Introduction to 85108A Pulsed Network Analyzers Pengcheng Jia Dec 13, 2001 Outline What is measurement? What is 85108 Pulsed Network Analyzer? What are the Connectors & Cables? What are the Calibration Kits? What is Non-Insertable Calibration? How to do Pulsed Measurement? How to Control by Computer? What is measurement Define Measurement: f, P, T, expected values and tolerance. Define System Configuration Configure Hardware: User setting, Gauging, Always use ESD Protection Check Hardware: DUT, Cables, Connectors Calibrate Verify Measure Analyze and Archive Measure S Parameters 85108A System Configuration System Cabling Diagram LO source: 10M-20 G RF source: 10M-50 G Test Set: Test Set works from 1-50 GHz in 2 Bands: 1-20 GHz Use directly LO source. 6-50 GHz Use third harmonic of LO source. 1-50 GHz 85110A Test Set Signal Flow Simplified NA - 8722D •50 MHz to 40 GHz •100 dB of dynamic range •Excellent measurement accuracy •Two measurement channels •Four display channels Difference: 85108A—8722D Freq Range 8722D 50M-40G 85108A 1-20G,6-50G Accuracy N/A 0.01dB Pulsed RF N Pulse Profile (Time Domain) N Sweep Mode Ramp Mode IF Bandwidth 3KHz Y Y Step Mode 1.5MHz, 10KHz High Power Meas. Y N Avoid Large Time Delay of DUT Frequency Sweep Step Mode: 85108A f t 200 ms Ramp Mode: 8722D f t DUT =10 ns Original 3KHz BW Ramped 8722D with 3K IF BW 4KHz f Pulse Meas.- Freq Domain Point-in-Pulse •Freq Domain Meas. •Use 1.5M Broadband IF filter (Normal meas. use 10K IF filter) •Adjustable Duty Cycle Default: 10% •Adjustable Pulse Width Default: 10 ms, min 1us •Trigger time: 5ms from 0 Pulse Meas.-Time Domain •Display Time Domain Pulse Profile •Choose Pulse Profile in Domain menu •Measure only at a single frequency High Power Meas. +43 dBm +17 dBm +43 dBm +17 dBm Never Exceed the Power Limitation on the Test Port Port1 & Port2: +43 dBm RF Input Port: +17 dBm Connectors for NA Connector Dielectric Freq Mate with Wrench Note SMA Yes 18G 3.5, 2.9 5lb.in Not recommend to mate with 3.5, 2.9 3.5 No 26.5G SMA,2.9 8lb.in HP PSC connector 2.9(K) No 40G SMA,3.5 8lb.in HP PSC, Wiltron K 2.4 No 50G 1.85 HP PSC 1.85(V) Yes 65G 2.4 8lb.in PSC: Precision Slotless Connector HP PSC, Wiltron V Cables for NA •Cables (only following cables can be used for NA): Semi-Rigid with K connectors Flexible—Thin Jacket with K connectors (Blue, from Gore) Flexible—Thick Jacket with 2.4 mm connectors (Black, from Agilent) • Connectors on Port 1 & 2 is 2.4 mm • 2.4 to 2.9 (K) adaptors are put to mate with K connector cables, Never Remove Them • Always Keep Connects & Adaptors Clean • Treat Them Gentle and Put Back to Box Cautions: • Select Correct Cal Kit and Load Calibration Constants before Calibration • Never Remove a Calibration Standard from the Box or Use for Any Purpose Other Than Calibration • Never Drop the Standards to Ground • Always Use Special Wrench Correctly • Put the Cal Kits Back to Cabinet after Use • Report Any Discrepancy Load Configuration Disk Load the Test Set Configuration Disk for 85110A with H60/61 option (Disks are stored in Drawers, never remove them) Disk->Storage is Internal Load->More->Machine Dump MD_1_20_NP, 1-20 GHz, Non-Pulsed MD_1_20_PL, 1-20 GHz, Pulsed MD_6_50_NP, 6-50 GHz, Non-Pulsed MD_6_50_NP, 6-50 GHz, Pulsed Calibration Procedure • Press Cal menu • Choose Calibration Kit, if not installed in equipment, load the cal constant from disk • Choose in the Calibration Menu LRM for on-wafer Calibration, Full 2-port for cable calibration • Follow instruction, connect standards and measure • Save calibration data • You will see “Correction On” Error Model for Full 2-Port Cal Calibration Non-Insertable DUT Swap Equal Adaptors Swap Equal BetweenSeries Adaptors Swap Equal In-Family Adaptors Adaptor Removal • Calibration with adaptor on Port 2, save as Cal Set 1 • Calibration with adaptor on Port 1, save as Cal Set 1 • Modify Cal Set->Adaptor Removal •Refer the additional document for more details •More calibration methods are included in that file Calibration with On-Wafer Kit • Use WinCal on Controller-Dizzy • Calibration Constant is set in program • Choose LRM, since short is not good • LRRM is recommended for more accuracy How to do Pulsed Meas.(1) Frequency Domain (Point-in-Pulse) • Domain->Frequency • System->More->Pulse Config->Detector: Wide BW • Select Maximum Points • Calibration • System->More-> Pulse Config-> High or Low • Set Pulse Width, Duty Cycle • Stimulus->More->Trigger Mode->Trigger Delay • Measure How to do Pulsed Meas.(2) Pulse Profile Domain - Single Frequency • Domain->Frequency • Stimulus->Step mode & Start->Set desired frequency • Domain->Pulse Profile • Select Maximum Points • Calibration • Measure • Repeat for Next Frequency How to do Pulsed Meas.(3) Pulse Profile Domain – Frequency List • Cal->Correction off • Domain->Frequency • System->More->Pulse Config->Detector: Wide BW • Stimulus->More->Edit List • Add->Start->3 G/n, Number of Points->51x1->Done • Add->Start->4 G/n x1->Done • Add->Start->5 G/n x1->Done • Done Frequency List, and Proceed with Calibration How to do Pulsed Meas.(4) Using Frequency List – Continue: •When begin to Measure: Domain-> Pulse Profile Stimulus->Frequency List->Single Segement The last selected segment will be active and ready for measurement Calibration will be effective for each frequency point • Set Pulse Polarity, Pulse Width and Duty cycle if needed Computer Control • 2 GPIB Connectors are on the back of 8510C 1 for 85108 System bus (don’t touch) 1 for computer controller (already connected) • Dizzy is the controller • Use WinCal for On-Wafer Calibration and SParameter Measurement Choose the correct VNA first Config->Option->VNA->8510 or 8722 • Use Labview Program for 85108 configuration