Design of a Programmable Filter For Macromolecules by Byron Miguel Stancil 8ARr'e OF TECHNOLOGy 0 f 2 b L1BRARIES B.S.E., University of Maryland Baltimore County (1999) Submitted to the Department of Mechanical Engineering in partial fulfillment of the requirements for the degree of Master of Science in Mechanical Engineering at the MASSACHUSETTS INSTITUTE OF TECHNOLOGY June 2002 @ Byron Miguel Stancil, MMII. All rights reserved. The author hereby grants to MIT permission to reproduce and distribute publicly paper and electronic copies of this thesis document in whole or in part. Author ........ .. Departnzint of Mechanical Engineering May 24th, 2002 Certified by ........... Accepted by ................ 7 Kamal Youcef-Toumi Professor of Mechanical Engineering Thesis Supervisor ...... Ain A. Sonin Chairman, Department Committee on Graduate Students j Room 14-0551 MITLibraries Document Services 77 Massachusetts Avenue Cambridge, MA 02139 Ph: 617.253.2800 Email: docs@mit.edu http://libraries.mit.edu/docs DISCLAIMER OF QUALITY Due to the condition of the original material, there are unavoidable flaws in this reproduction. We have made every effort possible to provide you with the best copy available. If you are dissatisfied with this product and find it unusable, please contact Document Services as soon as possible. Thank you. The images contained in this document are of the best quality available. Design of a Programmable Filter For Macromolecules by Byron Miguel Stancil Submitted to the Department of Mechanical Engineering on May 24th, 2002, in partial fulfillment of the requirements for the degree of Master of Science in Mechanical Engineering Abstract The focus of this thesis is the design of a device that separates biologically-active macromolecules by particle size. The final apparatus design functions by pumping molecules in an aqueous solution between two surfaces with flatness on the nanoscale. The varying gap width between the two surfaces will determine what molecules will pass through to the solution collector. A controller is used to change the gap width by utilizing two piezoactuators and readings from two capacitance probes. The goal of this project is to be able to develop a new method of particle separation utilizing the best qualities of present methods and eliminating their worst qualities. Particles should be filtered very quickly without any contaminants on a particle range of 0.5 nanometers to 0.5 microns. Thesis Supervisor: Kamal Youcef-Toumi Title: Professor of Mechanical Engineering 2 Acknowledgments I would like to extend my gratitude to various people for their support. My advisor, Professor Kamal Youcef-Toumi, has mentored me throughout my stay here. I have benefited greatly from his guidance, knowledge and persistence in seeing this project to its completion. Dr. Manzooh Shah, CEO of Alpine Pharmaceutical Co., sponsored and guided this project from the beginning. Furthermore, Dean Isaac Colbert, Dean Blanche Staton, Associate Dean Roy Charles, Brima Wurie, Ed Ballo, George Brennan and Heather Fry in the Graduate Student Office provided financial, emotional, academic and spiritual support in many ways. They've been like family to me here. Professors George Barbastathis, Jung-Hoon Chun, Dave Pritchard, Peter So, Alex Slocum, and David Trumper provided invaluable insight and/or assistance from their labs and through conversations during the course of this project. I also want to thank Gerry Wentworth, Stephen Haberek and everyone else from the Central Machine Shop, the Laboratory for Manufacturing and Production and the Pappalardo Undergraduate Laboratory for allowing me to utilize their machine shops and for their insight and assistance. I cannot forget Alex Cronin, Patrick Anquetil, James Tangorra, Bryan Crane, Phoebe Kwan and Andrew Stein, whom have assisted me in many different ways here. Also my colleagues in the d'Arbeloff's Laboratory have provided social and academic support. Especially, Bernardo Aumond, Osamah El Rifai, Vidi Saptiri, Eric Hoarau, Belal Helal, and Eric Wade have assisted me extensively throughout my stay here. I want to thank all my family, friends, and colleagues from MIT, UMBC, Maryland and many other places for their "prayers and wishes". I cannot forget the members of the Black Graduate Student Association for all their love and support. Plus, I cannot forget Leslie Regan, Joan Kravit, and Carolyn Skeete for going beyond the call of duty to assist me in many ways. Of course, I would like to thank Malo Huston, Reginald Hutchinson, and Mike Johnson for being there for me in many ways. Besides being the best of colleauges to me, they've been like brothers to me. I cannot go any further without giving thanks to Treena Boyd, my lovely fiance,, and my parents for 3 all their love, encouragement and emotional and spiritual support. Finally, but first and foremost, I would like to thank God, because without him, all of this would not be possible. 4 Contents 1 15 Introduction 1.1 Background . . . . . . . . . . . . . . 15 1.2 Objectives and Technical Issues . . . 16 1.3 Filtration Categories . . . . . . . . . 16 1.4 Conventional Methods of Separation 17 1.5 Approach: Mechanical Programmable Filter (MP F) 19 1.6 1.5.1 Theory of MPF . . . . . . . . 19 1.5.2 Advantages of MPF . . . . . . 20 1.5.3 Challenges of MPF . . . . . . 20 Thesis Outline . . . . . . . . . . . . . 21 25 2 Design Alternatives 2.1 Introduction . . . . . . . . . . . . . . . 25 2.2 Dual Flexure Design . . . . . . . . . . 25 2.2.1 Material Selection . . . . . . . . 25 2.2.2 Flexure Descriptions . . . . . . 26 2.2.3 Flexure Design Process..... 26 2.3 2.4 Tubular Filtration . . . . . . . . . . . . 28 2.3.1 Material Selection . . . . . . . . 28 2.3.2 Tubular Description . . . . . . 28 2.3.3 Tube Filtration Process . . . . 28 Implementation/Sealing of Fused-Silica Quartz Plates 2.4.1 Major Issues . . . . . . . . . . . 5 29 29 2.5 2.6 3 2.4.2 Sealant Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 2.4.3 Fixture Design . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Fluid Delivery System . . . . . . . . . . . . . . . . . . . . . . . . . . 30 2.5.1 Reynolds Number . . . . . . . . . . . . . . . . . . . . . . . . . 30 2.5.2 Inlet Pressure Calculations . . . . . . . . . . . . . . . . . . . . 31 2.5.3 Outlet Pressure Calculations . . . . . . . . . . . . . . . . . . . 32 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Summary 43 Experimental Setup 3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 3.2 Experimenting with the Probes and Piezoactuators . . . . . . . . . . 43 3.2.1 Capacitance Probes . . . . . . . . . . . . . . . . . . . . . . . . 43 3.2.2 Piezoactuators . . . . . . . . . . . . . . . . . . . . . . . . . . 45 3.2.3 System's Resolution, Range, and Noise . . . . . . . . . . . . . 46 . . . . . . . . . . . . . . . . . . . . . . . 47 3.3.1 Characterization of System . . . . . . . . . . . . . . . . . . . . 47 3.3.2 Dynamic Characteristics . . . . . . . . . . . . . . . . . . . . . 48 3.3.3 Designing Controller . . . . . . . . . . . . . . . . . . . . . . . 50 3.3.4 DSPACE Control . . . . . . . . . . . . . . . . . . . . . . . . . 52 3.3 3.4 DSPACE Controller Design Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69 4 Experimental Results 5 53 4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69 4.2 Flow Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69 4.3 Actuation and Sensing Test . . . . . . . . . . . . . . . . . . . . . . . 70 4.4 Filtration Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71 81 Conclusion and Recommenations A Company Addresses 89 B Schematics 91 6 107 C Controller Design C.1 Sensor, Piezoactuator, and Dynamic Signal Analyzer Data . . . . . . D Material List and Component Characteristics 7 107 129 8 List of Figures 1-1 Electrodialysis process [9] . . . . . . . . . . . . . . . . . . . . . . . . 22 1-2 Reverse osmosis process [20] . . . . . . . . . . . . . . . . . . . . . . . 22 1-3 High pressure liquid chromatography process [19] . . . . . . . . . . . 23 1-4 Capillary electrophoresis process [22] . . . . . . . . . . . . . . . . . . 23 1-5 Plates filtering solution (orange and green are large particles, respectively .) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 2-1 Side view of dual flexure design . . . . . . . . . . . . 38 2-2 Top view of dual flexure design . . . . . . . . . . . . 38 2-3 Different types of notch hinges: a) circular, b) elliptic, and c) leaf[23] 39 2-4 Filter tube design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 2-5 Stress versus strain plot of an elastic material [4] . . . . . . . . . . . 40 2-6 Hysteresis plot [3] . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 2-7 MPF final filtration design . . . . . . . . . . . . . . . . . . . . . . . . 41 3-1 Digital picture of probe . . . . . . . . . . . . . . . . . . . . . . . . . . 56 3-2 Micrometer stage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 3-3 Voice coil in a speaker [2] . . . . . . . . . . . . . . . . . . . . . . . . 57 3-4 Digital picture of piezo-amplifier . . . . . . . . . . . . . . . . . . . . . 57 3-5 Sensor output versus generator input (stiffness test) . . . . . . . . . 58 3-6 Simulink model of open loop "unlumped" system . . . . . . . . . 58 3-7 Simulink model of open loop "lumped" system . . . . . . . . . . . . . 59 3-8 Piezoactuator B and flexure F / sensor D 3-9 Piezoactuator B and flexure F / sensor 9 . magnitude bode diagram D phase bode diagram . . . 59 60 3-10 Piezoactuator A and flexure E / sensor D slope in magnitude bode diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 / sensor D E / sensor C 3-11 Piezoactuator A and flexure E slope in phase bode diagram 61 3-12 Piezoactuator A and flexure magnitude bode diagram . 61 . . . . 62 3-13 Piezoactuator A and flexure E/ sensor C phase bode diagram 3-14 Piezoactuator A and flexure E / sensor C slope in magnitude bode diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . / 62 sensor C slope in phase bode diagram 63 3-16 Simulink model of a closed loop system . . . . . . . . . . . . . . . . . 63 3-17 Asymptotic curves for basic terms of a transfer function [10] . . . . . 64 3-15 Piezoactuator A and flexure E 3-18 The optimum coefficients based on the ITAE criterion for a step input [10] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65 3-19 Root locus of top system . . . . . . . . . . . . . . . . . . . . . . . . . 65 3-20 Root locus of bottom system . . . . . . . . . . . . . . . . . . . . . . . 66 3-21 DSPACE controller model . . . . . . . . . . . . . . . . . . . . . . . . 66 3-22 DSPACE card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67 3-23 DSPACE panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67 3-24 Sensor output with DSPACE controller . . . . . . . . . . . . . . . . . 68 . . . . . . . . . . . . . . . . . . . . . . . . 4-1 Filtration tube 76 4-2 Sensor output(V) versus piezoelectric output from amplifier 4-3 Magnitude Bode diagram for MPF tube filtration design 77 4-4 Phase Bode diagram for MPF tube filtration design . . . . 77 4-5 Root locus of closed loop system 1 . . . . . . . . . . . . . . 78 4-6 DSPACE controller for filration experiment . . . . . . . . . 78 4-7 Sensor output and square wave input for open loop . . . . 79 4-8 Sensor output from square wave input for open loop . . . 79 4-9 Sensor output from square wave input for closed loop . . . 80 (V) 76 B-1 Isometric view of small flexure A (mm) . . . . . . . . 92 B-2 Front view 1 of small flexure A (mm) . . . . . . . . . 92 10 B-3 Front view 2 of small flexure A (mm) . . . . . . . . . . . . . . . . . . 93 B-4 Top view of small flexure A (mm) . . . . . . . . . . . . . . . . . . . . 93 B-5 Side view of small flexure A (mm) . . . . . . . . . . . . . . . . . . . . 94 B-6 Isometric view of large flexure B (mm) . . . . . . . . . . . . . . . . . 94 B-7 Front view of large flexure B (mm) . . . . . . . . . . . . . . . . . . . 95 B-8 Top view of small flexure B (mm) . . . . . . . . . . . . . . . . . . . . 95 B-9 Side view of small flexure B (mm) . . . . . . . . . . . . . . . . . . . . 96 B-10 Isometric view of piezo-holder M (inches) . . . . . . . . . . . . . . . . 96 B-11 Front view 1 of piezo-holder M (inches) . . . . . . . . . . . . . . . . . 97 B-12 Front view 2 of piezo-holder M (inches) . . . . . . . . . . . . . . . . . 97 B-13 Side view of piezo-holder M (inches) . . . . . . . . . . . . . . . . . . . 98 B-14 Top view of piezo-holder M (inches) . . . . . . . . . . . . . . . . . . . 98 . . . . . . . . . . . . . . . 99 . . . . . . . . . . . . . . . . . 99 . . . . . . . . . . . . . . . . . . 100 B-18 Top view 1 of Sensor-Holder N (inches) . . . . . . . . . . . . . . . . . 100 B-15 Isometric view of sensor-holder N (inches) B-16 Front view of sensor-holder N (inches) B-17 Side view of sensor-holder N (inches) B-19 Top view 2 of Sensor-Holder N (inches) . . . . . . . . . . . . . . . . . 101 . . . . . . . . . . . . . . . . . 101 . . . . . . . . . . . . . . . . . . . 102 B-22 Side view of tube claw 0 (inches) . . . . . . . . . . . . . . . . . . . . 102 B-23 Top view of tube claw 0 (inches) . . . . . . . . . . . . . . . . . . . . 103 B-24 Isometric view of sensor contact P (inches) . . . . . . . . . . . . . . . 103 B-20 Isometric view of tube claw 0 (inches) B-21 Front view of tube claw 0 (inches) B-25 Front view 1 of sensor contact P (inches) . . . . . . . . . . . . . . . . 104 B-26 Front view 2 of sensor contact P (inches) . . . . . . . . . . . . . . . . 104 B-27 Side view of sensor contact P (inches) . . . . . . . . . . . . . . . . . . 