44th Annual Meeting of the AAPM, July 15th, 2002 CE-IMRT 1: Quality Assurance of IMRT Delivery Systems - Elekta Acknowledgements • • • • • M.B. Sharpe, Ph.D. Department of Radiation Oncology William Beaumont Hospital Royal Oak, MI, 48073-6769 Kevin Brown – Elekta Oncology Systems Geoff Budgell – Christie Hospital Ted Thorson – Elekta Oncology Systems Peter Williams – Christie Hospital John Wong – William Beaumont Hosptial msharpe@beaumont.edu William Beaumont Hospital is customer and collaborator of ADAC Laboratories and Elekta Oncology Systems MBS June 4th, 2002 Outline • • • • • • Introduction Introduction Anatomy of an Elekta Accelerator and MLC MLC Calibration Beam Stability Dosimetry of Small Fields Summary • • • • MBS June 4th, 2002 Step-and-shoot approach with MLC. Non-uniform intensity profiles generated with a series of discrete, uniform beam segments. MLC leaves are stationary when beam is on. Beam is off when leaves are moving. MBS June 4th, 2002 Introduction Elekta Linear Accelerators • • • • • • Considerations: • • • • • Geometric calibration and constraints of MLC. Beam stability for short irradiation times (few MUs). Dosimetry of small fields (i.e., as small as 1x1 cm2). Small fields offset from central axis. Traveling wave accelerator Mounted on drum structure. Magnetron RF supply. Diode electron gun: no grid. Integrated MLC Fewer segments and MLC backup jaws allow less concern for inter-leaf leakage, tongue & groove effects, and curved leaf face. MBS June 4th, 2002 MBS June 4th, 2002 1 Fast Tuning Magnetron (FasTraq) • • • • New magnetron with instantaneous frequency capture Reduces initial start up and inter segment times Standard on new machines Available as upgrade Elekta MLC • • • • • • 40 leaf pairs (10mm pitch) Replaces upper jaws Backup jaws (30mm W) 40 x 40 cm2 field size Total of 32.5 cm leaf travel 12.5 cm travel over central axis Target Primary collimator Filter Ion chamber W edge • Opposing leaf-pairs do not touch • • Optical tracking (real time). Integral auto-wedge Reflector K Leaf Leaf Y back-up diaphragm X Diaphragms Adaptor Ring MBS June 4th, 2002 MBS June 4th, 2002 Optical Leaf Tracking System Elekta MLC End View Side View direction of leaf motion ILLUMINATION SYSTEM Reflected Light ILLUMINATION SYSTEM Incident Light Camera Camera Camera Camera 75/25Beam 75/25Beam Splitter Splitter 75/25Beam 75/25Beam Splitter Splitter Wedge Wedge Mirror Mirror Adjustable Adjustable Mirror Mirror Adjustable Adjustable Mirror Mirror Projector Projector Assembly Assembly MBS June 4th, 2002 Optical Leaf Tracking System Wedge Wedge Mirror Mirror Adjustable Adjustable Mirror Mirror Adjustable Adjustable Mirror Mirror Projector Projector Assembly Assembly Leafbank Leafbank Leafbank Leafbank MBS June 4th, 2002 Minimum Leaf Gap Requirement Desired Field MLC Conformation central axis “Flag pole” (10-12% transmission) Area to be exposed MBS June 4th, 2002 MLC X-Diaphram Y-Diaphram MBS June 4th, 2002 2 Considerations: 0.08 • • 0.06 0.04 Series1 Series2 0.02 Series3 40 37 34 31 28 25 22 19 16 13 7 0.00 10 Deviation (cm) MLC calibration MLC reproducibility • Beam stability for short irradiation times (few MUs). • Small fields (i.e., as small as 1x1 cm2). • Fields offset from central axis. 4 • Repositioning accuracy 1 IMRT QA and Commissioning -0.02 -0.04 Leaf number MBS June 4th, 2002 An Clinically “Accepted” MLC MBS June 4th, 2002 0.5 mm Gap: -4.56% MBS June 4th, 2002 MBS June 4th, 2002 1mm Gap: -9.01% MBS June 4th, 2002 No Gap, No Overlap MBS June 4th, 2002 3 Beam Symmetry and Flatness Beam Stability: Dose Rate • • • With step-and-shoot delivery, there is the potential for short irradiation times (MUs). Dose rate stability influences the treatment precision. Measure dose per MU versus total MU. Check short, and long term stability. • For > 2MU, dose rate is within +/-2% (2s). • Angle Change MBS June 4th, 2002 1.04 1.01 1.000 • 0.998 0.996 0.994 Stability of flatness and symmetry affects dose rate for small fields directed off the central axis. For an open 20x20cm2 field, measure profiles for irradiations ranging from 1 to 100 MU. • 0.992 0.990 100 MU/min 200 MU/min 0.988 1.00 • 1.002 Output factor 1.02 Beam Stability: Flatness, Symmetry Fast Tuning Magnetron SL20 - 18MV Initial 2 weeks later 6 months later, (new monitor chamber) 0.986 400 MU/min • • Sun Nuclear Profiler (46 diodes, 10 profiles/sec). Flatness is +/-3% if more than 5MU delivered. Symmetry +/-3% if more than 4MU delivered. 