NON-FLAME AAS 3 3.1 Limitations of flame AAS In conventional atomic absorption spectroscopy (FAAS) a flame is used as the means of atomisation. The advantages of this method of atomisation are: simplicity speed most metals atomise in readily available flames sensitivity suitable for general analyses (mg/L) Flame AAS was developed in the 1950s, and at this time, the environmental significance of ug/L levels of metals such as lead and mercury, was not really appreciated. However, this would soon change, and improvements in sensitivity were an ongoing part of instrumental development (think about the changes in polarographic instrumentation). A number of factors limit the sensitivity of flame AAS. CLASS EXERCISE 3.1 What factors or processes in flame AAS could contribute to sensitivity losses? To achieve sensitivity improvements of 100 times or more (to get accurate analysis of 10 ug/L levels) from atomic absorption spectrophotometry would require a considerably different approach: a system AA system based around a flame and nebuliser will never achieve the sensitivities required. There are two very different approaches, both of which dispense with the flame: vapour generation – where the analyte is converted into a volatile form electrothermal – where electrical heating is used 3.2 Basic principles of electrothermal atomisation As we have seen, the two major obstacles to improved sensitivity in FAAS were the flame and the nebuliser. In EAAS each of these is replaced. The sample is delivered as a single aliquot, rather than a constant flow, into a sample holder, which is heated electrically. As in other forms of spectroscopy, the absorption beam passes through the sample holder to the monochromator and detector. Table 3.1 summarises the key differences between FAAS and EAAS. TABLE 3.1 Improvements in EAAS Problem in FAAS Approach in EAAS Flame No flame; heat is provided electrically Nebuliser Aliquot of sample delivered to sample cell, rather than constant flow; nebuliser and burner is replaced by workhead 3. Non-flame AAS CLASS EXERCISE 3.2 How are the problems identified in Exercise 3.1 addressed by these changes? FAAS Problem Solution in EAAS 3.3 Equipment The burner/nebuliser in FAAS instruments is removed and the electrothermal workhead put in its place. Most modern instruments allow the two units to be interchangeable, because the rest of the instrument is identical: hollow cathode lamp monochromator photomultiplier detector signal processing and readout Electrothermal-capable instruments must have computer control to automate the sample delivery, and to monitor the heating and signal measurement. The workhead The workhead contains the sample cell and provides the services for controlled heating of the sample. The sample cell is a hollow graphite tube approximately 28 mm in length by 8mm in diameter. The other common name for EAAS is graphite furnace AAS, deriving from the graphite tube in which the sample is heated. The interior of the cylinder is coated with a layer of pyrolytic graphite, which has a greater resistance to heat than normal graphite. The sample is delivered into the graphite tube through a small hole in the barrel of the tube, as seen in Figure 3.1. There are two types of graphite tubes – partition and platform. The former has raised sections one-third of the length along from each end. These act like dam walls, preventing the droplet of sample spreading out too far. The platform is a separate piece of pyrolytic graphite sitting in the centre of tube. The droplet sits in this and supposedly is vaporised more predictably, because the heating only occurs by radiation (the gas inside) than by conduction (from the walls). The tube is positioned inside the workhead so that the delivery hole is FIGURE 3.1 pointing upwards. It rests on electrodes which are connected to a low-voltage Graphite furnace high current power supply which delivers upwards of 3.5kW to the cylinder AIT 3.2 3. Non-flame AAS walls. Because graphite is only a moderate conductor of electricity, the tube heats up, providing the conditions to vaporise and atomise the sample. The metal housing around the furnace is water cooled to enable rapid restoration of the furnace to ambient temperature after atomisation. Inert gas (generally argon) is flushed through and around the tube from the ends. This has three purposes: to remove steam and smoke vaporised during the heating to prevent oxidation of the graphite assist with cooling Figure 3.2 shows a typical workhead. Ar flows through and around tube cooling water coils HCL beam quartz windows FIGURE 3.2 Workhead Autosamplers Modern electrothermal atomisation units almost always use an autosampler to place the sample in the graphite tube. This is because the volumes involved are typically 5-20 uL, and therefore manual loading of graphite tubes (even with micropipettes) would be unreliable, with precision being greatly reduced compared to that obtainable with autosamplers. These also allow multiple samples to be run without an operator in attendance. Frequently, different solutions (blank, sample, standard, matrix modifier) are used in combination for the analysis. These are kept apart in the plastic tubing by air gaps. The sampler arm then swings over to the workhead and delivers the sample to the graphite tube. 3.4 The heating process The heating