Advanced Reconstruction Methods on Philips CT Systems Sandra Simon Halliburton, PhD

advertisement
Advanced Reconstruction
Methods on Philips CT Systems
Sandra Simon Halliburton, PhD
Director of Clinical Architecture
CT R&D
Optimization through system & statistic models
Philips Solution for Advance Reconstruction
4
iDose
PARADIGM
SHIFT
FBP
128 mAs
20 mAs
iDose
IMR
What is IMR?
๏‚ง Iterative Model Reconstruction
๏‚ง Formulates image reconstruction as
optimization of cost function, F ๐‘ฅ ,
where ๐‘ฅ is the image that minimizes F
๏‚ง Function is solved iteratively because
no explicit solution exists
function minimum
Cost Function [F(๐’™)]
System Model
Statistics Model
Detailed description of system geometry.
Representation of ideal measurement of object
based on Poisson model of x-ray transport.
Cost Function
F ๐‘ฅ = ๐ท ๐‘ฅ + ๐›ฝ โˆ™ ๐‘…(๐‘ฅ)
Data fit:
Difference between
estimated and acquired data
Regularization:
Noise penalty
Minimized when
๏ถestimated data most closely matches actual data
๏ถnoise is low
Optimization is balance between data fit and noise
Cost Function
F ๐‘ฅ = ๐ท ๐‘ฅ + ๐›ฝ โˆ™ ๐‘…(๐‘ฅ)
๏‚ง Defined by the desired image and the starting image
๏‚ง Desired image is controlled by xxxxxxxxxxxxxxxxxxxxxxx
taking into account xxxxxxxxxxxxxxxxxxxxxxxxxxxxx and
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
๏‚ง Starting image is xxxxxxxxxxxxxxxxxxxxxxxxxxx after
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx while preserving
xxxxxxxxxxxxxxxxxxxxxx
Noise Constraint
F ๐‘ฅ = ๐ท ๐‘ฅ + ๐œท โˆ™ ๐‘…(๐‘ฅ)
๐ท ๐‘ฅ
๐ท ๐‘ฅ
๐ท ๐‘ฅ
๐œท๐Ÿ โˆ™ ๐‘…(๐‘ฅ)
= too noisy
๐œท๐Ÿ โˆ™ ๐‘…(๐‘ฅ)
๐œท๐Ÿ‘ โˆ™ ๐‘…(๐‘ฅ)
= too smooth
= “just right”
Noise Modeling
Attenuation
Image
True Noise
(Monte Carlo)
Noise
Map
Noise Modeling
Attenuation
Image
Noise
Map
Static vs. Moving Object
๏‚ง Model-based IR attempts to create a final (optimized) image which is
closest match to projections through image voxels
y
x
t1
STATIC OBJECT
t2
1
0
1
0
2
0
1
0
1
Image
Projections
๏ƒผ Image matches projections
๏ƒผ Optimization achieved
๏ƒผ High-resolution, low-noise image
Static vs. Moving Object
๏‚ง Model-based IR attempts to create a final (optimized) image which is
closest match to projections through image voxels
y
x
t1
MOVING OBJECT
t2
Projections
1
0
0
0
1
0
0
0
1
Image
Projections
๏ƒ Image DOES NOT match projections
๏ƒ Optimization NOT achieved
๏ƒ Image “forced” to match, artifacts
Motion Sensitivity
๏‚ง Manifestation of motion on model/knowledge based IR is unpredictable
๏‚ง May compromise image quality since optimization “forces” data match
IMR w/o
motion compensation
Cardiac CT
Standard
Yuki et al., JCCT, 2014
iDose4
IMR Reconstruction Times
Number of
Images =
1201
Reconstruction time (minutes)
3.0
2.5
2.0
1.5
1.0
0.5
0.0
Brain
Brain CTA Carotid CTA Aorta CTA
Chest
Coronary CTA Abdomen
16 cm
15 cm
25 cm
70 cm
35 cm
14 cm
40 cm
0.9 x 0.45 mm…
0.9 x 0.45 mm…
0.9 x 0.45 mm…
0.9 x 0.45 mm…
0.9 x 0.45 mm…
0.9 x 0.45 mm…
0.9 x 0.45 mm…
Measured on iCT Elite
Benefits of IMR
↓ Noise
↓ Dose
↑ Low Contrast Detectability
↓ Noise
1 mm slice thickness
10 mGy CTDIvol
80
Noise (standard deviation in HU)*
Standard Recon
70
iDose4
iDose4 Level6
Standard
Recon
60
IMR Level 3
iDose4
iDose4: Level 6
50
40
30
20
Standard reconstruction
10
SD: 15.4 HU
0
0
200
400
600
Tube Current-Time Product
(mAs)
* Image noise as defined by IEC standard 61223-3-5. Assessed
using Reference Body Protocol on a CATPHAN phantom.
