FFDM -FCRm QC RequirementsWhat You REALLY Need to Know Melissa C. Martin, M.S., FACR, FAAPM, FACMP FDA Approval for Mammography • Fuji FCRm System approved for Mammography on July 10, 2006 by the FDA as a FFDM Digital Mammography System AAPM Annual Meeting July 28, 2008 CR Technology Fuji Mammo CR Protective Layer Photo-stimuable phosphor layer •Pixel size 50 micron Support •Use dual sided reading to DQE •Imaging Plate has thicker phosphor layer •Results in 40-50% increase in NEQ (DQE) •Post processing enhances image (e.g. filtering to recognize and enhance calcifications) J. Ed Barnes, Ph.D. - MTMI J. Ed Barnes, Ph.D. - MTMI 1 FCM Enlarged Images 1mm Standard System (100 micron pixel) 1mm J. Ed Barnes, Ph.D. - MTMI Fuji System Components ClearView-1m FCRm (50 micron pixel) Dual side scan Fuji Clearview CSm Reader Breast Imaging Review Station ClearView-CSm J. Ed Barnes, Ph.D. - MTMI • Four cassette stacker • Processes 20 screening exams/hour (or 100 CR plates/hour) • Multi-objective Frequency Processing (MFP) (enhances dense/peripheral tissue) • Pattern Enhancement Processing for Mammography (PEM) (enhances calcifications) Flash Plus IIPm J. Ed Barnes, Ph.D. - MTMI J. Ed Barnes, Ph.D., - MTMI 2 Other CR Mammography Venders - World Wide •Agfa •Kodak •Konica •Canon Kodak Mammography CR •Single sided plate •Thinner phosphor layer (improves resolution, reduces DQE) •Phosphor grain smaller and homogenous •BaFBr:Eu phospor •48.5 micron pixel •Read out speed reduced •Twice as slow •Increases SNR •Use conventional reader (modified) J. Ed Barnes, Ph.D. - MTMI Agfa CR Mammography J. Ed Barnes, Ph.D. - MTMI Konica CR Mammography •Single sided plate, rigid •Single sided plate, flexible •BaFI:Eu phosphor •BaSrFBrI:Eu phosphor •Smaller grain powdered phosphor •43.75 micron pixel •50 micron pixel •Utilize “Phased Contrast CR” •Scan slower than conventional CR •Expose breast on a mag platform to a 14x17 cassette •Wider light collection guide •Thinner plate to increase resolution J. Ed Barnes, Ph.D. - MTMI •Demagnify to a 24x30 cm image (air grid) J. Ed Barnes, Ph.D. - MTMI 3 PSP Radiography (CR) CR Mammography Systems • Currently the major technology available for large field-ofview digital imaging for AGFA, FUJI, KODAK, KONICA “ All measurements were equivalent to or exceeded the performance of a typical modern film/screen system” • Based upon the principles of photostimulated luminescence; 20+ years of experience • Operation emulates the screen-film paradigm in use and handling.. (flexible but labor intensive) • Manufacturing trends: – Smaller, faster, less expensive From Centre for Evidence-based Purchasing Report 06047, October 2006 J. Ed Barnes, Ph.D. - MTMI J. Seibert - UC Davis Computed Radiography “reader” reader” PSP Detector • Photostimulable Storage Phosphor (PSP) Phosphor Plate Fuji Cassette Holder Information panel 8 eV PSL e- 3 eV F center trap 2 eV ee Laser stimulation Plate stacker Incident xx-rays • Coating thickness: – Standard resolution: ~100 µm BaFBr – High resolution: ~50-70 µm BaFBr • Dual-side read; structured phosphor – CsBr J. Seibert, Ph.D. - UC Davis J. Seibert - UC Davis 4 CR “system” system”: more than the IP’ IP’s and the reader!! Image Acquisition Digital Image Networking • Modality Worklist Input (from RIS via HL-7) • Technologist QC Workstation CR QC Workstation DICOM / PACS Laser film printer CR Reader • PACS and DICOM – Digital Imaging Communications in Medicine – Provides standard for modality interfaces, storage/retrieval, and print • DICOM image output J. Seibert UC Davis Display / Archive J. Seibert - UC Davis Photostimulated Luminescence Incident Laser Beam Light