of the sample (or standard) in the graphite furnace does not simply involve an instant transition to the temperatures encountered in the flame 1800-3000C. Because there is a droplet of liquid in the tube, such heating would be disastrous for the analysis. CLASS EXERCISE 3.3 What do you expect would happen to a droplet of milk inside the graphite furnace it is was suddenly exposed to 2200C? AIT 3.3 3. Non-flame AAS Therefore, the heating process must occur gradually, and involves a number of carefully programmed steps, such as those shown in Table 3.2 and Figure 3.5: 1. Drying – removal of all volatiles at temperatures slightly above the boiling point of the solvent, and for 30-60 seconds 2. Ashing – the temperature is increased to 500-1000C and the organic matter is burnt away, like in a muffle furnace, for 5-40 seconds 3. Atomisation – now that all the non-elemental matter is gone, the analyte and any other inorganic elements in the matrix are heated to the temperatures used in FAAS (1800-3000C) for 3-4 seconds; measurement of absorbance occurs here TABLE 3.2 & FIGURE 3.5 Temperature program for Cu in water Step 1 2 3 4 5 6 7 8 9 Temp. (C) 85 95 120 800 800 800 2300 2300 2300 Time (s) 5 40 10 5 1 2 1.1 2 2 Gas ON ON ON ON ON OFF OFF OFF ON Designing the temperature program Each step in the cycle of dry/ash/atomise is important and the correct time and temperature conditions must be carefully selected. Both the analyte and the matrix will affect the times and temperatures that are needed. The instrument manual will provide a standard program for all elements, which is fine for that element in pure water or dilute acid, such as that for copper in Table 3.2. This can be used as a starting point for a real sample. In most instrument manuals, there will also be sections covering selected real analyte/sample situations, and these give guidance about the type of changes to the temperature that are needed. CLASS EXERCISE 3.4 Does the nature of the analyte/matrix have any effect on the steps in the temperature program? Analyte Drying time Drying temperature Matrix No Yes Ashing time Ashing temperature Atomisation time Atomisation temperature AIT 3.4 3. Non-flame AAS These notes are intended only give some general guidelines about the effect on the temperature program of the analyte, but more importantly, the sample matrix. When setting up a method for a new sample, it is very much a matter of trial and error until the most efficient sequence of times and temperatures is determined. Drying stage This step is designed to remove all volatiles without losing analyte by over-vigorous boiling, which could cause the mixture to splatter. CLASS EXERCISE 3.5 List some factors that would influence the temperature required for drying. List some factors that would influence the time required for drying. Usually the drying stage is divided into a number of steps: increase from room temperature to boiling point of solvent remain at this temperature for a time increase up to about 50ºC above b.p. to remove last volatiles The main problem likely in the drying stage is incomplete drying which can lead to splattering of the remaining material as the temperature increases to ashing point. You will hear the fizz as the temperature goes up. CLASS EXERCISE 3.6 Describe the differences in the drying stage, compared to the standard program for lead in water, where the matrix is: (a) petrol (b) AIT milk 3.5 3. Non-flame AAS Ashing stage This step is designed to remove the non-volatile organic components of the matrix by burning them away as in a muffle furnace. CLASS EXERCISE 3.7 List some factors that would influence the temperature required for ashing. List some factors that would influence the time required for ashing. The ashing stage will normally be divided up into at least three steps: increase from the final drying temperature to the ashing temperature (normally 5-10 seconds) remain at ashing temperature for required time in the last few seconds before atomisation, the argon gas flow is turned off so that the atmosphere inside the tube is still for atomisation Chemical modifiers are used where the analyte is relatively volatile and would be lost during the ashing stage, eg lead which is volatile above 500ºC. If a modifier wasn’t used, then the ashing temperature would have to be quite low, meaning a much longer time would be required. The modifier for lead is phosphate, since lead phosphate has a much higher boiling point than other lead salts, such as chloride and nitrate. However, the modifier must not hold onto the analyte during atomisation! Some of the volatile elements are lead, cadmium, mercury (of course), antimony and arsenic. The instrument documentation will indicate whether a modifier is needed. The main problem encountered with incomplete ashing is the generation of smoke during atomisation, which causes a high background reading due to scattering. You can observe this at the start of the atomisation step as a puff of smoke coming out through the delivery hole. CLASS EXERCISE 3.8 Describe the differences in the ashing stage, compared to the standard program for copper in water, where the analyte/matrix is: (a) lead in water (b) AIT aluminium in milk 3.6 3. Non-flame AAS Atomisation stage This step is designed to drive the analyte atoms into an atomic vapour in the path of the HCL beam. CLASS EXERCISE 3.9 List some factors that would influence the time required for atomisation. List some factors that would influence the temperature required for atomisation. The argon gas is turned on after the brief atomisation step to flush the tube and avoid it turning into CO2. 