Gas Monitoring - Pheonix India

advertisement

Gas Monitoring

Presented by: Dr. Meenal Aggarwal

Moderator: Dr. Dara Negi

Definitions

• Delay time: Time to achieve 10% of a step change in reading at the gas monitor

• Rise time/response time: Time required for a change from 10% to

90% of the total change in a gas value

• Total system response time: DT + RT

Monitor types

• 2 types: sidestream (diverting) or mainstream (nondiverting)

Mainstream:

• Sensor located directly in the gas stream (only for oxygen and CO

2

)

• Carbon dioxide: Using infrared technology with the sensor located between the breathing system and the patient

• Also available for the non-intubated patient: sensor attaches to a disposable oral and nasal adaptor

• Oxygen sensor: Uses electrochemical technology

 Usually placed in the breathing system inspiratory limb.

 Can measure both inspired and exhaled oxygen

Mainstream Infrared

CO 2 Analyser

• Advantages:

 Fast response times, no delay time (waveform has better fidelity)

 No gas is removed from the breathing system, so not necessary to scavenge these devices or to increase the fresh gas flow

 Water and secretions rarely cause a problem (although secretions on the windows of the cuvette can cause erroneous readings: problem with CO

2 sensor)

 Sample contamination by fresh gas is less likely

 Standard gas is not required for calibration (Oxygen sensor: calibrated by using room air)

 Use fewer disposable items than diverting monitors

• Disadvantages:

 Adaptor placed near the patient: cause traction/ weight

 Dead space

 Leaks, disconnections, circuit obstructions

 Condensed water, secretions, blood on the windows of the cuvette interferes with light transmission

 Sensor may become dislodged from the cuvette

 Expensive, vulnerable to damage

 Available only for oxygen and CO

2

 CO

2 sensor must be cleaned and disinfected between uses (potential for cross contamination), disposable become expensive

 Prolonged contact of the CO

2 sensor assembly with the patient could cause pressure injury

Sidestream:

• Aspirates gas from the sampling site, sensor (located in the main unit)

• Sampling tube: short- decreases delay time and more satisfactory waveforms

• Usually zeroed using room air and calibrated using a gas of known composition

• Traps, filters and hydrophobic membranes

• Water droplets and secretions may increase resistance/ affect the accuracy

• Clear or purge the sample line, or tube may be changed

• Accuracy at 20-40 bpm and short length, > 40 bpm decreases accuracy

• Sampling flow rate: less than 150 mL/minute should not be used.

Elevated baseline, falsely low peak readings on lower flows

Devices used:

• Face mask: Large dead space relative to tidal volume, difficult to obtain accurate end-tidal values

 can be attached to upper lip or placed in patient's nares or the lumen of an oral or nasopharyngeal airway under the mask

 With a breathing system, most often attached to a component between the mask and the breathing system

• Tracheal Tube: Sampling site must be between the patient and the breathing system (measurement for both exp and insp)

 Sampling site should be away from the fresh gas port (in mapleson circuits can l/t errors)

 Tracheal tubes that incorporate a sampling lumen that extends to the middle or patient end of the tube are available

 Tracheal tube connectors with an attachment or hole for a sampling tube are available or can be created

• Supraglottic devices: sampling tube can be inserted through the connector (preferred site is the distal end of the shaft)

 may be inserted into a nasal airway

• Oxygen Supplementation Devices:

 OxyArm: allow simultaneous administration of oxygen and carbon dioxide monitoring

 In both nose and mouth breathers

 A nasal cannula can be modified to accept a sampling tubing (may l/t choking hazard)

 Sampling tube may be connected to mask outlet, inserted through a vent hole or a slit in the mask

 Accuracy affected by: Mouth breathing, airway obstruction, and oxygen delivery through the ipsilateral nasal cannula

OxyArm

• Jet Ventilation: an injector incorporating a sampling lumen or a sampling tube placed in the airway may be used

 Ventilatory frequency may need to be lowered to measure the endtidal CO

2

• Other ways:

 Sampling line can be placed in front of or inside the patient's nostril

 In mouth breather: in the nasopharynx or hypopharynx

 Catheter placed in the trachea after extubation for CO

2 monitoring

 Bite block can be modified to accommodate a sampling line

 Sampling line can be placed over a tracheostomy stoma

• Advantages of Diverting devices:

 Calibration and zeroing usually automatic (Occasional calibration is necessary, usually easy)

 Added dead space is minimal.

