Determination of Metals in Alumina Using ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Introduction Deen Johnson The use of alumina is very common in the petroleum, petrochemcial and pharmaceutical manufacturing. Alumina is used as a substrate for many of the catalysts used in the refining or manufacturing process. The alumina is chemically bound to, or coated with, active catalytic agents to produce the desired reactions. Typical catalytic agents are; platinum, palladium, rhodium, titanium, tungsten and tantalum [1,2]. As substrate material, the alumina must be sufficiently pure. The alumina may contain other metals in the lattices of the structure which can result in contamination [3]. It is not desirable to have such contaminates in the alumina reacting with the catalytic agent or the products. These metallic impurities could alter desired reaction rates, or contaminate the final product. It is therefore necessary to analyze the alumina to determine the major and minor metallic components. The scope of this study was to determine six metals in alumina. The metals determined are; lanthanum, silicon, iron, sodium, titanium and lead. The primary elements of interest in the alumina are Si, Fe, Na and Ti. Additionally, the study looked at three different acid mixes for the digestion of the alumina. The acid mixtures used were; 1:1 HNO3, 1:1 HCl and a mix called the Alcoa #1 mix. The Alcoa #1 mix contained 350 mL water, 250 mL of 1:1 sulfuric acid, 200 mL concentrated hydrochloric acid and 200 mL concentrated nitric acid. The results were checked against both colorimetric and atomic absorption analysis. A 3% lanthanum nitrate solution was also analyzed for lead. All the samples were analyzed in the same analytical sequence. Experimental Table 1. Instrumentation Nebulizer pressure (kPa) Stabilization time (sec) Sample delay time (sec) Rinse time (sec) Snout purge 200 15 12 10 OFF Integration time (sec) Replicates 3 3 PMT (V) Power (kW) Plasma (L/Min) Auxiliary (L/Min) Pump speed (rpm) 650 1.20 12.0 1.50 10 Background correction mode Dynamic Concentration units Max curve order C.C. limit mg/L 4 0.9950 The Liberty 200 ICP-OES was used in this study. The instrument was optimized for the viewing height, scan and search windows, power, plasma flow, auxiliary plasma flow, pump speed and nebulizer pressure. The use of dynamic background correction was used for all the wavelengths in the determination. Analytical Procedure The samples were received in solution having already been prepared in the three acid mixtures. Approximately 0.20 grams of sample had been weighed. 10 mL of the selected acid was added for digestion. The overall dilution ratio was approximately 1:1000. The analytical blank was analyzed for each acid mixture to correct for any contaminates present. The samples were analyzed using standards prepared in both 0.014 M (1% v/v) nitric acid and 0.1 M (1% v/v) hydrochloric acid. The samples were also analyzed with 10,000 mg/L aluminum spiked into each standard to determine whether matrix matching of samples and standards is required. Table 2. Parameters Common to All Elements in the Alumina Analysis Element Specific Parameters for the Metals La Wavelength (nm) View height (mm) 8 The common parameters used in the analysis of the alumina are presented in Table 1. Si Fe Na Ti Pb 379.478 251.611 259.940 589.592 323.904 220.353 6 6 12 8 8 0.040 0.080 0.026 0.027 Windows Search (nm) 0.080 0.040 Scan (nm) 0.120 0.060 0.060 0.120 0.030 0.040 Filter position 6 6 6 7 6 1 Order 1 2 2 1 2 3 Results Standard 1 (units mg/L) 50.00 5.000 5.000 5.000 5.000 5.000 The results for the analysis of the alumina using the different acid mixtures are shown in Table 3. The results have been corrected for the blank contribution. Standard 2 (units mg/L) 100.0 10.00 10.00 10.00 10.00 10.00 The titanium was the element of primary interest in this analysis. The sample digested with the nitric acid exhibited the highest recovery for titanium. The hydrochloric acid gave the next best recovery, followed by the Alcoa #1 acid mixture. The recovery may vary