Design of An Assist Device for... Repositioning of Bedridden Patients

Design of An Assist Device for Automated Rolling and
Repositioning of Bedridden Patients
by
Arin Basmajian
B.S., Mechanical Engineering
Massachusetts Institute of Technology, 2000
Submitted to the Department of Mechanical Engineering
in Partial Fulfillment of the Requirements for the Degree of
Master of Science in Mechanical Engineering
BARKER
OF TECHNOLOGY
at the
Massachusetts Institute of Technology
June 2002
OCT 2 5 2002
LIBRARIES
@ Massachusetts Institute of Technology, 2002. All Rights Reserved.
Author ...............................
Depaarennte0 Mechanical Engineering
May 6, 2002
Certified by ..........................
Ilaruhiko H. Asada
Ford Professor Of Mechanical Engineering
Thesis Supervisor
Certified by ..............................
Emesto E. Blanco
Adjunct Professor of Mechanical Engineering
Thesis Supervisor
Accepted by ................................
Ain A. Sonin
Chairman, Department Committee on Graduate Students
2
Design of An Assist Device for
Automated Rolling and Repositioning of Bedridden Patients
by
Arin Basmajian
Submitted to the Department of
Mechanical Engineering
on May 6, 2002
in partial fulfillment of the requirements for the degree of
Master of Science in Mechanical Engineering
A novel design for rolling and repositioning a bedridden patient is presented. A pair of actuated rollers are attached to both sides of the bedsheet in order to lift and manipulate the
patient body. The patient is gently supported on a hammock-like active bedsheet, creating virtually no shear forces on the patient body. Therefore, the patient can safely and easily be
maneuvered with minimal physical assistance by the caregiver. This prevents development of
painful bedsores and of pneumonia. In this thesis, functional requirements and design issues
are addressed, followed by basic design concept, mechanism, and control strategy. A simple
kinematic model is built to create trajectories for generation of desired body motion. A prototype is built and initial experiments demonstrate that a human body is rolled and manipulated
by the assist device, providing proof of concept. Extended functionalities including bed-tochair and bed-to-bed transfers, are described at the end.
Thesis Supervisor: Haruhiko H. Asada
Title: Ford Professor of Mechanical Engineering
Thesis Supervisor: Ernesto E. Blanco
Title: Adjunct Professor of Mechanical Engineering
3
4
Acknowledgements
I would like to thank my grandfather who told me that I had to follow my
dream. I wish he could be here to see it come true. I would like to thank my
parents and brother, who contributed greatly to my being here today; without
their unconditional love -- even during times when I was driving them crazy
-- I could not have made it through. Thanks also to my aunt, Houry, for her
endless support and patience.
A warm thank you goes to both Professor Harry Asada and Professor Ernesto
Blanco for giving me such an opportunity to work on a truly exciting project
and especially for teaching and guiding me along the way. I would also like to
thank Professor Sarma for taking time out and being a mentor to me during
this eventful year.
This thesis would not have been written without the support of my roommates, Kavita and Janelle. You were my strength when I ran out of my own.
Thank you.
A big thank you to all my friends, you know who you are, who reminded me
of life beyond work, who believed in me, and who stood by the various decisions I made this year, yes, Jomaldo and Dinesh, that goes to you in particular.
5
Thanks to Petros, for his help, both in the formatting of this thesis and otherwise. I would be very ungrateful if I did not thank my labmates in the d'Arbeloff Lab, in particular, KyuJin -- for his intense dedication to research and for
heading the camera crew -- Binayak, Eric, Mikii, Phil and everyone else for
being labmates, friends, advisors, critics. Oh, and for tolerating my out-oftune singing!
Thanks to the machinists in the Central Machine Shop, and Steve in the Pappalardo Lab, for instructing me not just on machining and the art of painting
but for all their work and support too. I truly appreciate it.
I may have typed this thesis on my own but getting here was a team effort.
Thank you!!!
6
Table of Contents
11
CHAPTER 1
Introduction and Background .....................
CHAPTER 2
FunctionalRequirements and Design Concept ........ 15
(1) Functional Requirements ........
..........................
(2) Design Concept ...............
. ..
(3) Operation.....................
Airborne Rolling................
Pivotal Rolling .................
Horizontal Translation .............
Changing Bedsheets ...............
CHAPTER 3
. 15
....... 16
........... ........... ....... 19
. . . . . . ..... ..... 1. 20
..... ......
....... 21
........... ...........
....... 21
........... ...........
....... 22
..
..
..
. . ..
..
..
..
.
System Modeling and Analysis ....................
Modeling the Problem .................................
Determining Location of Patient............................
25
.......
.......
25
29
7
CHAPTER 4
Motion Planning..............................
33
... 33
.... 33
.... 36
.... 38
Trajectory Generation ....................................
Rolling ...................................................
Bed-Bed/Chair Transfer......................................
Control ofActuators.........................................
CHAPTER 5
Implementation of FirstPrototype ..........
.. . . . . 41
41
(1) Mechanical Design and Electronics ..........................
Selection of Materials and Determination of Design....................
Choosing the drive shaft .......................................
Electronics ....................................................
43
43
47
(2) Verification of Concept................................
50
(3) Verification of System Model...............................
55
System Verification..............................................56
PatientTranslation..............................................59
CHAPTER 6
(4) Control Software Implementation ...........................
60
Control Code for Rolling and Transfer ..............................
63
Conclusion and Future Work .....................
65
Discussion ..............................................
Future Work.............................................
Future Work regardingDesign and Structure of Device ..............
Future Work with respect to Reinforcing Bedsheets..................
Future Work regardingElectrical Circuitry........................
.. 65
.. 66
... 66
... 68
... 69
References....................................71
APPENDIX A
8
HistoricalRecord of Approaches ..................
73
Chronology of Designs ....................................
.. 73
Honeycomb M otion...........................................
Belt and Belt................................................
... 73
... 76
Tilt and belt...................................................
Tilt and tilt....................................................
ChangingBedsheets ............................................
76
78
81
Axial Positioning...............................................
82
APPENDIX B
Assembly Drawings............................
85
APPENDIX C
PartDrawings and Listing........................
89
Part Info...................................................89
APPENDIX D
Hardware,Software and Common Operation Tips ....
123
Tips on Using the Hardware and Software of Machine..............123
To Improve Performanceof Gearboxes.............................
HardwareConsiderations ......................................
Operation Considerations.......................................
123
125
125
9
10
CHAPTER 1
Introduction and
Background
As people get older, their mobility becomes impaired and they need
assistance performing everyday tasks. In many instances, they are bedridden
and may develop, in particular, bedsores and pneumonia. Bedsores are formed
due to the continuous application of pressure on a person's body part. Therefore, redistribution of pressure applied to the body may be taken as a preventive measure. Also, a patient's own motion helps to reestablish blood flow and
so prevents cell death.
Repositioning the patient is not only advantageous for prevention of
bedsores but also for preventing pneumonia. Pneumonia is an infection of one
or both lungs in which fluid and damaged lung cells fill the air spaces in the
lungs, making it difficult to breathe. Viral pneumonia, generally milder than
the bacterial form, is the result of lower respiratory infection and has been the
cause of more than 90% of deaths for individuals over 65 [1]. Stagnant mucus
in the lower airways is a medium for bacterial growth, should pathogens reach
11
Introduction and Background
the lower airways. Routine turning and positioning assists in mobilization of
secretions. Patients having breathing difficulties may find relief by being
turned slightly on their side [3]. Caregivers move patients manually, a highly
labor-intensive task. According to a survey by Garg et al [6], a nursing assistant performs over 50 such tasks per eight-hour shift. This results in alarmingly high numbers of back injuries that cost $24 billion each year [8].
In this thesis, a new methodology will be developed to assist patients
in turning and repositioning. Both patients and nursing personnel will benefit
from the proposed design. Bedsheets - with the aid of the marionette device
presented in this paper - may be used to roll patients. Turning helps prevent
bedsores. The patient is turned with almost no physical exertion or risk of
injury. Only one caregiver will be able to turn the patient easily. By tilting a
sheet laterally, rolling a patient becomes an easy task. It not only causes a
change in the person's position, thereby altering the pressure but also stimulates blood flow. The patient may be moved to either edge of the bed. Bedsheets can easily be changed and nursing personnel no longer need to pull
patients on the bed so reducing their own backache.
There are a few commercially available beds that fulfill some requirements of patient and nursing personnel. Hill-Rom [4] and Kinetic Concepts
Inc. (KCI) [5] are some of the companies that make hospital beds. Beds range
from manually-operated to electrical ones, some even equipped with fluidized
therapy units that provide pressure relief. However, their cost is exceedingly
high and the beds are big and bulky. The device designed has the main advantage of having almost no set-up time. The hoists that are currently on the market, such as the Sara-Lift', often remain unused in hospitals and nursing
12
homes. Nursing personnel complain that the setup time is too long. Caregivers
do not wish to go through the trouble of setting up the equipment, which can
sometimes take up to 20 minutes when the actual task would take a fraction of
that time. Furthermore, patients feel unsafe being completely airborne when
using these lifts.
This thesis explores a new approach and presents a device that uses
bedsheets to maneuver the human body. Using the bedsheet is beneficial to the
patient because it does not produce large stress concentrations across the
body, and it reduces the harmful shear forces. In addition, it is flexible so it
adheres to the shape of the person. The remainder of this thesis is organized as
follows. Chapter 2 deals with the methodology of the approach and the actual
design. Chapter 3 presents a kinematic model. Chapter 4 introduces the analysis and control of the machine. Chapter 5 illustrates its implementation. Chapter 6 suggests extended functionalities with varying configurations of the
design, and summarizes the important aspects of the design and how it may be
improved respectively.
13
Introduction and Background
14
CHAPTER 2
Functional Requirements
and Design Concept
(1) FunctionalRequirements
Caregivers roll patients from one side of the bed to the other when
changing bedsheets because rolling patients is less laborious for nursing personnel than pulling or lifting. Rolling is also used to reposition patients every
few hours. The patient may be positioned on one side and may then be rolled
onto his back and onto his other side to change the pressure being applied to
him. Therefore, rolling is a common practice that has been widely used in caring for bedridden patients. The objective of the project was to replace this
manual operation by a robotic device to assist nursing personnel in rolling and
repositioning patients. The design must satisfy the following requirements
whilst maintaining patient comfort:
* Patients must be repositioned to redistribute the pressure applied to their
bodies
15
Functional Requirements and Design Concept
* Patients must be transported from side to side to allow the bedsheets to be
changed easily
Shear forces applied to patients' bodies must be minimized. Patients
may accidentally get sheet burns as caregivers may pull them from one side of
the bed to the other to transport them. One of the laborious tasks caregivers
have to face is that of changing bedsheets. Another desirable function of the
device is the transfer of a patient from his/her current location to another one,
that is, from the bed he/she is lying on to another bed or chair.
Some of the important factors that should also be taken into consideration are quick setup times, safety and acceptability. The setup time of the
device should be small because caregivers may not wish to spend too much
time setting up a machine for doing a task that requires a fraction of the time
required to set it up. Patients' safety is a priority. Therefore, the device should
conform to safety standards. It should also be deemed acceptable by patients
and caregivers in particular in terms of its appearance and function.
(2) Design Concept
Figure 1 shows the basic design concept for rolling and translating the
patient. It consists of a pair of powered rollers to which the bedsheet is
attached on top of which the patient lies upon. The patient is suspended and
supported between bedsheets as on a stretcher. Unlike a stretcher, the bedsheets are automatically wound off the rollers.
16
(2) Design Concept
FIGURE 1.
Patient suspended on Hammock-like Sheet
The design consists of a pair of parallel, independently-actuated rollers that are positioned above the bed. The rollers may be positioned using rods
or arms that may be moved in two-dimensional vertical planes. The structure
may either be part of the bed itself or may form a separate portable system.
Each end of a bedsheet will be attached to a roller.
R2
Al
FIGURE 2.
A2
Hammock Skeleton
17
Functional Requirements and Design Concept
It was determined that arms with one degree of freedom causing
rotary motion would be sufficient for the required set of tasks that are previously mentioned. Figure 2 shows a schematic of two crossed arms supporting
rollers onto which the sheet is attached. The arms would be actuated at AIA 2
and the rollers at R1 R 2. The system has four degrees of freedom, the rotation
of the arms, 01 and 02, and the rotation of the rollers, $1
01
FIGURE
3.
and 02-
1'0
Person lying flat waiting to be rolled.
