ab210965 – Mouse Myoglobin SimpleStep ELISA Kit

advertisement
ab210965 – Mouse
Myoglobin SimpleStep
ELISA® Kit
Instructions for use:
For the quantitative measurement of Myoglobin in mouse
serum, plasma, and cell culture supernatant, cell and
tissue extract samples.
This product is for research use only and is not intended
for diagnostic use.
Table of Contents
INTRODUCTION
1
1.
BACKGROUND
1
2.
ASSAY SUMMARY
2
GENERAL INFORMATION
3
3.
PRECAUTIONS
3
4.
STORAGE AND STABILITY
3
5.
MATERIALS SUPPLIED
3
6.
MATERIALS REQUIRED, NOT SUPPLIED
4
7.
LIMITATIONS
4
8.
TECHNICAL HINTS
4
ASSAY PREPARATION
6
9.
6
REAGENT PREPARATION
10.
STANDARD PREPARATION
11.
SAMPLE PREPARATION
12.
PLATE PREPARATION
7
9
12
ASSAY PROCEDURE
13
13.
13
ASSAY PROCEDURE
DATA ANALYSIS
15
14.
CALCULATIONS
15
15.
TYPICAL DATA
16
16.
TYPICAL SAMPLE VALUES
18
17.
ASSAY SPECIFICITY
23
18.
SPECIES REACTIVITY
24
RESOURCES
25
19.
TROUBLESHOOTING
25
20.
NOTES
26
INTRODUCTION
INTRODUCTION
1. BACKGROUND
Myoglobin
in
vitro
SimpleStep
ELISA®
(Enzyme-Linked
Immunosorbent Assay) kit is designed for the quantitative
measurement of Myoglobin protein in mouse serum, plasma, and cell
culture supernatant, cell and tissue extract samples.
The SimpleStep ELISA® employs an affinity tag labeled capture
antibody and a reporter conjugated detector antibody which
immunocapture the sample analyte in solution. This entire complex
(capture antibody/analyte/detector antibody) is in turn immobilized via
immunoaffinity of an anti-tag antibody coating the well. To perform the
assay, samples or standards are added to the wells, followed by the
antibody mix. After incubation, the wells are washed to remove
unbound material. TMB substrate is added and during incubation is
catalyzed by HRP, generating blue coloration. This reaction is then
stopped by addition of Stop Solution completing any color change from
blue to yellow. Signal is generated proportionally to the amount of
bound analyte and the intensity is measured at 450 nm. Optionally,
instead of the endpoint reading, development of TMB can be recorded
kinetically at 600 nm.
Myoglobin is an oxygen- and iron-binding protein involved in the
transportation of oxygen within muscles. Upon muscle injury,
myoglobin can also be found in the bloodstream.
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
1
INTRODUCTION
2. ASSAY SUMMARY
Remove appropriate number of
antibody coated well strips.
Equilibrate all reagents to room
temperature. Prepare all
reagents, samples, and
standards as instructed.
Add standard or sample to
appropriate wells.
Add Antibody Cocktail to all
wells. Incubate at room
temperature.
Aspirate and wash each well.
Add TMB Substrate to each well
and incubate. Add Stop Solution
at a defined endpoint.
Alternatively, record color
development kinetically after
TMB substrate addition.
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
2
GENERAL INFORMATION
GENERAL INFORMATION
3. PRECAUTIONS
Please read these instructions carefully prior to beginning the
assay.
All kit components have been formulated and quality control tested to
function successfully as a kit. Modifications to the kit components or
procedures may result in loss of performance.
4. STORAGE AND STABILITY
Store kit at 2-8ºC immediately upon receipt.
Refer to list of materials supplied for storage conditions of individual
components. Observe the storage conditions for individual prepared
components in the Reagent and Standard Preparation sections.
5. MATERIALS SUPPLIED
10X Mouse Myoglobin Capture Antibody
600 µL
Storage
Condition
(Before
Preparation)
+4ºC
10X Mouse Myoglobin Detector Antibody
Mouse Myoglobin Lyophilized Recombinant
Protein
Antibody Diluent 4BI
600 µL
+4ºC
2 Vials
+4ºC
6 mL
+4ºC
10X Wash Buffer PT
20 mL
+4ºC
5X Cell Extraction Buffer PTR
10 mL
+4ºC
50X Cell Extraction Enhancer Solution
1 mL
+4ºC
TMB Substrate
12 mL
+4ºC
Stop Solution
12 mL
+4ºC
Sample Diluent NS
Pre-Coated 96 Well Microplate (12 x 8 well
strips)
Plate Seal
50 mL
+4ºC
96 Wells
+4ºC
1
+4ºC
Item
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
Amount
3
GENERAL INFORMATION
6. MATERIALS REQUIRED, NOT SUPPLIED
These materials are not included in the kit, but will be required to
successfully utilize this assay:

Microplate reader capable of measuring absorbance at 450 or
600 nm.

Method for determining protein concentration (BCA assay
recommended).

Deionized water.

Multi- and single-channel pipettes.

Tubes for standard dilution.

Plate shaker for all incubation steps.

Optional: Phenylmethylsulfonyl Fluoride (PMSF) (or other protease
inhibitors).
7. LIMITATIONS

Assay kit intended for research use only. Not for use in diagnostic
procedures.

Do not mix or substitute reagents or materials from other kit lots or
vendors. Kits are QC tested as a set of components and
performance cannot be guaranteed if utilized separately or
substituted.
8. TECHNICAL HINTS

Samples generating values higher than the highest standard
should be further diluted in the appropriate sample dilution buffers.

Avoid foaming
components.

Avoid cross contamination of samples or reagents by changing tips
between sample, standard and reagent additions.

Ensure plates are properly sealed or covered during incubation
steps.
or
bubbles
when
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
mixing
or
reconstituting
4
GENERAL INFORMATION

Complete removal of all solutions and buffers during wash steps is
necessary to minimize background.

As a guide, typical ranges of sample concentration for commonly
used sample types are shown below in Sample Preparation
(section 11).

All samples should be mixed thoroughly and gently.

Avoid multiple freeze/thaw of samples.

Incubate ELISA plates on a plate shaker during all incubation
steps.

When generating positive control samples, it is advisable to
change pipette tips after each step.

The provided 5X Cell Extraction Buffer contains phosphatase
inhibitors and protease inhibitor aprotinin.
Additional protease
inhibitors can be added if required.

The provided Antibody Diluents and Sample Diluents contain
protease inhibitor aprotinin. Additional protease inhibitors can be
added if required.

The provided 50X Cell Extraction Enhancer Solution may
precipitate when stored at + 4ºC. To dissolve, warm briefly at
+ 37ºC and mix gently. The 50X Cell Extraction Enhancer Solution
can be stored at room temperature to avoid precipitation.

To avoid high background always add samples or standards
to the well before the addition of the antibody cocktail.

This kit is sold based on number of tests. A ‘test’ simply
refers to a single assay well. The number of wells that
contain sample, control or standard will vary by product.
Review the protocol completely to confirm this kit meets your
requirements. Please contact our Technical Support staff
with any questions.
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
5
ASSAY PREPARATION
ASSAY PREPARATION
9. REAGENT PREPARATION

Equilibrate all reagents to room temperature (18-25°C) prior to
use. The kit contains enough reagents for 96 wells. The sample
volumes below are sufficient for 48 wells (6 x 8-well strips);
adjust volumes as needed for the number of strips in your
experiment.