105 B-28 Schematic of model 2800 series probe [1] . . . . . . . . . . . . . . . . 105 B-29 Piezoactuator Schematic [7] . . . . . . . . . . . . . . . . . . . . . . . B-30 Spherical top piece (steel) on moving end [7] . . . . . . . . . . . . . . C-1 Noise plot for sensor output . . . . . . . . . . 11 106 106 109 C-2 Sensor output with bottom piezoactuator at -30 Vdc . . . . . . . . . 109 C-3 Sensor output with top piezoactuator at -30 Vdc . . . . . . . . . . . . 110 C-4 Sensor output with bottom piezoactuator at 0 Vdc . . . . . . . . . . 110 C-5 Sensor output with top piezoactuator at 0 Vdc . . . . . . . . . . . . .111 C-6 Sensor output with bottom piezoactuator at 1 Vdc . . . . . . . . . .111 C-7 Sensor output with top piezoactuator at 1 Vdc . . . . . . . . . . . . . 112 C-8 Sensor output with bottom piezoactuator at +150 Vdc . . . . . . . . 112 . . . . . . . . . . 113 . . . . . . . . . 113 C-11 Sensor output with top piezoactuator at +1 Vdc . . . . . . . . . . . . 114 C-12 Piezoactuator B and flexure F 114 C-13 Piezoactuator / sensor C magnitude bode diagram . B and flexure F / sensor C phase bode diagram . . . . B and flexure F / sensor C slope in magnitude bode 115 diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115 C-15 Piezoactuator B and flexure F/ sensor C slope in phase bode diagram 116 C-16 Piezoactuator A and flexure E/ sensor D magnitude bode diagram . . 116 C-9 Sensor output with top piezoactuator at +150 Vdc C-10 Sensor output with bottom piezoactuator at +1 Vdc C-14 Piezoactuator C-17 Piezoactuator A and flexure E / C-18 Piezoactuator A and flexure E sensor D phase bode diagram . . . . / sensor D slope in magnitude bode diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-19 Piezoactuator A and flexure E / sensor 12 117 117 D slope in phase bode diagram 118 List of Tables 2.1 Parts list A 2.2 Aluminum comparison ..... ... ... .. .. .. ... .. .. 2.3 Flexure design results A . . . . . . . . . . . . . . . . . . . . . . . . 35 2.4 Flexure design results B ..... .... .. ... . .. ... .. .. 2.5 Flexure design results C . . . . . . . . . . . . . . . . . . . . . . . . 35 2.6 Parts list B for filter tube . . . . . . . . . . . . . . . . . . . . . . . . 36 2.7 Parts list C for MPF . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 6 2.8 Reynolds number results . . . . . . . . . . . . . . . . . . . . . . . . . 3 6 2.9 Pressure results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 35 35 2.10 Fluid definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 7 . . . . . . . . . . . . . . . . . . 54 . 54 3.1 Sensor outputs 3.2 Amplifier noise outputs versus voltage output 3.3 Sensor outputs and noise due to piezoactuator A versus amplifier volt... age output................. 3.4 .... . . . .p. . . . . . .. 55 Sensor outputs and noise due to piezoactuator B versus amplifier voltage output . . . . . . . . . . . . . . . . . . . . . 55 4.1 Filtration tube chart . . . . . . . . . . . . . . . . . . . . . . . . . . . 73 4.2 Open loop response due to a square wave input . . . . . . . . . . . . 74 4.3 Closed loop response due to a square wave input . . . . . . . . . . . . 74 4.4 Filtration test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 C. 1 Elasticity test of system . . . . . . . . . . . . . . . . . . . . . . . . . 108 13 C.2 Dynamic signal analysis of flexure design la . . . . . . . . . . . . . . 119 C.3 Dynamic signal analysis of flexure design lb . . . . . . . . . . . . . . C.4 Dynamic signal analysis of flexure design 2a 120 . . . . . . . . . . . . . . 121 C.5 Dynamic signal analysis of flexure design 2b . . . . . . . . . . . . . . 122 C.6 Dynamic signal analysis of flexure design 3a . . . . . . . . . . . . . . 123 C.7 Dynamic signal analysis of flexure design 3b . . . . . . . . . . . . . . 124 . . . . . . . . . . . . . . 125 C.9 Dynamic signal analysis of flexure design 4b . . . . . . . . . . . . . . 126 C.10 Dynamic signal analysis of MPF 1 . . . . . . . . . . . . . . . . . . . . 127 C.11 Dynamic signal analysis of MPF 2 . . . . . . . . . . . . . . . . . . . . 128 P arts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130 D.2 Probe characteristics [1] . . . . . . . . . . . . . . . . . . . . . . . . . 130 C.8 Dynamic signal analysis of flexure design 4a D .1 D.3 Piezoactuator characteristics [7] . . . . . . . . . . . . . . . . . . . . . 14 131 Chapter 1 Introduction 1.1 Background In the pharmaceutical industry, a major issue is the separation of biologically active macromolecules. The present methods take too long, are too costly, or have contamination issues. So, the industry is racing to find a new method of separation. Normally, you need a means of detecting, collecting, and purifying samples in three different stages. Hopefully this coupled-process can be eliminated or the combination of some of these processes can be accomplished. Usually, the sample, the medium or both are stained with some type of chemical coloring. This process affects the purification because the coloring and other contaminants have to be separated from the sample too. In collecting the sample, an intrusive means is not desired because contaminants could be added to the sample and/or the sample could be physically damaged. An attraction method could be used where the sample is collected on a film. Furthermore present methods are not efficient. Gels are reliable but still interfere with decontaminating samples. Using radiation and electricity (i.e., ultraviolet rays or electric charges) possibly has effects on the properties of molecules. 15 1.2 Objectives and Technical Issues The goal of this project is to design and construct an Autonomous Robotic Purification System (ARPS) using unconventional methods for purification of biologically active macromolecules. Gels are not desired in this project. Purification will be accomplished by separating particles by diameter size. Particles of diameters of 0.5 nanometers to 0.5 micrometers are desired for separation. Overall, the goal is to find a method that reduces the number of steps for purification, minimizes process time, eliminates contaminants and byproducts, records and stores data, and has a remote control function from a computer through the web (teleoperation). The maximum dimensions for this device are approximately 11 centimeters x 11 centimeters (4.33 inches x 4.33 inches). The positive aspects of High Performance Liquid Chromatography (HPLC) and Capillary Electrophoresis (CE) are desired. Serious thought must be put into how samples are held, how the device will interface with the computer, what type of software to use, what types of experiments to conduct, building a device from scratch versus coupling devices, and the design and layout. This project has to have universal application. Plus, FDA approval is also needed for implementation into the pharmaceutical market. 1.3 Filtration Categories There are five categories for filtration where the properties of the molecules are used to exploit the concept. Properties of interest are molecular mass and weight, charge, and particle size. The major areas are macro, micro and submicron particles [6]. Particles between 50-100 microns in size are macro particles. Particle filtration is applicable which includes pre-coat and depth filters and screen. These particles are usually visible to the naked eye. Examples of these macro particles are sand, hair, pollen, flour, and mist. An optical microscope is needed to see the next category of particles, micro particles. These range in size from 0.05 microns to 2 microns. Examples of these particles are red blood cells, coal dust, yeast cells and latex. Microfiltration is 16 used for filtering these particles. Surface, depth, and pleated filters are used here, as well. Submicron particles are broken down into subdivsions of macro molecular, molecular, and ionic particles. All of these methods involve semipermeable membrane usage. In the macro molecular and molecular range, particle sizes range from 0.05 microns to 1 micron and 0.002 microns and 0.02 microns, respectively. The macro molecular range, dimension-wise, is the small resolution end of the micro range. It is mentioned because on the high end of the micro range, semipermeable membranes are not always necessary. These particles are viewed using a scanning electron microscope (SEM). Microfiltration and ultrafiltration is utilized to separate macro molecular and molecular particles such as tobacco smoke, asbestos, various paint pigments and some bacteria, and synthetic dyes, viruses, and endotoxins, respectively. Ionic particles are usually under 0.001 microns in size and viewed with a scanning tunneling microscope (STM). Hyperfiltration methods, such as reverse osmosis, are used to separate aqueous salts, metal ions, and atomic radius-sized particles. Distillation and deionization are other means of filtering ionic particles but these methods are not optimal due to the high amount of chemicals and energy required. When coupled with reverse osmosis, cost can be minimized using distillation and deionization. 1.4 Conventional Methods of Separation Many conventional methods of separation are used in the pharmaceutical field. Hydrolysis is the splitting of a compound into fragments by the addition of water; the hydroxyl group being incorporated in one fragment and the hydrogen atom in the other [5]. In biological application, water is added to a starch to break the chains to get glucose. H 2 SO4 is the catalyst and NaOH (salt) is the byproduct. Also, membranes can be implemented into this method to increase the amount of separation. Even though this method is one of the simplest, the problem with this process is that too many contaminants are present. First, the water added to the sample must be checked for purity. Second, the byproduct, salt, must be filtered out, which adds 17 another step to the process. Most importantly, the catalyst, H 2 SO4 is not desired because it is a very strong solution. Ultrafiltration uses cell membranes to separate large and small molecules. The only problem with this method is that you have to clean the sample due to the gel [27]. Electrodialysis is an electric current-induced process where a solution is separated into its ionic components, cations (negatively charged particles) and anions (positively charged particles) [9]. Once separated, these particles migrate through a membrane that allows cations or anions to pass through. The membranes that allow anion or cations to pass are cation- or anion-exchange membranes, respectively. Usually, two oppositely charged electrodes are on placed on both ends of the current path. Between the two electrodes, multiple metal plates, which are the cation and anion exchange membranes, are placed to gather the ions at different locations in a tank or container. Depending on the type of membrane, anions and cations will gather on its opposite sides while some will gather on the electrodes. The accumulation of a salt will be on one side while cations and anions will be on another. So there will be a removal of salts and water, leaving a diluted solution in certain set of chambers while a concentrated solution will be in the ions' chamber. Problems with this method include plate erosion from the salt byproduct and it only separates by particle charge (particle size and charge do not always correlate). Figure 1-1 shows the Electrodialysis Process. Reverse osmosis utilizes the osmotic pressure, which is an applied pressure that must overcome the chemical driving potential driving force [20]. This action results in the pure solvent being driven from the solution to the other side of a semipermeable membrane. This process is "reverse" osmosis because in "regular" osmosis, the solvent naturally flows through a membrane into solution comprised of a solvent and a solute. Usually, reverse osmosis is utilized in the desalination of water where fresh water is driven from a solution (i.e., salt water). For the separation of various particles, you have to have more than one type of membrane. So, this method does not have universal application. Figure 1-2 shows the Reverse Osmosis Process. Chromatography is the process of separating a biological mixture into its individual components due to their physical (i.e., structure) and chemical (i.e., composition) 18 differences to the solvent in the mobility phase and the column packing in the stationary phase [19]. Physical differences are based on molecular property based methods that use charge (Ion Exchange) and size (gel permeation and size-exclusion). Usually, the separation of the solution is based on their mobility or separation speed through some sort of membrane or porous media. Two types of chromatography are manual and mechanical (i.e., liquid and gas, respectively). Mechanically High Pressure Liquid Chromatography (HPLC) applies a pressure and separates due to mobility and charge. If high pressures are involved, the structure must be designed properly to accommodate for the forces. In addition, various types of mediums are needed to separate a wide range of molecules. Figure 1-3 shows the High Pressure Liquid Chromatography Process. Electrophoresis uses an electric field to move ions and charged macromolecules through a medium [22]. The mobility of the molecules depends on molecular size, weight and shape of particles, charge carried, applied current and medium's resistance. In most setups, molecules move faster if they are smaller and highly-charged. Newer methods used in electrophoresis are Capillary Gel Electrophoresis and Capillary Isoelectric Focusing Electrophoresis. CE brings speed, quantitation, reproducibility, and automation to the conventional method. This method is one of the more optimal choices but due to the use of membrane, contamination and universal application are still major factors. Figure 1-4 shows the Capillary Electrophoresis Process. 