The trajectory caused by the rotation of the arms is represented by the
dashed arc shown in Figure 3 which is a side view of the bed system. The
arms will be independently actuated so the horizontal and vertical distances
between the rollers may vary, from almost touching one another to being at
the ends of the bed.
There are two modes of operation: rolling - airborne and pivotal and horizontal translation of the patient for both repositioning and transfer.
18
(2) Design Concept
(3) Operation
Rolling
Refer to Figures 4 and 5 for sketches of rolling and how it would
occur.
P
FIGURE
4.
Person tilted by sheet action
Initially, the person lies flat on the bed. The bedsheets on either side
of the person are attached to the rollers. This may be accomplished by attaching strips of VelcroTM to the ends of the sheets and the rollers.
Once the sheets are attached to the respective rollers, then say the left
one, is actuated in the counter clockwise direction. See Figure 4. This allows
the sheet to be wrapped and the left end of the sheet decreases in length causing the person to tilt and pivot about point P as shown in Figure 4. As the
roller is actuated, the sheet wraps further and the person is tilted at a greater
angle.
19
Functional Requirements and Design Concept
FIGURE
5.
Person rolled by motion of left roller towards the right.
In order to roll the person completely, the left roller may be moved
towards the right as shown in Figure 5. This results in the patient lying on his
side. Further motion of the left roller towards the right would allow complete
rolling of the patient. It would require coordination with the right roller that
would have to unwind, providing slack in the sheet so the patient would be
able to roll onto it. Figures 4 and 5 show that rolling and repositioning of the
patient may be accomplished.
Airborne Rolling: This is the case when the person is completely lifted off
the bed surface. No contact is made by the person onto the bed surface at
point P unlike Figures 4 and 5. Airborne rolling allows the person to be lifted
and transported to either edge of the bed if either of the rollers is translated.
The lifting motion eliminates harmful shear forces exerted on a patient's back
as compared to him being dragged from one edge of the bed to the other.
However, airborne rolling is scary for patients even though it allows complete
control due to the cradling effect of the bedsheets. Airborne rolling may also
be used to roll the patient in place, that is, he would not need to be translated
to create space for rolling. Even if his initial position lies close to the edge of
20
(2) Design Concept
the bed, he may roll in place since the rolling action will take place off the bed
surface and he would then be lowered to his original position on the bed.
Pivotal Rolling: Pivotal rolling may be more acceptable for patients. It is
demonstrated in Figure 4. During pivotal rolling, the patient pivots about a
shoulder which remains in contact with the bed surface. Patients feel comforted by the fact that they are not completely airborne though they will experience normal and shear forces.
Furthermore, using the bedsheet as the "tilting agent" results in the
forces exerted on the body being distributed over a larger surface area thereby
lowering the stress concentration. Supporting the patient in the sheet accounts
for patient comfort and safety. The concept allows the device to be on casters
hence making it portable and allowing a single module to be used for a number of hospital beds.
Horizontal Translation
Horizontal translation of the patient is achieved by coordinating the
movement of the arms and the rollers. Figure 6 shows the patient initially
positioned on the left edge of the bed and finally translated to the right
through coordination of both roller and arm motions.
The angles through which the arms rotate are controlled by the actuators at the base of the bed. Assuming the initial lengths of the sheet on either
side of the person are known, then the lengths of the sheets are known at every
instant in time since the roller rotations are controllable. Thus, four parameters are controlled: 0,
02,
P1 and 02 The position of the rollers is uniquely
21
Functional Requirements and Design Concept
P1
FIGURE 6.
P2
Horizontal Translation of the Patient.
determined as is the location of the patient. The actuators at the rollers alone
can cause rolling of the patient.
If the arms are long enough to be positioned at the edge of a chair that
is in line with the bed, then combining the motion of the arms may allow
transfer of the patient to either edge of the bed and then onto the awaiting
chair. The sequence of events is described in the next section where a preliminary model is introduced to obtain an understanding of the combination of
actuator motions required to yield the desired motions of the patient.
Changing Bedsheets
An added functionality of the device is that it may be used to change
bedsheets. The rolling actions described above may be used to change them.
The new bedsheet may be loaded onto a roller. The old sheet would be lifted
and attached via VelcroTM strips to the new sheet. The rollers would be actuated in the same direction, either both clockwise or both anticlockwise. This
would enable the patient to be continuously rolled a few times until the old
sheet is removed and he is lying on top of the new sheet. At this instant, the
22
(2) Design Concept
rollers would slowly be brought to rest, resulting in the patient lying on top of
the new sheet.
23
Functional Requirements and Design Concept
24
CHAPTER 3
System Modeling and
Analysis
Modeling the Problem
The kinematic model developed will be used to analyze the behavior
of the system and obtain relationships between actuator movements and
patient motion.
As shown in Figure 7, the reference axes are fixed at the center of the
bed, at the height where the actuators to control the arms are placed. As mentioned in a previous chapter, the rotation of the arms are given by 01 and 02,
and the roller rotations by $ and 02-
25
--
I
-
-
-
-.---~
-
System Modeling and Analysis
D
9P2
(P
I 2
S
FIGURE 7.
S
Preliminary Model
The difference in height between the rollers is defined as H, and the
horizontal distance between them is D. The angles made by the sheet on either
side of the patient are respectively a and P . His angle of tilt is y with his
width given by w. L, and L 2 are the lengths of the sheet on either side of the
patient and T, and T2 are the tensions in the sheet. Point M is the point where
the forces, that is, the weight of the patient and the tensions in the sheet, pass
through.
26
--- ~-
.-
-
Modeling the Problem
The following table summarizes the parameters used.
TABLE 1. List of Parameters
Symbol
Description
cc
angle made by left edge of person with left roller
P
angle made by right edge of person with right roller
LI
length of bedsheet from right edge of person to
right roller
L2
length of bedsheet from left edge of person to left
roller
01
angle made by left arm with the horizontal
02
angle made by right arm with the horizontal
T,
tension in left sheet
T2
tension in right sheet
1
length of each arm
Ws
width of sheet
w
total width of person
y
tilting angle of person
XRR
x-coordinate of right roller
YRR
y-coordinate of right roller
s
horizontal distance from pivot point of either arm
to center of bed
D
center-to-center horizontal distance between rollers
H
center-to-center vertical distance between rollers
01
rotation of left roller
02
rotation of right roller
27
System Modeling and Analysis
Symbol
Description
XCM
x-coordinate of patient's center of mass
YCM
y-coordinate of patient's center of mass
z
height from origin, to meeting point, M, of the
three forces
Although the problem is actually three-dimensional, it has been
reduced to a two-dimensional problem by taking a vertical cross-section. A
limitation is that if the patient is positioned diagonally across the bed, then the
model does not hold as the force distribution is altered. The model is based on
a number of assumptions. It is assumed that the patient suspended by the bedsheets is in quasi-static equilibrium, that is, the acceleration of the person is so
small that inertial force is considered to be negligible. It is also assumed that
the person is airborne. Both normal and shear forces acting by the bed on the
body are equal to zero as the body is suspended by the bedsheets alone. This
assumption will later be relaxed to obtain patient reassurance as some contact
between the patient and the bed is established. Bedsheet elongation is
assumed to be negligible. The final assumption is the initial knowledge of the
values for the lengths LI, L 2 , and the width w. The geometry of the given
problem results in the following kinematic relations. Considering the vertical
distance between rollers,
H = wsiny+L 2 sinf-Ljsin c
,
(1)
where H is a known value given by the geometry to be
H =
28
l(sinO 1 - sinO2 )
,
(2)
Modeling the Problem
and 1, 01 and
02
are known variables. Considering the horizontal distance
between rollers, we obtain the relation
D = wcosy + L 2cosP+ L, cosa
,
(3)
where D is known in terms of known variables as
D = l(cos0 1 + cos02 )-2s
.
(4)
The y-coordinate of the person's center of mass is given by:
YCM = -Lisinca+
Wsiny
(5)
Determining Location of Patient
If the location of the patient is known at every instant, then we can
control the trajectory of the patient and position him as desired. Two variables
would be enough to define the patient's location. The position of his center of
mass, yCM, and his angle of tilt, y, would fully define his position.
Taking moments about point P, the following equation is obtained:
T 1wsin(cc-y)-MgwcosF = 0,
(6)
A force balance in the vertical direction gives
Tisina+ T 2 sinp = Mg,
(7)
A force balance in the horizontal direction yields
Ticosa = T2 cos$.
(8)
29
System Modeling and Analysis
Given three equations, that is, Equations (6), (7), and (8), and three
unknowns, the unknowns may be solved for. The known parameters are M, w,
TI, and T2. The latter two parameters will be known as they will be measured
using sensors and they will be estimated as a fraction of the person's weight.
They will alter as functions of the angles made by the person with either side
of the sheet which is how they help to obtain these values for different
instances in time. The unknowns are, a,
P, and y. The most important infor-
mation is the value of y which is thus obtained.
In terms of the given variables, the x-coordinate for the center of mass
of the patient is obtained by the following equation,
XCM
=
xRR(
L 2 COSP+
(9)
WCOSIJ,
where XRR, the x-coordinate of the right roller, is known to be
XRR =
(10)
lcosO 1 - s
due to the positioning of the arms which depend on the angles
01 and 02-
Since the y-coordinate of the right roller is also known given as
YRR
=
lsinO,
(11)
then the length, L 2 , may be obtained from the equation,
L2 =
sin@
.
(12)
Therefore, two additional equations, Equations (9) and (12), are
obtained with two additional unknowns, L 2 and xCM. They may be solved for
and the location of the person's center of mass is defined.
30
Modeling the Problem
The overall equations used are rearranged as follows:
wsiny +
L, sinox =
YRR -
WCosy + YRRT + L,
ST2,
l(sin0,
- sin0 2 )
coscx = (CosOI+Cos02) - 2s
a = asin MgWCosY +y
(13)
(14)
(15)
2T
where T, is a function of y and is also a fraction of the patient's weight. The
equations obtained are non-linear. Therefore, they may be solved by either
numerical analysis or by iterative methods for angle, y, and length L, in terms
of known parameters, 01 and
02.
Since L, and L2 will be known, then the
rotation of the rollers are related using the following equation
+02+
L, + L2 = Ws --
,(16)
where wS is the width of the sheet. The equations shows that, to control the
device, it is necessary to have only three degrees of freedom. The roller rotations are coupled. Therefore, there exists an additional -- redundant -- degree
of freedom.
This analysis allows the complete monitoring of the patient and determination of his position. The model is rather simplistic, in particular since it
does not take into account shear forces that are experienced by the patient at
the pivoting contact area as well as the normal forces acting against the
patient's body when he is in contact with the bed surface. However, it will be
shown in Chapter 5 that this model does indeed come close to the actual posi-
31
System Modeling and Analysis
tioning of the patient. The system model developed here will be modified to
accommodate for a more empirical model in Chapter 5.
The additional degree of freedom provided by the design may be useful as a controllable parameter may be altered to provide greater accessibility
of the patient by nursing personnel. It may also be used to improve the efficiency of the system, that is, the line of force for one of the sheets may act
along the corresponding arm thereby decreasing the torque applied by the
sheet to zero, and minimizing the energy required for the system.
There are two modes of operation, namely rolling and horizontal
translation. Horizontal translation of the patient will be used to position the
patient on the bed for rolling to occur.
The patient must be accurately positioned on the bed in order to avoid
tipping him over the edge of the bed when rolling occurs. Rolling is desired.
This too, may be subdivided into two parts - complete rolling and half-rolling. The former is when the person is rolled through 180 degrees, that is, from
lying on his back to lying on his stomach or vice versa, and the latter is when
the person is rotated by 90 degrees and is positioned on his side. Horizontal
translation requires the use of all four actuators that will be coordinated
together to achieve the desired motion. Rolling only requires the use of the
actuators at the rollers, that is, the control of 01 and 02. The parameters may
be mapped using the Jacobian matrix to relate the controllable parameters,
[01
02 01
2] '
to the coordinates of the desired ones [XCM y c L2] . The fol-
lowing chapter discusses the planning of the motion based on the model presented here.
