Use of Volumetric Modulated Arc Therapy for Intra and Extra Cranial Stereotactic Radiosurgery Part III – SmartArc & QA Peter Balter, Ph.D. Pinnacle SmartArc • Clinically released in Ver 9 • Supports multiple linac vendors • New parts of photon machine setup • New SmartArc tab in IMRT screen SmartArc Optimization • Optimized around the entire arc simultaneously • Uses pinnacle DMPO framework • Optimization includes “high quality” convolution dose calculations (important in area of disequilibrium ie Lung SBRT) • Constraints are not allowed Optimization approach • SmartArc starts with 24o spacing around the arc • At each position apertures are set to the target and then optimized fluences are generated by gradient search Bzdusek, K, et al “Development and evaluation of an efficient approach to volumetric arc therapy planning,” Med. Phys. 36, 6, June 2009 Optimization approach • At each angle 2-4 segments are generated • Largest 2 segments are chosen and the other 1-2 discarded • The remain 2 are distributed to the left and the right (1/3 of the initial spacing = 8o) in such a way as to minimize leaf travel • The MLC settings at the original angles are filled in by creating new MLC positions by linear interpolation of the distributed segments. • This results in segments every 8o = 45 control points for Machine Parameter optimization • More interpolated segments are inserted to achieve the user-selected final spacing (generally 4o) these are used for dose calculation but are not independently optimized Machine Characterization Flexible settings to allow support of multi-vendor VMAT Settings should come from linac vendor’s documentation We set these values based on defualts in Eclipse (RapidArc), these are more conservative than the example values in the earlier referenced paper by Bzdusek (table 3). Other important machine setting • Maximum MU setting – 999 MU limit may prevent optimization for some beams • Allow beam MU to exceed maximum should be used (any beams above maximum must be split prior to export) • Varian sells a >999 MU upgrade – important for SBRT MOSAIQ Setup Small changes in machine characterization • Rapid Arc machine • Conformal Arc Machine • IMAT (Gantry Index) = YES • IMAT (Gantry Index) = Yes • VMAT (MU Index) = YES • No VMAT section (allows conformal arc and VMAT) • Allows conformal arc only SmartArc tab in IMRT parameters • Arcs/beam • Final gantry spacing (deg) • Start/Stop angle • Compute intermediate dose • Compute final dose • Maximum delivery time • Constrain leaf motion Pinnacle Smartarc IMRT objectives • Same interface we know and love • “Constrain” checkbox cannot be used with SmartArc • Objectives can now be adjusted during optimization Comparison SBRT Case Example RTOG 0915 34Gy x 1 • • • • Clinical SBRT (used) 11 beams – 2 non coplanar Planning time (3 hours) Total MU: 8548 Delivery time: 23 minutes SmartArc (Planning Study) • 2 Arcs coplanar • Planning Time (30 minutes) • Total MU: 7969 • Estimated delivery time 12 minutes Comparison: SBRT Case Clinical IMRT Plan SmartArc Plan Comparison: SBRT Case Clinical IMRT Plan SmartArc Plan Comparison: SBRT Case Clinical IMRT Plan SmartArc Plan Comparison: SBRT Case Solid: Clinical Plan Dashed: SmartArc Both plans meet the RTOG 0915 Guidelines for 34 Gy in 1 fx Patient setup for VMAT SBRT treatment • Direct Tissue targeting should be used – CBCT can target soft tissue • Can be used with BH gating • Cannot be used (in current form) with dynamic gating (but neither can RapidArc) – Fiducials • Can be used with traditional gating or BH How does VMAT affect PTV margins • Random and Systematic components: – Systematic • Alignment of kV(CBCT) and MV isocenters ( 1 mm) • Random sampling nature of initial sim PTV CTV GTV – Random • Interpretation of daily CT vs planning CT • Patient motion during treatment • Deformation of tumors/organs • Uncertainties in respiratory management ITV How does VMAT affect PTV margins • Only the patient motion component is affected: – Shorter time between imaging and treatment should allow shrinking of PTV margins – No data to demonstrate this for lung SBRT • Patient’s more willing to comply if treatment time is presented as shorter • It is possible that simpler immobilization (or no immobilization) could be used – simplify simulation process • This needs to be studied further Alignment accuracy for SBRT • Alignment of