Bruce J. Gerbi, Ph.D. TGTG-70: Clinical electron beam dosimetry: supplement to TGTG-25 Bruce J. Gerbi, Ph.D. University of Minnesota, Minneapolis, MN John Antolak, Antolak, Ph.D. David Followill, Followill, Ph.D. Michael Herman, Herman, Ph.D. Dimitris Mihailidis, Mihailidis, Ph.D. Ellen Yorke, Yorke, Ph.D. Ph.D. F. Christopher Deibel, Deibel, Ph.D. Patrick D. Higgins, Ph.D. M. Saiful Huq, Huq, Ph.D. David. W. O. Rogers, Ph.D. TGTG-70 Task Group of the Radiation Therapy Committee of the AAPM Formally charged in April, 2001 Inception date: July 22, 2001 Sunset date: December, 2005 Consultants: Consultants: Faiz Khan, Ph.D. Kenneth Hogstrom, Ph.D. TGTG-70: Goal of the Task Group Maintain the original intent of TGTG-25 – provide a useful set of procedures and processes for the practicing clinical physicist for the use of clinical electron beams in the energy range from 55-25 MeV – Not simply a rere-write of TGTG-25 • TGTG-25 was very well written and extensive • Much of the information is still very useful TG-51 Electron Beam Calibration TGTG-70 Goals (continued) To define clearly the tasks that a physicist needs to perform with regards to highhighenergy electrons To supplement the material of TGTG-25 To cover topics that are new since TGTG-25 or that were not fully developed in that report 1 Bruce J. Gerbi, Ph.D. TGTG-70: Table of Contents I. INTRODUCTION II. NOTATION AND DEFINITIONS III. DOSE MEASUREMENTS A. Calibration protocol, TGTG-51 B. Electron beam quality specification C. Dosimetry equipment 1. Ionization chambers 2. Phantoms D. Measurement of central axis percentage depth dose in water 1. Measurements using cylindrical ionization chambers 2. Measurements using planeplane-parallel ionization chambers in water 3. Measurements using diodes in water 4. Water phantom considerations TGTG-70: Table of Contents (2) E. Output factors for electron beams F. Dose determination in small, irregular electron fields G. Non-water phantoms: Conversion of relative dose measurements from non-water phantoms to water 1. Measurements using cylindrical ionization chambers in nonwater phantoms 2. Measurements using plane-parallel ionization chambers in non-water phantoms 3. Film dosimetry 4. Measurement of central axis percentage depth dose using non-water phantoms IV. ELECTRON BEAM ALGORITHMS V. ICRU 71 – Prescribing, Recording, and Reporting electron beam therapy TGTG-70: Table of Contents (3) VI. CLINICAL APPLICATIONS OF ELECTRON BEAMS A. Heterogeneities in electron treatments B. The use of bolus in electron beam treatments C. Electron field abutment VII. LIBRARY OF CLINICAL EXAMPLES A. Intact Breast- Tangent plus Electrons (IMC), Electron Boost B. Chest Wall - Tangent plus Electrons, Electrons only, Conformal Bolus C. Electron Arc D. Total Scalp E. Parotid F. Nose G. Eye – Eyelid, Retinoblastoma, Orbit H. Posterior Cervical Nodes I. Craniospinal J. Intraoperative K. Total Skin Electron Treatment (TSET) L. Total Limb III. A. Calibration Protocol, TGTG-51 TGTG-51 accomplished its two main objectives: 1. Incorporate the new absorbed dose to water standard – Absorbed dose is more robust than the Air Kerma Standard – Dose to water is closer to the dose to tissue 2. Simplify the calibration formalism (as much as possible) Defines dose at one point, the reference point VIII. REFERENCES TG-51 Electron Beam Calibration 2 Bruce J. Gerbi, Ph.D. TGTG-51 TGTG-51: Calibration equation for electrons 60 III.B. Electron beam quality specification Electron beam quality specified by R50 , the 50% depth of dose maximum instead of E0 R50 can be obtained: – from the depth ionization curve in terms of I50 • Correct the raw depth ionization curve for depth by offoff-setting the depth by 0.5 rcav toward surface for cylindrical chamber, no offset for p-p chamber. • Then use the following equations: – R50 = 1.029 I50 – 0.06 (cm) 2 I50 10 cm I50 > 10 cm – R50 = 1.059 I50 – 0.37 (cm) • The measurement must be made at 100 cm SSD. DwQ = MPgrQ k R' 50 kecal N D ,Co w [Gy ] 60 Co Uses the new absorbed dose standard, N D ,w Full calibration to be done in water only Reference depth, dref = 0.6R 0.6R50 - 0.1 cm rather than dmax Protocol uses realistic electron beam data for stopping powers of water to air Uses new factors, k’R50 and kecal DwQ = dose to water at beam