Finding the Calibration Factors For Multi-Product Registers
Remove the front calibration plate (Item 1) from the front of the register,
Now move switch # 6 from its current setting (“REMEMBER WHERE IT WAS”) to position # 9. Then reset the register by moving the toggle switch (#7) to the full right position. This is a spring loaded reset, the toggle switch will return to the centre compensated or “NET” position.
WARNING: NEVER MOVE SWITCH #7 (COMPENSATOR) TO THE LEFT OR GROSS POSITION,
YOU RISK A CALIBRATION OVERWRITE, AND YOUR TRUCK MAY HAVE TO BE RE-INSPECTED
IF YOU DO NOT KNOW YOUR CALIBRATION FACTORS !
The register will now display the calibration factor for the product that appears on the MID:COM 8000 cab mounted computer on line 2 of the display. Record the product and calibration factor below.
_____________________
(product on line 2, (gasoline or distillate) from the MIDCOM cab mounted computer screen)
Now write down what is being displayed on the MIDCOM register below.
_____________________
Now “switch” products by pressing the line 2 button on the MIDCOM cab mounted computer (the display should change on line 2 of the screen to the other product, gasoline or distillate).
The MIDCOM register will now be displaying the calibration factor for the other product. Now as above record the information below.
_____________________
(product on line 2, (gasoline or distillate) from the MIDCOM cab mounted computer screen)
Now write down what is being displayed on the MIDCOM register below.
_____________________