 Potential for cross contamination between patients low.

 Several gases can be measured simultaneously, allows automatic correction for nitrous oxide and oxygen.

 Sampling port can be used to administer bronchodilators

 These devices can be used when the monitor must be remote from the patient (e.g., during MRI)

• Disadvantages of Diverting devices:

 Leaks, sampling tube obstruction, or failure of the aspirator pump or can kink (use elbow connector)

 Sampling line can be connected to the wrong place

 Leak in sampling line can l/t mixing with air and so dilution of sample

 Aspirated gases must be either routed to the scavenging system (need to inc fresh gas flow) or returned to the breathing system

 Some delay time

 Supply of calibration gas

 Disposable items (adaptors and catheters) used

 Fresh gas dilution

 More variable differences between arterial and end-tidal CO

2 level

Technologies

• Infrared Analysis

 Black body Radiation technology

 Microstream technology

• Paramagnetic oxygen analysis

• Electrochemical oxygen analysis

 Galvanic cell

 Polarographic electrode

• Peizoelectric Analysis

• Refractometry

Infrared Analysis:

• Most common technology in use

• Principle: Gases with two or more dissimilar atoms in the molecule

(nitrous oxide, CO

2

, and the halogenated agents) have specific and unique infrared light absorption spectra.

• Amount of infrared light absorbed is proportional to the concentration of the absorbing molecules, the concentration can be determined

• Nonpolar molecules cannot be measured

• 2 technologies available:

 Black body radiation

 Microstream technology

Blackbody Radiation Technology:

• Utilizes a heated element called a blackbody emitter as the source of infrared light, produces a broad infrared spectrum (redundant radiation to be removed using filters)

• Optical detectors must be calibrated to recognize only infrared radiation that is modulated at a certain frequency by using a spinning chopper wheel

• Analyzer selects the appropriate infrared wavelength, minimize absorption by other gases that could interfere with measurement of the desired component

• Then an electrical signal is produced and amplified, and the concentration is displayed

• For halogenated agents: separate chamber to measure absorption at several wavelengths (single-channel, four-wavelength infrared filter photometers) have filter for each anesthetic agent and one to provide a baseline for comparison

Diverting type:

• Gas to be measured is pumped continuously through a measuring chamber

• Filtered and pulsed light is passed through the sample chamber and also through a reference chamber (has no absorption characteristics)

• Light is focused on an infrared photosensor (partly absorbed by the sample gas acc to conc)

• Changing light levels on the photosensor produce changes in the electrical current running through it

• Provides hundreds of readings for each respiratory cycle (Continuous waveform produced)

• Monochromatic analyzers use one wavelength to measure potent inhalational agents (unable to distinguish between agents)

• Polychromatic analyzers use multiple wavelengths to both identify and quantify the various agents

• Measuring cell is calibrated to zero (using gas that is free of the gases of interest, usually room air) and to a standard level (using a calibration gas mixture)

Non Diverting Type:

• Gas stream passes through a chamber (cuvette) with two windows, placed b/w the breathing system and the patient

• Sensor (has both the light source and detector) fits over the cuvette

• Sensor is heated slightly above body temperature (to prevent condensation)

• Infrared light passes through window on one side of the adaptor, sensor receives the light on the opposite side

• Then light goes through three ports in a rotating wheel, containing

(a) a sealed cell with a known high CO

2 concentration

(b) a chamber vented to the sensor's internal atmosphere

(c) a sealed cell containing only nitrogen

• Then passes through a filter (to isolate CO

2 information)

