Instrument Science Report STIS 2000-03 Sensitivity Monitor Report for the STIS First-Order Modes -II Ed Smith, David Stys, Nolan Walborn and Ralph Bohlin September 21, 2000 ABSTRACT An analysis of all low and medium resolution observations of the STIS Sensitivity Monitoring Programs made prior to December 1999 shows continuation of trends correlated with temperature and with time. We provide the first detailed report on additional trends in sensitivities revealed by the medium resolution modes of the instrument. Only the G750L mode shows no significant variation in sensitivity. All other modes, when averaged over a broad band of the wavelength coverage of each, show evidence for an ~1%/yr decline, except for G140L and G230L (after ~1998.7) being ~1.4%/yr, G140M being ~2.5%/yr, and G430L declining at ~0.5%/yr. A bimodal, before and after ~1998.7, trend previously found in the G230L is found in the NUV-MAMA G230M. Similar trends are also evident in the CCD in the G230LB and the two G230MB monitoring modes. The temperature dependence of the sensitivity in the G140M modes is found to be consistent with the -0.25%/˚C correlation measured in the G140L mode. The mean “relative gain”, gain1/gain4, of the CCD between 1998.3 and 1999.7 was 4.039 ±0.006 and did not change significantly in that period. Recommendations are made for changes in the ‘pipeline’ calibration software to correct for the time and temperature variations in all L modes and in the M modes of both MAMA detectors. . Instrument Science Report STIS 2000-03 1. Introduction This is an update to the analyses of the sensitivity monitoring observations of the STIS first order modes. In particular, we report on trends in sensitivities of the medium resolution modes of the instrument. The analyses now include all Cycle 7 data and all Cycle 8 observations made prior to the November-December 1999 gyro failure and servicing mission hiatus. Prior reports of sensitivity trends measured using these monitor programs were given by Walborn and Bohlin in ISR STIS 98-27 (WB1998) and by Bohlin in ISR STIS 99-07 (B1999). 2. Observations Appendix 1 lists all data obtained by the STIS sensitivity monitoring proposal’s low resolution (L) and medium resolution (M) modes. The NUV-MAMA and FUV-MAMA L mode observations are of GRW+70D5824, while all MAMA M mode and all CCD exposures are of AGK+81D266. The data are from CAL/STIS proposals 7673 and 8424 for the MAMA detectors and proposals 7672 and 8418 for the CCD. Sufficient data points are now available from these programs that we do not include measures of other standard reference stars, nor measurements from other STIS observing programs. All observations are taken using the 52X2 slit. All CCD data are CR-SPLIT=2 and GAIN=1. Except that, beginning in March, 1998, both a GAIN=1 and a GAIN=4 exposure is made for each visit of the G430M at the 4194 Å central wavelength setting. The MAMA L modes are monitored monthly while the M mode measures are taken at two month intervals. The CCD L modes are also monitored at two month intervals while the CCD M modes are checked every four months. G750L and G750M observations have contemporaneous ‘fringe-flat’ exposures. The normal MSM shifting is disabled for these monitoring observations in order to minimize variations due to spatial displacements of the spectra. To see spectra representative of those produced by each mode in this program please refer to Figures 1-10 of WB1998. Position change (+3˝ to -3˝) of FUV-MAMA spectra. On March 15, 1999, the default position of the first order mode spectra on the FUVMAMA detector [i.e., G140L and M spectra] was changed from a position ~+3˝ above the center line of the detector to a position at -3˝. This moves the spectra away from the higher dark counts present in the ‘upper left’ region of the detector. The off-center positioning is desired in order to avoid the Repeller Wire that is positioned across the center of the detector. In section 3b we describe the changes in the monitoring analyses due to the change in position of the spectra. 