SD: 8.7 HU
SD: 1.9 HU
↓ Noise + ↓ Dose + ↑ LCD
Task-Based Image Quality Assessment
IMAGE
GENERATION
TASK
OBSERVER
FIGURE OF
MERIT
FDA CLEARANCE FOR CLAIMS OF DOSE
REDUCTION
IMAGE
GENERATION
• MITA Test
Phantom
TASK
OBSERVER
3mm
+14HU
FIGURE OF
MERIT
IMAGE
GENERATION
TASK
• MITA Test
Phantom
๏‚ง Scanned 200x
(100x @ each dose)
FIGURE OF
MERIT
OBSERVER
• 4AFC
• Human
• N=36
Mode
Helical
Gantry Rotation Time
750 ms
Beam Collimation
64x 0.625 mm
Pitch
0.6
Tube Potential
๏‚ง Reconstructed 200x
120 kV
Tube Current-Time Product
153 mAs
31 mAs
CTDIvol
10 mGy
2 mGy
FBP
IMR
Reconstruction algorithm
Slice Thickness
0.8
Slice Increment
0.4
IMAGE
GENERATION
• MITA Test
Phantom
TASK
• 4AFC
๏‚ง Created 200 sets of 4 images
(100 @ each dose level)
– Each image with same dose and
recon algorithm
– 1 image ๏‚ฎ 3 mm pin
3 images ๏‚ฎ uniformity section
– Random position of image
containing pin
๏‚ง Task = choose image w/ pin
OBSERVER
• Human
• N=36
FIGURE OF
MERIT
IMAGE
GENERATION
• MITA Test
Phantom
TASK
• 4AFC
OBSERVER
FIGURE OF
MERIT
• Human
• N=36
๏‚ง 36 human observers
๏‚ง Each observer executed 100 trials per dose level
TOTAL TRIALS PER DOSE LEVEL = 3,600
IMAGE
GENERATION
TASK
• MITA Test
Phantom
FIGURE OF
MERIT
OBSERVER
• 4AFC
• Human
• N=36
Receiver Operating Characteristic (ROC)
True
Positive
Rate
False
Positive
Rate
B
C
D
(Sensitivity) (TFP)
Frequency
A
C
B
A
(1- Specificity) (FPF)
Test Output
IMAGE
GENERATION
TASK
• MITA Test
Phantom
FIGURE OF
MERIT
OBSERVER
• 4AFC
• Human
• N=36
• Detectability
Index, d’
Detectability Index (d’)
๐‘ก๐‘‡๐‘ƒ − ๐‘ก๐น๐‘ƒ
๐‘‘ =
๐œŽ
′
๐‘ก๐น๐‘ƒ
Mean of measurement
of signal present
Mean of measurement
of signal absent
Probability Density
๐‘ก๐‘‡๐‘ƒ
1
d'
0.9
True Positive Rate
False Positive Rate
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
๐œŽ
Standard deviation, assumed
same for both distributions
Burgess, Med Phys, 1995
0
-6
-4
-2
0
2
t
4
6
8
10
IMAGE
GENERATION
TASK
• MITA Test
Phantom
FIGURE OF
MERIT
OBSERVER
• 4AFC
• Human
1
• N=36
0.9
ROC
ROC
ROC
0.8
0.55
0.82
0.96
0.7
0.9
0.6
0.8
0
1
2
3
0.50
0.76
0.92
0.98
TPR
AUC
TPR
0.25
d’
(Sensitivity) (TFP)
Percent
Correct
1
0.5
0.7
0.4
0.6
0.3
0.5
0.2
0.4
0.1
• Detectability
Index, d’
3
d'=0
d'=1
d'=2
d'=3
2
1
d'=0
d'=1
d'=2
d'=3
0
0.3
0
0.2
0.4
0.6
0.8
0.8
1
FPR
0.2
0.1
0
0.2
0.4
0.6
FPR
(1- Specificity) (FPF)
d’ = 0 ๏‚ฎ Unable to distinguish images with low contrast object
present from image with low contrast object absent
d’ = ∞ ๏‚ฎ Always able to distinguish image with low contrast object
