r \ : OMEGA MODEL 873 Hand Held Digital Thermometer

advertisement
r \ : OMEGA
^ ^
^ ^ H EIMGIIMEERIIMCIIMC.
FiyniiyFFDii\iG i i u r
I n 0>lf."f» t l i r w u p C o m p a n y
MODEL 873
Hand Held Digital
Thermometer
Operator's Manual
WARRANTY
OI\^EGA warrants this unit to be free of defects in materials and workmanship
and to give satisfactory service for a period of one (1) year from date of purchase.
If the unit sfiould malfunction, it must be returned to ttie factory for evaluation.
Our Customer Service Department will issue an Authorized Return (AR) number
immediately upon phone or written request. Upon examination by OMEGA, it
the unit is found to be defective it will be repaired or replaced at no charge.
However, this WARRANTY is VOID if the unit shows evidence of having been
tampered with or shows evidence of being damaged as a result of excessive
current, heat, moisture, vibration, or misuse, (Components which wear or which
are damaged by misuse are not warranted. These include contact points, fuses,
and triacs.
THERE ARE NO WARRANTIES EXCEPT AS STATED HEREIN. THERE ARE NO
OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT
LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND OF
FITNESS FOR A PARTICUUR PURPOSE. IN NO EVENT SHALL OMEGA
ENGINEERING, INC. BE LIABLE FOR CONSEQUENTIAL, INCIDENTAL OR
SPECIAL DAMAGES. THE BUYER'S SOLE REMEDY FOR ANY BREACH OF THIS
AGREEMENT BY OMEGA ENGINEERING, INC. OR ANY BREACH OF ANY
WARRANTY BY OMEGA ENGINEERING, INC. SHALL NOT EXCEED THE
PURCHASE PRICE PAID BY THE PURCHASER TO OMEGA ENGINEERING, INC.
FOR THE UNIT OR UNITS OR EOUIPMENT DIRECTLY AFFECTED BY SUCH
BREACH.
Return Requests/Inquiries
Direct all warranty and repair requests/inquiries to OMEGA Customer Service
Department, telephone number (203) 359-1660. Before returning any instrument, please
contact the OMEGA Customer Service Department to obtain an authorized return (AR)
numtier. The designated AR number should then be marked on the outside of the return
package.
To avoid processing delays, also please be sure to Include:
1. Returnee's name, address, and phone number.
2. Model and Serial numbers.
3. Repair instructions.
i^OMEGA
A .
. ^ H ENGiniEERING, IIMC
One Omega Drive, Box 4047
Stamford, Connecticut 06907-(X)47
(203)359-1660 Telex: 996404 Cable: OMEGA
FAX: (203)359-7700
OMEGA'-' is a registered trademark of OMEGA ENGINEERING, INC.
-, Copyright 1984 OMEGA ENGINEERING, INC. Alt rights reserved including
illustrations. Nothing in this manual may be reproduced in any manner, either wholly
or in part for any purpose whatsoever without written permission from
OMEGA ENGINEERING, INC.
Printed in U.S.A.
M230/025
TABLE OF CONTENTS
GENERAL INFORMATION
Unpacking and Inspection
Optional Accessories
Preparation for Use
Safety Symbols and Terms
Environmental Conditions
OPERATION
Safety Precautions
SERVICING INFORMATION
Functionality Check
Performance Verification
Environmental Conditions
Disassembly Instructions
Battery Installation/Replacement
Calibration
Probe Matching
Troubleshooting . . . :
Special Handling of Static Sensitive Devices
Component Layout
Schematic Diagram
1
1
2
3
4
4
5
6
9
9
9
10
10
12
13
15
15
18
21
23
GENERAL INFORMATION
The OI^EGA® IVIodels 873C and 873F are Hand Held Digital Thermometers. Both instruments have large easy to read liquid crystal displays
(LCD). The input connection consists of a subminiature, type K thermocouple, female connector (SMP-K-F). This input accommodates any
OMEGA type K thermocouple probe fitted with an SMP-K-M connector. Cold junction electronic circuitry in the 873F/873C automatically
compensates for ambient temperature changes. These instruments
also provide piece-wise linearization conformity for thermocouple nonlinearity. The instruments are powered by a 9V battery. Continuous
operation of 350 hours is typical with an alkaline battery or 200 hours
typical with a carbon-zinc battery.
