October October 1999 1999 QC R. QC for for Stereotactic Stereotactic Breast Breast Biopsy Biopsy -- R. Pizzutiello Pizzutiello Stereotactic Breast Biopsy Timely Tips for Medical Physicists Acknowledgements Thanks to ACR SBBAP Physics Sub-Committee, especially Maynard High, Ph.D. Allen Hrejsa, Ph.D. Andrew Maidment, Ph.D. For contributing valuable slides to this presentation Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc. Side View Lorad Focal Spot Side View Front View Focal Spot X-ray Tube X-ray Tube Small area Collimator Small area Collimator Compressed Breast Compressed Breast Phosphor Fiber Optic CCD Mirror CCD CCD Lens Front View Fischer Phosphor A/D Optical coupling/mirror system Light reflection from phosphor DMA A/D 2:1 fiberoptic taper demagnification Light transmission through phosphor Digital Image Quality Contrast Blur Noise Artifacts Dose copyright copyright Upstate Upstate Medical Medical Phyics, Phyics, Inc. Inc. (716) 0350 (716) 924 924--0350 DMA Noise Noise decreases (improves) with increasing mAs Images may be produced using any mAs technique (from 10 - 500 mAs) Window and level controls can be used to make the image “appear” properly exposed System noise will change 11 October October 1999 1999 QC R. QC for for Stereotactic Stereotactic Breast Breast Biopsy Biopsy -- R. Pizzutiello Pizzutiello 113 mrad 58 mrad Factors Affecting Breast Dose kVp, mAs thickness breast composition (dense or fatty) multiple exposures breast 28 kVp digital image processing does NOT affect dose optical density of film (if hardcopy is used) does NOT affect dose Let’s take a look at kVp dependence… 339 mrad 227 mrad 30 kVp 28 kVp To Minimize Breast Dose Develop and maintain a good technique chart Obtain manufacturer’s suggested techniques Evaluate image quality at different mAs values (Technologist and Medical Physicist) Moderately higher mAs will reduce image noise, but increase dose Insufficient mAs will produce a noisy (grainy) image, but can be made to appear “well exposed” with window/level control Excessive mAs images may also appear “OK” with window/level adjustment Minimize retakes 120 mrad 32 kVp 34 kVp QC Tests Unique to SBB Minimum Testing Frequencies Zero Alignment Test Before each patient (only on some units) Localization Accuracy Test (in Air) Daily Phantom Image Quality Test Weekly Hardcopy Output Quality Monthly (if hard copy is produced from digital data) Visual Equipment Check Repeat Analysis Compression Force Test Zero Alignment Test Perform before each patient that zero coordinate is accurate Assures that stereotactic unit is not improperly installed Verify Monthly Semi-annually Semi-annually copyright copyright Upstate Upstate Medical Medical Phyics, Phyics, Inc. Inc. (716) 0350 (716) 924 924--0350 B RT 22 October October 1999 1999 QC R. QC for for Stereotactic Stereotactic Breast Breast Biopsy Biopsy -- R. Pizzutiello Pizzutiello Setting Z zero on Lorad: Phantom Image Quality Evaluation Nuclear Associates Digital Mini Phantom Maynard High, Ph.D. Fibers Specks Masses ACR Accreditation NA Digital 1.56 1.12 0.8 0.75 0.54 0.54 0.4 0.32 0.24 0.16 2 1 0.75 0.5 0.25 x x 0.93 0.74 0.54 0.54 x 0.32 0.24 0.2 x 1 0.75 0.5 0.25 Mammography Accreditation Phantom RT Minimum Passing Phantom Image Scores Fibers Specks Masses ACR-MAP Accreditation Phantom MiniPhantom Screen/film Digital Digital 4.0 3.0 3.0 5.0 4.0 3.5 3.0 3.0 2.5 Be sure to use only an approved phantom Phantom Imaging: a common avoidable failure NAD Digital Mini Phantom 1st image (image quality) 2nd image (TLD) Mammo Accreditation Phantom 4 images for image quality 5th image for TLD OK to window/level digital images grid (or not) per clinical technique Use copyright copyright Upstate Upstate Medical Medical Phyics, Phyics, Inc. Inc. (716) 0350 (716) 924 924--0350 W 33 October October 1999 1999 QC R. QC for for Stereotactic Stereotactic Breast Breast Biopsy Biopsy -- R. Pizzutiello Pizzutiello Hardcopy Output Quality Measure optical density M Record or plot data Laser or multiformat camera SMPTE Test Pattern, if available Record window width, level Produce hardcopy Measure OD at a consistent location on each image Record and monitor for consistency Image Jan Feb Mar Evaluate RT Repeat Analysis S 1.62 1.62 1.63 3 1.59 1.57 1.61 4 1.61 1.60 1.58 5 1.63 1.64 1.62 6 1.60 1.61 1.62 SBB Annual Medical Physics Survey Count repeated and rejected film by category and tabulate Use a log of images repeated Document analysis and corrective action even if your repeat rate is low Repeat rate will typically be higher than for mammography MP LOOK at THIS BEFORE YOU START ! 