1.5 Approach: Mechanical Programmable Filter (MPF) 1.5.1 Theory of MPF MPF follows a simple methodology. Two parallel plates, with extremely flat contact surfaces, are adjusted so that various particles can move between them in an aqueous solution. The plates are moved by some means of actuation to be discussed later in the thesis. After being calibrated, a sensor is used to check the gap width between 19 the plates. Then the sensor readings will correlate to a certain gap width. Figure 1-5 shows the plates filtering a solution. 1.5.2 Advantages of MPF As mentioned before, MPF methodology is easy to understand. Secondly, the system will require more of an understanding of mechanical engineering instead of too many disciplines (excluding of course the interaction of the particles at the gap). In addition, a mechanical system will be more robust than an optical or chemical because of less electronic interference and stability due to environmental changes (i.e., temperature, lighting, vibrations, etc.). Since it is a physical and tangible system, this methodology hopes to be more intuitive; easier to comprehend and apply. Furthermore, no mechanical means of separation has been used on the nanoscale. 1.5.3 Challenges of MPF Even though the methodology is simple, this method has many challenges. First, the contact surfaces of the plates need a "flatness" less than 0.5 nanometers for this project to be applicable across the desired range. Secondly, the plates must be aligned "perfectly" so that only the desired particles pass through the gap. In addition, on a nano-scale, small temperature gradients may affect the reliability of the plates' gap (i.e., a temperature change of less than one Fahrenheit may affect the system.). Another factor to consider is pressure. For such a small orifice, a large pressure may be required to drive and/or suck the particles through the gap. The system will have to be designed properly to withstand the pressure distributed through its entire system. Even excluding everything else, particle behavior under high pressures and at the interface will be unknown. Since the plates cannot be attached to the actuators, it will be necessary to design some means of distributng and possibly amplifying the displacement. The design step for moving the plates is critical for proper resolutions and stability. 20 1.6 Thesis Outline In design alternatives, two concepts are investigated for application in this project. Also, this section involves the structural design and experimental process for filtration. The experimental setup section covers means of sensing, actuation, and control. Then, the experimental results sections uses all the information from the previous section to test the methodology of filtration for this project. Conclusions and recommendations are given to give a summary of the project and to discuss recommendations for future research using this concept. 21 8r*~e ________________________ I S-ot 10 be IQ-iof d I A I AAnd C A t A I, I I A I, A 00 - - -I I I I N-Z -6- I a I I I I I &an De-lonied Sot Figure 1-1: Electrodialysis process [9] REVERSE OSMOSiS OSMOTIC EQUILIBRIUM NORMAL OSMOSIS P OSMOTIC PRESSURE FRESH WATER SALINE WATER SEMIPERMEABLE MEMBRANE Figure 1-2: Reverse osmosis process [20] 22 Iw -Is _____ D1frnntlta migration -- A Mobil. phase (Mn) C As Stationary phase Cs' Figure 1-3: High pressure liquid chromatography process [19] U S Electroosmotic flow ueo 0 0 0C -Sj' 0Sl-+S 0 0 p 0 [22 30808 Figure 1-4: Capillary electrophoresis process [22] 23 Wl'' I eW Figure 1-5: Plates filtering solution (orange and green are large particles, respectively.) 24 Chapter 2 Design Alternatives 2.1 Introduction This chapter will look at two approaches for filtrating biological solutions. One method transmits displacements through a flexure element while another uses a flexible element. While both have similarities, they have their own advantages and disadvantages. At the end of this chapter, one of the methods will be selected as the optimal filtration design. In addition, a fluid mechanic analysis will be performed to see if sufficient flow can be obtained with feasible parameters for the design. 2.2 2.2.1 Dual Flexure Design Material Selection A dual flexure design was designed for this project. Ramco Machine Inc. used the ProE drawings for this project and machined a flexure from Al 7079-T6 (Figure 21 and Figure 2-2). Aluminum was chosen because it is a relatively inexpensive, malleable, and flexible metal. It had to be malleable enough to be machined easily and flexible enough to transmits displacements through the structure. Al 7079-T6 was chosen instead of Al 6061-T6, because the selected material is not as dense and stiff. With a lower modulus of elasticity, the design is more flexible for the desired 25 application but still stiff enough to resist external disturbances (i.e., small vibrations and noise). 2.2.2 Flexure Descriptions In order to change the gap width of the plates, flexure F is used to transmit the displacement from piezoactuator B. In case of a misalignment of the plates during assembly, piezoactuator A transmits motion through the flexure E in order to align the plates together by creating a yaw motion. The fixed plate is curved 2 degrees because this design eliminates the need for the plates to close "perfectly" face-to-face, making this a 2-dimensional motion control issue instead of a 3-dimensional case. 2.2.3 Flexure Design Process The first step in designing was to study different types of flexure elements (i.e., cantilevers, small pivotal designs, etc.). Initially, the maximum force from the piezoactuator, 500 Newtons, was divided by the maximum range of travel, 0.5 microns, to get a spring constant of 1E+9 N/m. Then, a ProE model was created and this number was implemented into that system and applied to many possible designs. For completion, a simulation in ProMechancia was run on the model to see if the range of travel was accomplished. The tolerances for this system are ± 1E-5 m. It is very important to get as close to these tolerances as possibly because 1. during machining, accumulating errors will occur as you make each side of a piece, 2. alignment is a very critical issue here and you want design errors to be minimized for optimal plate placement, 3. during assembly, accumulating errors will lead to misalignment issues. After looking at the behavior of various flexure elements, a specific type had to be chosen. For this project, a notch hinge leaf type flexure was desirable because of its popularity in the field of flexure elements; more information cited about this 26 flexure element [23]. A circular hinge geometry was selected to make machining easier; drilling two holes close to each other to form a small pivotal area or flexure element. Figure 2-3 shows the three different elliptic geomeotry types for notch hinges: a) circular, b) elliptic, and c) leaf. In designing notch type flexures, the angular stiffness can be derived from ([23]) (2.1) Kz = M Oz and 2 2*E*b*t/ Mz M * E 2**bax5 *( 9*7r Oz where Mz is the bending moment, 6 2 .2 ) is the angular deflection about the neutral z-axis, E is the elastic modulus, b is hinge depth, and a, is half the hinge length. The true stress at can simply be found with from the stress concentration factor Kt, the previous information, and the hinge thickness t to be ([23]) = K = 6 * Kt * M( bb ** t2 (1 + (2.3) (2.4) 0)9/20 and t (2.5) 2 * ax The angular deflection at the hinge can be derived from the translation necessary for the desired gap width. From this information, you can calculate the stress on the flexure element to see if the material will yield. The hinge stress E *Oz * (1+ ah 9/20) is ([23]) (2.6) f * #2 where f is a dimensionless compliance factor. Instead of solving for the compliance factor and the hinge stress, from the ay and Kt equations, t is ([23]) y=9 * a, * (7r * ay * *(4* E*Kt)2 27 oz)2(27 (2.7) Substituting the maximum angular deflection Omax for 0 and the yield stress ay in for o-, the maximum stiffness is ([23]) K Kz,max The numbers from Tables 2.3, b * a2 * E4 * Y)5 (19 * ir 4 * Kt * z) 5 (2.8) 2.4, and 2.5 were used to make sure that the dimensions of the flexures were obtainable for our experiment. 2.3 Tubular Filtration In this section, we will look at the design that utilizes a flexible element. With this design, we wanted to eliminate the sealant issues associated with the last design. So instead of the piezoelectric actuators directly distributing displacement through connected parts to the plates, the displacement is indirectly distributed through neighboring, non-connected, parts. Furthermore, any misalignment issues should be eliminated, hopefully, making this simplier than the 2-DOF criteria in the flexure design. 2.3.1 Material Selection Silicone and polypropylene tubing was selected for the main housing due to its flexibility, biologically/chemically inert characteristics, waterproof nature, and low cost. 2.3.2 Tubular Description Here is the assembly design for this project in Figure 2-4. In this method, the cross sectional area of a tubing will be changed in order to create some filtering points. 2.3.3 Tube Filtration Process This process simply filters the solution inside of a flexible tubing. An aqueous solution will enter the tubing and be filtered at a desired, internal location due to the change in cross sectional area. Since the tubing surface is not flat enough to filter small particles, some "filtering" pieces will be inside the tubing. A sealant will be used 28 to eliminate all external and internal leaks. As mentioned before, it will have to be highly flexible for displacemnt transmission form the actuators. The filtering parts will be sealed inside the tubing in a closed position. When actuation is applied to one side of the tubing's outer surface, this will cause the plates to move apart. Sensors will be placed parallel to a metal plate attached to the moving walls of interest, so that the gap width can be determined. 2.4 Implementation/Sealing of Fused-Silica Quartz Plates 2.4.1 Major Issues Both methods mentioned before will require filtering pieces to separate the particles. Fused silica quarz plates were chosen for the filtering parts. They're chemically and biologically inert to aqueous solutions and they can be optically-polished to a very fine flatness of 2-3nm. Due to the flatness limit, the resolution of this project raised from 0.5 to 10 nanometers. Implementation and sealing of the fused-silica quartz plates is probably the most critical portion of this project. The two fused silicaquartz plates (5mm x 12.5mm x 5mm) were implemented into the design last. A fixture for the plates had to hold them with no slippage or the filter resolution would be unreliable. Moreover, the contact between the fixture and the plates cannot have high stress concentrations, because the plates are very brittle due to their glass-like nature. Furthermore, a sealant is needed to come into contact with the fixture, plates, and surrounding regions to make sure there is no leakage. The applied sealant has to be elastic enough for the plates to move in any desired direction and return to their initial positions when there is no displacement applied. 2.4.2 Sealant Test In order to make sure that the sealant's bond to all moveable parts is really elastic, a stress versus strain test can be conducted (or load versus displacement). A voltage 29 would be applied to the piezoelectric actuators which would cause them to apply a displacement or load through the system. Then, the plates would move a certain distance and the analog outputs of the capacitance probes could be converted into a displacement. By varying the input voltage to a maximum value then lowering it back to zero, the output readings are converted into displacements and plotted. If the sealant is "truly" elastic, the results should look like Figure 2-5. Actually, the results will probably look like Figure 2-6. This test will be performed in the Results section. 2.4.3 Fixture Design Aquarium silicone was selected for the sealant. It is very elastic, waterproof and biologically/chemically inert. The solution seals water tanks, comes into contact with fresh and salt water, and does not bother the bio-environment of the aquarium (i.e., fish are alive and unaffected). Here is the assembly design for this project in Figure 2-7. 2.5 2.5.1 Fluid Delivery System Reynolds Number When studying fluidic systems, the type of flow has to be known. Before calculations are made, some assumptions must be stated [12]: 1. Steady flow 2. Incompressible flow 3. Flow along a streamline 4. No friction 5. Uniform velocity in regions before and after filtration 6. No streamline curvature at entrance regions, so pressure is uniform 30 To describe the behavior, one calculates the Reynolds Number, Re [12]. Re= V*D p*V*D (2.9) V P _Q (2.10) V = A 4*b*h D = 4*bh(2.11) 2(b+h) and L-_0.06*p*V*D D (2.12) A where p is the density, V is the flow velocity, D is the equivalent diameter for a rectangular orifice, b is the average gap width, h contact length of plates, A is the absolute viscosity, v is the kinematic viscosity, Q is the volume flow rate, L is the entrance length of the tube and A is surface area. If Re is less than 3000, flow is considered laminar. Turbulent flow occurs when Re is greater than 3000. Then Re,t is used [12]. 