MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms. HST.583: Functional Magnetic Resonance Imaging: Data Acquisition and Analysis, Fall 2008 Harvard-MIT Division of Health Sciences and Technology Course Director: Dr. Randy Gollub. MR physics and safety for fMRI Lawrence L. Wald, Ph.D. Massachusetts General Hospital Athinoula A. Martinos Center Wald, fMRI MR Physics Outline: Wed. Sept 24 (LLW): MR signal, Gradient and spin echo Basic image contrast Mon. Sept 29 (LLW): Encoding the image Wed Oct 1 (LLW): Fast imaging for fMRI, artifacts fMRI BOLD Wald, fMRI MR Physics What is NMR? NUCLEAR MAGNETIC RESONANCE A magnet, a glass of water, and a radio wave source and detector…. Wald, fMRI MR Physics What is NMR? Nuclear magnetism M = Wald, fMRI MR Physics E B ΔE= hυ protons (N↑ – N↓)/NTOT = 1 – exp(-ΔE/kT) ≈ 10-4 Earth’s Field N E W S Wald, fMRI MR Physics compass Compass needles Earth’s Field υ z Main Field Bo North N E W y S x Freq = γ B Wald, fMRI MR Physics 42.58 MHz/T Gyroscopic motion Main Field Bo z North • Proton has magnetic moment M y • Proton has spin (angular momentum) >>gyroscopic precession x Larmor precession freq. = 42.58 MHz/T Wald, fMRI MR Physics υ = γ Bo EXCITATION : Displacing the spins from Equilibrium (North) Problem: It must be moving for us to detect it. Solution: knock out of equilibrium so it oscillates How? 1) Tilt the magnet or compass suddenly 2) Drive the magnetization (compass needle) with a periodic magnetic field Wald, fMRI MR Physics Excitation: Resonance Why does only one frequency efficiently tip protons? Resonant driving force. It’s like pushing a child on a swing in time with the natural oscillating frequency. Wald, fMRI MR Physics z is "longitudinal" direction x-y is "transverse" plane RF Field (B1) applies a torque to the spins… Static Field, B0 z y Mo Applied RF Field (B1) x The RF pulse rotates Mo the about applied field Wald, fMRI MR Physics "Exciting" the Magnetization: tip angle z Static Field, B0 45° Wald, fMRI MR Physics z y y x x 90° Detecting the Magnetization: Faraday’s Law z A moving bar magnet induces a Voltage in a coil of wire. (a generator…) 90° y x υo The RF coil design is the #1 determinant of the system SNR V(t)= -dΦ/dt Φ = n Bspins A Wald, fMRI MR Physics Detecting the NMR: the noise Noise comes from electrical losses in the resistance of the coil or electrical losses in the tissue. z 90° For a resistor: Pnoise = 4kTRB y x υo • V(t) Wald, fMRI MR Physics • • Noise is white. >>Noise power α bandwidth Noise is spatially uniform. R is dominated by the tissue. >> big coil is bad. The NMR Signal RF time Voltage (Signal) time υ υo Bo z z 90° z y Mo y y x x Wald, fMRI MR Physics υo V(t) x Signal to Noise Ratio in MRI • Most important piece of hardware is the RF coil. • SNR α voxel volume (# of spins) • SNR α SQRT( total time of data collection) • SNR depends on the amount of signal you throw away to better visualize the brain (gain image contrast) Wald, fMRI MR Physics Physical Foundations of MRI NMR: 60 year old phenomena that generates the signal from water that we detect. MRI: using NMR signal to generate an image Three magnetic fields (generated by 3 coils) 1) static magnetic field Bo 2) RF field that excites the spins B1 3) gradient fields that encode spatial info Gx, Gy, Gz Wald, fMRI MR Physics Three Steps in MR: 0) Equilibrium (magnetization points along Bo) 1) RF Excitation (tip magn. away from equil.) 2) Precession induces signal, dephasing (timescale = T2, T2*). 3) Return to equilibrium (timescale = T1). Wald, fMRI MR Physics Magnetization vector during MR RF encode time Voltage (Signal) Mz Mxy Wald, fMRI MR Physics Three places in process to make a measurement (image) 0) Equilibrium (magnetization points along Bo) 1) RF Excitation (tip magn. away from equil.) 