02.01)ThirdEdition
ANSI/UL61010-1ThirdEdition
CAN/CSA-C22.2No.61010-1ThirdEdition
EN61010-2-030:2010(1stEdition)
ANSI/ISA-61010-2-030(82.02.03)FirstEdition
ANSI/UL61010-2-030FirstEdition
CAN/CSA-C22.2No.61010-2-030FirstEdition
RefertoDeclarationofConformityforcurrentrevisions
MeasurementCategoryIIto300V
OthernonMAINScircuitsto1000Vpk
PollutionDegree2
EMC
IEC61326
EN61326
CISPR
ICES-001
AS/NZS2064.1
RefertoDeclarationofConformityforcurrentrevisions
Acousticnoise(nominal)45dBA
Acousticnoise(nominal) 45dBA
Triggering conditions
Externalinput
Low–powerTTLcompatibleinputprogrammableedgetriggered
Delay:
<1µs
Jitter:
<1µs
Minimumpulsewidth: 1µs
Maximumrate:
Upto1kHz(34461A),upto300Hz(34460A)
Voltmetercompleteoutput 3.3Vlogicoutput
Polarity:
Programmableedgepulse
Pulsewidth:
Approximately2µs
Computer interfaces
LXI(rev1.4)
USB
GPIB
Language
10/100Base-TEthernet(Sockets,VXI-11protocol,
Webuserinterface)(Optionalon34460A)
USB2.0(USB-TMC488&MTPprotocol)
OptionalGPIBIEEE-488
SCPI-1999,IEEE-488.2,34401Acompatible
Front-panel USB host port
SupportsUSB2.0high-speedmassstorage(MSC)classdevices
Capability:import/exportinstrumentconigurationiles,savevolatilereadings
andscreencaptures
14
System speeds (averages)
Benchmark
GPIB
Functionchange1
50/s
100/s
Rangechange2
USB 2.0
50/s
100/s
VXI-11
50/s
100/s
Sockets
50/s
100/s
1.Ratetochangefrom2-wireresistancetoanyotherfunction
2.Ratetochangefromonerangetothenexthigherrange,≤10V,≤10MΩ
Triggering and memory
Samplespertrigger
Triggerdelay
Externaltriggerdelay
Externaltriggerjitter
Volatilereadingmemory
1to1,000,000
0to3600sec(~1µsstepsize)
<10µs
<1µs(DCixedrange)
10,000(34461A),1,000(34460A)
Probe hold
Captureandnavigatestablelistofreadings
Internal flash file system
80MBtotalcapacity
Savereadingmemorytonon-volatilememoryinCSVformat
Storeandrecalluser-deinedstates,power-offstate,1andpreferenceiles
SavescreencapturesinBMPorPNGformats
1.Power-offstateonlywhenpower-downisinitiatedviafront-panelpowerswitch.
Math functions
Perfunctionnull,min/max/avg/Sdev,dB,dBm,span,count,limittest,histogram
Display
4.3”colorTFTWQVGA(480x272)withLEDbacklight
Supports:basicnumber,barmeter,trendchart(34461Aonly),histogramviews
User-deinedpower-onmessage,displaylabel,andselectablescreencolors
Integrated,context-sensitivesystemhelpthroughpress-and-holdbuttons
Real-time clock/calendar
Setandread,year,month,day,hour,minute,seconds(Note:secondsnotsettable)
BatteryCR-2032coin-type,replaceable,>10-yearlife(typ)
Software available
IOLibraries:www.agilent.com/find/IOLibraries
DMMConnectivityUtilitysoftware:www.agilent.com/find/DMMutilitysoftware
OPTIONS & ACCESSORIES
34460A/34461A
Options
Standalone product numbers
34460A Digital multimeter, 6½ digit, basic Truevolt DMM
LAN
RearpanelLAN/LXIwebinterface,externaltriggeringfor34460A–
factoryenabled
SEC
NISPOMandilesecurityforTruevoltSeriesDMMs–factoryenabled
Z54
Certiicateofcalibration–ANSI/NCSLZ540.3-2006,printed
GPB
GPIBinterfacemoduleforTruevoltSeriesDMMs–factoryinstalled
ACC
Accessorykitfor34460A–documentationCDs,testleads,USBcable;
shippedwithunitfromfactory
Ordered as standalone to be installed by the distributor or customer
3446LANU Upgrade:
Enable rear panel LAN/LXI web interface, external triggering for 34460A
3446SECU Upgrade:
EnableNISPOMandilesecurityforTruevoltSeriesDMMs
3446GPBU Upgrade:
GPIBuser-installableinterfacemoduleforTruevoltSeriesDMMs
3446ACCU Accessory kit for 34460A:
DocumentationCDs,testleads,USBcable
34461A Digital multimeter, 6½ digit, 34401A replacement, Truevolt DMM
SEC
NISPOMandilesecurityforTruevoltSeriesDMMs–factoryenabled
Z54
Certiicateofcalibration–ANSI/NCSLZ540.3-2006,printed
GPB
GPIBinterfacemoduleforTruevoltSeriesDMMs–factoryinstalled
Rack mount kits
Accessories
34190A
34191A
Accessories included
34460A:
Powercord
Calibrationcertiicate
34138Atestleadsetwithprobes,inetipprobes,SMTgrabbers
andminigrabberattachments
Powercord
DocumentationCD
IOLibrariesCD
USBcable
Calibrationcertiicate
Accessories available
34461A:
11059A
11060A
11062A
34131A
34133A
34134A
34136A
34138A
34151A
34152A
34153A
34162A
34171B
34172B
34308A
34330A
E2308A
Y1133A
Kelvinprobeset
Surface-mountdeviceprobe
Kelvinclipset
Transitcase
Precisionelectronictestleads
DC-coupledcurrentprobe
High-voltageprobe
Testleadset
ThreeSignalWedgeProbeKit
PT100/RTD4-WireClassASensorKit
PT100/RTD4-WireClassSensorElements
Accessorypouch
Inputterminalblock
Calibrationshort
Thermistorkit
30-Acurrentshunt
Thermistortemperatureprobe
Low-thermalexternaldigitalmultimeterscanningkit
34194A
Rackmount kit:
Useformountingone2Uinstrumentbyitself,withoutanotherinstrument
laterallynexttoit.Includesoneracklangeandonecombination
racklange-illerpanel.
