TIPS Meeting 19 June 2003, 10am, Auditorium 1. NIRSPEC Operations Concept ` Mike Regan 2. WFC3 Optical Alignment Massimo Stiavelli 3. Increased Thermal Background for the post-NCS NICMOS Megan Sosey Next TIPS Meeting will be held on 17 July 2003. NIRSpec Operations Concept Michael Regan(STScI), Jeff Valenti (STScI) Wolfram Freduling(ECF), Harald Kuntschner(ECF), Robert Fosbury (ECF) Ops Concept: What is it? • Explain how the instrument will be used – – – – – What are the observing modes? What types of calibrations are required? How are observations planned? What are the actions required to perform an observation? Which actions are performed in flight software, flight hardware, ground planning software, and ground pipeline software? – How often will mechanisms be used? NIRSpec Optical Layout Filter Wheel Pick-off Optics Micro-Shutter Array Grating/Prism/Mirror Wheel Fore-optics Collimator Camera Detector Array Target Acquisition • Need to have maximal light from science targets going through all the slits formed by shutters – This requires getting both the correct pointing and the correct roll – After acquisition both the pointing and the roll must be held relatively constant throughout the observation. Target Location Tolerance • Assure that the ensemble throughput is not reduced by more than 10% for 95% of the observations – Leads to a two sigma error of 25 mas. – Therefore, one sigma we must be within 12 mas of desired location. – Both pointing and roll errors contribute to this error How do they interact? • Sin(roll_error) < sqrt(12mas2-pointing_error2)/100” Roll Angle Acquisition • User will be given a range of roll angles after visit has been preliminarily scheduled – User will select a roll and design their shutter mask – Chosen roll angle and shutter mask will be put into visit file – Spacecraft will use star trackers to move telescope to required roll angle Positional Acquisition • Uncertainties in the locations of stars in the GSC2 are much larger than the required (<10mas) – Have to take acquisition image to get an offset to the correct location. Microshutter Grid and Point Source Location • Microshutter grid will lead to biases in the centroid of a point source ~14mas. – More sophisticated algorithms can reduce this • Only by dithering one source or using multiple reference objects can this be averaged out. • With 9 targets get final error of 5 mas. Roll requirement • With a 5 mas positional uncertainty – Allowed roll error is ~15 arcseconds • Even with perfect positional accuracy – Allowed roll error is ~20 arcseconds • Note that this error includes the user’s uncertainty in being able to determine the required roll angle • Therefore, for now, we are assuming that roll will need to be adjusted. Image Stability • Around 1/3 of the science will be one day per grating selection – Need to be stable on this time scale – Otherwise, will have to reacquire and recalibrate – Spacecraft roll about FGS star will need to be stable to within ~1.7” per day • Smaller due to larger radius to FGS star Steps in a Target Acquisition • Assume wheels at home locations or move them: – filter wheel at closed location – grating wheel at mirror location • Turn on calibration lamp • Take image of MSA plane (uncertain mirror location) • 1D – Centroid each fixed slit – Store away the difference between expected and actual position • Turn off lamp • Open all MSA shutters [except those around bright objects in field] • Move filter wheel to requested acquisition filter • Take acquisition images and centroid • Find ∆x, ∆y, and ∆roll • Offset pointing and roll to correct location Correcting for MSA and Cosmic Ray Effects • Two method for getting required positional precision – One object, multiple dithers, (slower but more flexible) – Multiple objects, one dither, (faster but more restrictive) • At each location take acquisition image (up-the-ramp,3) – Form two differences (1-0, 2-1) – Flat field two differences? – Take minimum pixel value (CR reject) • Add mirror offset and acquisition offset • Pass delta pix to FGS • Repeat if accuracy is not good enough. Contemporaneous Calibrations • • • • After target acquisition Switch to a long pass filter Configure MSA for observation Take a short direct image – This will help pipeline processing • • • • • • • Switch to requested grating/prism