Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 1 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Signature Page Author’s Signature: Trevor Brasel PhD__________________ Typed Name of Author Acceptance by Subcontracting Institution: Robert Sherwood_PhD______________ Typed Name of Subcontracting PI Acceptance by the University of New Mexico: __C. Rick Lyons MD. PhD__________ Principal Investigator Acceptance by NIAID: _Patrick Sanz______3/5/10_________ Typed Name of NIAID Project Officer Heidi Holley ___________________ Typed Name of NIAID Contract Officer __3/9/2010___ Date Accepted Page 1 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 2 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Table of Contents 1 Milestone Summary ...................................................................................................................................................2 2 Milestone Objectives..................................................................................................................................................3 3 Methods, Critical Reagents and SOPs........................................................................................................................3 3.1 Methods Summary......................................................................................................................................3 3.2 Critical Reagents ...................................................................................................................................... 17 3.3 Standard Operating Procedures ................................................................................................................ 17 4 Salient Original Data, Results, Interpretation, Quality Control ............................................................................... 18 4.1 Original Data and Results (Rationale, Tables/Figures with legends and location annotations) ............... 18 4.2 Interpretation ............................................................................................................................................ 18 4.3 Quality Control ......................................................................................................................................... 39 5 Deliverables Completed ........................................................................................................................................... 46 6 Appendices ............................................................................................................................................................... 47 6.1 Appendix 1: Original Data Tables and Figures ....................................................................................... 47 6.2 Appendix 2: Quality Assessment of Milestone Completion and Report ................................................. 53 6.3 Appendix 3: Additional Data/Figures not included in the Text of the Milestone Completion Report (Section 4) .............................................................................................................................................. 56 1 Milestone Summary The goal of Milestone #3 was to optimize Francisella tularensis (F. tularensis) bioaerosol procedures, including bacterial proliferation and aerosol generation/collection techniques. Completion of these tasks was critical in that all future vaccine efficacy testing in nonhuman primates (as described under the Tularemia Vaccine Development Contract) will be dependent upon an optimal means of F. tularensis bioaerosol challenge. Cumulative results from live vaccine (LVS) and SCHU S4 strain testing demonstrated that an optimal F. tularensis challenge bioaerosol was achieved under the following parameters: (1) challenge material prepared from a 24h subculture of F. tularensis in Chamberlain’s broth subcultured from 48h growth on blood cysteine glucose agar (BCGA) (2) bacterial suspension aerosolized using the Collison 3-jet nebulizer at a normalized flow rate of 7.5 L/min (3) relative humidity in the aerosol chamber maintained at 6090% (4) bioaerosol material collected into brain heart infusion broth (BHIB) plus antifoam A using the AGI-4 all-glass impinger (5) bacterial culture for purity and titer performed on BCGA. Because of the observed repeatability, bacterial viability retention, and spray factor efficiencies, these procedures will be incorporated into all future F. tularensis nonhuman primate aerosol challenge studies. Page 2 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Version: 2.0 Page 3 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 2 Accepted Date:4/23/09 Milestone Objectives Bioaerosol technique selected and optimized. This milestone consisted of testing of a variety of selected aerosol generation devices as agreed upon by LBERI and UNM in order to optimize aerosol delivery of bacteria to mammals. Minimal endpoints included determination of the conditions that optimize bacterial survival during aerosolization (pre- and post-), reproducibility and accuracy of targeted inoculums, and size of droplets generated. This Milestone Completion Report (MSCR) for Milestone 3 summarizes the data generated. 3 Methods, Critical Reagents and SOPs 3.1 3.1.1 Methods Summary NOTE: Because F. tularensis is a select agent, all manipulations were performed under ABSL-3 conditions using aseptic techniques in a Class II biological safety cabinet. Comparison of Solid and Liquid Media for the Culture of F. tularensis Prior to initiation of MS3, growth conditions of F. tularensis LVS were optimized on solid agar and in liquid broth media. The following nine solid agar media types were tested: Chamberlain’s agar supplemented with hemoglobin (CAH), cysteine heart agar supplemented with 2 g/L hemoglobin (CHAH-2), cysteine heart agar supplemented with 10 g/L hemoglobin (CHAH-10), cysteine heart agar (CHA), blood cysteine glucose agar (BCGA), cysteine heart agar with blood (CHAB), supplemented BCGA (SBCGA), glucose cysteine agar with thiamine (GCAT), and Chamberlain’s agar (CA). Components of each agar are detailed in Table A1 in Appendix 3. For each agar type, a frozen LVS working stock vial (see Section 3.1.2) was thawed and used to inoculate a 90-mm agar plate via the standard quadrant streak method as depicted in Figure 1. The plates were allowed to incubate in a 37°C ± 2°C incubator for 24 or 48h after which purity and phenotype were evaluated. Additionally, as a positive control, a 48h culture of LVS grown in Chamberlain’s broth (procedure detailed below) was serially diluted and plated onto eight of the nine types of agar media (GCAT was removed from this portion of the study). Plates were allowed to incubate in a 37°C ± 2°C incubator for 72h after which colonies were enumerated (Table 6). Relative recoveries of LVS for each media type were compared. The following five broth media types were tested as well: modified cysteine partial hydrolysate (MCPH), modified Mueller-Hinton broth (MMHB), Chamberlain’s broth (CB), Chamberlain’s broth supplemented with hemoglobin (10 g/L; CBH), and MCPH with isovitale-X (MCPHI). Components of each broth are detailed in Table A2 in Appendix 3. Broth cultures were grown with shaking (225 rpm) in a 37°C ± 2°C incubator for 48h in 250- or 500-mL Erlenmeyer flasks. The lower volumes were inoculated with 20 μL of LVS working stock while 500-mL flasks were inoculated with 40 μL. Bacterial suspensions were serially diluted and cultured Page 3 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 4 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 onto 90-mm CHAB plates. The plates were allowed to incubate in a 37°C ± 2°C incubator for 72h after which purity, phenotype, and colony counts were determined. 3.1.2 Preparation of Francisella tularensis Seed and Working Stocks F. tularensis seed and working stocks were prepared according to SOP LBERI-1.1, Preparation of Francisella tularensis Seed and Working Stocks. Bacterial seed stocks were prepared using the following primary stock vials: LVS Lot 703-0303-016 (13DEC05, Fisher Biosciences) SCHU S4 Submaster Project 110419.2 (9MAR06, Fisher Biosciences) o SCHU S4 Sublot#1 was received at UNM 9MAR06 and transferred to LBERI o SCHU S4 Sublot#1 (NOV 05 to FEB 06, Midwest Research Institute) to Fisher o SCHU S4 parent strain originated from the Salk Institute in Swiftwater, PA, Lot 623-42 (6SEP86) o The Salk Institute received the seed stock from Fort Detrick, MD. Briefly, upon receipt, the vial was thawed and used to inoculate a 90-mm BCGA plate via the standard quadrant streak method as depicted in Figure 1 below: Figure 1. Standard microbiological quadrant streak method for bacterial cultures. The plates were allowed to incubate in a 37°C ± 2°C incubator for 48-72h after which purity and phenotype were determined. After 48-72h of growth on BCGA, F. tularensis LVS and SCHU S4 colonies are 3-4 mm in diameter, gray-white to slightly blue and opaque, demonstrate little or no hemolysis, present a raised, convex appearance, and have a wet, shiny surface as depicted in Figure 2 below: Page 4 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 5 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Figure 2. Francisella tularensis LVS growth on BCGA. This image represents colony morphology after 48h of incubation at 37°C (see method summary in section 3.1.1) Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06\FY06-078_TUL-03\LVS Growth characterization\48h on BCGA.jpeg Upon verification of purity and phenotype, one isolated colony was picked from the BCGA plate and aseptically inoculated into 100 mL of sterile Chamberlain’s broth medium in a sterile, plastic 250-mL Erlenmeyer flask. The broth was incubated with shaking at 37°C ± 2°C, 225 ± 25 rpm for 48h ± 8h. Following incubation, the culture was aseptically aliquoted into two sterile 50-mL centrifuge tubes in a Class II biosafety cabinet (BSC). Suspensions were centrifuged in a sealed rotor at 2,400 x g, 4-6°C, for 20 min after which the supernatant was decanted from each tube. To each bacterial pellet, 20-25 mL of sterile Chamberlain’s broth was added (depending on the desired number of seed stock vials). The solutions were mixed thoroughly using a vortex mixer to ensure complete resuspension of the culture. An equal volume of 20% (w/v) sucrose was then added to each tube resulting in a final sucrose concentration of 10%. Aliquots (0.5 mL) of this suspension were aseptically added to 1.8 mL capacity cryovials and stored in a -80°C freezer. These cryovials served as the F. tularensis seed stock. The working stock was prepared in the same manner using 100 L of bacterial suspension from a seed stock vial for the initial Chamberlain’s broth inoculum. These stocks, at a working volume of 1.0 mL, were stored in a -80°C freezer for a maximum of six months due to significant titer decrease observed after this time period. 3.1.3 Generator and Impingement Solution Optimization In addition to growth media, generator and impingement solutions were tested in order to determine the optimal aerosolization and collection media for F. tularensis. Focus for this study Page 5 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 6 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 was placed on the media type and bacterial viability as a function of time and temperature. The following four media types were evaluated: Chamberlain’s broth, phosphate buffered saline (PBS), brain-heart infusion broth (BHIB), and 1% (w/v) peptone (See Figure 9). Each solution was inoculated with F. tularensis LVS from frozen working stock to a final concentration of 2.0 or 3.0 x 107 CFU/mL in 20 mL. Suspensions were allowed to sit at room temperature (20°C) or 4°C for four hours during which a 500 L aliquot was removed hourly, serially diluted, and cultured onto 90-mm CHAB plates. We have observed previously with other bacterial agents that storage temperature significantly impacts bioaerosol stability. For example, with Yersinia pestis, bioaerosol stability (i.e., bacterial viability) greatly decreases if the generator suspension is not stored cold. For F. tularensis, this variable was unknown thus requiring the need to test the two temperatures stated above. Following inoculation, plates were allowed to incubate in a 37°C ± 2°C incubator for 72h after which purity, phenotype, and colony counts were determined. 