Outline Effect of Scan Parameters in Cardiac Imaging with MDCT • • • • • • Mahadevappa Mahesh, M.S., Ph.D. The Russell H. Morgan Department of Radiology and Radiological Science Johns Hopkins University, Baltimore, MD Introduction Fundamentals of Cardiac CT Imaging Temporal and Spatial Resolution Pitch, Geometric Efficiency,… Effect of scan parameters on image quality Conclusions 47th Annual Meeting, Seattle, WA (Yr 2005) Essentials for Cardiac Imaging Key Issues in Cardiac Imaging with CT • • Fast imaging - High temporal resolution to coronary arteries located close to heart muscles freeze cardiac motion and avoid artifacts • that show strong movement during cardiac cycle • • • Fine imaging - High spatial resolution to resolve small lesions in any plane • Radiation dose - the consequences of CT Diastolic Phase versus Heart Rate 400 200 Diastole Exposure time Least cardiac motion is observed during diastolic phase • Diastole phase narrows with increasing heart rate • Desired temporal resolution for motion free cardiac imaging 100 ms 0 40 • 500 ms 250 ms ms Rapid movement is present during systole phase Imaging should be performed during diastole phase Image acquisition and reconstruction are to be synchronized accurately with heart movement imaging 600 High temporal resolution is key for imaging 60 80 100 Heart rate (bpm) 120 • − ~ 250 ms for heart rates ~ 70 bpm − ~ 150 ms for heart rates ~ 100 bpm Motion-free imaging during other phases requires temporal resolution ~50 ms Mahadevappa Mahesh, MS, Ph.D. Johns Hopkins University, Baltimore, MD Essentials for Cardiac Imaging • Most proximal coronary segments (RCA, LAD) require high submillimeter isotropic spatial resolution • Sufficient contrast-to-noise ratio is required to resolve small and lowcontrast structures such as plaques • Low-contrast resolution with limited radiation exposure at shortest exposure time is key RCA LAD 1 High Quality Coronary CTA images • Cardiac imaging is a high demanding application of CT • Temporal, spatial and contrast resolution are to be optimized and Axial Coronal Sagittal also radiation exposure are to be limited Cardiac Images from 16 section MDCT* Temporal Resolution *Mahesh M, Clini Cardio Vasc Img Textbook, pp 1-77, 2004 Prospective ECG Triggering Approaches for stopping heart motion • Temporal resolution 200 – 250 msec Radiation dose minimized Limited data set Acquire all data fast enough to stop cardiac motion • Achieved in MDCT by − Prospective ECG triggering − Retrospective ECG gating Conventional Axial “ Partial Scan ” (Step and Shoot) Preset Delay X-ray ON moving couch-top Preset Delay X-ray ON ECG Mahadevappa Mahesh, MS, Ph.D. Johns Hopkins University, Baltimore, MD 2 Partial vs Segmented Reconstruction Retrospective ECG Gating Temporal Resolution 200 - 250 msec Radiation dose higher than prospective triggering 㻨㻦㻪 Continuous spiral scan Retrospective reconstruction (partial scan: 180o + fan angle) Continuous recording of spiral scan and ECG TR: 200 msec 䝿䚭䝿䚭䝿 ECG moving couch-top Time / Pos. 