4D CT Just kidding! Room Shielding for IMRT

advertisement
4D CT
Just kidding!
Daniel A. Low
Room Shielding for IMRT
Mallinckrodt Institute of Radiology
Washington University School of Medicine
St. Louis, Missouri 63110
Daniel A. Low
Washington
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Outline
•
•
•
•
•
Mallinckrodt Institute of Radiology
Washington University School of Medicine
St. Louis, Missouri 63110
General facility planning
General facility planning
Shielding formalism
Conventional shielding example
IMRT changes in formalism
Neutrons
• IMRT versus Conventional
– Tomotherapy
– Conventional Linac – SMLC
– Conventional Linac – DMLC
• Tomotherapy:
– Power, cooling, data transfer similar to
conventional RT
• Conventional Linac
– Power, cooling, data transfer is same as
conventional RT
• Shielding?
Washington
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Conventional Shielding Formalism
• Start with the conventional formalism
• Concept – determine the amount of shielding
to limit dose/dose rate beyond a shielding
barrier or barriers
• Geometry (distances, angles)
• Multiple barriers
• Different shielding materials
– reflection, transmission coefficients
• Different particles (photons, neutrons)
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Conventional Radiation Therapy
Room Shielding Formalism
• NCRP Reports No. 49 (≤
(≤10 MeV),
MeV), 51, and 79
• Primary and secondary (leakage and scatter) barriers
• Barrier thickness based on:
Where:
B = barrier transmission
P = weekly design exposure rate
W = workload
U = use factor
T = occupancy
d = distance from radiation source
m = correction factors for
secondary barriers
Pd 2
B=
•m
WUT
McGinley, Shielding Technologies for Radiation Oncology
Facilities, second edition, Med. Phys. Pub. 2002
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Typical High Energy Installation
6'-7'concrete
7'-8'
concrete
7'-8'
concrete
• The fraction of transmission that the barrier
requires to reduce the radiation intensity to
an acceptable value
• Unitless
1.3 m
secondary
barrier
primary
barrier
isocenter
primary
barrier
linac rotation plane
Barrier Transmission B
Pd2
B=
•m
WUT
Washington
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Design Exposure Rate P
Workload
• The dose rate allowed by regulatory agencies
• Typically weekly
• The amount the linear accelerator is
operating, defined at 1 m (typically isocenter)
• Nominally equivalent to the total weekly dose
delivered
• Units: Gy week-1
– Some regulations have “instantaneous” dose rates
as well
• Units = Sv week-1
B=
Pd2
•m
WUT
B=
Pd2
•m
WUT
Washington
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Use Factor U
Occupancy Factor T
• The fraction of time that a linear accelerator
is pointing towards the barrier
• Relevant only for “primary” barriers (take
as 1 for secondary barriers)
• Unitless
• Fraction of time a location is occupied by a
human
• Typically only used for “uncontrolled areas”
• Most relevant to locations such as hallways,
staircases, etc.