guide CR PointPoint-scan readout PSL Signal PMT Exposed Imaging Plate – Image manipulation processing reconciliation – Reconciliation and image QA Light Scattering Photostimulated Luminescence Reference detector f-θ lens Laser Source Light channeling guide Polygonal Mirror Output Signal ADC Protective Layer Phosphor Layer Cylindrical mirror Laser beam: Scan direction PMT x= 1279 To image y= 1333 processor z= 500 Laser Light Spread "Effective" readout diameter Base Support J. Seibert - UC Davis Plate translation: SubSub-scan direction J. Seibert - UC Davis ERASURE Reuse 5 Phosphor Plate Cycle PSP SubSub-scan Direction Typical CR resolution: 35 x 43 cm -- 2.5 lp/mm (200 µm) 24 x 30 cm -- 3.3 lp/mm (150 µm) 18 x 24 cm -- 5.0 lp/mm (100 µm) Base support x-ray exposure Plate translation plate exposure: create latent image reuse laser beam scan plate readout: extract latent image light erasure Screen/film resolution: Scan Direction plate erasure: remove residual signal 7-10 lp/mm (80 µm - 25 µm) Laser beam deflection J. Seibert - UC Davis J. Seibert - UC Davis CR Innovations Reference detector Laser Source • Dual side readout Mirror • High-speed line scan systems (<10 sec) Light Erasure • Low cost table-top CR readers J. Seibert - UC Davis Output Signal To image processor ADC To image processor • Structured PSP PMT ADC CR cassette • Mammography applications f-θ Cylindrical mirror lens Light channeling guide Erasure Stage Transport J. Seibert - UC Davis 6 CR Mammography CR Plates • Dual-side readout capabilities (increased DQE) • Fine laser beam spot size, 50 µm (10 lp/mm) • Mammography-specific image processing Conventional Mammo system CR reader QC workstation Low attenuation cassette, special IP Fuji CRm System Transparent base CR Plate Reader 50 µm spot 24x30 18x24 2 light guides CR cassette Two cassette sizes Scanning laser beam Medical Physics Annual Tests Section F = Medical Physicist’s QC Tests • • • • • • • Mammographic Unit Assembly Evaluation Collimation Assessment S-Value Confirmation System Resolution kVp Accuracy and Reproducibility HVL Measurement Breast Entrance Exposure and Glandular Average Dose • Radiation Output Rate • Phantom Image Quality Evaluation - Printed Images - Softcopy 7 Medical Physics Annual Tests • • • • • • • • • • AEC System Performance System Artifact Evaluation CR Reader Scanner Performance Dynamic Range Primary Erasure - Additive and Multiplicative Lag Effects Inter Plate Consistency Viewing and Viewing Conditions Review Workstation Tests Printer Tests Technologist’s QC FCRm Collimation Test for Routinely Used: Collimator, Compression Paddle, Bucky Combinations 1. Mammographic Unit Assembly Evaluation Free-standing unit is mechanically stable All moving parts move smoothly, without obstructions to motion All locks and detents work properly Image receptor holder assembly is free from vibrations Image receptor slides smoothly into holder assembly Image receptor is held securely by assembly in any orientation Compressed breast thickness scale is accurate to +/-0.5 cm, reproducible to +/-2 mm Patient or operator is not exposed to sharp or rough edges, or other hazards Operator technique control charts are posted Operator protected during exposure by adequate radiation shielding All indicator lights working properly Auto decompression can be overridden to maintain compression (status displayed) Manual emergency compression release can be activated in the event of a power failure Pass/Fail/NA Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Comments: The following four items satisfy MQSA Final Regulations: 1.) Upon power interruption, the gantry remains stationary and stable. 