32
CHAPTER 4
Motion Planning
Trajectory Generation
A number of programs have been written in order to control the
sequence of motion. The two main types are either for complete rolling of the
patient or to transfer a patient from the bed he is laying on to another bed or
chair. The programs may have their values adjusted for different patients. The
following sketches show the sequence of events in each mode of operation
upon which the formulated trajectory is based.
Rolling
Initially, the device is at rest. The arms are positioned in their home or
zero position, that is, they are extended to the maximum.
33
Motion Planning
I
z
~
p1
FIGURE 8.
p2
Arms at Home Position.
Then, the arms are positioned according to the location of the patient
on the bed as well as his width.
i-a---
Pi
FIGURE 9.
P2
Arms Positioned
The actuators controlling the bedsheet motion are switched on. They
rotate in the same direction causing the patient to be slowly pivoted about one
shoulder. As shown in Figure 10, both the rollers rotate clockwise.
34
Trajectory Generation
P1
P2
FIGURE 10. Rollers actuated in the same direction.
The sheet is taut at the right edge of the person and correspondingly,
slack is introduced in the left end of the sheet to allow the person to roll over
onto the left end. The next figure shows that while the right roller causes tension in the sheet, the left continues introducing slack and meanwhile, the right
arm may move left to aid the cradling effect for the patient and allow slow yet
precise rolling of the patient.
FIGURE 11. Person begins to tilt
35
Motion Planning
The program would then command the entire system, that is, all four
actuators to come to a complete stop to allow caregivers to detach the bedsheets from the rollers. The arms then return to their home positions.
Bed-Bed/Chair Transfer
As in the case of rolling, initially, the arms are placed in their home
positions.
P
P2
FIGURE 12. Patient at "Home" Position
Then, one arm sweeps out to the edge of the bed/chair to which the
patient will be transferred. See Figure 13. The rollers rotate in the same direction. This causes translation of the patient. The sketches show that the rollers
move clockwise, causing the patient to move rightwards.
36
Trajectory Generation
P, P2
FIGURE 13. Roller Actuation causes Translation.
As the patient nears the edge of the bed, the right arm moves further
right to cause cradling of the patient and to enable transfer onto the bed/chair.
See Figure 14.
P,
P
2
FIGURE 14. Coordination of Roller and Arm motion allows Translation.
The right roller continues to turn clockwise, pulling the patient. This
motion is coordinated with the motion of the arm which travels further
37
Motion Planning
towards the right and aids the transfer of the patient from the bed he was originally laying on to the new bed/chair.
Control of Actuators
Each roller and arm is assigned to an actuator. The actuators may be
separately controlled in manual mode. Or, a pre-programmed file may be executed for the simultaneous motion of both arms and rollers. If a program
sequence is to be used, then all the actuators are first initialized. Equations
(13), (14), and (15) have been differentiated with respect to time and the
velocities are given by the following relations.
(-LI)cosa
cosina-+y)
Cos(a-Y)
sina
wcosy
-sin
-wsiny
coscc
-cos(c-y)+Mg
Wcosy
0
_
0
+L
(7Cos)
cosy
For different values of the angle of tilt where the weight is assumed to
be evenly distributed, that is, T,
=
T 2 , a number of data points have been theo-
retically obtained. They illustrate that the coupled roller rotations yield a magnitude in the range of 6 to 8 inches for a subject of width, 15 inches, and the
positioning of the arms equally at 600. Sample trajectory has been generated.
38
Trajectory Generation
Sample Trajectory for Rolling, Arms Fixed at Specified Angle
7
6
0
W
-3
0
2
1
u'
1
2
3
4
5
Rotation of Right Roller, Phil
6
7
8
FIGURE 15. Sample Trajectory for Rolling Patient
If AC,, corresponds to the amount of sheet wrapped on either roller,
then initially, AC 1 and AC
2
are positive to enable the sheets to be taut on either
side of the patient. Once the sheet is tight, the patient is firmly supported, then
AC 1 increases whereas AC
2
decreases. As sheet is wrapped around one roller,
slack is introduced in the other. Slack introduction appears as a negative
amount of wrap.
Figure 16 gives a simplified version of bed transfer where it is
assumed that if both arms are kept horizontal and the rollers are actuated, then
the sheet wrapping acts almost as a conveyor belt along the width of the bed.
39
Motion Planning
Sample Trajectory for Translation, Arms Fixed at Specified Angle
20
*
18
*
*
16
14
*
*
12
10
16
00
*
8
*
0f
6
*
4
*
2
0
2
4
6
8
10
12
Rotation of Right Roller, Phil
14
16
18
20
FIGURE 16. Sample Trajectory for Transferring Patient
The graph shows that as the sheet is wrapped around one roller, it is
correspondingly unwrapped from the other exhibiting a simple linear relationship.
For a more interactive version of this prototype, communication
between the sensors and the program commands will be established to enable
response to various actions of the patient.
40
CHAPTER 5
Implementation of First
Prototype
This chapter will consist of the following sections:
(1) Mechanical Design and Electronics
(2) Verification of Concept
(3) Verification of System Model
(4) Control Software Implementation
(1) Mechanical Design and Electronics
An initial prototype of the device has been built and tested. It does
indeed allow rolling of patients. It has almost no setup time, only requiring the
sticking of the bedsheet onto the roller via VelcroTM. It demonstrates both the
ease with which the device may be set up above the bed and with which the
patient may be transported.
41
Implementation of First Prototype
5
7
(2y
1 Frame
2 Gearmount
3 Roller
4 Arm, Type 1
5 Arm, Type 2
6 Gearbox, LP-90
7 Gearbox, LP- 155
8 McMC Spur Gears
9 Bottom Joi ner
10 Edge Joiner
1 1 Leg, Type 1
12 Leg, Type 2
FIGURE 17. Design of Rolling Bed.
42
(1) Mechanical Design and Electronics
The design shown in the figure above has been chosen because it
allows variation in both width and height without a significant increase in the
cost of actuation. See Appendices B and C for complete assembly and part
drawings. When the arms are rotated, not only do they change the height
between the rollers and the patient but they also vary the width between the
rollers. The former facilitates bed-bed or bed-chair transfer and the latter
accommodates for different-sized patients.
Selection of Materials and Determination of Design
A box frame is built for this prototype to demonstrate how the
machine would work on a regular bed. The frame is made in order to have a
box-spring and mattress placed upon it. It is made up of square hollow steel
tubes welded to one another to cover the drive tube, support its load via a
bushing, and prevent any objects from either jamming or interfering in any
manner with the driving of the arms. The final prototype could have the
machine on casters so that it may be portable and may be wheeled around different beds.
Choosing the drive shaft: The drive shaft is made of hollow stainless steel
tube with either end welded to a solid steel shaft having the same diameter,
1.75 in, but different lengths. The front and back solid steel shafts have keyways to allow transmission of torque to either of the arms as well as to the follower gears. Both the arm ends and the gears have set screws to locate the
position of the arms and gears on the drive shaft.
The hollow steel drive shaft was chosen to minimize deflection at the
furthest end. The values were determined using the following equation,
43
Implementation of First Prototype
0 =
GJ
,
(18)
where 0 is the deflection measured in radians, T is the torque, I is the
length of the shaft, G is the modulus of rigidity of the material, and J, the
moment of inertia is given by the equation
=
-
.
(19)
DOWt is the outer diameter of the tube and Din is the inner diameter of
the tube. A hollow tube is chosen instead of using a solid shaft because a hollow tube has a larger moment of inertia than a solid one.
Steel is chosen for the tube because its properties suit the application.
It has a large modulus of rigidity. Hardened steel would be ideal to minimize
deflection. However, due to the large difference in cost, stainless steel was
decided upon as a more reasonable option.
The deflection equation yields a range of deflection between 1-2*.
Given that the length of the arm is 41.60 in, then, using the tangent of the
deflection gives the deflection of one arm with respect to the other resulting in
an absolute deflection in a range of 0.7-1.4 in. This amount of deflection, and
hence misalignment between either end of the rollers, is deemed acceptable. It
is expected to be absorbed by the limited elasticity of the aluminum rollers,
the self-aligning bearings, as well as the flexible couplings.
Sleeves may be placed on the end of the drive shaft on the gearmount
side to prevent the shaft from slipping away. Sleeves will not be placed on the
other side of the shafts so as not to overconstrain it.
44
(1) Mechanical Design and Electronics
Hollow rectangular steel tube is used for the arms to minimize both
torsion and weight in the arms. The lower ends of the arms are made of steel
and are welded to the hollow rectangular steel arms. The upper ends are also
made of steel for welding. However, care is taken to ensure that they weigh as
little as possible. This is to minimize the torque at the base of the structure.
Even a small weight difference would cause a significant increase in the
torque because, when multiplied by a large moment arm, the torque value
shoots up.
On the gearmount side, the upper ends of the arms, shaped as cubes,
have access windows cut out, both as slots and as squares. This is done to
decrease the weight and also keeping assembly in mind. The access window is
necessary because the flexible coupler consists of three separate pieces. One
end of the coupler attaches to the gearbox shaft and the other to the shaft
extending out of the roller. The gearbox shaft is not long enough to drive the
roller itself. The access window allows one to attach the flexible coupler to the
extending shaft easily.
The upper ends of the arms that are not on the gearmount side only
contain the bearings for the roller to roll on. The arms are staggered from one
another to prevent interference during their motion. The arms are maintained
at the staggered distance by locating set-screws.
Holes may be drilled at the top of the hollow arms to allow electrical
cables into the tube. The cables run from the gearbox and motors placed for
roller actuation to the controller. Inserting the cables into the tubes would
house the cables and significantly reduce chances of operators/ patients tripping or getting caught in the cables.
45
Implementation of First Prototype
The rollers are made out of hollow aluminum tubing. This ensures
that they are lightweight yet their diameters are large enough to allow easy
wrapping of the sheet. The roller has a cap - gudgeon - at either end of it and
a steel shaft going through it. The cap is long enough to drive the roller and a
shaft extends out of it in order to go into the coupler. A rolled pin holds the
cap and shaft in place.
Yaskawa motors were selected to be placed at the base for maximum
torque. Gearbox LP-155 (Alpha Gear Manufacturers) with a ratio of 100:1 is
compatible with Yaskawa motors to increase torque. Also, gearbox LP-90,
with a ratio of 50:1 is compatible with Yaskawa motors. Actual gears were
chosen from the McMaster Carr catalogue to effectively double the torque
output at the gearbox shaft. Cost-effective and easily available gears were
chosen. The base spur gear ratio was 2.38:1. This, along with the gearbox,
results in a total gear reduction at the base of 238:1.
The gearmount - used to mount both LP-155 and the gears from
McMaster Carr - is made of aluminum, strong enough to bear the load of the
gearbox, yet light so as not to increase the overall weight of the machine. Bottom and edge joiners are welded to the box frame and then attached by bolts
to the gearmount to minimize twisting when the forces between the drive and
follower gears act to push them away from one anther. The bolt attachment
allows for minor misalignments that may occur due to imprecision either in
parts and machining, or during assembly.
The gearbox, LP- 155, has a spacer between it and the gearmount
which serves a dual purpose. First, it positions the output shaft of the gearbox
far enough away so as not to interfere with the motion of the arms without
46
(1) Mechanical Design and Electronics
causing a significant increase in the weight of the gearmount. Also, the spacer
has a shoulder to allow the gearbox to rest onto it and to locate the gearbox
with its location holes. The gearbox, LP-90, is mounted onto the arms on the
side of the gearmount, which is where the drive shaft is driven to minimize
flexure at the end of the shaft.
Adjustable feet are attached to the bottom of the box frame to account
for any misalignments that may be present on the floor. Also, the feet accommodate imprecision in either the legs or the attachment of the gearmount. This
reduces the cost since high precision is not needed during machining and
assembly.
Electronics
Software, ServoWorks40 and the MCQuad Package, was obtained
from Soft Servo Systems, Inc. It allows 4 axes actuator control and may be
extended to 16 axes. This would be beneficial for when the system is extended
to various functionalities, having a larger number of actuators for additional
control of the patient and also for patient monitoring.The following figure
more clearly illustrates the architecture of the ServoWorks software.