2 Instrument Science Report STIS 2000-03 3. Data Reduction and Analysis. All data were calibrated with STSDAS CALSTIS Version 2.6 using the most recent updates to the various calibration reference files (bias, dark, etc) as of March 15, 2000. The NET counts for each spectrum are extracted from the X1D product for analysis of the MAMA data and the SX1 product for the CCD analysis. Small offsets in wavelength coverage in a given mode induced by small variations in position of the spectra on the detector are dealt with by extracting a consistent wavelength range for each mode. Problematic wavelength regions such as geo-coronal Lyman Alpha and G750L fringing are also avoided by the choice of the wavelength passbands being extracted. Those passbands are listed in TABLE 1. Our analyses of sensitivity changes vs time will be made relative to the Two additional steps in reduction of the FUV-MAMA data are applied. There is a correction applied to account for variations in sensitivity that correlate with temperature. Also, all data recorded with the spectra at the -3˝, ‘below-the-repeller-wire’ position on the detector are corrected for the difference in sensitivity at that position relative to the +3˝ position. After those steps, descibed below, we examine trends in the “relative sensitivity”, i.e the sensitivity relative to a reference spectrum. For each mode that reference is the first observation in the series for that mode, and for the FUV-MAMA the reference spectrum is at the +3˝ position on the detector at a detector temperature of 36˚C. 3a. FUV-MAMA: Correction for Temperature dependence. WB1998 deduced a correlation between the G140L sensitivity measurements and the temperature recorded at the charge amplifier (OM1CAT in the.spt file) of the FUVMAMA detector. With additional measurements in hand, that relationship was updated slightly in B1999. The reported effect is a change in sensitivity of -0.25% per degree (˚C) change in OM1CAT. We perform a similar analysis of the other FUV-MAMA first order data, G140M. We find that for G140M(1150-1190Å) the variation is -0.42%/˚C ±0.08 (rms=0.33) and for G140M(1542-1592Å) it is -0.22 ±0.09 (rms=0.40). We combine the G140M data by normalizing both datasets to 1 at a temperature which they have in common, 34 ˚C. This yields a fit of -0.31%/˚C ±0.06 (rms=0.39), in good agreement with the G140L analysis. We use the -0.25%/˚C, G140L result, to remove temperature dependent trends from both the FUV-MAMA L and M mode data, normalizing each observation’s sensitivity to a reference temperature of 36 ˚C, i.e.: FUV-MAMA Relative Sensitivity = Relative Sensitivity + 0.0025 *(OM1CAT -36.0) 3b. FUV-MAMA: Correction for +3˝ to -3˝ repositioning of G140 L, and M, Spectra Presumably due to lack of a low order flatfield calibration [the extraction traces were checked and found to be nominal], there is a small but significant change in sensitivity at 3 Instrument Science Report STIS 2000-03 the new default position of the first-order spectra on the FUV-MAMA detector. We remove that difference to arrive at a uniform set of data covering the epochs before and after the change. For G140L data, we use the correction derived and described in the “Time Correlations” section in B1999. For each set of G140M observations, CENWAVE 1173 and 1567, we perform a derivation similar to that for the G140L. After extracting the net counts spectrum from each reduced, X1D, observation, we sum those counts that are within the broadband wavelength ranges listed in column 4 of TABLE 1. Using just the data taken from the +3˝ position, a linear fit is made of the trend of those total net counts versus time. That fit is then used to normalize all, both +3˝ and -3˝, datapoints to the date of the earliest data point. With the time variation removed, a ratio spectrum is produced by dividing the average of the +3˝ by the average of the -3˝ spectra. That measure of the difference in sensitivity at the two detector positions is presented in FIGURES 1 and 2. The “noisy” line in each figure is the ratio spectrum while the smooth line is a cubic spline fit (5 splines) to the ratio. We apply these corrections in wavelength space and plot those wavelengths along the x-axis. However, in assuming that these are flatfield variations, it is appropriate to consider the x-axis to be related to the pixel to pixel column position across the detector. The broadband average of the ratio spectrum is used as correction factor to normalize all -3˝ datapoints to the +3˝ values. That correction is applied to the net count, X1D, data prior to the following analyses of trends in time. Figure 1: Ratio of the G140M(λc 1173Å) average at +3˝ to average at -3˝ after temperature and time corrections to sensitivity. 4 Instrument Science Report STIS 2000-03 Figure 2: Ratio of the G140M(λc 1567Å) average at +3˝ to average at -3˝ after temperature and time corrections to sensitivity. 