The Model 873C measures temperature from - 55°C to 1000°C on
two ranges. The 200°C range measures temperature from -55.0°C to
199.9°C with a resolution of 0.1°C. The 1000°C range measures
temperature from -55°C to 1000°C with a resolution of 1°C. The accuracy is ± (0.25% rdg -i- 1.0°C) from -40°C to 1000°C and ± (0.25%
of rdg -l-2°C) for -40°C to -55°C.
The Model 873F measures temperature from - 60°F to 200°F on two
ranges. The 200°F range measures temperature from -60.0°F to
199.9°F with a resolution of 0.1°F. The 2000°F range measures
temperature from -60°F to 1999°F with a resolution of 1°F. The accuracy of the Model 873°F is ±(0.25% of rdg -i-2.0°F).
UNPACKING A N D INSPECTION
Remove the Packing List and verify that all equipment has been
received. If there are any questions about the shipment, please call
OMEGA Customer Service Department at (203) 359-1660.
Upon receipt of shipment, inspect the container and equipment for
any signs of damage. Take particular note of any evidence of rough
handling in transit. Immediately report any damage to the shipping
agent.
NOTE
The carrier will not honor any claims unless all shipping
material is saved for their examination. After examining and
removing contents, save packing material and carton in the
event reshipment is necessary.
OPTIONAL ACCESSORIES
The following optional accessories can be used with the Models 873C
and 873F to enhance the performance of the instrument. These
accessories are available:
Part Number
Description
KS871
Tilt Stand/Belt Clip/Probe Holder; can
be used as a tilt stand or clipped to
your belt. It will hold one compact
probe.
SC800
Soft Carrying Case
880A2
Deluxe Carrying Case
MN1604
9V Replacement Battery
SMP-K-M
Subminiature Thermocouple Connector, Male
SMP-K-F
Subminiature Thermocouple Connector, Female
OT-201-1
1 oz jar of High Thermal Conductivity
Paste for fast surface measurements.
PREPARATION FOR USE
The Models 873F and 873C are supplied ready for use with battery
installed. Connect the appropriate probe to the miniature TC connector located on the top of the instrument (just above the display).
SAFETY SYMBOLS A N D TERMS
The symbol / ! \ on the instrument denotes that the user should refer
to the opjerating instructions.
The WARNING used in this manual explains dangers that could result
in personal injury or death.
The CAUTION used in this manual explains hazards that could
damage the instrument.
ENVIRONMENTAL CONDITIONS
Operation of the Models 873C and 873F should take place at an ambient temperature of -10°C to 50°C (14°F to 122°F), less than 80%
relative humidity up to 35°C, lineariy derate 3% R.H./C from 35°C to
60°C.
OPERATION
The 873C and 873F Hand Held Digital Thermometers are easy to use
There is only one control (located on the side of the instrument) that
turns the instrument on/off and selects the range. (See Figures 1 and
2). Use the following procedure to operate the instrument:
1. Turn on the power to the instrument by selecting the desired range.
Model 873C:
A. 200°C range is the center position of the ON/OFF range switch.
B. 1000°C range is the upper position of the ON/OFF range switch.
Model 873F:
A. 200°F range is the center position of the ON/OFF range switch.
B. 2000°F range is the upper position of the ON/OFF range switch.
2. Select the appropriate temperature probe and insert it into the input
jack.
3. Make the measurement:
A. Touch the probe tip to the material to be measured.
B. Read the display. The reading is in Celsius for the Model 873C.
The reading is in Fahrenheit for the Model 873F.
NOTE
An open sensor or overrange condition is indicated by:
CAUTION
Do not attempt to measure temperatures beyond
the range of the probe being used. Probe damage
may occur.
WARNING
A shock hazard exists on the Models 8730 and 873F
Input jacks when probes or sensors are exposed to
voltage levels greater than 42V peak to earth ground.
SAFETY PRECAUTIONS
1. Do not touch the probe tip when measuring excessively high or low
temperature.
2. Do not exceed 42V peak from the probe tip or input jack to earth
ground.
