UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/312_02 STANDARD OPERATING PROCEDURE ® HANNA BENCH-TOP PH METER HI 8521 Valid for: 2 years from approval Table of Contents Page 1. 2. 3. 4. 5. 6. 7. 8. 9. Scope Objective Definitions Responsibility Procedure 5.1. Diagram of Front Panel 5.2. Diagram of Rear Panel 5.3. Installation 5.4. Operation 5.5. Calibration 5.6. Maintenance 5.7.Troubleshooting 5.8.Average time taken to carry out Procedure 5.9. Flow charts Precautions References Appendices Revision History Page 1 of 14 2 2 2 3 3 3 3 4 4 5 6 7 8 8 13 13 13 14 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/312_02 STANDARD OPERATING PROCEDURE ® HANNA BENCH-TOP PH METER HI 8521 Valid for: 2 years from approval 1. Scope This Standard Operating Procedure (SOP) applies to the staff and students using the Hanna® Bench-Top pH Meter HI 8521 in the laboratories of the Pharmacy Department, University of Malta. 2. Objective To describe the procedure for installation, operation, calibration, maintenance and troubleshooting of the Hanna® Bench-Top pH Meter HI 8521. 3. Definitions 3.1. ↑ oC and ↓ oC Keys: To set the temperature for Manual Temperature Compensation. 3.2. Automatic Temperature Compensation (ATC): A feature in which the temperature probe is able to detect the temperature of the solution in which it is immersed in. 3.3. Bayonet Neill-Concelman (BNC) Electrode connector (1): A cable connector used to connect the pH or an ORP electrode to the meter. 3.4. CAL Key: To enter the calibration mode. 3.5. CON Key: To confirm the calibration data. 3.6. Display: Digitally displays the pH, temperature and millivolt (mV) results. 3.7. DC Power Socket (5): To connect the meter with the mains power supply. 3.8. Manual Temperature Compensation (MTC): A feature in which the temperature of the sample solution is manually inputted when this is already known. 3.9. On/Off Key: To turn the meter on or off. 3.10. RANGE Key: To select between pH, millivolt or temperature results. Page 2 of 14 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/312_02 STANDARD OPERATING PROCEDURE ® HANNA BENCH-TOP PH METER HI 8521 Valid for: 2 years from approval 3.11. Reference Jack Socket (2): To connect a reference electrode to the meter. 3.12. Reset Button (3): To be pressed whenever erroneous messages are displayed due to strong electrical interference or when the meter is disconnected from its power supply before being switched off properly. 3.13. Temperature Probe Socket (4): The socket at which the temperature probe is connected to the meter. 4. Responsibility 4.1. The members of the Department of Pharmacy (staff and students) are responsible for following this SOP. 4.2. The designated Laboratory Officer or Laboratory Assistant is responsible for ensuring that this SOP is followed. 5. Procedure 5.1. Diagram of Front Panel 5.2. Diagram of Rear Panel Page 3 of 14 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/312_02 STANDARD OPERATING PROCEDURE ® HANNA BENCH-TOP PH METER HI 8521 5.3. Installation 5.3.1. 5.3.2. 5.4. Valid for: 2 years from approval Connect the power cable to the DC Power Socket and to the mains supply. Connect the pH electrode and the temperature probe to the BNC and temperature sockets respectively. Operation 5.4.1. Measuring pH 5.4.1.1. 5.4.1.2. 5.4.1.3. 5.4.1.4. Before use, ensure that the meter is being calibrated, to obtain more accurate measurements (see section 5.5). Press the On/Off Key to switch on the meter, which will automatically enter into pH mode. Submerge the pH electrode and the temperature probe into the sample to be tested and keep them close together, if automatic temperature compensation is desired. Disconnect the temperature probe from the meter if manual temperature compensation is desired whenever the temperature of the sample to be tested is already known. 5.4.1.4.1. Press the RANGE Key until the display changes to temperature mode. 5.4.1.4.2. Adjust the displayed temperature by using the Up and Down Keys respectively. 