Prepare only as much reagent as is needed on the day of the
experiment. Capture and Detector Antibodies have only been
tested for stability in the provided 10X formulations.
9.1
1X Cell Extraction Buffer PTR (For cell and tissue extracts
only)
Prepare 1X Cell Extraction Buffer PTR by diluting 5X Cell
Extraction Buffer PTR and 50X Cell Extraction Enhancer
Solution to 1X with deionized water. To make 10 mL 1X Cell
Extraction Buffer PTR combine 8 mL deionized water and
2 mL 5X Cell Extraction Buffer PTR. Mix thoroughly and
gently. If required protease inhibitors can be added.
9.2
1X Wash Buffer PT
Prepare 1X Wash Buffer PT by diluting 10X Wash Buffer PT
with deionized water. To make 50 mL 1X Wash Buffer PT
combine 5 mL 10X Wash Buffer PT with 45 mL deionized
water. Mix thoroughly and gently.
9.3
Antibody Cocktail
Prepare Antibody Cocktail by diluting the capture and
detector antibodies in Antibody Diluent. To make 3 mL of
the Antibody Cocktail combine 300 µL 10X Capture Antibody
and 300 µL 10X Detector Antibody with 2.4 mL Antibody
Diluent. Mix thoroughly and gently.
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
6
ASSAY PREPARATION
10. STANDARD PREPARATION
Prepare serially diluted standards immediately prior to use. Always
prepare a fresh set of positive controls for every use.
The following section describes the preparation of a standard curve for
duplicate measurements (recommended).
10.1 For serum, plasma and cell culture supernatant samples
follow these instructions:
10.1.1 Reconstitute the Myoglobin standard sample by
adding 100 µL water. Mix thoroughly and gently.
Hold at room temperature for 10 minutes and mix
gently. This is the 50,000 pg/mL Stock Standard
Solution.
10.1.2 Label eight tubes, Standards 1– 8.
10.1.3 Add 475 μL Sample Diluent NS into tube number 1
and 150 μL of Sample Diluent NS into numbers 2-8.
10.1.4 Use the Stock Standard to prepare the following
dilution series. Standard #8 contains no protein and
is the Blank control:
25 µL
150 µL 150 µL 150 µL 150 µL 150 µL 150 µL
µ
50,000
pg/mL
2,500
pg/mL
XX µL
µL
µL
XX µL
µ
µ
µ
µ
625
1,250
312.5 156.25
pg/mL pg/mL pg/mL pg/mL
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
µ
0
78.12 39.06
pg/mL pg/mL pg/mL
7
ASSAY PREPARATION
10.2 For cell and tissue extract samples follow these
instructions:
10.2.1 Reconstitute the Myoglobin standard sample by
adding 100 µL water. Mix thoroughly and gently.
Hold at room temperature for 10 minutes and mix
gently. This is the 50,000 pg/mL Stock Standard
Solution.
10.2.2 Label eight tubes, Standards 1– 8.
10.2.3 Add 490 μL 1X Cell Extraction Buffer PTR into tube
number 1 and 150 μL of 1X Cell Extraction Buffer
PTR into numbers 2-8
10.2.4 Use the Stock Standard to prepare the following
dilution series. Standard #8 contains no protein and
is the Blank control:
10 µL
150 µL 150 µL 150 µL 150 µL 150 µL 150 µL
µ
µ
µ
µ
L
µ
µ
50,000
pg/mL
1,000
pg/mL
250
125
500
62.5
pg/mL pg/mL pg/mL pg/mL
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
0
31.25 15.62
pg/mL pg/mL pg/mL
8
ASSAY PREPARATION
11. SAMPLE PREPARATION
TYPICAL SAMPLE DYNAMIC RANGE
Sample Type
Range
Mouse Plasma - Heparin
1% - 20%
Mouse Plasma - EDTA
1:64 – 1:4
Mouse Plasma - Citrate
2% – 38%
Mouse Serum
1:128 – 1:8
Heart Supernatant
1:50,00 – 1:1600
C2C12 Cell Extract
8 – 125 μg/mL
Skeletal Muscle Extract
20 – 300 ng/mL
11.1 Plasma