4*Q Re, (2.13) Table 2.8 shows the Reynolds Number results. The results show that a laminar flow is possible with feasible dimensions. NOTE: All calculations are based upon the minimum flow rate required of Q = 1.67E-11 1 S (equivalent to 1 mL per minute). 2.5.2 Inlet Pressure Calculations P1 -P 2 is the pressure needed to drive the fluid through the plates. It can be found by using [12] Patm- P, = p * g * (hi- h 2 ) (2.14) where Patm is the atmospheric pressure coming from the top of the tank, P is the pressure before the plates, p is the pressure of the medium, g is the acceleration due to gravity, and hi - h2 is the change in elevation from the regions before and after the 31 (assuming it is water), g = 9.81 m, and plates. With Patm = 101.3 kPa, p = 1000 1 h, - h2= 15 mm (approximately equal to the plate height plus a few millimeters), P = 101.15 kPa. 2.5.3 Outlet Pressure Calculations In order to find P2 , a relationship has to be stated between P and P2 . This part is where the use of fluid mechanics in an orifice becomes useful. Using Bernoulli's equation [12], Pi -- P1 + V12 2 2 + g1h1 = constant = P2 -- + V22 P2 2 + g2 h 2 (2.15) where P is a pressure, p is the density, V is the fluid velocity, h is the elevation height, and 1 and 2 refers to the inlet and outlet region, respectively, the continuity equation [12], V1A 1 V2 A 2 (2.16) m = p* V * A (2.17) = and the mass flow rate equation [12] the theoretical mass flow rate is [12] rnht= A 2 [( 2 * p(Pi - P 2 )) thA21 = _ (A9)2 0 5 (2.18) A1 With this equation, P2 and, evidently, P - P2 , the change in pressure equal to the pump's induced pressure, was calculated. Now the theoretical flow rate has to be compared with the actual. The calculated pressures will be used to do a reverse calculation with a different equation for flow rate. Using the discharge coefficient C [12], C = actualmassflowrate/theoreticalmassflowrate 32 (2.19) the actual mass flow rate is [12] (2.20) mhac =CAt[ (2p(Pi [1-B- P 2]))0. [1 - B34] 2.0 where At and Dt, the surface area and diameter at the plates, are equal to A 2 and D 2 when Dt is a very small compared to D 1 . C is obtained from [11] or by using the discharge coefficient equation for Re greater than 4000 [12] C = Cinf + b(2.21) ReDi * n where Cinf is the coefficient discharge at an infinite Reynolds number, Rei is the Reynolds number at D 1 , and b and n are scaling constants. A more general equation for coefficient discharge is [12] C = 0.5959 + 0.0312B2. 1 - 0.184B 8 + 91.71B 2 5 (2.22) (ReD1 )0.75 and B--- D (2.23) This discharge can be used to find and check the flow rates for the system. Listed in Table 2.9 are the results. NOTE: Before using any of the equations, the Reynolds number must be calculated because most of these equations are flow-dependent (laminar versus turbulent). 2.6 Summary From this chapter, the tubular filtration design was chosen over the dual-flexure design. Spacing and sealant issues were more of a concern with the latter. The tubular method minimized the amount of sealant needed, provided more sufficient space for assembly, minimized the alignment issue and displacement distribution issues through the materials. In addition from the fluid analysis, the actual and theoretical flow 33 rates differed but, feasible parameters are possible within the system. To compensate for this possible error, two variables pumps would be used in conjunction for this project. One would assist in pushing the fluid through the plates while the second would provide additional suction at the orifice. 34 Table 2.1: Parts list A Name A B C D E F Variable Top Piezoactuators Bottom Piezoactuators Master Sensor Slave Sensor Top Flexure Bottom Flexure Figure Location 2-1 2-1 2-2 2-2 2-1 2-1 Table 2.2: Aluminum comparison Al Types Al 6061-T6 Al 7079-T6 Density(kg/m 3 ) 2700 2740 Elastic Modulus(Pa) 7.31E+10 7.142E+10 Poisson Ratio 0.33 0.33 Table 2.3: Flexure design results A Variable M, Oz Kz Units radians N GPa Value 1335.056 1.5281 873.6577 b E m 71.42 m 5.02158E-2 Table 2.4: Flexure design results B Variable Units Values t ax Kt m 2.032E-3 m 2.921E-3 O-t GPa 1 M 1.1438 44.19 3.4783E-1 Table 2.5: Flexure design results C Variable Units Values ixh f Og ty Kz,max GPa N/A N/A N/A GPa 71.143 m 34.47E16 N*m ras 35 2.2045E+3 Table 2.6: Parts list B for filter tube Name G H I J K Variable Tube Housing Sealant Region Silica Plates Tube Claw Sensor Contact Table 2.7: Parts list C for MPF Name L M N 0 P Variable Tube Housing Piezo-Holder Sensor-Holder Tube Claw Sensor Contact Table 2.8: Reynolds number results P(f ) 1000 D (n) D(m) 5.1E-7 A(m 2 ) 2.04E-13 V( ) 6.55E-3 b(m) h(m) 255E-9 12E-3 N*s V(2 Q( ) p ) vf) 1E-3 1E-6 Re Re,t 3.34E-3 N/A n 1.67E-11 L(m) 1.02E-10 Table 2.9: Pressure results 9.81 rth~f 1.67E-11 B 2.68E-5 h-h 2 (m) 15E-3 Patm(Pa) 101.3E+3 A,1(M2) 2.85E-4 C 0.5959 P2 (Pa) 101.15E+3 rmac(m) 7.12E-21 36 P,(Pa) 101.15E+3 D1 (m) 0.01905 N/A N/A Table 2.10: Fluid definitions Units N/A p V Definition Reynolds number Fluid density Fluid velocity Q Volume flow rate A D p v b h L Re,t hrh2 Surface area Orifice diameter Absolute viscosity Kinematic viscosity Average gap width Plate contact length Entrance length Turbulent Reynolds number Atmospheric pressure Pre-plate pressure Post-plate pressure acceleration due to gravity Change in elevation height above and below plates Th Mass flow rate nht D_ A1 Di Theoretical mass flow rate Actual mass flow rate Surface area at Plates Diameter at Plates Surface area at Plates Diameter at Plates C C Cinf n ReDi Variable Re Patm P1 P2 g mac At M Values 3.34E-3 1000 6.55E-3 : 1.67E-11 m2 m 2.04E-13 5.1E-7 1E-3 1E-6 2.55E-7 1.2E-3 1.02E-10 N/A 101.3 101.15 101.15 9.81 15E-3 y N/A m S m m m N/A kPa kPa kPa 1.67E-11 m2 m m2 7.12E-21 N/A N/A m 2.85E-4 1.905E-2 Discharge coefficient N/A 05959 Discharge coefficient Discharge coefficient at Large Re N/A N/A 05959 N/A Scaling constant Reynolds number at D 1 N/A N/A N/A N/A 37 Figure 2-1: Side view of dual flexure design Figure 2-2: Top view of dual flexure design 38 a) b) (half c) widt) (Lpth) r 14 0 Y, 61 ' 14 Z~ Figure 2-3: Different types of notch hinges: a) circular, b) elliptic, and c) leaf[23] F u. 2 using Silica 0 Pla s I or t Tubae Claw ,J Figure 2-4: Filter tube design 39 Sa an* Sion B (77 -- -- M 6 Figure 2-5: Stress versus strain plot of an elastic material [4] r P A / r F, /)0 Figure 2-6: Hysteresis plot [3] 40 Sensor-Holder N D( Trub C Plezo-Holder M Figure 2-7: MPF final filtration design 41 42 Chapter 3 Experimental Setup 3.1 Introduction Now that the design and calculations are done for the filtration process, the controller must be implemented in order for the system to be programmable. This step will involve implementing a sensing and actuation element to make the system a closedloop system. Due to the interest of time and the ongoing design process on the Tube Filtration method, the dual-flexure was used to check the functionality of the sensors and actuators and for familiarization in designing a controller. All Tube Filtration information pertaining to the sensors and probes will be mentioned in the Results and Conclusion section. Therefore this entire chapter is geared towards implementation of the dual-flexure for analysis and observation purposes. 3.2 Experimenting with the Probes and Piezoactuators 3.2.1 Capacitance Probes To obtain the gap width, capacitance sensors were implemented. Capacitance sensors depend on the equations for "the ratio of charge to potential difference" and "potential difference between two plates" [26]. These equations are listed below, respectively 43 [26]. C= - (3.1) V =Q*d (3.2) V and eo * A So eo * A * V2 C (33) where C is the capacitance, Q is charge magnitude, V is voltage between the plates, d is the gap width, A is plate surface area, and e, is a universal constant. The 3800 Model OEM Gaging system was ordered for this project. Its maximum bandwidth is five kHz but this setting was lowered to 100Hz so high frequency noise would not interfere with the system. The 2805-1 probes were ordered for measurements. Their range was +/- 50 microns (or 100 microns). After proper calibration, the range of the probes corresponds to +/- 10 Vdc from the analog output. Figure 3-1 shows the Digital Picture of Probe. If calibrated correctly, -10 V (or +10 V) is the near standoff distance, zero V is the nominal standoff distance, and +10 V (or -10 V) is the far standoff distance. So, as the gap width goes from 0.5 microns to 0.5 nanometers, the analog output should become more negative. If the calibration is correct, then the calibration factor is ±50pm ±l0Vdc _ 100pum 20Vdc _ 5pum Vdc (34) Therefore, 0.5 nanometers should be achieved at 0.1 mVdc. This system has two capacitance probes which are used differently for the two different designs. Using the dual flexure design shown in Figure 2-1, if the sensors' analog outputs are not equal, then the two edges of the flexure and plates are not parallel. So, the sensors are used to check if the plates are parallel. Then, the yaw for alignment can be controlled by looking at the analog outputs. Secondly, the gap width can be determined by using the calibration factor. From the analog output, the 44 gap width can be determined and varied. With the Tube Filtration method, the two plates will be moved from a closed position. So the absolute change in output from both probes will give a value that correlates to the gap width using the calibration factor. In calibrating the probes, a micrometer incorporated in a stage was used to move the sensors. Figure 3-2 shows the Stage, Micrometer, Mounting Piece and Intermediate Piece. 3.2.2 Piezoactuators In choosing a means of actuation, the two choices were between a voice coil and a piezoactuator. A voice coil was not selected because they have too much vibration and electrical noise. Figure 3-3 shows the voice coil in speaker. Due to vibrations, a voice coil would have to be slightly modified for this application. Usually, the permanent magnet is fixed to the housing while the voice coil's core is allowed to move in the vertical direction. In this application, the core would need to be fixed and the magnet would be free to move vertically. The magnet's heavy weight would aid in vibration reduction. Since a piezoactuator would require no modifications, this means of actuation is more desirable. A PZT 150/4/20 VS09 and 150/5/20 VS1O stack piezoactuators were chosen for this application. They both had a 20-micron stroke, a maximum input voltage of +150 Vdc and a mechanical compressive load of 50ON-1000N, depending on which actuator is used. See Table D.3 for all primary specifications. The actuators were powered by a triple channel amplifier with an input and output voltage range of -1 to +5 Vdc and -30 to +150 Vdc, respectively. From the maximum distance and voltage input, the resolution is 2O"' +150V or = 1.33E-lm V nanometers, is obtained at 3.75 mVdc. 45 133.33nm +lvdc So the desired resolution, 0.5 3.2.3 System's Resolution, Range, and Noise After calibration, the linearity of the probes was checked by varying the frequency of a frequency generator and reading the outputs from the probes. The input signal was fed into the piezoactuator B from Figure 2-1. Table 3.1 shows the sensor outputs at various frequencies, but Figure 3-5 shows the data on a plot. From this plot, the information looks "linear" and reliable over a small voltage range, 15 Vdc. Now that the piezoactuators' and capacitance probes' functions have been checked and both are ready to be implemented in either design, now one must check to see if the desired resolution and range of the system are obtainable. In addition, noise in the system must be checked to see if it is negligible. In order for the noise to be negligible, the peak-to-peak voltage reading from analog outputs of the probes must be less than 0.1 mVdc. The manufacturers use a voltmeter and measure the output voltage Vac for noise. From the purchased merchandise, when the two electronics are coupled, a t 1 mVac, equivalent to a 10 nanometer error. Then, the peak-to-peak Vdc was plotted from an oscilloscope. ± 2 mVac was the input noise from the power source in the breadboard. Figure C-1 shows the noise plot for sensor output. In the following analysis, the data is differentiated from outputs produced by piezoactuators A and B from Figure 2-1, which corresponds to the top and bottom, respectively, and outputs from the master and slave probes, C and D, respectively, shown in Figure 2-2. The master probe is connected to the power source, while the slave gets its power from the master. In addition, the slave's electronics had to synchronized to the master's. As shown in Figure C-1, a peak-to-peak noise of 2 mVdc, equivalent to a 10 nanometer error, came from both probes. The master probe was closer to the parellel plate at -0.892 Vdc, while the slave was at 0.852 Vdc. These voltage readings were the initial location of the probes. There was no motion involved in this test. Since, a peak-to-peak noise of 2 mVdc is equivalent to +/-1 Vdc, the voltmeter and the oscilloscope gave the same error. The noise found in the system is probably from the power source but that needs to be tested. 