ELEG 479 Lecture #12 Magnetic Resonance (MR) Imaging Mark Mirotznik, Ph.D. Associate Professor The University of Delaware Physics of Magnetic Resonance Summary Protons and electrons have a property called spin that results in them looking like tiny magnets. In the absence of an external magnetic field all the magnetic dipole are oriented randomly so we get zero net magnetic field when we add them all up. = No Net Magnetization Random Orientation Physics of Magnetic Resonance Summary When we add a large external magnetic field we can get the protons to line up in 1 of 2 orientations (spin up or spin down) with a few more per million in one of the orientations than the other. This produces a net magnetization along the axis of the applied magnetic field. Bo 2 2 Mo PD 4kT When we add a large external magnetic field we cause a torque on the already spinning proton that causes it to precess like a top around the applied magnetic field. The frequency is precesses , called its Larmor frequency., is determined from Larmor’s equation. Physics of Magnetic Resonance Summary from Last Lecture The net magnetization vector is the sum of all of these little magnetic moments added together. This is what we measure. z z z z z z xy y z x z Net Magnetization Vector xy xy y y x y x z z xy xy x y x N M (t ) n ( xn , yn , zn , t ) n 1 M (t ) M xy (t ) M z zˆ Physics of Magnetic Resonance Summary from Last Lecture However since all the spinning protons are precessing out of phase with each other this results in zero net magnetization in the transverse plane. =0 M (t ) M xy (t ) M z zˆ This is bad news since M xy (t ) is where are signal comes from! Somehow we need to get these guys spinning together! Physics of Magnetic Resonance Summary from Last Lecture To get them to all spin together we add a RF field whose frequency is the same as the Larmor resonant frequency of the proton and is oriented in the xy or transverse plane. B1 RF Excitation time B1 Physics of Magnetic Resonance Summary from Last Lecture z B1 M a x M xy y a B1Dt Tip Angle Amplitude of RF Pulse Time of Application of RF Pulse Physics of Magnetic Resonance Summary from Last Lecture z e 1 B (t ) M a M xy y x e 1 B (t ) = envelope of In general a B (t )dt 0 e 1 the RF signal Physics of Magnetic Resonance Summary from Last Lecture To get the signal out we place a coil near the sample. A timevarying transverse magnetic field will produce a voltage on the coil that can be digitized and stored for processing. d V (t ) M xy (t ) dt V (t ) K o M o sin( a ) recall Bo 2 2 Mo PD 4kT and o Bo Relaxation Processes Physics of Magnetic Resonance Relaxation After the RF field is removed over time the spin system will return back to it’s equilibrium state due to several relaxation processes. Physics of Magnetic Resonance Relaxation After the RF field is removed over time the spin system will return back to it’s equilibrium state due to several relaxation processes. These are: (1) Spin-Spin relaxation (also called the T2 relaxation): Due to random processes in which neighboring proton spins effect each other spin system will lose coherence and Mxy will decay. This is an irreversible process. (2) Spin-Lattice relaxation (also called T1 relaxation): Due to another random process the Mz will begin to recover back to it’s original equilibrium state. Also irreversible. (3) T2* relaxation: Due to inhomogenities in the external Bo field Mxy will decay much faster than T2. This is a reversible process. T1 Relaxation T2 Relaxation (FID) T2 Decay RF RF * T2 Relaxation T2* Decay: Dephasing due to field inhomogeneity z' Mxy y' =0 x' T2* relaxation is dephasing of transverse magnetization