5.4.1.4.3. Press the RANGE Key again until the display changes back to pH mode. 5.4.1.4.4. Submerge pH electrode into the sample to be tested. 5.4.1.5. 5.4.1.6. Shake the sample briefly and wait for reading to stabilise before taking the reading. Read off the displayed stable pH reading compensated for temperature. Page 4 of 14 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/312_02 STANDARD OPERATING PROCEDURE ® HANNA BENCH-TOP PH METER HI 8521 5.4.1.7. 5.4.2. Rinse off the pH electrode with some solution from the next sample if more than one sample is to be successively tested. Measuring Temperature 5.4.2.1. 5.4.2.2. 5.4.2.3. 5.4.2.4. 5.4.2.5. 5.5. Valid for: 2 years from approval Press the On/Off Key to switch on the meter. Press the RANGE Key until the display changes to temperature mode. Dip the temperature probe into the sample. Allow 1-2 minutes for the reading to stabilise. Read off the displayed stable temperature reading. Calibration 5.5.1. 5.5.2. 5.5.3. 5.5.4. 5.5.5. 5.5.6. 5.5.7. 5.5.8. 5.5.9. 5.5.10. 5.5.11. 5.5.12. 5.5.13. 5.5.14. 5.5.15. Pour small quantities of pH 7.01 and pH 4.01 or pH 10.01 buffer solutions into two separate clean beakers and use another two separate beakers for rinsing. Press the On/Off Key to switch on the meter. Rinse the pH electrode with some pH 7.01 buffer solution. Immerse the pH electrode into the pH 7.01 buffer solution and stir gently. Ensure that the pH electrode is kept submerged approximately 4cm into the solution. Ensure that the temperature probe is located as close to the pH electrode as possible. Press the CAL Key. Wait until the pH symbol, next to the pH buffer value, stops flashing to indicate that reading is stable. Press the CON Key to confirm the calibration. If [E 4] is flashing, ensure that the correct buffer solution is being used. Wait until [E 5] is displayed to indicate that the electrode is still in the pH 7.01 buffer, but expecting a pH 4.01 or 10.01 buffer. Rinse the pH electrode with distilled water followed by the second buffer solution to be used. Immerse the pH electrode into the second buffer solution (pH 4.01 or pH 10.01) and stir gently. Wait until the pH symbol, next to the pH buffer value, stops flashing to indicate that the reading is stable. Press the CON Key to confirm the calibration. Page 5 of 14 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/312_02 STANDARD OPERATING PROCEDURE ® HANNA BENCH-TOP PH METER HI 8521 Valid for: 2 years from approval 5.5.16. If [E 4] is flashing, ensure that the correct buffer solution is being used. 5.5.17. Wait until the meter returns to the normal operating mode to ensure that the calibration process was successful. 5.6. Maintenance 5.6.1. 5.6.2. 5.6.3. 5.6.4. 5.6.5. 5.6.6. 5.6.7. 5.6.8. 5.6.9. Inspect the cables of the electrodes for any signs of broken insulation. Inspect the pH electrode for any cracks in its stem or bulb. Rinse off any salt desposits with water. Soak the pH electrode in Hanna HI 7061 or HI 8061 General Cleaning Solution for approximately ½ hour to perform general cleaning. Soak the pH electrode in Hanna HI 7073 or HI 8073 Protein Cleaning Solution for 15 minutes if protein deposits are present on the membrane and/or junction of the electrode. Soak the pH electrode in Hanna HI 7074 or HI 8074 Inorganic Cleaning Solution for 15 minutes if inorganic deposits are present on the membrane and/or junction of the electrode. Rinse the pH electrode with Hanna HI 7077 or HI 8077 Oil and Fat Cleaning Solution if oil films or deposits are present on the membrane and/or junction of the electrode. Rinse the pH electrode thoroughly with distilled water after performing any of these cleaning procedures. Refill the reference chamber with fresh electrolyte and soak it in HI 70300 or HI 80300 Storage Solution for at least 1 hour before taking any measurements. Page 6 of 14 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/312_02 STANDARD OPERATING PROCEDURE ® HANNA