3.1.4 Generator Suspension Preparation Throughout the milestone testing period, various methods of F. tularensis growth were tested for use as the generator suspension material. Of these, two methods were finalized: one for frozen material and one for fresh bacterial growth. The following sections describe these optimized procedures. Frozen F. tularensis Separate bacterial suspensions were prepared for each bioaerosol run. Because of an observed titer decrease over time, all working stock vials used for frozen F. tularensis bioaerosol testing were less than or equal to six months in age. Briefly, a working stock cryovial of confirmed titer (i.e., culture was performed shortly before testing to obtain the viable bacterial titer) was removed from frozen storage, thawed and used to directly prepare the suspensions needed for the day of aerosol testing. Bioaerosol material was prepared in sterile BHIB to optimal working volumes (based on manufacturer recommendations and reservoir capacity) of 5 mL for the Aeromist nebulizer, 10 mL for the Collison and Micropump generators, and 20 mL for the sparging generator reservoir. Prior to delivery to the Aerosol laboratory, bacterial suspensions were weighed in order to establish a baseline for post-bioaerosol volume calculations. Fresh F. tularensis Separate bacterial suspensions were prepared for each bioaerosol run. Briefly, a working stock cryovial was removed from frozen storage, thawed and used to inoculate three BCGA plates using the standard bacterial quadrant streak method. Plates were allowed to incubate at 37°C for 48h or until the colonies were 3-4 mm in diameter. Following incubation, plates were removed and purity was verified. Using a 1-L inoculating loop, colonies (generally n=10) were removed from the BCGA and inoculated into 4.5 mL of CB. This suspension was mixed thoroughly using Page 6 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 7 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 a vortex mixer and analyzed at 600 nm (OD600) using a spectrophotometer. Data were used to normalize the bacterial suspension to a final OD600 value of 0.100 ± 0.01. Three new, sterile 500mL baffled flasks containing 100 mL of CB were each inoculated with 400 L of the normalized suspension. Flasks, sealed with filtered stoppers, were allowed to incubate in a 37°C incubator with shaking (200 RPM) in the dark for 24h. Using a growth curve prepared from similarly prepared material, bioaerosol material was diluted from the 24h culture in sterile BHIB to final volumes of 5 mL for the Aeromist nebulizer and 10 mL for the Collison and Hospitak generators. Prior to delivery to the Aerosol laboratory, bacterial suspensions were weighed in order to establish a baseline for post-bioaerosol volume calculations. (See LBERI SOP 2.1) 3.1.5 Impingement Solution Preparation Impingement solutions were prepared according to SOP LBERI-2.1, Preparation of Francisella tularensis for Bioaerosol Studies. In brief, 20 mL of sterile BHIB were aseptically dispensed into sterile, plastic 50 mL centrifuge tubes (n=1 per bioaerosol run). To minimize foaming, one drop (10-20 L) of Antifoam A was added to each tube prior to delivery to the Aerosol laboratory. Allglass impinger (AGI) solutions were then weighed in order to establish a baseline for postbioaerosol volume calculations. 3.1.6 Aerosol Exposure System and Bioaerosol Generator Operation The LBERI NHP head-only exposure system (Figure 3) was used for the F. tularensis bioaerosol runs conducted throughout MS3. The main components of the system were as follows: The bioaerosol generators, operated using HEPA-filtered compressed air at selected delivered pressures, were used to aerosolize the F. tularensis at stable, controlled flow rates. The dilution/delivery line was used to add dilution air to the bioaerosol at a controlled rate and to deliver the agent to the head-only chamber. Dilution air was humidified using an ultrasonic humidification chamber. Humidity was maintained between 60% and 90%. A head-only exposure chamber provided primary containment of the aerosolized F. tularensis and ensured a homogenous bioaerosol. Mass flow controllers (MFCs) were used to monitor and control the aerosol generator, dilution makeup air, impingers, and exposure chamber exhaust. A RH/temperature sensor and a percent oxygen transducer were used to measure the relative humidity and temperature and oxygen concentration, respectively, with the exposure chamber. The AGI sampling system was used to collect the bioaerosol from the exposure chamber during each run in order to assess the aerosol concentration and other critical parameters. Page 7 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 8 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 A DustTrak aerosol monitor was used to indicate the relative real-time chamber concentration of the test aerosol. An Aerodynamic Particle Spectrometer (APS) was used to measure the particle size distribution of a representative sample during each exposure period. A Class III biosafety cabinet provided secondary containment of the exposure system. All manipulations were performed therein. The LabVIEW monitoring/control system was used to electronically control and monitor aerosol parameters including nebulizer and AGI flows, humidity, DustTrak data, exhaust flow and to display summary data of the bioaerosol conditions. Figure 3. LBERI NHP Head-Only Exposure System. Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06\FY06078_TUL-03\TUL-03 data files\Exposure line schematic.ppt Page 8 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 9 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 All F. tularensis testing was conducted using the Aeromist, Collison 3-jet, Aeroneb® Lab Micropump, and Hospitak Nebulizers, and/or the sparging generator. Operating procedures for each unit are described in SOP LBERI 3.1 and in the following sections. Page 9 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 10 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Aeromist Nebulizer The Aeromist Nebulizer is depicted in Figure 4: Figure 4. Aeromist Nebulizer. In addition to the Collison and Hospitak nebulizers, the Aeromist was one of the three bioaerosol generators down-selected for extensive LVS and SCHU S4 testing. The Aeromist Nebulizer was filled with 4.5-5.0 mL of the F. tularensis suspension (preparation procedures described above) and was operated by delivering compressed air at 5-20 pounds per square inch, depending on the objectives of the study. This provided a nebulizer flow rate of 3.99.0 L/min. The bacterial aerosol was diluted with humidified air and delivered to the head-only chamber. Total chamber flow was regulated at 16 L/min and main exhaust was adjusted to maintain chamber pressure slightly negative (-0.1 to -0.3 inches water column) with respect to the Class III biosafety cabinet (a means to prevent inadvertent release of the aerosol). Samples were extracted from the head-only chamber to quantify the bioaerosol concentration (via AGI-4 sampling), size distribution (via 30-sec APS measurement) and relative particle concentration (via continuous DustTrak sampling). AGI sampling was performed at flows of ranging from 5.1 to 5.4 L/min during the aerosol trials. Following aerosol trials, AGI suspensions were transferred to the Microbiology laboratory for quantitative analyses as described below (Determination of F. tularensis Bioaerosol Parameters). The nebulizer and AGI flowrates were calibrated on the day of and prior to all of the aerosol trials. Page 10 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 11 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Collison 3-Jet Nebulizer The Collison 3-Jet Nebulizer is depicted in Figure 5: Figure 5. Collison 3-Jet Nebulizer. In addition to the Aeromist and Hospitak nebulizers, the Collison was one of the three bioaerosol generators down-selected for extensive LVS and SCHU S4 testing. It is considered to be the standard generator throughout the bioaerosol field. The Collison nebulizer (BGI, Inc., Waltham, MA) was filled with 10 mL of the F. tularensis suspension (preparation procedures described above) and was operated by delivering compressed air at 24.5 to 26.0 pounds per square inch. This provided a nebulizer flow rate of 7.5 ± 0.1 L/min. The bacterial aerosol was diluted with humidified air and delivered to the head-only chamber. Total chamber flow was regulated at 16 ± 0.2 L/min and main exhaust was adjusted to maintain chamber pressure slightly negative (-0.1 to -0.3 inches water column) with respect to the Class III biosafety cabinet (a means to prevent inadvertent release of the aerosol). Samples were extracted from the head-only chamber to quantify the bioaerosol concentration (via AGI-4 sampling), size distribution (via 30-sec APS measurement) and relative particle concentration (via continuous DustTrak sampling). AGI sampling was performed at flows of ranging from 5.3 to 5.5 L/min during the aerosol trials. Following aerosol trials, AGI suspensions were transferred to the Microbiology laboratory for quantitative analyses as described below (Determination of F. tularensis Bioaerosol Parameters). The nebulizer and AGI flowrates were calibrated on the day of and prior to all of the aerosol trials. Page 11 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 12 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Aeroneb® Lab Micropump Nebulizer The Micropump Nebulizer is depicted in Figure 6: Figure 6. Aeroneb® Lab Micropump Nebulizer. The Micropump nebulizer was used only in LVS testing throughout the milestone study period. The Aeroneb® Lab Micropump Nebulizer (Aerogen, Inc., Galway, Ireland) was filled with 10 mL of the F. tularensis suspension (preparation procedures described above) and was operated by delivering compressed air at a flow rate of 7.5-10.0 L/min. The bacterial aerosol was diluted with humidified air and delivered to the head-only chamber. Total chamber flow was regulated at 16 ± 0.2 L/min and main exhaust was adjusted to maintain chamber pressure slightly negative (0.1 to -0.3 inches water column) with respect to the Class III biosafety cabinet (a means to prevent inadvertent release of the aerosol). Samples were extracted from the head-only chamber to quantify the bioaerosol concentration (via AGI-4 sampling), size distribution (via 30-sec APS measurement) and relative particle concentration (via continuous DustTrak sampling). AGI sampling was performed at flows of ranging from 5.2 to 5.3 L/min during the aerosol trials. Following aerosol trials, AGI suspensions were transferred to the Microbiology laboratory for quantitative analyses as described below (Determination of F. tularensis Bioaerosol Parameters). The nebulizer and AGI flowrates were calibrated on the day of and prior to all of the aerosol trials. Page 12 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 13 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 HospitakTM Nebulizer The Hospitak Nebulizer is depicted in Figure 7: Figure 7. Hospitak Nebulizer. In addition to the Collison nebulizer, the Hospitak was one of the two bioaerosol generators down-selected for extensive LVS and SCHU S4 testing. The Hospitak Nebulizer (Unomedical, a/s, Birkeroed, Denmark) was filled with 10.0 mL of the F. tularensis suspension (preparation procedures described above) and was operated by delivering compressed air at 10 pounds per square inch. This provided a nebulizer flow rate of 4.8-5.2 L/min. The bacterial aerosol was diluted with humidified air and delivered to the head-only chamber. Total chamber flow was regulated at 16 L/min and main exhaust was adjusted to maintain chamber pressure slightly negative (-0.1 to -0.3 inches water column) with respect to the Class III biosafety cabinet (a means to prevent inadvertent release of the aerosol). Samples were extracted from the head-only chamber to quantify the bioaerosol concentration (via AGI-4 sampling), size distribution (via 30-sec APS measurement) and relative particle concentration (via continuous DustTrak sampling). AGI sampling was performed at a flow rate of 5.0 ± 0.5 L/min during the aerosol trials. Following aerosol trials, AGI suspensions were transferred to the Microbiology laboratory for quantitative analyses as described below (Determination of F. tularensis Bioaerosol Parameters). The nebulizer and AGI flowrates were calibrated on the day of and prior to all of the aerosol trials. Page 13 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 14 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Sparging Generator The sparging generator is depicted in Figure 8: Figure 8. Sparging Generator. The sparging generator, similar in design to the Collison nebulizer, was used only in LVS testing throughout the milestone study period. The Sparging Generator (CH Technologies, Inc., Westwood, NJ) was delivered a F. tularensis suspension (preparation procedures described above) at a rate of 1.0 ± 0.1 L/min and operated by providing a carrier air flow of 8.0 L/min. This provided a generator flow rate of 8.0 ± 0.1 L/min. The bacterial aerosol was diluted with humidified air and delivered to the head-only chamber. Total chamber flow was regulated at 16 ± 0.2 L/min and main exhaust was adjusted to maintain chamber pressure slightly negative (-0.1 to -0.3 inches water column) with respect to the Class III biosafety cabinet (a means to prevent inadvertent release of the aerosol). Samples were extracted from the head-only chamber to quantify the bioaerosol concentration (via AGI-4 sampling), size distribution (via 30-sec APS measurement) and relative particle concentration (via continuous DustTrak sampling). AGI sampling was performed at flows of ranging from 5.1 to 5.2 L/min during the aerosol trials. Following aerosol trials, AGI suspensions were transferred to the Microbiology laboratory for quantitative analyses as described below (Determination of F. tularensis Bioaerosol Parameters). The generator and AGI flowrates were calibrated on the day of and prior to all of the aerosol trials. 