too but it turns out to be reversible Animation of T2* Dephasing Spin- Echo Spin Echo Summary of Relaxation Processes MRI Image An MRI image is determined by two things three intrinsic properties of the tissue. These are: T1, T2 and Pd. (two relaxation time constants and the density of protons) the details of the external magnetic fields (Bo, B1 and the gradient magnetics (have not talked about these yet)). How they are configured and how we turn them on and off (pulse sequence) effects what the image looks like. By varying the pulse sequence we can control which of the intrinsic properties to emphasize in the image. Tissue Contrast TR 180 degree RF pulses 0.5TE 90 degree RF pulses 0.5TE 0.5TE 0.5TE TE TE White matter T1=813 ms T2=101 ms TR 180 degree RF pulses CASE I: TR>>T1 , TE<T2 What do we measure? 90 degree RF pulses TE TE White matter T1=813 ms T2=101 ms TR 180 degree RF pulses 90 degree RF pulses CASE I: TR>>T1 , TE<T2 Short TE means that the signal has not decayed much due to T2relaxation. Long TR means that by the next pulse the system is back at equilibrium (Mz due to T1 relaxation has fully recovered) So what are we measuring? TE TE White matter T1=813 ms T2=101 ms TR 180 degree RF pulses 90 degree RF pulses TE CASE I: TR>>T1 , TE<T2 Short TE means that the signal has not decayed much due to T2relaxation. Long TR means that by the next pulse the system is back at equilibrium (Mz due to T1 relaxation has fully recovered) So what are we measuring? PD Weighted imaging! White matter T1=813 ms T2=101 ms TE TR 180 degree RF pulses CASE II: TR>>T1 , TE~T2 90 degree RF pulses TE TE White matter T1=813 ms T2=101 ms TR 180 degree RF pulses 90 degree RF pulses CASE II: TR>>T1 , TE~T2 TE on the order of T2means that the signal is proportional to the T2relaxation constant. Long TR means that by the next pulse the system is back at equilibrium (Mz due to T1 relaxation has fully recovered) So what are we measuring? TE TE White matter T1=813 ms T2=101 ms TR 180 degree RF pulses 90 degree RF pulses CASE II: TR>>T1 , TE~T2 TE on the order of T2means that the signal is proportional to the T2relaxation constant. Long TR means that by the next pulse the system is back at equilibrium (Mz due to T1 relaxation has fully recovered) So what are we measuring? T2 Weighted imaging! TE TE White matter T1=813 ms T2=101 ms TR 180 degree RF pulses 90 degree RF pulses TE CASE III: TR~T1 , TE<T2 What do we measure? TR 180 degree RF pulses 90 degree RF pulses TE CASE III: TR~T1 , TE<T2 TE is shorter than T2means that the signal is not heavily weighted on the T2relaxation constant. TR on the order of T1 means that by the next pulse the system is not back at equilibrium (Mz due to T1 relaxation has not fully recovered) So what are we measuring? TR 180 degree RF pulses 90 degree RF pulses CASE III: TR~T1 , TE<T2 TE is shorter than T2means that the signal is not heavily weighted on the T2relaxation constant. TR on the order of T1 means that by the next pulse the system is not back at equilibrium (Mz due to T1 relaxation has not fully recovered) So what are we measuring? T1 Weighted imaging! TE Tissue Contrast Summary TE PD weighted T1weighted T2 weighted TR TE<T2 (short TE) TR>>T1 (long TR) TE<T2 (short TE) TR~T1 (short TR) TE~T2 (long TE) TR>>T1 (long TR) Bloch Equations Full Bloch equation including relaxation ( M x (t) x̂ M y (t) ŷ) ( M z (t ) M o )ẑ dM M (t ) γB(t) dt T2 T1 precession, RF excitation transverse magnetization B (t) Bo B1 (t ) includes Bo and B1 longitudinal magnetization Example: Solve for the transverse components of M after a 90 degree pulse. ( M x (t) x̂ M y (t) ŷ) ( M z (t ) M o )ẑ dM M (t ) γB(t) dt T2 T1 1 M x (t) M y (t ) Bo M z (t ) B y (t ) T2 M x (t ) d 1 M ( t ) γ M ( t ) B M ( t ) B ( t ) M (t) y x o z x y T dt 2 M x (t ) B y (t ) M y (t ) Bx (t ) M z (t ) 1 M z (t) - M o T1 Example: Solve for the transverse components of M after a 90 degree pulse. 