3.5 Signal measurement This is much more complex than in flame AAS, because the analyte atoms are only present for a few seconds once per run, rather than being present whenever the nebuliser is in the flask. Therefore, the signal is a short-lived peak (like in a GC or HPLC), as in Figure 3.6. This means that signal capture by the computer is essential. The alternative is to wait through two minutes of drying and ashing and then make sure you are watching when the atomisation step occurs. To make matters more difficult, the steam and ash generated in the first two stages cause scattering of the beam, and an apparent absorbance which varies over time, and has nothing to do with the real analyte absorbance. In the software, this is avoided by not having the detector responding until the atomisation steps. Absorbance EAAS FAAS Time FIGURE 3.6 Signals in electrothermal and flame AAS AIT 3.7 3. Non-flame AAS 3.6 Practical aspects Checking instrument performance In flame AAS, the sensitivity actually had a numerical value: it was the concentration that would give an absorbance of 0.0044. In electrothermal AAAS, where a discrete volume of sample is analysed, concentration is only one of the factors that affects absorbance: the other is the volume. Therefore, the mass of analyte (concentration x volume) is the variable that determines the absorbance. The characteristic mass is the mass that produces the absorbance of 0.0044. This can be used as shown below to determine the expected absorbance of a solution, and also the working performance of the instrument. A exp 0.0044ms mc Eqn 3.1 where mc is the characteristic mass, most often in pg, ms the mass of analyte in the sample/standard (in the same units) and As the expected absorbance of the standard. Calculating masses in pg (10-12 g) from uL and ug/L might seem a good candidate for getting the decimal place wrong, but in fact, nothing could be easier, as shown in Equation 3.2: no unit conversion required at all! pg uL x ug / L Eqn 3.2 CLASS EXERCISE 3.10 (a) Calculate the expected absorbance for 10 uL of 20 ug/L Fe, if the characteristic mass for Fe is 1.2 pg. (b) Comment on the instrument performance, if a 10 uL aliquot of 50 ug/L Pb gives an absorbance of 0.24, given the characteristic mass is 5.5 pg. Memory effects If a metallic element is present in the analysed aliquot at concentrations above 250 ug/L, the atomisation stage will not be sufficient to atomise and then remove all of it. This will occur whether the species is the analyte or in the matrix. If it is the analyte, then some proportion will remain in the tube when the next run commences. This will produce a higher absorbance than it should, and is known as a memory effect. AIT 3.8 3. Non-flame AAS The solution to this problem is a number of tube cleans, which take the temperature up to above 2000C very rapidly. Depending on the element and the concentration, it may require three or four tube cleans before continuing. CLASS EXERCISE 3.11 How could you check whether the tube is ready to be used after a number of tube cleans? Solutions A typical analysis run may use at least three different solutions: sample, added standard, blank, modifier. The total volume of liquid should be between 15 and 30 uL, and should be consistent for all solutions. Tube performance The graphite tube has a finite lifetime of around 500 normal temperature cycles. This will be reduced by various factors. CLASS EXERCISE 3.12 To be done in your own time. Download the Varian publication Factors affecting the analytical performance and lifetime of graphite tube atomizers from the AIT webpage. Read it and answer the following questions. They are examinable. (a) (b) (c) (d) (e) (f) What should be checked when changing tubes? What should be done after installing the new tube? What problems are encountered with solutions of different surface tension? Of high viscosity? What circumstances would require the use of air in the ashing stage? What limitations should be made on its use? Why is argon preferred to nitrogen? Why is a consistent liquid volume recommended? Background correction Background correction is a means of correcting for sources of non-analyte reduction in beam intensity. These sources include: smoke particles molecular species These are more significant (especially the smoke) in EAAS than FAAS. When it occurs, the measured absorbance is the sum of analyte and background absorbance (as in Eqn 3.3). An ideal system would measure the total absorbance, and also the background by itself. At = Aa + Ab AIT Eqn 3.3 3.9 3. Non-flame AAS There are two commonly employed systems for background correction: deuterium lamp – by aid of a chopper, the signal reaching the detector alternates between the HCL, which measures At, and a deuterium lamp (like in a UV-VIS) which only detects Ab; this is a simple system, but requires that the HCL and D2 beams are exactly aligned through the tube, which is difficult Zeeman effect – a magnetic effect which changes the characteristics of the analyte absorption line, meaning that when