AbstractID: 10138 Title: Setup Margins Requirements for Stereotactic Lung Radiotherapy Setup Margins Requirements for Stereotactic Lung Radiotherapy Introduction The success of radiation therapy depends on the delivery of the prescription dose to the intended target while minimizing dose to the surrounding normal tissue. Higher doses delivered in fewer fractions such as in stereotactic body radiation therapy (SBRT) demand improved accuracy in target positioning. The uncertainties associated with target positioning are taken into account in the CTV to PTV expansion by the use of margins. A widespread approach is to use a standard setup margin for a patient population defined by their disease site. This work presents an analysis of setup margins for stereotactic lung radiotherapy patients for one clinical practice. Methods and Materials Forty-five patients were treated with stereotactic lung radiotherapy using Cone Beam CT imaging (Varian Medical Systems) OBI for daily target localization in routine clinical practice. At each treatment fraction at least three CBCTs were acquired to assess the patient position. The first CBCT determined any necessary shifts required after laser alignment (representing the amplitude of offsets for patient set up using only laser alignment), the second to verify the position before treatment (corresponding to the residual offsets after correctional shifts) and the third to estimate the intra-fraction motion. The CBCT images were matched with the planning CT using soft tissue contrast. This provided the tumor offset relative to the simulation position for every fraction before and after the correctional shifts. Setup margins were extracted from calculated systematic and random errors [1] and compared to the RTOG 0618 SBRT lung protocol requirements. Results and Discussion The setup error breakdown is shown in table 1. A large reduction in both systematic and random errors can be observed when CBCT-IGRT is used. The margins required to give 95% of the prescribed dose to 90% of the patient population (M 95%Dose 90%Pop) as well as the margins required to give 99% of the prescribed dose to 99% of the patient population are detailed below. Table 1 SETUP ERROR BREAKDOWN Systematic Laser Pre Tx Post Tx Random Laser Pre Tx Post Tx Vertical (mm) 5.1 0.9 1.3 5.9 0.7 1.5 Longitudinal (mm) 5.7 1.4 1.5 8.1 0.6 1.8 Lateral (mm) 7.6 1.1 0.7 8.3 0.6 1.1 MARGIN REQUIREMENTS M 95%Dose 90%Pop Laser Pre Tx Post Tx M 99%Dose 99%Pop Laser Pre Tx Post Tx Vertical (mm) 17.0 2.8 4.3 22.9 3.7 5.8 Longitudinal (mm) 20.0 3.9 5.0 26.9 5.3 6.7 Lateral (mm) 24.8 3.2 2.6 33.4 4.3 3.5 These numbers provide a guideline for an evidence-based choice of setup margins when CBCT is used in stereotactic lung radiotherapy. This demonstrates the accuracy of volumetric imaging for daily tumor positioning. Conclusions The amplitude of the 95%Dose-90%Population margin (3-5 mm depending on direction) is consistent with the RTOG requirements of 5-10 mm. Larger margins might be required if higher dose or population coverages are intended. References [1] M. van Herk et al., Int. J. Radiation Oncology Biol. Phys. 47, 1121 (2000) AbstractID: 10138 Title: Setup Margins Requirements for Stereotactic Lung Radiotherapy 40 30 25 Laser alignment 30 # of fractions # of fractions 35 25 20 15 10 Laser alignment 20 15 10 5 5 0 0 -40 -30 -20 -10 0 10 20 30 -40 -30 -20 -10 40 16 18 14 16 laser alignment 10 8 6 4 20 30 40 Laser alignment 12 10 8 6 4 2 2 0 0 -40 -30 -20 -10 0 10 20 30 40 0 10 20 30 40 50 60 3D magnitude of error (mm) Lateral positioning error (mm) 120 120 Pre Tx CBCT Post Tx CBCT 80 60 40 80 60 40 20 20 0 0 -15 -5 5 Pre Tx CBCT Post Tx CBCT 100 # of fractions 100 # of fractions 10 14 12 # of fractions # of fractions Vertical positioning error (mm) -15 15 -5 5 120 90 80 Pre Tx CBCT Post Tx CBCT 70 # of fractions 100 80 60 40 60 Pre Tx CBCT Post Tx CBCT 50 40 30 20 20 10 0 0 -15 -5 5 Lateral positioning error (mm) 15 Longitudinal positioning error (mm) Vertical positioning error (mm) # of fractions 0 Longitudinal positioning error (mm) 15 0 10 3D magnitude of error (mm)