4D CT Just kidding! Daniel A. Low Room Shielding for IMRT Mallinckrodt Institute of Radiology Washington University School of Medicine St. Louis, Missouri 63110 Daniel A. Low Washington Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Outline • • • • • Mallinckrodt Institute of Radiology Washington University School of Medicine St. Louis, Missouri 63110 General facility planning General facility planning Shielding formalism Conventional shielding example IMRT changes in formalism Neutrons • IMRT versus Conventional – Tomotherapy – Conventional Linac – SMLC – Conventional Linac – DMLC • Tomotherapy: – Power, cooling, data transfer similar to conventional RT • Conventional Linac – Power, cooling, data transfer is same as conventional RT • Shielding? Washington Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Conventional Shielding Formalism • Start with the conventional formalism • Concept – determine the amount of shielding to limit dose/dose rate beyond a shielding barrier or barriers • Geometry (distances, angles) • Multiple barriers • Different shielding materials – reflection, transmission coefficients • Different particles (photons, neutrons) Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Conventional Radiation Therapy Room Shielding Formalism • NCRP Reports No. 49 (≤ (≤10 MeV), MeV), 51, and 79 • Primary and secondary (leakage and scatter) barriers • Barrier thickness based on: Where: B = barrier transmission P = weekly design exposure rate W = workload U = use factor T = occupancy d = distance from radiation source m = correction factors for secondary barriers Pd 2 B= •m WUT McGinley, Shielding Technologies for Radiation Oncology Facilities, second edition, Med. Phys. Pub. 2002 Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Typical High Energy Installation 6'-7'concrete 7'-8' concrete 7'-8' concrete • The fraction of transmission that the barrier requires to reduce the radiation intensity to an acceptable value • Unitless 1.3 m secondary barrier primary barrier isocenter primary barrier linac rotation plane Barrier Transmission B Pd2 B= •m WUT Washington Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Design Exposure Rate P Workload • The dose rate allowed by regulatory agencies • Typically weekly • The amount the linear accelerator is operating, defined at 1 m (typically isocenter) • Nominally equivalent to the total weekly dose delivered • Units: Gy week-1 – Some regulations have “instantaneous” dose rates as well • Units = Sv week-1 B= Pd2 •m WUT B= Pd2 •m WUT Washington Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Use Factor U Occupancy Factor T • The fraction of time that a linear accelerator is pointing towards the barrier • Relevant only for “primary” barriers (take as 1 for secondary barriers) • Unitless • Fraction of time a location is occupied by a human • Typically only used for “uncontrolled areas” • Most relevant to locations such as hallways, staircases, etc. • Unitless Pd2 B= •m WUT Pd2 B= •m WUT Washington Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Distance d Secondary Barriers • Typically distance measured from isocenter to point of calculation • Divided into leakage and scatter • Leakage = head leakage – Location from which distance is measured depends on source of radiation – Typical specification is that leakage is 0.1% of primary beam intensity – Distances measured from linac head (use closest approach for d) • Very important component (squared) • When outside barrier, distance overestimates dose falloff – barrier acts as closer source – dose falloff is faster than using d – conservative estimate • Units: meters B= 2 Pd •m WUT • Scatter – Patient – Wall/barrier Washington Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Leakage POI secondary barrier Scatter – Patient B= 2 1000 Pd sec B= WT reflection coefficient isocenter primary barrier field size (cm2) at scatterer scatterer-POI target-scattering distance surface (1m) secondary barrier dsec primary barrier P 2 2 400 d sec d sca aWT F dsec isocenter primary barrier linac rotation plane primary barrier linac rotation plane Washington Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Scatter - Patient Scatter - Barrier Scattering barrier-POI distance target-scattering barrier distance • a = ratio of scattered radiation at 1m from the patient to the primary radiation at 1m from the target • F = field size at patient • Scattered beam has different energy than primary B= reflection coefficient 2 2 Pd sec d sca αAWTU field size (cm2) at scattering surface secondary barrier dsec – Compton scattering primary barrier isocenter dsca linac rotation plane primary barrier