present from image with low contrast object absent
1
↓ Noise + ↓ Dose + ↑ LCD
IMAGE
GENERATION
• MITA Test
Phantom
TASK
FIGURE OF
MERIT
OBSERVER
• 4AFC
• Human
• N=36
• Detectability
Index, d’
Median % Median d’
Correct
FBP
50%
0.821
10 mGy
IMR
2 mGy
70%
1.475
80%
CTDIvol
80%
Benefits of IMR
↓ Noise 70-83%
↓ Dose 60-80%
↑ Low Contrast Detectability 43-80%
In clinical practice, use of IMR may reduce CT patient dose depending on the
clinical task, patient size, anatomical location, and clinical practice. A consultation
with a radiologist and physicist should be made to determine appropriate dose to
obtain diagnostic image quality for particular clinical task.
Image Quality Improvement with IMR
IMR provided
๏‚ง Lowest noise
๏‚ง Best low-contrast
detectability,
including at the
lowest dose
๏‚ง Improved resolution
and lowered noise
simultaneously
L¨ove A, et al. Six iterative reconstruction algorithms
in brain CT: a phantom study on image quality at
different radiation dose levels. Br J Radiol 2013;
86:20130388.
Low-kVp Made Routine with IMR
Standard
iDose4
Standard
Oda S. et al, Iterative model reconstruction:
Improved image quality of low-tube-voltage
prospective ECG-gated coronary CT angiography
images at 256-slice CT, Eur J Radiol (2014)
p-value
iDose4
Sub-mSv Imaging for Nodule Assessment with IMR
Conclusions: Sub-mSv IMR improves delineation of lesion margins
compared to standard-dose FBP and sub-mSv iDose4.
Khawaja R. et al, CT of Chest at <1 mSv:
An Ongoing Prospective Clinical Trial of
Chest CT at Sub-mSv Doses with IMR and
iDose4 , JCAT 2014;38: 613–619
Improved Detection of PE with IMR
Standard
Kligerman. et al, Detection of PE on CT:
Improvement using Model-Based Iterative
Reconstruction compared with FBP and
Iterative Reconstruction. J Thorac Imaging
2015;30:60–68
Kligerman. et al, Detection of PE on CT:
Improvement using Model-Based Iterative
Reconstruction compared with FBP and
Iterative Reconstruction. J Thorac Imaging
2015;30:60–68
iDose4
IMR Highlights
Proven benefits
Rigorous human observer studies evidence
simultaneously lowering noise, dose, and improving low
contrast detectability
Fast reconstruction times
Majority of reference protocols reconstructed in < 3 min
Model-based solution for cardiac
IMR is currently only model-based iterative algorithm
available for cardiovascular image reconstrcution
Installs
250 sites with IMR by end of 2015
Scientific papers
30 peer-reviewed publications
3T MR
Download