Collect plasma using citrate, EDTA or heparin. Centrifuge
samples at 2,000 x g for 10 minutes. Dilute samples into
Sample Diluent NS and assay. Store un-diluted plasma
samples at -20ºC or below for up to 3 months. Avoid
repeated freeze-thaw cycles.
11.2 Serum
Samples should be collected into a serum separator tube.
After clot formation, centrifuge samples at 2,000 x g for
10 minutes and collect serum. Dilute samples into Sample
Diluent NS and assay. Store un-diluted serum at -20ºC or
below. Avoid repeated freeze-thaw cycles.
11.3 Cell Culture Supernatants
Centrifuge cell culture media at 2,000 x g for 10 minutes to
remove debris. Collect supernatants and assay. Or dilute
samples into Sample Diluent NS and assay. Store un-diluted
samples at -20°C or below. Avoid repeated freeze-thaw
cycles.
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
9
ASSAY PREPARATION
11.4 Preparation of extracts from cell pellets
11.4.1 Collect non-adherent cells by centrifugation or
scrape to collect adherent cells from the culture
flask. Typical centrifugation conditions for cells are
500 x g for 5 minutes at 4ºC.
11.4.2 Rinse cells twice with PBS.
11.4.3 Solubilize pellet at 2x107 cell/mL in chilled 1X Cell
Extraction Buffer PTR.
11.4.4 Incubate on ice for 20 minutes.
11.4.5 Centrifuge at 18,000 x g for 20 minutes at 4°C.
11.4.6 Transfer the supernatants into clean tubes and
discard the pellets.
11.4.7 Assay samples immediately or aliquot and store at
-80°C. The sample protein concentration in the
extract may be quantified using a protein assay.
11.4.8 Dilute samples to desired concentration in 1X Cell
Extraction Buffer PTR.
11.5 Preparation of extracts from adherent cells by direct
lysis (alternative protocol)
11.5.1 Remove growth media and rinse adherent cells 2
times in PBS.
11.5.2 Solubilize the cells by addition of chilled 1X Cell
Extraction Buffer PTR directly to the plate (use
750 µL - 1.5 mL 1X Cell Extraction Buffer PTR per
confluent 15 cm diameter plate).
11.5.3 Scrape the cells into a microfuge tube and incubate
the lysate on ice for 15 minutes.
11.5.4 Centrifuge at 18,000 x g for 20 minutes at 4°C.
11.5.5 Transfer the supernatants into clean tubes and
discard the pellets.
11.5.6 Assay samples immediately or aliquot and store at
-80°C. The sample protein concentration in the
extract may be quantified using a protein assay.
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
10
ASSAY PREPARATION
11.5.7 Dilute samples to desired concentration in 1X Cell
Extraction Buffer PTR.
11.6 Preparation of extracts from tissue homogenates
11.6.1 Tissue lysates are typically prepared by
homogenization of tissue that is first minced and
thoroughly rinsed in PBS to remove blood (dounce
homogenizer recommended).
11.6.2 Homogenize 100 to 200 mg of wet tissue in
500 µL – 1 mL of chilled 1X Cell Extraction Buffer
PTR. For lower amounts of tissue adjust volumes
accordingly.
11.6.3 Incubate on ice for 20 minutes.
11.6.4 Centrifuge at 18,000 x g for 20 minutes at 4°C.
11.6.5 Transfer the supernatants into clean tubes and
discard the pellets.
11.6.6 Assay samples immediately or aliquot and store at
-80°C. The sample protein concentration in the
extract may be quantified using a protein assay.
11.6.7 Dilute samples to desired concentration in 1X Cell
Extraction Buffer PTR.
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
11
ASSAY PREPARATION
12. PLATE PREPARATION