• Unitless
Pd2
B=
•m
WUT
Pd2
B=
•m
WUT
Washington
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Distance d
Secondary Barriers
• Typically distance measured from isocenter
to point of calculation
• Divided into leakage and scatter
• Leakage = head leakage
– Location from which distance is measured
depends on source of radiation
– Typical specification is that leakage is 0.1% of
primary beam intensity
– Distances measured from linac head (use closest
approach for d)
• Very important component (squared)
• When outside barrier, distance overestimates
dose falloff
– barrier acts as closer source
– dose falloff is faster than using d
– conservative estimate
• Units: meters
B=
2
Pd
•m
WUT
• Scatter
– Patient
– Wall/barrier
Washington
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Leakage
POI
secondary
barrier
Scatter – Patient
B=
2
1000 Pd sec
B=
WT
reflection
coefficient
isocenter
primary
barrier
field size (cm2)
at scatterer
scatterer-POI target-scattering
distance
surface (1m)
secondary
barrier
dsec
primary
barrier
P
2
2 400
d sec
d sca
aWT
F
dsec
isocenter
primary
barrier
linac rotation plane
primary
barrier
linac rotation plane
Washington
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Scatter - Patient
Scatter - Barrier
Scattering barrier-POI
distance
target-scattering
barrier distance
• a = ratio of scattered radiation at 1m from
the patient to the primary radiation at 1m
from the target
• F = field size at patient
• Scattered beam has different energy than
primary
B=
reflection
coefficient
2
2
Pd sec
d sca
αAWTU
field size (cm2)
at scattering surface
secondary
barrier
dsec
– Compton scattering
primary
barrier
isocenter
dsca
linac rotation plane
primary
barrier
Washington
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Scatter - Barrier
Parameters
• T Occupancy factor
– occupational = 1
– uncontrolled
» 1.0 = full occupancy, offices, shops, labs
» ¼ = partial occupancy, corridors, rest rooms
» 1/16 = occasional occupancy, waiting rooms,
toilets, stairways, elevators
• α = reflection coefficient of scattering barrier
– barrier material
– beam energy
– scattering angle
• A = field size at scattering barrier (cm2)
• Scattered beam has different energy than
primary
– Compton scattering
B=
• U Use factor (primary)
– 0.25 walls, ceiling
– 1.0 floor
• W Workload (photon treatments only)
2
2
Pd sec
d sca
αAWTU
– LowLow-energy 350 Gy week-1
– HighHigh-energy 250 Gy week-1
Washington
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Parameters
Parameters
• a (scatter fractions)
• P (occupational)
– Scattering angle and beam energy
– Taylor and Rodgers (1999)
– Maximum permissible dose rate
– 0.05 Sv per year
– ALARA = 0.1 mSv per year
• P (uncontrolled)
– 1.0 mSv per year
– 0.02 mSv in any one hour (cannot irradiate annual
limit too quickly!)
– This is satisfied if either:
» Dose to an occupyable space (T=1) at max
exposure <0.5 mSv in one year and 0.02 mSv
in one hour (T = 1)
» Dose to most irradiated person limited to 1.0
mSv in one year (implies occupancy factor)
– Other statestate-defined limitations!
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Parameters
6 MV
10 MV
18 MV
24 MV
10
1.04 x10-2
1.66 x10-2
1.42 x10-2
1.78 x10-2
20
6.73 x10-3
5.79 x10-3
5.39 x10-3
6.32 x10-3
30
2.77 x10-3
3.18 x10-3
2.53 x10-3
2.74 x10-3
45
1.39 x10-3
1.35 x10-3
8.64 x10-4
8.30 x10-4
60
8.24 x10-4
7.46 x10-4
4.24 x10-4
3.86 x10-4
90
4.26 x10-4
3.81 x10-4
1.89 x10-4
1.74 x10-4
135
3.00 x10-4
3.02 x10-4
1.24 x10-4
1.20 x10-4
150
2.87 x10-4
2.74 x10-4
1.20 x10-4
1.13 x10-4
Scattering fraction
decreases with
increasing beam
energy
Scattering fraction decreases
with increasing scatter angle
Table II: Scatter fractions (a) at 1 m from a human-sized phantom, target-to-phantom distance of 1 m and
field size of 400 cm2. Data from McGinley(McGinley 2002) and Taylor and Rogers(Taylor and Rodgers
Washington
1999).
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Barrier Transmission to Thickness
• NCRP 49 and 51
• Can use TVLs from NCRP 51
• Scattered mean beam energy (MeV
(MeV))
Beam
E
(MV)
Angle (deg)
Scatter Angle (deg)
0
10
20
30
40
50
70
90
6
1.6
1.4
1.2
0.9
0.7
0.5
0.4
0.2
10
2.7
2.0
1.3
1.0
0.7
0.5
0.4
0.2
18
5.0
3.2
2.1
1.3
0.9
0.6
0.4
0.3
24
5.6
3.9
2.7
1.7
1.1
0.8
0.5
0.3
Scattered beam
energy decreases
with increasing
scattering angle
Incident
Electron
Energy
(MeV)
MeV)
6
18
24
Table III: Mean energy of patient-scattered radiation as a function of scattering angle and
megavoltage beam energy. Data from McGinley(McGinley 2002) and Taylor and Rogers(Taylor
and Rodgers 1999).
