IOP PUBLISHING PHYSICS IN MEDICINE AND BIOLOGY Phys. Med. Biol. 54 (2009) 6277–6286 doi:10.1088/0031-9155/54/20/016 A quantitative assessment of the depth sensitivity of an optical topography system using a solid dynamic tissue-phantom Teresa Correia, Anil Banga, N L Everdell, Adam P Gibson and Jeremy C Hebden Department of Medical Physics and Bioengineering, University College London, Gower Street, London WC1E 6BT, UK E-mail: tcorreia@medphys.ucl.ac.uk Received 17 June 2009, in final form 15 August 2009 Published 1 October 2009 Online at stacks.iop.org/PMB/54/6277 Abstract A solid dynamic phantom with tissue-like optical properties is presented, which contains seven discrete targets impregnated with thermochromic pigment located at different depths from the surface. Changes in absorption are obtained in response to localized heating of the targets, simulating haemodynamic changes occurring in the brain and other tissues. The depth sensitivity of a continuous wave optical topography system was assessed successfully using the phantom. Images of the targets have been reconstructed using a spatially variant regularization, and the determined spatial localization in the depth direction is shown to be accurate within an uncertainty of about 3 mm down to a depth of about 30 mm. 1. Introduction Diffuse optical imaging (DOI) is a non-invasive technique for monitoring haemodynamic activity in biological tissues (Leef et al 2008, Zeff et al 2007, Gibson et al 2005). It provides three-dimensional (3D) maps of the spatial distribution of optical properties of tissue, which can reveal changes in the concentration of oxygenated and deoxygenated haemoglobin, from measurements of light attenuation at near-infrared wavelengths. An array of sources and detectors is placed on the surface of the object of study. Where measurements of transmitted light are involved, the technique is usually known as optical tomography, whereas if reflectance measurements are made, it is often referred to as optical topography (Gibson et al 2005). For the latter the source–detector separations are relatively small (<3 cm), which restricts imaging to relatively superficial signals resulting in a degradation in spatial resolution with increasing depth. However, it is usually based on continuous-wave (CW) technology that permits fast acquisition times, which makes it optimal for real-time imaging applications. Nevertheless, 0031-9155/09/206277+10$30.00 © 2009 Institute of Physics and Engineering in Medicine Printed in the UK 6277 6278 T Correia et al for these types of systems, depth resolution can be improved by using a high-density array of optical fibres (Zeff et al 2007), overlapping measurements at different source–detector separations (Boas et al 2004) and a spatially variant regularization parameter (Culver et al 2003). Zhao et al (2007) used a depth variant regularization, which follows a sigmoidal decay function with depth, to reconstruct images from simulated data. They showed that depth contrast is improved and position errors of less than 3 mm could be obtained for objects located at depths from the surface of 1–3 cm. A similar study, using a regularization parameter proportional to the maximum singular value of the forward matrix for each depth layer, yielded identical results (Niu et al 2008). Endoh et al (2008) assigned a different regularization parameter to each voxel by setting the regularization parameter proportional to the square root of the diagonal elements of the forward matrix times its transpose (refer to section 3.1), hence compensating for both depth and lateral sensitivity decrease. They successfully reconstructed simultaneously two simulated objects located at different depths. Simulations, if correctly implemented, are good approximations to reality, and are inexpensive, relatively easy and fast to perform. However, they usually reproduce an ideal and simplified system performance, where, for example, all sources have identical noise levels. To fully characterize a specific imaging system, appropriate phantom measurements are essential. Over the last decade, the increasing popularity of DOI prompted the development of tissue-like optical phantoms to evaluate the performance of optical imaging systems, optimize existing systems and compare between systems and image reconstruction algorithms (Pogue and Patterson 2006). Most optical phantoms employ liquids, commonly based on an intralipid emulsion mixed with an absorbing element, such as near-infrared dyes (Flock et al 1992). Dynamic liquid phantoms have been used to mimic blood flow, or localized changes in haemoglobin concentration by flowing liquid through tubes embedded within a scattering solution (Franceschini et al 2006, Selb et al 2006, Kurth et al 1995). The disadvantages of using such phantoms are the relatively poor stability over time, the lack of repeatability with high accuracy, and the fact that they are not easily transportable. Hence, comparison between the different optical systems in different laboratories is difficult using this type of phantom. A solid dynamic phantom containing a reversible, electrically activated region has been developed by NIRx Medical Technologies (USA). Their phantom consists of realistic head and breast geometries with embedded