3c.Sensitivity Variations with Time After the G140L data are normalized to 36 ˚C and all FUV-MAMA data are normalized to a +3˝ position on the detector, we examine each of the monitoring modes for any variation in sensitivity with time. For each observation the total of the net counts in the wavelength regions given in Table 1 is divided by the value from the first, earliest observation in that mode. A linear fit, using iraf.stsdas.ttools.tlinear, is made of the trend of these relative sensitivities per year of time. Following the deduction of B1999 we split the G230L, NUV-MAMA, spectra into two epochs, before and after ~1998.7, and apply separate linear fits to each epoch. To further examine this “two-epoch” trending we perform a similar analysis of the other modes whose spectra are within the wavelength range of the G230L, i.e, G230M from the NUV-MAMA and G230MB and G230LB from the CCD. Figures 4 through 22 present the relative sensitivity vs. time for each monitoring mode, plus the linear fits to those points. The slope, i.e. per cent per year change in sensitivity, and 1 σ uncertainty in the fits is given at the bottom of each figure. The 1 σ rms(%) of the data residuals from their linear fit values is given by the SIGMA value also at the bottom of each figure. A summary of those values for each mode is contained in TABLE 1. 5 Instrument Science Report STIS 2000-03 Table 1. Analysis Parameters and Results of Linear Fits to Sensitivity vs. Time. Epoch (All, unless noted) Extracted Wave_range (Å) %/yr ± Sigma (%) (rms of fit residuals) G140L 1300-1500 -1.36 0.15 0.53 5 G140Mλ1 1150-1190 -3.51 0.48 0.69 6 G140Mλ2 1542-1592 -2.19 0.24 0.49 7 G230L 2200-2600 -0.12 0.16 0.55 8 G230L <1998.7 2200-2600 1.49 0.37 0.36 8 G230L >1998.7 2200-2600 -1.36 0.15 0.20 9 G230M 2775-2860 0.02 0.16 0.33 10 G230M <1998.7 2775-2860 1.13 0.22 0.11 10 G230M >1998.7 2775-2860 -0.58 0.18 0.18 11 G230LB 2000-3000 -0.14 0.16 0.39 12 G230LB <1998.7 2000-3000 0.72 0.21 0.16 12 G230LB >1998.7 2000-3000 -1.25 0.31 0.24 13 G230MBλ1 1920-2070 -0.11 0.43 0.70 14 G230MBλ1 <1998.7 1920-2070 0.84 0.64 0.38 14 G230MBλ1 >1998.7 1920-2070 -3.06 0.83 0.27 15 G230MBλ2 2340-2490 -0.67 0.27 0.45 16 G230MBλ2 <1998.7 2340-2490 0.35 0.53 0.32 16 G230MBλ2 >1998.7 2340-2490 -1.91 0.01 0.01 17 G430L 3100-5500 -0.26 0.26 0.63 18 G430Mλ1 3050-3300 -0.87 0.11 0.18 19 G430Mλ2 4060-4320 -1.07 0.08 0.12 20 G430Mλ2 4060-4320 -1.31 0.12 0.12 21 G750L 5600-7000 0.07 0.16 0.39 22 G750M 7000-7500 -1.21 0.13 0.21 Figure # Mode 4 GAIN 4 A graphical summary of TABLE 1 is provided by Figure 3. For each mode we plot the derived yearly change in sensitivities vs. the central wavelength of the grating. The diamonds are the M mode data and the x’s are the L mode. The vertical error bars are the 1 σ uncertainties in the sensitivity trends. The horizontal “error bars” are the range in wavelength extracted for analysis of the given mode. The numbers at the left of each data point 6 Instrument Science Report STIS 2000-03 are the index (figure) number for that mode as provided in the first column of Table 1. The cluster at ~2500Å of non-negative sensitivity trends are the pre-1998.7 two-epoch measures. Figure 3: Summary of Rates of Change in Sensitivity per Instrument Mode and Wavelength. 4. Discussion of Sensitivity vs. Time. The current results are in general agreement with previous reports; however, additional trends are suggested, particularly by extension of the two-epoch analyses of B1999 to modes other than G230L, as well as by comparing overall trends separately in the L and in the M modes. 4a. Present results vs. previously reported trends. The rates of change in sensitivity with time (%/yr column in Table 1) are consistent with those found for matching instrument modes (i.e., the L modes) presented in B1999 and WB1998. The values are within approximately 1 σ with respect to B1999. We make 7 Instrument Science Report STIS 2000-03 that comparison by averaging the fit values for those wavelength bins in B1999 which cover the same wavelength range as the ‘single bin’ passbands of the current analyses. The M mode time variations in Table 1 are also consistent with those from similar wavelength regions in the L mode fits of B1999. However, that consistency generally requires allowing 2 σ variations in the fits. In one particular mode, G750M (7000-7500Å), the change in sensitivity versus time is 6 σ different from a similar passband in the G750L analysis in B1999 ( -1.21 ±0.13 vs. 0.28 ±0.08). 