The FP-50 is a dual-link VersioBus Master Board inserted into a host
CPU and used to communicate with both the Servo and 1/0 Network. The
DC-120 is a servo interface module that controls 4 axes, and upto four DC120s may be daisychained to increase the control capacity to 16 axes. The IM200 us a general 1/0 module that provides 64 points of 1/0. It too may be daisychained with a total of 4 IM-200s.
47
Implementation of First Prototype
ISA Bus
FP-50
VersioBuTM Servo Network
FIGURE 18.
j
VersioBuTM
1/O Network
DC-1201
IM-200
Servo Drives
1/0 Devices
ServoWorks Architecture
Additional circuitry was designed to enable the control of the device
in case of emergencies such as loss of control by the ServoWorks controller or
by computer failures.
Limit switches were purchased to ensure that the arms did not crush
the gearmount if the controller lost control. The following circuit was
designed using a series of relays and switches as shown in Figure 19. It is currently being implemented.
48
(1) Mechanical Design and Electronics
FIGURE 19. Electrical Circuit designed for Limit Switches
An external circuit connected to the motor supply with relays was
used to control the power to the motors. If either of the limit switches is activated, then both motors at the base of the bed will be switched off. Therefore,
the relays are connected in series. The switching of both motors may, at first
glance, seem unnecessary since the arms are never supposed to cross. However, they are not allowed to cross if they are under the controller's control. If
this is lost, then the right roller may actually try and travel towards the left
limit switch. It may collide with the left roller, thereby causing a very unpleasant situation. So, if one limit switch is reached by either roller, the power to
both motors is switched off preventing further mishap. When a limit switch is
49
Implementation of First Prototype
pressed, then the circuit is broken, the relay loses contact and the power to the
motors is lost.
The limit switches used are normally closed (NC) switches to allow
detection of any breaks in the circuit. If normally open switches were used,
then, if there did exist another break in the circuit, one would not know about
it unless a limit switch was activated and it did not cut the power to the
motors. This, however, would defeat the purpose of the limit switch because
we would only know that it was not working after it did not work.
A "start", normally open (NO) switch is used to manually restart the
system once the limit switches are cleared. If the limit switches are not
cleared, then pressing the start switch will not turn the motors on because the
circuit would still be open by the limit switch.
An additional emergency stop "mushroom" switch could be included
in this circuit design but it currently is not because the handwheel device provided by Soft Servo Systems contains an emergency switch, that if needed,
stops all the motors from functioning.
(2) Verification of Concept
Implementation of the first prototype verified the proposed and
designed concept.
50
.
-------......
--....
........
...........
............
.. .........
(2) Verification of Concept
The photographs illustrate the device in action. Initially, the machine
is at its home position with the arms almost horizontal as shown.
FIGURE 20. Machine at Rest Position
51
.
.. .............
.........
... .....
Implementation of First Prototype
The patient lies on the bed as shown and the arms are actuated to
position the rollers above the patient.
FIGURE 21. Patient lying on back in Bed
52
.
............
.
... .........
............
.....................
(2) Verification of Concept
The patient is lying on her back and the sheets are attached on either
FIGURE 22. Arm Positioned
side to the rollers via VelcroTM. The right roller - as viewed from the foot of
the bed - is rotated clockwise. Figure 23 shows the person slowly beginning
to roll onto her side.
53
Implementation of First Prototype
FIGURE 23. Patient beginning to tilt.
Since the left bedsheet has so far remained stationary, the patient
tends to lean onto it. Therefore, as slack is introduced in the left bedsheet the
patient rolls completely - with the aid of the momentum gained - onto her
stomach.
54
(3) Verification of System Model
FIGURE 24. Patient Completes a Full Roll onto Stomach
The steps show that the device does indeed cause the patient to roll. If
patients are cooperative, then rolling is easily achieved. However, uncooperative patients, especially those with dementia, may require a modified
approach wherein the rollers are translated in the direction of rolling as discussed in Chapter 2, Figure 5.
(3) Verification of System Model
The assumptions made in the model presented in Chapter 3 were
rather crude and may have caused error in the system modeling phase. How-
55
Implementation of First Prototype
ever, as will be demonstrated, the errors are allowable and patient trajectory
may be obtained. Quasi-static equilibrium was assumed. This is the point
when the person just becomes airborne. This assumption will need to be
relaxed because the patient is never totally raised from the bed as patients tend
to feel unsafe when completely airborne. Therefore, there will be shear and
normal forces acting on the patient that are currently ignored.
The patient is modeled as a rigid body. This simplifies the analysis.
However, it does not take into account the elasticity of patients. In fact, during
rolling, one must realize that the patient appears to contract in width. This is
due to the cushioning effect of human tissue and fat that flows and is partially
compressible. The model presented does not take into account the elongation
of the bedsheet as it helps roll the patient.
Measurements were estimated for certain sheet lengths and distances.
Taking accurate measurements was relatively difficult. In addition to this,
determining the "width" of the person was an issue. The patient was measured
at the shoulders and the hips and an average of the two values were taken.
There are inevitable errors in measuring because the rollers themselves are
assumed to be points. The sheet does not really meet the roller at its center,
rather there is the distance that is actually the radius of the roller.
System Verification
Prior to the start of verifying the system model, the apparatus was calibrated to determine the torque constant, k,. The motor manufacturers did provide a value for this. However, it was for an AC motor but for my purposes,
the motor becomes DC. A Hall effect sensor was placed at the motor and
56
(3) Verification of System Model
readings were obtained with (a) no weight applied to the roller, and (b) with a
mass of 18.9kg attached to the roller. The following was obtained:
Mass
Input Current
(Peak Value)
No Mass
I, = 0.3 A
18.9kg
12 =
A
Using the following equation,
, = kI
(20)
where T is the torque, and I is the current provided to the motor, two
equations may be obtained. They are
Fir = kJ11
(21)
(F 1 + F)r = k,1 2
(22)
where F1 is the weight of the system with no additional weight, F is
the added weight to the overall weight of the system, and r is the radius of the
roller which is 1 in. Subtracting one from the other and rearranging the above
equations gives
k,
Fr
(23)
I2(-11
Therefore, k, is determined to be 6.72Nm/A. So, the tension, TI, may
be written as
T, = 2651
(24)
57
Implementation of First Prototype
The theoretical value of T, is expected to be 490N -- when T2 is equal
to zero -- which is the weight of the author, the test subject used. The experimental value obtained at the moment of rolling of the subject was measured to
be 424N.
TI, experimental
=
0.87Ti, theoretical
(25)
This shows that there is a close correspondence of the experimental
value obtained to the theoretical prediction!! Furthermore, a sharp drop in T
was observed at the moment of roll during experimentation. This gave clear
indication that the person had indeed shifted weight from the left to the right.
Verification of the model allows the trajectory generation for patients given
their weights and widths.
A larger data set will be obtained. Currently, the model verified the
concept, and demonstrated that the model obtained is satisfactory. It allows
the building of a generic control program to generate trajectory motion. However, this needs to be perfected and an extended model should be designed.
Furthermore, additional human subject testing will help in verification.
L, was expected to increase as slack is introduced. L 2 appears to
decrease because it is measured from the right end of the patient who, while
turning, moves closer to the right roller as it is lifted off the bed surface.
Rollers are placed as close to the width of the patient as possible. This
was not predicted prior to the implementation of the first prototype. However,
it was realized that placing the rollers with minimal horizontal distance
between them helps in caressing the patient between the sheets and in placing
58
(3) Verification of System Model
him wherever desired. It enables the patient to easily be placed on his side.
The sheets act to provide greater stability while being placed in this position.
As demonstrated by experimentation, the angle, a , reaches 900 when
the person is tilted towards a side. This is to be expected. Since the rollers are
placed as close as possible, then, as the patient is turned onto his side, the left
sheet meets the patient in an almost parallel direction.
Corresponding to the 900 value reached by angle, ux,
P drops signifi-
cantly in my calculations. This is because the angle is measured from the
patient's right end that is closer to the right roller thereby decreasing the
angle. However, this value was never really verified because obtaining a measurement for L 2 was not a simple task. There was too much slack in the sheet.
Also, it would be too subjective a decision to take as to where the sheet ended
- at the right end of the patient or underneath him.
The motor torques at the base may be monitored by the current input
into the system. The torques would provide further information with respect
to obtaining an extended model that accounts for shear and normal forces acting on the body.
Patient Translation
Patient translation was also verified. The arms were positioned at
their rest, "home" position and the rollers were simultaneously actuated. The
position of the person is located on the bed, as described by previous analysis
and the person is translated laterally as in a conveyor belt system.
59
Implementation of First Prototype
When translating the patient sideways, the roller moves in the direction of translation and the fact that it moves upwards - in an arc - increases
the tension in the sheet helping the person to be translated.
(4) Control Software Implementation
The position loop gain and the smoothing time was tuned. PD control
was used. The ServoWorks software and the handwheel device provided
allows both the manual yet separate control of the motors. If desired, trajectories may be generated by programming in G-code using the system model
developed and verified earlier. The NC files created for the programs allow the
simultaneous control of all four actuators.
The roller feedrates were adjusted to 1200mm/min. deemed to be a
suitable rate to roll patients. Preliminary experiments showed that slow and
steady motion was a priority in control so the positioning of the arms was also
set to a relatively low speed of 1200mm/min. G-code was written for a person
weighing 110 lbs and having an average width of 15 inches. The value for the
program command of the roller, PCrollerwas determined using the equation:
PCroller =
s
27tr
x dpm x
GRroier,
(26)
where s is the amount of sheet wrapped around the roller, r is the
radius of the roller, which is 1 in. for this prototype, dpm is the distance per
motor revolution which is set to 8192 jm, and GRroller is the gear ratio which
is 50 for the rollers. The program command for the arm, PCarm, is given as
60
(4) Control Software Implementation
PCarm =
x dpm x GRrm,
(27)
where a is the arc length that the arm travels, R is the radius of the circular trajectory, effectively the length of the arm which is 41.60 in. GRarm is
the gear ratio which, as previously mentioned, is 238.
Motor 1, X - right roller
Motor 2, Y - left roller
Motor 3, Z - left arm
Motor 4, A - right arm
(for right roller)
(for left roller)
The arms were initialized to home position. The length of arm travel
and roller wrap was determined and input into a program which was then executed.
The motions are defined as tabulated below.
TABLE 2. Motion Definition
Motion
Description
X positive
Clockwise rotation of right roller
Y positive
Clockwise rotation of left roller
Z positive
Clockwise rotation of motor, counterclockwise rotation of
left arm (supporting right roller), that is, arm moves up
A positive
Clockwise rotation of motor, counterclockwise rotation of
right arm (supporting left roller) that is, arm moves down
It was determined that the arms could be positioned at the given
parameters in the program, that is, close to the width of the patient and then
the rollers could be actuated. The preliminary code used to successfully roll
the author is shown as follows. Initially, since the bedsheet used is a full-sized
sheet, first the left end of the sheet is attached to the left roller which wraps up
some of the sheet in order to have enough sheet to later introduce slack to.
This is performed in the second line (N200) of the following code.
61
Implementation of First Prototype
Using the given software, the machine was taught a sequence of
events. In fact, this was done inversely by first specifying the points in the
code and then observing the behavior of the device. The desired coordinates
were stored for use in generic programming. The home position was set to
where the sheet would be attached to the roller. The left roller's home position
was after two complete wraps of the sheet onto it. This allows enough sheet to
introduce slack while rolling. The right roller's home position was set directly
above the patient. The desired coordinates were recorded for when the sheet
would be relatively tight at either end of the patient prior to commencement of
rolling. This teaching was for patients ranging in widths of 15-19 inches.
The code for rolling the patient includes positioning of the arms, and
initial uptake of slack in the sheet on either side of the patient.
The code for transferring the patient from one end of the bed to
another assumes that the maximum length of translation of the patient will be
half the width of the bed, from the center of the bed to either edge. Therefore,
the maximum length of travel was assumed to about 20 in. The first line of
code, however, is used to uptake any slack in the sheet prior to translation.