3.1.7 Comparison of Pre- and Post-Spray F. tularensis Generator Suspension Concentrations In order to assess bacterial viability within the generator suspension, bacterial concentrations were analyzed prior to the bioaerosol run and compared to the titer in the residual generator Page 14 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 15 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 suspension following the test. This was performed only for the Aeromist, Hospitak, and Collison nebulizers. For each sample tested in this manner, suspensions were diluted into sterile 1% peptone to three final culture dilutions (dependent upon the initial generator concentration) using standard ten-fold microbial dilution procedures. Aliquots from each dilution (100 L) were plated in triplicate onto BCGA and incubated at 37°C ± 2°C for 48h ± 8h after which purity was determined and F. tularensis colonies were counted. Post-spray analyses were performed similarly and are described in the following section. 3.1.8 Determination of F. tularensis Bioaerosol Parameters Following each bioaerosol exposure, generator and impinger suspensions were aseptically aliquoted into 50 mL centrifuge tubes and transferred to the Microbiology laboratory for immediate processing. Briefly, suspensions were weighed to assess the volume generated during the bioaerosol run and diluted into sterile 1% peptone to three final culture dilutions (dependent upon the initial generator concentration) using standard ten-fold microbial dilution procedures. Aliquots from each dilution (100 L) were plated in triplicate onto BCGA and incubated at 37°C ± 2°C for 48h ± 8h after which purity was determined and F. tularensis colonies were counted. Using colony enumeration data, relevant data, including CFU/L of air and spray factors, were determined. Total CFU sprayed was determined using Equation 1: CFU Aerosolized = CFU/mL (post-bioaerosol generator suspension)*Vol sprayed (mL) CFU/mL (post-bioaerosol generator suspension) was calculated using Equation 2: CFU/mL = (Mean of plate CFU counts)*dilution factor The dilution factor was inclusive of the 10-fold dilution resulting from the 100 L plating volume. Volume sprayed (mL) was calculated using Equation 3: Vol sprayed (mL) = Initial weight (g) – Final weight (g) It was assumed that the specific gravity of the suspension fluid was equal to that of water. The CFU sprayed data were used to indicate the accuracy of the suspension preparation, concentration of total CFU sprayed during the bioaerosol run, and to provide insight into the bioaerosol stability during aerosolization. Page 15 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 16 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 CFU/L of air was calculated using Equation 4: CFU/L of air = total CFU recovered in the AGI (sample time [min]*flow rate [L/min]to the AGI) The total CFU recovered in the AGI was calculated using Equation 5: Total CFU recovered in AGI = Final AGI vol (mL)*CFU/mL CFU/mL represents the calculated CFU/mL of the AGI suspension following the bioaerosol run and was calculated using Equation 2. The final AGI volume (mL) was calculated using Equation 6: Final AGI vol (mL) = Initial vol (mL)-(Initial weight [g] – Final weight [g]) For each bioaerosol run, the initial volume aliquoted into the AGI was 20 mL. It was assumed that the specific gravity of the suspension fluid was equal to that of water. Sample times and flow rates to the AGI were recorded electronically using the LabView software. CFU/L data were used to establish the bacterial atmosphere concentration. The spray factor (SF), a unitless measurement defined as the ratio of the aerosol concentration to the starting concentration (in the generator suspension) was calculated using Equation 7: SF = (CFU/L)/(CFU/mL [post-bioaerosol generator suspension]*1000) The spray factor is primarily a function of the generator, the agent being tested and the system flow and is used to compare efficacy between bioaerosol runs and, for this milestone, generator types. Page 16 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Version: 2.0 Page 17 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 3.2 Critical Reagents Preparation of the critical reagents listed below are described in detail in SOP LBERI 4.1. Reagent Manufacturer Catalog # Agar n/a (components below) Remel n/a (components below) 451012 Peptone BD/Difco 211677 Meat Extract EMD/Merck 1.03979.0500 Sodium Chloride Sigma Aldrich S3014-500G L-histidine Sigma Aldrich H8000-25G Glucose Sigma Aldrich G5767-500G L-cysteine Sigma Aldrich 168149-25G Sterile Defibrinated Sheep Blood PML A0407 Sucrose Sigma Aldrich S0389-500G Peptone BD/Difco 211677 Antifoam A Sigma A5633-100G Brain Heart Infusion Broth Remel R452472 Chamberlains Medium Teknova C0711 BCGA 3.3 Accepted Date:4/23/09 Standard Operating Procedures SOP Number1 SOP Title LBERI-1.1 Preparation of Francisella tularensis Seed and Working Stocks LBERI-2.1 Preparation of Francisella tularensis for Bioaerosol Studies LBERI-3.1 Operation of the Aeromist, Hospitak, Collison 3-jet, Aeroneb® Micropump, and Sparging Generators During Francisella tularensis Bioaerosol Studies Page 17 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Version: 2.0 Page 18 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 SOP Number1 LBERI-4.1 Accepted Date:4/23/09 SOP Title Preparation of Reagents and Media for Francisella tularensis Studies 1 Individual Standard Operating Procedures will be reviewed separately and accepted by the Subcontracting PI and UNM PI 4 Salient Original Data, Results, Interpretation, Quality Control 4.1 4.1.1 Original Data and Results (Rationale, Tables/Figures with legends and location annotations) Comparison of Solid and Liquid Media for the Growth of F. tularensis Results from solid and liquid media optimization analyses using LVS are summarized in Tables 1-3: Table 1. Qualitative Comparison of Francisella tularensis LVS Growth on Various Solid Agar Media Media Qualitative Analysis Quadrant Colony Description CAH Poor growth 1 Circular, raised, entire, glistening, smooth, opaque, grey/white CHAH-2 Good growth 2 Circular, raised, entire, glistening, smooth, opaque, white CHAH-10 Fair growth 2 Circular, raised, entire, glistening, smooth, opaque, white/green/pink CHA Fair growth 1 Circular, raised, entire, glistening, smooth, opaque, white BCGA Fair growth 2 Circular, raised, entire, glistening, smooth, opaque, grey/white CHAB Good growth 3 Circular, raised, entire, glistening, smooth, opaque, grey/blue/white SBCGA Fair growth 2 Circular, raised, entire, glistening, smooth, opaque, grey/blue/white GCAT No growth n/a n/a CA No growth n/a n/a Page 18 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 19 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Table 1. Qualitative Comparison of Francisella tularensis LVS Growth on Various Solid Agar Media (see method summary in section 3.1.1). Results demonstrated that F. tularensis LVS qualitatively grew best on CHAB and CHAH-2, though fair growth was also noted on CHAH-10, CHA, BCGA and SBCGA. Colony morphologies were similar among the different agars. The predominant colony morphology difference between the media, when present, was in colony color. No growth was observed on GCAT or CA. From these data, the decision was made to proceed with consistent use of CHAB and secondary use of BCGA as the solid media for growth of LVS, the latter due to the ease of preparation versus other media types. Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\Working Group\F. tularensis media comparison.pdf Table 2. Quantitative Comparison of Francisella tularensis LVS Growth on Various Solid Agar Media % Recovery Media Type Mean CFU/mL SBCGA CHAH-2 CAH CHA BCGA CHAB CHAH-10 1.05E+08 2.00E+08 1.02E+08 2.50E+08 2.00E+08 2.00E+08 5.35E+07 SBCGA 1.05E+08 100% 53% 103% 42% 53% 53% 196% CHAH-2 2.00E+08 190% 100% 197% 80% 100% 100% 374% CAH 1.02E+08 97% 51% 100% 41% 51% 51% 190% CHA 2.50E+08 238% 125% 246% 100% 125% 125% 467% BCGA 2.00E+08 190% 100% 197% 80% 100% 100% 374% CHAB 2.00E+08 190% 100% 197% 80% 100% 100% 374% CHAH-10 5.35E+07 51% 27% 53% 21% 27% 27% 100% Table 2. Quantitative comparison of Francisella tularensis LVS growth on various solid agar media (see method summary in section 3.1.1). Here, each media type is compared as row versus column (numerator versus denominator, respectively). For instance, the percent recovery of LVS on SBCGA versus CHAH-2 was 53%, indicating that LVS quantitatively grew better on CHAH-2 among these two media types. Overall, peak recoveries were noted on CHAH-2, Page 19 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 20 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 CHA, BCGA, and CHAB. GCAT and CA were excluded from these tests as qualitative analyses demonstrated the absence of growth (Table 1). These data supported our decision to move forward with consistent use of CHAB and secondary use of BCGA as the solid media for growth of LVS. Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\Working Group\F. tularensis media comparison.pdf Table 3. Quantitative Comparison of Francisella tularensis LVS Growth in Various Nutrient Broth Media CFU/mL Broth Medium Flask #1 Flask #2 Average MCPH 0 (at 1:1000 dilution) 0 (at 1:1000 dilution) 0 (at 1:1000 dilution) MMHB 3.00E+05 1.20E+05 2.10E+05 CB 2.00E+07 2.00E+07 2.00E+07 CBH 3.00E+07 Not tested 3.00E+07 MCPHI 0 (at 1:1000 dilution) 2.00E+07 Inconclusive Table 3. Quantitative comparison of F. tularensis LVS growth in various liquid broth media (See method summary in section 3.1.1). Data demonstrated that bacteria grew optimally in CBH. However, this testing was not repeated and it was noted that CBH was difficult to prepare. Because of this, CB, the next best liquid broth medium based on quantitative LVS growth, was chosen for all subsequent testing. Though growth was noted in MMHB, and MCPHI, inconsistencies were observed and/or titers were significantly lower when compared to CB. No growth was observed in MCPH following 72h of incubation at 37°C. All broth suspensions were serially diluted and sub-cultured onto CHAB. Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\Working Group\F. tularensis media comparison.pdf 4.1.2 Generator and Impingement Solution Optimization Results from generator and impingement solution optimization testing using LVS are summarized in Figure 9: Page 20 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 21 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Francisella tularensis LVS Stability Testing: Effect of Time, Temperature, and Media on Viability and Culturability CFU/mL (log10) 9.5 9.0 8.5 8.0 7.5 7.0 0 1 2 3 4 Time (h) Chamberlains, RT Chamberlains, 4 deg C PBS, RT PBS, 4 deg C BHIB, RT BHIB, 4 deg C 1% peptone, RT 1% peptone, 4 deg C Figure 9. Effect of time, temperature, and media on F. tularensis LVS viability/culturability (See method summary in section 3.1.3). Bacterial suspensions in various liquid media were allowed to sit at room temperature or refrigerated for up to four hours. Aliquots were cultured hourly and the resulting data were used to evaluate the effects of media, time and temperature on LVS viability/culturability. Though results were similar among the groups, BHIB at room temperature demonstrated optimal viability and culturability maintenance. It was for this reason that BHIB was chosen as the final medium for bioaerosol material resuspension. All samples presented an increase in titer over the first two hours indicating bacterial proliferation upon initiation of the testing. Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\LVS Growth characterization\Stability\Effect of time and temp on LVS recovery_18April07.xls 4.1.3 Comparison of Pre- and Post-Spray F. tularensis Generator Suspension Concentrations Results from pre- and post-bioaerosol run generator suspension analyses using the Aeromist, Hospitak and Collison Nebulizers are presented in Figures 10-13 below: Page 21 of 62 5 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 22 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Figure 10. Effect of delivered pressure to the Aeromist Nebulizer on F. tularensis LVS viability/culturability (see method summary in section 3.1.6 and 3.1.7). During select bioaerosol runs, varied pressures ranging from 5.0 to 20.0 psig were delivered to the Aeromist while testing frozen and fresh LVS cultures. Qualitatively, results indicated that as pressure delivered to the generator increased, viability/culturability decreased following the bioaerosol runs. Statistically (paired t-test performed on log-transformed data), there was no difference between pre- and posttest generator suspensions using frozen LVS at 5.0 or 10.0 psig; however, due to poor bioaerosol output at 5.0 psig, 10.0 psig was chosen as the final delivered pressure parameter. Remaining data, consisting of fresh LVS tested at 10.0, 15.0, and 20.0 psig, could not be statistically analyzed due to insufficient replicates, though data suggested little or no difference between frozen and fresh bacteria. Continued testing was halted with LVS because of our clearance to work with SCHU S4, a critical step for this milestone. All NHP challenges were planned with SCHU S4 aerosols; however, LBERI staff started aerosol testing with LVS because they were waiting for USAMRIID to provide the LVS vaccinations to the LBERI staff. LBERI could only prepare LVS aerosols until LBERI staff were LVS vaccinated as another layer of protection from Page 22 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 23 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 SCHU S4 infection. After the LBERI staff were LVS vaccinated, the LBERI management allowed the LBERI staff to switch to SCHU S4 aerosols. Switching to SCHU S4 aerosols was more important than generating statistically significant aerosol data with LVS. Results obtained with the experiments presented above served as a baseline to initiate the SCHU S4 aerosol optimization. Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\FY06-078_TUL03_Cumulative MS3 data.xls Page 23 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 24 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Figure 11. Effect of delivered pressure to the Aeromist Nebulizer on F. tularensis SCHU S4 viability/culturability (see method summary in sections 3.1.6 and 3.1.7). During select bioaerosol runs, varied pressures (either 7.5 or 10.0 psig) were delivered to the Aeromist while testing frozen and fresh SCHU S4 cultures. SCHU S4 testing was performed following all LVS testing. We knew with LVS, that delivered pressures of 15 and 20 psi were too high and likely would destroy SCHU S4 as well. We also knew that 5 psi was too low; the low pressure was very gentle on the bacteria, but not enough to actually generate the aerosol. For LVS, we concluded 10 psi was optimal. We tested 7.5 psi for SCHU S4 because it was desirable to attempt to be slightly gentler to the SCHU S4 bacteria, but that pressure, like 5 psi, was not high enough to generate the aerosol. Fresh SCHU S4 was tested at 10 psi because that was concluded to be the optimal delivered pressure for aerosol generation and bioaerosol stability. Essentially, output was better with 10 psig; 7.5 psig was simply not enough to generate sufficient numbers of particulates. Page 24 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 25 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Statistically (paired t-test performed on log-transformed data), there was a difference between pre- and post-spray SCHU S4 suspensions (frozen and fresh) when tested at 7.5 or 10.0 psig, with values demonstrated better recovery (i.e., increased viability retention) when a fresh bacterial culture was used. From these data, we concluded that optimal SCHU S4 bioaerosols were achieved with the Aeromist using fresh bacteria and a delivered pressure of 10 psig. Animal exposures for ensuing studies could be conducted while observing these optimized procedures. Please note: Initial target CFU/mL values were purposely varied to see the effect of concentration on viability during the aerosol testing. Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\FY06-078_TUL03_Cumulative MS3 data.xls Page 25 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 26 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Figure 12. Effect of aerosol generation using the Hospitak Nebulizer on F. tularensis SCHU S4 viability/culturability (see method summary in section 3.1.6 and 3.1.7). The X-axis represents replicates performed on seven separate days. The Hospitak was viewed as a potential replacement for the Aeromist since production of the Aeromist was ceased by the manufacturer during the course of this milestone. During all bioaerosol runs, a standard pressure of 10.0 psig (to mimic the optimized procedures established for the Aeromist) was delivered to the Hospitak while testing fresh SCHU S4 cultures. Statistically (paired t-test performed on log-transformed data), there was no difference between pre- and post-test generator suspensions using fresh SCHU S4 at 10.0 psig. Following these experiments, it was decided that the Hospitak behaved very similarly to the Aeromist and could be used since LBERI could no longer purchase the Aeromist. Data location: \Saturn\absl3\Agent and Study Specific Data \STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\FY06-078_TUL-03_Cumulative MS3 data.xls Page 26 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 27 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Figure 13a. Effect of delivered pressure to the Collison Nebulizer on F. tularensis LVS and SCHU S4 viability/culturability (see method summary in section 3.1.6 and 3.1.7). X axis represents replicates. During all bioaerosol runs, pressure delivered to the Collison was maintained between 20 and 30 psig (as required per the Collison manufacturer) to establish a flow rate of 7.5 L/min. Pre- and post-generator suspension analyses were performed at a delivered pressure of approximately 25 psig. Statistically (paired t-test performed on log-transformed data), there was no difference between pre- and post-test generator suspensions using frozen LVS or SCHU S4 at 25.0 psig (though it should be noted that there were only three replicates of each). Analyses did, however, demonstrate a statistical difference between pre- and post-spray values using fresh SCHU S4. Data location: \\Saturn\absl3\Agent and Study Specific Data\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\FY06-078_TUL-03_Cumulative MS3 data.xls NOTE: The Collison required 20-30 psig per the manufacturer; however, LBERI tested lower psig for other aerosol generators with the goal of increasing bacterial viability with lower pressures. The Collison’s potential weaknesses were variability between different Collison units, the potential for the 3 Page 27 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 28 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 jets to become blocked, the glass construction as a safety risk in BSL3 environment, and potential suboptimal bacterial viability due to the higher required pressures. Figure 13b (continuation of 13a). Effect of delivered pressure to the Collison Nebulizer on fresh F. tularensis SCHU S4 viability/culturability (see method summary in section 3.1.6 and 3.1.7). During all bioaerosol runs, pressure delivered to the Collison was maintained between 20 and 30 psig to establish a flow rate of 7.5 L/min. Pre- and post-generator suspension analyses were performed at a delivered pressure of approximately 25 psig. Statistical analyses are described above under Figure 13a. From the experimental data presented in Figures 13a and 13b, it was decided that the Collison, while operated under manufacturer specifications using fresh SCHU S4, produced results similar to those optimal for the Aeromist and Hospitak. Essentially, based on viability retention alone (i.e., pre- versus post-bioaerosol generator suspension cultures), the Aeromist and Hospitak operated at a delivered pressure of 10 psig and Page 28 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 29 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 the Collison operated at a delivered pressure of 25 psig demonstrated no significant differences with fresh SCHU S4 testing. Thus, after months of testing multiple aerosol generators, LBERI in consultation with NIAID and UNM chose the field standard Collison aerosol generator for all future NHP challenges with fresh SCHU S4. Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\FY06-078_TUL03_Cumulative MS3 data.xls 4.1.4 Bioaerosol Data All F. tularensis bioaerosol optimization tests were performed first with LVS under the premise that LVS data would linearly predict the SCHU S4 data (i.e., it was hypothesized that the two strains would behave the same as bioaerosols). Also, LBERI initially could work only with LVS and not SCHU S4, until the LBERI staff received the LVS vaccination. LBERI management would not allow LBERI to work with SCHU S4 aerosols, until the LBERI staff were vaccinated. UNM, LBERI and NIAID agreed that LBERI should start with LVS aerosols until SCHU S4 aerosols were permitted by the LBERI management.Results from these initial studies were used to down-select to optimal conditions for SCHU S4 bioaerosol testing. These data are summarized in following graphs and tables. Frozen F. tularensis LVS Page 29 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 30 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Spray Factors vs CFU/L: Frozen Francisella tularensis LVS Testing with the Aeromist, Collison, Micropump and Sparging Nebulizers 6.0 CFU/L (log10) 5.0 4.0 3.0 2.0 1.0 0.0 -4.5 -1.0 -5.5 -6.5 -7.5 -8.5 -9.5 -10.5 Spray Factor (log10) Aeromist, 5 psi Aeromist, 10 psi Aeromist, 15 psi Aeromist, 20 psi Collison, 20 psi Collison, 25 psi Micropump, 20 psi Sparging Figure 14. Cumulative spray factor versus aerosol concentration data using frozen F. tularensis LVS and the Aeromist, Collison, Micropump and Sparging Nebulizers (see method summary in sections 3.1.6, 3.1.7, and 3.1.8). Data are presented as log10 values. Tested nebulizers and associated delivered pressures are presented in the accompanying legend. A decrease in spray factor is proportional to a decrease in bioaerosol efficiency (i.e., bioaerosol efficiency decreases from left to right). These data demonstrate that the Aeromist operated at a delivered pressure of 10-20 psi performed most consistently and efficiently with frozen LVS among the four tested nebulizers. The Collison operated under normal operating conditions (20-25 psi) performed well overall, but presented with some poor spray factors. The Micropump and Sparging generators were least efficient with frozen LVS under the specified operating parameters. Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\FY06-078_TUL03_Cumulative MS3 data.xls Page 30 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 31 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Table 4. Frozen LVS Testing with the Aeromist, Collison, Micropump, and Sparging Nebulizers: Relationship between Delivered Pressure, Aerosol Concentration, and Spray Factor Generator Pressure psig N CFU/L Spray Factor Geo Mean Geo SD Geo Mean Geo SD Aeromist 5 6 3.26E+02 9.78E+00 5.98E-08 1.58E+00 Aeromist 10 3 1.08E+03 1.50E+00 3.38E-07 1.28E+00 Aeromist 15 2 1.07E+03 2.11E+00 2.00E-06 2.54E+00 Aeromist 20 30 1.70E+02 9.26E+00 4.61E-07 2.62E+00 Collison 20 8 4.80E+02 4.78E+01 4.61E-07 1.78E+00 Collison 25 18 7.96E+01 3.60E+01 1.96E-07 1.44E+01 Micropump 20 11 2.67E+01 6.21E+00 3.34E-08 1.48E+01 Sparging n/a** 12 6.55E+01 8.98E+00 6.01E-09 8.17E+00 *N, number of replicates **n/a, not applicable as carrier air was supplied to the Sparging for bioaerosol generation Table 4. Summary spray factor and aerosol concentration data using frozen F. tularensis LVS and the Aeromist, Collison, Micropump and Sparging Nebulizers. Data are presented as geometric means and standard deviations (see method summary in section 3.1.6, 3.1.7, 3.1.8). These data are presented graphically in Figure 14. For this milestone, adjustments based on available data were made by the TVDC and LBERI teams in real-time as to how to proceed. Replicates were not originally planned with statistical analyses in mind, rather, decisions were made to conduct further testing on generator/pressure combinations that demonstrated promising results with respect to bioaerosol optimization. It should be noted here, that further testing was ceased with LVS following LBERI’s clearance to work with SCHU S4 (again, a critical step for this milestone); this is reflected in the small number of LVS replicates at 5, 10, and 15 psig delivered for the Aeromist as well as 20 psig delivered to