46 To test the sensor output at full range it was time to incorporate the piezoactuators. First, the output noise of the amplifier was tested with a power supply at a 13.618 Vdc setting. The peak-to-peak noise from the supply was 60 mV (0.15 pm error). Then, the peak-to-peak amplifier noise was taken up to the maximum input for the oscilloscope, 100 Vdc. Table 3.2 shows the results of this test for amplifier noise versus its output. The behavior of the noise is a "stepping-average" where, as the output increases, its noise is relatively-constant over a certain range of values until it reaches a maximum. So the noise does not change linearly. The next series of tests involved inputting a variety of voltages and checking the probes' analog outputs and the corresponding noise. Each tables' data is from a different piezoactuator, but both sensors are used in each. Table 3.3 and Table 3.4 shows the sensor outputs and noise versus amplifier voltage output. As expected, the noise increased with the output voltage. This observation proves that the amplifier affects the amount of noise in the sensors. Even though the noise will affect the resolution, the sensors are working fine. To find the resolution and filter the noise, a test will have to be done in Matlab with the controller to fine tune the voltage output. Some noise-filtering methods will be needed in eliminating the noise (i.e., better low-pas filter, shorter cables, minimizing external noise, etc.). The amplifier's output is too large to output a small enough voltage to get even close to 10 nanometers. Figures C-2 through C-11 show the sensor output with different inputs to the piezoactuators A and B. 3.3 3.3.1 DSPACE Controller Design Characterization of System A dynamic analyzer was used to characterize entire open loop system, which included the flexures, the piezoactuators, the capacitance probes and the sensor electronics. Characterization is a very useful method, because an "unlumped" system can be viewed as a "lumped" system; meaning the overall components of a system are viewed 47 as one system. Figures 3-6 and 3-7 show samples of a simulink model of an open loop "unlumped" and "lumped" system. The dynamic analyzer inputs a sinusoidal signal into the system through the piezoactuators and the data from the capacitance probes' analog outputs will be converted into a bode diagram. Using control knowledge from [10], the transfer function for the lumped system can be calculated. 3.3.2 Dynamic Characteristics Two controllers will be needed to control the translation motion for closing the plates, and to control the yaw motion for aligning the plates. The dynamic analyzer will input a sinusoidal wave into piezoactuators A and B shown in Figure 2-1, separately and independently. In addition, data will be collected from sensors C and D shown in Figure 2-2, separately and independently. This process is shown in Figure 3-6. From all these components, four sets of bode plots can be found to ultimately create two controllers, one for opening and closing the plates and the other for turning the plates. Listed below are the combinations of parts that will be used in this experiment: 1. Piezoactuator B, flexure F, and sensor C (Figures 2-1 and 2-2), 2. Piezoactuator B, flexure F, and snesor D (Figures 2-1 and 2-2), 3. Piezoactuator A, flexure E, and sensor C (Figures 2-1 and 2-2) and 4. Piezoactuator A, flexure E, and sensor D (Figures 2-1 and 2-2). Using the data collected from the dynamic analyzer and Excel, the bode diagrams were found by plotting the magnitudes and phase from the digital analyzer versus the frequency. Each bode diagram has a corresponding slope plot which was part of the process of approximating of Bode plots. Figures C-12, C-13, C-14, C-15, 3-8, 3-9, 3-10, 3-11, 3-12, 3-13, 3-14, 3-15, C-16, C-17, C-18, and C-19 show the Bode plots and their corresponding slope plots. The slope equations were used to calculate how much the data dropped in decibels per decade. Using the equations and conversion tools from Figure 3-17, the transfer 48 functions were found. A transfer function of Parts A, E and C from Figure 2-1 was chosen over a lumped system of Parts A, E and D, because a clearer was signal was obtained from the former results. Since both were the same system with opposite readings for the sensors, there was no need to make two transfer functions. With Part B, there should be no difference in readings from Parts C or D, because they are next to each other. Figure 2-2 shows how when Part B is closing the gap, both sensors should be reading the same displacement. When Part A is turning the plates, one sensor will read the flexure getting closer while the other is reading an increasing displacement. Figure 3-17 showed how to calculate a transfer function from the bode diagrams using straigh line approximation. With these figures, the cutoff frequencies, w1, was estimated to be 500 rad. This value was found by making a trendline in Excel for the latter values. Then by setting the independent variable (magnitude) equal to zero in the trendline, the dependent variable (frequency) is found. Using the trendline, data, interpolation and Figure 3-17, the magnitudes were approximated to be dropping -40 decibels (dB) over every a decade (dC). This observation meant that two poles existed for every -IdB. So the general transfer function, G(s), was a multiplication of the "gain times the multiple pole functions" or [10] G(s) K * -N D k (3.5) K = 10 T (3.6) N = 1 (3.7) and D = (1 + 49 W1 Y(3.8) )s where k is the value on the bode plot when the slope of the magnitude is zero, K is the DC gain, p is the number of poles, and s is the LaPlace Transform complex variable. The transfer functions correspond to a sinusoidal input from the analyzer and an analog output from the sensors. During application, the input will be a DC voltage input to the piezoactuators. Including the higher order responses, the transfer functions were G ~7.08E - 1(39 topS) = 1.6E - u1s 4 + 3.2E - 8s 3 + 2.4E - 5S2+ 8E - Is + and Gbottom (S) =+8 3.2E - 14s 3.3.3 5 8.91E .1 JS - 2 +8E- 11s 4 +8E - 8s 3 + 4E - 5s 2 +1E - 2s + 1 (3.10) Designing Controller The next step involved designing the controller, Gc(s), for the top and bottom figure and making a closed-loop system (Figure 3-16). A Proportional-Integral-Derivative (PID) controller was selected for this application where [10], Gc(s) K, + K + Kds s (3.11) or G,(s) = KdS2 + K~s + K (3.12) where Kp, Ki, and Kd are the proportional, integral, and derivative gain coefficients. Then, using general control theory knowledge [10] Gc(s) * G(s) 1 + Gc(s) * G(s) and values for the system and control transfer functions, the top and bottom closed loop transfer functions were found to be 50 T"(s) [At * (Kps 2 + Kis + Kd)] (B * S5 + Ct * S4 + Dt * S3 + (Et + Ft * Kd)s 2 + (1 + Gt * Kp)s + Ht * Kj) (3.14) where At= 7.08E-1, Bt = 1.6E-11, Ct = 3.2E-8, Dt = 2.4E-5, Et = 8E-1, Ft = 7.08E-1, G = 7.08E-1, and Ht = 7.08E-1 and Tbottom (S) =- (B * S6 + C1 * S5 + Dt * [At * (Kps 2 + Kis + Kd)] S4 + Et * S3 + (Ft + Gt)s 2 + (1 + Ht * Kd)s + It * Ki) (3.15) where At = 8.91E-2, Bt = 3.2E-14, Ct = 8E-11, Dt = 8E-8, Et = 4E-5, Ft = 1E-2, Gt = 8.91E-2, Ht = 8.91E-2, and It = 8.91E-2. Even though the transfer function for the entire feedback system is found, the gain coefficients, Kp, Ki, and Kd, are still unknown. The optimal gain coefficients can be found using the characteristic equations based on the "integral of time multiplied by absolute error"(ITAE) criterion for a step input found in Figure 3-18. ITAE minimizes large initial errors and steady-state errors. Matching the characteristic equations in Figure 3-18 with the closed loop transfer functions of the same order, values for the coefficients were found. When it was possible to get two values for a coefficient, the average of the two were used. Kp,top, Ki,top, Kd,top, Kp,bottom, Ki,bottom, and Kd,bottom were found to be 10.77, 2260.88, 1.94E-1, 137.32, 23947.2, and 3.28E-1, respectively. So, by using the ITAE criterion for characteristic equations and averaging multiple values from ITAE, the optimal PID controller was calculated to be 2 Gecs) G,(s) == 1.94E - Is + 10.77s + 2260.88 (3.16) 3.6 and 3.28E - is 2 + 137.32s + 23947.2 Gc(s) = M (3.17) Using a Matlab tool rltool, the root locus was found for the top and bottom closed 51 loop system. Using this tool, you can find the ideal gain for the system for the best stability. For the top and bottom, the ideal gain was found to be 9.64E-4 and 1.24E3, respectively. Figures 3-19 and 3-20 shows the Root Locus of Top and Bottom System. 3.3.4 DSPACE Control DSPACE is a Matlab tool that allows the user to create a digital controller from a frequency domain or z domain transfer function. With the controllers found in previous section, the following DSPACE controller was constructed. Figure 3-21 shows the DSPACE Controller Model. The capacitance probes send signals to the Analog-to-Digital converters (ADC) from their analog outputs. Signals go into the DSPACE card, shown in Figure 3-22, and into the digital controller in the DSPACE software. A virtual control panel can be created for operating your signal (Figure 3-23). The digital control consists of the "gain block", which contains the gain value obtained from rlTool, and the PID controller. Since the derivative component of the PID controller was much smaller than the other gain values, the controller was simplied into a PI. Due to the derivative coefficient being very small relative to the other coefficients, this deletion was not considered to have a major effect. After the controller processes the information, the signals leave the DSPACE card through the Digital-to-Analog converters (DAC) into the piezoactuators. The controller did not seem to increase the stability of the signals. The noise was 6.13 mV and 4.5 mV for the sensors C and D, respectively. These results were not too different from the previous results from experimenting with the sensors. Further tests need to be done for adjusting the constants and gains and seeing the response of the system. Figure 3-24 shows the sensor output with DSPACE Controller . 52 3.4 Summary The capacitance probes, amplifier and actuators are functioning properly. A signicant amount of noise seems to be affecting the system (10 nanometer error from oscilloscope and sensors and 0.15 micrometer error from the amplifier). A means of minimizing the noise will need to be proposed or the desired resolution of the project will have to be modified even further. For the final design, an attempt will be made to use a new micrometer-integrated stage to check the sensors' calibrations. The entire range will be used in case of problems with the system beyond 0.5 micrometers. In addition, a new analyzer test needs to be done to characterize the new system. So far, the controller does not perform adequately in adjusting the gap width or reducing the noise, but results were affected by the DSPACE card's location (not on an air table). In addition, an error may have been made in calculating the transfer functions from the bode diagram, but this step was more for familiarization with the bode calculations. An over-approximation may have occured using the straight line approximation method from Figure 3-17 over too large a frequency range. Since low frequency behavior is of primary interest, high frequency responses will be neglected in the next chapter. Plus, the DSPACE card and its computer could not be moved at the specific time. For testing the methodology, all components will be moved to the air table for better results. 53 Table 3.1: Sensor outputs Freq Input(V) 1.04 2.14 3.32 4.42 7.06 8.11 8.93 10.8 12.2 14.5 Sen Output(mV) 121 272 463 663 1142 1146 1450 1684 1850 1965 Sen Output(p m) 0.0605 0.136 0.2315 0.3315 0.571 0.573 0.725 0.842 0.925 0.9825 Table 3.2: Amplifier noise outputs versus voltage output Input 13.618 13.618 13.618 13.618 13.618 13.618 13.618 13.618 13.618 13.618 13.618 13.618 13.618 13.618 13.618 13.618 13.618 13.618 13.618 13.618 13.618 13.618 13.618 13.618 Input Noise 60 60 60 60 60 60 60 60 60 60 60 60 60 60 60 60 60 60 60 60 60 60 60 60 Amp setting 0 1 5 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 105 Offset Output(V) 232 1.1294 2.0081 5.313 10.371 15.375 20.692 25.874 30.588 35.231 40.522 45.777 50.630 55.483 60.329 65.523 70.410 75.201 80.833 85.223 90.451 95.371 100.71 105.17 54 Output Noise (mV) 13 13 15 60 130 130 190 250 250 250 500 500 500 500 500 500 500 500 500 500 500 500 500 500 Table 3.3: Sensor outputs and noise due to piezoactuator A versus amplifier voltage output Amp Out (Vdc) -30 0 1 +150 Parts A and C/Noise(V) 1.1448+/-0.0058 -0.06147+/-0.00595 -1.3367+/-0.00625 -8.2671+/-0.0065 Parts A and D/Noise(V) N/A 0.68739+/-0.00543 0.84282+/-0.0078 2.9305+/-0.0083 Table 3.4: Sensor outputs and noise due to piezoactuator B versus amplifier voltage output Amp Out (Vdc) -30 0 1 +150 Parts B and C/Noise(V) 1.0761+/-0.005 -0.04756+/-0.0058 -1.2095+/-0.00565 -8.2337+/-0.00705 55 Parts B and D/Noise(V) N/A 0.682+/-0.0166 0.87335+/-0.0078 2.8826+/-0.00815 4: Figure 3-1: Digital picture of probe Figure 3-2: Micrometer stage 56 Figure 3-3: Voice coil in a speaker [2] Figure 3-4: Digital picture of piezo-amplifier 57 1.2 y = 0.0737x + 0.003 0 0 0 2 6 8 10 Generator Input (V) 12 16 14 Figure 3-5: Sensor output versus generator input (stiffness test) 24~s+ ~.s+B~ Jj] , L E3. 