Miniature TC Connector
OMEGA 873C THERMOMETER
Low Battery Indicator
LO BAT
-loao
Minus Sign Displayed •
Plus Sign Implied
3V4 Digit D i s p l a y —
200
On/Off Range Switch
TYPE K THEHMOCOUPLE
NiCr-N<AI
RANOE: - I B ' C to lOOO'C
^
42V MAX - ^
Figure 1. Model 873C Control Layout
4
Miniature TC Connector
OMEGA 873F THEHMOMETER
Low Battery Indicator^
Minus Sign DisplayedPlus Sign Implied
VA Digit Display-
r DM I
-\QO.U
^=
On/Off Range Switch
TYPE K THEIIMOCOUPI.E
NiCr-NiAl
RANOE. -tO'E to 2000*F
A 42V MAX
+
Figure 2. Model 873F Control Layout
^
SERVICING INFORMATION
This section contains information needed to maintain the Models
873C and 873R The following information is included: functional
check, performance verification, disassembly instructions, battery installation/replacement, calibration, probe matching, troubleshooting,
parts list, schematic diagram and component layout.
WARNING
The information presented in this section is intended for use by qualified personnel only. Turn off
the power and remove all probes from the instrument before replacing the battery.
FUNCTIONALITY CHECK
At room temperature hold the tip of a probe sensor between your
thumb and index finger. A reading between 25°C and 40°C (77°F and
104°F) should be observed to confirm functionality of the instrument.
PERFORMANCE VERIFICATION
Accuracy Check
1. Prepare a distilled water ice bath or use an OMEGA* TRC III Ice
Point Reference Chamber.
2. Connect a TC probe to the input jack of the instrument.
3. Immerse the TC probe into the TRC III or distilled water ice bath
and allow 10 minutes for thermal stabilization,
4. Turn on the instrument and use Table 1 to verify that the readings
on each range are within specifications.
Table 1. Performance Verification
Range
873°C
200
0.0 ± 1°C
0 ± 1°C
1000
2000
873°F
32,0 ± 2°F
32 ± 2°F
ENVIRONMENTAL CONDITIONS
The Models 873C and 873F should be at an ambient temperature of
23°C ±1°C and a relative humidity of less than 80%.
DISASSEMBLY INSTRUCTIONS
WARNING
Turn off the instrument and disconnect the
temperature probe before removing the bottom
cover.
NOTE
When disassembling the instrument, handle the PC
board by the edges. Body oil, dirt, etc can degrade circuit
performance. Keep the PC board clean and free of contaminants.
1. Place the unit face down on a bench or other similar surface and
remove the screws from the bottom cover.
2. Separate the bottom cover from the rest of the instrument by
grasping the top of the case (just above the display) and carefully
lifting it away from the display.
3. Remove the battery.
4. Remove the standoff (ST-166-1) that secures the PC board to the
case. The standoff is located in the center of the PC board.
5. Remove the switch cap.
6. Lift the PC board out of the top cover (front panel)
7. Remove the shield.
8. The LCD is held in place on the PC board by a shock resistant
mounting and is assembled as shown in Figure 3.
9. To reassemble the unit, reverse the above procedure.
CAUTION
When removing the LCD from the PC board be careful
not to spread the four retaining hooks on the clear plastic
bezel out too far. The plastic bezel hooks may fracture.
WARNING
When reassembling the instrument, be sure to
reinstall the switch cover. Common mode voltage
may be present on the switch creating a possible
hazard if the cover Is not installed.
f U a K 7 / V RLUSTtn
a HEQUIRED)
BOTTOM COVER
Figure 3. Model 873 Exploded View
BATTERY INSTALLATION/REPLACEMENT
To install or replace the battery:
1. Remove the bottom cover of the instrument (refer to the
disassembly instructions).
WARNING
Turn the instrument off and disconnect the input
probe before removing the bottom cover. Reinstall
the cover before resuming usa of the instrument.
2. Remove the old battery.
3. Place the new battery in the battery compartment. Be sure to
observe the proper polarity. Refer to Figure 4.
4. Reinstall the bottom cover before resuming use of the instrument.
BArrenv TERMINALS
Rgure 4. Battery Installation
CALIBRATION
Required Test Equipment
1. DC voltage calibrator wrth ±0.01% of setting ±20/iV accuracy
2. 100: 1 divider- 1.98k and 20n wirewound resistors, 0.02% ratio
accuracy or better.