4b. Two-epoch analysis of modes covering 1900-3000Å. The G230L (MAMA) mode was shown by B1999 to exhibit a bimodal time variation in sensitivity. It increases in sensitivity prior to ~1998.7 and decreases thereafter. This two-epoch variation was suggested by the apparent non-random scatter from a single epoch linear fit and by the improvement in the fit-residuals when splitting the fit into two epochs (see Figs. 7 and 8). We expand this two-epoch analysis to all modes with wavelength coverage within that of the G230L. While a single linear fit to all of the G230M (MAMA) data (Fig. 9) indicates, with relatively large uncertainty, no variation with time, two-epoch fits (Fig. 10) yield slopes significant to 3 σ and reduce the fit residuals by a factor of 2. The suggested variations in sensitivity agree with the equivalent passband in B1999’s G230L analysis, within ~1 σ before ~1998.7 and ~2 σ after that date. Similarly, two-epoch fits to G230LB(2000-3000Å) (CCD), and to the two separate central wavelength datasets of G230MB (CCD) yield significant reductions to the fit uncertainties and residuals (compare Figs. 11 through 16). [Though this report includes only 3 post-1998.7 datapoints for G230MB, quicklook analyses (Stys and Walborn, private communication, 19Jul00) using two additional recent proposal 8418 spectra find the post-1998.7 trend continuing in both G230MB modes. If additional data validate these G230MB trends, these two modes exhibit the largest rate of decline in sensitivity, at ~ -2 to -3 %/yr.] Overall the data in all MAMA and CCD modes covering 1900-3000Å suggest a trend in sensitivity that is increasing prior to ~1998.7 and which is decreasing thereafter. The data from modes redward of 3000Å, G430<*> and G750<*>, suggest a single linear trend over the entire monitoring period. These trends suggest that the changes affecting G230L are not particular to that mode or detector, but, instead, are more generic processes which are wavelength dependent and strongest in the 2000-3000Å range. 4c. M modes versus L modes. Except for G230M, our linear fit measures of the rates of loss in sensitivity for all M modes are greater than those of the ‘comparable’ (same detector and equivalent central 8 Instrument Science Report STIS 2000-03 wavelengths) L modes. This can be seen in comparisons using the %/year change in sensitivity values in Table 1, Figure 3, and the averages of the equivalent passbands in the L modes of the B1999 analyses. However, given the uncertainties of the fits, the actual significance of these M vs. L mode differences varies from <1 σ to >5 σ. For most modes the differences are marginal (<3 σ). But, the difference in rates of change are more significant, 3 σ and >5 σ respectively, for the longer wavelength(4060-4320A) G430M vs. G430L, and the G750M vs. G750L. It is expected that any sensitivity changes may vary with wavelength. It is very difficult to ascribe any possibility of an overall M vs. L mode difference to a single effect because of the differences in the optical elements in the light path of each mode. However, for the CCD modes, where the difference in rate of change of sensitivity is most significant, the only difference in the optical elements (besides the gratings) is that the L modes use different order sorting filters than the M modes. Further monitoring and a more detailed, wavelength dependent, analysis should be the next step in tracing the source of these differences. From the present data and analyses, the most definite conclusions are that the MAMA M modes track the corresponding L mode wavelength regions, while the CCD G430M and G750M modes exhibit significantly greater changes than any corresponding L mode regions. 4d. Gain 4 vs. Gain 1. Figure 23 shows the ratio of each of the 5 pairs of gain 1,gain 4 G430M(4060-4320) measurements covered by this report. The error bars are the propagated poissonian errors and a linear fit is overplotted. There was no significant change within the time period covered. The mean value is 4.039 ±0.006 (WB1998 found 4.034 ±0.010). 4e. Repeatability If one obtains spectra in the same observing mode, with comparably high signal, the ‘SIGMA’ values in Table 1 represent the 1 σ uncertainty, or repeatability, in the absolute flux assignments. That is, assuming that the temperature and/or time variations in sensitivity are corrected. 