1 M x (t) M y (t ) Bo M z (t ) B y (t ) T2 M x (t ) d 1 M ( t ) γ M ( t ) B M ( t ) B ( t ) M (t) y x o z x y T dt 2 M x (t ) B y (t ) M y (t ) Bx (t ) M z (t ) 1 M z (t) - M o T1 After 90 degree pulse the RF field is shut down and only Bo is non-zero 1 M x (t) M x (t ) M y (t ) B0 T2 d γ - M (t ) B 1 M (t) M ( t ) y 0 x T2 y dt M z (t ) 0 1 M z (t) - M o T1 Example: Solve for the transverse components of M after a 90 degree pulse. After 90 degree pulse the RF field is shut down and only Bo is non-zero 1 M x (t) M x (t ) M y (t ) B0 T2 d 1 M ( t ) γ M ( t ) B M (t) y 0 x T2 y dt M z (t ) 0 1 M z (t) - M o T1 Initial conditions for 90 degree pulse: M (0) 0 x M y ( 0) M o M z ( 0) 0 Example: Solve for the transverse components of M after a 90 degree pulse. After 90 degree pulse the RF field is shut down and only Bo is non-zero 1 M x (t) M x (t ) M y (t ) B0 T2 d 1 M ( t ) γ M ( t ) B M (t) y x 0 y T dt 2 M z (t ) 0 1 M z (t) - M o T1 M x (t ) M o sin( ot )e Solutions t M y (t ) M o cos(o t )e M z (t ) M o (1 e t T1 ) T2 t T2 Example: Solve for the transverse components of M after a 90 degree pulse. After 90 degree pulse the RF field is shut down and only Bo is non-zero 1 M x (t) M x (t ) M y (t ) B0 T2 d 1 M ( t ) γ M ( t ) B M (t) y x 0 y T dt 2 M z (t ) 0 1 M z (t) - M o T1 M xy (t ) M x (t ) jM y (t ) Solutions M xy (t ) M o e j o t M z (t ) M o (1 e e t t T1 T2 ) Example: Solve for the transverse components of M after an arbitrary flip angle (a) 1 M x (t) M y (t ) Bz M z (t ) B y (t ) T2 M x (t ) d 1 M ( t ) γ M ( t ) B M ( t ) B ( t ) M (t) y x z z x y T dt 2 M x (t ) B y (t ) M y (t ) Bx (t ) M z (t ) 1 M z (t) - M o T1 After an arbitrary RF pulse the RF field is shut down and only Bo is non-zero 1 M x (t) M x (t ) M y (t ) B0 T2 d γ - M (t ) B 1 M (t) M ( t ) y 0 x T2 y dt M z (t ) 0 1 M z (t) - M o T1 Example: Solve for the transverse components of M after an arbitrary flip angle (a) After an arbitrary RF pulse the RF field is shut down and only Bo is non-zero 1 M x (t) M x (t ) M y (t ) B0 T2 d γ - M (t ) B 1 M (t) M ( t ) y 0 x T2 y dt M z (t ) 0 1 M z (t) - M o T1 Initial conditions for 90 degree pulse: M x (0) 0 M y (0) M o sin( a ) M z (0) M o cos(a ) Example: Solve for the transverse components of M after an arbitrary flip angle (a) 1 M x (t) M x (t ) M y (t ) B0 T2 d 1 M ( t ) γ M ( t ) B M (t) y x 0 y T dt 2 M z (t ) 0 1 M z (t) - M o T1 M xy M x jM y M xy (t ) M xy (0 )e jBot e t T2 Solutions M o sin( a )e jot e t T2 M z (t ) M o (1 e t M o (1 e T1 t ) M z (0)e T1 t T1 ) M o cos(a )e t T1 Solve full Bloch equation with only B=Bo Solution for transverse components Mx and My jBo t t T2 jo t t T2 M xy (t ) M xy (0 )e M xy (0 )e e e M xy (0 ) M z (0 ) sin( a ) M z (t ) M o (1 e t T1 ) M o cos(a )e t T1 Where a is the flip angle after RF excitation Signal Detection Signal Detection via RF coil Signal Detection via RF coil Transverse magnetization at t=0. s (t ) A z max z min y max xmax y min xmin M xy ( x, y, z ,0 )e jBot e t / T2 ( x , y , z ) dxdydz Coils oriented as shown above will only respond to changes in the transverse magnetic field (this is what we want) Assuming the magnetic fields are homogenous the signal will be a weighted integration of all the protons within the coil. The waiting will be based on the total magnetization at location x,y,z at the start of the pulse (Mxy(x,y,z,0)) and the tissue decay time T2(x,y,z) This