the magnet is on, the analyte no longer absorbs the HCL; this only requires the HCL lamp, but is much more expensive CLASS EXERCISE 3.13 Why would the analyte absorption not be detected in the deuterium beam? 3.7 Vapour generation The vapour generation technique relies on the formation of a volatile species containing the analyte metal. This may be the pure metal itself, as in the case of mercury (cold vapour), or the readily decomposed hydride of metals, such as arsenic, antimony and tin (hydride generation). The volatile form of the metal is created by chemical reaction with the sample in a reaction vessel which has a stream of gas (eg nitrogen) passing through it. The volatile species is carried along in the gas into the measurement part of the AAS where a quartz tube is seated (where the burner would be in a flame AAS). It has optical windows at either end to allow the beam from the hollow cathode lamp to pass through and out to the detector (see Figure 3.7). The analyte absorbs the radiation in the normal way. heated quartz tube Hollow Cathode Lamp Monochromator & Detector Gas stream in Gas stream out (with analyte) Reaction vessel (sample + reagents) FIGURE 3.7 Schematic diagram of vapour generation apparatus In the case of mercury, the analyte is already in the chemical form required – Hg(g) – after chemical treatment in the reaction vessel by reduction of ionic mercury by tin chloride, so no heating of the quartz tube is required, hence the name “cold vapour”. AIT 3.10 3. Non-flame AAS In the hydride generation method, the analyte is reduced, usually by sodium borohydride (NaBH4), producing the molecular hydride. When this is flushed out in the gas stream, it is not in the atomic form, and requiring heating in the tube to decompose the hydride into the atomic metal. This procedure is applicable to As, Se, Te, Sb, Bi and Sn. The tube may be heated electrically or with a flame. What You Need To Be Able To Do define important terminology explain reasons for limitations of flame AAS sensitivity outline the changes in NFAAS that allow improved sensitivity draw a schematic diagram of the workhead of an EAAS instrument and explain the function of the components, including the various services (eg water) describe a typical temperature program outline factors affecting the design of the temperature program outline how the nature of the analyte and matrix affects the temperature cycle explain the purpose of a modifier, and give an example of an analysis when it may be necessary explain the purpose of background correction, and give an example of an analysis when it may be necessary describe the principle of one type of background correction perform calculations related to characteristic mass list optimisation steps for routine EAAS analysis explain how vapour generation AAS works Revision Questions 1. (a) Give THREE factors that limit the sensitivity of flame AAS. (b) Explain how electrothermal AAS overcomes these. 2. What services other than electricity are required for EAAS? What function(s) do they serve? 3. Why is an autosampler an essential part of an EAAS instrument? 4. Why is the gas switched off at the end of the ashing stage? 5. What changes would you make to the basic “analyte in water” temperature program for the following analyses: (a) aluminium in milk (b) lead in petrol 6. Why is a phosphate modifier recommended for lead analyses? How does it work? 7. The characteristic mass for aluminium is 6.1 pg. What is the expected absorbance for a 5 uL aliquot of 50 ug/L standard? 8. What type of samples would most require background correction? 9. Explain the following for measurements of lead in a wine sample. Using the standard temperature program for lead in water, the sample gives an absorbance of 0.6 (a) When the ashing time is increased from 5 to 60 seconds, the absorbance decreases to 0.3. (b) When the drying time is reduced from 60 seconds to 5, the absorbance decreases to 0.1. (c) When the modifier is not used, there is no absorbance. 10. Why is tin chloride required in mercury analysis by vapour generation AAS? 11. Why is heat needed in the vapour generation analysis of arsenic? AIT 3.11 3. Non-flame AAS Answers to revision questions Where the answer can be found directly in your notes, only a reference to it will be provided. 1. (a) inefficiency of nebuliser, reaction in the flame, formation of oxides, background colour of flame (b) Table 3.1 2. p3.3 3. p3.4 4. top of p3.7 5. (a) increase drying & ashing times (b) decrease drying temp, small increase in ashing time 6. p3.7 7. mass in aliquot is 50 x 5 pg. Abs = 0.0044 x 250 ÷ 6.1 = 0.18 8. High organic content, high levels of elements giving molecular species eg Al, Ca 9. (a) significant amounts of smoke in the standard conditions run causing higher absorbance, this is removed during the longer ashing (b) insufficient drying time means that when the temperature goes up to the ashing point, the remaining moisture is rapidly lost, causing the sample to fizz and spray out of the tube. (c) the lead becomes volatile during the ashing stage and is lost before measurement 10. For this analysis, mercury must be in the elemental form (Hg0) so the tin chloride reduces the +1 and +2 ionic forms. 11. The volatile form of arsenic is AsH3, which must be decomposed (by heat) to form As to absorb the HCL radiation. AIT 3.12