quality, Q M = corrected meter reading PgrQ = gradient correction k R' 50 = energy/chamber dependent factor k ecal = factor tha t maps dose at 60Co energy into dose at reference electron energy 60 N D ,Co Co energy w = absorbed dose to water chamber calibration factor at 60 III.C. Equipment, TGTG-70 Ionization chambers – calibration and relative measurements - Both cylindrical and planeplane-parallel chambers are acceptable Diodes – for measurement of %dd %dd data (checked v. ion chambers for accuracy) Phantom material – – – directly from % Depth Dose curve using diode system (checked for agreement with ion chamber data) TG-51 Electron Beam Calibration Water is preferred whenever possible NonNon-water materials are allowed (but not for absolute calibration, as per TGTG-51) 3 Bruce J. Gerbi, Ph.D. PlanePlane-Parallel Chambers PlanePlane-parallel chambers are recommended to calibrate electrons of energy R50 ≤ 2.6 cm Not many waterproof pp-p chambers – Markus, NACP, Memorial, Exradin – p-p chambers cannot be waterproofed easily Interface problems between pp-p chamber and surrounding water medium have not been addressed Backscatter from p-p chamber material Cylindrical Ion Chambers Commonly available Used routinely: calibration & automated scanning systems Gradient correction required Fluence correction required Corrected Meter Reading, M Make measurement with chamber at dref to obtain Mraw – Correct Mraw for Pion, PTP , Pelec , Ppol to get M M = Pion PTP Pelec Ppol M raw – So, of course, you need these correction factors TG-51 Electron Beam Calibration 4 Bruce J. Gerbi, Ph.D. Gradient Correction, Pgr Pgr Electron Beam Dosimetry, k´R50 Based on TGTG-21 formalism (remember Prepl = PgrPfl) – depends on the user’ user’s beam – must be measured for each beam k 'R50 = M raw (d ref + 0.5rcav ) k ' R50 = PgrQ = M raw (d ref ) = 1.00 ) ) L water ρ air L water ρ air (cylindrical) cylindrical) • Prepl • Pcel |evaluated at arbitrary electron quality R50 • Prepl • Pcel |evaluated at electron quality R50 =7.5 cm The energy/chamber dependent factor which relates dose at an arbitrary electron energy, expressed as R50, to the reference energy, R50 = 7.5 cm. -Uses L/ρ for an electron spectrum representative of realistic electron beams (plane(plane-parallel) -The values of k’R50 are appropriate only at dref Determination of k´R50 Electron Beam Dosimetry, kecal From available figures, or using analytical fits (to within 0.2%) Farmer cylindrical chambers, 2 ≤ R50 ≤ 9 cm: k R' 50 (cyl ) = 0.9905 + 0.0710e ( − R50 3.67 ) PlanePlane-parallel chambers, 2 ≤ R50 ≤ 20 cm: k R' 50 ( pp ) = 1.2239 − 0.145(R50 ) 0.214 TG-51 Electron Beam Calibration ) L water k ecal = ρ air ) • Prepl • Pcel |evaluated at electron quality R50 =7.5 cm L water ρ air • Prepl • Pwall • Pcel |evaluated at cobalt energy -Factor that maps dose at CoCo-60 energy into dose at a reference electron energy (that energy whose depth dose falls to 50% at 7.5 cm depth, ~18 MeV) -Allows a specific electron chamber calibration factor for cylindrical chambers (at a later date, when/if available) -Allows the calibration of pp-p chambers by intercomparison with a cylindrical chamber in a highhigh-energy electron beam -kecal is independent of energy 5 Bruce J. Gerbi, Ph.D. Determination of kecal Cylindrical chambers – look up the value in available table PlanePlane-parallel chambers – Can use tabled values but crosscross-calibration versus a cylindrical chamber is recommended kecal , PlanePlane-Parallel Ionization Chambers kecal determined from crosscross-calibration with cylindrical chamber at dref in water using highhighenergy electrons (recommended): (k 60 Co ecal N D , w ) pp = = TGTG-51: Dose Equation for Electrons DwQ = MPgrQ k R' 50 kecal N D ,Co w [Gy ] (Dw )cyl (Mk ) pp ' R50 (MP k Q ' gr R50 ecal k 60 N D ,Co w (Mk ) ) cyl pp ' R50 TGTG-51 Calibration 60 DwQ = dose to water at beam quality, Q M = corrected meter reading PgrQ = gradient correction k R' 50 = energy/chamber dependent factor So now you have the calibration, or the dose rate, at one point, dref However, according to ICRU specifications, the prescription dose is to be reported at the dmax point k ecal = factor tha t maps dose at 60Co energy into dose at reference electron energy 60 N D ,Co Co energy w = absorbed dose to