electrochromic cells, whose optical properties can be programmed to vary over time (Barbour et al 2006). Koh et al (2009) proposed a phantom based on a liquid crystal display (LCD) inserted in a multi-layer resin phantom. Each pixel in the LCD can be activated individually, meaning that it can reproduce highly localized optical dynamic changes. Recently we presented an electrically activated tissue-like solid phantom, based on a thermochromic substance that changes colour in response to a change in temperature, which consequently produces changes in absorption (Hebden et al 2008a). A portable, batteryoperated prototype phantom containing two thermochromic targets was described, and its optical and temporal characteristics were investigated. An alternative design, based on an array of semiconductor diodes, allows the sequential activation of an arbitrary number of different thermochromic regions within a phantom (Hebden et al 2008b). This type of phantom can provide controlled and localized changes in the optical properties, and can mimic, for example, haemodynamic changes in the brain, which is essential to assess the performance of optical topography systems. One of the main limitations of optical topography is the low spatial accuracy, i.e. poor ability to accurately locate a change in the optical properties, in particular in the depth direction. In this paper we introduce a new design of electrically activated phantom, which has been developed to evaluate the ability of A quantitative assessment of the depth sensitivity of an optical topography system 6279 4.5mm 5mm z 4.7 k thermistor 4 3 5mm 115mm (a) (b) z 2 5mm 23mm 6 5 10 0m m y 5m m 12 resistor 7 41mm 50mm x x 1 5mm (c) Figure 1. (a) A target with thermistor and resistor embedded. The resistor is connected to one of the wires of the thermistor; (b) schematic of the phantom containing seven targets impregnated with thermochromic dye located at different depths. The targets are electrically activated through the sockets on top of the phantom; (c) top view of the phantom showing the target arrangement. Targets are at depths from 5 mm to 35 mm in intervals of 5 mm. an optical topography system to detect and localize discrete regions of absorption at different depths below the surface. The phantom is described, and its use to characterize the performance of an imaging system is reported. 2. Phantom design The phantom is based on a recipe described in detail in Hebden et al (2008a). It consists of a solid block of polyester resin (Alec Tiranti Ltd, UK) mixed with titanium dioxide (TiO2 ) particles, which provide tissue-like optical scattering, with embedded discrete R powder, Thermographic targets containing a black thermochromic pigment (ChromaZone Measurements Co. Ltd, UK). When heated, the pigment changes from black to white, with a nominal activation temperature of 47 ◦ C. Seven targets were generated by inserting a 12 surface mount resistor together with a 4.7 k bead thermistor (to monitor the temperature) inside a small plastic cylindrical tube and filling it with a mixture of thermochromic dye and polyester resin. The concentration of pigment was 16 mg cm−3, which produces a change in absorption of approximately 0.6 mm−1 when heated above the activation temperature, according to our previous measurements at 690 nm. When extracted from the tube, each target was a 4.5 mm diameter cylinder with a length of 41 mm. To reduce the number of wires and therefore the possibility of a short circuit within the narrow targets, the thermistor and heating resistor shared a common wire, as shown in figure 1(a). The thermistor was placed 5 mm from one end, to allow heat to flow along the length of the target without being blocked by the thermistor. The phantom was constructed in four stages. The first stage involved filling a polyethylene box with a 23 mm thick layer of polyester resin mixed with TiO2 powder to produce a transport scattering coefficient μs = 0.8 ± 0.05 mm−1, absorption coefficient μa = 0.001 ± 0.0002 mm−1 and dimensions of 115 mm × 100 mm. Once hardened, the block was removed from the box and the targets were attached to it with the arrangement illustrated in figure 1(b). This target arrangement minimizes the imaging array dimensions necessary to cover all targets simultaneously and confines the targets to a central position, which takes advantage of the higher sensitivity in the centre of the imaging array. The smallest spacing T Correia et al 6280 Figure 2. Imaging array with eight sources and eight detectors. D1–D15 are the source–detector separations. (This figure is in colour only in the electronic version) between targets is 11.2 mm, which is sufficient to ensure that heating of a target does not cause a colour change in its neighbours. The block with attached targets was placed in the box again, and the remaining space (50 mm) was filled with the same resin mixture described previously. Finally, the block was removed from the box and machined to provide a flat imaging surface, where the target nearest the surface was at depth 5 mm and the deepest at 35 mm, as shown in figure 1(c). The target closest to the surface will be referred to as target 1, the second target closest to the surface as target 2 and so on. Sockets were attached to the wires and mounted on a plastic plate fixed on the top of the phantom. Tests on a target embedded in a small volume of clear resin indicate that 3.5 V delivers a heating power of approximately 1.2 W, which is sufficient to produce a change in the absorption of the target after approximately 70 seconds, at room temperature. 