Isocenter of setup modality vs treatment beams must be known • Initial Verification – End to end test • Monthly QA – Alignment check (winston-Lutz type test) • Daily QA – Check for gross changes ( 2 mm level) End-to-end check • A phantom is scanned, sent to TPS, planned, sent to R&V, sent to Linac, Imaged, treated. • RPC credentialing for RTOG SBRT protocols requires this test. • This should be done for all patient orientations used (head-first supine, head first prone, etc). • This should be done prior to first patient treatment Some type of end-to-end check should be done for any setup system being used for SBRT patients Phantom designed to check congruence of isocenters: MV, kV, CBCT, Light Field, OGP. Enables end-to-end testing using treatment conditions. Align using kV / CBCT to center on cerrobend dot. Place Gafchromic film and treat SRT beams. Scan the film to ensure treatment isocenter matched the setup imaging techniques Charles Mayo, Ph.D Monthly check should also be done to verify nothing has changed BB Radionics target pointer Radionics BB on a stick used to check MV, kV and CBCT isocenter Radiation isocenter is determined using the the center of a MLC created shape imaged at multiple angles Software automatically determines the accuracy of the kV and MV imager and the CBCT. 1. Du W, Yang J, Luo D and Martel M. A simple method to quantify the coincidence between portal image graticules and radiation field centers or radiation isocenter. Med Phys. 2010; 37: 2256-2263. Daily QA • Daily QA – Look for gross changes due to sudden failures (or unexpected maintenance) – Should have achievable (2 mm) accuracy while not significantly slowing down the morning warm-up procedure Pre-treatment patient specific QA – Record and Verify • All parameters are correct for all fields – Imaging • Integrity check of data in 3D-3D matching system • Verification of reference images for 2D matching – Dosimetry • Measurement based IMRT QA done, reviewed and signed by physicist and attending before the first fraction Per fraction QA • Physicist present (at a minimum) first day • Attending present for all fractions – required for billing • MV portal images for isocenter verification can serve as a final check – Verification at the 3-5 mm level – Checks that • Data in the R&V system for the isocenter was correct • Couch shifts were done correctly • Patient did not move during after the CBCT • Not gross error occurred in the entire process SBRT OAR Guidelines SBRT is an ablative procedure: Care must be taken to control normal tissue toxicity OAR Guideline should be: 1) A function of the number of fractions 2) Based on published guidelines based on heterogeneous calculations when ever possible 3) Lower than traditional fractionation 4) Updated based on follow-up Example: RTOG 0915 34 Gy x 1 12 Gy x 4 Change in fractionation can cut in half the dose constraints Respiratory Management with VMAT • Methods of respiratory motion management – Dynamic gating – Breath-hold gating (either voluntary or ABC) – ITV (treating the track of tumor motion) • Abdominal compression is a special case of ITV as it reduces rather than eliminates motion Breath-hold gating • Current RapidArc machines – Can perform CBCT under a series (2-4) of breath-holds – Can deliver RapidArc under a series of BHs • Probably not recommended by the manufacturer • We were not able to measure any dose differences between an uninterrupted delivery and one started and stopped many times • This should be done with caution ITV treatments • Margins expanded to cover track of tumor motion • Based on 4DCT • Interplay affects are minimal (as pointed out in previous talk) • Expected to be lower than for traditional IMRT due to • Slow gantry rotation • Smooth motion of MLC SAMS Question 1 Which method of respiratory motion management can be currently implemented with Rapid Arc 20% 20% 20% 20% 20% 1. 2. 3. 4. 5. Tumor tracking using MLCs Tumor tracking using couch motions Gating during normal respiration Breath hold gating or 4DCT based ITV targeting No motion management is possible 10 SAMS Question 1 • 1. 2. 3. 4. 