Switch to closed filter wheel Turn on emission line lamp Take a wavecal image Turn off emission line lamp Switch to filter wheel long pass filter Begin science exposures Detector Operations • NIRSpec will be detector noise limited in R>1000 modes • Up-the-ramp/Multiaccum sampling has been shown to be better than Fowler for detector noise limited observations • In addition, up-the-ramp sampling is more robust against cosmic rays Signal Level Baseline Readout Mode Reset Samples T2 T2 T2 T2 T2 Groups TIME Signal Level Alternative Readout Mode (depends on noise characteristics of flight electronics & detector) Reset Samples T2 T2 T2 T2 T2 Groups TIME Readout Parameters • Time between storing of groups on SSR = 50 sec • Samples per group = 1,4 • Number of groups = exposure_time/50 Other parameters • Sub-array readout – Minimum 12 second exposure time is too long for many sources – Sub-array readout will be needed – Only one sub-array at a time – Readout time = 12*(number of pixels in subarray/8 million) seconds Electronic Gain • Goal is to have only one gain setting for NIRSpec – Maximum gain is set by Nyquist sampling single sample read noise (~9e-) or ~4 e-/ADU – Would like to be able to use entire full well ~90K -200K e– 16 bit A/D values lead to 64K dynamic range – Saturated values can be reconstructed from early reads in up-the-ramp. – A single gain of 1.5 e- to 2.5 e- will work Calibration • Assumptions – NIRSpec will have internal line and continuum sources – Line sources will reach required S/N is a 60 sec exposure – There will be NO parallel calibration • Although it should not be ruled out – Wavelength zero point calibration are required every time the grating wheel is moved. – MSA-to-detector calibration is required every time the mirror is moved in. Monitoring Calibrations • Two types – Parallel Capable (do not require dedicated visit) • • • • • Dark current/read noise/gain Hot pixels Shutter throughput Fixed slit throughput Small scale flat field variations – Dedicated (frequency depends on stability of detectors and geometry of optical bench) • • • • • Linearity Persistence Geometric distortions Large scale flat field Wavelength solution Mission Lifetime Usages • Assumptions – – – – 5 year lifetime NIRSpec is used 50% of the time 70% of the time we are doing science All observations are multi-object MSA spectroscopy Type of projects Type Fraction of NIRSpec Time Average Time for one visit (sec) Number of visits per year Short 0.10 3K 470 Medium 0.60 20K 430 Long 0.30 100K 43 Usage for each visit Type Filter Wheel Grating Wheel MSA Lamp Short 7 4 8 4 Medium 6 3 4 3 Long 5 2 4 2 Total Usage Type Filter Wheel Grating Wheel MSA Lamp Short 5*370*9 5*370*4 5*370*8 5*370*4 Medium 5*330*7 5*330*3 5*330*6 5*330*3 Long 5*33*5 5*33*2 5*33*4 5*33*2 Total 29K 13K 25K 13K Optical alignment and image quality testing at GSFC • Test lead: Bill Eichorn, GSFC Test conductors: George Hartig, Sylvia Baggett, Massimo Stiavelli • Full suite of alignment and image quality tests performed twice: before and after vibe test. 1st epoch : 21 Mar-18 Apr ’03, 2nd epoch ongoing – WFC3 with non-flight detectors (IR MUX, UVIS surrogate) and flight POM installed in RIAF with OS (CASTLE) providing OTAlike point source illumination at 633 nm (only) – During 2nd epoch flat fields in F336W, F439W, and F625W were also obtained – Over ~4000 images obtained on both IR and UVIS channels May 22, 2003 WFC3 Report at the June 03 TIPS 1 Optical alignment and image quality testing at GSFC • Test objectives: – Demonstrate Optical Stimulus (CASTLE) performance over field; develop techniques and procedures – Obtain pre-vibration (of OA) alignment baseline and compare it with post-vibration data – Compare with results obtained at Ball with Mini-Stimulus – Measure residual surrogate detector alignment offsets for use in flight detector alignment – Assess image quality and wavefront performance of the completed optical assembly May 22, 2003 WFC3 Report at the June 03 TIPS 2 Optical Stimulus (CASTLE) Rm 150 Bld 29 Clean room May 22, 2003 WFC3 Report at the June 03 TIPS 3 CASTLE Schematic M4 Steering Mirror Precision Encoders M3 Steering Mirror Precision Encoders • Inverted Cassegrain with 46% Central Obscuration M1 Stimulus Relay Optics • Images positioned over field with 2 gimballed flats produces pupil