2U dual flange kit:
Useformountingtwo2Uinstrumentsside-by-side.
Includestwostandardracklanges.
Note: Mounting two instruments side-by-side will require the
34194A dual-lock link kit and a shelf for the instruments to sit on.
Dual lock link kit:
Forside-by-sidecombinationsofinstrumentsandincludeslinks
forinstrumentsofdifferentdepths.
Standard 3-Year Warranty
Definitions
Specification (spec)
Thewarrantedperformanceofacalibratedinstrumentthathasbeenstoredfora
minimumof2hourswithintheoperatingtemperaturerangeof0–55°Candaftera
60-minutewarmupperiod.Allspeciicationsincludemeasurementuncertaintyand
werecreatedincompliancewithISO-17025methods.Datapublishedinthisdocument
arespeciications(spec)onlywherespeciicallyindicated.
Typical (typ)
Thecharacteristicperformance,which80%ormoreofmanufacturedinstrumentswill
meet.Thisdataisnotwarranted,doesnotincludemeasurementuncertainty,andis
validonlyatroomtemperature(approximately23°C).
Nominal (nom)
Themeanoraveragecharacteristicperformance,orthevalueofanattributethatis
determinedbydesignsuchasaconnectortype,physicaldimension,oroperatingspeed.
Thisdataisnotwarrantedandismeasuredatroomtemperature(approximately23°C).
Measured (meas)
Anattributemeasuredduringdevelopmentforpurposesofcommunicatingtheexpected
performance.Thisdataisnotwarrantedandismeasuredatroomtemperature(approximately23°C).
TCAL
Thetemperatureatwhichtheinstrumentwascalibrated.
15
www.agilent.com
www.agilent.com/find/truevolt
myAgilent
myAgilent
www.agilent.com/find/myagilent
Apersonalizedviewintotheinformationmost
relevanttoyou.
www.axiestandard.org
AdvancedTCA®ExtensionsforInstrumentation
andTest(AXIe)isanopenstandardthat
extendstheAdvancedTCAforgeneralpurpose
andsemiconductortest.Agilentisafounding
memberoftheAXIeconsortium.
Three-Year Warranty
www.agilent.com/find/ThreeYearWarranty
Agilent’scombinationofproductreliabilityandthreeyearwarrantycoverageisanotherwaywehelpyou
achieveyourbusinessgoals:increasedconfidence
inuptime,reducedcostofownershipandgreater
convenience.
Agilent Advantage Services
www.agilent.com/find/AdvantageServices
Accuratemeasurementsthroughoutthelifeofyour
instruments.
Agilent Electronic Measurement Group
www.lxistandard.org
LANeXtensionsforInstrumentsputsthepower
ofEthernetandtheWebinsideyourtestsystems.AgilentisafoundingmemberoftheLXI
consortium.
www.pxisa.org
PCIeXtensionsforInstrumentation(PXI)modular
instrumentationdeliversarugged,PC-based
high-performancemeasurementandautomation
system.
Agilent Channel Partners
www.agilent.com/find/channelpartners
Getthebestofbothworlds:Agilent’smeasurement
expertiseandproductbreadth,combinedwith
channelpartnerconvenience.
DEKRA Certified
ISO 9001:2008
Quality Management Sys
System
www.agilent.com/quality
FormoreinformationonAgilentTechnologies’
products,applicationsorservices,pleasecontactyour
localAgilentoffice.Thecompletelistisavailableat:
www.agilent.com/find/contactus
America
Canada
Brazil
Mexico
United States
(877) 894 4414
(11) 4197 3600
01800 5064 800
(800) 829 4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Other AP Countries
1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
(65) 375 8100
Europe & Middle East
Belgium
Denmark
Finland
France
32 (0) 2 404 93 40
45 45 80 12 15
358 (0) 10 855 2100
0825 010 700*
Germany
Ireland
Israel
Italy
Netherlands
Spain
Sweden
United Kingdom
49 (0) 7031 464 6333
1890 924 204
972-3-9288-504/544
39 02 92 60 8484
31 (0) 20 547 2111
34 (91) 631 3300
0200-88 22 55
44 (0) 118 927 6201
*0.125 €/minute
For other unlisted countries:
www.agilent.com/find/contactus
(BP-3-1-13)
Productspecificationsanddescriptionsinthisdocument
subjecttochangewithoutnotice.
©AgilentTechnologies,Inc.2013
PublishedinUSA,August20,2013
5991-1983EN