the Collison. Once the LBERI staff were allowed to use SCHU S4 aerosols, LVS aerosol testing ceased Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\FY06-078_TUL-03_Cumulative MS3 data.xls Page 31 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 32 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Fresh F. tularensis LVS Spray Factors vs CFU/L: Fresh Francisella tularensis LVS Testing with the Aeromist, Collison, and Micropump Nebulizers 6.0 CFU/L (log10) 5.0 4.0 3.0 2.0 1.0 0.0 -4.5 -5.0 -5.5 -6.0 -6.5 -7.0 -7.5 -8.0 -8.5 -9.0 Spray Factor (log10) Aeromist, 10 psi Aeromist, 15 psi Collison, 30 psi Micropump, 20 psi Aeromist, 20 psi Collison, 20 psi Collison, 25 psi Figure 15. Cumulative spray factor versus aerosol concentration data using fresh F. tularensis LVS and the Aeromist, Collison, and Micropump Nebulizers (see method summary in section 3.1.6, 3.1.7, 3.1.8). The Sparging generator was removed from further testing shortly following frozen LVS experiments due to poor results in conjunction with the difficulties encountered during setup in the LBERI ABSL-3 Facility; because of these factors, it was not pursued for further fresh LVS testing. Setting up the Sparging generator made it technically difficult for the personnel working the constrained space in the ABSL- 3 conditions so LBERI did not pursue the Sparging generator with fresh LVS. Data are presented as log10 values. Tested nebulizers and associated delivered pressures are presented in the accompanying legend. A decrease in spray factor is proportional to a decrease in bioaerosol efficiency (i.e., bioaerosol efficiency decreases from left to right). These data demonstrate that the Aeromist operated at a delivered pressure of 10-20 psi performed most consistently and efficiently with fresh VS among the three tested nebulizers. The Collison operated under normal operating conditions (20-30 psi) performed consistently, but was less efficient than the Aeromist. The Micropump performed poorly with fresh LVS under the specified operating parameters. Data location: Page 32 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 33 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\FY06-078_TUL-03_Cumulative MS3 data.xls Page 33 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 34 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Table 5. Fresh LVS Testing with the Aeromist, Collison, and Micropump Nebulizers: Relationship between Delivered Pressure, Aerosol Concentration, and Spray Factor Generator Pressure psig N* CFU/L Spray Factor Geo Mean Geo SD Geo Mean Geo SD 2 2.38E+03 1.32E+00 7.36E-07 1.70E+00 Aeromist 10 Aeromist 15 2 1.67E+03 1.56E+00 1.71E-06 1.80E+00 Aeromist 20 20 6.62E+02 5.55E+00 3.79E-06 2.59E+00 Collison 20 3 5.96E+03 1.43E+00 3.75E-07 1.32E+00 Collison 25 21 1.29E+03 7.48E+00 4.16E-07 2.44E+00 Collison 30 9 4.11E+00 2.17E+00 2.09E-07 2.14E+00 Micropump 20 12 9.37E+01 1.06E+01 2.67E-08 1.02E+01 *N, number of replicates Table 5. Summary spray factor and aerosol concentration data using fresh F. tularensis LVS and the Aeromist, Collison, and Micropump Nebulizers. Data are presented as geometric means and standard deviations (see method summary in section 3.1.6, 3.1.7, 3.1.8). These data are presented graphically in Figure 15. For this milestone, adjustments based on available data were made by the TVDC and LBERI teams in real-time as to how to proceed. Replicates were not originally planned with statistical analyses in mind, rather, decisions were made to conduct further testing on generator/pressure combinations that demonstrated promising results with respect to bioaerosol optimization. It should be noted here, that further testing was ceased with LVS following LBERI’s clearance to work with SCHU S4 (again, a critical step for this milestone); this is reflected in the small number of replicates at 10 and 15 psig delivered for the Aeromist as well as 20 psig delivered to the Collison. Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\FY06-078_TUL-03_Cumulative MS3 data.xls Frozen F. tularensis SCHU S4 Page 34 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 35 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Spray Factors vs CFU/L: Frozen Francisella tularensis SCHU S4 Testing with the Aeromist and Collison Nebulizer 6.0 CFU/L (log10) 5.0 4.0 3.0 2.0 1.0 0.0 -4.5 -5.0 -5.5 -6.0 -6.5 -7.0 -7.5 -8.0 Spray Factor (log10) Aeromist, 7.5 psi Aeromist, 10 psi Collison, 25 psi Figure 16. Cumulative spray factor versus aerosol concentration data using frozen F. tularensis SCHU S4 and the Aeromist and Collison Nebulizers. Data are presented as log10 values (see method summary in section 3.1.6, 3.1.7, 3.1.8). Tested nebulizers and associated delivered pressures are presented in the accompanying legend. A decrease in spray factor is proportional to a decrease in bioaerosol efficiency (i.e., bioaerosol efficiency decreases from left to right). These data demonstrate that the Aeromist operated at a delivered pressure of 10 psi performed best with frozen SCHU S4 among the four tested nebulizers. The Collison operated under a normal operating condition of 25 psi performed similar to the Aeromist at 10 psi. The Aeromist operated at a delivered pressure of 7.5 psi performed poorly with frozen SCHU S4 likely due to the inability to effectively generate particles at this low pressure. Replicates were minimal for frozen SCHU S4 testing based on data from frozen LVS testing and due to the knowledge that fresh cultures were chosen for future bioaerosol optimization. Data location: \Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL03\TUL-03 data files\FY06-078_TUL-03_Cumulative MS3 data.xls Page 35 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 36 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Table 6. Frozen SCHU S4 Testing with the Aeromist and Collison Nebulizers: Relationship between Delivered Pressure, Aerosol Concentration, and Spray Factor Generator Pressure psig N* CFU/L Spray Factor Geo Mean Geo SD Geo Mean Geo SD Aeromist 7.5 3 1.86E+01 1.51E+00 1.02E-07 1.45E+00 Aeromist 10 3 1.74E+02 1.78E+00 1.09E-06 1.82E+00 Collison 25 3 9.63E+01 1.56E+00 6.22E-07 1.86E+00 *N, number of replicates Table 6. Summary spray factor and aerosol concentration data using frozen F. tularensis SCHU S4 and the Aeromist and Collison Nebulizers. At this point in the milestone, TVDC and LBERI teams narrowed further bioaerosol testing down to the Aeromist and Collison nebulizers. This decision was based on the overall good viability retention of both units in conjunction with efficient spray factors. Here, the number of replicates (n=3) was small because it was concluded previously that frozen bacteria as a starting generator suspension material was suboptimal for bioaerosols. Data are presented as geometric means and standard deviations (see method summary in section 3.1.6, 3.1.7, 3.1.8). These data are presented graphically in Figure 16. Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\FY06-078_TUL-03_Cumulative MS3 data.xls Page 36 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 37 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Fresh F. tularensis SCHU S4 Figure 17. Cumulative spray factor versus aerosol concentration data using fresh F. tularensis SCHU S4 and the Aeromist, Hospitak, and Collison Nebulizers. Data are presented as log10 values (see method summary in section 3.1.6, 3.1.7, 3.1.8). Tested nebulizers and associated delivered pressures are presented in the accompanying legend. A decrease in spray factor is proportional to a decrease in bioaerosol efficiency (i.e., bioaerosol efficiency decreases from left to right). Here, testing with the Hospitak nebulizer was initiated due to the termination of Aeromist production by the manufacturer (described previously). These data demonstrate that that the Aeromist, Hospitak and Collison nebulizers performed consistently with one another, a result that, based on stability data, was expected. Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\FY06-078_TUL-03_Cumulative MS3 data.xls Page 37 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 38 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Fresh SCHU S4 Testing with the Aeromist, Hospitak, and Collison Nebulizers: Relationship between Delivered Pressure, Aerosol Concentration, and Spray Factor Generator Pressure psig N* CFU/L Spray Factor Geo Mean Geo SD Geo Mean Geo SD Aeromist 10 13 1.21E+02 8.89E+00 1.14E-07 2.27E+00 Hospitak 10 27 1.73E+03 1.60E+01 1.11E-07 4.03E+00 Collison 25 35 1.03E+03 1.54E+01 1.08E-07 2.41E+00 *N, number of replicates Table 7. Summary spray factor and aerosol concentration data using fresh F. tularensis SCHU S4 and the Aeromist, Hospitak, and Collison Nebulizers. Data are presented as geometric means and standard deviations (see method summary in section 3.1.6, 3.1.7, 3.1.8). These data are presented graphically in Figure 17. Replicates for the Aeromist operated at a delivered pressure of 10 psig were minimal due to ceased production of the unit by the manufacturer. From these results, the decision was made that either the Hospitak or Collison could be used for animal challenges, though more weight was placed on the Collison due to its historical use, published literature, and consistent product availability over time from the manufacturer. Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\FY06-078_TUL-03_Cumulative MS3 data.xls Particle Size Data Table 8 summarizes particle size data, as determined using the APS, for the Aeromist, Collison, and Micropump nebulizers. Due to poor aerosol production, similar data were not obtained for the sparging generator. Note: Particle size is inherent to the device; delivered pressure has very little effect on the size, particularly at the pressure ranges utilized. Additionally, most of the particles generated are from the diluent itself and not the bacterium; therefore, particle size is not dependent on the use of LVS vs. SCHU S4. Table 8. Particle Size Data for the Aeromist, Hospitak, Collison, and Micropump Nebulizers MMAD Nebulizer m GSD Page 38 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Version: 2.0 Page 39 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Aeromist 1.60 1.56 Hospitak 2.22 1.71 Collison 1.49 1.57 Micropump 1.49 1.67 Accepted Date:4/23/09 Table 8. Particle size data for the Aeromist, Hospitak, Collison, and Micropump Nebulizers. During bacterial aerosol generation, representative samples were measured using an aerodynamic particle sizer (APS), a device that, through a sophisticated laser system, measures the drop sizes produced by the generator. Briefly, during testing, a 30-second sample from the exposure line was pulled through the instrument; readings were automatic and the device was calibrated using monodisperse polystyrene latex particles prior to use. Results demonstrate similar performance between the four nebulizers, though the Hospitak generated slightly larger particles (P<0.05, ttest). For our purposes, the acceptable size was defined as < 4 m MMAD since particles in this range demonstrate high deep lung deposition rates. For these types of tests, it should be noted that particle size is inherent to the device; delivered pressure has very little effect on the size, particularly at the pressure ranges we worked within. Additionally, most of the particles generated are from the diluent itself and not the bacterial presence; therefore, particle size during these tests was not dependent on the use of LVS versus SCHU S4. MMAD, mass median aerodynamic diameter; GSD, geometric standard deviation. Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06078_TUL-03\TUL-03 data files\FY06-078_TUL-03_Summary particle size data.xls 4.2 4.2.1 Interpretation Comparison of Solid and Liquid Media for the Culture of F. tularensis Two solid media failed to support the growth of LVS (GCAT and CA). CAH supported only poor growth of LVS. Several media, CHAH-2, CHAH-10, CHA, and BCGAH) presented fair to good growth of LVS (qualitatively based on morphology, size, and concentration), but the colony morphology was atypical primarily due to coloration; specifically, colonies were small, in few number, and grew pink, green, and brown rather than the expected blue/gray pigmentation. Several solid media provided fair to good growth of LVS with expected colony morphology (BCGA, SBCGA, and CHAB). SBCGA, BCGA, and CHAB also supported good percent recovery of LVS when compared to other media types. Page 39 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 40 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Based on these studies alone, it was concluded that CHAB was the optimal solid agar medium for F. tularensis LVS growth. In studies conducted post-optimization of agar medium, however, it was noted that F. tularensis growth on CHAB was inconsistent, when present, and growth took a significant amount of time (minimum of 72h). Certain lots of CHAB supported growth when others would not, even though preparation and quality control procedures were consistent. Taken together, CHAB was concluded to be sub-optimal in context of data management. Contrary to this, growth on BCGA was consistently present following 48h of incubation (Figure 18). Based on these observations, LBERI decided to use BCGA for the majority of the bioaerosol optimization testing (including bacterial proliferation and generator/AGI suspension cultures). Figure 18. Growth of F. tularensis LVS on CHAB (left) and BCGA (right). As depicted here, LVS growth was optimal on BCGA after 48h of incubation (see method summary in section 3.1.1) Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06\FY06-078_TUL-03\LVS Growth characterization\48h on CHAB.jpeg and \48h on BCGA.jpeg MCPH broth provided poor growth of LVS; less than 1000 CFU/mL was observed. MMHB grew LVS to approximately 2.1 x 105 CFU/mL. Bacterial titers in three of the five media (CB, CBH, and MCPHI) were greater than 1.0 x 107 CFU/mL. One flask of MCPHI demonstrated no growth while the second presented a bacterial titer greater than 1.0 x 107 CFU/mL. The reasons for this were inconclusive. Because of the ease of preparation for CB and CBH as Page 40 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 41 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 compared to MCPHI (i.e., one or two components, respectively, versus the seven needed for MCPHI), the use of CB or CBH was concluded to be optimal. CB and CBH supported near identical F. tularensis LVS growth characteristics; however, CB appeared as a clear solution which allowed for easy optical density titer characterization. It was therefore decided that CB was the optimal broth medium for F. tularensis LVS culture and CB would be used for SCHU S4 growth, until SCHU S4 growth optimization is completed under Milestone 9, the Qualification of SCHU S4 bioaerosols. 4.2.2 Generator and Impingement Solution Optimization Based on LVS viability and culturability, data from these studies demonstrated that the optimal medium and storage conditions for use during F. tularensis (LVS tested; SCHU S4 assumed) bioaerosol studies was brain heart infusion broth at room temperature.(see Figure 9) BHIB is a nutrient-rich medium capable of maintaining bacterial viability more than minimal media such as PBS and 1% peptone. Though Chamberlain’s broth contains sufficient nutrients to allow for bacterial growth, it did not demonstrate the capacity to stabilize a F. tularensis suspension under sub-optimal conditions. During bacterial aerosolization, the liquid droplet in which a bacterium is contained significantly desiccates to the point where the resulting particulate delivered to the exposure chamber is likely composed only of the organism and the dried remnants of the suspension medium. It is hypothesized that a nutrient-rich medium such as BHIB forms a protective coating on the bacterium upon desiccation. Solutions such as PBS and 1% peptone may not do this to the same degree. BHIB was chosen and was used for the generator and impinger solutions for all ensuing bioaerosol tests throughout Milestone 3. 4.2.3 Comparison of Pre- and Post-Spray F. tularensis Generator Suspension Concentrations Production of a bioaerosol via compressed air generation, as was performed throughout these studies, places a high degree of stress on bacteria within a suspension. During aerosolization, significant shear forces are generated as a result of collisions between bacterial particles and between bacteria and the exposure system (i.e., the generator walls, exposure line, etc.). Though these forces are unavoidable, our hypothesis was that they could be minimized in order to optimize bacterial viability and, consequently, the bioaerosol. The objective was to screen multiple bioaerosol generators for reproducible, bacterial viability in the pre- and post-sprays. The decision to conduct these experiments in Milestone 3 was not finalized until generator down-selection for SCHU S4 bioaerosol optimization was made to the Aeromist, Hospitak, and Collison Nebulizers. This decision was based primarily on spray factors, repeatability and ease of use as discussed in the sections below. Results from these tests demonstrated that bacterial viability/culturability in the Aeromist suspensions was highly dependant upon pressure delivered to the generator; specifically, as delivered pressure increased, bacterial viability in the Page 41 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 42 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 suspension decreased. The Hospitak was tested only at a delivered pressure of 10 psig; however, because of its similarity to the Aeromist, it can be hypothesized that results would be similar at higher delivered pressures. This phenomenon was not observed in the Collison for which a relatively high pressure (25 psig) was tested. Though bacterial viability and culturability decreased after each Collison run, it was not as marked as observed with the Aeromist even though a higher pressure was used to generate the bioaerosol with the Collison (a requirement as specified by the Collison manufacturer). The observed difference between the two types of generators was likely due to the manufacturing of the units; the Aeromist and Hospitak nebulizers utilized one jet for delivery of compressed air whereas the Collison evenly distributed the supplied air across three jets. This distribution decreased the shear forces during operation. To clarify, each jet of the 3-jet Collison MRE exhibited 1/3 of the total pressure provided to the unit. Results from these tests demonstrated that decreased pressure delivery to the aerosol generators increased bacterial viability. Though valuable information, this attribute cannot predict how the bioaerosol will performed since many other factors such as come into play following bacterial nebulization. This is discussed in the following section. 4.2.4 Bioaerosol Optimization Frozen F. tularensis LVS Based on spray factor values, data from these studies demonstrated that optimal bioaerosol performance for frozen LVS was established with the Aeromist operating at a delivered pressure of 15 psig. The geometric mean calculated spray factor under these parameters was 2.00 x 10-6. Further interpretation, however, indicated that these data were misleading. As discussed in the previous section, a significant decrease in viability/culturability was observed in the Aeromist generator suspension at a delivered pressure of 15 psig. The spray factor is calculated using post-bioaerosol run concentrations; a decreased titer falsely improves the spray factor that would not otherwise be noted if very little or no change in titer were observed. Though the calculated spray factor was optimal, the viability of the bioaerosol was sub-optimal. Taking immune responses into consideration, it was undesirable to generate a bioaerosol containing an increased concentration of non-viable (and potentially immune-stimulating) bacteria at the expense of a better spray factor. The Aeromist operating at a delivered pressure of 10 psig did not demonstrate this significant decrease in bacterial viability and presented spray factor values comparable to the Collison at 20-25 psig. Poor bioaerosol parameters were observed when testing the Aeromist at 5 psig. The geometric mean spray factor for these experiments was 5.98 x 10-8. Though bacterial viability was maintained in the generator suspension, the delivered pressure was not sufficient to physically aerosolize the agent, thereby negating this approach for further F. tularensis bioaerosol optimization testing. The Micropump and Sparging generators demonstrated poor bioaerosol efficacy when testing frozen LVS. Geometric mean spray factor values were 3.34 x 10-8 and 6.01 x 10-9 for the two units, respectively. Taken together, data indicated that the Aeromist operating under a delivered Page 42 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 43 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 pressure of 10 psig was optimal for generation of frozen LVS. The Collison was comparable, but several poor spray factors were noted for unknown reasons. In addition, we noticed early on in the milestone that the 3-jet nozzle of the Collison required routine cleaning to prevent clogging, an aspect that was deemed as less desirable for optimized testing. Fresh F. tularensis LVS Overall, data from these studies demonstrated that fresh LVS performed more efficiently as a bioaerosol when compared to spray suspensions prepared from frozen LVS stock vials. Based on spray factor values, the optimal bioaerosol performance was established with the Aeromist operating at a delivered pressure of 20 psig. However, as discussed previously, this pressure was concluded to be unacceptable due to the significant drop in viability/culturability in the generator suspension. As was observed with frozen LVS testing, the Aeromist at 10 psig and the Collison at 20-25 psig resulted in improved bioaerosol parameters (i.e., spray factors were more efficient and viability retention was increased ). Geometric mean spray factor values were 7.36 x 10-7 and 4.16 x 10-7 for the two units, respectively. The Micropump nebulizer performed poorly demonstrating a geometric mean spray factor of 2.67 x 10-8. Taken together, data indicated that the Aeromist and Collison nebulizers operating under delivered pressures of 10 and 25 psig respectively were optimal for generation of fresh LVS. There are no defined criteria for what is considered to be an optimal/acceptable spray factor. LBERI considers any spray factor above 1 x 10-7 to be acceptable; values greater than or equal to 1 x 10-6 could be considered optimal, though this is subjective and will vary between microbes and bioaerosol procedures. Values less than 1 x 10-7 (i.e., in the 1 x 10-8 range) are generally considered to be poor spray factors, again depending on the organism and aerosol parameters. Frozen F. tularensis SCHU S4 Results from these studies demonstrated that frozen SCHU S4 performed more efficiently (based on spray factor values) as a bioaerosol when compared to frozen LVS. Optimal bioaerosol performance was established with the Aeromist operating at a delivered pressure of 10 psig. The geometric mean spray factor value was 1.09 x 10-6. When a lower delivered pressure was tested (7.5 psig), the performance decreased significantly from a geometric mean spray factor of 1.09 x 10-6 to 1.02 x 10-7. This was due to the same reason that poor spray factors were observed when testing frozen LVS at 5 psig: efficient bacterial aerosolization was not established at this low-end parameter. The Collison, at 25 psig, performed slightly less efficiently than the Aeromist with respect to spray factors, though better than frozen LVS under the same conditions. The associated geometric mean spray factor value was 6.22 x 10-7. Page 43 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 44 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Fresh F. tularensis SCHU S4 Results from these studies demonstrated that fresh SCHU S4 performed less optimally (i.e., spray factors were lower overall) as a bioaerosol when compared to fresh LVS. The reasons for this are currently unknown, but can be hypothesized to be due to structural differences between the two strains. Based on spray factor values, there was no difference between the Aeromist, Hospitak and Collison nebulizers operated at 10, 10, and 25 psig, respectively, with fresh SCHU S4. Additionally, data demonstrated consistency between the three units under these operating conditions. The geometric mean spray factor for all three units was approximately 1.1 x 10-7. Particle Size Data Particle size data indicated that for all tested nebulizers, the droplet size generated (1.49 to 2.2 m) during bioaerosol runs was well within the respirable range (less than 4.0 m MMAD). Conclusions Based on cumulative results from LVS and SCHU S4 testing, the following parameters were concluded to be optimal for F. tularensis SCHU S4 growth, bioaerosol generation, delivery, maintenance, collection, and analysis: Challenge material (SCHU S4) will be cultured in Chamberlain’s broth medium as follows: 24h at 37°C, 200 rpm, in the dark. The broth will be inoculated with a normalized suspension (OD600 = 0.100 ± 0.010) of harvested SCHU S4 from a 48h BCGA culture grown at 37°C. Because a significant titer drop (> 1log10) was observed in the frozen stocks after six months of -80°C storage, the use of fresh cultures was optimal for bioaerosol studies. The Collison 3-jet Nebulizer will be used to generate F. tularensis SCHU S4 in challenge