2) Precession induces signal, allow to dephase for time TE. 3) Return to equilibrium (timescale =T1). proton density weighting T2 or T2* weighting T1 Weighting Wald, fMRI MR Physics Contrast in MRI: proton density Form image immediately after excitation (creation of signal). Tissue with more protons per cc give more signal and is thus brighter on the image. No chance to dephase, thus no differences due to different tissue T2 values. Magnetization starts fully relaxed (full Mz), thus no T1 weighting. Wald, fMRI MR Physics T2*-Dephasing Wait time TE after excitation before measuring M. Shorter T2* spins have dephased z z z y y y vector sum x initially Wald, fMRI MR Physics x at t= TE x T2* Dephasing Just the tips of the vectors… Wald, fMRI MR Physics T2* decay graphs Transverse Magnetization 1.0 T2* = 200 0.8 Tissue #1 0.6 T2* = 60 0.4 0.2 0.0 0 Wald, fMRI MR Physics Tissue #2 20 40 60 Time (milliseconds) 80 100 T2* Weighting Phantoms with four different T2* decay rates... There is no contrast difference immediately after excitation, must wait (but not too long!). Choose TE for max. inten. difference. Wald, fMRI MR Physics Gradient Echo (T2* contrast) Dephasing is entirely from a spatial difference in the applied static fields. z 90° y t=0 x B o + Gx x z y x t=T Red arrows processes faster due to its higher local field Wald, fMRI MR Physics x Gradient Echo (T2* contrast) Dephasing is entirely from a spatial difference in the applied static fields. z 90° y t=0 x z z y t=T z Wald, fMRI MR Physics x x x y y t=T B o + Gx x x t = 2T B o + Gx x x Gradient Echo RF excitation t Gx t S t Boring! Wald, fMRI MR Physics 7T 32ch MGH array 2D FLASH, TR/TE=3 500/30 0.22 x 0.22 x 1mm (48nl) 8min acq Wald, RSNA 2007 Courtesy of Dr. Christopher J. Wiggins. Used with permission. A.A. Martinos Center, MGH Radiology 7T 32ch MGH array Courtesy of Dr. Christopher J. Wiggins. Used with permission. G. Wiggins, C. Wiggins, Martinos Center MGH Wald, RSNA 2007 2D FLASH, TR/TE=3 500/30 0.22 x 0.22 x 1mm (48nl) 8min acq A.A. Martinos Center, MGH Radiology 7 Tesla 230um 2D FLASH 3 0.23 x 0.23, x 1.5mm 8min acq Courtesy of Dr. Christopher J. Wiggins. Wald, Munich, 2008 Used with permission. 7 Tesla 230um 2D FLASH, 3 0.23 x 0.23 x 1.5mm 8min acq Wald, Munich, 2008 Courtesy of Dr. Christopher J. Wiggins. Used with permission. 7 T, 32ch 200um x 200um x 1mm an 2D T2* weighted 200um x 200um x 1mm (1024x1024 matrix) Wald, Munich, 2008 “U-fiber” ?? Courtesy of Dr. Christopher J. Wiggins. Used with permission. Spin Echo (T2 contrast) Some dephasing can be refocused because its due to static fields. 180° z 90° y t=0 x z t=T Echo! z z y y y x x x t = T (+) Blue arrows precesses faster due to local field inhomogeneity than red arrow Wald, fMRI MR Physics t = 2T Spin Echo 180° pulse only helps cancel static inhomogeneity The “runners” can have static speed distribution. If a runner trips, he will not make it back in phase with the others. Wald, fMRI MR Physics NMR Signal T2 weighed spin echo image gray white Time to Echo , TE (ms) Wald, fMRI MR Physics Other contrast for MRI In brain: (gray/white/CSF/fat) Proton density differ ~ 20% T1 relaxation differ ~ 2000% How to exploit for imaging? Vary repetition rate - TR Wald, fMRI MR Physics T1 weighting in MRI (w/ 90o excite) TR RF encode encode encode Voltage (Signal) Mz grey matter (long T1) Wald, fMRI MR Physics white matter (short T1) time T1-Weighting 1.0 white matter T1 = 600 Signal 0.8 grey matter T1 = 1000 0.6 CSF T1 = 3000 0.4 0.2 0.0 0 1000 2000 TR (milliseconds) Wald, fMRI MR Physics 3000 T1 weighting in MRI (w/ 30o excite) TR RF encode encode encode Voltage (Signal) Mz white matter (short T1) Wald, fMRI MR Physics time Image contrast summary: TR, TE Long Proton Density T2 TR Short T1 poor! Short Wald, fMRI MR Physics Long TE Source of T1 and T2 contrast in brain: Myelin content Layer 1: no cell bodies, moderate myelination Image removed due to copyright restrictions. Diagram showing the arrangement of nerve cells and fibers in layers and sublayers parallel to the surface in a vertical section through the human striate area or visual cortical center. Determine functional Layer 4b: thick region with myelination (line on boundaries based of Gennari) MR strucure alone… White matter: heavy myelination Myelin differences are the primary source of T1 and T2 contrast of gray/white matter. Nissel stain: Weigert stain: fMRI MR Physics cellWald, bodies fibers Cortical layers in Monkey at 7T Intensity along line perpendicular To V1 MPRAGE 250um x 250um x 750um (4 hours) Wald, fMRI MR Physics