Washington Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Scatter - Barrier Parameters • T Occupancy factor – occupational = 1 – uncontrolled » 1.0 = full occupancy, offices, shops, labs » ¼ = partial occupancy, corridors, rest rooms » 1/16 = occasional occupancy, waiting rooms, toilets, stairways, elevators • α = reflection coefficient of scattering barrier – barrier material – beam energy – scattering angle • A = field size at scattering barrier (cm2) • Scattered beam has different energy than primary – Compton scattering B= • U Use factor (primary) – 0.25 walls, ceiling – 1.0 floor • W Workload (photon treatments only) 2 2 Pd sec d sca αAWTU – LowLow-energy 350 Gy week-1 – HighHigh-energy 250 Gy week-1 Washington Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Parameters Parameters • a (scatter fractions) • P (occupational) – Scattering angle and beam energy – Taylor and Rodgers (1999) – Maximum permissible dose rate – 0.05 Sv per year – ALARA = 0.1 mSv per year • P (uncontrolled) – 1.0 mSv per year – 0.02 mSv in any one hour (cannot irradiate annual limit too quickly!) – This is satisfied if either: » Dose to an occupyable space (T=1) at max exposure <0.5 mSv in one year and 0.02 mSv in one hour (T = 1) » Dose to most irradiated person limited to 1.0 mSv in one year (implies occupancy factor) – Other statestate-defined limitations! Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Parameters 6 MV 10 MV 18 MV 24 MV 10 1.04 x10-2 1.66 x10-2 1.42 x10-2 1.78 x10-2 20 6.73 x10-3 5.79 x10-3 5.39 x10-3 6.32 x10-3 30 2.77 x10-3 3.18 x10-3 2.53 x10-3 2.74 x10-3 45 1.39 x10-3 1.35 x10-3 8.64 x10-4 8.30 x10-4 60 8.24 x10-4 7.46 x10-4 4.24 x10-4 3.86 x10-4 90 4.26 x10-4 3.81 x10-4 1.89 x10-4 1.74 x10-4 135 3.00 x10-4 3.02 x10-4 1.24 x10-4 1.20 x10-4 150 2.87 x10-4 2.74 x10-4 1.20 x10-4 1.13 x10-4 Scattering fraction decreases with increasing beam energy Scattering fraction decreases with increasing scatter angle Table II: Scatter fractions (a) at 1 m from a human-sized phantom, target-to-phantom distance of 1 m and field size of 400 cm2. Data from McGinley(McGinley 2002) and Taylor and Rogers(Taylor and Rodgers Washington 1999). WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Barrier Transmission to Thickness • NCRP 49 and 51 • Can use TVLs from NCRP 51 • Scattered mean beam energy (MeV (MeV)) Beam E (MV) Angle (deg) Scatter Angle (deg) 0 10 20 30 40 50 70 90 6 1.6 1.4 1.2 0.9 0.7 0.5 0.4 0.2 10 2.7 2.0 1.3 1.0 0.7 0.5 0.4 0.2 18 5.0 3.2 2.1 1.3 0.9 0.6 0.4 0.3 24 5.6 3.9 2.7 1.7 1.1 0.8 0.5 0.3 Scattered beam energy decreases with increasing scattering angle Incident Electron Energy (MeV) MeV) 6 18 24 Table III: Mean energy of patient-scattered radiation as a function of scattering angle and megavoltage beam energy. Data from McGinley(McGinley 2002) and Taylor and Rogers(Taylor and Rodgers 1999). Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Material TVL1 (m) first tvl TVLe (m) subsequent tvls concrete steel lead concrete steel lead concrete steel lead 0.35 0.099 0.055 0.47 0.108 0.51 0.109 - 0.035 0.099 0.057 0.43 0.108 0.46 0.109 - Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Example Calculations • • • • • • S d IMRT Modifications Primary shield, occupational (T=1), ordinary concrete 20 MV, 4.4 m away, lateral barrier (U=1/4) W = 500 Gy week-1 at isocenter P = 0.0001 Sv week-1 (ALARA) B = 1.55 x 10-5 Three ways to calculate (S = shield thickness): – NCRP 51 curve E.8 S = 2.21 m – NCRP 51 TVL table » T1 = 0.48m, Te = 0.44 m » S = 2.16m – Varian has TVL data (0.457 m) » S = 2.20m B= Pd 2 WUT • All methods have very similar results! – Total TD – Total MUs W = TD = Ec • MU ≅ MU E ≡ TD MU • Ec = conventional radiation therapy delivery efficiency factor, assumed to be equal to 1 • The workload value is the same for primary and secondary barrier calculations Washington Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS IMRT Room Shielding Formalism IMRT Room Shielding Formalism • Depending on the complexity of the target volume and the intensity modulation device Ec >> E IMRT MU c << MU IMRT Conventional efficiency • Workload is either IMRT efficiency • Because EIMRT<< 1 – decouple workloads for primary, leakage, and scatter shielding calculations • Occupancy factor (T) remains unchanged • Weekly design dose rate (P) remains unchanged • Primary, leakage, and scatter components still separated • For workload estimates, the t total delivered weekly monitor units and the total delivered weekly dose are decoupled • Look at – MLC IMRT – sequential tomotherapy delivery Washington Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Primary Barrier FormalismFormalism-MLC • W = TD • U approximately