1.63 1.61 1.60 2 Set limits (e.g.) 1.60 + 0.05 RT 1 SBB Unit Assembly Evaluation Collimation Assessment Focal Spot Performance and System Limiting Resolution kVp Accuracy and Reproducibility Beam Quality Assessment (HVL) Automatic Exposure Control System Performance Uniformity of Screen Speed or Digital Field Breast ESE, AGD, AEC Reproducibility Image Quality Evaluation (phantom) Artifact Evaluation Localization Accuracy Getting Started If you have an upright stereo, LET TRAINED TECHNOLOGIST MAKE ALL CONNECTIONS ! NEVER remove or install cables when power is on. copyright copyright Upstate Upstate Medical Medical Phyics, Phyics, Inc. Inc. (716) 0350 (716) 924 924--0350 44 October October 1999 1999 QC R. QC for for Stereotactic Stereotactic Breast Breast Biopsy Biopsy -- R. Pizzutiello Pizzutiello Assembly Evaluation Free-standing unit is mechanically stable All moving parts move smoothly, without obstructions to motion All locks and detents work properly Image receptor holder is free from vibrations Image receptor is held securely by assembly in any orientation MP Image receptor slides smoothly into holder assembly Compressed breast thickness scale is accurate to ± 0.5 cm, reproducible to ± 2 mm Patient or operator is not exposed to sharp or rough edges or other hazards Operator technique charts are posted Operator protected by adequate radiation shielding MP Collimation Assembly Evaluation X-Ray Field should extend beyond Image Receptor on all 4 sides X-ray Field should extend < 5 mm on any side (in plane of image receptor) Note: X-rays beyond the digital image receptor will not be seen on the monitor Does the biopsy window align with the image field of view? Lorad Collimation Assessment: Film cassette behind compression paddle BIOPSY WINDOW in compression paddle MP Collimation Assessment: Digital Image Measure visible diameter of coin with TOOLS/CALIPERS (Lorad). Anterior missing image is 19.0 – 17.6 = 1.4 mm Should be <5 mm BUT… How about the X-Ray Field ? CHEST WALL Incorrect or misadjusted collimator aperture 17.6 mm 19.0 mm Anterior copyright copyright Upstate Upstate Medical Medical Phyics, Phyics, Inc. Inc. (716) 0350 (716) 924 924--0350 Film behind compression paddle (does not show full extent of x-ray field) 55 October October 1999 1999 QC R. QC for for Stereotactic Stereotactic Breast Breast Biopsy Biopsy -- R. Pizzutiello Pizzutiello Focal Spot Size Performance System Limiting Resolution Line Pair Test Pattern Use film (x-ray machine) Use CRT image (“system”) Technique, clinical kVp Scoring the image Film - Lines distinct over 1/2 length CRT - Lines distinct, correct # over any part of pattern Need to measure in BOTH: 512 matrix 1024 matrix Need to measure BOTH: Must ZOOM Image: Parallel to A-C axis Perpendicular to A-C Typical with this Test object: kVp Accuracy Reproducibility 5.6 lp/mm (512 mode) 7.1 lp/mm (1024 mode) Beam Quality (HVL) Thickness of aluminum to reduce radiation exposure by one-half Affects contrast and dose Used in dose calculation minimum = kVp/100 No compression paddle lucite in the beam Verify that actual kVp’s are the same as the indicated kVp’s Range of clinical kVp values Accuracy within 5% Reproducible CV < 0.02 MP copyright copyright Upstate Upstate Medical Medical Phyics, Phyics, Inc. Inc. (716) 0350 (716) 924 924--0350 MP 66 October October 1999 1999 QC R. QC for for Stereotactic Stereotactic Breast Breast Biopsy Biopsy -- R. Pizzutiello Pizzutiello AEC System Performance MP Lorad: Matrix Size and Patient Dose AEC available on some digital SBB units Performance Capability Record signal level as function of thickness and technique Monitor exposure time Performance Capability (4,6,8 cm) Performance Criteria: a) 6, 8 cm DR# should be within 