Slide 1 Slide 2 Introduction Optimizing Dose and Image Quality in Digital Mammography Eric A. Berns, Berns, PhD Certified Statistics – 15 Months April 1, 2005 August 1, 2006 Difference Total Certified Mammo Facilities 8,929 8,829 -100 Total Accredited Mammo Units 13,640 13,556 -84 Certified Facilities with FFDM units 607 1,130 +523 Accredited FFDM units 819 1,604 +785 eberns@radiology.northwestern.edu Northwestern University Medical School Lynn Sage Comprehensive Breast Center Chicago, IL Slide 3 Slide 4 Introduction FFDM vs. Date Fuji FCRm App: Jul ‘06 Detector Types 12% Siemens Novation App: Aug ‘04 % Fac w Dig 10% % Units FFDM 6% Cesium Iodide with Silicon Diode Array (GE) Lorad Selenia App: Oct ‘02 Lorad CCD App: Mar ‘02 8% Selenium with Silicon Diode Array (Lorad (Lorad,, Siemens, Fischer Senoscan App: Sept ‘01 GE Essential App: Apr ‘06 4% Kodak, Agfa) GE 2000D App: Jan 2000 2% 0% Jul-98 Slot Scanning CCD Array (Fischer) GE DS App: Feb 2004 Dec-99 Apr-01 Sep-02 Jan-04 May-05 Oct-06 Computed Radiology (Fuji, Kodak, Konica, Agfa) 1 Slide 5 Slide 6 Equipment GE Senographe 2000D Slide 7 Equipment GE Senographe DS Slide 8 Equipment Equipment GE Senographe 2000D & DS & Essential – FOV: 19.2 x 23.0 cm & 24.0 x 31.0 cm – Spatial resolution: 100 microns (5.0 lp/mm) lp/mm) GE Senographe Essential 2 Slide 9 Slide 10 Equipment Equipment Siemens Mammomat NovationDR Lorad Selenia Slide 11 Slide 12 Equipment Equipment Lorad Selenia & Siemens Novation – FOV: 24 x 29 cm – Spatial resolution: 70 microns (7.14 lp/mm) lp/mm) Fischer Senoscan 3 Slide 13 Slide 14 Equipment Equipment Computed Radiology Fischer Senoscan – Fuji – FDA approved – FOV: 21 x 29 cm (std), 11 x 15 cm (high res) – Kodak – 54 & 27 microns – 9.3 lp/mm lp/mm in 54 µm mode – Konica – Scan time: time: 5.2 seconds – Agfa Slide 15 Slide 16 Equipment Equipment Fuji FCRm Fuji FCRm – FOV: 18 x 24 and 24 x 30 cm – Spatial resolution: 50 microns (10.0 lp/mm) lp/mm) – 60 to 80 imaging plates per hour Flash Plus IIPm FCSm 4 Slide 17 Slide 18 Equipment Equipment Kodak DirectView CR Mammography System Sectra Microdose – SlotSlot-scanning photon counter detector – FOV: 24 x 26 cm – Spatial resolution: 50 microns (10.0 lp/mm) lp/mm) Slide 19 Slide 20 Equipment Equipment • FDAFDA-Approved Laser Imagers ~ 40 µ m spot •Agfa LR5200 Laser Imager (Wet Chemistry) •Agfa DS4500M •Kodak 8600 Laser Imager •Kodak 8610 Laser Imager •Kodak 8900 •Fuji Drypix 4000, 5000, 7000 •Fuji Drypix FMFM-DP L •Codonics Horizon Ci, Ci, GS, SF Planmed Nuance – Amorphous Selenium – 85 µm •Konica Minolta DryPro 793 5 Slide 21 Slide 22 Workstations Workstations What is a Multimodality Review Workstation? Until recently, workstations came with acquisition units Those days are over Spurred by FDA approval of third party RWS, high resolution displays, and PACS As a result, FFDM is becoming more a la carte to allow best of each (acquisition, RWS, PACS, etc.) Slide 23 