ACR CT Accreditation Pitfalls to Avoid in Phantom Testing Phantom Images • One set of phantom images (2 films) per CT • Completed data sheets Tom Payne PhD Abbott Northwestern Hospital Minneapolis, MN First Step: Table 1 • Ensure that the data matches what they do clinically • If AEC, need to determine average mAs • If “effective” mAs, need to calc mAs and mA • N x T = beamwidth z-axis collimation (T) = the width of the tomographic section along the z-axis imaged by one data channel. In multi-detector row (multi-slice) CTscanners, several detector elements may be grouped together to form one data channel. number of data channels (N)= the number of tomographic sections imaged in a single axial scan. maximum numb er of data channels (Nmax) = the maximumnumber oftomographic sections for a single axial scan. increment (I) = the table increment per axial scan or the table increment per rotation of the x-ray tube in a helical scan. mA= the actual tube current used (averaged over one rotation) exposure time = the time required for one complete rotation of the x-ray source. Pitch • Use IEC definition – Pitch = I/N*T • I = table increment/speed • N = number of data channels used (not detector elements) • T = z-axis collimation (thickness of one data channel) – Not always stated correctly on the CT system • Know underlying detector configuration Phantom Positioning Use CT alignment lasers Optional base can make life easier Use bubble level to verify pitch and roll Ensure teflon rings off module centers Scan with slice >5mm to check “longest” wires High-resolution chest technique Must see all four BBs (in Modules 1 & 4) Longer wires should be visible on both ramps Alignment adjustment WW = 1000 WL = 0 MODULE 1 Polyethylene -97 HU “Bone” +910 HU Water 0 HU Acrylic +120 HU If FAIL, do not pass “go”, do not collect accreditation Film 1, Boxes 4 - 12 CT Number Accuracy and Slice Thickness • Align Module 1 to lasers • Scan with “Axial” version Adult Abdomen protocol • Record slice thickness and all CT numbers – Water, air, polyethylene, bone and acrylic • Scan varying slice thickness (Hi Res 3, 5, & 7 mm) – Measure water CT number and slice thickness • Scan varying kVp (scan with all available) – Measure water CT number and slice thickness Air -1000 HU Axial Conversion • If an axial acquisition cannot be made using that selection of N and T, keep T the same as described in Table 1 and use the next smallest allowed value of N. Example: Siemens Sensation 16 system with N = 16 and T = 1.5 mm and reconstructed helical scan width = 5 mm. Axial images cannot be acquired using N = 16. Use the same value of T (1.5 mm) but the next lowest allowed value of N, which would be 12. Thus the 12 x 1.5 mm detector configuration would be used for the axial version of the spiral adult abdomen protocol with N = 16 and T = 1.5 mm. This is similarly true for the 16 x 0.75 mm detector configuration (use an axial 12 x 0.75 mm detector configuration). Adult abdomen technique Low contrast = 6 HU ± 0.5 HU Poly “Bone” MODULE 2 2 mm 25 mm “Use 120 kVp” Provide a note if regular technique is 140 kVp CT number measured in 5 materials 3 mm 6 mm Water 4 mm Acrylic Air 5 mm WW = 400 WL = 0 Image Quality – Low Contrast Low Contrast Resolution • Scan Module 2 – Adult Abdomen protocol - Helical – Adult Head protocol - Axial • Record the diameter (mm) of the smallest set of LCR rods seen Adult abdomen and head techniques Record diameter of the smallest set of rods for which all 4 rods can be seen ROI check of absolute contrast using large rod WW = 100 WL = 100 High Contrast Resolution Bar patterns: lp/cm MODULE 4 12 4 10 6 • Scan Module 4 8 – Adult abdomen protocol – Helical – Hi Resolution Chest protocol 9 5 • Record limiting resolution (lp/cm) 7 Adult abdomen, adult head and high-resolution chest techniques The End Record the first highest frequency bar pattern for which the bars and spaces merge Remember – in completing the ACR Physics Tests “Artificial intelligence is no match for natural stupidity” WW = 100 WL 1100