0.984 1 0.99 1 10 MU 100 10 100 1000 Number of MU MBS June 4th, 2002 MBS June 4th, 2002 12 6 Beam Stability: Flatness, Symmetry Symmetry 0.10 5 Flatness vs time (s) after beam on 3 2 GUN-TARGET DIRECTION CROSS-PLANE DIRECTION 0.05 4 1 0 0.00 0 0.16 0.32 0.48 0.64 0.8 0.96 1.12 1.28 1.44 1.6 1.76 1.92 sec 0 0.16 0.32 0.48 0.64 0.8 0.96 1.12 1.28 1.44 1.6 1.76 1.92 1.8 -0.05 0.10 Flatness Relative Dose Rate Energy Change MBS June 4th, 2002 Beam Stability: Dose Rate 1.03 Position Change 1.6 1.4 Symmetry vs time (s) after beam on 0.05 1.2 1 0.8 0.6 0.00 0.4 0.2 -0.05 1 MBS June 4th, 2002 10 Total MUs Delivered 100 0 sec 4 Relative Output in Water (6MV) Relative Dose Rate vs Field Size • • Collimation of head scatter affects dose rate “in-air”. Dose rate in vivo is further affected by photon scatter and e- transport. Beam weights calculated by inverse planning must be adjusted to account for head scatter. Courtesy of Wolfram Laub head scatter Relative Output Factor • 1.1 1.0 0.9 phantom scatter & e-- transport 0.8 0.7 "In-Air" In-Phantom 0.6 0.5 0 2 4 6 8 10 Side Length of Square [cm] MBS June 4th, 2002 MBS June 4th, 2002 Output Factors for Small Fields Output Factors for Small Fields 1.1 • • • SL20 -18MV 10x10cm 2 Profiles measured in water (10cm depth) using a pinpoint ion chamber (0.015cm3). Calculated with pencil-beam convolution and extrafocal source model. Verified radiation field size and offsets. Calculations & measurement agree to within 1% on average, 2.5% max. 1.0 Relative Dose Rate • 0.9 5x5cm 2 0.8 2x2cm 2 0.7 MEASURED: GUN-TARGET PLANE MEASURED: CROSS PLANE CALCULATED 0.6 2 1x1cm 2 0.5 -7.5 -5.0 -2.5 0.0 2.5 5.0 7.5 Distance Off-Axis [cm] MBS June 4th, 2002 MBS June 4th, 2002 • • Typically measure in water (10cm depth) using a pinpoint ion chamber (0.015cm3) and beam scanning system Scan profiles to verify leaf positioning radiation field size and offsets %Change in Relative Ouput Output Factors for Small Fields 20 6MV 10 0 Nominal Field Size 1cm 2cm 5cm 10cm -10 -20 -3 (a) MBS June 4th, 2002 IMRT QA with EPID (iViewGT) -2 -1 0 1 2 3 • • • • • • Flat-panel imager 41cm2 a-Si panel 26cm2 field at isocentre 15cm offset in all quadrants 16 bit gray scale resolution 3 frames per sec readout • Primary function is to verify patient position Deviation from Field Size [mm] MBS June 4th, 2002 5 iViewGT: IMRT QA Possibilities • • • • Small MU verification Dosimetry QA QA beam flatness and symmetry MLC calibration QA leaf position accuracy … 40 mu ___ 20x2 MU integrated MBS June 4th, 2002 MBS June 4th, 2002 Small MU verification Courtesy of Geoff Budgel, Ph.D. MLC Calibration • • • Extremely time-consuming using film, especially if adjusting gain & offset. Possibly requires tighter tolerances / more frequent checking for IMRT iViewGT has 0.25 mm resolution at isocentre – ought to be sufficient for MLC calibration 20x2 mu MBS June 4th, 2002 Courtesy of Geoff Budgel, Ph.D. MLC gain and offset calibration • • IMRT requires 10 cm over-travel measurements for both leaf banks and jaws Requires accurate definition of central axis of rotation MBS June 4th, 2002 Matched step & shoot fields Film (40x32cm) iViewGT (24x24cm) Central 24 leaves -12 cm -10 cm MBS June 4th, 2002 12 cm 10 cm Courtesy of Geoff Budgel, Ph.D. Courtesy of Geoff Budgel, Ph.D. 6 Spot the errors Subtraction images 0.5 mm 0.5 mm 1 mm 0.3 mm 1 mm 0.3 mm 0.7 mm 0.3 mm 0.7 mm 0.3 mm 0.5 mm 0.5 mm 0.2 mm 0.2 mm 1 mm 0.7 mm Courtesy of Geoff Budgel, Ph.D. MBS June 4th, 2002 Effect of set-up on error detection MBS June 4th, 2002 No positional change Re-set to same position • • Head angle 0.7 deg Courtesy of Geoff Budgel, Ph.D. Conclusions • No errors 1 mm 0.7 mm Low MU images and template matching enable easy patient position verification Require access to raw images for dosimetry applications Looks promising for QC applications and MLC calibration With errors No positional change MBS June 4th, 2002 Re-set to same position Head angle 0.7 deg MBS June 4th, 2002 Courtesy of Geoff Budgel, Ph.D. Summary • Commissioning: • • • Quality Assurance: • • Range of clinical operation to 1x1 cm2 fields, and 1-2 MU irradiations Verify beam stability for short exposures Enhance monthly QA to test MLC operation, accuracy and precision. A-Si EPID: Intergated QA device. MBS June 4th, 2002 7