20% of 4 cm. b) DR# should meet Lorad target of 4000 (512) / 6000 (1024). Provide suggested technique chart For same kVp & mAs, and same amplifier gain, DR# in 1024 will be 1/4 DR# in 512. Lorad boosts amplifier gain 2X in 1024, so DR# in 1024 is 1/2 DR# in 512 at same dose Thus, for equal DR#, mAs needs to be doubled in 1024 mode. However… Develop a Technique Chart 6. Automatic Exposure Control (AEC) System or Manual Exposure Performance Lorad recommends HIGHER target DR# for 1024: Performance Capability DR# = 4000 in 512 mode DR# = 6000 in 1024 mode Thickness Compensation Imaging mode Digital 512 - Manual mode Focal spot: Large Prepare for both 512 and 1024 modes (Lorad) mA: Phantom thickness Therefore input exposure must be increased in 1024 mode by: 2 * (6000/4000) = 3 times !! mAs Target 28 56 Mo Mo 4269 6 cm 30 140 Mo Mo 4361 8 cm 34 175 Mo Mo 4491 4 cm value 4269 Uniformity of Screen Speed or Digital Field Image a uniform phantom Screen Film systems Each cassette produces the same optical density under the same conditions Digital Systems Digital detector produces uniform signal values across the field of view Mean Signal Value kVp 4 cm Filter E v a lua tio n Density or Signal Range Allowable Range 4269 - 4491 3415.2 - 5122.8 Pass Digital Receptor Uniformity: Lorad Protocol 28 kVp mAs for DR# =4000 Measure SNR’s with TOOLS/STATS at specified locations. 32 x 32 pixel ROI – set with trackball. Lorad spec +/-20% of SNR(center). SNR SNR UL UR (100, 100) (400, 100) SNR Centr (256, 256) SNR LR (100, 400) (400, 400) MP copyright copyright Upstate Upstate Medical Medical Phyics, Phyics, Inc. Inc. (716) 0350 (716) 924 924--0350 SNR LL 77 October October 1999 1999 QC R. QC for for Stereotactic Stereotactic Breast Breast Biopsy Biopsy -- R. Pizzutiello Pizzutiello Lorad “Hidden” Statistics Functions: NT System ROI Statistics: S Press “S” displayed 1) 2) key on keyboard when ROI is to obtain: STANDARD DEVIATION SIGNAL / NOISE NT System ROI Statistics: Phantom Image Quality Same procedure as for technologists Medical Physicist reviews scoring procedure and checks for consistency Uses technique factors for dose determination MP Breast Entrance Exposure, AGD Artifact Evaluation Unwanted irregularity not caused by structures of interest Causes (Digital) Digital Image Receptor Data per technique chart Measure ESE HVL determines DgN AGD = ESE * DgN AGD < 300 mrad Dose and Optical Density corrected by white-fielding MP copyright copyright Upstate Upstate Medical Medical Phyics, Phyics, Inc. Inc. (716) 0350 (716) 924 924--0350 Dust (camera, screen, lens, mirror) Pixel defects (dropouts) Non-uniformities (light pipe stucture, vignetting, linear shading) Clipping (dose too high) MP 88 October October 1999 1999 QC R. QC for for Stereotactic Stereotactic Breast Breast Biopsy Biopsy -- R. Pizzutiello Pizzutiello On Lorad, “White-Field” is a service procedure: Dust Speck Artifact: - Dust Moved after White-Fielding Can correct some non-uniformities Consists of the average of 10 exposures on 4-cm lucite phantom Correction formula is: (Image – Dark Field) (White Field – Dark Field) “Moved” Dust Artifacts “Moved” Dust Artifacts White Speck next to Black Speck White Speck next to Black Speck Targeting Accuracy Gel Phantom Pre-Fire Images: Correct placement for needle: Performed annually by technologist under supervision of medical physicist Position gel-type phantom Image, target and sample Result: was the lesion collected? MP copyright copyright Upstate Upstate Medical Medical Phyics, Phyics, Inc. Inc. (716) 0350 (716) 924 924--0350 99 October October 1999 1999 QC R. QC for for Stereotactic Stereotactic Breast Breast Biopsy Biopsy -- R. Pizzutiello Pizzutiello Role of the Surgeon in QC Understand the importance of QC in SBB Assures that personnel remain qualified Support QC activities Allow enough time for QC Provide for QC training Periodically check that QC is done as required Confer with medical physicist annually Assure that follow-up is done if the QC program indicates corrective action is required Accreditation copyright copyright Upstate Upstate Medical Medical Phyics, Phyics, Inc. Inc. (716) 0350 (716) 924 924--0350 10 10