3+~s+&s+H3 EA+F2s24 2.r12 Analog: Outputl Sine Wive Input Moom or Top Piezoaduator Laigevo Small R~enei Ma~er or Slave Prob and Ele0ronis Figure 3-6: Simulink model of open loop "unlumped" system 58 As 3+8~s 4Cs~D Anaiog "OuPIut Sinre Wave em FLu-mped p p Figure 3-7: Simulink model of open loop "lumped" system 0 -10 A M -20 -30 -40 -M -50 -60 +- -70 -80 - I ------ -- - ----- I---LIJ Figure 3-8: Piezoactuator B and flexure F 59 --El I Frequency(rad/s) / nOO rad/a sensor D magnitude bode diagram 0 PUU I -60 -100 -160 ill -200 0-250 -300 -350 -400 - Frequency(radls) Figure 3-9: Piezoactuator B and flexure F / sensor D phase bode diagram LA -10 f -20 i -30 0) -40 I.:+ -60 -70 -80 firH H --V-------Frequency(rad/s) Figure 3-10: Piezoactuator A and flexure E diagram 60 / y =O:Ndft'd179 sensor D slope in magnitude bode 0 r suu -50 -100 -15n -200 U2 I 0-250 -300 -350 -400 y Frequency(rad/s) Figure 3-11: Piezoactuator A and flexure E / = -'045' 5.96 sensor D slope in phase bode diagram 0 FfU0 -L -10 -r -20 wL - K A~ IT -30 -40 -50 -....................... T I... ..... .......... ....... .................. ..... I..... ..... r.... ....... -60 -70 Frequency (rad/s) Figure 3-12: Piezoactuator A and flexure E 61 / 500 rad/s sensor C magnitude bode diagram 0 inn 41- -50 -100 150 -200 - %k -250 -300 [-LJ -35 Frequency (radls) Figure 3-13: Piezoactuator A and flexure E/ sensor C phase bode diagram 0 I LL I I I 1 1 100 -10 -20 -30 If -40 A I -50 -60 -70 F r111111y I sI Freq uency(radls) Figure 3-14: Piezoactuator A and flexure E diagram 62 / .y.... Y -0.01 13.412 sensor C slope in magnitude bode 0 6;w -50 -100 -150 4) 0..-250- -300 -350 -400 I I I .1 I yLI isI Frequency(radls) Figure 3-15: Piezoactuator A and flexure E / y .... ...... sensor C slope in phase bode diagram Sine WaveI C s+G Controller s+1 Gain XY Graph System Figure 3-16: Simulink model of a closed loop system 63 1. Gain, G(jw) Phase, 4(w) Magnitude 20 log GI Term K 40 904 201 45& 0 0' dB -0 2. Zero, 40 90 201 45 0 dB 4x --- -20 1 4 40 90 3.Pole - 49T ------ 4 G(Aj ) -01 (I + job) 4--- -- .(.. - w4 2-01 0(.1w, 4.Pole at the origin 1/jo G(jw) - 00 0 dB - l0o, 1 - 010 0) 0w 1 10 100 10 100 90, 4 4. d) 0 dB 0.01 - 0" 940 - --- 0A 100 10 001 . .... 0.1 18 40 80 5. Two complex poles 0.1 < < 1. G(jw)= (1 + i24u -u2) 20 dB 0 &(W) 00 20 90,* 40 0.01 18_ 0.1 1 10 U 100 0.01 - 0.1 1 U Figure 3-17: Asymptotic curves for basic terms of a transfer function [10] 64 S+( s3.5~ 2 +215os+( s4& +27is+ W4 s4±2.1gs +W S6 4 s28~ S J~)S3+5s+34S + O 5+ A6WS+86Oos + AS., +3.95(ons +(On 325L)s Figure 3-18: The optimum coefficients based on the ITAE criterion for a step input [10] 400 - 200- < 0 / - - - - - - --.......... -............... - + - - ----- ---------- --- -- - -- - ------ -200 - -400- -600- -8001 -1500 -1000 -500 Real Axis Figure 3-19: Root locus of top system 65 0 2000 1000 - 07 -1000- -2000- -3000 -3000 -2500 -2000 -1500 -1000 -500 Real Axis 0 500 1000 1500 2000 Figure 3-20: Root locus of bottom system Ba d Unk L - - SII - DS1103MUX.A.CCON1 Top3 C Top Confol DS113AC C Top Constant 137.3s+230000 DS1103MUX.ADC-CON2 Bottom Gain Bottomn DS1103DACC2 Controi Bottom Constant Figure 3-21: DSPACE controller model 66 Figure 3-22: DSPACE card Figure 3-23: DSPACE panel 67 Figure 3-24: Sensor output with DSPACE controller 68 Chapter 4 Experimental Results 4.1 Introduction This chapter is partitioned into three sections. The first section will involve testing the flow rate of water through the partially closed plates. The mock plates are "partiallyclosed", because they do not have a good resolution of flatness. In the second section, the system will be tested to see if the proper displacement are being distributed and detected through the system. Plus, the elasticity of the sealant can be viewed from the results. In the last section, the filtration test will see what range of particles can be separated. 4.2 Flow Test This test involved seeing if water could flow properly throughout the system. Peristaltic pumps were used for this process because of their unique pumping feature. The pump contracts and expands the tubing that runs through it. This contraction and expansion creates a suction by increasing and decreasing the inner cross sectional area. By the fluid staying in the tube and not contacting any portion of the pump, this method is contaminant-free. A VWRbrand medium-flow, peristaltic pump was used to draw the solution from its container to the filter. This pump's bidirectional flow rate ranged from 4 to 85 69 . Using some tube reducers and tubing, the filtration chamber was built (Figure ??). The components are described in Table 4.1. To reduce the risk of bottlenecking mn and clogging at the filter, another pump with a much larger flow rate was placed after the filter. A Barnant E-Seires peristaltic pump was used to suck the solution through the filter and into a container. Its flow rate was 375 1. min For the flow test, the actual filter plates were not used. Plexiglass plates were being tested instead of the fused silica quartz. The results of this test was sufficient flow throughout the system. A 0.6" buildup in water occurred after about 45 seconds but it quickly diminished. As expected, the large flow pump eliminated the potential clogging of the system and its flow rate dominated the process. One observation was that the plates were in a closed position, but fluid was still flowing between the "mock" plates. This was not a surprise because the plates were not "perfectly" flat at the edges. The test did show the sufficient flow through a small gap width was possible. Instead of the minium of 1 -, 4.3 the results were approximately 50 -!. Actuation and Sensing Test The micrometer-stage setup seen in Figure 3-2 was not accurate enough for our application, because the probes kept slipping in the fixture to the stage. Due to other more pertinent issues, this test was not completed so the manufacturer's calibration factor was used. The next step was to verify if the sensors could detect any displacements applied on the outer tubing by the piezoactuators. In addition, output readings will be taken at random input voltages to test for repeatability of the system. This is the sealant test mentioned in Chapter 2. If repeatability is good, that means that the elastic nature of the silicone sealant, vinyl tubing, and rubber cement for the displacement distributor is working properly. Then the data plot should resemble either Figure 2-5 (if optimal) or Figure 2-6 (more realistic). The master sensor (Part C) in Figure 2-2 was tested first. If the master test is successful, the results would also be considered for the slave (Part D). Figure 4-2 shows the result of this test. An input of 13.54 V 70 with a peak-to-peak noise of 90 mV was driven into the amplifier by a 15 V power source. The amplifier's output depends on its settings. The peak-to-peak noise from the sensors ranged from 23 to 30 mV. Then the setup was reinvestigated and it was discovered that the filter piece was slipping. That is why consistent readings were not given. The piezo- and sensor-holders' tube claw were not sufficient for support, so a stand with a two 3-prong clamps was used to hold the filter. Then the test was performed again with much better results. Table C.1 shows the data. The input voltage was increased to its maximum value then decreased back to its initial value (0 V). Figure 4-2 shows that it resembles a hysteresis plot but its loop is not large (top curve: increasing and bottom curve: decreasing). In addition, it returns to its initial value. The change in voltage from the minimum to the maximum value was 0.77 V. Going by the calibration factor mention in Chapter 2, this is equivalent to 3.85 micrometers. 4.4 Filtration Test Using a dynamic signal analyzer, the MPF Tube Filtration design was characterized into a transfer function. Table C.10 and Table C.11 shows the data for this analysis. From this information, bode diagrams were plotted in Figure 4-3 and Figure 4-4. The magnitude plot started at a gain of -29 dB and dropped 40 dB over 2 dC. This observation is equivalent to a first order equation. Since higher frequency responses are not interest of this analysis, this assumption will be taken. Referring to Figure 3-17 and using equations from the previous chapter, the system's lumped transfer function was Gf(s) = Kf * (1 + -29 (4.1) -) WC where Kf equals 109 dB (gain) and w, (cutoff frequency) equals 314.16 a r. Using the same assumptions and general equations for the control equation, the overall transfer was 71 Tf (s) = s Kf * (Kpf * s + Kif) (Z*Lpf +-w,) *s+Z*Kf(4.2) Before solving the characteristic equation, the natural frequency needs to be found. For a second-order equation, a settling time criterion of ± 2 percent is desired. Since IT = - (4.3) wc (= 0.7 (4.4) and for the ± 2 percent criterion 4 *T = 4(4.5) ( * Wn the natural frequency w,, equals 314.16 Id. With this value and matching the coefficients for a second-order characteristic equations from Figure 3-18 with the coefficients in the denominator of the overall transfer function, Kpf and Kif are 3541.6844 and 2781632.45824, respectively. Now using "rltool" in Matlab, the root locus is shown in Figure 4-5. The figure seems very stable so the controller appears to be appropriate for this application. The next stage involves testing the open and closed loop system to see if it the will work in an actual experiment. A step input will be applied using a DC power supply to apply a voltage. Looking at the settling time, steady state error, and steady state value, the controller will be evaluated. First, a DSPACE model of the controller was constructed in Simulink (Figure 4-6), but the open-loop case was investigated prior to the closed-loop. Then a square wave input was applied to the open loop system and Figures 4-7 and 4-8 show the results. Table 4.2 shows the performance of the open loop system. The closed loop system performed much better(Figure 4-9 and Table 4.3). Since the signal settled so quickly, most of the performance values could not be obtained and the oscilloscope could not record all the data (nanosecond reaction). 72 After running all the mechanical and electrical test, the final step is to test the methodology of the MPF. The test particles to be separated are of diamater sizes of are 0.079, 0.304, 0.482, 0.093, and 1.03 microns. They came in 10 ml vials. First, water would run through the system at 0 V to test to see if the tests are properly closed. If this preliminary test suceeded, then the test would continue by trying to separate a solution of 1.03 micron particles from water. Once filtrated, the voltage reading would be recorded and the next smallest size would be tested. This process would continue until all particle solutions are filtrated or no more could be possibly filtrated. Table 4.4 was to show the results of the filtration test, but unfortunately, results were not obtainable. At 0 V, water passed through the filter region when the plates were supposed to be closed. In addition, a major leak sprang from one side of the filter where the tubing connected with the contact plates. So this test had to be concluded prematurely. 73 Table 4.1: Filtration tube chart Q Part Name 0.75" ID filter chamber R 0.75" to 0.375" tubing reducer S 0.375" to 0.25" tubing reducer T pump connector Part Label Table 4.2: Open loop response due to a square wave input Initial Input(V) 2.81mV Initial Time (s) Steady State Value (V) Steady State Time(V) Max-Min(V) 0.1 percent (V) 0.1 percent (s) 0.9 percent (V) 0.9 percent(s) Rise Time (s) Settling Time (s) Peak Value (V) Peak-to-Peak Value (V) Percent Overshoot Os -9.69mV 20ms -12.5mV 1.56mV 0.005ms -8.44mV 4.5ms 4.445ms 5.025ms -10.25mV 23mV 5.779 per. 74 Table 4.3: Closed loop response due to a square wave input Initial Input(V) Initial Time (s) Steady State Value (V) Steady State Time(V) Max-Min(V) 0.1 percent (V) 0.1 percent (s) 0.9 percent (V) 0.9 percent(s) Rise Time (s) Settling Time (s) Peak Value (V) Peak-to-Peak Value (V) Percent Overshoot 5.63mV 109.2ns 5mV 142ns -0.63mV N/A N/A N/A N/A N/A 30ns 5mV 8mV 0 per. Table 4.4: Filtration test Particle Size(microns) 0.079 0.304 0.482 0.093 01.03 Input Voltage (V) N/A N/A N/A N/A N/A 75 Master Sensor (V) N/A N/A N/A N/A N/A Slave Sensor (V) N/A N/A N/A N/A N/A S T Figure 4-1: Filtration tube 11.7 11.6 - 11.5 11.4 11.3 . 