䝿䚭䝿䚭䝿 䝿䚭䝿䚭䝿 Time/position Segmented reconstruction Factors affecting Temporal Resolution • • Gantry rotation speed Prospective triggering − Retrospective gated • Partial or segmental • • Pitch • Partial scan reconstruction − • continuous segment of projection data at single heartbeat Segmented reconstruction − Post-processing algorithms Different segments of projection data from same phase of cardiac cycle at successive heartbeats used Partial Scan Reconstruction • • • Data from prescribed time range during one cardiac cycle is selected for reconstruction Multisegment Reconstruction • Each heart cycle provides segment of data required for partial scan reconstruction • For ‘M’ segments (‘M’ heart cycles), maximum temporal resolution is 200 to 270 ms temporal resolution is achieved for 0.4 s gantry rotation Works best with very low pitch (p<0.25) • Mahadevappa Mahesh, MS, Ph.D. Johns Hopkins University, Baltimore, MD TR: <100 msec Retrospective Reconstruction Image Reconstruction − 䝿䚭䝿䚭䝿 − ‘TR/2M’ where TR is gantry rotation time − 100 ms with 2 segments and 50 ms with 4 segments for TR of 0.4 sec For M segment, temporal resolution vary between TR/2 to TR/2M 3 Temporal Resolution: Partial vs Segmented ECG gated reconstruction Temporal Resolution • Depends on gantry rotation time - • 0.5 - 0.37 second with 64 section MDCT scanners Up to 80 - 250 ms achieved through partial scans or sub-segment data reconstruction • Improves with sub-segment data reconstruction, but spatial resolution decreases and motion Half-scan reconstruction Temporal resolution: 250 msec Segmented reconstruction Temporal resolution: ~105 msec artifacts increases Courtesy Toshiba Coronary Angiography vs CT Angiography Right coronary artery showing calcification Spatial Resolution CT angiography Coronary Angiogram Volume rendered CT angiogram of right coronary artery acquired at 16x0.75 mm and 0.42 sec rotation time Hoffmann et.al., AJR: 182, March 2004 Factors affecting Spatial Resolution • • • • • • SDCT versus MDCT MDCT detector array designs Section thickness/Section collimation/ Effective section thickness X-ray Tube Tube Collimator Collimated Slice Pitch Reconstruction Increment Detector Collimator Reconstruction algorithms Patient motion … 1-Row Detector Single row detector CT (SDCT) 8-Row Detector Multiple row detector CT (MDCT) *Mahesh M, RadioGraphics, 22: 949-962, 2002 Mahadevappa Mahesh, MS, Ph.D. Johns Hopkins University, Baltimore, MD 4 Detector Element Arrays* in 4-section MDCT scanners 16 x 1.25 mm How are detector elements used in MDCT? Uniform 20 mm 4 x 1.25 mm Detectors 5 2.5 1.5 1 1 1.5 2.5 5 Non-uniform 20 mm 20 mm 4 x 0.5 Hybrid 15 mm Switching Array 4-section scanners collect 4 simultaneous channels of data 15 mm 32 mm Z-axis *Mahesh M, RadioGraphics, 22: 949-962, 2002 Detector elements in 16 section MDCT scanners* 4 x 1.25 mm Detector elements in >32 section MDCT scanners 4 x 1.25 mm 4 x 1.2 mm 32 x 0.6 mm 4 x 1.2 mm 16 x 0.625 mm Siemens - Sensation 64 GE - Lightspeed 16 28.8 mm 20 mm 4 x 1.5 mm 16 x 0.75 mm 64 x 0.5 mm 4 x 1.5 mm Toshiba - Aquilion 64 Siemens - Sensation 16 Philips - Mx8000 IDT 32 mm 6 x 1.25 mm 24 mm 40 x 0.6 mm 6 x 1.25 mm Philips - Brilliance 40 16 x 0.5 mm 12 x 1 mm 40 mm 12 x 1 mm 64 x 0.625 mm Toshiba - Aquilion 16 GE - Lightspeed 64 32 mm 40 mm *Mahesh M, Clini Cardio Vasc Img Textbook, pp 1-77, 2004 Pitch redefined for MDCT Beam Pitch = I Pitch: Definition, Confusion… Detector Pitch = W T Beam Pitch = Detector Pitch I - Table feed (mm/rotation) W - Beam width (mm) N I W I T = I N*T = Pitch† T - Single DAS channel width (mm) N - Number of active DAS channels † IEC Part 2-44, 2003 Mahadevappa Mahesh, MS, Ph.D. Johns Hopkins University, Baltimore, MD 5 Dose in MDCT varies as: Why Cardiac CT protocols use Low Pitch†? Z-position • Pitch >1 implies extended imaging and reduced patient dose with lower axial Helical scan direction • • data overlap Z-position • Pitch <1 implies overlapping