Introduction PET Scanner Performance, Quality Assurance and Acceptance Testing Quality Control Procedures required to ensure that the distribution of radiation emitted from a patient is accurately reflected in the measured raw data. Facets Brad Kemp, PhD Department of Radiology Mayo Clinic – quality assurance: – verify scanner is operating properly – identify problems prior to scanning patients – acquisition of corrections to compensate for known imperfections in data measurement TU-A-319-01 Review: Types of Coincident Events Data Acquisition Sinogram Sort parallel LORs True Scatter Angle Radial Distance Randoms Gamma ray (Singles) Line of Response (LOR) Prompts = True + Scatter + Randoms 1 Block Detectors Block Detectors 2D array of crystals attached to 4 PMTs 2D array of crystals attached to 4 PMTs Crystal element B A D C z PMT x X= Light guides define the pattern of light distribution in the block Determines light output (sharing) to PMT's System Correction: Normalization Purpose: Corrects for the variations in efficiency in LORs in each slice of the sinogram • A uniformity correction • For 2D: Direct Measurement using a low activity source • For 3D: Indirect Measurement using a uniform phantom • Acquired quarterly or after system maintenance B+D A+ B+C + D Z= C+D A+ B+C + D Each crystal produces a unique combination of signals in the PMTs Compare signals X and Z to preset values in a 2D lookup table - map X and Z to a single crystal System Correction: Normalization Sinogram of a Uniform Cylinder Before Normalization After Normalization 2 Blank Scan Quality Assurance Quality Assurance Test Requirements • Well defined regimen of measurements – Quick and easy to conduct • On stand-alone PET systems, used with transmission images to create attenuation correction factors • Acquired daily - good source of QA data • Emission scan of uniform cylinder Good blank Bad block Miscoded event – Sensitive to modes of failure of the scanner – Preferably quantitative, not qualitative Quantitative Analysis of Blank Scans • The quantitative analysis of blank scans is an important aspect of the QA of a PET scanner Overview: Daily QA Daily QA analysis calculates a global and slice normalized mean square error (NMSE) by using a reference scan; the reference scan is defined as the blank scan acquired immediately after normalization • Quantitative analysis of blank scans: - validates system calibration - provides information on crystal, block or module (bucket) efficiencies with respect to system average - monitors system stability - ensures validity of normalization 3 Overview: Advanced QA Advanced QA analysis computes the bias in efficiency for each crystal, block and module. Those crystals, blocks and modules with a bias greater than a preset limit are listed in a table. Deviation of module bias with respect to the reference scan is also shown. Quality Assurance Detector and Electronics Characterization – Singles mode detector calibration • Crystal map • PMT gain adjustment • Energy map – Coincidence timing calibration Did the engineer calibrate the scanner properly? Crystal Map Verification of Crystal Map QA of crystal maps created during calibration of the scanner Purpose: Map the position of the detected event to a specific crystal Obtained quarterly or after detector maintenance Count distribution in a block Crystal Map Look Up Table 4 PMT Gain Adjustment Coincidence Timing Calibration Purpose: Purpose: Balances the gain characteristics of the PMTs in a block Compensates for PMT gain drift with temperature, age, … Acquired weekly (daily) or after detector maintenance Adjust for timing delays so events from all blocks are timestamped equivalently Acquired weekly or after detector maintenance 0.0 Update Gains Update Gains Block Bias (%) -1.0 -2.0 Morning -3.0 Afternoon -4.0 γ γ -5.0 Coincidence ? Yes Register Event -6.0 28-Feb 7-Mar 14-Mar 21-Mar 28-Mar Date Coincidence Timing Calibration Daily Quality Assurance Blank scan created with incorrect CTC calibration Can the Scanner be Used Today? Potential Problems • System stability, drifts • Detector module / PMT/ preamp failure • Loose cables, connectors • Inoperable gantry motors, source loader Daily QA Regimen • Automatic PMT gain adjustment • Blank scan • QA will detect but not prevent these problems 5 Update Gains and Block Bias Example: NMSE and Block Bias Advanced QA analysis can be used to assess the stability of the system and the effectiveness of daily Update Gains QA must be sensitive to modes of failure of the scanner Update Gains Block Bias (%) -2.0 Morning -3.0 Afternoon -4.0 -5.0 -6.0 28-Feb 7-Mar 14-Mar 21-Mar 28-Mar 6 Global NMSE Slice NMSE Block Bias 0.008 4 2 0.006 0 -2 0.004 -4 Block Bias (%) Update Gains Normalized Mean Square Error 0.0 -1.0 0.010 0.002 -6 Date 0.000 -8 10-Dec-02 29-Jan-03 20-Mar-03 9-May-03 28-Jun-03 17-Aug-03 Date Example: Blank Scan Count Rate The acquisition count rate is used as an estimate of system sensitivity. Daily QA: Loose Cable Loose Cable Better 800 Acquisition Rate (kcps) 700 600 500 400 300 200 100 0 Nov-01 Feb-02 May-02 Sep-02 