Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments Root-Causes of CT Number Inaccuracies Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison • Nature of the X-ray Physics • New Technology scanner – Beam Hardening – Scatter – Aliasing – Helical – Cone Beam – Incomplete Sampling • Patient • Operator operator – Motion – Photon Starvation – Protocols (scan thin, recon thick) – Partial Volume patient 1 2 Nature of Physics: Beam Hardening measured poly-energetic x-ray photon energy-dependent m(E) measured projection error polynomial correction N p' n p n n 1 channel 0 cm water • The attenuation characteristics of bone is significantly different from that of soft tissue. Similar observation can be made for contrast agents. 1000 m (1/cm) • • • • • m(E) intensity ideal 10000 Bone Beam Hardening 100 bone 10 water 1 0.1 0 50 100 150 energy (keV) energy, E 15 cm water 30 cm water no correction with correction 3 Water BH Error Bone BH 4 1 Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments Dual Energy Correction Impact of Iodine Contrast • Dual energy takes advantage of the two types of x-ray interaction with matter: photoelectric and Compton. • Use of the additional information can effectively remove BH. • Iodine contrast – Cardiac phantom experiments – Dual energy solution Conventional 120kVp Dual Energy 70keV 100000 10000 m (1/cm) 1000 Iodine 100 10 Bone 1 Soft-tissue 0.1 0 80 kVp 140 kVp Perfusion 100 150 Cardiac Phantom energy (keV) Projection-space DECT ww=300 Regular 140 kVp 50 5 6 DE Mono 70 keV Metal-induced Error • Perfusion is important to test the viability of a tissue • • Perfusion requires accurate quantitation of CT numbers • • BH is one of the major sources of error CT is sensitive to the presence of high-density and highfrequency objects. Special care must be taken to avoid placing foreign objects inside the scanner FOV. 0.625mm thickness, 1.0mm apart 140kVp Images courtesy of Drs. T. Lee and A. So of St. Mary’s Hospital, London, OT, Canada metal reinforcement 7 8 2 Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments Nature of Physics: Scatter • • Scatter-to-Primary Ratio (SPR) increases almost linearly with the z-coverage. SPR decreases with the increase in kVp SPR increases with the object size • • X-ray source Scatter Artifacts • • Scatter-to-Primary Ratio (SPR) increases almost linearly with the z-coverage. SPR decreases with the increase in kVp SPR increases with the object size z scatter scatter M e a s u re d S P R in V C T /H D a t 4 0 m m c o v e ra g e z ghost image 0 .5 120kVp 48cm Poly 0 .4 30 25 0 .2 SPR 8 0 kVp 1 0 0 kVp 0 .1 1 2 0 kVp 20 15 10 0 shading 5 15 25 35 45 55 w a te r d ia m e te r (c m ) 0 0 50 100 Beam Width (mm) 150 200 9 Lung CT Number Accuracy Inspiration Level Beam Hardening – Increased cone beam effect – Increased scatter -980 Sagittal view -985 32mm Helical -990 Scatter – Increased volume coverage – Insufficient post-patient collimation 10 • Air CT number can be impacted by – Patient centering – Variation of object size – Presence of non-water materials • 160mm Detector Air-Lung Accuracy Lung Volume • • 32mm Detector -995 Axial CT Number CT Number (HU) S PR 35 0 .3 -1000 -1005 -1010 -1015 160mm Axial -1020 JP Sieren, PF Judy, DA Lynch, JD Newell, HO Coxson, and EA Hoffman, “QIBA Quantitative CT: Towards routine quantitative CT in obstructive lung disease,” QIBA COPD/Asthma Subcommittee -1025 -1030 1 11 51 101 151 201 12 3 Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments Truncation Issue Scatter Correction • • • • • For x-ray radiography, truncation does not cause artifact inside FOV. • Truncated projection will cause significant artifacts for CT Increased collimator aspect ratio: H/W 2D collimation Algorithmic correction possible Phantom calibration COPDGene Phantom (CTP657) W Chest X-ray H z x Full FOV original 14 Size Matters • CT is sensitive to both high density object and objects extended outside the FOV. • Patient arms need to be positioned outside the FOV. Truncated corrected 13 Photon Starvation 50cm FOV • Beer’s law indicate that the amount of attenuation increases exponentially with path length. I e mL I0 • • Scout Image CT Image 15 At low signal level, the noise in the projection is no longer dominated by the x-ray photon. Convolution filtering operation will further amplify the noise and streak artifacts will result. 16 4 Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments Protocol Optimization Protocol optimization – – – – – – Scan speed Helical pitch kVp selection Slice thickness Reconstruction kernel mA modulation m (1/cm) • • Adaptive filtering for streak reduction 10 Bone 1 soft-tissue 0.1 0 30 60 90 120 150 energy (keV) 80kVp, 6.4mGy adaptively filtered 140kVp, 6.3mGy 17 Iterative Reconstruction 120 kV, 10 mA, 0.5 rot time 18 Sinogram Cardiac Motion 16-slice Examples Dose report FBP original Algorithmic Correction 100 MBIR 0.625mm slice thickness • • • Complexity of cardiac motion Artifacts from several factors In-plane vs. cross-plane temporal resolution .78mSv* phase mis-registration sub-optimal gating Images courtesy of Dr Barrau, CCN, France *Obtained by EUR-16262 EN, using an abdomen factor of 0.015*DLP and a pelvis factor of 0.019*DLP In-Plane sub-optimal table speed Cross-Plane 20 5 Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments Artifact Correction • Advanced Technology-induced Artifact In-Plane Temporal Resolution Improvement: – – – – • • • • • Phase optimization Faster scan speed Multiple source-detector Advanced reconstruction Half-scan: p + fan angle acquisition. Dual source reduces the acquisition by 40-45% Two detectors have different size Object outside small FOV may cause artifacts. Compensation algorithm can reduce the impact centered phantom half-scan sub-optimal phase • • • • optimal phase 25 g at 0.35 s 8X safety margin 200 g 76 g at 0.2 s 8X safety margin 612 g • • Cone Beam Axial Axial Truncation desired ROI detector ww=600 z source trajectory phase mis-registration 22 • Exact cone-beam reconstruction does not exist. • Large cone angle can lead to cone-beam artifact. Increase z-coverage will reduce phase mis-registration artifacts. Entire heart covered by a single rotation. Step-and-shoot mode artifacts are due to: – Missing frequencies – Mis-handled frequencies – Z-truncation smaller detector FOV 21 Cross-Plane Temporal Resolution: Cone Beam CT • off-centered phantom 23 Small Detector Helical Mode Large Detector Step-and-shoot 24 6 Cone Beam Axial Temporal Resolution Improvement: Axial, Phantom Advanced Reconstruction Algorithms • Exact cone-beam reconstruction does not exist for step-and-shoot mode. • Large cone angle can lead to cone-beam artifact. • Active research in on-going Characterize cardiac motion Compensate for motion Coronal, Simulation Advanced Axial, Phantom • • Conventional Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments Conventional Rest/Low HR 25 Conventional Stress/High HR Advanced Stress/High HR High HR + motion correction 26 Conclusion • There are many sources of error for quantitative CT – – – – Nature of Physics New Technologies Patient Operator • Many approaches are taken to reduce or eliminate the error. • It takes a combined efforts from manufactures, CT operators, and patient corporation to reduce the impact of these pitfalls to a minimum. 27 28 7