5. Which method of respiratory motion management can be currently implemented with Rapid Arc Tumor tracking using MLCs Tumor tracking using couch motions Gating during normal respiration 4DCT based ITV targeting or breath hold gating No motion management is possible Answer: 4) 4DCT based ITV targeting or breath hold gating Wilko F.A.R. Verbakel, WF,” Rapid delivery of stereotactic radiotherapy for peripheral lung tumors using volumetric intensitymodulated arcs.” Radiotherapy and Oncology 93 (2009) 122–124 and Measurements done at MDACC (not yet published) SAMS Question 2 How can VMAT affect the PTV margins used for IGRT 20% 4. It may allow PTV reductions due to the shorter delivery time It may require PTVs to be expanded due to the smearing of dose by the rotation It does not affect PTV margins as PTVs are a function of setup techniques and immobilization and are included in the planning process The rapid falloff allows smaller PTV margins 20% 5. The rapid falloff requires larger PTV margins 20% 20% 20% 1. 2. 3. 10 SAMS Question 2 How can VMAT affect the PTV margins used for IGRT 1. It may allow PTV reductions due to the shorter delivery time 2. It may require PTVs to be expanded due to the smearing of dose by the rotation 3. It does not affect PTV margins as PTVs are a function of setup techniques and immobilization and are included in the planning process 4. The rapid falloff allows smaller PTV margins 5. The rapid falloff requires larger PTV margins Answer: 3) It does not affect PTV margins as PTVs are a function of setup techniques and immobilization and are included in the planning process Sonke, J.J. et al “Frameless Stereotactic Body Radiotherapy for Lung Cancer Using Four-Dimensional Cone Beam CT Guidance”, Int J. Radiation Oncology Biol Phys, Vol 74, No 2, pp 567-574, 2009 SAMS Question 3 Planning guidelines for SBRT in the lung 20% 1. Are the same as for traditionally fractionated therapy 20% 20% 20% 20% 2. 3. 4. 5. Don't need to be evaluated due to the rapid dose falloff from VMAT Should be taken from RTOG 0236 (The initial SBRT lung trial) Can be reduced when rapid arc is used for treatment delivery Should be taken from published data calculated with heterogeneity corrections and with the same number of fractions 10 SAMS Question 3 Planning guidelines SBRT in the lung 1. Are the same as for traditionally fractionated therapy 2. Don't need to be evaluated due to the rapid dose falloff from rapid arc 3. Should be taken from RTOG 0236 (The initial SBRT lung trial) 4. Can be reduced when rapid arc is used for treatment delivery 5. Should be taken from published data calculated with heterogeneity corrections and with the same number of fractions Answer 5) Should be taken from published data calculated with heterogeneity corrections and with the same number of fractions • Ref: Videtic, G.M, Singh, A.K., Chang, J.Y, Quynh-Thi, L, Parker,W, Olivier, K.R., Schild, S.E., Bae, K. “RTOG 0915: A RANDOMIZED PHASE II STUDY COMPARING 2 STEREOTACTIC BODY RADIATIONTHERAPY (SBRT) SCHEDULES FOR MEDICALLY INOPERABLE PATIENTS WITHSTAGE I PERIPHERAL NON-SMALL CELL LUNG CANCER”, Radiation Therapy Oncology Group, Philadelphia, PA, 2009 Conclusions (Senan & Verbakel) • Fast lung SBRT in <6.5 mins delivery time – Total linac time (+ CBCT set-up) 20 minutes – Superior OAR sparing possible – Less chance for intrafraction motion • No interplay effect between moving tumor and moving leaves • More time for appropriate and efficient IGRT Summary • VMAT is a new technology that will require new vigilance in QA – Synchronization of dose rate controls, gantry and MLC speed – Thing we don’t know yet that may go wrong • Hypo-fraction requires increased vigilance in QA – Tight margins – Limited (or no) averaging of random setup errors – Limited opportunities to recover if an error is caught mid-treatment Final thoughts • Hypo-fractionated radiation therapy – Has the potential to do real immediate harm to the patient – May result in more severe morbidity than traditional XRT – Has little room for error – Is more akin to surgery than XRT and the same degree of care is required • VMAT and CBCT are enabling technologies that make hypo-fractionation easier to implement Acknowledgments • • • • • • • • • • Charles Mayo, Ph.D. Lei Dong, Ph.D. Weilang Du, Ph.D. Milos Vicic, Ph.D. Rebecca Howell, Ph.D. Song Gao, Ph.D. Ramaswamy Sadagopon, MS Oleg Vassiliev, Ph.D Karl Prado, Ph.D. Issac Rosen, Ph.D. • • • • • • • • Kara Bucci, M.D. Joe Chang, M.D., Ph.D. James Cox, M.D. Thomas Guerrero, M.D., Ph.D. Melinda Jeter, M.D., M.P.H. Ritsuko Komaki, M.D. Zhongxing Liao, M.D. Mary Frances McAleer, M.D, Ph.D. • James Welsh, M.D.