The feedrate was increased to 4500mm/min because during testing, it was
deemed that although slower is better for the skin on patients, an increase in
the speed decreases a feeling of dizziness experienced by the author. This
parameter may be left as a matter of preference and may be adjusted according to each person's wishes.
62
(4) Control Software Implementation
Control Code for Rolling and Transfer
Codefor Rolling
% Rolling (Generic)
N100 G91 G01 Z202000 A-204000 F1500
N200 G01 X456270 Y-293316 F2500
N300 G04 P3000
N400 G01 X325907 Y114067
N500 G04 P2000
N400 G01 X325907 Y130363
N500 G04 P2000
N400 G01 X325907 Y195544
N500 G04 P2000
N600 G01 Y651814
M02
Codefor Translation
%Translation
N100 G91 G01 X570337 Y-407384 F2500
N200 G04 P5000
N300 G01 X-1271037 Y-1271037 F4500
M02
These programs were used to automatically and successfully control
the implementation of the first prototype.
63
Implementation of First Prototype
64
CHAPTER 6
Conclusion and Future
Work
Discussion
The prototype built verified the proposed design concept. The device
will indeed help to transport bedridden patients easily and comfortably. The
benefits of using the bedsheet as the transport medium are manifold. First,
patients experience lower stress concentrations. During initial pivotal rolling,
the forces are distributed on a patient's body through the active sheet. This
may be followed by a short period of airborne rolling helping patients to complete the roll as they lose contact with the bed surface. Second, shear forces
exerted on patients, when moved from one edge of the bed to the other, are
almost eliminated as they become airborne on the sheet and are moved over.
Lifting does not introduce increased setup times as in currently available
products. Since patients will scarcely be lifted off the bed surface, their fears
are also reduced. Finally, the cradling aspect of the design ensures that the
patient is transported both safely and comfortably.
65
Conclusion and Future Work
Additional actuated prototypes will be built. The design may be
improved using the jointed arm configuration discussed below in order to test
the proposed extended functionalities. Further experiments and in-depth analysis will be conducted to ensure safe and easy repositioning of patients.
Future Work
Future Work regarding Design and Structure of Device
The stiffness of the arms may be increased by adding a rib-like structure to the hollow rectangular steel arm in its weakest direction. In fact, the
structure may be made such that the second arm does not collide with it and
slides within it.
Currently, strips of VelcroTM are used to attach the sheet to the bed.
This may be extended to using a band of VelcroTM on either side of the sheet
to ensure that it remains attached to the roller. A clamping mechanism or even
eyelets may be used to hold the bedsheets instead of VelcroM.
Patient transport has been accomplished in this paper. However, the
concept and first prototype may be extended to include certain functionalities
such as bed-to-chair transfer, and bed-to-bed transfer.
The device may be made such that is easily adapted to accommodate
different-sized beds and people. It may be portable. If so, then in order to minimize deflection, it may be mounted on solid rubber casters used to wheel it
around. The device may be adapted to a U-shaped frame to allow it to be
wheeled around the bed. It may have an attachable chain or flaps to anchor it
66
Future Work
to the ground on the open-ended side if necessary. If the device is indeed portable, then it also allows axial repositioning of patients by initially suspended
the patient on the bedsheet and then wheeling the device to the desired axial
location of the patient and then placing him back onto the bed.
Figure 25 illustrates that the device can be adjusted by additional piv-
FIGURE
25. Jointed Arms for Easy Transfer.
oting about joints A and B. This allows it to be adjusted to beds of different
widths. Roller motions are also independent. Since each side may be pivoted,
the patient may be lifted off the bed and positioned where desired, for example near the edge of the bed.
The device may also be used to transfer a patient from a bed to a
chair. A reclined chair may be positioned next to the bed. The pivoting marionette configuration would allow the distance between the two rollers to
encompass both the bed and the chair. This may be accomplished if, as shown
in Figure 25, the right roller pivots about joint B. Coordination of roller actuation would allow patient transfer from the bed to the chair. Once the patient is
transferred to a chair, the sheet needs to be tucked into the chair so that it may
be used to return the patient to the bed. Using the bedsheet to transfer the
67
Conclusion and Future Work
patient is advantageous because it helps move the patient over discontinuities
such as bed edges during transfer.
The jointed arms would also allow the arms to be positioned underneath the bed. They would be "tucked in" to their home position as shown.
)
A,
A2
0
FIGURE 26. Home Position of Jointed Arm Design
One of the drawbacks of the current prototype is the noticeable vibration of the arms as they are positioned. This is attributed to the discretization
of the motion due to the spur gears being used to drive the arms. The small
discretization is amplified as the arm is long. For future prototypes, helical
gears could be used to alleviate this problem.
Future Work with respect to Reinforcing Bedsheets
The sheets that are currently being used for demonstration are standard off-the-shelf sheets with VelcroTM strips sewn along the length of either
edge. For patients with larger weights than the inventor, the sheets may be
68
Future Work
reinforced laterally by having fibers sewn into them. Cotton strips may be
attached along the width of the sheet. Other materials, such as a type of plastic
filament may be sewn into the sheet. Cotton is preferable for now because, it
is not yet clear how, say, kevlar-reinforced sheets would affect the skin of the
patient. Furthermore, cotton strips would still allow the sheets to be easily
washed. There is also another material that is currently under investigation -clothing used by people in the navy that is known for its strength.
Bedsheets should constantly be inspected for holes. Appearance of
holes could cause the design to not yield the desired results. When the sheet is
under tension and a hole exists, then the sheet will tend to tear at the hole, the
weakest point. In order to avoid this, the easiest solution would be to use new
sheets at the onset of holes. Another option would be to use a patch or a clasp.
These would allow the sheets to be used for a longer period of time. However,
if there are holes in sheets, then it should anyway be a clear indication to caregivers that the time has arrived for new sheets!
Future Work regarding Electrical Circuitry
Currently, an emergency switch is available on the handwheel controlling the device so the power may be immediately cut off during testing.
Limit switches for the rollers will be provided for a second prototype. A beltpulley or gearing system may be used to control the limit switch. For every n
turns of the roller, the follower gear or pulley will turn once and hit a limit
switch at the end of its travel. The number of turns of the roller is chosen by
determining a relatively "safe" number of turns for the roller, small enough so
as not to totally wind the patient but also large enough so as not to interfere
69
Conclusion and Future Work
with the various motions required for rolling and repositioning the patient.
This switch will also be easily reset manually to allow flexibility in operator
control. Also, sensors may be mounted to determine the location of the gear or
pulley and hence monitor the absolute position of the patient given the amount
of sheet that is wrapped onto each roller.
This work may be extended to make one arm of the device longer
than the other. This would enable the arms to cross one another without colliding. It would also allow a patient to be transferred easily to another bed or
chair without significantly increasing the workspace required by the device
when it is in use.
70
References
[1] www.encyclopedia.com
[2] Nursing Strategies to Prevent Ventilator-Associated Pneumonia,
AACN Clinical Issues: Advance Practice in Acute and Critical
Care, Vol. 9, No 1. February 1998. Shelby Hixson, RN, MSN,
CCRN; Mary Lou Sole, RN, PhD, CCRN, FAAN; Tracey King, RN,
BSN, CCRN
[3]
http://seniors-site.com/copin2/sickdvin.html
[4] http://www.hill-rom.com
[5] http://www.kcil.com
[6] A. Garg, B.D. Owen, B. Carlson, An ergonomic study of nursing
assistants' job in a nursing home, Ergonomics, vol. 35, no. 9, 979995, 1992
[7] A. Garg, B. D. Owen, D. Beller, J. Banaag, A biomechanical and
ergonomic evaluation of patient transferring tasks: bed to wheelchair and wheelchair to bed, Ergon omics, vol. 34, no. 3, 289-312,
1991
71
References
[8] Garrett, Singiser, and Banks, 1992. Back injuries among nursing
personnel: the relationship of personal characteristics, risk factors,
and nursing practices. AAOHN Journal,40 (11), 510-516.
[9] Spano, J., Asada, H., 2000, Kinematic Analysis and Design of Surface Wave Distributed Actuators with Application to a Powered
Bed for Bedridden Patients, accepted for publication in IEEE
Transactions on Robotics and Automation, February 2000.
[10] S.H. Crandall, N.C. Dahl, T.J. Lardner, An Introduction to the
Mechanics of Solids, 2 nd Edition, McGraw-Hill Book Company.
[11] Shigley, and Mischke, Mechanical Engineering Design, 5 th Edition, McGraw-Hill, Inc., 1989
[12] Machine Design - Power and Motion Control Reference Volume
#12, Penton Publication, June 1989.
[13] F.E. Giesecke, A. Mitchell, H.C. Spencer, I.L. Hill, and J.T.
Dygon, Technical Drawing, 7 th Edition, Macmillan Publishing
Co., Inc., NY 1980.
[14] ServoWorksTM Documentation, Version 0.3, Soft Servo Systems,
72
APPENDIX A
Historical Record of
Approaches
Chronology of Designs
The conceptual phase of the design was crucial to arriving at the final
design. This stage consumed a large amount of time and creative energy! This
chapter presents the designs as they were arrived at chronologically. An
attempt is made here to depict the evolution of concepts, without disguising
the inevitable imperfections and wishful thinking that may have violated the
laws of physics during conceptualization.
Honeycomb Motion
Initially, flexible honeycomb structures were examined for twodimensional motion due to their ability to expand and contract in either direction. Honeycomb sheets for medical purposes are commercially available.
Two methods of motion were considered using the honeycomb sheet. They
were the inching and buckling motions. Figure 27 shows the inching motion.
73
Historical Record of Approaches
In this case, the length of the honeycomb structure would be contracted
towards the fixed end, hence moving the person from his position in Figure
27(a) to that of Figure 27(b). Assuming that the contracted segment is small
enough to leave the patient stationary at his newly acquired position, the contraction could be propagated underneath the person. Once the segment is
propagated, as seen in Figure 27(d), the free end may be extended and the
motion repeated.
FIGURE
27. Inching Motion
The second approach considered was buckling motion, illustrated in
Figure 28, which would utilize the buckling nature of the honeycomb. Similar
to the mechanism utilized in inching motion, the honeycomb would be contracted towards the fixed end, causing the patient to move. A buckle would
develop which would then be propagated under the person and extended on
the other (free) end allowing repetition of the cycle.
74
Chronology of Designs
FIGURE 28. Buckling Motion
A person lying in the center of the bed could easily be at the edge just
by contracting the honeycomb sheet. In this way, the person would not really
move to the edge, instead, the edge would move to the person.
Honeycomb sheets were found to be unsuitable for this application
due to their lack of load bearing ability. Controlling the propagation of a contracted segment precisely was also an issue. A solution would be to insert rods
into the honeycomb sheet and actuate the rods, either individually or as sets
using linear actuators. Rods would also increase the load bearing capability of
the sheets. However, they would cause discomfort to the patient. Tests were
conducted and it was observed that not only were the rods uncomfortable but
there existed too much friction between the rods and the honeycomb sheet
during expansion and contraction. Also, for precise control of the propagated
segment, many rods would need to be actuated resulting in too many actuators
being used. Therefore, both the inching and buckling motions were reevaluated.
75
Historical Record of Approaches
Belt and Belt
Use of modular conveyor belts was investigated.
FIGURE 29. Modular Belt Design
Figure 29 illustrates a single conveyor belt in the y-direction with
modules consisting of mini-conveyor belts in the x-direction. Each module is
made up of a flexible material attached rigidly at the ends that are shown as
bold lines in the figure above. The material for each module needs to be flexible to enable it to make the turns at the top and foot of the bed. In this manner,
motion in two dimensions would be obtained but problems would arise in
accessing the modules for actuation. Furthermore, the design did not satisfy
the requirement of stimulating the patient - it only provided patient translation.
Tilt and belt
An offshoot of "Belt and Belt" was developed into "Tilt and Belt",
which consisted of two main modules, tilt and y-belt. See Figure 30. Tilting
would occur both along the x-axis and the y-axis. Tilting the bed along the y-
76
Chronology of Designs
axis - plates 1 and 2 allow this in either direction - would aid in rolling the
person from side to side. Tilting along the x-axis using plate 3 would allow
the patient to be reconfigured into a sitting position, enabling him to get out of
bed easily.