and higher patient dose with higher axial resolution 1 Pitch† • (mAs/rotation) Time Slope: Table feed speed • N - Number of active DAS channels TR - Gantry Rotation Time (ms) TRR - Time for one heart beat (ms) M - Number of subsequent heart cycles − For ex: For heart rates 45-100 bpm, with TR 0.5 s, TQ 250-360 ms, P = 0.375 to 0.875 − P = 1.48 CTDI = 20.6 mGy 45% lower Mahadevappa Mahesh, MS, Ph.D. Johns Hopkins University, Baltimore, MD For ex: For heart rate 60 bpm, with TR 0.4 s, N = 16 and M = 2, P = 0.21 For ex: For heart rate 60 bpm, with TR 0.4 s, N = 16 and M = 3, P = 0.15 Section Collimation (SC) vs Section Width (SW) Effect of Pitch on Dose and Image Quality P = 0.83 CTDI = 37 mGy Data gaps with higher pitch Pitch is further restricted by the number of segments used in reconstruction N - Number of active DAS channels TR - Gantry Rotation Time (ms) TRR - Time for one heart beat (ms) TQ - Partial scan rotation time (ms) P = 0.64 CTDI = 47.8 mGy 30% higher Typical pitch: 0.20 - 0.4 Multiple Segment Reconstruction Retrospective ECG-gating with single segment (partial scan) reconstruction requires limiting the pitch dependent on heart rate − Hence pitch is low, and radiation dose is high • Single Segment Reconstruction • High quality 3D with minimal artifacts requires resolution Dose ∝ Higher pitch produces gaps • Section collimation is total beam collimation divided by number of active detector channels • Section width is the true thickness of reconstructed image, measured as FWHM of slice sensitivity profile • SW has to be larger than or equal to SC • SW ≈ 1.3 (±0.2) SC Affected by collimation, pitch, reconstruction algorithm, z-filter … Slice Sensitivity Profiles: conventional and spiral acquisition 6 Reconstruction Interval (RI) • • • • Defines degree of overlap between axial scans Reconstruction Interval • Overlapping results in large number of images but ensures optimum lesion display and improves MPR and 3D images without increasing patient dose For routine applications including small structures detection a 30% section overlap is sufficient • • For MPR and 3Ds, at least 50% overlap is desirable Large RI yields fewer images but provides suboptimum lesion detection • Limited by reconstruction time, number of images to interpret and storage space Independent of section collimation or section width Effect of Reconstruction Interval Theoretic optimum is smaller than half the section width with minimal added value in clinical practice Effect of Reconstruction Interval SW 0.5 mm, RI 0.3 mm SW 0.5 mm, RI 5.0 mm SW 0.5 mm, RI 0.3 mm SW 0.5 mm, RI 0.5 mm 301 images 19 images 301 images 184 images Effect of Reconstruction Algorthims: Spatial Resolution and Image Noise Trade-offs Effect of Slice Thickness on Image Noise n lutio eso lR atia Sp 10 mm Smooth Medium 5.0 mm 0.5 mm Sharp Reconstruction Filters Smoothing filters results in minor reductions in spatial resolution but require less dose for constant image noise ©Johns Mahadevappa Mahesh, MS, Ph.D. Johns Hopkins University, Baltimore, MD Scan at thin collimation (raw data) for high spatial resolution, but reconstruct thick sections with overlaps to improve image noise Hopkins 32 slice MDCT 7 High Contrast Spatial Resolution in Z direction • Phantom - 0.5 mm spacing • Scan modes - 32 x 0.5 mm • Technique - 120 kVp, 50 mAs Spatial Resolution • Axial or In-plane Resolution − • MDCT: 10 - 20 lp/cm (resolvable ~ 0.5- 0.25 mm) Z-axis Resolution - influenced by section collimation 0.6 mm 0.5 