ENGINEERING-43 RC Series Circuits Lab-13 Lab Data Sheet – ENGR-43 Lab-13 Lab Logistics Experimenter: Recorder: Date: Equipment Used (maker, model, and serial no. if available) Directions 1. Check out a DMM an oscilloscope a Signal/Function Generator. Cables and Leads 2. Go to the side counter, collect resistors, a capacitor, “bread board”, and leads required to construct the circuit shown in Figure 1. See also Figure 2. 3. See the Instructor to use the LCR meter to measure the actual value of the Capacitor 4. Use the DMM to measure the actual value of the Resistor © Bruce Mayer, PE • Chabot College • 291211011 • Page 1 Figure 1 • RC Series Circuit. Vs = 14 Vpp (Vamplitude = 7V) or Vs = 7V0º f = per Table I or Table IV. R = 350-750 Ω (470 Ω nominally). C = 100 nF (0.1 µF) nominally. Digital-Meter Actual-Values R= C= 5. Make the Measurements and Calculations needed to complete Table I. Reveal the Vs-GND signal on the Scope Connect CHANNEL-1 between the Vs and GND nodes for the circuit shown in Figure 1 (measures Vs-GND) Press the CH1 Button to bring the signal onto the display Set up the waveform for each frequency using the TDS340 Oscope Set the Frequency, f: [MEASURE Menu] → [Frequency side menu-1, button-2]. See Figure 3. Set the Peak-to-Peak Amplitude, Vpp: [MEASURE Menu] → [Pk-Pk side menu3, button-3] Confirm the Period, T : [MEASURE Menu] → [Period side menu-1, button-1] Reveal the Vc-GND signal on the Scope Connect CHANNEL-2 between R and C (measures Vc-GND) Press the CH2 Button to bring the signal onto the display o Both the CH1 and CH2 traces should be simultaneously displayed Use the Scope to Measure the Capacitor Phase-Difference, , in Terms of TIME © Bruce Mayer, PE • Chabot College • 291211011 • Page 2 Expand the SEC/DIV scale on the scope to make a maximally precise measurement of the time-based phase difference o See the Instructor if you are unsure about this procedure Measure using the cursors – See Figure 4 and Figure 5. o Activate the Vertical-Bar cursors: [CURSOR button] → [Side Menu = V Bars] o Use the SELECT button and the GENERAL PURPOSE knob to position the Left cursor at the peak of the left-most trace. o Use the SELECT button and the GENERAL PURPOSE knob to position the Right cursor at the peak of the left-most trace. o Read the “Δ” measurement in the upper-left corner of the display Be Sure to Note if LEADS or LAGS the source which is DEFINED to have ZERO phase angle Measure Vc,pp using the cursors o Activate the Horizontal-Bar cursors: [CURSOR button] → [Side Menu = H Bars] o Use the SELECT button and the GENERAL PURPOSE knob to position the top cursor at the peak of the CH2 trace. o Use the SELECT button and the GENERAL PURPOSE knob to position the Bottom cursor at the trough of the CH2 trace. o Read the “Δ” measurement in the upper-left corner of the display Use The DMM in “TRUE RMS” mode to measure the rms values for V & I Be sure readjust the amplitude every time the frequency is changed Table I – Capacitance Measurements Frequency, f T for Vc Vc,pp Vc,rms Ic,rms 1000 Hz 3333 Hz 10000 Hz 6. Perform the Calculations needed to complete Table II Use the Signal Period and the Capacitor Phase-Difference Time to calculate the capacitor phase ANGLE, , in DEGREES (°) relative to the base-line value for Vs: LEAD 360 sec T sec LAG Using measured values calculate: |XC| = Vrms/Irms From measured Vrms & Irms values Calculate C from the expression for XC |XC| = Vrms/Irms = 1/(C) Use the calculation and the LCR meter data to determine the % for the Capacitor as C-% = 100x(Ccalc – CLCR)/CLCR © Bruce Mayer, PE • Chabot College • 291211011 • Page 3 Table II – Capacitance Calculations for Vc Frequency, f |XC| Ccalc C-% 1000 Hz 3333 Hz 10000 Hz 7. Use the Digital-Meter Actual Values for R & C, and the rms & scope measurements to complete Table III. Specifically CALCULATE the Capacitor impedance, Zc (specify in RECTANGULAR form): Zc 1 j 1 1 j jX c , LCR Note that X c jC C C C Single-loop phasor circuit current Ic using Ohm’s law (specify Ic in POLAR form): Ic Vs Vs Z series R Z c Single-loop phasor Voltage across the Capacitor, Vc, by the Voltage Divider method (specify Vc in POLAR form): Vc Vs Zc Zc Vs Z series R Zc All Phase Angles MUST stated in degrees (°) To Calculate the Δ%’s, Let “Q” = Ic or Vc, and “Brms“ = Irms or Vrms, then: Q% Q 2 Brms Brms Table III – Impedance-Based Calculations Frequency, f Zc Ic Vc 1000 Hz 3333 Hz 10000 Hz © Bruce Mayer, PE • Chabot College • 291211011 • Page 4 Ic-% Vc-% 8. Make the Measurements and Calculations needed to complete Table IV Use the Scope to measure Vc,pp Use the Scope MATH menu MEASURE DIRECTLY VR,pp Press the MATH button o On the side menu press button-2 to activate the Ch1-Ch2 measurement With the MATH WaveForm selected Press the MEASURE Button o On the side menu press the button needed to activate the Pk-Pk function. See Figure 6. Use the Scope CURSOR function to measure the TIME-based phase difference, , for Vc as was done in step 5 Expand the SEC/DIV scale on the scope to make a maximally precise measurement of the time-based phase difference Be sure to note whether is LEADING or LAGGING Using determine the Capacitor Phase Angle in DEGREES (°) Table IV – Series RC Potential Measurements Sweep Frequency, f Vs,pp 100 Hz 14V0º 400 Hz 14V0º 1250 Hz 14V0º 2.5 kHz 14V0º 8.0 kHz 14V0º 25.0 kHz 14V0º 80 kHz 14V0º Vc,pp VR,pp for Vc for Vc . Note: Low frequencies such as 100Hz can be difficult to measure on older scopes 9. Use MATLAB or EXCEL to create two SemiLog plots of the data contained in Plot the Data Table IV. In both plots the frequency, f, will be plotted on the Logarithmic scale Plot-1 Independent variable = log(f) TWO dependent variables on the same plot: Vc,pp and VR,pp Plot-2 Independent variable = log(f) Dependent variable = Attach both plots to this lab report ANALYZE the trends shown in the plots, and comment on the physical CAUSE of the observed trends HINT: Consider the Behavior of the Circuit in these extreme cases o →0 o →∞ © Bruce Mayer, PE • Chabot College • 291211011 • Page 5 10. Return all lab hardware to the “as-found” condition Figure 2 • Connection of the circuit components and instrumentation. © Bruce Mayer, PE • Chabot College • 291211011 • Page 6 Figure 3 • Using the TDS430 MEASURE functions to set the frequency of the CGF250. © Bruce Mayer, PE • Chabot College • 291211011 • Page 7 Figure 4 • Using the TDS430 CURSOR function to measure the sign time difference, . In this case the space between the traces must be Expanded using the HORIZONTAL SCALE to improve the measurement accuracy. © Bruce Mayer, PE • Chabot College • 291211011 • Page 8 Figure 5 • Using the TDS430 CURSOR function to measure the sign time difference, . When compared to Figure 4 note how the horizontal scale has been expanded from 50 μs to 25 μs to increase the peak-to-peak distance to improve the measurement accuracy. See Figure 5. © Bruce Mayer, PE • Chabot College • 291211011 • Page 9 Figure 6 • Using the TDS430 MEASURE functions measure directly the Ch1–Ch2 signal difference. Run Notes/Comments Print Date/Time = 29-May-16/04:00 © Bruce Mayer, PE • Chabot College • 291211011 • Page 10