is not an image!! Signal Detection via RF coil s (t ) A z max z min s (t ) A z max z min s (t ) e j o t y max xmax y min xmin y max xmax y min xmin M xy ( x, y, z ,0 )e jBot e t / T2 ( x , y , z ) dxdydz M xy ( x, y, z ,0 )e jot e t / T2 ( x , y , z ) dxdydz z max y max xmax A M xy ( x, y, z ,0 )e t / T2 ( x , y , z ) dxdydz z min y min xmin After demodulation: so (t ) A z max z min y max xmax y min xmin M xy ( x, y, z ,0 )e t / T2 ( x , y , z ) dxdydz Creating an Image Creating an Image To create an image using NMR we need to figure out a way to encode the proton spins spatially in three dimensions. But how? Frequency and Phase Are Our Friends in MR Imaging q= t q The spatial information of the proton pools contributing MR signal is determined by the spatial frequency and phase of their magnetization. Gradient Coils z z z y y x x X gradient y Y gradient x Z gradient Gradient coils generate spatially varying magnetic field so that spins at different location precess at frequencies unique to their location, allowing us to reconstruct 2D or 3D images. Gradient Coils Purpose: Spatially alter magnitude of B0 (not direction) Sounds generated during imaging due to mechanical stress within gradient coils. Vector Notation G Gx aˆ x G y aˆ y Gz aˆ z r x aˆ x y aˆ y z aˆ z Bz ( x, y, z ) Bo G r Larmor frequency within a gradient field G Gx aˆ x G y aˆ y Gz aˆ z r x aˆ x y aˆ y z aˆ z Bz ( x, y, z ) Bo G r (r ) Bo G r Slice Selection Slice Selection Gradient BG Coil 2 Coil 1 Helmholtz Coils Z-Gradient Fields ( z ) Bo Gz z ( z ) 0 Gz z By adding a z-gradient field we cause a variation in the resonant frequency from head to toe. Example A sample is put inside a 1.5T magnet. A zgradient of 3 gauss/cm is applied. If we wish to image a 2 ft in length section of a person what is the range of resonant frequencies we will encounter? Example A sample is put inside a 1.5T magnet. A z-gradient of 3 gauss/cm is applied. If we wish to image a 2 ft in length section of a person what is the range of resonant frequencies we will encounter? Bz ( x, y, z ) Bo G r Bo Gz z Bzmin Bzmax 3 1.5 z 10000 3 1.5 12 2.54 1.490856 T 10000 3 1.5 12 2.54 1.509144 T 10000 Example A sample is put inside a 1.5T magnet. A z-gradient of 3 gauss/cm is applied. If we wish to image a 2 ft in length section of a person what is the range of resonant frequencies we will encounter? 3 B 1.5 12 2.54 1.490856 T 10000 3 max Bz 1.5 12 2.54 1.509144 T 10000 ( z ) B( z ), 42.58 MHz / tesla min 42.58 1.490856 63.48 MHz min z max 42.58 1.509144 64.26 MHz A Field Gradient Makes the Larmor Frequency Depend upon Position 1.500 T 1.501 T B0 63,480,000 Hz 64,260,000 Hz Z B(Z) B o G Z * Z (z) B(z) Gradient in Z Slice Selection (-) 62 MHz 63 MHz 64 MHz G 65 MHz 66 MHz (+) Slice Selection How do we determine the slice width and center? z z-gradient Bo (slice center) z After z selection gradient and excitation Dz (slice width) x Determining slice thickness Resonance frequency range as the result of sliceselective gradient: ( z ) Bo Gz z min Bo max Bo z min , z max Gz Gz max min Bo Bo Dz z max z min Gz D Dz Gz Changing slice thickness There are two ways to do this: (a) Change the slope of the slice selection gradient (b) Change the bandwidth of the RF excitation pulse D Dz G z Both are used in practice, with (a) being more popular Example Suppose we wish to have a slice thickness of 2 mm and we are using a z-gradient of 1.0 G/cm ? What range of RF frequencies should we use? Example Suppose we wish to have a slice thickness of 2 mm and we are using a z-gradient of 1 G/cm ? What range of RF frequencies should we use? D Dz Gz D 0.2cm 1 42.58 10000 4 D 8.516 10 MHz 851.6 