ACCEPTANCE TESTING GOALS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY FLUOROSCOPIC EQUIPMENT HAS CHANGED • • • • FLAT PANEL IMAGE RECEPTORS MORE X-RAY BEAM FILTRATION SELECTABLE PULSED FLUOROSCOPY MULTIPLE “ABC” PROGRAMS • • • • PRIOR TO 1st CLINICAL USAGE VERIFY PURCHASE ORDER CHECK ARCHITECTURAL DESIGN & FLOW MOST COMPREHENSIVE TESTING! – – – – – X-RAY GENERATOR & TUBE IMAGE RECEPTOR & IMAGE QUALITY RADIATION MEASUREMENTS MECHANICAL SYSTEMS SOFTWARE • MAKE REPAIRS & ADJUSTMENTS • UNDERSTAND ABC OPERATION • ESTABLISH BASELINE VALUES FOR QC I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS – ANATOMICAL or STUDY SELECTIONS – DOSE BASED – PEDIATRIC vs ADULT • DIFFERENT VIDEO MONITORS • MORE SOFTWARE IMAGE PROCESSING • 3-D ROTATING FLUOROSCOPY 1 X-RAY BEAM QUALITY (HVL) I. PHYSICS TESTS • • • • • • • • RADIATION OUTPUT HALF VALUE LAYER (HVL) kVp ACCURACY mA or mAs REPRODUCABILITY & LINEARITY PULSE WIDTH & RATE ACCURACY FOCAL SPOT SIZE MEASUREMENT WAVEFORM DISPLAY ABC PERFORMANCE EFFECTS OF INCREASE X-RAY FILTRATION HVL ISSUES • OFTEN CAN NOT USE METHOD OF KEEPING TOTAL FILTRATION CONSTANT WHILE MOVING FILTERS FROM BEHIND DETECTOR TO IN FRONT OF DETECTOR 80 kVp X-RAY SPECTRA WITH DIFFERENT COPPER FILTRATION 7.00E+04 6.00E+04 #X-RAYS / keV • kVp & FILTER MAY BE AUTOMATIC • FILTER RANGE: 0.0 – 0.9 mm Cu • HVL DEPENDS UPON kVp, CONTOUR & XRAY FILTER • FILTER MAY BE DIFFERENT IN FLUORO & RECORD • TYP. HVL= 3.5 - 9.0 mm @ 80 kVp & CAN BE HIGHER 5.00E+04 Al FILTERS 4.00E+04 3.00E+04 DETECTOR 2.00E+04 1.00E+04 Al FILTERS 0.00E+00 0 10 20 30 40 50 60 70 X-RAY ENERGY (keV) 0 mm Cu 0.3 mm Cu 0.6 mm Cu 80 X-RAYS 0.9 mm Cu 2 COMPARISON OF HVL vs. ADDED FILTRATION HVL MINIMUM REGULATORY MEASURED HVL ( mm Al EQ. ) 9 HVLmin > 3.5 (kVp/100) + 0.08 in mm aluminum 8 7 6 IEC Standards (new regulations ??) for interventional labs. 5 4 3 @ 80 kVp HVL MAXIMUM EMPIRICAL 2 1 HVL max < 4.0 to 9.0 mm Al @ 80 kVp 0 0 0.2 0.4 0.6 0.8 1 ADDED FILTRATION ( mm COPPER ) SIEMENS GE COMMENTS ABOUT FOCAL SPOTS MEASURMENTS FILTER & kVp ACCURACY • DIFFERENT METHODS: INVASIVE, NONINVASIVE, TEST POINT • AUTOMATIC CONTROLS ARE METER DEPENDENT • FILTER & WAVE DEPENDENT • +/- 3 to 5% + 1 kVp correction factors for filtration kVp meter reflection of meter patient table x-ray tube • COMBINED SPATIAL RESOLUTION MAY BE BETTER TEST • FILM MAY NOT BE AVAILABLE • GEOMETRY & # DEG. IMPORTANT • INTENSITY DISTRIBUTION? STAR PATTERN IMAGE OF X-RAY TUBE FOCAL SPOT D =blur diameter 3 II. PHYSICS TESTS II. IMAGE QUALITY ASSESSMENTS • SPATIAL RESOLUTION – LINE PAIRS OR MESH? – FOV & GEOMETRY – SOFTWARE • LOW CONTRAST VISIBILITY • CONTRAST RATIO & VEILING GLARE • UNIFORMITY & DISTORTION • FOV ACCURACY • DISPLAY MONITOR • TEMPORAL RESOLUTION • 3-D ROTATIONAL FLUORO IMAGING SPATIAL RESOLUTION WITH BAR PATTERN SUBJECT CONTRAST OF BAR PATTERN WITH DIFFERENT FILTRATION AT 80 kVp 27 CM FoV 14 CM FoV CALCULATED % SUBJECT CONTRAST OF BARS 70 • FLUORO & RECORD MODES DIFFERENCES • FoV DEPENDENCE • FILTRATION DEPENDENCE • BAR PATTERN or MESH • TV or DEL DEPENDENT • GEOMETRICAL MAGNIFICATION & FOCAL SPOT DEPENDENT 60 50 40 30 20 10 0 0 0.2 0.4 0.6 0.8 1 THICKNESS OF COPPER X-RAY FILTER (mm) 0.05 mm Pb 0.10 mm Pb 4 FLAT PANEL MATRIX BAR PATTERN (1/2) PIXEL SHIFT BAR PATTERN ALIGNED WITH PIXELS BARS AT 45 DEGREES TO PIXELS 5 SUMMARY OF BAR PATTERN PLACEMENT • PARTIAL PIXEL SHIFT EITHER VERTICAL OR HORIZONTAL CAN ALTER MEASURED RESOLUTION – RESOLUTION BEST = 1 / [ 2 x PIXEL SIZE] – OR, ZERO FOR 50% PIXEL SHIFT – OR, ANYWHERE IN BETWEEN • 45 DEGREE ALIGNMENT OF BAR PATTERN INCREASES SPACING BY 1.41 x’s GE INNOVA SOFTWARE EFFECTS MINIMUM SOFTWARE SETTINGS STANDARD SOFTWARE SETTINGS – INCREASES RESOLUTION FOR BOTH FLAT PANEL & I.I. + TV MAXIMUM SOFTWARE SETTINGS COMPARISON RESOLUTION: MESH# vs LP/mm MESH PATTERN FOR RESOLUTION 80 MESH# VISIBLE 70 MESH # / 25.4 mm 60 50 40 30 y = 18.602x + 5.9194 2 R = 0.8273 20 10 0 MESH # / INCH 0 1 2 #LP/mm VISIBLE 3 4 6 5 % NOISE SPATIAL RESOLUTION AT INPUT SURFACE OF IMAGE RECEPTOR NO NOISE MAGE NOISE I 0 0.5 1 1.5 2 2.5 3 SPATIAL FREQUENCY (LP/mm) COMPARISON OF INHERENT DETECTOR SPATIAL RESOLUTION SPATIAL RESOLUTION (LP/mm) M TF VALUE MTF GRAPH & SPATIAL RESOLUTION LIMIT 1 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0 SPATIAL RESOLUTION AT DISTANCES AWAY FOR INPUT SURFACE OF IMAGE RECEPTOR 3.8 3.6 3.4 3.2 3 2.8 2.6 2.4 2.2 2 10 12 14 16 18 20 22 24 FIELD-of-VIEW (LP/mm) Axiom Artis I.I. Axiom Artis Flat Panel Innova Flat Panel 7 SYSTEM SPATIAL RESOLUTION SPATIAL RESOLUTION (LP/mm) • DEPENDS UPON GEOMETRICAL MAG ( M ), FOCAL SPOT ( f ) & INHERENT IMAGE RECEPTOR RESOLUTION ( νir ) • LESS THAN THE LOWEST OF: – Mx νir – M / [ (M - 1) x f ] • AT LOW “M”, νir DOMINATES 3.8 3.6 3.4 3.2 3 2.8 2.6 2.4 2.2 2 0 • AT HIGH “M”, FOCAL SPOT BLUR DOMINATES 5 10 15 20 25 30 35 40 45 DISTANCE FROM ENTRANCE SURFACE (cm) Axiom Artis I.I. Calc Axiom Artis Flat Panel Calc Innova Flat Panel Calc EFFECTS OF DISTANCE ON SPATIAL RESOLUTION COMPARISON OF SPATIAL RESOLUTION FOR SMALL Fov (I.I. = 13 cm, SFPD = 16 cm & GEFPD = 13 cm) 4 SPATIAL RESOLUTION (LP/mm) COMPARISON FLUORO SPATIAL RESOLUTION IN LARGE Fov (I.I.