rotation M2 Switchin g Mirror Slides in plane for Extended Sources Integrating Sphere Target Plane Upper Reticle Point Sources WFC3 POM • Point source XYZ stage permits accurate focus adjustment Fiber Feed RIAF OTA Latch Plane ABC Latches Lower Ret icle OTA Sta 198.440 Alignment Monitor May 22, 2003 Flux Monitors WFC3 Report at the June 03 TIPS 4 CASTLE/RIAF in SSDIF May 22, 2003 WFC3 Report at the June 03 TIPS 5 CASTLE Test Image Field Locations • Measurements made at 16-17 points, uniformly-distributed over IR and UVIS fields (4-5 used at Ball) May 22, 2003 WFC3 Report at the June 03 TIPS 6 IR MUX Image at Field Center May 22, 2003 WFC3 Report at the June 03 TIPS 7 UVIS Image Near Field Center May 22, 2003 WFC3 Report at the June 03 TIPS 8 UVIS alignment focus scans with WFC3 corrector and CASTLE May 22, 2003 WFC3 Report at the June 03 TIPS 9 IR channel focus scans feasible with CASTLE only May 22, 2003 WFC3 Report at the June 03 TIPS 10 Phase Retrieval • Phase retrieval (PR) analysis measures low-order aberration content of the system at the detector, including focus • Sets of monochromatic images obtained at multiple focus settings, on both sides of best focus, and fit simultaneously to remove ambiguity • IDL PR routines developed by Burrows & Krist (ca 1990), used for COSTAR, STIS, NICMOS, and ACS optical verification and alignment, and adapted for WFC3 • Only 13 lowest-order terms in Zernike expansion used for wavefront fit; generally provides good indicator of alignment state and final optical performance May 22, 2003 WFC3 Report at the June 03 TIPS 11 IR Channel Alignment Phase Retrieval Pupil/WFE map (cold mask) Measured MUX image Modeled (fit) image Weighted difference image May 22, 2003 WFC3 Report at the June 03 TIPS 12 UVIS Channel Alignment Phase Retrieval Pupil/WFE map (OS obsc/spiders) Measured MUX image Modeled (fit) image Weighted difference image May 22, 2003 WFC3 Report at the June 03 TIPS 13 Optical alignment summary • • Phase retrieval and focus scan (CASTLE fiber or WFC3 corrector) techniques yield similar, repeatable results Residual detector tip/tilt after alignment at Ball corroborated with CASTLE measurements using alternative focal surface alignment technique – Alignment of both detectors should be improved with shim offsets as flight detectors are aligned in transfer fixture at Ball • UVIS detector roll off nominal by 0.8 degrees, position off by ~50 pixels – Set incorrectly at Ball, will be compensated for flight detectors • • • IR detector roll nearly perfect (<0.1 degrees from nominal) IR focus is essentially identical to measurement at Ball UVIS focus differs by ~0.5 mm (STOPT) from Ball measurement – Believed to be due to offset in model positions for Mini-stimulus • Vibration testing did not alter alignment or image quality apart from a small shift – identical in both channel – probably due to latches tolerance May 22, 2003 WFC3 Report at the June 03 TIPS 14 UVIS Encircled Energy Measurements vs. Model • CASTLE EE measurements require correction for 0.46 central obscuration ratio (OTA is 0.33), spider width and other differences from OTA, as well as limited measurement radius, to predict on-orbit EE performance. • CEI Spec: EE(.25 arcsec diam.) > 0.75 (goal 0.80) at 633 nm UVIS,??=633 MS (Ball) OS (GSFC) OTA Model: Perfect ? =0.015 ? =0.25 0.67 0.86 0.58 0.85 0.65 0.84 Model: 0.03? RMS ? =0.15 ? =0.25 0.63 0.84 0.55 0.83 0.63 0.82 Actual Measured ? =0.15 ? =0.25 0.62-0.66 0.82-0.85 0.55-0.58 0.80-0.83 ?SMOV? ?SMOV? • EE spec will readily met, goal will likely be achieved onorbit, if flight detector performance and alignment are as expected. May 22, 2003 WFC3 Report at the June 03 TIPS 15 Flat Field Analysis C-D amp May 22, 2003 WFC3 Report at the June 03 TIPS 16 Flat Field Analysis • Many negative pupil images seen in the flat fields. They do not depend on the filter and have sizes compatible with what would be espected from dust particles on the surrogate UVIS detector. • Vignetting has been found on one corner of the field of view. It is a triangular area about 400 pixels long and 30 pixels wide at the largest point. Over this area the flux rolls down by more than a factor 2. – Not caused by filters, pickoff or channel select mechanism: too sharp (pickoff) and unaffected by rotating filters/CSM to offnominal positions – It may