Slide 24 Multimodality Workstations Workstations FDA Approved Multimodality Workstations – AGFA IMPAX MA3000 Current challenges – Sectra IDS5/mx – Kodak DirectView PACS System – Device to device connectivity – McKesson’ McKesson’s PACS Mammo Station – iCad Second Look 500M – DICOM incompatibilities between acquisition and – Cedara I-ReadMammo – Fuji Synapse displays – GE Seno Advantage & Seno Adv. 2 – Fischer Senoview Plus (Cedara (Cedara)) • Can result in image degradation – Hologic SecureView DX – Siemens MammoReportPlus – PACS connectivity 6 Slide 25 Slide 26 Workstations Detector Size (cm) AEC System Function vs. Image size (pixels) 5625 GE Essential GE (2000D, DS, Essential) Fischer Mo & Rh Targets, Mo & Rh Filters Lorad & Siemens Lorad & Siemens 4100 3100 31 29 Fischer 23 Monitor 3 Modes 2560 2304 GE 2000D & DS – Mo/Mo, Mo/Rh Mo/Rh,, Rh/Rh GE Essential – Con – Thin, less dense (2000D) GE 2000D & DS – Std – Intermediate (DS & Essential) – Dos – Thick, dense breasts 19 21 24 24003400 4096 1920 2048 Slide 27 Slide 28 AEC System Function Lorad – Mo target, Mo & Rh Filters – Mo/Mo, Mo/Rh Mo/Rh Modes – – – – AutoAuto-Filter – AEC sensor, exposure adjustment AutoAuto-kV – Filter, AEC sensor, exposure adjustment AutoAuto-Time – kV, filter, AEC sensor, exposure adjustment TEC – (Tissue Exposure Control) - Breast density AEC System Function Siemens – Mo & W Targets, Mo & Rh Filters – Mo/Mo, Mo/Rh Mo/Rh,, W/Rh W/Rh Modes – OPDOSE Mode (recommended using W/Rh W/Rh)) • Compress breast to given compression • (enhanced manual mode) – Manual Recommended Mode – AutoAuto-Filter • T/F & kVp prepre-selected by vendor lookup tables – Manual 7 Slide 29 Slide 30 2000 FFDM Variability Data AEC System Function AEC Summary How Does Digital Compare to ScreenScreen-film Mammography in Terms Of: – Manufacturer only recommends, user decides • Exposure – Ultimately up to the Radiologist and/or Physicist Times? • Breast Dose? to decide which mode to use • Detection of LowLow-contrast Lesions? – Make mode choice based on knowing effect on 3 AOP Modes: dose and image quality Slide 31 Contrast Standard O Dose + - Slide 32 Methods ContrastContrast-Detail image analysis Methods Contrast Detail Phantoms – Acquire images at recommended techniques • 2, 4, 6, 8 cm • Calculate Dose • Score contrastcontrast-detail image for image quality ACR Phantom – Calculate Dose – Scores Compare to screenscreen-film data 8 Slide 33 Slide 34 Methods Methods Decreasing Diameter 0.25 mm 4.0 mm 4% Decreasing Contrast 0.25% Slide 35 Slide 36 Methods ACR Phantom ACR Phantom 9 Slide 37 Slide 38 Average Glandular Dose vs. Thickness 38 ScreenScreen-film Units, 18 FFDM Units Exposure Time vs. Thickness 38 ScreenScreen-film Units, 18 FFDM Units 700 6 CMAP ACR Phantom CMAP 6 cm 500 CMAP 8 cm FFDM - Con/Auto 400 74% of FSM > 2 s CMAP 2 cm CMAP 4 cm Exposure Time (seconds) Average G landular Dose (m rad) CMAP 2 cm 600 3 mGy Limit for