= 23 cm, SFPD=25 cm & GEFPD=20cm) 4 3.8 • DISPLACING THE PATTERN AWAY FROM ENTRANCE SURFACE IMPROVES RESOLUTION UP TO 20 – 25 cm 3.6 3.4 3.2 – SID=100 cm USED MAG FACTOR ~ 1.30 – FLUORO FOCAL SPOTS 3 2.8 2.6 2.4 2.2 2 0 5 Axiom Artis I.I. Calc I 10 15 20 25 30 DISTANCE FORM ENTRANCE (cm) Axiom Artis Flat Panel Calc S 35 40 Innova Flat Panel Calc G 45 • LONGER DISTANCE RESULT IN FOCAL SPOT BLUR DEGRADING RESOLUTION • FLAT PANEL SYSTEMS BETTER AT LARGE FoV’S & BOTH SYSTEMS TESTED SIMILAR • IMAGE INTENSIFIER SYSTEM BETTER AT ALL SMALL FoV’S 8 LOW CONTRAST DISCRIMINATION PENETRAMETER LOW CONTRAST (TOO EASY FOR ANGIO & CARDIAC) POLAR BEAR CHASING A WHITE SEAL IN A SNOW STORM C-D PHANTOM DESIGN DEPTH HOLES (INCH) DIAMETER HOLES (INCH) 9 ACR ACR FLUOROSCOPY FLUOROSCOPY PHANTOM PHANTOM ACR FLUOROSCOPY PHANTOM MESH LOW CONTRAST CONTRASTDETAIL 19.3 cm ACRYLIC + 4.6 mm ALUMINUM LOW CONTRAST DEPENDS UPON: • • • • kVp USED X-RAY BEAM FILTRATION THICKNESS OF ATTENUATOR TYPE OF LOW CONTRAST PHANTOM IS USED • SOFTWARE: FRAME AVERAGING, EDGE ENHANCEMENT, ETC • DOSE (ABC) PROGRAM • FoV USED CONTRAST RATIO (RCC) • MEASURES DEGRADATION DUE TO LIGHT & PHOTON SCATTER • CAN DEGRADE VISIBILITY OF SMALL VESSELS • USE LIGHT ON MONITOR • RC > 60 : 1 10 EXCESSIVE SOFTWARE EDGE ENHANCEMENT LOWERS CONTRAST RATIO TEMPORAL RESOLUTION • 6 WIRES 0.13 - 0.56 mm DIAMETER AT 30 RPM • FLUORO & CINE`/ DSA • STATIONARY vs. ROTATING • WATER PHANTOM SIMULATES PATIENT ATTENUATION • 2 - 3 SPOKES VISIBLE MOVING ROTATING SPOKE PATTERN UNDER PULSED FLUOROSCOPY SMPTE PATTERN FOR VIDEO DISPLAY MONITORS bar pattern for resolution max white gray scale blocks max black low contrast visualiztion 11 III. RADIATION DOSES III. PHYSICS TESTS • TYPICAL PATIENT ENTRANCE RADIATION LEVELS I.I. – PEDIATRIC, SMALL ADULT, AVERAGE ADULT & LARGE ADULT – FLUOROSCOPY & RECORD MODES – DIFFERENT FoV’s & ABC PROGRAMS FIXED 30 cm ATTENUATION MATERIAL T RADIATION DETECTOR TABLE SID = 100 cm X-RAY TUBE NEW YORK STATE NEW FLUORO REGULATIONS FOR PATIENT ENTRANCE EXPOSURE • • • • • MAXIMUM FLUORO ENTRANCE DOSE RATE IMAGE RECEPTOR ENTRANCE DOSE RATES DAP METER CALIBRATION CHECK SCATTERED RADIATION LEVELS RADIATION PROTECTION SURVEY NEW YORK STATE FLUORO ENTRANCE EXPOSURE RATES • PEDIATRIC = 0.75 inches of Al • SMALL ADULT = 1.50 inches of Al • AVERAGE ADULT = 1.50 inches of Al plus 0.50 mm Cu • LARGE ADULT = 1.50 inches of Al plus 2.10 mm Cu • USE AT TYPICAL CLINICAL SETTINGS IN ALL FoV’s USED CLINICALLY 12 COMPARISON OF ACRYLIC WITH ALUMINUM AND COPPER FOR FLUORO OTHER PHYSICS CHECKS FLUORO ENTRANCE EXPOSURE RATE (R/min) 12 10 GE INNOVA – all FoV’s 8 6 4 2 0 0 5 10 15 20 25 30 35 THICKNESS OF ACRYLIC (cm) acrylic 1.5" Al + 0.5 mm Cu 0.75" Al 1.5" Al +2.1 mm Cu 1.5" Al Expon. (acrylic) CHECK LIST ITEMS • ELECTRICAL LEAKAGE & GROUNDING TESTS • “EMERGENCY” ELECTRICAL POWER • ELECTRICAL RECEPTICAL POLARITY • DATA ACHIVEAL AND RETRIEVAL • 5 MINUTE FLUORO TIMER • ROOM LIGHTING • ROOM SHIELDING • REGULATORY COMPLIANCE 13 14