11.2 Increasing Volts w Decreasing Volts J11.1 11 10.9 10.8 10.7 0 20 40 60 80 Ampimffer 100 120 140 160 Input(V) Figure 4-2: Sensor output(V) versus piezoelectric output from amplifier (V) 76 LOU U *.0 - Vak~s (Y) AXIS Maor GrdkI 0 a'~ Frequency (Hz) so Hz Figure 4-3: Magnitude Bode diagram for MPF tube filtration design i i i !I . rob a 0~ wu FU -1 4- p i i i I1 Frequency (Hz) Figure 4-4: Phase Bode diagram for MPF tube filtration design 77 Figure 4-5: Root locus of closed loop system 1 --- 3.5416844e+003s+2.78163246824e4006 Bad nk> -+ S DS03ADCC17 Gain 7 Bad Link DS1103DACCl Controller Sine Wave Figure 4-6: DSPACE controller for filration experiment 78 3.00E+00 2..0 00 1.OGE 00 - a, + 0) 0 -6.0 QE-02 -4.OOE-02 2.OOE-02 0.001 -+00 -2.00E-02 4.OOE-02 6.00 E-02 Square Input 1 m Sensor Output 2 1.AGE an Time(seconds) Figure 4-7: Sensor output and square wave input for open loop 4:SOE-62- 0. S~ -6.01 dE-02 -4.OOE-02 -2.OOE-02 0.00 4 2.OOE-02 4.OOE-02 6.00f E-02 0 C- 0 1.96E 82 co 1.50E 82 time(seconds) Figure 4-8: Sensor output from square wave input for open loop 79 ------------- Figure 4-9: Sensor output from square wave input for closed loop 80 Chapter 5 Conclusion and Recommenations Even though the Tube Filtration method was not actually tested, all the preliminary steps were investigated and this method seems very possible with some modifications. First, due to the 2-3 nanometer flatness of the plates, the 0.5 nanometer gap width had to be increased to 10 nanometers. Also, there was a 5 nanometer error from the sensors. The sensors are supposed to be able to achieve a 0.5 nanometer resolution. After further test on the equipment and seeing how the error increased to 150 nanometers when the setup was completely assembled, it was concluded that the error was due to electronic noise from the equipment. A better low-pass filter needs to be designed for this system. All wiring needs to be secured and minimized in length. An air-table (or any vibration-reducing table) is vital because the sensors are very sensitive. Furthermore, external noise may play some small role in the error as well. For optimal results, this experiment should be conducted in a noise-free environment. The controller was very effective in minimizing noise in the actual test with the filter tube. Due to the mock test being performed on a counter (DSPACE computer could not be moved at that time), the results were flawed. The air table assisted tremendously the filter tube controller test. Reducing the noise from 23 mV to 8 mV means that a ± 0.5 nanometer error exists. So the controller will enable resolutions below 10 nanometers to be obtained. The dual-flexure design had potiental for achieving the resolutions, but problems evolved when trying to implement the filtering part. Initially, the silica plates were 81 to be attached to Part E (top flexure) and have a chamber built around them. This idea failed when too many leakage issues arose and there was now a major spacing contraint (flexure was already built). Instead of investing the time into designing another flexure, another method was implemented to simplify this application. The tube filtration method was ideal. The plates were already inside the tubing. With more forethought about design around the filter, the leakage and spacing issues were eliminated. With the fluid analysis through a tubing, it seemed very feasible to achieve the minimum flow rate. When the actual flow test was conducted, a flow of 50 ML was obtained and that far exceeded the 1 4 minimum. A problem still persists min min in sealing the plates inside the tube. In the mock test, the plates were sealed while they were inside the sensor contacts. While the sealant cured, the sensor contacts were mounted to a metal mounting bar that supported the tube housing and plates. In the methodology test, the sensor contacts had to be modified at the last minute and they no longer fit the mounting bar for sealing. In this case, the tubing was supported by a chemistry stand and the sensor contacts were attached to the tube housing. While sealing, the plates probably shifted because the tubing was too elastic and did not provide a firm support. In the sensing and actuation test, there was a hysteresis effect observed. This effect was probably due to the rubber cement wearing after multipe usage. The tube claw started to slip out after about a week. At a maximum input voltage of 150 V for the piezoactuaters, going by the calibration factor, only 3.85 micrometers were obtained for motion. This value was more than what was needed (limit of 0.5 micrometers). So the motion factor from the amplifier is 25.67 "'. Since the piezoactuators have a maximum stroke of 20 micrometers, this shows a loss in performance. Probably the elasticity of the tubing and sealant absorbed some of the motion. Still this filtration method seems possible. For further work on the Tube Filtration method, a few guidelines and suggestions need to be followed. It is crucial to cut the openings "perfectly-parallel" so that both sensors on each side can get proper reading. Second, make sure that the tube claws are perfectly aligned with the tube's center and each, or the motion will not be straight. 82 In addition, some method in DSPACE or electronically needs to be designed to be able to get voltage outputs from the amplifier lower than 1 V. The amplifier's resolution was too high to obtain submicron motion. A lever to decrease the displacement output of the piezoactuators may mechanically enable the resolutions to be obtained. 83 84 Bibliography [1] http://www.adetech.com/2800.shtml [2] http://www.aa.washington.edu/controls/classes/448w96/lab5bc.jpg [3] http://fibec.flight.wpafb.af.mil/fibec/hysteresis.html [4] http://www.mse.cornell.edu/courses/engri111/modulus.htm [5] http://www.ndif.org/Terms/hydrolysis.html. [6] http://www.osmonics.com/products/Page710.htr. [7] http://www.piezomechanik.com [8] M. C. Bekker, J.P. Meyer, L. Pretorius, and D.F. Van Der Merwe "Separation of solid-liquid suspensions with ultrasonic acoustic energy", Water Resources, Vol. 31, No. 10, (March 1997). [9] G. Belfort (ed.), "Synthetic Membrane Processes: Fundamental and Water Applications", Academic Press, Orlando(1984). [10] R.C. Dorf and R.H. Bishop, "Modern Control Systems", Addison-Wesley, Menlo(1998). [11] J.A. Fay, "Introduction to Fluid Mechanics", MIT Press, Cambridge(1998). [12] R.W. Fox and A.T. McDonald, "Introduction to Fluid Mechanics, 4 th Ed.", John Wiley & Sons, Inc., New York(1992). 85 [13] T. Hanai (ed.), "Liquid Chromatography in Biomedical Analysis", Elsevier, Amsterdam(1991). [14] W.S. Hancock (ed.), "High Performance Liquid Chromatography in Biotechnology", John Wiley & Sons, New York(1990). [15] J.J. Hawkes and W.T. Coakley "A continuous flow ultrasonic cell-filering method", Enzyme and Microbial Technology, Vol. 19, No. 1, (July 1996). [16] J.J. Hawkes, J.J. Cefai, D.A. Barrow, W.T. Coakley, and L.G. Briarty "Ultrasonic manipulation of particles in microgravity", Journal of Applied Physics, Vol. 31, No. 14, (July 1998). [17] A. Henschen, K.P. Hupe, F. Lottspeich, and W. Voelter (ed.), "High Performance Liquid Chromatography in Biochemistry", VCH, Weinheim(1985). [18] I.L. Holwill, G.B. Davies, N.J. Titchenerhooker, and M. Hoare "Particle Manipulation by ultrasonic standing-wave fields to complement dynamic light-scattering experiments", Particle & ParticleSystems Characterization,Vol. 12, No. 3, (June 1995). [19] I.S. Krull, R.L. Stevenson, K. Mistry, and M.E. Swartz, "Capillary Electrochromatography and Pressurized Flow Capillary Electrochromatography: An Introduction", HNB Publishing, New York(2000). [20] Parekh(ed.), "Reverse Osmosis Technology: Applications for High-Purity-Water Production", Marcel Dekker Inc., New York(1988). [21] M. Saito, T. Daian, K. Hayashi, and S. Izumida "Fabrication of a polymer composite with periodic structure by the use of ultrasonic waves", [22] H. Shintani and J. Polonsky, "Handbook of Capillary Electrophoresis Applications", Blackie Academic & Professional, London(1997). [23] S.T. Smith, "Flexures: Elements fo Elastic Mechanisms", Gordon and Breach Science Publishers, Australia(2000). 86 [24] S. Sourirajan and T. Matsuura(ed.), "Reverse Osmosis and Ultrafiltration", American Chemical Society, Washington, D.C., 1985. [25] K.S. Spiegler and A.D.K. Laird(ed.), "Principles of Desalination: Part A., 2n Ed.", Academic Press, New York(1980). [26] R. Wolfson and J.M. Pasachoff, "Physics: with Modern Physics for Scientists and Engineers, 2 nd Ed.", Harper Collins College Publishers, New York(1995). [27] L.J. Zeman and A.L. Zydney, "Microfiltration and Ultrafiltration: Principles and Applications", Marcel Dekker Inc., New York(1996). 87 88 Appendix A Company Addresses 1. Active Electronics, 73 First St., Cambridge, Massachusetts 02141, Tel: (617)8643588, Fax: (617)864-055 2. ADE Technologies, 77 Rowe St., Newton, Massachusetts 02466, Tel: (617)8318000, Fax: (617)243-4400, www.adetech.com 3. Coghlin Electric/Electronics, A Div. Of Wesco Distribution Inc., 35 Otis St./PO Box 5100, Westboro, Massachusetts 01581-5100, Tel: (508)870-5000, Fax: (508)8705157 4. The Home Depot, 75 Mystic Ave, Somerville, MA 02145, Tel: (617)623-0001 5. Manostat: Division of Barnant Co., 28W092 Commercial Ave, Barrington, IL 60010-2392, Tel: (800)637-3739, Fax: (847)381-7053 6. McMaster-Carr Supply Company, 473 Ridge Rd, Dayton, NJ 08810-0317, Tel: (732)329-3200 7. Power-One, Inc., 740 Calle Plano, Camarillo, California 93012, Tel: (805)9878741, Fax: (805)388-0476 8. RadioShack.com, 3131 West Bolt St., Fort Worth, TX 76110-5813, Tel: (800)4645365, http://www.radioshack.com 89 9. Ramco Machine, Inc., 416 Calbot St., Beverly, Massachusetts 01915, Tel: (978)9214600, Fax: (978)921-8448 10. Ted Pella, Inc., 4595 Mountain Lakes Blvd, Redding, CA 96003-1448, Tel: (530)243-2200 11. Thorlabs, Inc., 435 Route 206, Newton, NJ 07860, Tel: (973)579-7227, Fax: (973)383-8406 12. US EuroTek, Inc., 25315 Costeau St., Laguna Hills, California 92653, Tel: (949)458-6794, Fax: (949)916-9121, useurotek~juno.com 13. Valley Design Corp., East Coast Division: 63 Power Rd., Westboro, Massachusetts 01866, Tel: (978)692-1971, Fax: (978)692-9549, eastcvalleydesign.com; West Coast Division: 151-D Harvey West Blvd., Santa Cruz, California 95060, Tel: (831)430-0595, Fax: (831)430-0592, west@valleydesign.com 14. VWR Scientific Products, 200 Center Square Rd., Bridgeport, New Jersey 08014, Tel: (800)932-5000, www.vwrsp.com 90 Appendix B Schematics 91 ......... Figure B-i: Isometric view of small flexure A (mm) i - 22 -W--__---2 - - 06 35 50 - -- i 2....--- Figure B-2: Front view 1 of small flexure A (mm) 92 -03 10 - - - --(7~ _ _ _4 _ 3.5 R2 . 5 10 ... F.. 12 2 ] 20 ]0 I Figure B-4: Top view of small flexure A (mm) 93 ( . Lf Figure B-5: Side view of small flexure A (mm) -.......... .... ZMy .... Figure B-6: Isometric view of large flexure B (mm) I ~IZIf~-66 I Ii 1 70 ... ........ .... . 8 -0 f2 7 148 40 Figure B-7: Front view of large flexure B (mm) 'L 01 15 10 [ o- L 5 3.0 7 I '5 Figure B-8: Top view of large flexure B (mm) 95 * -I I ~ ................................................................................................................................................................ ML--- - 4.5 f>;9-s... ...................................... ..... II-OFT-T...... ..... ..... ..................... ........... ........ .............. ... ...... ....... z 3; 3 ........ ........ I -...... ..... Figure B-9: Side view of large flexure B (mm) Figure B-10: Isometric view of piezo-holder M (inches) 96 __1.00000000000 0 46968503937 1. 1.000 Figure B-11: Front view 1 of piezo-holder M (inches) 25 0000000-r0- 1.00 .5 ---- 1.2348425196 ---- IL. - Figure B-12: Front view 2 of piezo-holder M (inches) 97 .24 1.125 4 45O- .50 1.23484244*9- 1 00 0----,---- --t- 120 Figure B-13: Side view of piezo-holder M (inches) .325 010 2.00d 1.23484251969 .. I 675 1.49 - 1 - 2.0 00___. 00 Figure B-14: Top view of piezo-holder M (inches) 98 Figure B-15: Isometric view of sensor-holder N (inches) 1401 _____ 8 1.80' - .30 .00 .30 - Figure B-16: Front view of sensor-holder N (inches) 99 .50 .50 25590551 8 5UM1O*~ r- I L fA1 IT n 74 - Figure B-17: Side view of sensor-holder N (inches) i 0 I..... 111Il . ...... t 1......._...... 325 197456692913 1it: 2 0000110 0000 Fe. .675 2.000 -.32 I 0- Figure B-18: Top view 1 of Sensor-Holder N (inches) 100 2. 10 ©J12 _ © 3) .~ ... -01159 25590551 18 .8074 Figure B-19: Top view 2 of Sensor-Holder N (inches) .~.......... Figure B-20: Isometric view 101 of tube claw 0 (inches) W-. 150 ;; T *- ..... F 50 1.00 Figure B-21: Front view of tube claw 0 (inches) ~~1 Figure B-22: Side view of tube claw 0 (inches) 102 125 525000 -01.0000 .50 .28 Figure B-23: Top view of tube claw 0 (inches) Figure B-24: Isometric view of sensor contact P (inches) 103 I. 1 .0 67 4 OO 00 50000000o 3i Figure B-25: Front view 1 of sensor contact P (inches) 13657480315 -T -54 18 1 10k -362 - - - 05 4 1 89 Figure B-26: Front view 2 of sensor contact P (inches) 104 1.00000 ..... .875 2- 1.704 0000- 56 75 .1 L -45000- Figure B-27: Side view of sensor contact P (inches) GRO GO3CO T AREA. R j.0 4 15 M Al i.mm Imp 'I TT.IH~ ACVEEAIA, 5 mM (0.2') Figure B-28: Schematic of model 2800 series probe [1] 105 I~ 4 mm A~. ~ if"'' ~4aaO.S I t i I 6.1f P -t IT a a *s 4 mm mu e-vm s Figure B-29: Piezoactuator Schematic [7] 30 IIT f *1 1 1 I I I1 I 11 4r Figure B-30: Spherical top piece (steel) on moving end [7] 106 Appendix C Controller Design C.1 Sensor, Piezoactuator, and Dynamic Signal Analyzer Data NOTE: In dynamic signal analysis, -360 was added to phase once it was positive to plot. 107 Table C.1: Elasticity test of system Increasing Input(V) 0 7 10 17 26 32 43 52 59 63 72 75 80 89 95 108 113 120 125 140 145 N/A N/A N/A N/A N/A N/A N/A Output(V) 11.57 11.55 11.54 11.5 11.44 11.4 11.33 11.28 11.24 11.21 11.15 11.13 11.11 11.04 11.01 10.94 10.93 10.89 10.87 10.81 10.8 N/A N/A N/A N/A N/A N/A Decreasing Input(V) 140 135 129 125 116 109 104 94 87 83 80 75 67 61 55 54 50 44 36 30 24 26 18 13 5 1 0 108 N/A Output(V) 10.81 10.82 10.84 10.85 10.88 10.91 10.93 10.97 11.01 11.03 11.04 11.07 11.11 11.14 11.15 11.18 11.21 11.24 11.29 11.34 11.38 11.43 11.46 11.48 11.53 11.56 11.57 Figure C-1: Noise plot for sensor output -I-I -I-I--I-IA .. .. .... t~'Y AL 0, - - - .4 Figure C-2: Sensor output with bottom piezoactuator at -30 Vdc 109 II At~ *14I. I,1' '1 [" & i~i..i4 14.Ln k P I 17, . [.1 L I, 1'' 4...... ,Ag Ij p I r Figure C-3: Sensor output with top piezoactuator at -30 Vdc AtV~ I NW... * 4$-ti' IIl t WA Figure C-4: Sensor output with bottom piezoactuator at 0 Vdc 110 I ......... ........ ... t ~ W arr I Figure C-5: Sensor output with top piezoactuator at 0 Vdc Figure C-6: Sensor output with bottom piezoactuator at 1 Vdc 111 - Figure C-7: Sensor output with top piezoactuator at 1 Vdc Figure C-8: Sensor output with bottom piezoactuator at +150 Vdc 112 I ffiI~ I ....... 4 4 4"' 4' * I .............. A Figure C-9: Sensor output with top piezoactuator at +150 Vdc 'I-i-I-I-.