3. Phantom measurements 3.1. Optical imaging The UCL continuous-wave optical topography system described by Everdell et al (2005), was used to image changes in absorption for each target within the phantom. The sources are frequency modulated and illuminated simultaneously, so that the contribution from each can be isolated from the detected signal, sampled at 10 Hz, by Fourier transform. The imaging array consists of eight detectors and eight sources operating at a wavelength of 670 nm. Data were collected for all 15 possible source–detector distances, making a total of 64 measurements. The array configuration is shown in figure 2. The probe was placed on the centre of the xy surface at z = 0 (using the coordinate system shown in figure 1(b)), with the centre of the array immediately above the centre of target 2. Data were acquired for 20 s with the targets at room temperature, and then each target was heated in turn for approximately 70 s, followed by a cooling period of 120 s to allow the targets to return to room temperature. For each target and each source–detector channel, a reference measurement was calculated from the mean signal of the 20 s recorded at room temperature. Similarly, the activation measurement was obtained by averaging over the 20 s segment of data centred on the time at which the maximum temperature was reached. Differences between A quantitative assessment of the depth sensitivity of an optical topography system (g) (e) (c) 6281 (f) (d) (b) (a) Figure 3. Optical topography images of each target in the xz plane. Bottom images from right to left correspond to (a) target 1 located at true depth ztrue = 5 mm, (b) target 2 at ztrue = 10 mm and (c) target 3 at ztrue = 15 mm. Middle images from right to left show (d) target 4 at ztrue = 20 mm and (e) target 5 at ztrue = 25 mm. Top images from right to left show (f) target 6 at ztrue = 30 mm and (g) target 7 at ztrue = 35 mm. the reference measurements and the corresponding activation measurements were used to generate 3D images representing absorption changes occurring within the phantom. The linear image reconstruction approach relates the changes in the measured data, y, to changes in the optical absorption x through y = J x, where J is the Jacobian or sensitivity matrix. The Jacobian was calculated using the software package TOAST (temporal optical absorption and scattering tomography), which has been developed by Professor Arridge and Dr Schweiger at UCL, by solving the diffusion equation using the finite element method (FEM), for both the forward and adjoint solutions (Arridge et al 2000), assuming uniform optical properties (μs = 0.8 mm−1 and μa = 0.001 mm−1). In order to find x one needs to solve an ill-posed and under-determined inverse problem, and Tikhonov regularization of the Moore–Penrose inverse is used in order to obtain a stable solution. This is expressed as x = L−1 J̃ T (J̃ J̃ T +λI )−1 y, where I is the identity matrix, λ is the regularization parameter, which was set to 0.5% of the maximum singular value of J J T , σmax (found using the L-curve method (Correia et al −1 1/2 2009)), and J̃ = J L−1 , where the spatial variant regularization L = diag J J T + σmax . As mentioned previously, this type of regularization has been shown to improve the depth spatial accuracy for images reconstructed from simulated data (Endoh et al 2008). Figure 3 shows the images for each target in the xz plane at y = 33 mm. The targets in the image have different relative locations, and visually they seem to be approximately T Correia et al 6282 Figure 4. Graph showing how resolution varies with depth. accurate. It is also clear that the spatial resolution and contrast decreases with depth. A quantitative analysis necessary to assess the imaging performance is described in the following section. 3.2. Image analysis For the purpose of this analysis only the xz cross-sectional images of each target were used. This gives us information regarding apparent depth and lateral positions of the targets. Spatial resolution is often characterized by the full width at half maximum (FWHM) of the point spread function (PSF). The true PSF represents the image of an object of an infinitesimal size. The image cross-section in the xz plane for each target is, in fact, the convolution of the true target structure with the PSF. Therefore, the squared FWHM of the target image is approximately equal to the squared width of the true target (4.5 mm) added to the squared width of the PSF, i.e. FWHM2target = FWHM2PSF + (4.5)2 . Profiles containing the highest intensity pixel in the image were generated for both the x and z directions and the FWHM was calculated in each case. The estimates of the PSF width were calculated using FWHMPSF = FWHM2target − (4.5)2 . The error is given by the pixel size, which is 3.33 mm for the z direction and 3.54 mm for the x direction. Figure 4 shows the mean values of the PSF width for each target plotted as a function of depth, and as