2.) Under compression, the paddle deflection was: 18x24 cm: 3 mm; 24x30 cm: 4 mm. 3.) The collimator light illuminance was 420 lux. 4.) The compression paddle indicates the location of the AEC sensor. Compression is set at 31 pounds. The technique chart above reflects the results of our measurements and should be posted on the unit. The unit will be used clinically in the Auto-Filter mode. Fuji CRm System S-Value Confirmation Measure Exposure Rate at Surface of Breast Support Target Materials 8 1 - S Value Confirmation Date: November 21, 2007 Exposure Unit ID: Lorad M-IV Image Reader ID: Fuji ClearView CSm Is the 95% uji ClearView CS FCRm S Value Confirmation • Make exposure at 25 kVp with Mo/Mo to give exposure > 20 mR • Expose cassette at this technique - wait 10 minutes and process • Record S value • Calculate the “corrected S value” Action Limit: Corrected S value should not exceed the range of 120 +/- 20% (96 to 144) FCRm System Resolution kVp: mAs: Exposure 1 (mR): Exposure 2 (mR): Exposure 3 (mR): Average Exposure: 25 3.0 37.0 37.1 37.0 37.0 IP Exposure Time 1: IP Exposure Time 2: IP Exposure Time 3: 2.0 2.0 2.0 S Value 1: S Value 2: S Value 3: Average S Value: 66.0 68.0 65.0 66.3 Corrected S Value: 122.8 Corrected S Value Pass/Fail? Corrective Action Needed? Description of Corrective Action: Corrective Action Taken? (Yes, No?) Date of Corrective Action: Results of New Test: (Pass, Fail?) System limiting resolution must be 8 lp/mm +/- 2 lp/mm (6 to 10 lp/mm) IP Read Time 1: IP Read Time 2: IP Read Time 3: Yes 2 min 2 min 2 min PASS NO Comments: 2. Evaluation of System Resolution X-ray Tube Manufacturer: • Up to 10 lp/mm Test Pattern • 4 cm Acrylic Phantom • Parallel and Perpendicular to Tube with 3-5 degree angle Exposure readings within 5%? Model #: Varian (1/02) B-115/M-113R Fuji CRm Visible black points: Anode material Nominal kVp setting Nominal mA setting Density control setting mAs S Value (Digital) Magnification factor Limiting resolution in line-pairs per mm Action Limit: bars parallel to A-C axis bars perpendicular to A-C axis No No Mo Mo 27 30 0 57 58 1.80 10 10 25 100 0 124 63 CONTACT 7 7 For Fuji FCRm, if the limiting resolution with the bars parallel to the anode-cathode axis is < 8 +/2 line-pairs/mm or perpendicular to the anode-cathode axis, then a more detailed investigation of the reason should be made and corrective action should be taken. 9 FCRm Density Control Function Density Control Function: Imaging mode: mA: Phantom thickness: 4 cm Acrylic, Accreditation Phantom Technique Repeat at -2, -1, 0, 1, 2, etc. density settings Record mAs mAs Changes should be 5 to 15% per step Density Selector Setting 5 4 3 2 1 0 -1 -2 -3 -4 -5 Action Limit: FCRm Reproducibility and Image Mode Tracking Film/Screen Focal spot: 0.3 mm MO kVp: 4 cm BR-12 Cassette ID: QC Relative to Normal % mAs Density mAs Change Change 255.0 227.0 201.0 187.0 166.0 146.0 125.0 109.0 98.3 86.3 75.4 11.0% 11.5% 7.0% 11.2% 12.0% 0% 14.4% 12.8% 9.8% 12.2% 12.6% Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Each step should result in a 5 to 15% change in mAs. If not, seek service. 