BENCH-TOP PH METER HI 8521 5.7. Valid for: 2 years from approval Trouble Shooting Fault Possible Cause/s Solution Meter fails to calibrate or gives erroneous results pH electrode is defective Replace the electrode and repeat the calibration - Meter is slow in responding or gives erroneous readouts Electrode not working or diaphragm is clogged Meter does not accept 2nd buffer solution pH electrode defective or bad buffer solution - - Reading is drifting Temperature not stable - Display shows [E 1] pH scale is out of range - Temperature scale is out of range Display shows [E 2] - Display shows [E 3] Display shows [E 4] Display shows [E 5] - mV scale is out of range - Erroneous buffer solution used for first part of calibration Defective electrode - Erroneous buffer solution used for second part of calibration Defective electrode - Page 7 of 14 - - Leave electrode in storage solution after performing diaphragm cleaning If problem is not solved, replace electrode and recalibrate pH scale Clean the electrode Replace electrode and re-calibrate Use fresh buffer solution Wait longer since thermal equilibrium is still not reached Ensure pH electrode is connected Ensure pH of sample is between 0-14 Check the state of the electrode and the level of its filling electrolyte Ensure T probe is properly connected Ensure that T of sample lies between -10oC and 100oC Ensure that ORP probe is properly connected Ensure mV is within +/- 1999 range Ensure that buffer solution used is of pH 7 and replace if necessary Replace electrode and re-calibrate pH scale Ensure that correct pH buffer solution is being used (pH 4.01 or pH 10.01) Check electrode and replace if necessary UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/312_02 STANDARD OPERATING PROCEDURE ® HANNA BENCH-TOP PH METER HI 8521 5.8. Average time taken to carry out Procedure Procedure Section Average Time (mins) N/A Installation Measuring pH with Automatic 1 (for one pH reading) Temperature Compensation Measuring pH with Manual 2 (for one pH reading) Temperature Compensation 2 (for one T reading) Measuring Temperature 3 ( for two-point calibration) Calibration N/A Maintenance N/A Troubleshooting 5.9. Valid for: 2 years from approval Flow Charts 5.9.1. Installation Start Connect power cable to DC power socket and to mains supply Connect pH electrode and T probe to BNC and T sockets respectively End Page 8 of 14 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/312_02 STANDARD OPERATING PROCEDURE Valid for: 2 years from approval ® HANNA BENCH-TOP PH METER HI 8521 5.9.2. Operation Start 2 No No Measuring pH Measuring T Yes Yes No Press On/Off key to switch on meter Perform calibration before use (see section 5.8.3) Meter being before calibrated No Yes Display in T mode Press On/Off Key to switch on meter Yes No No Use ATC feature Yes Yes Submerge pH electrode and T probe into sample and keep them close together Allow 1-2 minutes for reading to stabilise Disconnect T probe from meter Press RANGE Key until display changes to T mode 1 Shake sample briefly and wait for reading to stabilise Adjust displayed T by using the Up and Down keys respectively Read off stable pH reading Press RANGE Key again until display changes back to pH mode No Test another sample Submerge pH electrode into sample to be tested Yes Rinse off electrode with next sample before further use Dip T probe MTC 1 2 Page 9 of 14 Read off displayed stable T reading End Press RANGE Key until display changes to T mode UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/312_02 STANDARD OPERATING PROCEDURE ® HANNA BENCH-TOP PH METER HI 8521 5.9.3. Valid for: 2 years from approval Calibration Start 1 Pour small quantities of pH 7.01 and pH 4.01 or pH 10.01 buffer solutions into 2 separate beakers Use another 2 separate beakers for rinsing Press On/Off Key to switch on meter Rinse pH electrode with some pH 7.01 buffer solution Immerse pH electrode into pH 7.01 buffer and stir gently No Submerged 4cm into solution