• Signal amplified and sent to the display module

• Calibration done using: low calibration cell contains 100% nitrogen, high cell contains a known partial pressure of CO

2

• Corrections for nitrous oxide and oxygen entered manually

Infrared Mainstream Analyser

Microstream Technology:

• Uses laser-based technology to generate infrared rays that match the absorption spectrum of CO

2

• Smaller sample cell, low flow rate

• Emission source: Glass discharge lamp coupled with an infrared transmitting window

• Electrons (generated by a radio frequency voltage) excite nitrogen molecules  Carbon dioxide molecules are excited by collision with the excited nitrogen molecules  These drop back to their ground state and emit the signature wavelength of CO

2

• This emission now passes through main optical detector and reference detector (compensates for changes in infrared output)

• Measurements made every 25 msec

• As low sample flow and small sample cell, useful for measuring:

 CO

2 in very small patients

 high respiratory rates

 low-flow applications

 unintubated patients

• Readings not affected by high concentrations of oxygen or anesthetic gases

Advantages of Infrared Analysis:

• Multigas Capability

• Volatile Agent Detection

• Small, compact, lightweight

• Quick response times (faster for CO

2

)

• Short warm-up time

• Convenience (no complicated calibrations)

• Lack of interference from other gases (argon, low conc NO)

• Detecting anaesthetic agent breakdown (desflurane to CO will show as wrong or mixed agent)

Disadvantages of Infrared Analysis:

• O

2 and N

2 cannot be measured

• Gas interference :

 O

2 causes broadening of CO

2 spectrum l/t lower readings

 N

2

O absorption spectrum overlaps with CO

2

(l/t higher vlues): so need either automatic or manual correction for N

2

O

 He l/t underestimation of CO

2

• Other substances l/t inaccuracies (ethanol, methanol, diethyl ether, methane): give high volatile agent reading, polychromatic less affected

• Water vapors: Absorb infrared rays (l/t lower values), use water traps, hydrophobic membranes

• Slow response time (with rapid resp rates)

• Difficulty in adding new volatile agents

Paramagnetic Oxygen Analysis:

• Paramagnetic substances: Substances which locate themselves in the strongest portion of the field when introduced into a magnetic field

• Oxygen is paramagnetic

• Principle: When a gas that contains oxygen is passed through a switched magnetic field, the gas will expand and contract, causing a pressure wave that is proportional to the oxygen partial pressure

• Pressure difference is detected by the transducer and converted into an electrical signal that is displayed as oxygen partial pressure or volumes percent.

• Short rise time so both inspired and end-tidal oxygen levels can be measured

• Desflurane disturbs the paramagnetic oxygen sensor and it reads higher than expected

Eletrochemical Oxygen Analysis:

• Consists of a sensor, analyzer box, display, and alarms

• Sensor: A cathode and an anode surrounded by electrolyte

• Sensor is placed in the inspiratory limb

• Gel held in place by a membrane (nonpermeable to ions, proteins, but is permeable to oxygen)

• Older ones respond slowly to changes in oxygen pressure, so cannot be used to measure end-tidal concentrations (not so with new analyzers)

• Technology:

 Galvanic cell/ fuel cell

 Polarographic electrode

Galvanic cell:

• Principle: Oxygen diffuses through the sensor membrane and electrolyte to the cathode, where it is reduced, causing a current to flow

• Current generated is proportional to the partial pressure of oxygen in the gas

Cathode: O

2

+ 2H

2

O + 4e → 4OH -

Anode: 4OH + 2Pb → 2PbO + 2H

2

O + 4e -

• Cathode is the sensing electrode, anode is usually consumed

• The current is strong enough to operate the meter so a separate power source is not required to operate the analyzer

.

• The chemical reaction is temperature dependent (a thermistor may be connected in parallel with the sensor.)