4. Automatic Exposure Control (AEC) System Performance #3 QC AEC position: Small cassette ID: Density control: Large cassette ID: 0 24 x 30 Performance Capability: • • • • • 4 cm Acrylic with Clinical Technique Position Ion Chamber in Beam Record mAs and Exposure Repeat 3 Times Repeat in each Mode (small, large, mag and no grid) Action Limits: Coefficient of Variation for Exposure must not exceed 0.05 No Significant Difference in Exposure between small and large bucky when using similar grid CNR PER OBJECT THICKNESS Image Mode: Focal spot: Density Control Phantom thickness 2 cm 4 cm 6 cm 8 cm Auto Time Large 0 Target/Filter kVp mAs (Auto) CNR CNR (relative to 4cm PMMA) Acceptable Level Mo/Mo Mo/Mo Mo/Rh 25 25 30 26.5 131 218 16.3 14.7 12.5 1.11 1.00 0.85 >100% 100% >75% PASS Image Mode Tracking mA: Phantom thickness: Anode/Filter: Image Mode small grid large grid Magnification/no grid 100 4 cm BR-12 Mo/Mo AEC MODE Target Filter kVp mAs Comments Auto Filter Auto Filter Auto kV Mo Mo Mo Mo Mo Mo 26 25 27 146.0 113.0 57.7 HTC Grid HTC Grid No Grid 10 Fuji CRm System FCRm Dynamic Range • To confirm dynamic range of reader and plate • Use 2 and 4 cm acrylic • Technique for 6 cm • View image to determine if 3 regions are visible and discernable Dynamic Range Test Fuji CRm System - CNR Test FCRm Reader Scanner Performance CNR PER OBJECT THICKNESS Image Mode: Auto Time Focal Spot: 3 Density Control: 0 AEC Mode: Auto Time Phantom Thickness Target Filter kVp mAs (Auto) mAs (Manual) CNR CNR (relative to 4cm PMMA) Acceptable level 2 cm Mo Mo 25.0 27.2 28.0 17.197 114.4% >110% 4 cm Mo Mo 25.0 135.0 130.0 15.030 100.0% 100% 6 cm Mo Rh 30.0 282.0 280.0 13.765 91.6% >75% • Establish that IP Reader and optics do not exhibit scan or print jitter • Non-Grid Exposure • Position rulers in “T” formation • Expose and Process Action Limit: Image must have smooth borders free from jagged edges or defects PASS 11 3. CR Reader Scanner Performance Test Date: November 21, 2007 Image Reader ID: Fuji ClearView CSm # 5722213 (Required for Annual Surveys; not required for Mammography Equipment Evaluations of new units. However, FCRm Primary Erasure Parallel to Chest Wall • To assess the erase performance of reader and plates Additive: Shoot phantom, process, wait 1 minute, reprocess, change S value to 10X original, inspect for visibility of phantom image Multiplicative: Shoot & process phantom, shoot 4 cm acrylic on same cassette, process and inspect for visibility of phantom image Acceptance Limits Pass No Pass No NA NA NA NA Workstation Monitor Corrective Action Needed? Laser Film Printer* Corrective Action Needed? Corrective Action Description? Corrective Action Taken? Date of Corrective Action Results of New Test Perpendicular to Chest Wall Acceptance Limits Pass No Pass No NA NA NA NA Workstation Monitor Corrective Action Needed? Laser Film Printer* Corrective Action Needed? Corrective Action Description? Corrective Action Taken? Date of Corrective Action Results of New Test *If hard copy is used for final interpretation. Cassette / IP Number A30389595C Additive Lag test Exposure conditions ACR MAPP LSG BH: F Target Mo Filter Mo kVp 25 S value of the ACR Phantom Image: Acceptable? CORRECTIVE ACTIVE NEEDED? CORRECTIVE ACTIVE DESCRIPT CORRECTIVE ACTIVE TAKEN? CORRECTIVE DATE: RESULTS OF NEW TEST: FCRm Inter-Plate Consistency 520 Artifact Evaluation Pass No NA NA NA NA Multiplicative Lag test Exposure conditions ACR MAPP A uniform 4 cm sheet image Printer: 52 S value applied to non-exposure image (10 Times Phantom Image S Value) Acceptable? CORRECTIVE ACTIVE NEEDED? CORRECTIVE ACTIVE DESCRIPT CORRECTIVE ACTIVE TAKEN? CORRECTIVE DATE: RESULTS OF NEW TEST: mAs 139 Target Mo Mo Filter Mo Mo kVp 25 25 mAs 139 132 Artifact Evaluation Pass No NA NA NA NA • To confirm x-ray absorption & SNR consistency • Exposure plates with 4 cm acrylic using clinical technique • Record mAs (must be within 10%) • Calculate SNR (must be within 15%) The mAs noted on the generator read-out must not change by more than ±15% from the previous test. Percent Change in mAs: Acceptable? 