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Material
TVL1 (m)
first tvl
TVLe (m)
subsequent
tvls
concrete
steel
lead
concrete
steel
lead
concrete
steel
lead
0.35
0.099
0.055
0.47
0.108
0.51
0.109
-
0.035
0.099
0.057
0.43
0.108
0.46
0.109
-
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Example Calculations
•
•
•
•
•
•
S
d
IMRT Modifications
Primary shield, occupational (T=1), ordinary concrete
20 MV, 4.4 m away, lateral barrier (U=1/4)
W = 500 Gy week-1 at isocenter
P = 0.0001 Sv week-1 (ALARA)
B = 1.55 x 10-5
Three ways to calculate (S = shield thickness):
– NCRP 51 curve E.8 S = 2.21 m
– NCRP 51 TVL table
» T1 = 0.48m, Te = 0.44 m
» S = 2.16m
– Varian has TVL data (0.457 m)
» S = 2.20m
B=
Pd 2
WUT
• All methods have very similar results!
– Total TD
– Total MUs
W = TD = Ec • MU ≅ MU
E ≡ TD MU
• Ec = conventional radiation therapy delivery efficiency
factor, assumed to be equal to 1
• The workload value is the same for primary and
secondary barrier calculations
Washington
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
IMRT
Room Shielding Formalism
IMRT
Room Shielding Formalism
• Depending on the complexity of the target volume
and the intensity modulation device
Ec >> E IMRT MU c << MU IMRT
Conventional
efficiency
• Workload is either
IMRT
efficiency
• Because EIMRT<< 1
– decouple workloads for primary, leakage, and scatter
shielding calculations
• Occupancy factor (T) remains unchanged
• Weekly design dose rate (P) remains unchanged
• Primary, leakage, and scatter components still
separated
• For workload estimates, the
t total delivered weekly
monitor units and the total delivered weekly dose are
decoupled
• Look at
– MLC IMRT
– sequential tomotherapy delivery
Washington
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Primary Barrier FormalismFormalism-MLC
• W = TD
• U approximately same as conventional
Primary Barrier Formalism
-Sequential Tomotherapy
• W = TD
• Average number of indexes (I) used to treat patients
• U has well defined, broad distribution in angle
– Define as product of
– fraction of gantry angle (κ) that the beam is incident on the
primary barrier
– λ is the primary barrier leakage correction factor
κ = θ i θ t WtUt =TD Iκλ
• The product Iκλ may be very close to conventional
therapy use factor
Washington
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Secondary Barrier Formalism - Leakage
• W = MUs
• Workload increases by >2 relative to
conventional
• There may be an increase in secondary
barrier thickness due to the increased
leakage component
Secondary Barrier Formalism - Scatter
• W = TD
Washington
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Experimental Verification
Experimental Verification
• DMLC delivery 14 cm wide, 40 cm long portal was used,
with 2, 4, 6, 8, 10, 12, and 14 cm wide sliding windows
• For tomotherapy a 3.7 x 20 cm2 open field (at isocenter)
was used
• 300 cm3, 6 atmosphere air-filled pressurized ionization
chamber located at 30 cm beyond the primary barrier
Center of the Field
Center of the Field
Center of the Field
Center of the Field
Dose Fraction = 0.000
Dose Fraction = 0.222
Dose Fraction = 0.778
Dose Fraction = 1.000
Washington
Varian Clinac 2300C/D
Measurement Point
30cm from the wall
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Fraction Dose of Open Field
4 cm Window Example
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Results - DMLC
0.7
Window
Width
14x40 cm2
2
4
6
8
10
12
14
4, 10x10 cm2
0.6
0.5
0.4
0.3
0.2
0.1
0.0
0.0
0.2
0.4
0.6
Fraction MUs
0.8
1.0
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
EIMRT
Beyond
Shield
0.194
0.264
0.343
0.394
0.448
0.492
0.521
0.326
E = efficiency
EIMRT
Isocenter
Ratio
±0.04
0.169
0.259
0.332
0.393
0.444
0.486
0.524
0.323
1.15
1.02
1.03
1.00
1.01
1.01
0.99
1.01
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Results - Tomotherapy
Lamda
1.2
1.7
1.0
1.6
1.5
0.8
1.4
0.6
λ
Relative Dose Rate
Results - Tomotherapy
1.3
0.4
1.2
0.2
1.1
0.0
-15
-10
-5
0
5
10
1.0
15
Incidence Gantry Angle (deg)
Normalized Dose Rate
Conservative
estimate!