ACR Phantom FFDM - Std/Auto FFDM - Dos/Auto 300 200 100 0 5 CMAP 4 cm CMAP ACR Phantom CMAP 6 cm 4 16% of FSM > 2 s CMAP 8 cm FFDM - Con/Auto 3 FFDM - Std/Auto 2 seconds FFDM - Dos/Auto 2 1 0 0 1 2 3 4 5 6 7 8 9 10 0 1 2 3 Thickness (cm) 5 6 7 8 9 10 Thickness (cm) Slide 39 Slide 40 Contrast Detail Scores vs. Thickness 38 ScreenScreen-film Units, 18 FFDM Units 18 Methods 2002 GE 2000D Opto Data 16 14 C D S co res 4 12 CMAP 2 cm CMAP 4 cm CMAP 6 cm CMAP 8 cm FFDM - Con/Auto FFDM - Std/Auto FFDM - Dos/Auto 10 8 6 0 2 4 6 Thickness (cm) 8 10 Medical Physics. March 2003. 30 (3) pages 334-340 10 Slide 41 Slide 42 Methods Results CD Score vs. kVp 2 cm Breasts 15 Objective: To determine optimized technique 14 Mean CD Score factors for fullfull-field digital mammography system (GE 2000D) for lowlow-contrast lesion detection 13 12 11 2 cm Mo/Mo 10 Optimization done under condition of matched 9 patient dose to screenscreen-film mammography 8 2 cm Mo/Rh Mo/Rh 2 cm Rh/Rh ScreenScreen-Film 20 25 30 Compare fullfull-field digital results to screenscreen-film results 35 45 50 Mo/Mo Trend pp-value = 0.0752 Mo/Rh p-value = 0.1369 Mo/Rh Trend pRh/Rh Trend pp-value = 0.0985 Optimized Technique Comparison p-value = 0.47 Slide 43 40 kVp Slide 44 Results Results CD Score vs. kVp 6 cm Breasts 15 15 14 14 Mean CD Score Mean CD Score CD Score vs. kVp 4 cm Breasts 13 12 11 4 cm Mo/Mo 10 4 cm Mo/Rh Mo/Rh 13 12 11 6 cm Mo/Mo 10 6 cm Mo/Rh Mo/Rh 6 cm Rh/Rh 4 cm Rh/Rh 9 9 ScreenScreen-Film ScreenScreen-Film 8 8 20 25 30 35 kVp Optimized Technique Comparison p-value = 0.013 40 45 50 Mo/Mo Trend pp-value = 0.2221 Mo/Rh Mo/Rh Trend pp-value = 0.5691 Rh/Rh Trend pp-value = 0.0123 20 25 30 35 kVp Optimized Technique Comparison p-value < 0.0001 40 45 50 Mo/Mo Trend pp-value = 0.5710 Mo/Rh Mo/Rh Trend pp-value = 0.6496 Rh/Rh Trend pp-value = 0.2735 11 Slide 45 Slide 46 Results Conclusions CD Score vs. kVp 8 cm Breasts 15 LowLow-contrast lesion optimization for FFDM (GE 2000D) 8 cm Mo/Mo Mean CD Score 14 8 cm Mo/Rh Mo/Rh 8 cm Rh/Rh 13 – Thin breasts (< 2 cm): Mo/Mo with low kVp ScreenScreen-Film 12 – Intermediate breasts (~4 cm): Insensitive to targettarget-filter and kVp 11 10 selection 9 – Thick breasts (>5 cm): Rh/Rh with higher kVp 8 20 25 30 35 40 kVp 45 50 Mo/Mo Trend pp-value < 0.0428 Mo/Rh p-value < 0.9453 Mo/Rh Trend pRh/Rh Trend pp-value < 0.0121 Optimized Technique Comparison p-value < 0.0001 Slide 47 Slide 48 ACRIN Data Compressed Breast Thickness vs. Dose - All GE Sites (5) 6 Compressed Breast Thickness vs. Dose - All Fischer Sites Mean dose is 32% lower with FFDM than SFM 6 Film-screen 5 Digital 5 4 Dose (mGy) Dose (mGy) 4 3 2 3 2 1 1 Film Screen Digital 0 0 1 2 3 4 5 6 7 8 9 Compressed Breast Thickness (cm) 10 0 0 Fischer FFDM doses were 32% lower than SFM doses 2 4 6 8 10 Compressed Breast Thickness (cm) 12 Slide 49 Slide 50 Compressed Breast Thickness vs. Dose - All Lorad Sites Compressed Breast