-.-.-.-. ~f I0%72 17-L 11 I-Ij111 4 4 4 *T -4. *' IL r * ... .A.. -.-..... 70 T~ T* V44 . ~ ur~ 0 -~-'~- i1ji"~ June Figure C-10: Sensor output with bottom piezoactuator at +1 Vdc 113 ___j Figure C-11: Sensor output with top piezoactuator at +1 Vdc 0 LPUU -10 -20 -30 ca -40 -50 -60 -70 -80 Frequency(radls) Figure C-12: Piezoactuator B and flexure F 114 / sensor C magnitude bode diagram 0 bo i I I DOG 000 -50 -100 I -:-150 -200 C -250 -300 -350 -400 -L L. Frequency(rad/s) Figure C-13: Piezoactuator B and flexure F / sensor C phase bode diagram 0 PUU -10 -20 -30 ~40 I I ) - -50 -60 -70 -80 Frequency(radls) Figure C-14: Piezoactuator B and flexure F diagram 115 / y = -0.01 38x - 32.771 sensor C slope in magnitude bode 0 -F0 4- -100 -150 -200 c 0.L -250 -I -300 I -350 II....... _ --. 1..L. 1 .. Frequency(rad/s) 400 Ly -- 1.82 O.O428~t- Figure C-15: Piezoactuator B and flexure F/ sensor C slope in phase bode diagram I I rI I .ILE KJU 0 -~~~~~~~~~ ThrIIzzi±±~h ± ±Hzz -10 I I IZLIUZL I I rtUU H iHi I I I -20 -30 at -40 fl$ -50 i -60 -70 . i i i i iii i i ! i i ! i H ii i i i i i i !i i [ !iHH! i i i i i i I I I Hi 1 1 1 11 I I 1111 1 Frequency(radls) .1.U Figure C-16: Piezoactuator A and flexure E/ sensor D magnitude bode diagram 116 - 200- .............. I......... ...... T1111 T1 1 I 150 100 50 - ]- 0 Po I 11 PI00 -50 100 150 -2 00 4- _1I ..... A - I j I I 1 1 . - 1- - , -1- A L IA . L- ..-... --- --- ---- --' - A --- Frequency(radis) Figure C-17: Piezoactuator A and flexure E / sensor D phase bode diagram 0 00 -10 -20 -30 C, -40 -50 -- 166x - 13.418 -60 -70 - I I I I[II: I - I L 1- ]I l Frequency(radls) Figure C-18: Piezoactuator A and flexure E diagram 117 / sensor D slope in magnitude bode 0 [Nil -20 -40 -60 -60 -100 iL -120 -140 -160 -180 - - - II I . . .I Frequency(radls) Figure C-19: Piezoactuator A and flexure E 118 / y =0- 3 8k;;47 078 sensor D slope in phase bode diagram Table C.2: Dynamic signal analysis of flexure design la Lt Probe/Tp Piezo INPUT (HZ) 5 14.9 24.8 34.7 44.6 54.5 64.4 74.3 84.2 91.625 92.863 94.1 104 115.138 125.038 134.938 144.838 154.738 164.638 175.775 185.675 195.575 205.475 N/A INPUT (RAD/S) 31.41592654 93.61946108 155.8229956 218.0265302 280.2300647 342.4335992 404.6371338 466.8406683 529.0442029 575.6968538 583.4754372 591.2477374 653.4512719 723.4333899 785.6369244 847.840459 910.0439935 972.2475281 1034.451063 1104.426897 1166.630432 1228.833966 1291.037501 119 N/A MAG (dB) -3.132 -4.15096 -5.19313 -6.30853 -7.48723 -8.75977 -10.1266 -11.5976 -13.1809 -14.4426 -14.6544 -14.8663 -16.6287 -18.6683 -20.4838 -22.3058 -24.0914 -25.8415 -27.524 -29.4045 -30.9247 -32.5452 -33.9815 N/A PHASE (Deg) -21.4564 -43.8148 -64.5002 -83.9115 -102.405 -120.121 -137.121 -153.41 -168.72 -179.527 -178.726 176.979 163.85 150.529 139.891 130.305 121.695 114 107.137 100.389 93.9279 89.5737 84.7065 Table C.3: Dynamic signal analysis of flexure design lb Lt Probe/Tp Piezo INPUT (HZ) 215.375 225.275 235.175 245.075 254.975 264.875 274.775 284.675 295.813 305.713 315.613 325.513 335.413 345.313 355.213 365.113 375.013 384.913 396.05 405.95 415.85 425.75 435.65 445.55 455.45 465.35 475.25 N/A INPUT (RAD/S) 1353.241036 1415.44457 1477.648105 1539.851639 1602.055174 1664.258708 1726.462243 1788.665777 1858.647895 1920.85143 1983.054964 2045.258499 2107.462033 2169.665568 2231.869103 2294.072637 2356.276172 2418.479706 2488.455541 2550.659075 2612.86261 2675.066145 2737.269679 2799.473214 2861.676748 2923.880283 2986.083817 120 N/A MAG (dB) -35.4416 -36.7826 -38.1748 -39.4201 -40.6861 -41.8613 -43.0625 -44.117 -45.4548 -46.5135 -47.518 -48 -49.3806 -50.6682 -51.4048 -52.2944 -53.1495 -54.1121 -54.8516 -55.8666 -56.6994 -57.2524 -58.234 -58.6173 -59.5667 -60.0306 -60.7755 N/A PHASE (Deg) 80.0955 75.9424 72.1972 68.906 65.6772 62.6932 59.9237 56.8906 53.7013 51.2112 50.3351 47.0217 43.4869 44.1102 41.8138 39.7233 38.8971 36.4525 32.8334 33.1369 31.3441 29.7661 27.6247 26.7657 27.2549 27.6169 26.6101 Table C.4: Dynamic signal analysis of flexure design 2a Lt Probe/Bt Piezo INPUT (HZ) 5 14.9 24.8 35.938 45.838 55.738 65.638 75.538 85.438 95.338 96.575 105.238 115.138 125.038 134.938 1146.075 155.975 165.875 175.775 185.675 195.575 205.475 N/A INPUT (RAD/S) 31.41592654 93.61946108 155.8229956 225.8051136 288.0086481 350.2121827 412.4157172 474.6192517 536.8227863 599.0263208 606.798621 661.2298554 723.4333899 785.6369244 847.840459 917.8162937 980.0198283 1042.223363 1104.426897 1166.630432 1228.833966 1291.037501 121 N/A MAG (dB) -21.2377 -22.2101 -23.1015 -24.1525 -25.163 -26.2143 -27.4276 -28.7924 -30.2249 -31.8147 -32.0152 -33.441 -35.1449 -36.7652 -38.3836 -40.52 -42.4958 -44.0519 -45.5873 -46.5025 -48.3214 -49.4819 N/A PHASE (Deg) -20.33 -40.7702 -60.1497 -81.0327 -98.9319 -116.437 -133.152 -149.492 -164.757 -179.313 178.995 167.628 155.487 144.42 134.509 119.805 116.428 110.33 105.894 99.7541 92.5037 89.3792 Table C.5: Dynamic signal analysis of flexure design 2b Lt Probe/Bt Piezo INPUT (HZ) 215.375 225.275 235.175 245.075 254.975 264.875 274.775 285.913 295.813 305.713 315.613 325.513 335.413 345.313 355.213 365.113 375.013 386.15 396.05 405.95 415.85 425.75 435.65 445.55 455.45 465.35 475.25 485.15 N/A INPUT (RAD/S) 1353.241036 1415.44457 1477.648105 1539.851639 1602.055174 1664.258708 1726.462243 1796.444361 1858.647895 1920.85143 1983.054964 2045.258499 2107.462033 2169.665568 2231.869103 2294.072637 2356.276172 2426.252006 2488.455541 2550.659075 2612.86261 2675.066145 2737.269679 2799.473214 2861.676748 2923.880283 2986.083817 3048.287352 122 N/A MAG (dB) -50.9592 -52.1374 -53.3294 -54.3999 -55.6418 -56.5107 -57.1387 -58.6616 -59.985 -60.9484 -62.0578 -62.1311 -64.4222 -64.7992 -65.2131 -66.5474 -67.1869 -68.191 -68.6242 -69.4766 -68.3737 -69.281 -69.7738 -71.0817 -69.4183 -71.933 -71.1978 -70.3661 N/A PHASE (Deg) 83.6187 78.311 76.0191 72.4009 65.5622 66.447 61.0581 51.593 52.3132 51.1797 47.1574 39.1374 31.359 42.3933 39.6273 38.8048 38.8289 36.7647 31.0064 34.1842 29.4043 22.2619 19.7137 32.1998 21.2463 19.2854 18.0418 25.7117 Table C.6: Dynamic signal analysis of flexure design 3a Rt Probe/Tp Piezo INPUT (HZ) 5 14.9 24.8 35.938 45.838 55.738 65.638 75.538 85.438 95.338 96.575 105.238 115.138 125.038 134.938 146.075 155.975 165.875 175.775 185.675 195.575 205.475 215.375 N/A INPUT (RAD/S) 31.41592654 93.61946108 155.8229956 225.8051136 288.0086481 350.2121827 412.4157172 474.6192517 536.8227863 599.0263208 606.798621 661.2298554 723.4333899 785.6369244 847.840459 917.8162937 980.0198283 1042.223363 1104.426897 1166.630432 1228.833966 1291.037501 1353.241036 123 N/A MAG (dB) -3.38552 -4.39704 -5.42789 -6.66989 -7.84981 -9.11407 -10.486 -11.961 -13.5578 -15.2447 n/a -17.0045 -18.8038 -20.6211 -22.4333 -24.4471 -26.1873 -27.842 -29.4869 -31.1203 -32.6378 -34.1135 -35.5623 N/A PHASE (Deg) 158.696 136.485 115.961 94.3273 76.0366 58.4834 41.5271 25.4256 10.2463 -3.87665 n/a -16.8387 -28.6408 -39.3137 -48.9045 -58.5677 -66.248 -72.922 -78.995 -85.5244 -90.6466 -94.6227 -98.9881 Table C.7: Dynamic signal analysis of flexure design 3b Rt Probe/Tp Piezo INPUT (HZ) 225.275 235.175 245.075 254.975 264.875 274.775 285.913 295.813 305.713 315.613 325.513 335.413 345.313 355.213 365.113 375.013 384.913 396.05 405.95 415.85 425.75 435.65 445.55 455.45 465.35 475.25 485.15 N/A INPUT (RAD/S) 1415.44457 1477.648105 1539.851639 1602.055174 1664.258708 1726.462243 1796.444361 1858.647895 1920.85143 1983.054964 2045.258499 2107.462033 2169.665568 2231.869103 2294.072637 2356.276172 2418.479706 2488.455541 2550.659075 2612.86261 2675.066145 2737.269679 2799.473214 2861.676748 2923.880283 2986.083817 3048.287352 124 N/A MAG (dB) -36.8816 -38.2135 -39.515b -40.7845 -41.9256 -43.1502 -44.3218 -45.6116 -46.5869 -47.5013 -48.6186 -49.6428 -50.4986 -51.3656 -52.3872 -53.3435 -54.1544 -54.9458 -55.9383 -56.7783 -57.1897 -58.1033 -58.869 -59.6022 -60.8774 -60.703 -62.1952 N/A PHASE (Deg) -103.226 -106.673 -110.508 -113.588 -116.57 -120.046 -122.492 -124.632 -126.46 -128.205 -131.881 -134.772 -135.447 -138.524 -140.362 -141.62 -143.598 -144.689 -147.078 -148.695 -148.284 -150.75 -150.735 -152.236 -156.178 -158.144 -156.388 Table C.8: Dynamic signal analysis of flexure design 4a Rt Probe/Bt Piezo INPUT (HZ) 5 14.9 24.8 35.938 45.838 55.738 65.638 75.538 85.438 95.338 96.575 105.238 115.138 125.038 134.938 144.838 155.975 165.875 175.775 185.675 195.575 205.475 215.375 N/A INPUT (RAD/S) 31.41592654 93.61946108 155.8229956 225.8051136 288.0086481 350.2121827 412.4157172 474.6192517 536.8227863 599.0263208 606.798621 661.2298554 723.4333899 785.6369244 847.840459 910.0439935 980.0198283 1042.223363 1104.426897 1166.630432 1228.833966 1291.037501 1353.241036 125 N/A MAG (dB) -21.3065 -22.2421 -23.1262 -24.1829 -25.2013 -26.257 -27.5069 -28.8663 -30.3509 -31.9162 -32.1167 -33.5418 -35.2745 -36.9033 -38.5241 -40.291 -42.6603 -43.8698 -45.8687 -46.8446 -48.7036 -50.1258 -51.0322 N/A PHASE (Deg) -20.2289 -40.7964 -60.2623 -81.2439 -99.2608 -116.667 -133.382 -149.893 -165.316 -179.63 178.69 167.226 155.252 144.363 134.169 122.119 116.8 109.124 101.564 99.2181 96.0367 88.3451 83.104 Table C.9: Dynamic signal analysis of flexure design 4b Rt Probe/Bt Piezo INPUT (HZ) 225.275 235.175 245.075 254.975 264.875 274.775 285.913 295.813 305.713 315.613 325.513 335.413 345.313 355.213 365.113 375.013 384.913 396.05 405.95 415.85 425.75 435.65 445.55 455.45 465.35 475.25 485.15 N/A INPUT (RAD/S) 1415.44457 1477.648105 1539.851639 1602.055174 1664.258708 1726.462243 1796.444361 1858.647895 1920.85143 1983.054964 2045.258499 2107.462033 2169.665568 2231.869103 2294.072637 2356.276172 2418.479706 2488.455541 2550.659075 2612.86261 2675.066145 2737.269679 2799.473214 2861.676748 2923.880283 2986.083817 3048.287352 126 N/A MAG (dB) -52.4258 -53.5037 -54.806 -55.3729 -57.0411 -57.3818 -58.8365 -60.0914 -60.5119 -61.8403 -61.9458 -62.9584 -65.2482 -65.7363 -66.7573 -67.3654 -67.9877 -68.7795 -67.6922 -69.9703 -69.3106 -70.3536 -69.5599 -70.6195 -72.3766 -69.9427 -70.5405 N/A PHASE (Deg) 78.1209 75.5326 69.9195 68.5061 65.4402 63.369 50.686 56.049 54.8751 52.8149 43.5044 31.8963 31.5821 32.5532 34.1446 33.9895 37.6903 33.5132 31.2689 24.671 17.1439 21.7782 18.8388 11.8233 24.4627 23.4534 12.8642 Table C.10: Dynamic signal analysis of MPF 1 Hz 0.1 0.1998 0.2994 0.4034 0.5045 0.6046 0.7093 0.8059 0.9061 1.008 1.097 1.372 1.929 2.337 3.115 3.855 4.977 5.715 6.923 7.7 8.938 9.837 10.94 11.42 15.38 21.17 24.57 Mag(Log db) 0.034822 0.0322643 0.031687 0.0308083 0.0294867 0.0293718 0.0291263 0.0290577 0.0289469 0.028719 0.0286832 0.0279589 0.0274643 0.0274348 0.0269128 0.0261942 0.0260554 0.0254092 0.0252998 0.0253432 0.0249462 0.024466 0.0238369 0.0234754 0.0229894 0.021087 0.0207078 Phae(Deg) 169.996 172.111 173.941 172.236 170.76 172.174 170.854 171.624 171.628 171.891 171.696 171.325 170.58 169.704 167.71 167.528 165.527 164.225 162.943 162.787 159.832 158.292 155.999 156.053 147.838 140.572 133.398 127 K(dB) 21.66965493 21.54241042 21.5137935 21.47030904 21.40507213 21.3994098 21.38731646 21.38393845 21.37848354 21.36726792 21.36550664 21.32990375 21.30562583 21.30417867 21.27858751 21.24340827 21.23661998 21.20504487 21.19970393 21.20182257 21.18245028 21.15904176 21.12841388 21.11083424 21.08722324 20.99505398 20.97673036 Table C.11: Dynamic signal analysis of MPF 2 32.07 35.67 39.26 46.06 53.46 61.39 64.75 73.57 77.6 88.17 107.9 112.6 117.5 122.6 129.4 133.5 148.6 189.8 217.9 242.4 293.6 337.2 371.1 430.8 469.1 500 0.0172708 0.0162759 0.0185403 0.0143372 0.0146369 0.0103712 0.00807178 0.0029962 0.00724308 0.0343488 0.0128517 0.00533539 0.0036433 0.0058238 0.00671831 0.00824914 0.0212462 0.00814071 0.00305248 0.00232058 0.00145735 0.00103618 0.000838793 0.000830028 0.000492234 0.000304348 126.725 126.854 124.429 110.705 103.392 87.7254 84.7341 145.187 177.805 127.416 129.2952 114.812 93.3213 80.5108 103.906 114.362 58.2951 -21.683 -42.724 -23.035 -42.918 -37.8 -47.391 -62.1 -87.954 -84.628 128 20.811376 20.76375499 20.87229939 20.67127172 20.68554157 20.48335989 20.37519509 20.13845717 20.33635318 21.6460569 20.60068662 20.24721925 20.16848589 20.27000219 20.31179504 20.38351774 21.00275158 20.37842924 20.14106707 20.10715268 20.06722618 20.04777482 20.03866517 20.03826075 20.02268106 20.01402066 Appendix D Material List and Component Characteristics 129 Table D.1: Parts Part Name Piezactuator A Piezoactuator B Sensor C Sensor D Flexure E Flexure F Tube Housing(0.75" ID, 1" OD) G Sealant Region H Plates(average size: 5 x 12 x 5mm) I Tube Claw J (s) Sensor Contact K Tube Housing(0.75" ID, 1" OD) L Piezo-Holder M Sensor-Holder N Tube Claw 0 Sensor Contact P 0.75" ID Filter Chamber Q 0.75" to 0.375" Tubing Reducer R 0.375" to 0.25" Tubing Reducer S Pump Connector T Material Type PZT PZT Stainless Steel Stainless Steel Aluminum Aluminum Vinyl Silicone Fused Silica Quartz Aluminum Aluminum Vinyl Aluminum Aluminum Aluminum Aluminum Vinyl Nylon Nylon Nylon Table D.2: Probe characteristics [1] Range +p m (in) Standoff pi m (in) Resolution nm (p in) Sensor Diameter (mm) Overall Diameter (mm) Minimum Recommended Target Size (mm) Capacitance transducers Non-contact No probe wear 130 50(0.002) 100(0.004) 0.5(0.02) 5 8 10 Coaxial passive probes Non-destructive Easily fixtured Table D.3: Piezoactuator characteristics [7] Type Max Voltage Input (V) PZT 150/5/20 VS1O +150 PZT 150/4/20 VS09 Max Stroke (p m) 20 20 Prestress Force (N) Mechanical Compressive Load (N) Length (mm) Capacitance (nF) Stiffness (L) Resonance Frequency (Hz) 150 1000 28 800 25 30 100 500 28 340 12 30 131 +150