3. OMEGA TRC III Ice Point Reference Chamber or distilled water ice
bath.
4 . OMEGA type K TRP Reference Probe or make a thermocouple
Transition junction (see Calibration Set-Up).
Calibration Set-Up
1. To make a thermocouple transition junction, silver solder a piece
of copper wire to each (type K) thermocouple wire and insulate the
junction with a waterproof jacket.
2. Connect the thermocouple wires from the type K transition junction to an OMEGA Miniature Thermocouple Connector SMP-K-M.
3. Turn on the voltage calibrator.
4 . Plug in the miniature connector (see steps 1 and 2) into the 873
input.
5. Connect the copper wires to the 100:1 divider as shown in Figure
5. Connect the copper/chromel wire to the positive sida
6. Connect the 100:1 divider to the DC voltage calibrator as shown.
7. Use an OMEGA TRC III or prepare the distilled water ice bath as
follows:
A. Drill a hole in the cap of a Dewar flask or Thermos™ to
accommodate the thermocouple,
B. Firmly pack the flask with pea-size ice cubes made from
distilled water and then fill the flask with distilled water.
10
C. Replace the melted ice with more ice while removing the excess
water.
D. Place the cap on the flask. Immerse the thermocouples into the
flask and allow 20 minutes for temperature stabilization.
DC CALIBRATOR
100:1 VOLTAGE DIVIDER
ICE BATH
873
*The 20(1 resistor must be at a uniform temperature during calibration.
Figure 5. Set-Up Procedure
Calibration Procedure
1. Remove the bottom cover as described in the disassembly instructions.
WARNING
To prevent a shock hazard, turn the instrument off
and remove the Input probe from the instrument
before removing the bottom cover.
2. Turn on the Model 873 and perform the calibration adjustments
listed in Table 2 (873C) or Table 3 (873F). For access to calibration adjustments refer to Figure 4.
11
Table 2. Modal 8730 Calibration Adjustments
Step
1
2
3
4
Adjustment
Zero
-i-Gain ( x l )
+ Gain(x10)
-Gain
Range
200°C
1000°C
200°C
200°C
Calibrator Trimmer Desired
Setting
Adjust Reading
0.00000
R112
00.0°C
+ 41.269mV
R115
1000°C
-H7.139mV
R104
175.0°C
-1.527mV
R105
-40.0°C
Table 3. Model 873F Calibration Adjustment
Step Adjustment
1
Zero
-i-Gain(xl)
2
3
+ Gain(x10)
4
-Gain
Range
200°F
2000°F
200°F
200°F
Calibrator Trimmer Desired
Setting
Adjust Reading
-0.692mV
R112
00.0°F
-i-44.645mV
1990°F
R115
3.589mV
190.0°F
R104
-1.929mV
-60.0°F
R105
PROBE M A T C H I N G
To improve the accuracy of the Model 873 while using a particular
proJTe, calibration can be adjusted so that probe errors in the region
of zero (0.0°C or 32°F) are compensated. Place the probe in an ice
bath and allow the reading to stabilize. Then reset the zero adjustment
(R112) to a 0.0°C or 32°F reading.
TROUBLESHOOTING
The following troubleshooting information is intended to be used by
qualified electronic maintenance personnel who are familiar with the
proper use of standard electronic test equipment.
To gain access to the PC board, refer to the disassembly instructions.
Use the parts list, schematic diagram and component layout to identify parts and check point locations.
13
DC Voltage Checks
1. Battery: V B A T > 7 . 5 V
2. -HV to (Analog Ground) = -1-2.9 ±0.3V
3. -fV to (Digital Ground) = +5V ±1V
4. Reference Diode
CR101 (Anode) to (Analog Ground) = -1.23V ±0.03V.
5. Negative reference divider (R114) referenced to analog ground. *
Pin 12
Pin 10
Pin 11
Pin 7
Pin 9
Pin 8
Pin 6
Pin 4
Pin 5
Pin 3
Pin 2
873C
= -1.23 V
= -0.209V
= -0.209V
= -0.206V
= -0.204V
= -0.204V
= -0.201V
= -0.200V
= -0.197V
= -0.195V
= -0.193V
Pin 12
Pin 11
Pin 10
Pin 9
Pin 8
Pin 7
Pin 6
Pin 5
Pin 4
Pin 3
Pin 2
873F
= -1.23 V
= -0.209V
= -0.208V
= -0.205V
= -0.204V
= -0.203V
= -0.198V
= -0.197V
= -0.196V
= -0.191V
= -0.188V
*The setting of Rl 15 and the tolerances of the resistors within the network (R114) can affect these levels by ±5%.