FIGURE 30. Sketch of "Tilt-N-Belt" Bed.
Additionally, the design includes a belt drive system in the y-direction, referred to as y-belt. The purpose of y-belt is to repeatedly position the
patient longitudinally in the center of the bed'. Y-belt would be made up of a
flexible material to allow tilting of the bed by plates 1, 2, and 3 in either direction without tearing it. As in regular beds, bedsprings would bear load but
unlike regular beds, the springs may be activated to cause tilting when
desired.
The extra degree of freedom provided by y-belt allows positioning of
the patient onto the actuated portion so that his center of mass coincides with
the location of the actuated portion.
1.
Patients have a tendency to slip to the foot of the bed while lying down.
77
Historical Record of Approaches
This design causes motion in two dimensions by repositioning and
hence stimulating the patient. However, one of the key assumptions of the
design is that the y-belt will be extremely flexible which may not really be the
case. Controlling the flexible belt would be imprecise due to the formation of
a hyperbolic paraboloid.
Tilt and tilt
A two-way folding bed was designed. Independently-actuated plates,
with n number of rows by m number of columns, would be used to cause both
lateral and longitudinal tilting of the bed. The plates would be able to slide
past each other for axial repositioning of patients. Rollers could be used to
assist in changing bedsheets by moving the patient from one end of the bed to
the other. Tilting would aid the patient to roll off the used bedsheet and onto a
new one. The photographs shown below illustrate how the concept would
work.
Initially, the person would be lying on top of the sheet to be changed.
See Figure 31. The sheet would then be pulled with the person lying on top of
it towards the edge. See Figure 32.
78
Chronology of Designs
FIGURE 31. Patient on sheet to be removed.
FIGURE 32. Sheet and Patient translated towards edge.
The mattress is exposed and the person is only on a fraction of the
sheet to be changed. The bed is now tilted to release the remaining portion of
the sheet.
79
Historical Record of Approaches
FIGURE 33. Bed tilted to release bedsheet
Therefore, the person is off the old bedsheet and if a new one is
attached to the ends of the old one, then he will be lying on the new sheet. In
the figure, he is shown finally to be lying just on top of the mattress.
FIGURE 34. Bedsheet removed with Patient on new sheet
The bed could be reconfigured into a sitting position as shown below.
80
Chronology of Designs
FIGURE 35. Bed reconfigured into Chair.
Changing Bedsheets
Bedsheets could be changed using rollers attached along the length of
the bed. Prior to use, the patient would be rolled over to one edge of the bed to
free up a portion of the sheet. This would be accomplished by tilting the bed.
The sheet would then be inserted between the rollers that would be pressed
together. The bottom roller would be actuated causing the bedsheet to be
rolled through it. This would result in pulling the patient lying on top of the
sheet to the edge.
Once the patient reached sufficiently close to the edge of the bed, the
bed would be tilted laterally allowing the patient to roll towards the opposite
edge, freeing the remaining portion of the sheet. In this way, changing bedsheets would be easier. Nursing personnel would no longer need to drag the
patient towards any side because the tilting would aid in rolling the patient
and the rollers in any pulling that may be necessary.
81
Historical Record of Approaches
Axial Positioning
Axial motion would be obtained by positioning the patient say, on the
right side of the bed, as shown in Figure 36. The right side then slides forward
and tilts. The patient will be on the left side and the right plate may slide back
to its original position. In this manner, the patient could be repositioned axially on the bed..
FIGURE 36. Sliding Motion for Axial Repositioning of Patient
The preliminary design is shown below.
FIGURE 37. Lateral Tilt of Two-Way Folding Bed.
The figure shows one side of the bed frame tilted laterally. The design
shows that the plates can be tilted either laterally or longitudinally. The plates
82
Chronology of Designs
may be tilted about the y-axis as shown above because they are attached to
rollers. This also gives the advantage of allowing the plates to slide past one
another along the y-direction. The sliding rollers, can translate the plates thus
enabling axial positioning of the patient as described previously in Figure 36.
The figure shows the plates moved as a unit. The patient would have been
positioned onto the moving plates, translated axially "higher" on the bed, and,
as Figure 38 shows, the plates would be tilted and the patient would be positioned in the desired axial location back onto the stationary plates. The moving plates would then slide back to their original position.
FIGURE
38. Sliding and Tilting Motions for Axial Repositioning
83
Historical Record of Approaches
84
APPENDIX B
Assembly Drawings
The following views of the overall design are given: isometric, top
and side.
85
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PRONIBITED.
3
x
66
8
10
4
5
7
1 Frame
2 Gearmount
3 Roller
4 Arm, Type 1
5 Arm, Type 2
6 Gearbox, LP-90
7 Gearbox, LP- 155
8 McMC Spur Gears
9 Bottom Joiner
10 Edge Joiner
1 Leg, Type 1
12 Leg, Type 2
2
U)
9
(U
I...
0
11a
.0
E
C)
U)
U)
<I
D2IMENSS HE IE INCECS
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLES
XX+.0I
+I
PA RTS LIST
USE.0CUT
APPROVAL$
MATERIAL
DO NOT
A
SCALIE:1:1
WEIGH:
CHECKED
'
USED ON
APPLICATION
ISOMETRIC VIEW
OF MACHINE
0
AssemblWG
DATE
'Ann Basmaoioct.2
o
FINISH
NEXT ASSY
MIT d'Arbeloff Laboratory
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
SCALE DRAWING
NG APPR.
1AFG APPR.
I HEET3OF 3
00
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
LII
--------------------------------
-------------------
:1
------------------ ---------------------------------- ----I
I
IQ
--
I
- -
-
-
- -
-
-
------------------
T
'
II
-- -
--
-
---
-
---------------------------- ------------------------------------------------I.-L4
UNLEUS OTHERISE SPECFIED
PARTS LIST
DIMENSONT ARE IN INCHUS
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLE
XX+.0
+1
CTERIAL
CHECKED
fl ISH
NEXT ASSY
USED ON
APPLICATION
NGAP
DO NOT
SCALE
DRAWING
MIT d'Arbeloff Laboratory
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
APPROVALS
D
00
---
DATE
smojio-E
snn
TOP VIEW OF
MACHINE
DWG NO
AssemblyQPR.
SCALE::::
jvfEIGHT:
V
SCAL~lyI
REV.
OISET
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
...
m
.0
0l
I-
E
UNLESS OTHERWISE SPECIIED
DIMENONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLE
+
XX+V50
xxxtoo1
+
+
FINISH
NEXT ASSY
DO NOT
SIDE VIEW
OF MACHINE
01
CHECKED
O
GAPPR.
USED ON
APPUCATION
MIT d'Arbeloff Laboratory
DATE
APPROVALS
DRA "BSOIf~t
MATERLAL
I
PARTS LIST
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
SCALE
DRAWING
MFG APPR.
.
DWG
SCALE: 1::I1
ISCALE:
NO
:
I
1WEIGHT:
REV
1OF
SHEET
I SHffT I OF 33
00
00
APPENDIX C
Part Drawings and
Listing
PartInfo
In this section, wherever possible, the parts have their part numbers listed and/or they are presented in separate drawings for machining.
89
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
1
2
3
5
6
7
8
4
0)
-i
0)
1
2
3
4
5
6
7
8
Motor
Gearbox, LP-90
Arm, Type 1
Flexible Coupler
Steel Shaft
Square Bearing
Gudgeon
Roller
UNLESS OTHERWISE SPECFIED
DIMENSIONS ARE IN
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLE
+
XX+.01
+1
INCHES
PARTS LIST
APPRVa
MATERXAL
APPLICATION
DO NOT SCALE DRAWING
ROLLER
ATTACHMENT
DATE
'rnO Bcisma iar ,,.2w
NG APPR.
NEXT ASSY USED ON
MIT d'ArbeloffLaboratory
CAI) GENERATED DRAWING.
DO NOT MANUALLY UPDATE
1FG APPR.
SIZE
A
DWG. NO.
.E15
TopOOO
EGH:SETIO
REV.
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
0
0
ol
0
0*9
-U
El-
.9
-N
0
%~0
c0
2.00
4.00
C%4
00J
%0
00
UNLESS OTHERWISE SPECFIED
DIMENSIONS ARE ININCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLES
XX+.0
+1
XXX+.005
PARTS LIST
APPEOVALS
-Arin
-ATERIAL
USED ON
APPLICATION
DATi __
GEARBOX;
LP-90
rrr
:NO
DO NOT SCALE DRAWING
MIT d'Arbeloff Laboratory
Basmajiar c.200
RNISH
NEXT ASSY
I
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
APPR.
MIGAPPR.
SIZ
A
WG.
scALE::2
NO.
ToO0l
WEIGHT
1
SEET
OF
I
IS
THE INFORMATION CONTAINED IN THIS DRAWING
THE SOLE PROPERTY Of
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBTED.
HOLLOW ALUMINUM TUBE
OD = 2.25"
ID = 1.75'
LENGTH = 72'
0.
PARTS LIST
UNLESS OTHERWISE SPECIED
DIMENSIONS ARTE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLES
+
+XX+.Ol
XXX+.005
1
APPROVALS
MATERtAt
FINISHI
NEXT ASSY
USED ON
APPLICATION
ROLLER
DATE
Arn BosmojiaOct
2001
CHECKED
N
DO NOT SCALE DRAWiNG
MIT d'Arbeloff Laboratory
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
p
SIE
A
-OCALE1
.
NO.
REV.
Top002
WEIGHT:
SHEET
IofI
Cl
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
00.75
-u
0
02.00
00
UNLESS OTHERWISE SPECFIED
PARTS LIST
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLE
X+.01
(y)
(X+
6
+
005
MATERIAL
FINISH
NEXT ASSY
USED ON
APPLICATION
APPROVALS
DRAWNa
Anin
MIT d'Arbeloff Laboratory
TOP BEARING
DATE
BasmaiiaOpe.2o0w
CHECKED
HG
DO NOT SCALE DRAWING
I
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
APPR.
MFG APPR.
ToG.
L:: | WIHI :
I,CSCALE:1:1
AE
SALE±
IWEG
W EIG
pTITISTlO
.OO3
REV
i
SHEET
SEIIoIOF
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY Of
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
00.75
0 1.75
0)
*0-
U)
-g
U>
I
I
I
I
S
I
I
I
I
I
I
I
I
I
I
I
I
S
C)
LC)
CN
c'.J
I
UNLESS OTHERWISE SPECFIED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
DECIMALS ANGLE!
.t
TUSTI
*
FRACTIONS
PARTS LIST I
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
TUU.1X5
I
I
___________ £ ____
MATER AL
FINISH
NEXT ASSY
ALUMINUM
ENG A PPR.
DO NOT
GUDGEON
GT
. 2ooi
CHECKED
O
USED ON
APPLICATION
APPROVALS
Arin Bosmaio
MIT d'Arbeloff Laboratory
SCALE
DRAWING
MFG APPR.
S G
SCALE:: I
NO
WErIGHT:
004
REV.
SHEET I OF I
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
0o-
S-
-U
0)
5
-I
I
I
-I'
0
0
C-)
UNLESS OTHERWISE SPECFIED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLE
XX+.01
+I
XXX+.005
MATERIAL
NEXT ASSY
CA
I
APPROVA
STEEL
USED ON
APPLICATION
PARTS LIST
DO NOT SCALE DRAWING
---
MIT d'Arbeloff Laboratory
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
DATE
ROLLER
SHAFT
pri~n Basmaiiacu.
CHECKED
NG A PPR.
MFGAPPRI
SIZE NO.
A
-CLE:1:2 IWDGT
REV.
TWG.
Top005
[SET
I
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
0 0.38
4 HOLES c
2.50
00
0
(D
ff-
0 1.00f
00.75
cy)
-0
C.4
0)
0.44
1.69
-J
C
052.38
3.38
-
UNLESS OTHERWISE SPECFIED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLE'
+
XX+.001
+
PARTS LIST
APPROVALS
MATERIAL
FINISH
NEXT ASSY USED ON
APPLICATION
I
DO NOT SCALE DRAWING
MIT dArbeloff Laboratory
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
AwiNn
Bsmoai
SQUARE
DATE
ct
2oo
CHECKED
NG APPR.
MFGAPPR.