Thickness vs. Dose - All Fuji Sites 6 6 Film-screen Filmscreen Digital 5 4 4 Dose (m Gy) Dose (mGy) Digital 5 3 2 2 1 1 0 0 0 2 4 6 8 0 10 2 4 6 8 10 Compressed Breast Thickness (cm) Compressed Breast Thickness (cm) Lorad FFDM doses were 2% higher than SFM doses Fuji FFDM doses were 5% lower than SFM doses Slide 51 Slide 52 Mean Dose Comparison by Digital Manufacturer 3.0 Screen-Film Overall: Digital Doses 22% Lower Than SFM Digital 2.6 mGy)) Mean Glandular Dose ((mGy 3 2.4 2.5 ARRS 2006 2.4 2.3 2.3 2.3 2.2 2.0 1.9 1.8 1.7 1.5 1.0 0.5 Comparison of Image Quality and Average Glandular Dose on Four FDA-approved FullField Digital Mammography Systems 0.0 Fischer Fuji GE Lorad All Sites 13 Slide 53 Slide 54 Purpose Methods To Measure and Compare 4 FDAFDA-approved FFDM systems – Image Quality – GE 2000D – Average Glandular Dose • On 4 FDAFDA-approved FFDM systems – GE Senographe DS • Across full range of breast thicknesses – Lorad Selenia • Using each manufacturers’ manufacturers’ recommended techniques Slide 55 – Siemens Novation Slide 56 Methods Results Average Glandular Dose Comparison To acquire images – Position the phantom like a clinical acquisition 4 DS 3.5 – Apply 10 dN compression force technique – Record technique factors – Measure HVL’ HVL’s and entrance exposures – Calculate AGD – Measure and calculate ContrastContrast-Detail scores Dose (mGy) – Acquire image using manufacturer’ manufacturer’s recommended Lorad 2000D 3 Lorad GE 2000D Siem 2.5 Range up to 2.9 Siemens 2 GE DS 1.5 1 0.5 0 0 2 4 6 Breast Thickness (cm) 8 10 14 Slide 57 Slide 58 Results Results Contrast-Detail Scores 17 16 Contrast-Detail Scores Average Glandular Dose Comparison 4 Lorad GE 2000D GE DS Siemens 14 18 DS 3.5 3 13 Lorad 12 Siemens GE DS GE 2000D 11 17 2000D Contrast-Detail Score 15 Dose (mGy) Contrast-Detail Score 18 Lorad Siem 2.5 2 1.5 1 0.5 15 14 13 12 11 10 9 8 0 10 16 0 2 4 6 Breast Thickness (cm) 8 10 0 2 4 6 Breast Thickness (cm) 8 10 9 8 0 2 4 6 Breast Thickness (cm) 8 10 Slide 59 Slide 60 Results Conclusions ACR Phantom Average Glandular Dose 3 – Average glandular doses varies by up to a factor of 2.9 2.5 AGD (mGy) Results indicate that 2 – There are significant differences in image quality 1.5 1 – Technique factors and automatic exposure mode selection can play an important role in clinical image quality and patient dose 0.5 0 Siemens GE DS GE 2000D Lorad 15 Slide 61 Slide 62 Digital Proposed Phantom Slide 63 Slide 64 Phantom Comparison ACR Phantom Object # ACR Proposed 1 1.56 2 1.12 3 0.89 4 0.75 5 0.54 0.54 6 0.40 0.40 Fiber Diameter (mm) 0.89 0.82 0.75 0.65 0.30 16 Slide 65 Slide 66 Phantom Comparison ACR Phantom Object # ACR Proposed Fiber Diameter (mm) ACR Proposed Speck Diameter (mm) Phantom Comparison ACR Phantom Object # ACR Proposed Fiber Diameter (mm) ACR 0.54 1 1.56 0.54 1.12 0.40 2 1.12 0.40 3 0.89 3 0.89 0.82 4 0.75 0.75 0.24 0.54 0.54 6 0.40 0.24 4 0.75 0.20 0.16 