expected the resolution decreases with depth, except for the discrepancy in the general trend between targets 3 and 4. The lower resolution of target 3 is almost certainly due to the lower sensitivity of the imaging array to perturbations located near the edges, as opposed to higher sensitivity to perturbations in central locations. Contrast is a measure of the ability to distinguish a feature based on its brightness in the image relative to the background. Image contrast in terms of μa can be expressed as C = μamax − μabkg μabkg , where μamax refers to the maximum absorption coefficient value in the image and μabkg is the mean background absorption coefficient. The error associated with derived contrast values corresponds to the propagation of the uncertainty associated with σμ , where σμ is the the calculation of the background absorption coefficient, i.e., ∂μ∂C abkg ab kg a bkg A quantitative assessment of the depth sensitivity of an optical topography system 6283 Figure 5. Graph showing how contrast varies with depth. standard deviation of μabkg . As shown in figure 5, contrast decreases with increasing depth up to 20 mm depth, and beyond this point contrast values are very low. This means that the reconstructed absorption changes become less accurate with depth. The error increases from 4% of the background value for the first target to 15% for the deepest target. The spatial accuracy is defined as a measure of the error between the true and the apparent positions of the target. It is assumed that the target closest to the surface was in the correct location, and therefore the positions of the other targets were calculated relative to this. The derived from weighted mean position along the PSF profiles, which apparent position xm is the n μ is calculated as xm = ni μai xi ai . Again, the uncertainty in position corresponds to i the intrinsic uncertainty of the pixel size. As shown in figure 6, the lateral spatial accuracy reduces with depth, due to the lower sensitivity of the measurements and the smaller number of overlapping measurements at deeper regions. As shown in figure 7, the calculated apparent position in the z axis reveals a very good agreement with the true depth of the target. From the results described before, one would expect the depth accuracy to be reasonable up to 20 mm depth. However, the apparent depth agrees with the true depth within the associated error, at all depths up to at least 30 mm. 4. Discussion One of the limitations of optical tomography as a brain imaging tool is the low sensitivity to haemodynamic activity occurring within deeper tissues compared to that occurring nearer the surface. This is due to the fact that larger source–detector separations are required to sample at greater depths, and large separation implies more attenuation and less diffusely reflected light. The volume of tissue sampled also increases with larger separation, and consequently spatial resolution and contrast also decrease with depth. Nevertheless, considerable effort has been devoted to the development of image reconstruction algorithms capable of generating 3D images from diffuse reflectance data, and schemes have been devised which aim to improve the ability of optical tomography to discriminate along the depth direction [18]. 6284 T Correia et al Figure 6. Graph showing the difference between the real target position and apparent position in the image, assuming that the target closest to the surface is in the correct position. Figure 7. Graph showing how the depth of the target in the image varies with true depth. The expected depth is represented by the straight line. In this paper we have described a dynamic tissue phantom for quantitative assessment of the depth sensitivities of optical topography systems. The phantom is easy to use, compact, involves no moving parts, and gives reliable and repeatable changes in absorption properties. Its portability enables it to be easily exchanged between users of similar systems, enabling appropriate comparisons of imaging performance to be made. The background absorption is deliberately relatively low (representing that of fatty tissue at near-infrared wavelengths), enabling depth sensitivity to be evaluated at large depths. The optical properties of the resin block are stable over many years, while the thermochromic pigment is known to maintain A quantitative assessment of the depth sensitivity of an optical topography system 6285 its properties for at least 12 months (unless subjected to excessive heating, well above the activation temperature). The study presented here indicates that while spatial resolution degrades roughly linearly with depth, contrast falls exponentially. It is also shown that, with an appropriate choice of regularization parameter, the apparent depth of a feature can match the true depth within the statistical uncertainty of the measurement. Relatively good resolution, contrast and spatial accuracy were obtained for images reconstructed of targets up to 20 mm depth. For targets located deeper, our system was still able to reconstruct sensible images but with lower quality. An increase in the number of overlapping measurements at deeper regions could potentially improve the quality of the images of deeper located targets. There are three