5.2% Pass 12 Inter Plate Consistency Lorad M-III Exposure Unite ID: Exposure conditions AEC-mode Target Filter kVp Auto Time Mo Mo 25 Group: Small Cassettes mAs limit is plus or minus 10 % from mean. SNR limit is plus or minus 15 % from mean. Cassette ID mAs Acceptable SNR Accept able A456, #3 132 Yes 70.946 Yes A234, #2 132 Yes 70.952 Yes A123, #1 132 Yes 70.152 Yes Yes A222, #4 132 Yes 70.166 A111, #5 132 Yes 69.521 Yes A777, #6 133 Yes 69.662 Yes FCRm CNR per Object Thickness • CNR using clinical techique for 2 cm • Repeat for 4 and 6 cm Action Limits: CNR of 2 cm relative to 4 cm must be > 100% CNR of 6 cm relative to 4 cm must be > 75% FCRm Contrast to Noise Ratio Test (CNR) CNR PER OBJECT THICKNESS Image Mode: Focal Spot: Density Control: AEC Mode: Auto Time L 0 Auto Time Phantom Thickness Target Filter kVp mAs (Auto) mAs (Manual) CNR CNR (relative to 4cm PMMA) Acceptable level 2 cm 4 cm 6 cm Mo Mo Mo Mo Mo Rh 25.0 25.0 30.0 26.2 134.0 227.0 26.0 135.0 225.0 13.270 13.165 10.890 100.8% 100.0% 82.7% >100% 100% >75% PASS • To examine consistency of CNR ratio measured over time • Use 4 cm Acrylic and 0.2 mm Aluminum • Manual Technique (Mo/Mo, 26 kVp, 125 mAs) • Calculate CNR using software Action Limit: +/- 20% of baseline 13 FCRm Viewing Conditions • Hardcopy Interpretation: Same conditions as to be used for screen-film mammography images. Film masking devices and hot lights must be available to the radiologist. • Softcopy Interpretations: Follow the monitor manufacturer’s QC procedures. If no monitor manufacturer’s QC procedures available, follow procedures outlined in Fuji QC Manual. Laser Printer must be able to print images on mammography laser film Fuji CRm - Monitor Tests Site: Room #: Date of Survey: 4/19/2007 Date of Installation: Apr-07 Med. Img Ctr of So Cal Reading Room DOME©E5 Serial Num bers Luminance Value Setting (cd/sq-m) Left M onitor Right M onitor 703PNKN00011 703PNKN00028 499.2 500.88 Target Value = 500 cd/sq-m Is the CXtra icon present in the taskbar and shown as a green check m ark? Yes No Manual Conform ance report for both displays attached? Yes No Is DICOM calibration Graph for both displays attached? Yes No FCRm Printer Annual QC • Follow the printer manufacturer’s QC program for test procedures, frequency, performance and corrective action. • If no printer manufacturer’s QC program, follow the QC program outlined in Fuji QC Manual. • Tests must be performed initially prior to interpreting hard copy mammograms and after major repairs and then annually. 14 OD Requirements for HiHi-Resolution Laser Imagers 10. Breast Entrance Exposure, AEC Reproducibility, Average Glandu Dose, and Radiation Output Rate Digital Fuji CR CR 18 x 24 Collimators Imaging mode: Screen type: Film type: Cassette size (cm): Field Restriction: Breast thickness (cm) Phantom Nominal kVp setting Anode material Filter AEC mode AEC density control setting Measured HVL (mm Al) 4.2 RMI 156-15914 25 Mo Mo Auto Filter 0 0.3 SID (cm): Source-detector distance (cm): Source-bucky distance (cm): Dosimeter used: Energy correction factor: 4.2 RMI 156-15914 26 Mo Mo Auto Time 0 0.31 4.2 RMI 156-15914 27 Mo Mo Auto Time 0 0.32 65 60 64 Keithley Triad 1.00 4.2 RMI 156-15914 28 Mo Rh Auto Time 0 0.43 Breast Entrance Exposure and AEC Reproducibility: Dmax > 3.5 OD Exposure #1 Exposure #2 Exposure #3 Exposure #4 Mean values Standard deviations (SD) Coefficient of variation (CV) Action limit: R mAs R mAs R mAs R mAs 1.581 1.498 1.521 1.501 1.525 0.039 147.0 139.0 142.0 140.0 142.0 3.56 1.480 121.0 1.231 89.1 0.705 75.2 1.480 121.0 1.231 89.1 0.705 75.2 0.025 0.025 If coefficient of variation for either R or mAs exceeds 0.05, seek service. Average Glandular Dose: 1.525 