0.89 0.32 0.82 0.28 0.65 5 0.32 0.75 Proposed 2.00 1.56 2 0.32 ACR Mass Thickness (mm) 1 0.89 Proposed Speck Diameter (mm) 1.00 0.32 0.75 0.28 0.24 0.65 0.24 0.50 0.20 0.16 0.16 0.75 0.68 0.50 0.38 0.16 5 0.54 0.54 0.25 0.40 0.13 6 0.40 0.40 0.13 0.25 0.20 0.30 0.10 0.30 0.10 0.15 Slide 67 17 Slide 72 Optimization 2006 Objective: To determine optimized technique factors for clinically available fullfull-field digital mammography systems On 6 FFDM Units (Siemens counts as 2) Using – ContrastContrast-Detail Phantoms 4, 3, 3 3, 1, 1 – ACR Phantom – Proposed ACR Digital Phantom 18 Slide 73 Slide 74 Optimization 2006 Optimization 2006 Measure dose at all recommended techniques Evaluate performance of the ACR phantom and and compare to ACRIN doses compare to our proposed ACR digital phantom Compare dose and image quality within each Evaluate image quality as a function of dose for system using each mode different modes Compare dose and image quality for each Ultimately find the optimum techniques to provide system using their recommended mode Slide 75 the highest image quality with the lowest dose Slide 76 HVL vs. kVp HVL vs. kVp 0.65 GE 2000D Mo/Mo GE 2000D Mo/Rh GE 2000D Rh/Rh GE DS Mo/Mo GE DS Mo/Rh GE DS Rh/Rh GE Essential Mo/Mo GE Essential Mo/Rh GE Essential Rh/Rh Lorad Mo/Mo Lorad Mo/Rh Siemens Mo/Mo Siemens Mo/Rh Siemens W/Rh 0.55 0.45 0.35 0.25 0.55 HVL (mm Al) HVL (mm Al) 0.65 0.45 GE 2000D Mo/Mo GE DS Mo/Mo GE Essential Mo/Mo Lorad Mo/Mo Siemens Mo/Mo 0.35 0.25 20 30 40 kVp 50 20 25 30 35 40 45 50 kVp 19 Slide 77 Slide 78 HVL vs. kVp 0.65 0.65 0.55 0.55 HVL (mm Al) HVL (mm Al) HVL vs. kVp 0.45 GE 2000D Mo/Rh GE DS Mo/Rh GE Essential Mo/Rh Lorad Mo/Rh Siemens Mo/Rh 0.35 0.45 GE 2000D Rh/Rh GE DS Rh/Rh GE Essential Rh/Rh Siemens W/Rh 0.35 0.25 0.25 20 25 30 35 40 45 50 20 25 30 kVp Slide 79 40 45 50 Slide 80 Entrance Skin Exposure vs. kVp Entrance Skin Exposure vs. kVp 45 45 GE 2000D Mo/Mo GE 2000D Mo/Rh GE 2000D Rh/Rh GE DS Mo/Mo GE DS Mo/Rh GE DS Rh/Rh GE Essential Mo/Mo GE Essential Mo/Rh GE Essential Rh/Rh Lorad Mo/Mo Lorad Mo/Rh Siemens Mo/Mo Siemens Mo/Rh Siemens W/Rh 35 30 25 20 15 10 5 40 ESE (mR/mAs) 40 ESE (mR/mAs) 35 kVp 35 30 25 20 GE 2000D Mo/Mo GE DS Mo/Mo GE Essential Mo/Mo Lorad Mo/Mo Siemens Mo/Mo 15 10 5 0 0 20 30 40 kVp 50 20 25 30 35 40 45 50 kVp 20 Slide 81 Slide 82 Entrance Skin Exposure vs. kVp Entrance Skin Exposure vs. kVp 45 45 40 GE 2000D Mo/Rh GE DS Mo/Rh GE Essential Mo/Rh Lorad Mo/Rh Siemens Mo/Rh 35 30 25 20 15 GE 2000D Rh/Rh GE DS Rh/Rh GE Essential Rh/Rh Siemens W/Rh 35 ESE (mR/mAs) ESE (mR/mAs) 40 30 25 20 15 10 10 5 5 0 0 20 25 30 35 40 45 20 50 25 30 35 Slide 83 40 45 50 kVp kVp Slide 84 Average Glandular Dose vs. Breast Thickness 4.0 GE 2000D - Std/Auto GE 2000D - Dos/Auto 3.0 AGD (mGy) Dose GE 2000D - Con/Auto 3.5 GE DS - Con/Auto GE DS - Std/Auto 2.5 GE DS - Dos/Auto GE Essential - Con/Auto 2.0 GE Essential - Std/Auto GE Essential - Dos/Auto 1.5 Lorad - Auto-Filter 1.0 Siemens - OpComp - Mo/Mo Siemens - OpComp - W/Rh 0.5 ACRIN - Film Dose ACRIN - Digital Dose 0.0 0 2 4 6 8 10 Breast Thickness (cm) 21 Slide 85 Slide 86 Average Glandular Dose vs. Breast Thickness Average Glandular Dose vs. Breast Thickness 4.0 4.0 GE 2000D - Con/Auto 3.5 GE 2000D - Dos/Auto GE DS - Std/Auto 3.0 AGD (mGy) AGD (mGy) 3.0 GE DS - Con/Auto 3.5 GE 2000D - Std/Auto 2.5 2.0 1.5 GE DS - Dos/Auto 2.5 2.0 1.5 1.0 1.0 0.5 0.5 0.0 0.0 0 2 4 6 Breast Thickness (cm) 8 0 10 4 6 8 10 Breast Thickness (cm) Slide 87 Slide 88 Average Glandular Dose vs. Breast Thickness Average Glandular Dose vs. Breast Thickness 4.0 4.0 GE Essential - Con/Auto 3.5 GE 2000D - Con/Auto 3.5 GE Essential - Std/Auto 3.0 GE DS - Std/Auto 3.0 GE Essential - Dos/Auto AGD (mGy) AGD (mGy) 2 2.5 2.0 1.5 2.0 1.5 1.0 1.0 0.5 0.5 0.0 GE Essential - Std/Auto 2.5 0.0 0 2 4 6 Breast Thickness (cm) 8 10 0 2 4 6 8 10 Breast Thickness (cm) 22 Slide 89 Slide 90 Average Glandular Dose vs. Breast Thickness Average Glandular Dose vs. Breast Thickness 4.0 4.0 3.5 Siemens - OpComp - Mo/Mo 3.0 GE DS - Std/Auto GE Essential - Std/Auto 3.0 Siemens - OpComp - W/Rh AGD (mGy) AGD (mGy) GE 2000D - Con/Auto Lorad - Auto-Filter 3.5 2.5 2.0 1.5 Siemens - OpComp - W/Rh ACRIN - Film Acrin-Digital 2.0 1.5 1.0 1.0 0.5 0.5 0.0 Lorad - Auto-Filter 2.5 0.0 0 2 4 6 8 10 0 2 4 Breast Thickness (cm) Slide 91 6 8 10 Breast Thickness (cm) Slide 92 Contrast-Detail Score vs. Breast Thickness 17 GE 2000D - Std/Auto Contrast-Detail Score CD Scores GE 2000D - Con/Auto 16 15 GE 2000D - Dos/Auto GE DS - Con/Auto 14 GE DS - Std/Auto 13 GE DS - Dos/Auto GE Essential - Con/Auto 12 GE Essential - Std/Auto 11 GE Essential - Dos/Auto Lorad - Auto-Filter 10 Siemens - OpComp - Mo/Mo 9 Siemens - OpComp - W/Rh 8 0 2 4 6 8 10 Breast Thickness (cm) 23 Slide 93 Slide 94 Contrast-Detail Score vs. Breast Thickness 17 17 16 16 Contrast-Detail Score Contrast-Detail Score Contrast-Detail Score vs. Breast Thickness 15 14 13 12 11 10 GE 2000D - Con/Auto GE 2000D - Std/Auto GE 2000D - Dos/Auto 9 15 14 13 12 11 10 GE DS - Con/Auto GE DS - Std/Auto GE DS - Dos/Auto 9 8 8 0 2 4 6 8 10 0 2 Breast Thickness (cm) Slide 95 6 8 10 8 10 Slide 96 Contrast-Detail Score vs. Breast Thickness Contrast-Detail Score vs. Breast Thickness 17 17 16 16 Contrast-Detail Score Contrast-Detail Score 4 Breast Thickness (cm) 15 14 13 12 11 GE Essential - Con/Auto 10 GE Essential - Std/Auto 9 15 14 13 12 11 Lorad - Auto-Filter 10 Siemens - OpComp - Mo/Mo 9 GE Essential - Dos/Auto 8 Siemens - OpComp - W/Rh 8 0 2 4 6 Breast Thickness (cm) 8 10 0 2 4 6 Breast Thickness (cm) 24 Slide 97 Slide 98 Contrast-Detail Score vs. Breast Thickness 17 ACR Phantom Contrast-Detail Score 16 15 14 13 12 GE 2000D - Con/Auto 11 GE DS - Std/Auto 10 GE Essential - Std/Auto Lorad - Auto-Filter 9 Siemens - OpComp - W/Rh 8 0 2 4 6 8 10 Breast Thickness (cm) Slide 99 Slide 100 Fibers