6. Cold Junction Voltage
(Ambient Temperature = 25°C ±3°C)
Across R102A approximately equal to ImV.
7. Waveform Checks (see Figure 6)
U104 pin 40 (OSC 1)
U104 pin 39 (OSC 2)
U104 pin 38 (OSC 3)
U104 pin 21 (Backplane)
13
U104 PIN 40 (OSC 1)
U104 PIN 39 (OSC 2)
H
25|isec
- ^ '
K\/
U104 PIN 38
K
= 25/(sec
5V
U104 PIN 21
M
= 20msec
BV
Figure 6. Waveform Checks
14
SPECIAL H A N D L I N G OF STATIC SENSITIVE DEVICES
CMOS devices are designed to function at high impedance levels.
Normal static charge can destroy these devices. Table 4 lists all the
static sensitive devices for the Models 873C and 873F. Steps 1
through 7 provide instructions on how to avoid damaging these
devices.
Table 4. Static Sensitive Devices
Schematic
Designation
U104
U105
U107
U108
U109
1. Devices should be handled and transported in protective containers, antistatic tubes or conductive foam.
2. Use a properiy grounded work bench and a grounding wriststrap.
3. Handle devices by the body only.
4. Printed circuit boards must be grounded to bench while inserting
devices.
5. Use antistatic solder removers.
6. Use grounded tip soldering irons.
7. After devices are soldered or inserted into sockets they are protected and normal handling can resume.
15
SPECIFICATIONS
MODEL 873F SPECIFICATIONS
TEMPERATURE SENSOR: Type K INICr-NiAl) Thermocouple,
RANGE
TEMPERATURE
SPAN
RESOLUTION
ACCURACY*
IBS to SyP; 2 years)
200°F
2000°F
-60.0° to 199.9°
- 6 0 ° to 1999°
0.1°
1^
±(0,25% rdg -f-2.0°F)
± (0.25% rdg -t- 2°F)
•Includes NBS confofmity, coW junction compensation, temperature coefficient 165° to
82<'FI and repeatability errors. Excludes therrtiocouple errors; however thermocouple errors around 32°F may be compensated for by an intemal zero adjustment,
REPEATABILITY: 0.3°F typical for 1 week at constant ambient
temperature.
TEMPERATURE COEFFICIENT: 65° to 82°F; included in accuracy
specification. From 14° to 65° and 82° to 122°F: less than ±(0,025%
of rdg -(-0.2°F)/°F,
INPUT CURRENT: 25nA typical, 50nA maximum.
MODEL 8730 SPECIFICATIONS
TEMPERATURE SENSOR: Type K (Ni&NiAl) Thermocouple,
RANGE
200°C
1000°C
TEMPERATURE
SPAN
RESOLUTION
ACCURACY*
(18 to 28°C: 2 yeara)
-40.0° to 199.9°
-55.0° to -40.0°
- 4 0 ° to 1000°
- 5 5 ° to - 4 0 °
0.1°
0.1°
1°
If
± (0.25% rdg -t- 1.0°C)
±(0.25% rdg -t- 2.0°C)
±(0.25% rdg -I- 1°C)
± (0.25% rdg + 2°C)
"Includes NBS conformity, cold junction compensation, temperature coefficient
(18° to 28°C) and repeatability errors. Excludes thermocouple errors; however tfiermocouple errors around 0°C may be compensated for by an intemal zero adjustment,
REPEATABILrrV: 0.2°C typical for 1 week at constant ambient
temperature.
TEMPERATURE COEFFICIENT: 18° to 28°C; included in accuracy
specification. From - 1 0 ° to 18° and 28° to 50°C: less than ±(0.025%
of rdg -(-0.1°C)/°C.
INPUT CURRENT: 25nA typical, 50nA maximum.
Specifications subject t o change without notice.