The 96 well plate strips included with this kit are supplied ready to
use. It is not necessary to rinse the plate prior to adding reagents.

Unused plate strips should be immediately returned to the foil
pouch containing the desiccant pack, resealed and stored at 4°C.

For each assay performed, a minimum of two wells must be used
as the zero control.

For statistical reasons, we recommend each sample should be
assayed with a minimum of two replicates (duplicates).

Differences in well absorbance or “edge effects” have not been
observed with this assay.
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
12
ASSAY PROCEDURE
ASSAY PROCEDURE
13. ASSAY PROCEDURE

Equilibrate all materials and prepared reagents to room
temperature prior to use.

It is recommended to assay all standards, controls and
samples in duplicate.
13.1. Prepare all reagents, working standards, and samples as
directed in the previous sections.
13.2. Remove excess microplate strips from the plate frame,
return them to the foil pouch containing the desiccant pack,
reseal and return to 4ºC storage.
13.3. Add 50 µL of all sample or standard to appropriate wells.
13.4. Add 50 µL of the Antibody Cocktail to each well.
13.5. Seal the plate and incubate for 1 hour at room temperature
on a plate shaker set to 400 rpm.
13.6. Wash each well with 3 x 350 µL 1X Wash Buffer PT. Wash
by aspirating or decanting from wells then dispensing 350 µL
1X Wash Buffer PT into each well. Complete removal of
liquid at each step is essential for good performance. After
the last wash invert the plate and blot it against clean paper
towels to remove excess liquid.
13.7. Add 100 µL of TMB Substrate to each well and incubate for
10 minutes in the dark on a plate shaker set to 400 rpm.
13.8. Add 100 µL of Stop Solution to each well. Shake plate on a
plate shaker for 1 minute to mix. Record the OD at 450 nm.
This is an endpoint reading.
Alternative to 13.7 – 13.8: Instead of the endpoint reading at
450 nm, record the development of TMB Substrate
kinetically. Immediately after addition of TMB Development
Solution begin recording the blue color development with
elapsed time in the microplate reader prepared with the
following settings:
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
13
ASSAY PROCEDURE
Mode:
Kinetic
Wavelength:
600 nm
Time:
up to 15 min
Interval:
20 sec - 1 min
Shaking:
Shake between readings
Note that an endpoint reading can also be recorded at the
completion of the kinetic read by adding 100 µL Stop
Solution to each well and recording the OD at 450 nm.
13.9. Analyze the data as described below.
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
14
DATA ANALYSIS
DATA ANALYSIS
14. CALCULATIONS
14.1 Calculate the average absorbance value for the blank
control (zero) standards. Subtract the average blank control
standard absorbance value from all other absorbance
values.
14.2 Create a standard curve by plotting the average blank
control subtracted absorbance value for each standard
concentration (y-axis) against the target protein
concentration (x-axis) of the standard. Use graphing
software to draw the best smooth curve through these points
to construct the standard curve.
Note: Most microplate reader software or graphing software
will plot these values and fit a curve to the data. A four
parameter curve fit (4PL) is often the best choice; however,
other algorithms (e.g. linear, semi-log, log/log, 4 parameter
logistic) can also be tested to determine if it provides a better
curve fit to the standard values.
14.3 Determine the concentration of the target protein in the
sample by interpolating the blank control subtracted
absorbance values against the standard curve. Multiply
the resulting value by the appropriate sample dilution factor,
if used, to obtain the concentration of target protein in the
sample.
14.4 Samples generating absorbance values greater than that of
the highest standard should be further diluted and
reanalyzed. Similarly, samples which measure at an
absorbance values less than that of the lowest standard
should be retested in a less dilute form.
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
15
DATA ANALYSIS
15. TYPICAL DATA
TYPICAL STANDARD CURVE – Data provided for demonstration
purposes only. A new standard curve must be generated for each
assay performed.
Standard Curve Measurements
Conc.
O.D. 450 nm
Mean
(pg/mL)
1
2
O.D.
0
0.092
0.095
0.094
39.06
0.176
0.153
0.165
78.12
0.226
0.188
0.207
156.25
0.284
0.296
0.290
312.5
0.510
0.525
0.518
625
0.845
0.979
0.912
1,250
1.734
1.980
1.857
2,500
3.427
3.615
3.521
Figure 1. Example of mouse Myoglobin standard curve in Sample Diluent NS.
The Myoglobin standard curve was prepared as described in Section 10. Raw
data values are shown in the table. Background-subtracted data values (mean
+/- SD) are graphed.
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
16
DATA ANALYSIS
Standard Curve Measurements
Conc.
O.D. 450 nm
Mean
(pg/mL)
1
2
O.D.
0
0.056
0.060
0.058
15.62
0.103
0.107
0.105
31.25
0.141
0.152
0.146
62.5
0.209
0.231
0.220
125
0.372
0.396
0.384
250
0.731
0.764
0.747
500
1.470
1.507
1.489
1,000