for the ACR Phantom 7 2.50 6 5 2.00 Fibers AGD (mGy) AGD for the ACR Phantom 3.00 1.50 1.00 3 2 0.50 0.00 4 1 Con Std 2000D Dos Con Std DS Dos Con Std Essential Auto Mo/Mo W/Rh Filter Siemens Lorad Dos 0 Con Std 2000D Dos Con Std DS Dos Con Std Essential Dos Auto Mo/Mo W/Rh Filter Siemens Lorad 25 Slide 101 Slide 102 Masses for the ACR Phantom 6 5 5 4 4 Masses 6 3 3 2 2 1 1 0 Con Std 2000D Dos Con Std DS Dos Con Std Essential Dos 0 Auto Mo/Mo W/Rh Filter Siemens Lorad Slide 103 Con Std Dos Con 2000D Std Dos DS Con Std Essential Dos Auto Mo/Mo W/Rh Filter Siemens Lorad Slide 104 AGD for "Proposed" ACR Digital Phantom Proposed ACR Digital Phantom 3.00 2.50 AGD (mGy) Speck Groups Speck Groups for the ACR Phantom 2.00 1.50 1.00 0.50 0.00 Con Std 2000D Dos Con Std DS Dos Con Std Essential Dos Auto Mo/Mo W/Rh Filter Siemens Lorad 26 Slide 105 Slide 106 Fibers for "Proposed" ACR Digital Phantom 8 7 7 6 6 Speck Groups 8 5 Fibers Speck Groups for "Proposed" ACR Digital Phantom 4 3 5 4 3 2 2 1 1 0 Con Std Dos Con 2000D Std Dos DS Con Std Dos Essential 0 Auto Mo/Mo W/Rh Filter Siemens Lorad Slide 107 Con Std 2000D Dos Con Std DS Dos Con Std Essential Dos Auto Mo/Mo W/Rh Filter Siemens Lorad Slide 108 Masses for "Proposed" ACR Digital Phantom Vary Dose from Rec. Techs. 8 7 Masses 6 5 4 3 2 1 0 Con Std 2000D Dos Con Std DS Dos Con Std Essential Dos Auto Mo/Mo W/Rh Filter Siemens Lorad 27 Slide 109 Slide 110 Contrast-Detail Scores vs. % Vendor AGD For 6 cm Breast Thickness Contrast-Detail Scores vs. % Vendor AGD For 6 cm Breast Thickness GE 2000D - Con/Auto 16.0 GE 2000D - Std/Auto 15.0 GE 2000D - Con/Auto 15.0 GE DS - Con/Auto 14.0 GE DS - Std/Auto 13.0 GE Essential Con/Auto GE Essential Std/Auto Lorad 12.0 11.0 Siemens - Mo/Mo 10.0 Contrast-Detail Scores Contrast-Detail Scores 16.0 GE 2000D - Std/Auto 14.0 Rh/Rh – 29 kVp 13.0 Rh/Rh – 31 kVp 12.0 11.0 10.0 Siemens - W/Rh 9.0 9.0 0% 50% 100% 150% 200% 0% 250% 50% 100% Slide 111 200% 250% Slide 112 Contrast-Detail Scores vs. % Vendor AGD For 6 cm Breast Thickness Contrast-Detail Scores vs. % Vendor AGD For 6 cm Breast Thickness 16.0 GE DS - Con/Auto GE DS - Std/Auto 15 14 Rh/Rh – 29 kVp 13 Rh/Rh – 29 kVp 12 11 10 9 Contrast-Detail Scores 16 Contrast-Detail Scores 150% % Vendor AGD % Vendor AGD GE Essential - Con/Auto 15.0 Rh/Rh – 29 kVp GE Essential - Std/Auto Rh/Rh – 29 kVp 14.0 13.0 12.0 11.0 10.0 9.0 0% 50% 100% 150% % Vendor AGD 200% 250% 0% 50% 100% 150% 200% 250% % Vendor AGD 28 Slide 113 Slide 114 Contrast-Detail Scores vs. % Vendor AGD For 6 cm Breast Thickness 16.0 Contrast-Detail Scores Lorad 15.0 Vary kVp Mo/Mo – 29 kVp Siemens - Mo/Mo 28 kVp Siemens - W/Rh 28 kVp 14.0 13.0 12.0 11.0 10.0 9.0 0% 50% 100% 150% 200% 250% % Vendor AGD Slide 115 Slide 116 Summary Contrast-Detail Scores vs. kVp At Matched Average Glandular Doses to Film and Digital GE 2000D - Mo/Mo - Film GE 2000D - Rh/Rh - Film 16.0 GE 2000D - Mo/Rh - Digital 15.0 