Hz Selecting different slices ( z ) Bo Gz z min Bo max Bo z min , z max Gz Gz z max z min max min 2Bo z 2 2Gz max min / 2 Bo o z Gz Gz o z G z Selecting different slices In theory, there are two ways to select different slices: (a) Change the position of the zero point of the slice selection gradient with respect to isocenter (b) Change the center frequency of the RF to correspond to a resonance frequency at the desired slice o z G z Option (b) is usually used as it is not easy to change the isocenter of a given gradient coil. RF Excitation (RF Pulse) Fo FT 0 Fo Fo+1/ t t Time Frequency Fo t Fo FT DF= 1/ t RF Excitation (RF Pulse) Fo t sin( D t ) j 2 t s(t ) AD e D t FT A D 1 2 v S ( ) A rect ( ) D v2 v1 v 2 RF Excitation: Flip angle Fo FT A D t 1 v S ( ) A rect ( ) D sin( D t ) j 2 t s(t ) AD e D t Envelope of the pulse a 2 t max e B 1 (t )dt t min v2 v1 v 2 RF Excitation: Flip angle Fo FT A D t 1 v S ( ) A rect ( ) D sin( D t ) j 2 t s(t ) AD e D t a v2 v1 v 2 t max t max t min t min e B 1 (t )dt 2 j 2 t s ( t ) e dt RF Excitation: Flip angle Fo FT A D t 1 2 v S ( ) A rect ( ) D sin( D t ) j 2 t s(t ) AD e D t v2 v1 v 2 sin( D t ) a B (t )dt AD dt D t t min t max e 1 v zz a A rect ( ) A rect ( ) D Dz RF Excitation: Flip angle (truncated sinc) p / 2 p /2 sin( D t ) j 2 t t ~ s (t ) AD e rect ( ) D t p p 2 a p sin( D t ) AD dt D t 2 zz a A p rect ( ) * sin c( p Gz ( z z )) Dz A potential problem Wait a minute! Dr. M I remember you telling us that if the magnetic field varied from place to place (inhomogeneous) then we would get rapid dephasing of spins. Since some spins are spinning faster than others they quickly get out of phase. That was the whole reason behind the spin echo stuff! Won’t that happen again? Wait a minute! Dr. M I remember you telling us that if the magnetic field varied from place to place (inhomogeneous) then we would get rapid dephasing of spins. Since some spins are spinning faster than others they quickly get out of phase. That was the whole reason behind the spin echo stuff! Won’t that happen again? Why yes it will! It is called gradient dephasing Good question! Spinning slow Spinning fast Why yes it will! It is called gradient dephasing. It will quickly kill our signal much faster than T2 or even T2* Any ideas on how to get around this? Spinning slow Spinning fast Localization in xy plane Lets Start with a Simple Flat Person (only xz plane) z Bo x Lets Start with a Simple Flat Person (only xz plane) z z-gradient After z selection gradient and excitation Bo x Lets Start with a Simple Flat Person Frequency Encoding Mathematical Analysis Recall that the signal we measure is given by: s (t ) A z max z min y max xmax y min xmin M xy ( x, y, z ,0 )e jBot e t / T2 ( x , y , z ) dxdydz Now we have selected only a single slice in z (z=zo) and we have no y dependence (flat person) s(t ) e j 2 o t A M xy ( x,0 )e t / T2 ( x ) dx After demodulation (envelope detection) so (t ) s (t ) e j 2 o t A M xy ( x,0 )e t / T2 ( x ) dx Lets Start with a Simple Flat Person Frequency Encoding Mathematical Analysis After demodulation (envelope detection) so (t ) s (t ) e Let j 2 o t A M xy ( x,0 )e t / T2 ( x ) dx f ( x) A M xy ( x,0 )et / T2 ( x) so (t ) f ( x)dx This is what we want to image (called the effective proton density) Lets Start with a Simple Flat Person (frequency encoding using x-gradient) z After z selection gradient and excitation Bo x Lets Start with a Simple Flat Person Frequency Encoding Mathematical Analysis Now lets apply a gradient in the x direction (Gx) s(t ) A M xy ( x,0 )e j 2 o Gx x t e t / T2 ( x ) dx s(t ) e j 2 o t A M xy ( x,0 )e j 2Gx x t e t / T2 ( x ) dx After