1.480 1.231 0.705 Dose conversion factor from Tables 1-3 (mrad/R) 153 159 164 217 Computed average glandular dose (mrad) 233 235 202 153 Inv Sq corrected skin exp Mid-density > 1.5 OD Action limit: If average glandular dose exceeds 300 mrad (3 mGy) for 4.2 cm effective breast thickness, seek service or technique adjustment. Corrective action must be taken before further examinations are performed if the test results fail MQSA regulations. Radiation Output Rate: 3 sec, 4.5 cm above breast support kVp Anode Filter 28 28 Mo Mo Mo Mo SID (cm) Exp (mR) mAs Time (sec) 65 4740 300 3 Kerma Rate (mR/s) (mGy/s) 1580 13.8 Air Kerma (mGy/sec) = Exp Rate (mR/s) x 0.00873 mGy/mR Action limit: If output rate is less than 800 mR/sec (7.0 mGy air kerma/sec), seek service. Phantom Image Quality Check Possible Items (6) (5) Proper Scoring Techniques of the ACR Mammography Accreditation Phantom (5) 15 Fibers Fibers • Full Point Score - Full length of fiber visible (+/- 1 mm) use first fiber as a reference for length - Correct location - Correct orientation • Half Point Score - Not all, but > 50% visible - Correct location - Correct orientation - Stop if orientation or borders not Fibers - Artifacts Fibers - Artifacts Deductions • In Wrong Location of Phantom Insert – Fiber-like – At least as apparent as last fiber scored – Diameter and length > last scored fiber visible • Deduct 1.0 - last fiber is scored as whole and artifact is > 75% of full fiber length • Deduct 0.5 - last fiber is scored as 1/2 and artifact is > 50% of full fiber length • Only deduct from last fiber scored Action Limits: Minimum of 4 fibers for all units except Hologic and Siemens FFDM units. Minimum of 5 fibers for Hologic and Siemens FFDM units 16 Speck Groups Speck Groups - Artifacts Full Point Score - > 4 specks visible - Correct locations • Noise or speck-like • As apparent as last specks scored (i.e. - at least as bright and > diameter) • Do not subtract bright artifacts caused by dust or emulsion pickoff Half Point Score - 2 or 3 specks visible - Correct locations Speck Groups - Deductions Speck Groups • Deduct 1-for-1 from scored specks • Only deduct from the specks(s) in the last group scored (with either a 1.0 or 0.5 score) Action Limit: Minimum score of 3 speck groups for all units except Hologic and Siemens FFDM units Minimum score of 4 speck groups for Hologic and Siemens FFDM units 17 Masses Score Masses Score • Full Point - Density difference visible - Generally circular (> 75% circumference) - Correct location • Half Point - Density difference visible - Not generally circular (< 75% circumference) - Correct location Masses Score - Artifacts Masses Scores In Wrong location of Phantom Insert - Mass - like - As apparent as last mass scored Deduction Only deduct from last mass scored Action Limit: Minimum score of 3 masses for all units except Hologic and Siemens FFDM units Minimum score of 4 masses for Hologic and Siemens FFDM units 18 Acceptable Phantom Image WW = 200 (suggestion) Description: Acceptable ACR Phantom image Minimum Acceptable Phantom Image Score for the Fuji FCRm system is: 4 fibers, 3 speck groups, 3 masses Image Quality Values • The Phantom Image Quality on the Fuji CRm system will reflect very closely what was achieved previously on the film/screen system in use - expect no significant change in quality. MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARY Full-Field Digital – Fuji Site Name Report Date Address Survey Date Medical Physicist's Name Signature X-Ray Unit Manufacturer Model Date of Unit Installation Room ID FFDM Image Receptor Mfr FFDM Model Fuji FFDM QC Manual Version: (check one; must