5. Summary We have added detailed M mode trends to the characterization of STIS sensitivity trends. While the known trends are generally confirmed we show the need to expand the G230L ‘two-epoch’ analysis to other modes. In particular, we found that: 9 Instrument Science Report STIS 2000-03 • Only the G750L shows no significant variation in sensitivity. For the G430L, our average over a broad band, 3100-5500Å, is within 1σ of the B1999 result of -0.56% ±0.16%. Most other modes show evidence for an ~1%/yr decline except for G140M being ~2.5%/yr and G140L being ~1.4%/yr (broad band average, but tracking the larger declines at the wavelengths corresponding to the two G140M modes that are monitored) and G230L (after ~1999.7) also declining at ~1.4%/yr . Also, if confirmed by additional data, the two monitored G230MB modes are declining at ~ 2 to 3%/yr after ~ 1998.7 • A bimodal, before and after ~1998.7, trend evident in the NUV-MAMA G230M is similar to that of the G230L. Similar trends are also evident in the CCD G230LB and the two G230MB monitoring modes. • The rate of decline in sensitivity is systematically greater in some of the M modes (G430M(4060-4320) and G750M) relative to the corresponding L modes. • The temperature dependence of the sensitivity in the G140M modes is consistent with the 0.25%/˚C correlation in the G140L mode. • The “relative gain”, gain1/gain4, of the CCD has not changed significantly. Between 1998.3 and 1999.7 the mean of this ratio was 4.039 ±0.006. • The corrections needed to normalize data taken at +3” versus -3” from the center line of the FUV-MAMA detector indicate that the lack of low order flatfielding leaves residual sensitivity variation of as much as 8%, peak to peak, between the two positions. 6. Recommendations • Pipeline corrections should be implemented to account for the FUV-MAMA variation of 0.25%/˚C in both G140L and G140M modes. Future derivations of that relation should include both sets of G140M data. • The wavelength-dependent correction for the time variations in G140L, G230L, and G430L (a null correction to G750L) recommended in B1999 should be implemented in the pipeline. • Our current suite of IRAF-based analysis software should be enhanced to provide wavelength dependent analysis of each mode. • When a low-order flatfield is available for the FUV-MAMA, the +3˝/-3˝ correction should be removed from these monitoring analyses (if the two prove to be consistent). Until then, that correction should be derived using the iterative procedure described in B1999. • When a ‘two-epoch’ wavelength-dependent analysis of the G230LB has been completed, a pipeline correction for time variation in that mode should be implemented in the same manner as the G230L. • The narrow passbands of the G140M, G230M, and G230MB modes should not require wavelength-dependent corrections for their time variation; thus, the correction algorithm in the pipeline (two-epoch for the latter two modes) should easily allow for a correction that is constant in wavelength for these modes. 10 Instrument Science Report STIS 2000-03 • With the additional data in hand as of June 2000, a wavelength-dependent comparison of the FUV-MAMA and NUV-MAMA M modes to the comparable L modes should be adequate to confirm that there is no significant difference in the time variations between those particular M and L modes. If that is confirmed, the L modes for those detectors should be used to derive the pipeline correction for wavelength-dependent time variation in the passbands of all of the G140M and G230M modes (two-epoch correction for G230M). It is preferable to use the L modes because they receive more frequent monitoring and will better track changes in time; their larger number of measurements yield better number statistics; and because we monitor only a few of the wavelength settings for the M modes, thus, corrections for only those few would be available. The monitored M modes should be used to periodically confirm that they are being tracked by the L modes. • We do not recommend similar pipeline corrections for the CCD M modes. The CCD L modes have not been shown to track changes in the wavelength regions of the corresponding M modes and there are only a few of the M modes being monitored and thus potentially correctable. However, the sensitivity calibration files currently being used for the CCD M modes should be updated. • The probable two-epoch time trends for the G230MB modes leave too few data points to adequately define either the pre- or post-1998.7 