Galvanic cell

Fuel cell Oxygen Analyzer

• Sensor comes packaged in a sealed container that does not contain oxygen

• Its useful life is cited in percent hours: the product of hours of exposure and oxygen percentage

• Sensor life can be prolonged by removing it from the breathing system and exposing it to air when not in use

• Whole sensor cartridge must be replaced when it becomes exhausted

Polarographic Electrode:

• Components: anode, a cathode, an electrolyte, and a gas-permeable membrane

• Needs power source for inducing a potential between the anode and the cathode

• Same principle as galvanic cell

• May be either preassembled disposable cartridges or units that can be disassembled and reused by changing the membrane and/or electrolyte

Advantages:

 Easy to use, low cost, compact

 No effect of argon

Disadvantages:

 Maintenance (more in polarographic)

 Need to be calibrated every day (every 8 hrs)

 Slower response time

Peizoelectric Analysis:

• Uses vibrating crystals that are coated with a layer of lipid to measure volatile anesthetic agents

• Principle: When exposed to a volatile anesthetic agent, the vapor is adsorbed into the lipid  resulting change in the mass of the lipid alters the vibration frequency

• By using an electronic system consisting of two oscillating circuits, one has an uncoated (reference) crystal and the other a coated (detector) crystal, an electric signal that is proportional to the vapour concentration is generated

• Diverting devices

Advantages:

 Accuracy

 Fast response time

 No need for scavenging

 Short warm up time

 Compact

Disadvantages:

 Only one gas measured

 No agent discrimination

 Inaccuracy with water vapours

Chemical Carbon Dioxide Detection

• Consists of a pH-sensitive indicator

• Principle: When the indicator is exposed to carbonic acid that is formed as a product of the reaction between CO

2 and water it becomes more acidic and changes color

Technology:

 Hygroscopic

 Hydrophobic

Uses:

• For confirming tracheal intubation when a capnometer is not available

• Disposable so it may be useful to confirm tracheal intubation in patients with respiratory diseases (e.g.SARS)

Advantages:

• Easy to use, small size, low cost

• Not affected by N

2

O, volatile anaesthetics

• Offers minimal resistance to flow

• CO doesn’t interfere

Disadvantages:

• Recommended to wait six breaths before making a determination

• False-negative results may be seen with very low tidal volumes

• Drugs instilled in the trachea or gastric contents can cause irreversible damage to the device

• False-positive results can occur if CO

2 in the stomach

• Semiquantitative, cannot give accurate measurement of CO

2

(So use limited to check endotracheal intubation)

Refractometry:

• Optical interference refractometer (interferometer): Light beam passes through a chamber into which the sample gas has been aspirated, also passes through an identical chamber containing air.

• Vapour slows the velocity of light, the portion passing through the vapor chamber is delayed

• Beams form dark and light bands, position of these bands yields the vapor concentration

• Used primarily for vaporizer calibration

• Sensitive to nitrous oxide (so cannot be used to measure halogenated agents in a O

2

, N

2

O, agent mixture)

Gas Measurement

Oxygen:

• Standard requirements:

 Oxygen readings shall be within ±2.5% of the actual level (min for 6hrs together)

 The high and low oxygen level alarms must be at least medium priority, oxygen concentration below 18% (should be high priority alarm)

 It shall not be possible to set the low oxygen alarm limit below 18%

• Technology used:

 Electrochemical Technology

 Paramagnetic Technology: Rapid response time, even for nonintubated

Applications of Oxygen Analysis:

Detecting Hypoxic or Hyperoxic Mixtures:

 Oxygen monitor provides earlier warning of inadequate oxygen than pulse oximetry

 Problems resulting from hyperoxygenation: patient movement during surgery, awareness, damage to the lungs and eyes, fires

Detecting Disconnections and Leaks:

 However not dependable

Detecting Hypoventilation:

 Normal: Difference b/w inspired and expired oxygen is 4% to 5%

End tidal Oxygen Measurement:

 Assess pt’s oxygen consumption (Malignant hyperthermia)