demodulation (envelope detection) s o (t ) A M xy ( x,0 )e j 2Gx x t e t / T2 ( x ) dx f ( x) A M xy ( x,0 )et / T2 ( x) s o (t ) f ( x)e j 2Gx x t dx What does this look like? Lets Start with a Simple Flat Person Frequency Encoding Mathematical Analysis After demodulation (envelope detection) s o (t ) f ( x)e j 2Gx x t dx Let u Gxt u F (u ) s o ( ) f ( x)e j 2ux dx Gx The received signal is related to the Fourier transform (THIS IS THE KEY!) Lets Start with a Simple Flat Person Frequency Encoding Mathematical Analysis After demodulation (envelope detection) s o (t ) f ( x)e j 2Gx x t dx f ( x) A M xy ( x,0 )et / T2 ( x) Let u Gxt u F (u ) s o ( ) f ( x)e j 2ux dx Gx We can now find our image as a function of x by taking an inverse Fourier Transform A Simple Example of Spatial Encoding with Frequency Encoding Constant Magnetic Field w/o encoding Varying Magnetic Field w/ encoding A Simple Example of Spatial Encoding with Frequency Encoding Frequency Decomposition Decays faster than T2* Extend this to a full 3D person Extend this to a full 3D person After slice selection we need to image in xy plane y x Spatial Encoding in xy plane Frequency Encoding Mathematical Analysis Now lets apply a gradient in the x direction (Gx) s (t ) A M xy ( x, y,0 )e j 2 o Gx x t e t / T2 ( x , y ) dxdy After demodulation (envelope detection) so (t ) A M xy ( x, y,0 )e j 2Gx xt e t / T2 ( x , y ) dxdy f ( x, y) A M xy ( x, y,0 )et / T2 ( x, y ) s o (t ) Effective proton density f ( x, y )e j 2Gx x t dxdy Spatial Encoding in xy plane Frequency Encoding Mathematical Analysis After demodulation (envelope detection) so (t ) A M xy ( x, y,0 )e j 2Gx xt e t / T2 ( x , y ) dxdy f ( x, y) A M xy ( x, y,0 )et / T2 ( x, y ) s o (t ) f ( x, y )e j 2Gx x t dxdy Let u Gxt u F (u, v 0) s o (t ) f ( x, y )e j 2ux dxdy Gx Spatial Encoding in xy plane Frequency Encoding Mathematical Analysis s o (t ) f ( x, y )e j 2Gx x t dxdy Let u Gxt u F (u, v 0) s o (t ) f ( x, y )e j 2ux dxdy Gx Corresponds to a single line or trajectory in the uv plane Spatial Encoding in xy plane Frequency Encoding Mathematical Analysis Now lets apply gradients in both the x direction (Gx) and y direction (Gy) After demodulation (envelope detection) so (t ) A M xy ( x, y,0 )e j 2Gx xt e j 2G y yt t / T2 ( x , y ) e dxdy Let f ( x, y) A M xy ( x, y,0 )et / T2 ( x, y ) s o (t ) f ( x, y )e Effective proton density j 2G x x t j 2G y y t e dxdy Spatial Encoding in xy plane Frequency Encoding Mathematical Analysis Now lets apply gradients in both the x direction (Gx) and y direction (Gy) s o (t ) Let f ( x, y )e j 2G x x t j 2G y y t e dxdy u Gx t , v G y t F (u , v) f ( x, y )e j 2uxe j 2vx dxdy Spatial Encoding in xy plane Frequency Encoding Mathematical Analysis Let u Gx t , v G y t F (u , v) f ( x, y )e j 2uxe j 2vx dxdy Polar Scanning Gradient Echo (A brief detour) Ts/2 Spatial Encoding in xy plane Gradient Echo Mathematical Analysis Spins will dephase very quickly (quicker than T2*) due to the gradient fields. After negative x-gradient ( x, t ) Gx xdt Gx xt p , p t p Ts 2 t p ( x, p Ts 2 ) Gx x Ts 2 Spatial Encoding in xy plane Gradient Echo Mathematical Analysis After negative x-gradient ( x, t ) Gx xdt Gx xt p , p t p Ts 2 t p ( x, p Ts 2 ) Gx x Ts 2 Now impose positive x-gradient for Ts ( x, t ) Gx x Ts 2 t p Ts 2 Gx xdt Gx x Ts T Gx x t p s 2 2 ( x, t ) Gx xTs Gx xt p , p Ts 2 t p 3Ts 2 Spatial Encoding in xy plane Gradient Echo Mathematical Analysis Now impose positive x-gradient for Ts ( x, t ) Gx xTs Gx xt p , p Ts 2 t p 3Ts 2 ( x, t p Ts ) Gx xTs Gx x p Ts p 0 Phase Encoding Pulse Repetition Image Contrast Image Quality Field of View and Resolution in MRI