3.049
3.112
3.081
Figure 2. Example of mouse Myoglobin standard curve in 1X Cell Extraction
Buffer PTR. The Myoglobin standard curve was prepared as described in
Section 10. Raw data values are shown in the table. Background-subtracted
data values (mean +/- SD) are graphed.
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
17
DATA ANALYSIS
16. TYPICAL SAMPLE VALUES
SENSITIVITY –
The calculated minimal detectable dose (MDD) is determined by
calculating the mean of zero standard replicates and adding 2 standard
deviations then extrapolating the corresponding concentration. The
MDD is dependent on the Sample Diluent buffer used:
Sample Diluent Buffer
Sample Diluent NS
1X Cell Extraction Buffer PTR
n=
Minimal Detectable
Dose
24
24
10.6 pg/mL
9.4 pg/mL
RECOVERY –
Three concentrations of Myoglobin were spiked in duplicate to the
indicated biological matrix to evaluate signal recovery in the working
range of the assay.
Sample Type
Mouse Serum
Mouse Plasma - Citrate
Mouse Plasma - Heparin
Mouse Plasma - EDTA
Mouse Heart Supernatant
Mouse Skeletal Muscle Extract
Average %
Recovery
88
112
105
100
115
112
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
Range (%)
76-99
100-120
100-113
95-106
110-123
108-119
18
DATA ANALYSIS
LINEARITY OF DILUTION –
Linearity of dilution is determined based on interpolated values from
the standard curve. Linearity of dilution defines a sample
concentration interval in which interpolated target concentrations are
directly proportional to sample dilution.
Native Myoglobin was measured in the following biological samples in
a 2-fold dilution series. Sample dilutions are made in Sample Diluent
NS.
37.5%
25%
20%
12.5%
1:1600
Dilution
Mouse Mouse Mouse
Interpolated value Mouse
Mouse Heart
Plasma Plasma Plasma
Factor
Serum
Supernatant
(Citrate) (EDTA) (Heparin)
2,757
pg/mL
896
1,452
2,226
1,260
Undiluted
100
% Expected value 100
100
100
100
1,296
pg/mL
386
689
965
673
2
94
% Expected value
86
95
87
107
596
pg/mL
200
341
529
298
4
86
% Expected value
89
94
95
94
281
pg/mL
108
162
211
149
8
82
% Expected value
96
89
76
95
150
pg/mL
60
75
109
85
16
87
% Expected value 109
83
78
108
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
19
DATA ANALYSIS
Native Myoglobin was measured in the following biological samples in
a 2-fold dilution series. Sample dilutions are made in 1X Cell Extraction
Buffer PTR
Dilution
Factor
Undiluted
2
4
8
16
Interpolated value
pg/mL
% Expected value
pg/mL
% Expected value
pg/mL
% Expected value
pg/mL
% Expected value
pg/mL
% Expected value
300 ng/mL Mouse 125 μg/mL
Skeletal Muscle
C2C12
Extract
Extract
1044
687
100
100
559
353
107
103
263
168
101
98
136
84
105
97
69
38
106
88
PRECISION –
Mean coefficient of variations of interpolated values of Myoglobin in 3
concentrations of mouse serum within the working range of the assay.
n=
CV (%)
IntraAssay
8
6.1
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
InterAssay
3
4.8
20
DATA ANALYSIS
Figure 3. Interpolated concentrations of native Myoglobin in mouse serum,
plasma and cell culture supernatant samples.
The concentrations of
Myoglobin were measured in duplicates, interpolated from the Myoglobin
standard curves and corrected for sample dilution. Undiluted samples are as
follows: serum 12.5%, plasma (citrate) 37.5%, plasma (heparin) 20%, plasma
(EDTA) 25% and heart supernatant 1:1600. The interpolated dilution factor
corrected values are plotted (mean +/- SD, n=2). The mean Myoglobin
concentration was determined to be 6.87 ng/mL in serum, 3.57 ng/mL in
plasma (citrate), 6.35 ng/mL in plasma (heparin), 7.76 ng/mL in plasma
(EDTA) and 4,187 ng/mL in heart supernatant.
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
21
DATA ANALYSIS
Figure 4. Interpolated concentrations of native Myoglobin in mouse skeletal
muscle based on a 300 ng/mL extract load and C2C12 extract based on a 125
μg/mL extract load. The concentrations of Myoglobin were measured in
duplicate and interpolated from the Myoglobin standard curve and corrected
for sample dilution. The interpolated dilution factor corrected values are
plotted (mean +/- SD, n=2). The mean Myoglobin concentration was
determined to be 1.08 ng/mL in skeletal muscle extract and 0.67 ng/mL in
C2C12 extract.
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
22
DATA ANALYSIS
17. ASSAY SPECIFICITY
This kit recognizes both native and recombinant mouse Myoglobin
protein in serum, plasma, and cell culture supernatant, cell and tissue
extract samples only.
CROSS REACTIVITY
Recombinant human Myoglobin was prepared at 50 ng/mL and 1
ng/mL and assayed for cross reactivity. No cross-reactivity was
observed.
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
23
DATA ANALYSIS
18. SPECIES REACTIVITY
This kit recognizes mouse Myoglobin protein.
Other species reactivity was determined by measuring 15% serum
samples of various species, interpolating the protein concentrations
from the mouse standard curve, and expressing the interpolated
concentrations as a percentage of the protein concentration in mouse
serum assayed at the same dilution.
Reactivity < 3% was determined for the following species:


Human
Cow
Please contact our Technical Support team for more information.
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
24
RESOURCES
RESOURCES
19. TROUBLESHOOTING
Problem
Cause
Solution
Inaccurate Pipetting
Check pipettes
Improper standard
dilution
Prior to opening, briefly spin
the stock standard tube and
dissolve the powder
thoroughly by gentle mixing
Incubation times too
brief
Ensure sufficient incubation
times; increase to 2 or 3 hour
standard/sample incubation
Inadequate reagent
volumes or improper
dilution
Check pipettes and ensure
correct preparation
Incubation times with
TMB too brief
Ensure sufficient incubation
time until blue color develops
prior addition of Stop solution
Plate is insufficiently
washed
Review manual for proper
wash technique. If using a
plate washer, check all ports
for obstructions.
Contaminated wash
buffer
Prepare fresh wash buffer
Low sensitivity
Improper storage of
the ELISA kit
Store your reconstituted
standards at -80°C, all other
assay components 4°C.
Keep TMB substrate solution
protected from light.
Precipitate in
Diluent
Precipitation and/or
coagulation of
components within
the Diluent.
Precipitate can be removed
by gently warming the
Diluent to 37ºC.
Poor standard
curve
Low Signal
Large CV
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
25
RESOURCES
20.
NOTES
ab210965 Mouse Myoglobin SimpleStep ELISA Kit
26
UK, EU and ROW
Email: technical@abcam.com | Tel: +44-(0)1223-696000
Austria
Email: wissenschaftlicherdienst@abcam.com | Tel: 019-288-259
France
Email: supportscientifique@abcam.com | Tel: 01-46-94-62-96
Germany
Email: wissenschaftlicherdienst@abcam.com | Tel: 030-896-779-154
Spain
Email: soportecientifico@abcam.com | Tel: 911-146-554
Switzerland
Email: technical@abcam.com
Tel (Deutsch): 0435-016-424 | Tel (Français): 0615-000-530
US and Latin America
Email: us.technical@abcam.com | Tel: 888-77-ABCAM (22226)
Canada
Email: ca.technical@abcam.com | Tel: 877-749-8807
China and Asia Pacific
Email: hk.technical@abcam.com | Tel: 108008523689 (中國聯通)
Japan
Email: technical@abcam.co.jp | Tel: +81-(0)3-6231-0940
www.abcam.com | www.abcam.cn | www.abcam.co.jp
Copyright © 2016 Abcam, All Rights Reserved. The Abcam logo is a registered trademark.
All information / detail is correct at time of going to print.
Discover more at www.abcam.com
Download