water chamber calibration factor at 60 TG-51 Electron Beam Calibration 6 Bruce J. Gerbi, Ph.D. Absorbed Dose at dmax from dref Determine the dose at dmax from that at dref using clinical %dd data %dd requires stoppingstopping-power ratios – In TGTG-51, realistic stoppingstopping-power ratios (SPRs (SPRs)) have been used instead of monomono-energetic SPRs as in TGTG25 – Expression from Burns et al. as a function of R50 should be used – Prepl (= Pfl X Pgr) is also required (TG21 & TG25) Burns formula for clinical work Accurate enough for clinical work Realistic StoppingStopping-Power Ratios for Water From Burns et al. al. (in water) L ρ w (R50 , z ) = air a + b(ln R50 ) + c (ln R50 ) + d ( z R50 ) 2 1 + e(ln R50 ) + f (ln R50 ) + g (ln R50 ) + h( z R50 ) 2 3 c = 0.088670 g = 0.003085 d = - 0.08402 h = - 0.12460 Where: a = 1.0752 e = - 0.42806 b = - 0.50867 f = 0.064627 These coefficients give an rms deviation of 0.4% and a max. deviation of 1.0% for z/R50 between 0.02 and 1.1. The max. deviation increases to 1.7% if z/R50 values up to 1.2 are considered. Measurement of CA %dd %dd Curve Usually done using automated water scanning system – Exception of magnetically swept beams – Regions close to the surface – Can use cylindrical or plane-parallel chambers – Each type has its advantages & disadvantages OR corrections that are needed Problem: if corrections are not applied in the scanner software, it is difficult to access data and apply corrections Rogers DW, Med. Phys. 2004 TG-51 Electron Beam Calibration 7 Bruce J. Gerbi, Ph.D. Cylindrical chamber corrections Apply gradient correction to raw depth ionization data Need to apply Realistic Stopping Power data to depth ionization data Need to apply fluence correction as a function of electron energy at depth II. B. Cone Factors for Electron Beams Should be determined at dmax – to minimize perturbations – to minimize positioning uncertainty Potential problems – significant contamination by low energy electrons moving dmax toward surface – dmax is broad for high energy electrons Possible solution (IPEMB, 1996) PlanePlane-parallel chamber corrections No gradient correction Need to apply Realistic Stopping Power data to depth ionization data No fluence correction except for Markus & Capintec chambers (TG39) Pwall small correction (1-2%) Ppol small effect but needs to be checked II. C. Measurements in nonnon-water phantoms Water is recommended for absolute calibration Output checks can be done in plastics Plastic phantoms may be more convenient in certain situations • low energy electron beams • use of planeplane-parallel chambers There is added complexity to convert dose in plastic to dose in water using ion chambers – Use dmax or 0.5R 0.5R50 , whichever is greater TG-51 Electron Beam Calibration 8 Bruce J. Gerbi, Ph.D. Use of nonnon-water phantoms Water substitutes should mimic water across the whole electron energy range – mainly in stopping and scattering powers • thus, both the electron density and the effective atomic number should be matched to water • in practical phantoms, this is difficult to achieve (due to the carbon in plastics) – offoff-thethe-shelf material can have large variations in density and scattering power • Must be careful when using these materials • Check the density of the plastic being used, composition is more difficult to verify Measurements of Absorbed Dose in nonnonwater Phantoms The SSD and field size are not to be scaled Chamber must be positioned with its effective point of measurement at the equivalent scaled reference depth in the plastic phantom TG-51 Electron Beam Calibration Use of nonnon-water Phantom Materials Depths need to be scaled Chamber readings need to be multiplied by an appropriate fluencefluence-ratio correction StoppingStopping-power ratios should be taken at the scaled depth Charge storage effects should be kept in mind using polystyrene and PMMA III. D. Dose Determination in Small/Irregular Fields Inherent Problems in Dosimetry of Small Electron Fields – depth of dmax becomes shallower – the output factor may be significantly different than the cone factor if the field size is smaller than ~ E (MeV) /2.5 cm. – isodose coverage is reduced in all directions as the field shrinks 9 Bruce J. Gerbi, Ph.D. Change in %DD v. Field Size 9 MeV Fractional Depth Dose vs Field Size When is special dosimetry required? 