I
BEARING
SL6.
A.
ToP006
SCALE;I:
IWEIGHT:
IREV.
SHEET I Of I
ON0
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
2.50
47.00
01.75
-4
0.43
00.38
4 HOLES
-U
F
110 -1
T
h.
I
i
d.R,
33W.
~*1
0.75
0
0
00.38
4 HOLES
2.00
0 3.0 _
0
0
10
co)
41.60
0
0
4.00
STD. KEY WAY
SLOT: 3/8 x 3/16
3.50
UNLESS OTHERWISE SPECFIED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
DECIMALS ANGLES
FRACTIONS
xx+.I
XXX+.005
MATERIAL
+1
STEEL
PARTS LIST
I
MIT d'Arbeloff Laboratory
CAD GENERATED DRAWING.
DC NOT MANUALLY UPDATE
APPROVALS
ARM,
TYPE 1
DATE
RAAiN
CHECKED
FINISH
NEXT ASSY
NG APPR.
USED ON
APPLICATION
DO NOT
SCALE DRAWING
14FG A PPR.
SUE OWG. NO.
A
REV
ArmO0l
SCALE:A20 IWEIGHT:
IS
tO I OF I
THE INFORMATION CONTAINED IN THIS DRAWING 1S THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
01.75
03.00
4
04.00
co
co)
02.00,
0)
C
(I)
-J
0
01.00
2.50
SEE DETAIL
*0
C
--r7-77
'A
0)
C
I-
0
1.00
I-
0~
UNLESS OTHERWISE SPECIRED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLE!
+XX*OI
0^+O
+1
-1
I
PARTS LIST
APPROVALS
MAERIAL
NEXT ASSY
STEEL
DATE
ARM,
TYPE 2
An Basmairio4c. moi
10E
USED ON
APPUCATION
MIT d'Arbeloff Laboratory
CAD GENERATED DRAwiNG.
DO NOT MANUALLY UPDATE
DO NOT SCALE DRAWING
I
DFN0
A
SCALEI-20
I
RE'
Arm002
WE GHT:
SHEET
I OF I
00
C\
IS
THE SOLE PROPERTY OF
THE INFORMATION CONTAINED IN THIS DRAWING
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
0
9
7
- ,
1 Motor
2
3
4
5
6
7
8
9
Gearbox, LP-155
Spacer
Bottom Bearing
Gearmount
Follower Gear
Drive Gear
Arm, Type 1
Arm, Type I
8
2
56
3 4
UNLESS OTHERWISE SPECFIED
DIMENSIONS ARE IN INCHES
TOLERANCEU ARE:
FRACTIONS DECIMALS ANGLES
+I
xx+I
+
SDIUA.005
PARTS LIST
APPROVALS
MATERIAL
m
Bosmaar
FINISH
NEXT ASSY
USED ON
APPUCATION
NG APPR.
00 NOT
DO
SCALE DRAWING
NOT SCALE DRAWINO
MIT d'Arbeloff Laboratory
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
MFGAPPR.
BASE,
GEARMOUNT END
DATE
1ot.2001
A
SCALE:1:A
SCAIE:l:I
-NO-BaseO O
|IWEIGHT:
D
.
WEIOAT:
ISHEET I OF
ISAEEEIDFI
I
I
I
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
0
0
00.75
It)
C)
C)
0
0
u-)
EC D
"DE O"SE
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLES
XX+.0i
+l
X(X+.005
TS LIST
PARRSLS
P
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
APPROVALS
DATE
MATERIAL
MIT d'Arbeloff Laboratory
S
GMOTOR
CHECKED
NEXTASSY
USED ON
APPUCATION
1NGAPPR.
DO
SCALE
NOT
DRAWING
DO NOT SCALE DRAWINO
A
SCALE:A:2
BseCOl
IWEIGHT:
ISHEET IOfI
EARLS TOP I
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
-16
0
Lo
-u
-I
0
LE)
(y)
3.13
c')
6.25
'04
L0
II
C))
LoI
Li
LI
')
Li
cy)l
14)i
'
00
UNLESS OTHERWISE SPECFIED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLES
+.I
+XX+.l
PARTS LIST
XX(X+.005
MATERIAL
MIT d'Arbeloff Laboratory
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
DRAW.P
APPROVA
I DATE
,-Ann Basmaiiarc. 2oid
CHECKED
NEXT ASSY
USED ON
APPLICATION
tNGAPPR.
DO NOT SCALE DRAWING
MFG APPR,
I
I
SIZE
WG
A
S
SC:1:
E
AL A:
GEARBOX,
LP- 155
B s
BaseOO2
REV
1.
WEiGH
W
HT:
I SHE
1
I OF I
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
00.50
4 HOLES
04.75
01.80
cl
Ito
TV)
0)
.4-'
U)
*0
C
Ru
U)
0)
C
------
0
4
2.76
2.76
-
C)I
L.
0
Ru
I
T
0~
I
I
UNLESS OTHERVUSE SPECIFIED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLES
-
+
X(X+.00
+5
PARTS LIST
APPROVALS
MATERIAL
MIT d'Arbeloff Laboratory
CAD GENERATED DRAWING.
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
DRNAN
Ain
T BOSMOI2
SPACER,
FOR LP-155
DATE
Oct.j
V 2001
CHECKES
NEXT ASSY
USED ON
APPUCATION
DO NOT
SCALE DRAWING
DNG APPR.
MFG^AP-
SIZ
A
Sw
SCALEA:I2
.
NO
.
REV.
BaseOO3
UWEIGHT:
IEET IOF
I
I
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
0
O
0
02.00
[
_
0_
_
_
1.
1 ________
UNLESS OTHERWISE SPECFIED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLE
XX+.
xxX+.UOS
I
I
NET ASSY
USED
APPICATION
+1
MATERIAL
ON
I
PARTS LIST
APPROVALS
rin
NISH
D
_NG APPR,
DO NOT
SCALE
DRAWING
MIT d'Arbeloff Laboratory
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
BOTTOM
BEARING
DATE
BosmoaIamc.2001
A
A
G
BE se
Base0O4
, | GH :HE
sCALE:1:1
T
REV.
Ii
IO
)OF
ISHEET
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
00.50
8 HOLES
2.76
0.25
23.00
01.80
2 HOLES 5.51
II(D
/0
uLO
-o
0.25
D
0
0
-r
0
0
0 2. 00
2 HOLES
1.50
1.50
3.00
13.50
CG
00.13
8 HOLES
.
0
cc
Cn
cc
Gearmount symmetric
00
C)LC
o
31.00
UNLESS OTHERWISE SPEC FED
DIMENSIONS ARE IT INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLE!
+XX+.01
+1
PARTS LIST
XXXV GJS
MATERIAL
ALUMINUM
FINLSH
NEXT ASSY
USED ON
APPUCATION
APPROVALS
DAwd7n Bosmojiar
GEARMOUNT
DATE
It.2oi
CHECKED
NG APPR.
DO NOT SCALE DRAWING
MIT d'Arbeloff Laboratory
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
A
DW. NO.
SCALE:l:10
Base005
IEREV.
WEIGHT:
0
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
ALL 4 PLATES WELDED
PARTS LIST
OTHERIE
SENECED
DNS
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLES
XX+.T
+I
xx__
+o005_
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
APPROVALS
DATE
BasmalianOct.2001
MATERIAL
ALUMINUM
NEXT ASSY USED
APPUCATION
ON
SCALE DRAWING
GEARMOUNT,
ISOMETRIC VIEW
CHECKED
;NG APPR.
DO NOT
MIT d'Arbeloff Laboratory
MFG
APPR.
SEE DWG. NOSCALE:l:I
BaseQ05'
WEIGHT:
REV
SHEET IOF
I
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
PROHIBITED.
-IS
01.00
0
CN
LO
02.50
00
*0
i
14 1/2 Deg Spur Gear 10 Pitch, 60 Teeth
McMaster Carr, Part 6867K84
Y
Gear broached for std.
a.
ANSI key, 1/4 x 1/8
TI
UNLESS OTHERWISE SPECHIED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLES
10
+
XX+.00
PARTS LIST
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
+
APPROVALS
xxxERIA
MATERIAL
MIT d'Arbeloff Laboratory
Follower Gear
DATE
DRAWN
Arnn Bosmojair
ct. 2001
CHECKED
FINISH
NEXT ASSY
USED ON
APPLCATION
DO NOT
zNG APPR.
APPR.
SCALE DRAWING MFG
A
G.No.
SCALE::2
BaseOO6
WEIGHT:
REV
SHEET I
Of
I
10
IS
THE INFORMATION CONTAINED IN THIS DRAWING THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBIT ED.
01.57
0
10
02.13
|
|
aI
ai
141/2 Deg Spur Gear 10 Pitch, 24 Teeth
McMaster Carr, Part 6867K76
Gear broached for Metric
Shaft, 0.4724 x 0.2362
'0
UNLESS OTHRWISE SPECIFIED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLES
XX+.O
+
+1
XX+.005
MATERAL
MIT d'Arbeloff Laboratory
PARTS LIST
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
APPROVALS
DATE
Drive Gear
)P~rn Basmaiiar oct.2ooi.
CHECKED
NEXT ASSY
USED ON
APPUCATION
:NG APPR.
DO NOT
SCALE DRAWING
Base007
A
MFG APR.SCALE1
I
WEIGHT:
SHEET I OF I
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
3
4
2
*a
0.
I
2
3
4
Solid Drive Shaft, Type 1
Hollow Drive Tube
Bushing
Solid Drive Shaft, Type 2
CC
PARTS UIST
DIMESION"SE EN NCECSD
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLES
+
XX+.01
+I
DRIVE SHAFT
DATE
XX+.0
Dn
MATERIAL
MIT d'Arbeloff Laboratory
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
APPROVAiS
A
smw arm
A TTA CHMENT
ANISHCHECKED
NEXT ASSY
NG APPR.
USED ON
APPUICATION
DO NOT SCALE DRAWING
"'
A77
NO'REV.
Basei'
C1:1
wEIGHiT:
SET1O
00
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
01.75
-----------------------0
01.00
0
0
STD. ANSI KEYWAY SLOTS
FOR DIA. 1.75", 3/8 x 3/16
FOR DIA. 1", 1/4 X 1/8
0)
0
Coi
UNLESS OTHERWISE SPECFIED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLE
S
XX+.01
+1
PARTS LIST
Xxx+A.15
MATER
L
MIT d'Arbeloff Laboratory
CAD GENERATED DRAWING.
DONOT MANUALLY UPDATE
APPROVALS
SOLID DRIVE SHAFT,
TYPE 1
DATE
DArin Basmajiaroct.221
STEEL
CHECXED
FINISH
NEXT ASSY
USED ON
APPLICATION
Q
NG APPR.
DO
NOT
SCALE
DRAWING
AFG APPR
BaG.
A
N9
SCALEAl:S
IWEIGHT:
REV-
ISHEET
IOf I
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
STAINLESS STEEL TUBE
OD = 2.38"
ID= 1.75"
LENGTH = 70.50"
0)
C
Rn
C
Ul)
0)
C
UNLESS OTNHERISE SPECIIED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIOM6 DECIMALS ANGLE!
+
XX+.005
MATERIAL
APPROVAL$
ENG APR.
DO NOT
HOLLOW DRIVE
TUBE
DATE
CHECKED
USED ON
APPUCATION
MIT d'Arbeloff Laboratory
DAnn BasmaiiarOct. 2w)1
FINISH
NEXT ASSY
PARTS LIST
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
SCALE DRAWING
MFG APPR.
I
"
DWvG NOI
TCAIE.I:ST
SCALEAl2
~
BaseO10
lH:
I WEIGHT:
1Of
SHEET
ISATET I OFT
1
I S14EET I Of I
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
TU
BRONZE BUSHING
OD = 2.75"
ID = 2.38"
LENGTH = 0.75"
UNLESS OTHERWISE SPECFIED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
DECIMALS ANGLES
FRACTIONS
XX+.I
+
+1
PARTS LIST
DO NOT MANUALLY UPDATE
XXX+.005
MATERIAL
APPROVALS
USED
APEPUC AT ON
ON
DO NOT
I
(ATE
BUSHING
An Bosmoja rct. 2W1
CHECXED
NETASSY
MIT d'Arbeloff Laboratory
CAD GENERATED DRAWING.