trends. Sufficient data points to establish reliable post-1998.7 trend coefficients should be in hand by summer 2001. Then, given the other CCD L modes which do not track the M mode variations, a review should be made of whether the G230MB mode time variation is significantly different than the G230LB. • Unlike the other M modes, the wider passbands of the CCD G430M and G750M modes may warrant a wavelength-dependent correction for time variations. Since the G430L mode was not shown to track the time variations in the G430M(4060-4320) mode, we recommend that all G430M time corrections be derived independently using the G430M data in hand as of June, 2000. The independence of the G430M and the equivalent G430L passbands should be revisited in summer 2001. Similarly, a G750M time correction should be derived from a wavelength-dependent analysis of the G750M data in hand at the date of this ISR. However, the lack of data at most M mode settings introduces a basic uncertainty about possible differences among them. • The analysis software should be modified to project all sensitivity trends back to the same date for all modes, e.g. 1997.38 as recommended by B1999, in order to more easily assess the total sensitivity changes and compare them mode to mode. • The larger scatter in the CCD L mode measurements relative to the M modes (Figs.1822) should be investigated. • The echelle mode data should be investigated. In particular, the medium resolution echelle, EM modes, provide complete spatial and wavelength coverage of the MAMA detectors, which may help resolve ambiguities remaining from the use of the firstorder data. 11 Instrument Science Report STIS 2000-03 7. REFERENCES Walborn, N., & Bohlin, R. 1998, Instrument Science Report, STIS 98-27, (Baltimore:STScI). (WB1998). Bohlin, R. 1999, Instrument Science Report, STIS 99-07, (Baltimore:STScI). (B1999). Thanks to Susan Rose for the FRAMEmaking! 12 Instrument Science Report STIS 2000-03 Figure 4: Linear fit to Relative Sensitivity vs. Time. With the %/yr slope of the fit, the 1 σ uncertainty in the fit and the 1 σ rms scatter about the fit. Figure 5: 13 Instrument Science Report STIS 2000-03 Figure 6: Linear fit to Relative Sensitivity vs. Time. With the %/yr slope of the fit, the 1 σ uncertainty in the fit and the 1 σ rms scatter about the fit. Figure 7: 14 Instrument Science Report STIS 2000-03 Figure 8: Linear fit to Relative Sensitivity vs. Time. With the %/yr slope of the fit, the 1 σ uncertainty in the fit and the 1 σ rms scatter about the fit. Figure 9: 15 Instrument Science Report STIS 2000-03 Figure 10: Linear fit to Relative Sensitivity vs. Time. With the%/yr slope of the fit, the 1 σ uncertainty in the fit and the 1 σ rms scatter about the fit. Figure 11: 16 Instrument Science Report STIS 2000-03 Figure 12: Linear fit to Relative Sensitivity vs. Time. With the %/yr slope of the fit, the 1 σ uncertainty in the fit and the 1 σ rms scatter about the fit. Figure 13: 17 Instrument Science Report STIS 2000-03 Figure 14: Linear fit to Relative Sensitivity vs. Time. With the %/yr slope of the fit, the 1 σ uncertainty in the fit and the 1 σ rms scatter about the fit. Figure 15: 18 Instrument Science Report STIS 2000-03 Figure 16: Linear fit to Relative Sensitivity vs. Time. With the %/yr slope of the fit, the 1 σ uncertainty in the fit and the 1 σ rms scatter about the fit. Figure 17: 19 Instrument Science Report STIS 2000-03 Figure 18: Linear fit to Relative Sensitivity vs. Time. With the %/yr slope of the fit, the 1 σ uncertainty in the fit and the 1 σ rms scatter about the fit. Figure 19: 20 Instrument Science Report STIS 2000-03 Figure 20: Linear fit to Relative Sensitivity vs. Time. With the %/yr slope of the fit, the 1 σ uncertainty in the fit and the 1 σ rms scatter about the fit. Figure 21: 21 Instrument Science Report STIS 2000-03 Figure 22: Linear fit to Relative Sensitivity vs. Time. With the %/yr slope of the fit, the 1 σ uncertainty in the fit and the 1 σ rms scatter about the fit. Figure 23: “Gain Ratio”: G430M(4060-4320) GAIN 1 / G430M(4060-4320) GAIN 4 22 Instrument Science Report STIS 2000-03 APPENDIX 1:OBSERVATION LOG for the STIS Sensitivity Monitoring Program (through December 1999). Observation Mode CenWave Date Time (UT) Propid Exptime (sec) G140L (CenWave =1425) o45901010 G140L 1425 08/12/97 9:18:10 7673 180 o45910010 G140L 1425 09/16/97 21:11:23 7673 180 o45911010 G140L 1425 10/06/97 13:36:39 7673 180 o45912010 G140L 1425 11/17/97 17:23:40 7673 