16 MeV Fractional Depth Dose vs Field Size 1.2 1.0 1.0 0.8 10x10 open 0.8 3.4 cm diam 4 cm diam fdd 3.4cm diam fdd 10x10 open 0.6 5 cm diam 0.6 When the minimum field dimension is less than the minimum radius of a circular field that produces lateral scatter equilibrium Req = 0.88 E p , 0 4 cm diam 0.4 0.4 where 0.2 0.2 E p , 0 = 0.22 + 1.98 R p + 0.0025R p2 0.0 0.0 0 1 2 3 4 5 6 0 depth (cm) 2 4 6 8 10 depth (cm) 12 Khan F & Higgins PD, PMB 44 (1999), N77N77-N80 Possible Methods for Small Field Dosimetry Models for Small Field Dosimetry Measurements of output and depth dose data Several publications give calculational methods to approximate output – film dosimetry – ionization chamber dosimetry Model to describe change in output, depth of dmax, and isodose changes TG-51 Electron Beam Calibration – Square Root Method (Phys Med Biol, 1981)) ( – Khan Model using Lateral Buildup Ratios (LBRs) LBRs) (Phys Med Biol. 1998 43:274143:2741-54, Phys Med Biol. 44:N7744:N77-N80, Phys Med Biol. 2001 46:N946:N9-14) – Jones approach (Br J Radiol. Radiol. 1990 63:5963:59-64) – Jursinic approach (Med Phys. 1997 24:176524:1765-9) 10 Bruce J. Gerbi, Ph.D. Square Root Method Rectangular electron field %dd determination Take the geometric mean of the percent depth doses for a square field of length (L) and width (W) % D (d ; LxW ) = % D(d ; LxL) ⋅ % D(d ;WxW Khan Model - Lateral Buildup Ratios (LBRs (LBRs)) Using this approach and a sector integration description for LBR, the depth dose, and dose per monitor unit for irregular shaped electron fields can be determined Uses Lateral Buildup Ratios (LBRs ), and (LBRs), Pencil Beam Model parameters σ r ( d ) = effective spread ( width ) Hogstrom KR et al., Phys Med Biol, 1981 Basic Procedure Separate Cone Factors from inin-phantom scatter contributions by measuring percentage depth doses (% (%dd), dd), normalized to the surface Measure %dds as a function of field size, ranging from very small (e.g. 1 cm radius) to “infinitely” infinitely” large fields (e.g. 20x20) TG-51 Electron Beam Calibration Basic Procedure - LBR For each depth, divide the %dd of the small field by that for the large (reference) field This ratio = Lateral Buildup Ratio (LBR), or LBR = % dd x (rx , d ) % dd ∞ (r∞ , d ) 11 Bruce J. Gerbi, Ph.D. LBR v. depth LBR vs Depth 20x20 Cone 16 MV 1.2 Result 1.0 2 cm 0.8 LBR 3 cm 4 cm 6 cm 0.6 8 cm 0.4 Use the values of σ r (d ) for 2 cm diameter cutout, Then, for any radius, the value of LBR can be calculated using: σ r2 (d ) = r 2 / ln 0.2 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1 1 − LBR(0, d ) Z/Rp Small or Irregular Electron Beams New methods of modeling electron depth dose distributions may enable irregular field calculations to become an option, option eliminating the need for many clinical measurements. IV. Electron Beam Algorithms Short history of electron algorithms Description of how data should be entered into treatment planning computers What should be done to commission these algorithm (being consistent with TG53) – Describe some of the pitfalls and limitations of electron algorithms – Discuss normalization of dose distributions for electron algorithms • Restricted field • Extended treatment distance • Plans involving inhomogeneities TG-51 Electron Beam Calibration 12 Bruce J. Gerbi, Ph.D. V. ICRU 71 – Prescribing, Recording, and Reporting electron beam therapy Regular treatments Intra Operative treatments Total Skin Electron treatments VI. Clinically Relevant Topics Inhomogeneities in electron treatments – The effects of inhomogeneities on dose distributions – Computer representation of the effects of dose inhomogeneities Use of bolus Field abutment – – – VII. Library of Clinical Examples Extensive list of clinical applications Each section to detail the following: ElectronElectron-electron, same & different energies ElectronElectron-photon, standard & extended distances Tertiary shielding for field abutment TGTG-70: Conclusion Task Group report to be finished by end of 2005 – Introduction and Purpose of the treatment – History and Description – Pre-requisites: special equipment, measurements – Treatment Planning & Delivery – Quality Assurance TG-51 Electron Beam Calibration 13