use version applicable to system tested; contact mfr if questions) 1st edition, 07.2006 897N0602, 2006 Accessory Equipment: Manufacturer MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARY (Fuji, continued) FCRm OTHER (write in): Model Location QC Manual Version On-site Off-site Review Workstation* Laser Film Printer* On-site Off-site *FDA recommends that only monitors and printers specifically cleared for FFDM use by FDA’s Office of Device Evaluation (ODE) be used. See FDA's Policy Guidance Help System www.fda.gov/CDRH/MAMMOGRAPHY/robohelp/START.HTM Survey Type: Mammo Eqpt Evaluation of new unit (include MQSA Rqmts for Mammo Eqpt checklist) Annual Survey Medical Physicist's QC Tests 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. Evaluation of Site's Technologist QC Program (Required for Annual Surveys; not required for Mammography Equipment Evaluations of new units. However, medical physicists must review the site's technologist QC program within 45 days and complete this section so that the facility may submit this form along with the entire Mammography Equipment Evaluation report with their phantom and clinical images to the ACR.) PASS/FAIL MQSA Regs PASS/FAIL S Value Confirmation ( < 120 + 20% [96 < corrected S value < 144]) System Resolution (8 lp/mm + 2 lp/mm in both directions) CR Reader Scanner Performance Imaging Plate (IP) Fog Mammography Unit Assembly Evaluation Collimation Assessment Test date if different from above: Chest wall edge of X-ray field extends to edge of IR Deviation between X-ray field and light field <2% of SID X-ray field does not extend beyond any side of the IR by more than 2% of SID Paddle chest wall edge not beyond IR by more than 1% of SID or appear on the image Automatic Exposure Control (AEC) System Performance Assessment AEC density control function meets Fuji performance criteria Reproducibility (CV) for either exposure or mAs is <0.05 Image mode tracking meets Fuji performance criteria CNR per object thickness meets Fuji performance criteria System Artifact Evaluation Phantom Image Quality Evaluation Fibers Specks Masses (at least 4 fibers, 3 speck groups & 3 masses) Phantom IQ (printed images) Click in boxes to use drop-down lists Phantom IQ (softcopy) Other tests meet Fuji performance criteria (mAs, OD & DD for hardcopy, S value for soft copy) Dynamic Range Primary Erasure Inter-Plate Consistency (variation of mAs within + 10%; SNR within + 15%) kVp Accuracy and Reproducibility Test date if different from above: Measured average kVp within +5% of indicated kVp kVp coefficient of variation <0.02 Dose (average glandular dose for average breast is < 3 mGy [300 mrad]) mrad Beam Quality Assessment & HVL Measurement Test date if different from above: HVL >kVp/100 mm Al Radiation Output Test date if different from above: Radiation output rate is >800 mR/sec (7.0 mGy/sec) at 28 kVp with Mo/Mo Viewing and Viewing Conditions Test date if different from above: Mammographic viewbox luminance >3000 cd/m^2 (nit) Room illuminance (viewbox surface & seen by observer) <50 lux Review Workstation (RWS) Tests** (for all RWS, even if located offsite) Overall Results ("Pass" means all tests pass; indicate "Fail" if any test fails) 1. 2. 3. 4. 5. 6. 