Submerge accordingly Yes Place T probe as close to pH electrode as possible No Meter in pH mode Press RANGE Key until display changes to pH mode Yes Press CAL Key No No [E 4] seen flashing [pH] symbol stopped flashing Yes Yes No Using correct buffer solution Press CON Key Use appropriate buffer solution Wait until [E 5] is displayed Yes 1 Replace electrode Page 10 of 14 2 Wait until it stops flashing UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/312_02 STANDARD OPERATING PROCEDURE ® HANNA BENCH-TOP PH METER HI 8521 Valid for: 2 years from approval 2 Rinse pH electrode with distilled water Rinse pH electrode with second buffer solution Immerse pH electrode into second buffer and stir gently No No [E 4] seen flashing [pH] symbol stopped flashing Yes Yes No Using correct buffer solution Press CON Key Use appropriate buffer solution Yes 1 Wait until meter returns to normal operating mode before further use Replace electrode End 1 Page 11 of 14 Wait until it stops flashing UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/312_02 STANDARD OPERATING PROCEDURE ® HANNA BENCH-TOP PH METER HI 8521 5.9.4. Valid for: 2 years from approval Maintenance Start Inspect cables of electrodes for any signs of broken insulation Inspect pH electrode for any cracks in stem or bulb Rinse off any salt deposits with water No No No Protein deposits present General cleaning Oil films / deposits present Yes Yes Yes Soak pH electrode in Hanna HI 7062 or HI 8061 General Cleaning Solution for ½ hour Soak pH electrode in Hanna HI 7073 or HI 8073 Protein Cleaning Solution for 15 minutes Soak pH electrode in Hanna HI 7074 or HI 8074 Inorganic Cleaning Solution for 15 minutes 1 1 1 No Inorganic deposits present Yes Soak pH electrode in Hanna HI 7074 or HI 8074 Inorganic Cleaning Solution for 15 minutes 1 Rinse pH electrode thoroughly with distilled water Refill reference chamber with fresh electrolyte Soak in HI 70300 or HI 80300 Storage Solution for at least 1 hr before use 2 End Page 12 of 14 2 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/312_02 STANDARD OPERATING PROCEDURE ® HANNA BENCH-TOP PH METER HI 8521 Valid for: 2 years from approval 6. Precautions 6.1. 6.2. 6.3. 6.4. 6.5. 6.6. 6.7. 6.8. 6.9. Always keep the pH electrode wet and rinse it thoroughly with the sample to be measured before use. In order to measure the pH accurately, the temperature factor must always be taken into consideration. If the sample temperature is quite different from the temperature at which the pH electrode is kept, allow a few minutes for thermal equilibrium to take place. When performing a two-point calibration, use pH 7.01 and pH 4.01 when measuring acidic samples and pH 7.01 and pH 10.01 for alkaline measurements. This provides more accurate readings to be taken. Do not be alarmed when salt deposits are present on the electrode since these normally form and can easily be rinsed off with water. If any air bubbles form inside the glass bulb, “shake it down” firmly to remove them. Always store the electrode in its appropriate storage solution and never store it in distilled or deionised water. Avoid high humidity environments and static discharges since these are detrimental for a stable pH reading. Perform a calibration: 6.9.1. 6.9.2. 6.9.3. 6.9.4. 6.9.5. 6.9.6. When pH electrode or temperature probe is replaced When reference electrode is changed At least once a month After testing aggressive chemicals When greater accuracy is required After any cleaning procedure 7. References Hanna® Instruments. Instruction Manual HI 8417 – HI 8519 – HI 8520 – HI 8521 Microprocessor Bench-Top pH Meters;1998. 8. Appendices N/A Page 13 of 14 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/312_02 STANDARD OPERATING PROCEDURE ® HANNA BENCH-TOP PH METER HI 8521 Valid for: 2 years from approval 9. Revision History Version Number 01 02 Page 14 of 14 Amendments/ Reasons for change Initial Release Inclusion of subheading titles of Procedure section in Table of Contents