studies. o Bacteria will be diluted into sterile BHIB for the generator suspension o An output flowrate of 7.5 L/min will be maintained during bioaerosol runs. Data from Milestone 3 demonstrated that this was, in general, established at a delivered pressure of 25 psig for the Collison generator. Significant deviation from this pressure may be indicative of a defective/clogged aerosol jet; if such a deviation is observed, a new jet or new Collison unit will be employed. Page 44 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 45 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 o 4.3 When compared to the Aeromist and Hospitak nebulizers, the Collison proved to be slightly less efficient and less cost-effective, but overall, more consistent and reliable. Additionally, the availability of historical data for the Collison far outweighs bioaerosol optimization data for the Aeromist and Hospitak nebulizers. Furthermore, the Aeromist is no longer commercially available. For nonhuman primate exposures, F. tularensis bioaerosols will be delivered to the head-only chamber via a 2-inch diameter, 24-inch length stainless steel inlet line as originally described by Henderson, 1952. Relative humidity of the exposure system will be maintained at 60-90%. The bioaerosol will be collected into BHIB plus antifoam A using the AGI-4 (All Glass Impinger). The flow to the impinger will be maintained at 5.0-5.5 liters per minute. Following bioaerosol runs, bacterial suspensions will be cultured onto BCGA plates to enumerate the viable CFU. F. tularensis SCHU S4 proved to grow more consistently and in a shorter period time on this medium versus other solid agar such as CHAB. Quality Control Use of Controls o Bioaerosol runs: Prior to generation of bacteria for each day of testing, sterile BHIB was aerosolized for 10 minutes and collected as described throughout this report in order to verify sterility of the exposure line and aerosol equipment (i.e., nebulizers, delivery lines, AGIs, etc.). This was defined as the negative control. Microbiological Media o Quality control was performed on all prepared media according to SOP LBERI 4.1. Briefly, a representative sample of each media batch was allowed to incubate at 37°C in a dark, humidified incubator for a minimum of 48h. Following the allotted time, media were qualitatively assessed for the presence of contaminants (e.g., turbidity, mycelia formation, etc.). All testing included the following: In-house prepared culture plates Broth Dilution blanks Water Antifoam A Page 45 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 46 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 o 5 Certificates of Analysis were received and filed for purchased media. Any batch suspected of being compromised (i.e., cracked plates), were not used. The manufacturer was immediately notified of such findings. Replicates and Statistical Analysis o When possible, bioaerosol testing with the different aerosol generator/F. tularensis strain (i.e., LVS or SCHU S4) combinations was replicated a minimum of three times over separate days between different scientists. Instances for which this approach was not followed were due to time and/or supply constraints, personnel availability, and/or at the request of the Study Director. For the latter, the Study Director made decisions primarily based on poor testing results or other factors such as difficulty in aerosol generator handling. o Statistical analyses were performed using Prism (GraphPad Software, Version 5.01). When necessary, data were transformed (e.g., log10 transformation) such that parametric calculations could be performed. Reproducible bioaerosols were defined as having no significant (p<0.05) difference between replicates with respect to aerosol concentrations and spray factors. o Acceptable bioaerosol criteria for GLP studies have yet to be defined. This will be addressed in Milestone 9 of the Tularemia Vaccine Development Contract in which F. tularensis SCHU S4 bioaerosols generated and sampled in the LBERI head-only primate exposure system will be validated. Deliverables Completed There are no reagent deliverables associated with Milestone 3. The MS#3 MSCR and 4 protocols (LBERI 1.1, LBERI 2.1, LBERI 3.1 and LBERI 4.1) for generating optimal aerosolization are the deliverables. Page 46 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Version: 2.0 Page 47 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 6 6.1 Accepted Date:4/23/09 Appendices Appendix 1: Original Data Tables and Figures Table/ Figure1 Title Notebook Location2 (Notebook # and page numbers) Electronic Location2 (Full Path & File Name) T-1 Qualitative Comparison of Francisella tularensis LVS Growth on Various Solid Agar Media n/a \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL03 data files\Working Group\F. tularensis media comparison.pdf T-2 Quantitative Comparison of Francisella tularensis LVS Growth on Various Solid Agar Media n/a \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL03 data files\Working Group\F. tularensis media comparison.pdf Page 47 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Version: 2.0 Page 48 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Table/ Figure1 Title Notebook Location2 (Notebook # and page numbers) Accepted Date:4/23/09 Electronic Location2 (Full Path & File Name) T-3 Quantitative Comparison of Francisella tularensis LVS Growth in Various Nutrient Broth Media n/a \\Saturn\absl3\Agent and Study Specific Data\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL03 data files\Working Group\F. tularensis media comparison.pdf T-4 Summary spray factor and aerosol concentration data using frozen F. tularensis LVS and the Aeromist, Collison, Micropump and Sparging Nebulizers n/a \\Saturn\absl3\Agent and Study Specific Data \STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL03 data files\FY06-078_TUL03_Cumulative MS3 data.xls [Worksheet entitled “Summary Tables”] T-5 Summary spray factor and aerosol concentration data using fresh F. tularensis LVS and the Aeromist, Collison, and Micropump Nebulizers n/a \\Saturn\absl3\Agent and Study Specific Data \STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL03 data files\FY06-078_TUL03_Cumulative MS3 data.xls [Worksheet entitled “Summary Tables”] T-6 Summary spray factor and aerosol concentration data using frozen F. tularensis SCHU S4 and the Aeromist and Collison Nebulizers n/a \\Saturn\absl3\Agent and Study Specific Data \STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL03 data files\FY06-078_TUL03_Cumulative MS3 data.xls [Worksheet entitled “Summary Tables”] T-7 Summary spray factor and aerosol concentration data using fresh F. tularensis SCHU S4 and the Aeromist, Hospitak, and Collison Nebulizers n/a \\Saturn\absl3\Agent and Study Specific Data \STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL03 data files\FY06-078_TUL03_Cumulative MS3 data.xls [Worksheet entitled “Summary Tables”] T-8 Particle size data for the Aeromist, Hospitak, Collison, and Micropump n/a \\Saturn\absl3\Agent and Study Specific Data \STUDY SPECIFIC Page 48 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Version: 2.0 Page 49 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Table/ Figure1 Title Notebook Location2 (Notebook # and page numbers) Nebulizers Accepted Date:4/23/09 Electronic Location2 (Full Path & File Name) DATA\FY06-078_TUL-03\TUL03 data files\FY06-078_TUL03_Summary particle size data.xls F-1 Standard microbiological quadrant streak method for bacterial cultures n/a \\Saturn\absl3\Agent and Study Specific Data \STUDY SPECIFIC DATA\FY06-078_TUL03\Quadrant_streak_method_imag e F-2 Francisella tularensis LVS growth on BCGA n/a \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06\FY06-078_TUL03\LVS Growth characterization\48h on BCGA.jpeg F-3 LBERI NHP Head-Only Exposure System n/a \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06\FY06-078_TUL03\TUL-03 data files\Exposure line schematic.ppt F-4 Aeromist Nebulizer n/a \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06\FY06-078_TUL03\TUL-03 data files\Aeromist nebulizer\Aeromist_nebulizer_ima ge F-5 Collison 3-jet Nebulizer n/a \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06\FY06-078_TUL03\TUL-03 data files\Collison generator\Collison_nebulizer_ima ge F-6 Aeroneb® Lab Micropump Nebulizer n/a \\Saturn\absl3\Agent and Study Page 49 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Version: 2.0 Page 50 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Table/ Figure1 Title Notebook Location2 (Notebook # and page numbers) Accepted Date:4/23/09 Electronic Location2 (Full Path & File Name) Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06\FY06-078_TUL03\TUL-03 data files\Micro Pump\Micropump_nebulizer_imag e F-7 Hospitak Nebulizer n/a \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06\FY06-078_TUL03\TUL-03 data files\Aeromist nebulizer\Hospitak_nebulizer_ima ge F-8 Sparging Generator n/a \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06\FY06-078_TUL03\TUL-03 data files\Sparging Generator\Sparging_generator_im age F-9 Effect of time, temperature, and media on F. tularensis LVS viability/culturability n/a \\Saturn\absl3\Agent and Study Specific Data \STUDY SPECIFIC DATA\FY06-078_TUL-03\LVS Growth\characterization\Stability\ Effect of time and temp on LVS recovery_18April07.xls F-10 Effect of delivered pressure to the Aeromist Nebulizer on F. tularensis LVS viability/culturability n/a \\Saturn\absl3\Agent and Study Specific Data \STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL03 data files\FY06-078_TUL03_Cumulative MS3 data.xls [Worksheet entitled “Aeromist Pressures”] F-11 Effect of delivered pressure to the Aeromist Nebulizer on F. tularensis SCHU S4 viability/culturability n/a \\Saturn\absl3\Agent and Study Specific Data \STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL03 data files\FY06-078_TUL- Page 50 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Version: 2.0 Page 51 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Table/ Figure1 Title Notebook Location2 (Notebook # and page numbers) Accepted Date:4/23/09 Electronic Location2 (Full Path & File Name) 03_Cumulative MS3 data.xls [Worksheet entitled “Aeromist Pressures”] F-12 Effect of aerosol generation using the Hospitak Nebulizer on F. tularensis SCHU S4 viability and culturability n/a \Saturn\absl3\Agent and Study Specific Data \STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL03 data files\FY06-078_TUL03_Cumulative MS3 data.xls [Worksheet entitled “Hospitak Pressures”] F-13a and b Effect of delivered pressure to the Collison Nebulizer on F. tularensis LVS and SCHU S4 viability/culturability n/a \\Saturn\absl3\Agent and Study Specific Data \STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL03 data files\FY06-078_TUL03_Cumulative MS3 data.xls [Worksheet entitled “Collison Pressures”] F-14 Cumulative spray factor versus aerosol concentration data using frozen F. tularensis LVS and the Aeromist, Collison, Micropump and Sparging Nebulizers n/a \\Saturn\absl3\Agent and Study Specific Data \STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL03 data files\FY06-078_TUL03_Cumulative MS3 data.xls [Worksheet entitled “Summary Graphs”] Page 51 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Version: 2.0 Page 52 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Table/ Figure1 Title Notebook Location2 (Notebook # and page numbers) Accepted Date:4/23/09 Electronic Location2 (Full Path & File Name) F-15 Cumulative spray factor versus aerosol concentration data using fresh F. tularensis LVS and the Aeromist, Collison, and Micropump Nebulizers n/a \\Saturn\absl3\Agent and Study Specific Data \STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL03 data files\FY06-078_TUL03_Cumulative MS3 data.xls [Worksheet entitled “Summary Graphs”] F-16 Cumulative spray factor versus aerosol concentration data using frozen F. tularensis SCHU S4 and the Aeromist and Collison Nebulizers n/a \\Saturn\absl3\Agent and Study Specific Data \STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL03 data files\FY06-078_TUL03_Cumulative MS3 data.xls [Worksheet entitled “Summary Graphs”] F-17 Cumulative spray factor versus aerosol concentration data using fresh F. tularensis SCHU S4 and the Aeromist , Hospitak, and Collison Nebulizers n/a \\Saturn\absl3\Agent and Study Specific Data STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL03 data files\FY06-078_TUL03_Cumulative MS3 data.xls [Worksheet entitled “Summary Graphs”] F-18 Growth of F tularensis LVS on CHAB and BCGA solid media n/a \\Saturn\absl3\Agent and Study Specific Data \STUDY SPECIFIC DATA\FY06\FY06-078_TUL03\LVS Growth characterization\48h on CHAB.jpeg and \48h on BCGA.jpeg Page 52 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Version: 2.0 Page 