with the acid mixture, due to the possibility of interferences in the plasma [4]. Titanium, silicon and lanthanum are refractory in nature which may affect the recoveries for the different acid mixtures used in digestion. These trends were similar to those found when the samples were analyzed by colorimetric and atomic absorption analysis [5]. Table 3. Element specific parameters, optimized for each of the metals analyzed, are shown in Table 2. Analytical Result (in mg/L) with Blank Correction Applied Acid Mixture La Si Fe Na Ti Pb HCl ND 0.104 0.038 0.018 3.81 0.015 HNO3 ND 0.086 0.032 0.028 4.06 0.176 Alcoa ND 0.102 0.035 0.014 3.16 0.029 La solution 97.34 0.702 1.75 0.033 3.00 0.100 ND = Non detected Titanium, silicon and lanthanum are refractory in nature which may affect the recoveries for the different acid mixtures used in digestion. These trends were similar to those found when the samples were analyzed by colorimetric and atomic absorption analysis. 2 No problems were found with background or spectral interferences with the sample dilutions used in this procedure, due to the high resolution of the spectrometer used. After dilution of the samples by approximately 1:1000, no problems were found either with background or spectral interferences from the aluminium. The dynamic background more than adequately compensated for the background levels. Because there were no problems with background or spectral interferences, there is the possibility of using a lower dilution than that used here. There were no spectral interferences encountered in the samples. For example, the high resolution for Pb in 3rd order (0.007 nm) easily resolved the very large Al peak near the Pb 220.353 nm line. Acknowledgement A Quality Control Standard (QCS) was analyzed during the analysis. The QCS, which was the low calibration solution, was analyzed after every 4 samples. The results of the QCS are reported in Table 4 and are the mean value of two determinations. The range shown represents plus or minus one standard deviation. Also shown are typical precision values (%RSD) for a within run determination of three replicates (Table 1). Table 4. We wish to acknowledge Ken Youngblood and David Bose of Vista Chemical (Austin, Texas) for supplying the samples. References 1. W. L. Leffler “Catalytic Reforming”. Petroleum Refining for the Non-Technical Person. PennWell Books, Tulsa, Oklahoma, U.S.A., 1979, p69. 2. P. Wiley “Catalytic Reforming”. Petrochemicals. John Wiley & Sons, New York, U.S.A., 1986, p90. 3. F. A. Cotton and G. Wilkinson “The Group IIIA (13) Elements: Al, Ga, In, Tl”. Advanced Inorganic Chemistry, 5th Edition,, John Wiley & Sons, New York, U.S.A. 1988, pp211-212. 4. G. A. Meyer, and P. N. Keliher “An Overview of Analysis by Inductively Coupled Plasma - Atomic Emission Spectrometry”. Inductively Coupled Plasma in Analytical Atomic Spectrometery. Second Edition, A. Montasser and D.W. Golightly, VCH Publishers, Inc, New York, U.S.A. 1992, p492. 5. Personal Communication Quality Control Standard Results and Recoveries Found Metal Within run concentration Recovery precision RSD mg/L % % La 49.16 ± 1.35 98.3 0.49 Si 4.85 ± 0.13 97.0 0.84 Fe 4.95 ± 0.09 99.1 0.83 Na 4.85 ± 0.09 97.0 0.71 Ti 5.41 ± 0.13 108.1 0.56 Pb 4.93 ± 0.008 98.5 1.75 Conclusions For More Information ICP-OES analysis of metals in alumina using the Liberty 200 is fast, accurate and precise. The ICP technique is much faster for this determination than the atomic absorption methods and considerably faster than the colorimetric methods. The ICP results were similar to those obtained using the other techniques. The low concentrations of iron, silicon, sodium and lead found in the solutions were measured with good reproducibility. For more information on our products and services, visit our Web site at www.agilent.com/chem Automated analysis with quality control solutions taken through the procedure, confirms the accuracy of the results. 3 www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. © Agilent Technologies, Inc., 1993 Printed in the USA November 1, 2010 ICPES-11