Dec-02 Mar-03 Jun-03 Oct-03 Date 6 QA Schedule Detector and Electronics Characterization – Crystal map – PMT gain – Coincidence timing Frequency Quarterly Weekly (daily) Weekly System Corrections Normalization – Scanner calibration – Blank Scan – Quarterly Quarterly (weekly) Daily NEMA NU 2-2001 NEMA NU 2-2001 PET Performance Measurements National Electrical Manufacturers Association. NEMA Standards Publication NU 2-2001: Performance Measurements of Positron Emission Tomographs Phantoms for NU 2-2001 NEMA NU 2-2001 Acceptance Testing Can we use our new scanner? Annual QA Is the scanner still performing within specification? Three phantoms: Scatter phantom 203x700mm phantom with activity in line source Sensitivity phantom Image quality phantom 7 NEMA NU 2-2001 • Measures performance of PET scanners under conditions that attempt to represent whole body studies – Phantom of greater length – Out of field activity • Standardizes oblique LOR manipulation in 3D acquisitions NEMA NU 2-2001 Performance Measures: • Spatial Resolution (Transaxial, axial) • Sensitivity • Scatter Fraction • Count Losses • Count Rate correction accuracy • Daube-Witherspoon ME, et al. PET Performance Measurements Using the NEMA NU 2-2001 Standard. J Nucl Med 2002;43:1398-1409. Spatial Resolution • Spatial resolution of a system represents its ability to distinguish between two points of radioactivity • F-18 point sources in air at six locations: – (0,1), (0,10) and (10,0) cm – Center of axial FOV and ¼ axial FOV from center • Reconstruct: image pixel < ⅓ expected FWHM • Profile width ~ 2 times FWHM • Report FWHM and FWTM in radial, tangential and axial directions • Image Quality Sensitivity • Sensitivity of a scanner represents its ability to detect annihilation radiation • Rate of true coincidence counts per unit radioactivity (expressed in cps/kBq) in absence of attenuating media • Rationale: need material around source to ensure annihilation of positrons, but this material also attenuates the annihilation photons • Based on technique by Bailey DL, Jones T, et al. Eur J Nucl Med 1991;18:374-379. 8 Sensitivity •Successive measurements with a 700 mm line source of F-18 surrounded by nested, known absorbers Sensitivity • Measure at radial locations of 0 and 10cm • Report system sensitivity and slice sensitivity profile •The count rate with no absorber is extrapolated from these measures Count Rate, R (cps) 1.5E+05 R0 R0 System Sensitivity = Activity 1.0E+05 Axial sensitivity (cps/kBq) 0.30 0.25 0.20 0.15 0.10 0.05 0.00 0 5.0E+04 10 20 30 40 Slice 0.0E+00 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 Sleeve Thickness (cm) Intrinsic Scatter Fraction • Scatter fraction is a measure of the relative system sensitivity to scatter SF = Scatter Scatter + Trues • Use 203 mm diameter polyethylene cylinder of length 700 mm, with activity located in a line source of diameter 2.3mm that is 4.5mm off axis • Measured with low activity (Randoms:Trues = 1%) to avoid random coincidences, deadtime and pulse pileup. Count Rate Performance • Measurement of count rate performance gives an indication of scanner performance as a function of activity • Use 700 mm long polyethylene cylinder • Measured with high initial activity of F-18 3D: 800 MBq; 2D: 5 GBq • Acquire data until randoms and deadtime losses are negligible (14 to 18 hrs) 9 Count Rate Performance Count Rate Performance 40 • Calculate Noise Equivalent Count Rate 35 R NECR (1R) NECR (2R) 30 RTrues + RScatter + k ⋅ RRandoms NECR (kcps) RNEC = 2 Trues RNEC: Figure of merit relating scanner performance to image SNR after randoms and scatter corrections. 25 20 15 10 For NEMA, k = 1 (calculated Randoms) 5 0 • Report peak NEC and effective activity concentration at peak 0 5 10 15 20 25 30 Activity concentration (kBq/cc) Image Quality Measurement • Standardized imaging situation that simulates a clinical whole body imaging condition • Phantom consists of a torso phantom with hot and cold lesions in a warm background • Scatter phantom abutted to image quality phantom Image Quality Measurement • • • • • • Hot spheres: 10, 13, 17, 22 mm id Cold spheres: 28, 37 mm id Lung insert Activity in hot spheres 8 and 4 times that of background Activity in background 5.3 kBq/ml Simulated acquisition 100cm in 60min 60 min Tacq = ∆z ≈ 7 min 100 cm • Repeat acquisition three times • Reconstruct using clinical protocol 10 Image Quality Measurement • Report image contrast and SNRs for hot and cold lesions, residual error in lung, variability in background • Visual inspection for artifacts NEMA and LSO • LSO is inherently radioactive • Background radiation gives rise to Randoms, some Trues • Implications for NEMA: cannot obtain Randoms:Trues ratio of 1% • For count rate measure – acquire delayed events • For sensitivity – acquire blank scan to estimate Trues • Watson CC, et al. NEMA NU 2 Performance Tests for Scanners with Intrinsic Radioactivity. J Nucl Med 2004;45:822-826. Summary • • • • Review of PET physics Need for QA regimen Calibrations and corrections required for PET imaging NEMA NU 2-2001 Performance Standard procedures 11