mm 0.4 mm 0.3 mm ©Johns − MDCT: 7- 15 lp/cm (resolvable ~ 0.7 - 0.3 mm) Hopkins CT Dose Index (CTDI) comparison: 4 vs 16* vs 64 MDCT¶ scanner Geometric Efficiency Sensation 64: Increase in CTDIw of nearly 4-19% compared to Sensation 16 as measured for head and body phantoms ¶Siemens * CTDIw(weighted average) normalized to 16x0.75 mm scan mode Geometric Efficiency Focal spot • Amount of radiation excluded (penumbra) Geometric Efficiency Beam collimator • Decreases with thinner sections and fewer detector elements Penumbra • Improves with thicker sections or with more active thin sections relative to the radiation collected by the detectors in forming an image • SDCT Penumbra caused by finite focal spot size − contributes to image and patient dose in SDCT − contributes to patient dose but not to image in MDCT Penumbra Penumbra Penumbra (over-beaming to ensure equal image quality) MDCT Mahadevappa Mahesh, MS, Ph.D. Johns Hopkins University, Baltimore, MD Penumbra 4 x 1.25 mm 4 x 2.5 mm 16 x 1.25 mm Strong effect with thin collimation Effect decreases with thicker collimation Effect diminishes further with more detectors 8 Advantage of Thin Sections Geometric Efficiency in MDCT • Dose in 4 slice scanners grows markedly with thin collimation but less so for 16 and 16+ scanners In general, thin sections yields higher z-axis resolution, improves partial volume, but requires higher tube current to reduce image noise, which leads to higher doses (especially with few detectors) Dose grows markedly with thin collimation and few active detector elements but less so for thicker collimation and more detectors ©Johns Hopkins ©Johns Hopkins MDCT scanners Effect of X-ray Beam Energy (kVp) Effect of kV and mAs on Image Noise 135 kV with 8.0 SD 37 mGy CTDI 120 kV with 10.0 SD 29 mGy CTDI ~22% dose reduction Effect of kV and mAs on Image Noise Artifacts Mahadevappa Mahesh, MS, Ph.D. Johns Hopkins University, Baltimore, MD 9 Pulsation Artifacts • Common artifacts due to cardiac pulsation • Multiple segment reconstruction is desirable Banding Artifacts Banding artifacts Artifacts due to increased heart rate starting at 51 bpm increased to 69 bpm Radiographics 2005 Radiographics 2005 Artifacts due to incomplete breath holding Streak Artifacts Coronal Sagittal Metallis structure but no artifact visible on axial images Streak artifacts due to earlier stent placement visible on thin MIP and MPR images Axial images show no artifacts Radiographics 2005 Impact of Dose Modulation: Chest CT MDCT 64* CT Dose Modulation Radiation dose: Lateral: 16% increase, AP: 25% reduction *Mahesh, Kamel & Fishman, Evaluation of ‘CareDose’ on Siemens Sensation 64 MDCT scanner Mahadevappa Mahesh, MS, Ph.D. Johns Hopkins University, Baltimore, MD 10 Impact of Dose Modulation: Abdominal CT MDCT 64* Radiation dose: Lateral: 28% increase, AP: 48% reduction *Mahesh, Kamel & Fishman, Evaluation of ‘CareDose’ on Siemens Sensation 64 MDCT scanner Conclusions Dose Modulation for varying Image Noise Index • Cardiac imaging is highly demanding application of MDCT and is possible due to technological advances • Understanding trade-offs between various scan parameters that affects image quality is key • Low SD - High dose 184 mA Medium SD & dose 69 mA High SD - Low dose 48 mA Cardiac CT has the potential to becoming reliable tool for noninvasive diagnosis and prevention of cardiac and coronary artery disease Toshiba Future with CT Mahadevappa Mahesh, MS, Ph.D. Johns Hopkins University, Baltimore, MD 11