16
MODEL 873F and 873C GENERAL SPECIFICATIONS
(unless specified):
DISPLAY: 314 digit LCD, 13mnfi (0.5") height. Polarity and decimal point
indication.
CONVERSION RATE: 2.5 readings per second.
OVERRANGE AND OPEN SENSOR INDICATION: 3 least significant
digits blanked,
MAXIMUM COMMON MODE VOLTAGE: 42V peak to earth.
NORMAL MODE REJECTION RATIO: Greater than 40dB at 50 & 60Hz.
COMMON MODE REJECTION RATIO: Greater than 120dB at DC, 50 &
60Hz,
ENVIRONMENTAL LIMITS FOR OPERATING (Model 873F): 14° to
122°F, less than 80% relative humidity up to 95°F; lineariy derate 1.5%
R.H./°F from 95° to 122°F.
ENVIRONMENTAL LIMITS FOR OPERATING (Model 873C): - 1 0 ° to
50°C, less than 80% relative humidity up to 35°C; lineariy derate 3%
R,H./°C from 35° to 50°C.
ENVIRONMENTAL LIMITS FOR STORAGE (Model 873F): - 3 0 ° to
140°F, less than 90% relative humidity up to 95°F; linearly derate 1.5%
R,H,/°F from 95° to 140°F,
ENVIRONMENTAL LIMITS FOR STORAGE (Model 873C): - 3 5 ° to
60°C, less than 90% relative humidity up to 35°C, linearly derate 3%
R.H./°C from 35° to 60°C.
THERMOCOUPLE LINEARIZATION: Multislope A/D with 11 piecewise
linear segments.
THERMOCOUPLE INPUT: Miniature TC connector.
POWER: 9V alkaline or carbon-zinc (NEDA 1604) battery.
BATTERY LIFE, CONTINUOUS: 350 hours typical with alkaline; 200
hours typical wrth carbon-zinc.
BATTERY INDICATOR: Display indicates "LO BAT" when less than 10%
of life remains.
DIMENSIONS: 160mm long x 69mm vinde x 31mm thick (6,3" x
2.7" X 1.2"). Net weight 210gm (7.5 oz.),
CONSTRUCTION: Heavy duty ABS plastic housing.
ACCESSORIES SUPPLIED: Battery, Instruction Manual and Model
TC-GG-K-24-36-SMP-f\/l Thermocouple Sensor
17
Model 8730 and 873F, Parts List
Sch
Desig.
Sch
Location
Description
9V Alkaline Battery, NEDA 1604
Capacitor, .^^lf, 50V, Ceramic
Capacitor, .l/iF, 50V, Ceramic
Capacitor, 33pF, 1000V, Ceramic
Capacitor, .VF, 50V, Ceramic
Capacitor, .IjjF, 50V, Ceramic
Capacitor, 47pF, 500V, Silver Mica or
Ceramic
Capacitor, .33f»F, 63V, Metalized
0107
Polyester
0108
Capacitor, .047^F, 50V, Metalized
Polypropylene
0109
Capacitor, 4.7/JF, 25V, Aluminum
Electrolytic
0110
Capacitor, .33MF, 63V, Metalized
Polyester
OR101 Silicon Diode, 1N914
J1001 Miniature TC Connector, Female
J1002 Battery Clip
J1003 Battery Clip
Transistor, NPN, 2N3904
Q101
Transistor, NPN, 2N3904
Q102
Transistor, PNP, 2N3906
Q103
Resistor, 10k, 5%, 1/4W, Composition
R101
R102* Thick Film Resistor Networic
R102** Thick Rim Resistor Network
Resistor, 10M, 5%, 1/4W, Composition
R103
R104* Potentiometer, Ik
R104** Potentiometer, 50012
Potentiometer, 2k
R105
Thick Film Resistor Network
R106
R107
Resistor, 191k, 1 % , 1/8W, Metal Film
Resistor, 390k, 5%, 1/4W, Composition
R108
R109
Thick Film Resistor Network
Factory Selected Resistor (Part of
R110
A31419)
Factory Selected Resistor (Part of
Rill
A31410)
R112
Potentiometer, 2k
Factory Selected Resistor (part of
R113
A31419)
BT101
C101
C102
C103
0104
C105
C106
18
Fl
Bl
C2
82
G1
G1
E2
F2
E2
G5
Dl
E3
A2
Fl
Fl
B2
82
E3
A2
Sev
Sev
Dl
Dl
Dl
D2
Sev
F2
F2
Sev
83
83
83
C3
Modei 873C and 873F, Parts Ust (Cont.)