be caused either by baffles close to the detector or by the camera head itself (currently at an incorrect position) May 22, 2003 WFC3 Report at the June 03 TIPS 17 Flat Field Analysis A-B Amp May 22, 2003 WFC3 Report at the June 03 TIPS 18 Conclusions • The various sets of measurements are consistent • The data will allow us to position the flight detectors at the correct position • Vibration testing did not significantly alter alignment or image quality • Small amount of vignetting in UVIS channel (still to be investigated) • EE field center goal on UVIS should be met over the whole field May 22, 2003 WFC3 Report at the June 03 TIPS 19 Increased Thermal Background for the post-NCS NICMOS TIPS – June 19, 2003 Megan Sosey NICMOS Cycle 11 Calibration Plan ♣ Parallel observations in NIC3:F222m, NIC2: F222m & F237M were crafted to re-measure the thermal background for cycles 11 and beyond -See proposals 9269 and 9702 ♣ Accounting for DQE increase, the average thermal background was predicted to be approximatly 20% higher than in Cycle 7 So, Where’s the extra signal? ♣ NO correlation was found between the variation in the thermal background and – – – – – telescope pointing sun angle prime instrument time since SAA passage or spacecraft orientation ♣ The T-1-1 temperature sensor shows the detectors are stable under operation of the NCS and do not change by more than 0.1K on long time scales T-1-1 Temperature Sensor vs. Time Telescope Warm Season 9/10/2002 Commanded Setpoint change - 5/30/2003 Thermal Stability of the NICMOS Enclosure and Telescope Assembly ♣NICMOS elements which affect the TB: re-imaging mirror (RM), FOM, PAM, filters, FDA(imaging mirrors), cold mask, bend mirrors, baffles, pupil ♣HST optics which affect the TB: primary mirror, secondary mirror, spiders,pads – these play significant roles in the TB because of their high temperatures (~290K) Representation of Cold Mask Misalignment See Robberto, Proc. SPIE v4013, p.386 and Krist, NICMOSISR-00-011 for more details on the cold mask alignment Rough Optical Path Diagram for the NICMOS Fore-Optics Thermal stability of the HST optics and aft shroud ♣The primary and secondary have remained thermally stable, introducing no extra thermal emission - however, the aft shroud has increased approximately 10K since Cycle 7 ♣There are multiple thermistor sensors (thermally sensitive resistors) in the aft shroud which are used to monitor temperature -TAFTBULK is a weighted average of the internal sensors located on the aft bulkhead -TASINAFB is an effective sink measurement from all the internal sensors, including some of the ones from the TAFTBULK average Thermal Environment of the HST Aft Shroud around NICMOS 2001-2002 Thermal Environment of the HST Aft Shroud around NICMOS 2001-2002 Aft Temperature Comparison of HST and NICMOS naftbtmp NIC3, F222m HST AFT TASINAFB TAFTBULK C7 & C11 NICMOS Aft Temps NICMOS Temperature Variations Avg. C7 Temp Avg. C11 Temp Delta T Key Descrip. nfob2tmp -5.82 C -0.48 C +5.34 C ndosftmp Fore-optical brkt. Dewar Fore -7.27 C -1.21 C +6.06 C ntrs3tmp Truss Aft 1.22 C 7.52 C +6.3 C ndosatmp Dewar Aft -1.60 C 5.30 C +7.24 C naftbtmp Aft NIC Encl. 8.5 C 17.5 C +9.0 C taftbulk HST Aft Shroud -17 C -7 C +10.0 C ♣The NIC fore-optics swing in temperature in the same manner as the aft end, but with a smaller amplitude Direct plot of thermal background vs fore-optics temperature Does this explain the increase? ♣The previous plot indicates that a 2K variation in temperature, translates to a 1.5DN change in the thermal background ♣An increase of 5K between Cycle 7 and Cycle 11 yields an additional 3DN of thermal background signal! - this makes up the difference between the predicted and observed thermal background Simple blackbody comparison of HST and NICMOS optical surfaces using NIC2 Optical Parameters Comparing to the NICMOS Thermal Background Code Estimates Conclusions The overall temperature environment of NICMOS and the operating temperature of the detectors has changed since cycle7 resulting in: • • • • Increase of ~17K in detector temperature General temp. increase in the aft HST Increased temperatures in the aft NICMOS Increased thermal background in long wavelength, or extra wide, camera 2 and camera 3 filters • Thermal background exposures are still affective in removing the excess signal