factors which contribute towards the poor recovery of true contrast values shown in figure 5. First, the linear reconstruction algorithm inherently assumes that all changes in optical properties are small (infinitesimal). Second, the Jacobian matrix used in the reconstruction algorithm assumes homogeneous optical properties (as is common practice in scientific applications of optical topography), while the phantom is actually highly heterogeneous. Although nonlinear methods (where the Jacobian is updated iteratively (Arridge 1999)) have been shown to provide more quantitatively accurate results, they are computationally complex and slow, and not suitable for real-time imaging of haemodynamic activity. Third, the decrease in spatial resolution with depth (shown in figure 4) implies that the contrast is spread over a greater volume at larger depths, and thus the maximum absorption value in the image is lower (hence the decrease observed in figure 5). Although the phantom presented here is of a generic design, the recipe can be adapted to produce phantoms with any background optical properties and incorporating different shapes and arrangements of targets. This offers the facility to develop dynamic, fully portable phantoms for a range of different tests of diffuse optical imaging systems. Acknowledgments The work has been supported in part by the EPSRC, and by a scholarship awarded to TC by Fundação para a Ciência e a Tecnologia, Portugal. This work has also been supported in part by funding from the EC’s seventh framework programme under grant agreement FP7-HEALTH-2007-201076. References Arridge S R 1999 Optical tomography in medical imaging Inverse Problems 15 R41–93 Arridge S R, Hebden J C, Schweiger M, Schmidt F E W, Fry M E, Hillman E M C, Dehghani H and Delpy D T 2000 A method for 3d time-resolved optical tomography Imag. Syst. Technol. 11 2–11 Barbour R L, Graber H L, Pei Y, Ansari R, Levin M B and Farber M 2006 Diffuse optical tissue simulator (dots): an experimental calibrating system for functional dot imaging Fifth Inter-institute Workshop on Optical Diagnostic Imaging from Bench to Bedside at the National Institutes of Health (Bethesda, MD, 25–27 September 2006) Boas D A, Chen K, Grebert D and Franceschini M A 2004 Improving the diffuse optical imaging spatial resolution of the cerebral hemodynamic response to brain activation in humans Opt. Lett. 29 1506–8 Correia T, Gibson A, Schweiger M and Hebden J 2009 Selection of regularization parameter for optical topography J. Biomed. Opt. 14 034044 Culver J P, Siegel A M, Stott J J and Boas D A 2003 Volumetric diffuse optical tomography of brain activity Opt. Lett. 28 2061–3 Endoh R, Fujii M and Nakayama K 2008 Depth-adaptive regularized reconstruction for reflection diffuse optical tomography Opt. Rev. 15 51–6 Everdell N L, Gibson A P, Tullis I D C, Hebden J C and Delpy D T 2005 A frequency multiplexed near infrared topography system for imaging functional activation in the brain Rev. Sci. Instrum. 76 093705 6286 T Correia et al Flock S T, Jacques S L, Wilson B C, Star W M and Vangemert M J C 1992 Optical properties of intralipid—a phantom medium for light-propagation studies Lasers Surg. Med. 12 510–9 Franceschini M A, Joseph D K, Huppert T J, Diamond S G and Boas D A 2006 Diffuse optical imaging of the whole head J. Biomed. Opt. 11 054007 Gibson A P, Hebden J C and Arridge S R 2005 Recent advances in diffuse optical imaging Phys. Med. Biol. 50 R1–43 Hebden J C, Brunker J, Correia T, Price B D, Gibson A P and Everdell N L 2008a An electrically-activated dynamic tissue-equivalent phantom for assessment of diffuse optical imaging systems Phys. Med. Biol. 53 329–37 Hebden J C, Correia T, Khakoo I, Gibson A P and Everdell N L 2008b A dynamic optical imaging phantom based on an array of semiconductor diodes Phys. Med. Biol. 53 N407–13 Koh P H, Elwell C E and Delpy D T 2009 Development of a dynamic test phantom for optical topography Advances in Experimental Medicine and Biology—Oxygen Transport to Tissue XXX vol 645, ed P Liss, P Hansell, D F Bruley and D K Harrison (New York: Springer) pp 141–6 Kurth C D, Watson R and Cope M 1995 A dynamic phantom brain model for near-infrared spectroscopy Phys. Med. Biol. 40 2079–92 Leef D R, Warren O J, Enfield L C, Gibson A, Athanasiou T, Patten D K, Hebden J, Yang G Z and Darzi A 2008 Diffuse optical imaging of the healthy and diseased breast: a systematic review Breast cancer Res. Treat 108 9–22 Niu H, Guo P and Jiang T 2008 Improving depth resolution of diffuse optical tomography with intelligent method ICIC ’08: Proc. of the 4th Int. Conf. on Intelligent Computing (Berlin: Springer) pp 514–20 Pogue B W and Patterson M S 2006 Review of tissue simulating phantoms for optical spectroscopy, imaging and dosimetry J. Biomed. Opt. 11 041102 Selb J, Joseph D K and Boas D A 2006 Time-gated optical system for depth-resolved functional brain imaging J. Biomed. Opt. 11 044008 Zeff B W, White B R, Dehghani H, Schlaggar B L and Culver J P 2007 Retinotopic mapping of adult human visual cortex with high-density diffuse optical tomography Proc. the Natl. Acad. Sci. USA 104 12169–74 Zhao Q, L Ji and Jiang T 2007 Improving depth resolution of diffuse optical tomography with a layer-based sigmoid adjustment method Opt. Express 15 4018–29