Sch
Desig.
R114*
RIM""
R115
R116
R117
R118
S101
U101
U102
U103
U104
U105
U106
U107
U108
U109
VR101
W101
Description
Thick Film Resistor Network
Thick Film Resistor Networi<
Potentiometer, 2k
Resistor, 100k, 5%, 1/4W, Composition
Resistor, 100k, 5%, 1/4W, Composition
Resistor, 121k, 1 % , 1/8W, Metal Film
Switch, 4P3T, On/Off Range Switch
Op Amp, LM308 (Factory selected; part
of A31419)
Current Source, LM334 (Factory selected;
part of A31410)
Low Power JFET Op Amp, TL062Cp
3-1/2 Digit A/D Converter, ICL7136
Quad 2 Input NOR Gate, 4001
Dual Synchronous Up Counter, 4518
12 Stage Binary Counter, 4020
Analog Multiplexer, 4051
Analog Multiplexer, 4053
Reference Diode, 1.23V
Jumper Wire
Sch
Location
Sev
Sev
F4
G4
86
Dl
Sev
C2
03
D2/G5
F2
Sev
05
05
D5
G5
85
•Model 873F
"Model 8730
NOTE: Mechanical Parts can be found in the expk>ded view (Figure 3).
•
19
SCHEMATIC DIAGRAM
+T]
! «f;r
a=
»a
^
^
il
15 ^» i -
i
ir^
'^H-
^""
I 11
8
*i
tiM
5» I
RII3(2fe7K)_
OR
RII3 (I2IK)-
n 1
SIOI
CIO!
ED
JIOO2
n n
JI003
o
o
TJ
o
z
(ZDCI02
•o
m
JIOOI
C109'
C106 Q
iftiilj—
Rmz>
RIQ&
RIGS
SIDE 2
JIOOI
^
" ^
RII5
5 H-y >
O
00
O
5
<
O
WIOl
UI06
c
ut
OiVIEGA^...
YOUR SOURCE FOR
SENSING AND CONTROL
This is the FREE
complete reference
book you will want to
keep within reach at
all times. Send for your
LATEST COPY now!
HMDilOOKAiio
BIEYiaOFEIllA
One Omega Drive, Box 4047
Stamford, CT 06907-0047
(203)359-1660 Telex: 996404
FAX: (203)359-7700
Phone: (203)359-1660
/^OMEGA
l A ^ . A H ENGiiUEERiNG. IIMC.
A n OIWEGA G r o u p Company
"The WORLD'S Leader in the Field of
Sensing and Control"
n
D
n
D
n
D
D
a
D
n
D
D
D
n
D
D
D
D
Thermocouple Probes and Assemblies
Industrial Thermocouple Assemblies
Infrared and Hand Held Instruments
Hand Held and Surface Thermocouple Probes
RTD Probes and Assemblies
Thermistor Probes and Assemblies
Connector Systems and Panels
Wire: Thermocouple, RTD and Thermistor
Cements, Coatings and Epoxles
Calibrators and Ice Point References
Analog Meters and Amplifiers
Digital Thermometers
Temperature and mV Recorders
Solid State Controllers and Relays
Digital Process Monitors
Data Acquisition Modules and Data Loggers
Computer Sensor Interface
Technical and Reference Books
Return Requests/Inquiries
Direct all warranty and repair requests/inquiries to OMEGA Customer
Service Department, telephone number (203)359-1660, Before
returning any instrument, please contact the OMEGA Customer
Service Department to obtain an authorized return (AR) number. The
designated AR number should then be marked on the outside of the
return package.
To avoid processing delays, also please be sure to include:
1. Returnee's name, address, and phone number.
2. Model and Serial numbers.
3. Repair instructions.
jncOMEGA
i ^
. A H ENGINEERING, INC.
One Omega Drive, Box 4047
Stamford, CT 06907-0047
(203)359-1660 Telex: 996404
FAX: (203)359-7700
Cable. OMEGA
Download