 To detect air embolism (inc ET O

2

)

Carbon Dioxide Analysis:

• Means for assessing metabolism, circulation, and ventilation

• ASA guidelines: Correct positioning of ET tube must be verified by identifying CO

2 in the expired gas

• Capnometry: Measurement of CO

2 in gas mixture

• Capnography: Recording of CO

2

Conc versus time

Standard requirements of Capnometer:

• CO

2 reading shall be within ±12% of the actual value or ±4 mm Hg

• Must have a high CO

2 alarm for both inspired and exhaled CO

2

Technology:

• Infrared Analysis

• Chemical colorimetric analysis

Clinical Significance of Capnometry:

• Metabolism

• Respiration

• Circulation

• Equipment Function

• Confirming endotracheal and enteric tube placement

• Dec ET CO

2

:

 Impaired peripheral circulation

 Impaired lung circulation (Pulmn embolus)

 Increased patient dead space

 Hyperventilation

 Hypothermia

 Increased depth of anaesthesia

 Use of muscle relaxants

 Leak in sampling line

 Leak around ET

• Increased ET CO

2

:

 Absorption of CO2 from peritoneal cavity

 Injection of NaHCO

3

 Convulsions

 Hyperthermia

 Pain, anxiety, shivering

 Increased muscle tone (reversal of muscle relaxation)

 Hyperventilation

 Upper airway obstruction

 Rebreathing

 Increased circulation from tissues to lung (release of tourniquet)

• Absent waveform:

 Esophageal intubation

 Disconnection

 Apnea

 Blockade of sampling line

Correlation between Arterial and End-tidal Carbon dioxide levels

• Normal: PaCO

2

– ET CO

2

= 2-5 torr

• Altered with:

 Reduced FRC (Pregnancy, Obese pt)

 Rebreathing

 Neurosurgical procedures

 During one lung ventilation

(In these cases transcutaneous CO

2 monitoring more accurate)

Capnography

• Examined for:

 Height (Depends on ETCO

2

)

 Frequency (R.R.)

 Rhythm

 Baseline (normally zero)

 Shape (Top hat or Sine wave is normal)

Capnography cont…

• Phase 1: E (Inspiratory baseline)

• Phase 2: B to C (Expiratory upstroke), S shaped- represents transition from dead space to alveolar space

• Phase 3: C to D (all from alveoli)

• End of Phase 3 (Point D): End tidal point (Max CO

2

)

• Alpha : Angle b/w Phase 2 & 3 (normal 100-110 degree)

• Beta: B/w end of phase 3 & Descending limb (90 degree)

• The slope of phase 3 (C to D) increases:

 With PEEP

 Airway obstruction

 V/Q mismatch

And so angle Alpha also increases

And angle Beta decreases

• Angle Beta increases with:

 Rebreathing

 Prolonged response time

Unusual waveforms:

• Leak in sample line: Brief peak at the end of plateau

• Partially paralysed (making intermittent resp effort) :Curare cleft

• Cardiogenic occilations:

Seen in pediatric pts

(d/t heart beating against

Lungs)

Volatile Anaesthetic Agents:

• Standard Requirements:

 Difference in value shall be within ) 0.2% vol%

 High conc alarm is mandatory, low conc alarm is optional

• Measurement technique:

 Infrared Analysis

 Refractometry

 Peizoelectric Analysis

• Significance:

 Assess vaporizer function and contents

 Information of patient uptake of the agent (insp & exp conc)

 Information on Anaesthetic depth

 Detecting contaminants/ disconnection

Nitrous Oxide:

• Measurement technique:

 Infrared Analysis

• Significance:

 Assess flowmeter function

Nitrogen:

• Previously measured using Raman spectroscopy or mass spectrometry

• Now no longer available

• Significance:

 Verifying adequate denitrogenation before induction (imp for pediatric pt, in lung ds, dec FRC)

 Detecting air emboli (Inc ET N2)

Thank You

Download