SCALE
DRAWING
NG APPR.
.FG ^PP-.
I
sAzE
DWG NOABaseOM1
WEIGHT:
scALE:l:1
S RSCALE:11
i WErGAI
REV.
1OF
SHEET
SHEET I OFAI
IS
THE SOLE PROPERTY OF
THE INFORMATION CONTAINED IN THIS DRAWING
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
--1
--------------------
01.75
V
C)
0
STD. ANSI KEY WAY
SLOT: 3/8 x 3/16
C
(A
C
1~
C
4.'
I-
0~
UNLESS OTH4ERWISE SPECIFIED
DMENSIONS AR EIN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLES
+
XX+.0US
PARTS LIST
APPROVALS
MATERIAL
STEEL
MIT d'Arbeloff Laboratory
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
ArIn Basmaiar
SOLID DIVE SHAFT,
TYPE 2
DATE
ct.2ooi
CHECKED
FIN$HA
NEXT
ASSY
USED ON
APPLICATION
DO NOT
SCALE
MG APPR.
DRAWING
A
W
Base012
SCALE:1.5
|WEIGHT:
-FsHEE 1-01 1
IN
THIS DRAWING IS THE SOLE PROPERTY OF
THE INFORMATION CONTAINED
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
0
3x3 HOLLOW STEEL TUBE
LENGTH = 73"
THK. = 1/8"
UNLESS OTHERNISE SPECID
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLES
+
XX+.S
+1
PARTS LIST
XXX+.005
MATERIAL
|
IFINISH
NEXT ASSYj USED ON
APPLICATION
DO NOT SCALE DRAWING
|
MIT d'Arbeloff Laboratory
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
APPROVALS
BOX BASE
DATE
Arin Basmaiiao ct.
|CHECKED
NG APPR
SBaseO
SCALEAl:20
wEiGw:
13
REAV
1i
SHEET
OF
ISH ET I Of I
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY Of
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBIT ED.
3x3 STEEL TUBE
LENGTH = 20"
THK. = 1/8"
U)
C0
C)
0>
cc
0.
UNLESS OTHERWISE SPECIFED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLES
+1
+XX+.Ol
xxx*-J 0
T
MATERIAL
APPROVALS
JOINING
DATE
BAR
CHECKED
NEXT ASSY USED ON
:NG APPR.
DO NOT
MIT d'Arbeloff Laboratory
Dr Basmoio cI. 2ELT
FINISH
AuPPUCATION
PARTS LIST
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
SCALE
DRAWING
MFG^APP-
A
SCA
No
I
BaseO 14
1E
WEIGHT:
ISHEET
O
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
0-
0
-0
00.13
4 HOLES
0
0-0
0
1.00C5.
1.00
U-)
04J
0.25
UNLESS DTHERWISE SPECEIRD
SPCIFID
UDNLS
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLES
OTHEREIE
+1XX+.01
+
PARTS LIST
APPROVALS
3.00
MATERIAL
STEEL
MIT d'Arbeloff Laboratory
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
KAriHn Basmoio
CHECKED
EDGE
JOINER
OATE
Ct 2t. 1
FINISH
NEXT ASSY
USED ON
APPLICATION
00
NOT
SCALE DRAWING
:NG APPR.
MFG APPK.
A
SCALE:I:i
BaseO 15
!WEIGHT:
ISHEET IOF I
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OP
MIT, ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
2.50
0.25
2.00
C14
0
C5
C
IF
- ----
0.50
*-0.50 -
--
-
-
-
-
-
-
--_-_ -
-
.-
C
-a
U,
V
C
0
0-
U)
)
0)
C
I..
0
-a
00.13
o
0;0
0
4 HOLES
0-
1..
a.
UNLESS OTHERWISE SPECIFED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE.
FRACTIONS DECIMALS ANGLO
XX+.Ol
+1
PARTS LIST
STEEL
STE
DRAWING.
BOVVOM
DATE
BOTTOM
CAD GENERATED
DO NOT MANUALLY UPDATE
XXX+ .005
MATERIAL
MIT d'Arbeloff Laboratory
CAD GENERATED DRAWING.
APPROVALS
JOINER
iAn BosmaiarOct. 2ml
CHECKED
ANISHN
NEXT ASSY
USED ON
APPUCATION
00 NOT
SCALE
DRAWING
NG APPR.
MFGAPPR.
A lwG
.
NO
BaseOl6
1EV
ISCALE:1;i
SCALE 1.1
SCALE: 1.1
I WEIGHT
WEIGHT
1 WEIGHT:
ISHEET I OH I
iNEET
1OF I
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
2.50
00.13
1/4
0.75
0.75
-u
a)
0
4 HOLES
-I
Lo~
00.50
-Ri
0
00.74
I
II\~
Lo
Nll
6
.o
i
t
o
0
Lo
c;
1.0
02.80
O
i,
|
UNLESS OTHERWISE SPECFIED
DIMENSIONS AE IN INCHEU
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLE
+I
XX+.I
PARTS LIST
XXX+.005
MATERIAL
NEXT ASSY
STEEL
USED ON
APPLICATION
DO NOT
SCALE DRAWING
DO NOT SCALE DRAWING
MIT d'Arbeloff Laboratory
CAD GENERATED DRAWING.
DO NOT MANUALLY UPDATE
APPROVALS
LEG,
TYPE 1
DATE
'Arn Basmajiaroct.200
CHECKED
ING APPR.
MFG AR.-
I
ISIZE
IDWG
WGNO.
A
SCALE;1:2
SCUIE~L2
0'BaseO 17
IW&GHt:
I WEV3HT:
REV.
SHEET I Of I
SAVE: 01
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
0.75
C)
0
-
0o
------~ ~------ --
------ ---- --L---- -----0
LO
0.50
1-
___
r-T
C
0)
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
S
I
I
I
I
I
I
I
I
S
I
I
I
I
1.00
C))
0
c'.J
'0
UNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLES
+1
XX+.O
XXX+.005
MAERIAt
STEEL
PARTS LIST
MIT d'Arbeloff Laboratory
CAT GENERATED DRAWING.
DO NOT MANUALLY UPDATE
APPROVALS
Ann Bosmaior
LEG,
TYPE 2
DATE
12oo1
CHECKED
FINISH
NEXT ASSY
USED ON
APPUCATION
DO NOT
SCALE DRAWING
SIZE
D
A
E
SCALE.! 2
G
NO -
BaseO
IWEIGHT:
REV .
E8
SHEET I Of I
00
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
MIT. ANY REPRODUCTION IN PART OR WHOLE WITHOUT WRITTEN PERMISSION
IS PROHIBITED.
00.75
----------------------------------I
0
00.50.
'I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
0
0
Inside threads 1/2-13
for adjustable feet
I
UNLESS OTHERWISE SPECF1ED
DMENSIONS ARE IN INCHES
TOLERANCES AREFRACTIONS DECIMALS ANGLE'
+XX+.01
+1
xxx+.005
MAIERINL
MATERIAL
STEEL
STEEL
APPROVALS
-
D'Arv1
SFINISH
NEXT ASSY
USED
NG APPR.
ON
APPUCATION
DO NOT
i
"C
SCALE DRAWING
MIT d'Arbeloff Laboratory
PARTS LIST
CAD GENERATED DRAWINGDO NOT MANUALLY UPDATE
MF PR
SATE
Basmojiir oci
LEG INSERT
2oo I
A
NO Baseol
SCALE:1:2 IWEIGHT:
9v
ISHEEI IOF I
Part Drawings and Listing
The table below gives the listing of the parts that were puchased.
TABLE 3. Off-the shelf Parts List
Name of
Part
Part
Number
Manufacturer/Supplier (if any)
Qty.
Gearbox,
R1:50
LP-90
Alpha Gear/Northeast Motion
2
Gearbox,
R1:100
LP-155
Alpha Gear/Northeast Motion
2
Roller
Motors
SGMAH-02
Yaskawa/Soft Servo Systems
2
Base Motors
SGMPH-15
Yaskawa/Soft Servo Systems
2
Bearings
60355K19
McMC
4
Adjustable
Feet
62805K42
McMC
4
Drive Gear
6867K76
McMC
2
Follower
Gear
6867K84
McMC
2
Flexible
Couplers
59985K25
McMC
2
Aluminum
Rollers
9056K166
McMC
2
Square
Bearings
6664K13
McMC
2
McMC = McMaster Ca:
Some of the parts labeled in the figure were machined out of stock that was purchased from
the MIT Central Machine Shop. A description is provided as to how the parts were made, what they
consisted of. A listing is given of different parts that are labeled as forming part of the machine.
Detailed drawings of each part are available. The listing helps keep track of the type and number of
parts. Note that though in the prototye that is built, both roller and basemotors used were the same
120
Part Info
due to the short lead times of Yaskawa motor, part number SGMAH-02, the base motors,
Yaskawa motor, part number SGMPH-15 arrived and is in the process of being tested as this thesis is being docmented.
TABLE 4. Machined Parts List
Part Name
Description
Quantity
Box Base
3x3 steel tube @ 7 ft.
2
Joining Bar
3x3 steel tube
2
Arms, Type 1
2x1 hollow steel tubes
2
Arms, Type 2
2x1 hollow steel tubes
2
Gearmount
4 sided aluminum box extrusion
1
Edge Joiner
steel attachment
2
Bottom Joiner
steel attachment
2
Bushing
bronze bushing for drive tube
2
Drive Tube
hollow stainless steel tube
2
Back Drive
Shaft
solid steel shaft
2
Front Drive
Shaft
solid steel shaft
2
Legs
hollow square steel tube
4
Steel Shaft
steel shaft to drive rollers
4
Gudgeon
aluminum insert for rollers
4
Spacer
aluminum spacer for LP- 155
2
121
Part Drawings and Listing
122
APPENDIX D
Hardware, Software and
Common Operation Tips
Tips on Using the Hardwareand Software of
Machine
To Improve Performance of Gearboxes
Lower Jog Feedrate
Default Value: 2000
Try 500-1000
Goto Parameter Mode
Goto tab NC Params.
Increase Smoothing Time
Default Value:60
Try 600
Do NOT exceed 5000ms
123
Hardware, Software and Common Operation Tips
Goto Servo Params.
Increase Position Loop Gain
Default Value: 5 Hz
Increase gradually by increments of 5 Hz
Do NOT exceed 20-30 Hz
Goto Servo Params
Tune the Amps
There are 4 black round little buttons on Amp.
When just powered, display should show "bb"
Push the "Mode/Set" button (1st from the left) twice. It shows PNOOO.
Change to PN1 10
Hold down data button
Make sure the 4th digit is "2". This is the control register and disables
autotuning.
Hold down data button till you get back to PN display.
Change to PN103
Hold down data button. You may see 100-150.
Increase it to 200-300. This is the inertia ratio of the gearbox to the
motor shaft as a %.
Hold down data button
The current settings for the motors controlling the rollers is 500 and
124
Tips on Using the Hardware and Software of Machine
for those controlling the arms is 1000.
Change to PN100
Hold down data button
Default value is either 25-40. This is the velocity loop gain. Make it
40.
To exit, push the "Mode/Set" button twice and you see the "bb" dis
play again.
Hardware Considerations
Check that the emergency button is not pressed on the handwheel.
This would cause the system to not respond to any commands from the PC.
Check that the power cables connected to the amps are still in place.
Loose connections are terrible!!!
Make sure that you are turning the arms in the correct direction. Failure to do so would result in the arms colliding with the gearmount thus either
activating the limit switches or causing an emergency shutdown of the software as the amps go into torque overload failure mode.
Operation Considerations
Ensure that the arms are initialized as soon as the software is loaded.
Since the encoders for the motors used are incremental, the software does not
store the absolute home position which is the almost horizontal position. Failure to do so will cause the arms to be initialized at an incorrect location.
125
Hardware, Software and Common Operation Tips
Therefore, when the program is loaded, this error will be propagated and the
desired motion of the patient will not be obtained.
Ensure that the sheet is wrapped one wrap around the roller prior to
the introduction of slack otherwise, as slack is introduced, the sheet will tend
to peel off the roller.
126
(4CVjC~ Q~A