Fourier Plane Spatial Domain Dx Du Dv Vcoverage FOVy Dy Ucoverage FOVx Nyquist Sampling Theorem: Review • Assume we have a continuous signal with maximum frequency of fmax • To avoid aliasing we must sample the signal at a sampling frequency of fs>=2 fmax • The sampling interval T=1/ fs • fmax<=1/(2T) Sampling in MRI • Slice selection direction: sampling in z-direction Slice thickness (Dz) controlled by RF excitation bandwidth (D) To avoid aliasing 1 2 f max, z Dz 1 Dz 2 f max, z Where fmax,z is the highest spatial frequency in along the z-axis Sampling in MRI • Within each slice: sampling in xy plan We sample in the Fourier domain (u,v) (called k-space in MRI literature, kx=u, ky=v) Rectilinear Scan Du depends on sampling interval T during readout (ADC) Du depends on sampling interval during this time Sampling in MRI • Within each slice: sampling in xy plan We sample in the Fourier domain (u,v) (called k-space in MRI literature, kx=u, ky=v) Rectilinear Scan Dv depends on spacing between phase encoding Dv depends on the integrated phase shift here Sampling in MRI • Within each slice: sampling in xy plan We sample in the Fourier domain (u,v) (called k-space in MRI literature, kx=u, ky=v) Polar Scan Angle scan depends on steps in Gy/Gx Angle scan depends on steps in Gy/Gx Sampling in MRI • Within each slice: sampling in xy plan We sample in the Fourier domain (u,v) (called k-space in MRI literature, kx=u, ky=v) Polar Scan Rho spacing depends on sampling interval T during readout rspacing depends on sampling interval during this time Dv X-gradient relates dimension x with Larmor freq by ( x) o Gx x To avoid aliasing only frequency given below are measured o fs f ( x) o s 2 2 fs f Gx x s 2 2 X-gradient relates dimension x with Larmor freq by ( x) o Gx x To avoid aliasing only frequency given below are measured fs f ( x) o s 2 2 fs fs x 2Gx 2Gx o fs f Gx x s 2 2 Field of view in the x-direction (FOVx) is thus given by FOVx xmax xmin FOVx fs fs fs 1 ( ) 2Gx 2Gx Gx GxT 1 1 FOVx GxT Du Dependant on the phase encoding gradient Gy. The amount of phase change is given by D Dv DG yTPE Field of view in y FOVy ymax ymin 1 1 GyTPE Dv While FOV is limited by the sampling interval in the UV plane (Fourier plane) the resolution is limited by the total extent of the UV plane being sampled. If we ignore high spatial frequency content we will have lower resolution (blur our image). Since MRI scans cover only a finite area of the Fourier space we can expect a finite resolution. Fourier space coverage in MRI U cov erage N x Du N xGxT Vcov erage N y Dv N yDGyTPE Fourier space coverage in MRI U cov erage N x Du N xGxT Vcov erage N y Dv N yDGyTPE Outside of this range we assume the contributions to be zero. This is equivalent to passing the actual image through a low-pass filter in the uv plane whose transfer function is given by u v H (u, v) rect ( ) rect ( ) U cov erage Vcov erage In the spatial domain this is then given by the point spread function (PSF) h( x, y ) U cov erage Vcov erage sin c(U cov erage x) sin c(Vcov erage y ) In the spatial domain this is then given by the PSF h( x, y ) U cov erage Vcov erage sin c(U cov erage x) sin c(Vcov erage y ) The full width half max (FWHM) resolution is given by the width of the sinc function’s main lobe FWHM x FWHM y 1 U cov erage 1 Vcov erage 1 1 N x Du N xGxT 1 1 N y Dv N yDG yTPE Increasing the U,V (coverage area in Fourier space) reduces blurring. Dx FOVx 1 1 M MDu U cov erage Dy FOV y N 1 1 NDv Vcov erage Field of View and Resolution in MRI Spatial Domain Fourier Plane Du Dx Dv Vcoverage FOVy Dy Ucoverage FWHM x FWHM y FOVx 1 U cov erage 1 Vcov erage 1 Du 1 FOV y Dv FOVx