same as conventional Primary Barrier Formalism -Sequential Tomotherapy • W = TD • Average number of indexes (I) used to treat patients • U has well defined, broad distribution in angle – Define as product of – fraction of gantry angle (κ) that the beam is incident on the primary barrier – λ is the primary barrier leakage correction factor κ = θ i θ t WtUt =TD Iκλ • The product Iκλ may be very close to conventional therapy use factor Washington Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Secondary Barrier Formalism - Leakage • W = MUs • Workload increases by >2 relative to conventional • There may be an increase in secondary barrier thickness due to the increased leakage component Secondary Barrier Formalism - Scatter • W = TD Washington Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Experimental Verification Experimental Verification • DMLC delivery 14 cm wide, 40 cm long portal was used, with 2, 4, 6, 8, 10, 12, and 14 cm wide sliding windows • For tomotherapy a 3.7 x 20 cm2 open field (at isocenter) was used • 300 cm3, 6 atmosphere air-filled pressurized ionization chamber located at 30 cm beyond the primary barrier Center of the Field Center of the Field Center of the Field Center of the Field Dose Fraction = 0.000 Dose Fraction = 0.222 Dose Fraction = 0.778 Dose Fraction = 1.000 Washington Varian Clinac 2300C/D Measurement Point 30cm from the wall Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Fraction Dose of Open Field 4 cm Window Example WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Results - DMLC 0.7 Window Width 14x40 cm2 2 4 6 8 10 12 14 4, 10x10 cm2 0.6 0.5 0.4 0.3 0.2 0.1 0.0 0.0 0.2 0.4 0.6 Fraction MUs 0.8 1.0 Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS EIMRT Beyond Shield 0.194 0.264 0.343 0.394 0.448 0.492 0.521 0.326 E = efficiency EIMRT Isocenter Ratio ±0.04 0.169 0.259 0.332 0.393 0.444 0.486 0.524 0.323 1.15 1.02 1.03 1.00 1.01 1.01 0.99 1.01 Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Results - Tomotherapy Lamda 1.2 1.7 1.0 1.6 1.5 0.8 1.4 0.6 λ Relative Dose Rate Results - Tomotherapy 1.3 0.4 1.2 0.2 1.1 0.0 -15 -10 -5 0 5 10 1.0 15 Incidence Gantry Angle (deg) Normalized Dose Rate Conservative estimate! 50 100 150 200 250 300 350 400 Arc Length (deg) Washington Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Efficiencies (1/E) Normalized Output Factors Inside and Outside the Room Isocenter and Inside Room 1.2 0 Conventional Beam Energy 1.0 Beam Intensity Modulated unwedged (MU/cGy) wedged (MU/cGy) MLC (MU/cGy) 6 MV 1.2 2.4 3.4 Serial Tomothera py (MU/cGy) 9.7 18 MV 1.0 1.5 2.8 8.1 25 MV 1.0 1.5 2.8 8.1 0.8 Outside Room 0.6 Outside Room Inside Room Isocenter 0.4 0.2 0.0 0 5 10 15 20 25 30 35 Ratio of MUs to tumor dose (1/E) for typical conventional and IMRT deliveries. Data from Followill et al.(Followill, Geis and Boyer 1997) 40 Field Size (cm) Washington Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Results - Shielding Neutrons • Neutron generation is function of MUs • IMRT • Data from Rodgers (2001) • r = percent IMRT r (%) 1/E 0 – MLC: efficiency Additional TVLs 0 TVL 50 4 0.40 TVL 50 10 0.74 TVL 100 4 0.60 TVL 100 10 1.0 TVL Field Size (cm x cm) Gantry angle (deg) Neutron Rem (nSv/MU) Photon Dose (nGy/MU) 0x0 0 1.12 0.77 20x20 0 0.98 0.70 40x40 0 0.78 0.73 90 0.60 0.80 270 0.83 0.80 (7 angles) 0.85 0.54 Abdomen IMRT 0 1.14 0.48 Breast IMRT 0 1.02 Neutron equivalent dose and photon dose 40x40 outside the treatment room door for some tested beam configurations. Data courtesy of 40x40 Boyer (private communication). Abdomen IMRT Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS 0.58 Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS Conclusions Conclusions • MLC – Significant Field Size Factor (0.35 - 0.80 relative to 40x40 cm2) – DMLC » 40x40 cm2 overestimates IMRT conformal therapy shielding requirements » Using W=TD - accurate within 10% to 2 cm window – Primary Barriers - Comparable to conventional shielding – Secondary Barriers - Possible increase due to leakage Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS • Sequential Tomotherapy » MU component (λ (λ) from 15% to 60% compared to geometric estimate TD * κ of single index workload. » 6 indexes, I κ λ ranges from 0.52 for 180° 180° to 0.30 for 360 ° » Very similar to Use factors of conventional! – Primary Barrier » Comparable to conventional shielding due to reduced number of treated patients and I κ λ product – Secondary Barriers » Increased due to leakage » MIR experience ≅ 3,000 MU/fraction » 300,000 MU/week for 20 patients » Similar to high energy W WtUt = TDIκλ Washington WASHINGTON• WASHINGTON•UNIVERSITY• UNIVERSITY•IN• IN•ST• ST•LOUIS