GE DS - Mo/Mo - Film 4.0 GE DS - Mo/Rh - Film 3.5 GE 2000D - Mo/Mo - Digital Average Glandular Dose vs. Breast Thickness Contrast-Detail Score vs. Breast Thickness GE 2000D - Rh/Rh - Digital 17 GE 2000D - Con/Auto GE DS - Mo/Rh - Digital GE DS - Rh/Rh - Digital GE Essential - Mo/Mo - Film 12.0 GE Essential - Mo/Rh - Film 11.0 GE Essential - Mo/Mo - Digital 10.0 GE Essential - Rh/Rh - Digital GE Essential - Rh/Rh - Film GE Essential - Mo/Rh - Digital Lorad - Mo/Mo - Film Lorad - Mo/Rh - Film 9.0 3.0 GE Essential - Std/Auto 2.5 Lorad - Auto-Filter Contrast-Detail Score GE DS - Mo/Mo - Digital 13.0 16 GE DS - Std/Auto GE DS - Rh/Rh - Film 14.0 AGD (mGy) Contrast-Detail Scores GE 2000D - Mo/Rh - Film 17.0 Siemens - OpComp - W/Rh 2.0 1.5 1.0 15 14 13 12 GE 2000D - Con/Auto 11 GE DS - Std/Auto 10 0.5 9 0.0 8 GE Essential - Std/Auto Lorad - Auto-Filter Siemens - OpComp - W/Rh 0 2 4 6 Breast Thickness (cm) 8 10 0 2 4 6 8 10 Breast Thickness (cm) Lorad - Mo/Mo - Digital Lorad - Mo/Rh - Digital 8.0 Siemens - Mo/Mo - Film 23 25 27 29 31 33 35 37 39 41 43 45 47 49 Siemens - Mo/Mo - Digital Siemens - W/Rh - Film Siemens - W/Rh - Digital kVp 29 Slide 117 Conclusions Conclusions What’ What’s responsible for different image Better calibration files on some digital manufacturer’ manufacturer’s systems quality scores? – Better calibration files on some digital manufacturer’ manufacturer’s systems fewer artifacts vs. – Different breast doses – Different postpost-processing algorithms Slide 119 Slide 120 Conclusions Dose Conclusions Image Quality – Digital in general has higher image quality scores – Digital in general has lower doses than film – For the same phantom, dose varies widely by mode – For the same phantom, image quality varies widely by mode – Breast thickness has big effect on image quality – Breast thickness makes big contribution to dose – Dose by thickness tracks pretty well across all vendors – Dose is affected by selection of mode being used – Image quality by thickness tracks pretty well across all vendors – Image quality is affected by selection of mode being used 30 Slide 121 Slide 122 Conclusions Conclusions Still to be done Take home messages – Phantom images scored by several readers – Dose makes more of a difference on image quality than – Analyzed for statistical differences and trends kVp – Digital has lower doses than Film What I didn’ didn’t mention – Some systems may set dose to low for their – Viewing conditions – Vendor QC recommended mode – Monitor calibration – Pay attention to what mode is used clinically – SNR Data Slide 123 Slide 124 Conclusions Take home messages Thank You – Digital mammography needs a more sensitive phantom – There is a wide range of image quality scores and doses across FFDM systems – Evaluate systems the way they are to be used clinically – Manufacturer only recommends AEC mode, user ultimately decides how the system is to be used 31