53 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 6.2 Accepted Date:4/23/09 Appendix 2: Quality Assessment of Milestone Completion and Report Assessment Criteria for Milestone Completion Evaluation of Milestone Completion Report Yes No N/A Comment Has the Milestone Completion Report format been used and all sections completed, including Milestone Summary, Milestone Objectives, Methods Reagents & SOPs, Salient Original Data Results Interpretation & Quality Control, Deliverables Completed, and Appendices? Does the Milestone Summary include the milestone’s goals, milestone results, an overall interpretation of the milestone’s data and conclusion? Do Methods, Critical Reagents and SOPs include summarized methods and details necessary to re-perform critical experiments? A list of critical reagents? The completed table of SOPs? Are salient negative and positive original data included in the Milestone Completion Report? Has the Deliverables Table been completed? X Have the Appendices been completed? Are the specific original data associations with experiments clearly annotated in the “Salient Technical Data” section of the Milestone Completion report? X X Evaluation of Data included Are the salient original data and results included in an organized, easily interpretable format? Is the rationale included? Do tables and figures have legends and original data location annotations? X X UNM accepted SOPs 1, 2, 3, 4 4/23/09 X X The deliverables are the MSCR and an SOP for aerosolization; UNM accepted SOPs 1, 2, 3, 4 and the MS3 MSCR on 4/23/09 Many of the data location cited are identical and yet the experiments show different data. LBERI added the worksheet name in the file reference for clarity Yes No N/A Comment X X X Many of the data location cited are identical and yet the experiments show different data. LBERI Page 53 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 54 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Assessment Criteria for Milestone Completion added the worksheet name in the file reference for clarity Is the data interpretation clear? Is the data storage location listed in Appendix 1 sufficient for data retrieval in the future? (E.g. notebook numbers and pages, electronic file locations including directory paths and file names). Are prior data locations cross-referenced to final data locations? X X Is the data backed up electronically or in hardcopy notebooks? X Is the data storage location secured either in a locked fireproof cabinet for hardcopy or on a server protected by firewall? X Has the data quality been assessed? How many replicates and how reproducible is the data? Has statistical analysis been performed on the data? What quality control has been utilized by the subcontractor during the data generation and assessment? X LBERI indicates all data is stored electronically. Per Trevor Brasel, Data files are backed-up daily and are located on a secure server. Access is limited to the data files by management. If a protein or peptide has been synthesized, how has the protein or peptide sequence been verified? What percentage of the sequences has been randomly verified? X If a genetic mutant has been made, how has the mutation been verified e.g. DNA sequencing, PCR sequence verification? How stable is the mutation? How has the impact of the genetic mutation on the bacterial growth been assessed? What is the sensitivity of the assay? X If an aerosol delivery system has been tested, how reproducible is the delivery to the animal? Have sufficient animal numbers been tested to determine reproducibility? X If UVA/psoralen treatment kills the bacteria but leaves them metabolically active, how is killing assessed? How sensitive is the assessment of killing? How is expression of bacterial epitopes determined? Do UNM and the subcontractor agree that the data supports the scientific interpretation of the milestone? X X Page 54 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 55 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Assessment Criteria for Milestone Completion Evaluation of Deliverables, as outlined in the Statement of Work Yes No N/A Comment Have Standard Operating Protocols have been written by subcontractor, reviewed by UNM, revised by subcontractor as requested, and accepted by UNM? The milestone completion report will not be accepted by UNM until all the SOPs are accepted by UNM. X Has the Milestone Completion Report been written by subcontractor, reviewed by UNM, revised by subcontractor as requested, and accepted by UNM? Has data from the milestone been submitted by the subcontractor, reviewed by UNM, data presentation revised by the subcontractor as requested for clarity, and accepted by UNM? For deliverable reagents, have the minimum number of vials, the minimum concentration, the minimum block size and the minimum weight of tissue been mutually agreed by UNM and the subcontractor? X X X Have bacterial strains and tissues been banked at the subcontractor’s institution and backup stocks and aliquots been received by UNM for long term storage? A minimum number of vials of -20C /-80C bacterial stocks at specified concentration in glycerol are stored at both institutions. A minimum size paraffin block or minimum weight of cryopreserved frozen tissues are stored at both institutions. Evaluation of SOPs Do SOPs contain standard sections e.g. purpose, list of supplies and equipment required including vendors and model numbers, reagent preparation, method, results expected, description of data interpretation, criteria for accepting or rejecting results, description of data storage location, date SOP is in service, names of people who prepared and reviewed the SOP? Can an independent scientist read and understand the standard operating procedure? UNM accepted LBERI SOPs # 1, 2, 3 and 4 on 4/23/09. The deliverables include a protocol for the optimized aerosolization. MS3 MSCR and 4 SOP’s accepted by UNM 4/23/09 The SOPs and MSCR are the deliverables X Yes No N/A Comment X X UNM accepted LBERI SOPs # 1, 2, 3 and 4 on 4/23/09. Page 55 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Version: 2.0 Page 56 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 6.3 6.3.1 Accepted Date:4/23/09 Appendix 3: Additional Data/Figures not included in the Text of the Milestone Completion Report (Section 4) Formulations for Solid Growth Media Table A1. Formulations for Solid Growth Media (per liter) Component BCGA SBCGA CHA BGCA GCAT CHAH2 CHAH10 CHAB CA CAH Agar Peptone Lab Lemco Peptone Meat extract NaCl L-cysteine L-histidine Glucose Horse blood, defibrinated IsovitaleX® Glucose L-cysteine HCl Lglutamine Adenine Thiamine pyrophosp hate Vitamin B12 GuanineHCl Fe(NO3)39H2O p-aminobenzoic acid ThiamineHCl 12.5 g 20 g 3g 1.9 g 12.5 g 20 g 3g 1.9 g - - 14 g - - - - 14 g - 14 g - 1.1 g 1.1 g - - - - - - - - 5g 1g 1g 25 g 5g 1g 1g 25 g - 1g 25 g 5g 25 g - - - - - 50 mL - - 50 mL 50 mL - - 50 mL - - - - 10 mL 1g - - - - 10 mL 1g - - - - 0.26 g - 1g - - 0.26 g - - - - 0.1 g - - - - 0.1 g - - - - 0.01 g - - - - 0.01 g - - - - 1 mg - - - - 1 mg - - - - 1 mg - - - - 1 mg - - - - 3 mg - - - - 3 mg - - - - 2 mg - - - - 2 mg - - - - 0.13 mg - - - - 0.13 mg - - - - 0.03 mg - 0.05 mg - - 0.03 mg - - Page 56 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 57 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Table A1. Formulations for Solid Growth Media (per liter) Component Cysteine Heart Agar Beef heart, solids from infusion Agar Glucose Proteose peptone NaCl L-cysteine Hemoglobin solution Nutrient agar Agar Pancreatic digest of gelatin Beef extract Papaic digest of soybean meal Beef heart for infusion Chamberlain’s medium BCGA SBCGA CHA BGCA GCAT CHAH2 CHAH10 CHAB CA CAH - - 51 g - - 51 g 51 g 51 g - - - - 10 g - - 10 g 10 g 10 g - - - - 15 g 10 g - - 15 g 10 g 15 g 10 g 15 g 10 g - - - - 10 g - - 10 g 10 g 10 g - - - - 5g 1g - - 5g 1g 5g 1g 5g 1g - - - 10 g 10 g - - 2g 10 g - - 10 g - - - 850 mL 12.8 g - - - - - - - - - 4.3 g - - - - - - - - - 2.8 g - - - - - - - - - - 10 g - - - - - - - - - 3g - - - - - - - - - - - - - 24.2 g 24.2 g Table A1. Formulations for solid growth media tested for optimal F. tularensis growth. As shown, solid agar media tested throughout Milestone 3 were diverse in components. Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\Working Group\F. tularensis media comparison.pdf Page 57 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Version: 2.0 Page 58 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 6.3.2 Accepted Date:4/23/09 Formulations for Liquid Broth Media Table A2. Formulations for Liquid Broth Media (per liter) Component MMHB MCPH MCPHI CB CBH NaCl - 6.25 g 6.25 g - - CaCl2 0.14 g - - - - MgCl2-6H2O 0.21 g - - - - Mueller Hinton broth Acid hydrolysate of casein Beef extract 10.5 g - - - - 8.8 g 12.5 g 12.5 g - - 1.5 g - - - - 0.8 g - - - - 0.006 g - - - - 2 mL - 10 mL - - Starch Ferric pyrophosphate IsovitaleX® Glucose 0.2 g - 1g - - L-cysteine HCl 0.052 g 0.1 g 0.26 g - - L-glutamine 0.02 g - 0.1 g - - Adenine Thiamine pyrophosphate Vitamin B12 2 mg - 0.01 g - - 0.2 mg - 1 mg - - 0.2 mg - 1 mg - - Guanine-HCl 0.06 mg - 3 mg - - Fe(NO3)3- 9H2O P-aminobenzoic acid Thiamine-HCl 0.04 mg - 2 mg - - 0.026 mg - 0.13 mg - - 0.0006 mg 25 mg 25 mg - - Bacto Yeast extract - 6.25 g 6.25 g - - K2HPO4 - 1.4 g 1.4 g - - KH2PO4 Chamberlain’s Medium Hemoglobin - 3.3 g 3.3 g - - - - - 24.2 g 24.2 g - - - - 10 g Table A2. Formulations for solid growth media tested for optimal F. tularensis growth. As with solid agar media, tested liquid broths were diverse in their constituents. Data location: \\Saturn\absl3\Agent Page 58 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 59 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL03\TUL-03 data files\Working Group\F. tularensis media comparison.pdf Comparison of Aerosol Parameters between Separate Aeromist, Collison, and Micropump Nebulizers Figures A1-A4 present results from testing observed when comparing separate Aeromist, Hospitak, Collison, and Micropump nebulizers to one another. All evaluations were performed using sterile water as the generator solution. Aeromist Water Output vs Pressure 27 Water Output ( L/L air) 6.3.3 25 23 Nebulizer #4 21 Nebulizer #5 19 Nebulizer #6 17 15 0 5 10 15 20 25 30 35 Pressure (psig) Figure A1. Water output versus delivered pressure for three separate Aeromist nebulizers. Results demonstrated little difference between the units at the tested parameters. Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\FY06-078_TUL-03_Summary Bioaerosol Generator Output Testing.xls Page 59 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 60 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Collison 3-Jet Nebulizer: Water Output vs Pressure 100 Water Output ( L/ L air) 80 Nebulizer #4 60 Nebulizer #5 40 Nebulizer #6 20 0 0 5 10 15 20 25 30 35 Pressure (psig) Figure A2. Water output versus delivered pressure for three separate Collison 3-jet nebulizers. Results demonstrated significant differences between the units at lower delivered pressures, though performances were very similar at standard operating conditions (20-30 psig). Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\FY06-078_TUL-03_Summary Bioaerosol Generator Output Testing.xls Page 60 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 61 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 Aeroneb® Micropump Nebulizer: Water Output vs Pressure Water Output ( L/L air) 100 80 Nebulizer #1 60 Nebulizer #2 40 Nebulizer #3 20 0 0 5 10 15 20 25 Pressure (psig) Figure A3. Water output versus delivered pressure for three separate Micropump nebulizers. Results demonstrated significant differences between the units at the tested parameters. Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL-03\TUL-03 data files\FY06-078_TUL-03_Summary Bioaerosol Generator Output Testing.xls Page 61 of 62 Tularemia Vaccine Development Contract Contract No. HHSN266200500040-C and ADB Contract No. N01-AI-50040 Prime Contractor: University of New Mexico Milestone Completion Report: MS #3 Institution: LBERI Author: Trevor Brasel MS Start Date:2/23/2006 MS End Date:8/31/2008 Report Date: 9/2/2008 Accepted Date:4/23/09 Version: 2.0 Page 62 of 62 Reviewed by : Barbara Griffith, 9/26/08, 10/23/08, 1/29/09, 3/3/09, 4/22/09, 6/26/09, 12/3/09,2/17/10,2/18/10 HospitakTM Nebulizer Flow Rate vs Water Output Water Output (mL/min) 0.2 0.18 0.16 0.14 0.12 0.1 4 4.5 5 5.5 6 6.5 7 Flow Rate (L/min) Output, mL/min Figure A4. Flow rate versus water output versus for 25 separate Hospitak nebulizers. Results demonstrated inconsistencies between different units. Data location: \\Saturn\absl3\Agent and Study Specific Data and Miscellaneous Documents\STUDY SPECIFIC DATA\FY06-078_TUL03\TUL-03 data files\FY06-078_TUL-03_Summary Bioaerosol Generator Output Testing.xls Page 62 of 62