50
100 150 200 250 300 350 400
Arc Length (deg)
Washington
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Efficiencies (1/E)
Normalized Output Factors Inside and Outside the Room
Isocenter and Inside Room
1.2
0
Conventional
Beam
Energy
1.0
Beam Intensity Modulated
unwedged
(MU/cGy)
wedged
(MU/cGy)
MLC
(MU/cGy)
6 MV
1.2
2.4
3.4
Serial
Tomothera
py
(MU/cGy)
9.7
18 MV
1.0
1.5
2.8
8.1
25 MV
1.0
1.5
2.8
8.1
0.8
Outside Room
0.6
Outside Room
Inside Room
Isocenter
0.4
0.2
0.0
0
5
10
15
20
25
30
35
Ratio of MUs to tumor dose (1/E) for typical conventional and IMRT
deliveries. Data from Followill et al.(Followill, Geis and Boyer 1997)
40
Field Size (cm)
Washington
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Results - Shielding
Neutrons
• Neutron generation is
function of MUs
• IMRT
• Data from Rodgers (2001)
• r = percent IMRT
r (%)
1/E
0
– MLC: efficiency
Additional
TVLs
0 TVL
50
4
0.40 TVL
50
10
0.74 TVL
100
4
0.60 TVL
100
10
1.0 TVL
Field Size
(cm x cm)
Gantry
angle (deg)
Neutron
Rem
(nSv/MU)
Photon Dose
(nGy/MU)
0x0
0
1.12
0.77
20x20
0
0.98
0.70
40x40
0
0.78
0.73
90
0.60
0.80
270
0.83
0.80
(7 angles)
0.85
0.54
Abdomen
IMRT
0
1.14
0.48
Breast
IMRT
0
1.02
Neutron equivalent dose and photon dose 40x40
outside the treatment room door for some
tested beam configurations. Data courtesy of 40x40
Boyer (private communication).
Abdomen
IMRT
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
0.58
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Conclusions
Conclusions
• MLC
– Significant Field Size Factor (0.35 - 0.80 relative to
40x40 cm2)
– DMLC
» 40x40 cm2 overestimates IMRT conformal therapy
shielding requirements
» Using W=TD - accurate within 10% to 2 cm window
– Primary Barriers - Comparable to conventional shielding
– Secondary Barriers - Possible increase due to leakage
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
• Sequential Tomotherapy
» MU component (λ
(λ) from 15% to 60% compared to geometric
estimate TD * κ of single index workload.
» 6 indexes, I κ λ ranges from 0.52 for 180°
180° to 0.30 for 360 °
» Very similar to Use factors of conventional!
– Primary Barrier
» Comparable to conventional shielding due to reduced
number of treated patients and I κ λ product
–
Secondary Barriers
» Increased due to leakage
» MIR experience ≅ 3,000 MU/fraction
» 300,000 MU/week for 20 patients
» Similar to high energy W
WtUt = TDIκλ
Washington
WASHINGTON•
WASHINGTON•UNIVERSITY•
UNIVERSITY•IN•
IN•ST•
ST•LOUIS
Download