180 o45913010 G140L 1425 12/19/97 23:54:46 7673 180 o45914010 G140L 1425 01/09/98 15:03:40 7673 216 o45915010 G140L 1425 02/12/98 14:55:32 7673 189 *o45916010 G140L 1425 03/11/98 9:21:19 7673 failed o45917010 G140L 1425 04/12/98 17:57:12 7673 204 o45940010 G140L 1425 05/07/98 2:51:18 7673 204 *o45941010 G140L 1425 05/27/98 19:16:27 7673 failed o45942010 G140L 1425 07/04/98 18:54:31 7673 204 o45943010 G140L 1425 08/05/98 7:38:58 7673 204 o45944010 G140L 1425 09/23/98 8:02:51 7673 204 o45945010 G140L 1425 10/14/98 22:40:39 7673 204 o45946010 G140L 1425 11/04/98 6:35:51 7673 204 o45947010 G140L 1425 12/14/98 14:51:39 7673 204 o45948010 G140L 1425 01/03/99 8:36:14 7673 204 o45949010 G140L 1425 02/11/99 12:54:13 7673 204 o45950010 G140L 1425 03/08/99 16:55:02 7673 204 o45951010 G140L 1425 04/11/99 5:52:14 7673 204 o45952010 G140L 1425 05/12/99 18:15:58 7673 204 o45953010 G140L 1425 06/03/99 2:08:16 7673 204 o5jj01010 G140L 1425 07/03/99 8:44:21 8424 201 o5jj02010 G140L 1425 08/03/99 0:32:16 8424 204 o5jj03010 G140L 1425 09/18/99 0:30:05 8424 201 o5jj99010 G140L 1425 10/01/99 18:14:28 8424 204 **o5jj99030 G140L 1425 10/01/99 18:35:41 8424 204 o5jj05010 G140L 1425 11/09/99 14:04:08 8424 201 23 Instrument Science Report STIS 2000-03 Observation Mode CenWave Date Time (UT) Propid Exptime (sec) G140M (CenWave=1173) o45921020 G140M 1173 05/17/98 18:48:20 7673 354 o45922020 G140M 1173 07/05/98 17:25:21 7673 354 o45923020 G140M 1173 09/13/98 12:50:06 7673 354 o45924020 G140M 1173 11/07/98 3:57:37 7673 354 o45925020 G140M 1173 01/16/99 0:38:55 7673 354 o45926020 G140M 1173 03/08/99 15:01:44 7673 354 o45927020 G140M 1173 05/14/99 15:25:53 7673 295 o5jj15020 G140M 1173 07/01/99 13:10:20 8424 210 o5jj16020 G140M 1173 09/11/99 20:10:46 8424 210 o5jj17020 G140M 1173 11/08/99 23:41:21 8424 210 G140M (CenWave=1567) o45902010 G140M 1567 09/19/97 18:13:12 7673 360 o45920010 G140M 1567 01/09/98 9:48:28 7673 360 o45921010 G140M 1567 05/17/98 18:35:17 7673 334 o45922010 G140M 1567 07/05/98 17:12:07 7673 354 o45923010 G140M 1567 09/13/98 12:36:52 7673 354 o45924010 G140M 1567 11/07/98 3:44:23 7673 354 o45925010 G140M 1567 01/16/99 0:25:46 7673 354 o45926010 G140M 1567 03/08/99 14:48:35 7673 354 o45927010 G140M 1567 05/14/99 15:12:53 7673 295 o5jj15010 G140M 1567 07/01/99 12:57:15 8424 300 o5jj16010 G140M 1567 09/11/99 19:57:41 8424 300 o5jj17010 G140M 1567 11/08/99 23:28:16 8424 300 G230L (CenWave=2376) o45901020 G230L 2376 08/12/97 10:14:48 7673 184 o45910020 G230L 2376 09/16/97 21:21:49 7673 184 o45911020 G230L 2376 10/06/97 13:47:05 7673 184 o45912020 G230L 2376 11/17/97 17:34:06 7673 184 o45913020 G230L 2376 12/20/97 0:05:12 7673 184 o45914020 G230L 2376 01/09/98 15:14:20 7673 216 o45915020 G230L 2376 02/12/98 15:05:45 7673 216 24 Instrument Science Report STIS 2000-03 Observation Mode CenWave Date Time (UT) Propid Exptime (sec) *o45916020 G230L 2376 03/11/98 9:31:32 7673 failed o45917020 G230L 2376 04/12/98 18:07:40 7673 204 o45940020 G230L 2376 05/07/98 3:01:46 7673 204 *o45941020 G230L 2376 05/27/98 19:26:46 7673 failed o45942020 G230L 2376 07/04/98 19:04:50 7673 204 o45943020 G230L 2376 08/05/98 7:49:17 7673 204 o45944020 G230L 2376 09/23/98 8:13:10 7673 204 o45945020 G230L 2376 10/14/98 22:50:58 7673 204 o45946020 G230L 2376 11/04/98 6:46:10 7673 204 o45947020 G230L 2376 12/14/98 15:01:58 7673 204 o45948020 G230L 2376 01/03/99 8:46:33 7673 204 o45949020 G230L 2376 02/11/99 13:04:32 7673 204 o45950020 G230L 2376 03/08/99 17:05:21 7673 204 o45951020 G230L 2376 04/11/99 6:02:33 7673 204 o45952020 G230L 2376 05/12/99 18:26:17 7673 204 o45953020 G230L 2376 06/03/99 2:18:35 7673 204 o5jj01020 G230L 2376 07/03/99 8:54:37 8424 204 o5jj02020 G230L 2376 08/03/99 0:42:35 8424 204 o5jj03020 G230L 2376 09/18/99 0:40:21 8424 204 o5jj99020 G230L 2376 10/01/99 18:24:47 8424 204 o5jj05020 G230L 2376 11/09/99 14:14:24 8424 204 G230M (CenWave=2818) o45902020 G230M 2818 09/19/97 18:27:32 7673 576 o45920020 G230M 2818 01/09/98 10:02:27 7673 576 o45921030 G230M 2818 05/17/98 19:02:08 7673 570 o45922030 G230M 2818 07/05/98 17:39:00 7673 570 o45923030 G230M 2818 09/13/98 13:03:45 7673 570 o45924030 G230M 2818 11/07/98 4:11:16 7673 570 o45925030 G230M 2818 01/16/99 0:52:34 7673 570 o45926030 G230M 2818 03/08/99 15:15:23 7673 570 o45927030 G230M 2818 05/14/99 15:38:33 7673 475 o5jj15030 G230M 2818 07/01/99 13:21:35 8424 480 25 Instrument Science Report STIS 2000-03 Observation Mode CenWave Date Time (UT) Propid Exptime (sec) o5jj16030 G230M 2818 