7. CNR Weekly Check Phantom Image Visual Checklist Repeat Analysis Compression Printer QC* Review Workstation QC-Overall* Weekly Weekly Monthly Quarterly Semi-annually See FDA guidance See FDA guidance Medical Physicist's Recommendations for Quality Improvement **FDA requires that all RWS comply with a QC program that is substantially the same as that recommended by the image receptor manufacturer . If the RWS is FDA-approved, the RWS's QC manual is considered to be "substantially the same" and you may follow it. (Check with the RWS manufacturer for their FDA clearance status and QC manual.) If the RWS is not FDA-approved for FFDM, you must follow the QC manual provided by the image receptor manufacturer. (Check with the image receptor manufacturer for their required tests.) *** YOUR MEDICAL PHYSICIST MUST SUMMARIZE HIS/HER RESULTS ON THIS FORM *** 19 Fuji Digital Mammography QC Technologist’s Tests TEST CNR-Contrast/Noise Phantom Image Printer/Monitor Visual Checklist Repeat Analysis Compression IP Fog FREQUENCY Weekly Weekly Per Manufacturer Monthly Quarterly Semi-annual Semi-annual Fuji Digital Mammography QC Technologist’s Tests Phantom Images Test Confirms image quality using ACR Phantom Tools required - QC Cassette & ACR Phantom Method - AEC Exposure, process IP using the Physics, ACR MAPP Menu Visually inspect the image. Same scoring method as MQSA Correction Period - Before any further examinations are performed Fuji Digital Mammography QC Technologist’s Tests Printer Tests No film in cassette Imaging Plate should be erased after 8 hours of non-use Imaging Plate should be processed or erased after each exposure Printers are monitored according to the manufacturer’s approved QC program Fuji Digital Mammography QC Technologist’s Tests CNR (Contrast/Noise Ratio) Test Evaluates Noise and Contrast using a fixed xray exposure Tools needed - 4 cm acrylic block & 0.2 mm Al Method - Fixed manual technique CNR Measurement must be within +/- 20% Correction Period - Before any further examinations are performed 20 Fuji Digital Mammography QC Technologist’s Tests Tests in Accordance with ACR/MQSA Visual Checklist Repeat Analysis Compression Test Monthly Quarterly Semi-Annual Same Action Limits and Corrective Action as MQSA IP Fog replaces Darkroom Fog test - Semi Annual FULL-FIELD DIGITAL MAMMOGRAPHY QUALITY CONTROL CHECKLIST – FUJI FCRm Daily and Weekly Tests Year Month Date Initials CNR Weekly Check (weekly) Phantom Image (weekly) Printer QC (See FDA guidance) Review Workstation QC (See FDA guidance) Visual Checklist - Technologist QC Year Month JAN FEB MAR APR MAY JUN JUL AUG SEP OCT NOV DEC Visual Checklist (monthly) Repeat Analysis (≤2% change) (quarterly) Compression (25-45 lb) (semi-annually) Review Workstation QC (See FDA guidance) Fuji Digital Mammography QC Technologist’s Tests Imaging Plate Fog Confirm that the storage conditions of the exposure room protects the IP cassettes from scatter Tools required - IP Cassette Method - Tape coin to front of cassette and place in storage location during test exposures Visually inspect image for visibility of coin Correction Period - Before any further examinations are performed 21 Fuji Digital Mammography QC Technologist’s Tests Inter-Plate Consistency - As Needed The Inter-plate Consistency Test replaces the Screen Speed Test Must be performed when new IPs and/or cassettes are introduced Can be performed by the QC technologist under Qualified Medical Physicist oversight Fuji Digital Mammography QC Technologist’s Tests Inter-Plate Consistency - As Needed Confirm that Imaging Plates used in the facility are similar in sensitivity and image quality Tools Required - new and existing IPs and Cassettes, 4 cm acrylic block Method - Expose cassettes using AEC (time) Mode Acceptable results: mAs +/- 10%, SNR +/- 15% Correction Period - Within 30 days of the test Thank You !!! Melissa C. Martin, M.S., FACR Therapy Physics Inc. 879 West 190 St., Ste 419 Gardena, CA 90248 Office Phone: 310-217-4114 Cell Phone: 310-612-8127 e-mail: Melissa@TherapyPhysics.com 22