09/11/99 20:22:01 8424 480 o5jj17030 G230M 2818 11/08/99 23:52:36 8424 480 G230LB (CenWave=2375) o45a01010 G230LB 2375 08/03/97 0:31:02 7672 144 o45a03010 G230LB 2375 10/01/97 5:49:13 7672 144 o45a04010 G230LB 2375 12/01/97 1:43:39 7672 144 o45a05010 G230LB 2375 02/04/98 7:59:36 7672 144 o45a06010 G230LB 2375 04/01/98 11:00:23 7672 144 o45a12010 G230LB 2375 05/31/98 16:13:52 7672 172 o45a13010 G230LB 2375 07/31/98 21:27:24 7672 172 o45a14010 G230LB 2375 10/12/98 12:29:56 7672 172 o45a15010 G230LB 2375 12/06/98 22:46:56 7672 172 o45a16010 G230LB 2375 02/07/99 0:36:26 7672 172 o45a17010 G230LB 2375 04/03/99 15:30:29 7672 172 o45a18010 G230LB 2375 06/03/99 13:29:36 7672 172 o5ig01010 G230LB 2375 08/01/99 1:16:42 8418 172 o5ig02010 G230LB 2375 10/03/99 2:14:59 8418 172 G230MB (CenWave=1995) o45a02010 G230MB 1995 09/01/97 11:56:05 7672 240 o45a07010 G230MB 1995 01/01/98 8:37:29 7672 240 o45a08010 G230MB 1995 04/18/98 0:49:06 7672 240 o45a09010 G230MB 1995 08/24/98 15:39:43 7672 240 o45a10010 G230MB 1995 01/13/99 6:23:58 7672 240 o45a11010 G230MB 1995 05/01/99 2:06:28 7672 240 o5ig08010 G230MB 1995 09/11/99 13:31:28 8418 240 G230MB (CenWave=2416) o45a02020 G230MB 2416 09/01/97 12:06:20 7672 180 o45a07020 G230MB 2416 01/01/98 8:47:44 7672 180 o45a08020 G230MB 2416 04/18/98 0:59:21 7672 180 o45a09020 G230MB 2416 08/24/98 15:49:58 7672 180 o45a10020 G230MB 2416 01/13/99 6:34:19 7672 180 o45a11020 G230MB 2416 05/01/99 2:16:49 7672 180 o5ig08020 G230MB 2416 09/11/99 13:41:49 8418 180 26 Instrument Science Report STIS 2000-03 Observation Mode CenWave Date Time (UT) Propid Exptime (sec) G430L (CenWave=4300) o45a01020 G430L 4300 08/03/97 0:39:41 7672 144 o45a03020 G430L 4300 10/01/97 5:57:52 7672 144 o45a04020 G430L 4300 12/01/97 1:52:18 7672 144 o45a05020 G430L 4300 02/04/98 8:08:15 7672 144 o45a06020 G430L 4300 04/01/98 11:09:02 7672 144 o45a12020 G430L 4300 05/31/98 16:23:01 7672 172 o45a13020 G430L 4300 07/31/98 21:36:33 7672 172 o45a14020 G430L 4300 10/12/98 12:39:05 7672 172 o45a15020 G430L 4300 12/06/98 22:56:11 7672 172 o45a16020 G430L 4300 02/07/99 0:45:41 7672 172 o45a17020 G430L 4300 04/03/99 15:39:44 7672 172 o45a18020 G430L 4300 06/03/99 13:38:51 7672 172 o5ig01020 G430L 4300 08/01/99 1:25:57 8418 172 o5ig02020 G430L 4300 10/03/99 2:24:14 8418 172 G430M (CenWave=3165) o45a02030 G430M 3165 09/01/97 12:16:05 7672 120 o45a07030 G430M 3165 01/01/98 8:57:29 7672 120 o45a08050 G430M 3165 04/18/98 1:20:29 7672 120 o45a09050 G430M 3165 08/24/98 16:11:06 7672 120 o45a10050 G430M 3165 01/13/99 6:55:39 7672 120 o45a11050 G430M 3165 05/01/99 2:38:09 7672 120 o5ig08030 G430M 3165 09/11/99 13:51:40 8418 120 G430M (CenWave=4194, GAIN=1) o45a02040 G430M 4194 09/01/97 12:24:00 7672 120 o45a07040 G430M 4194 01/01/98 9:05:24 7672 120 o45a08040 G430M 4194 04/18/98 1:12:34 7672 120 o45a09040 G430M 4194 08/24/98 16:03:11 7672 120 o45a10040 G430M 4194 01/13/99 6:47:38 7672 120 o45a11040 G430M 4194 05/01/99 2:30:08 7672 120 o5ig08040 G430M 4194 09/11/99 13:59:41 8418 120 27 Instrument Science Report STIS 2000-03 Observation Mode CenWave Date Time (UT) Propid Exptime (sec) G430M (CenWave=4194, GAIN=4) o45a08030 G430M 4194 04/18/98 1:09:06 7672 120 o45a09030 G430M 4194 08/24/98 15:59:43 7672 120 o45a10030 G430M 4194 01/13/99 6:44:10 7672 120 o45a11030 G430M 4194 05/01/99 2:26:40 7672 120 o5ig08050 G430M 4194 09/11/99 14:03:09 8418 120 G750L (CenWave=7751) o45a01030 G750L 7751 08/03/97 0:48:20 7672 360 o45a03030 G750L 7751 10/01/97 6:06:31 7672 360 o45a04030 G750L 7751 12/01/97 2:00:57 7672 360 o45a05030 G750L 7751 02/04/98 8:16:54 7672 360 o45a06030 G750L 7751 04/01/98 11:17:41 7672 360 o45a12030 G750L 7751 05/31/98 16:32:10 7672 432 o45a13030 G750L 7751 07/31/98 21:45:42 7672 432 o45a14030 G750L 7751 10/12/98 12:48:14 7672 432 o45a15030 G750L 7751 12/06/98 23:05:26 7672 432 o45a16030 G750L 7751 02/07/99 0:54:56 7672 432 o45a17030 G750L 7751 04/03/99 15:48:59 7672 432 o45a18030 G750L 7751 06/03/99 13:48:06 7672 432 o5ig01030 G750L 7751 08/01/99 1:35:12 8418 432 o5ig02030 G750L 7751 10/03/99 2:33:29 8418 432 G750M (CenWave=7283) o45a02050 G750M 7283 09/01/97 12:32:45 7672 120 o45a07050 G750M 7283 01/01/98 9:14:09 7672 120 o45a08060 G750M 7283 04/18/98 1:29:14 7672 120 o45a09060 G750M 7283 08/24/98 16:19:51 7672 120 o45a10060 G750M 7283 01/13/99 7:04:30 7672 120 o45a11060 G750M 7283 05/01/99 2:47:00 7672 120 o5ig08060 G750M 7283 09/11/99 14:12:00 8418 120 * Data not usable due to on-orbit problems. ** o5jj99030 taken at +3" position. 28