Ultrasound Physics
Artifacts
Our Hospital Physics Group
George David, M.S.
Associate Professor of Radiology
Nothing to do with
Anything
Artifacts
• Assumptions can cause artifacts
when assumed conditions are
not true
sound travels at 1540 m/s
sound travels in a straight line
All sound attenuation exactly
0.5 dB/cm/MHz
Distance from Transducer
• Calculation of Distance
scanner accurately measures
time delay between sound
generation & echo reception
Distance = Assumed Speed X Measured Delay / 2
Actual Distance to interface
1380 m/s X 58usec / 2 = 4 cm
Calculated Distance to interface
58 usec
V = 1380 m/s
1540 m/s X 58usec / 2 = 4.47 cm
(Average speed of sound in soft tissue)
4
cm
Distance from Transducer
• Echo positioning on
image
distance from transducer
calculated from assumed speed
of sound
can place reflector too close to
or too far from transducer
can alter size or shape of
reflector
V = 1380 m/s
Actual Object Position
X Position of Object on Image
V = 1540 m/s
X
Attenuation
• For all scanning your scanner
assumes
soft tissue attenuation
» .5 dB/cm per MHz
• Your scanner’s action
compensate for assumed
attenuation
allow operator fine tuning
» TGC
Shadowing
• Clinical Manifestation
reduction in imaged reflector amplitude
• Cause
object between this reflector & transducer
attenuates ultrasound more than assumed
Attenuates
assumed compensation not enough to
more than .5
provide proper signal amplitude
dB/cm/MHz
intensity under-compensated
• Opposite of Enhancement
Shadowed
Reflector
Shadowing
Attenuates
more than .5
dB/cm/MHz
Shadowed
Reflector
http://raddi.uah.ualberta.ca/~hennig/teach/cases/artifact/noframe/imag2-f2.htm
Enhancement
• Clinical Manifestation
increase in imaged reflector amplitude
• Cause
object between reflector & transducer
attenuates ultrasound less than
assumed
assumed compensation more than
needed to provide proper signal
amplitude
intensity over-compensated
• Opposite of Shadowing
Attenuates
less .5
dB/cm/MHz
Enhanced
reflector
Enhancement
Attenuates
less .5
dB/cm/MHz
Enhanced
reflector
http://raddi.uah.ualberta.ca/~hennig/teach/cases/artifact/noframe/imag6-f1.htm
Shadowing / Enhancing
• these artifacts not necessarily
bad
• can help in determining nature of
masses “upstream of artifact
which caused shadowing /
enhancing
Scanner Assumptions
• Echo positioning on
image
direction of all sound travel
assumed to be direction that
sound was transmitted
Actual Object Position
X Position of Object on Image
X
Refraction
Refraction Artifact
• refraction alters beam
direction
• direction of sound
travel assumed to be
direction sound
transmitted
Actual Object Position
X Position of Object on Image
X
Refraction
Refraction Artifact
• refraction alters beam direction
• scanner places dot in wrong location
along line of assumed beam direction
• can alter reflector shape
Lobe Artifacts
• Side Lobes
beams propagating from a
single element transducer
in directions different from
primary beam
reflections from objects
here will be placed on main
sound transmission line
• Grating Lobes
same as above except for
transducer arrays
X
Range Ambiguity
• Reflection from 1st
pulse reaches
transducer after 2nd
pulse emitted
scanner assumes this is
reflection from 2nd pulse
places echo too close & in
wrong direction
1
2
Range Ambiguity
• To improve any 1 of
3, at least 1 of other
2 must be reduced.
• many scanners
automatically
reduce frame rate
as depth increases
Lines / Frame
Depth
Range
Ambiguity
Trade-off
Frames / sec
(dynamics)
Scanner Assumptions
Multipath
Artifact
Actual Object Position
X Position of Object on Image
X
Multiple Reflection Scenario
• reflection from reflector
“B” splits at “A”
• some intensity rereflected toward “B”
• Result
1 2
3
A
B
later false echoes heard
scanner places dots behind
reflector “B”
1
2
3
real
false
Artifacts
• Reverberation (multiple
echo) artifact
“comet tail” effect is 1 example
can have dozens of multiple
reflections between
» transducer & reflector
» 2 reflectors
• Mirror Image
common around diaphragm
& pleura
Real
Mirror
Artifacts
http://raddi.uah.ualberta.ca/~hennig/teach/cases/artifact/noframe/imag1-f1.htm
Caused by Shotgun Pellets
Multiple Reflection Scenario
Real
Mirror
http://raddi.uah.ualberta.ca/~hennig/teach/cases/artifact/noframe/imag5-f2.htm
Resolution Artifacts
• Axial and Lateral Resolution
Limitations
results in failure to resolve 2 adjacent
structures as separate
minimum image size equal to resolution in
each direction
Section Thickness Artifact
• anatomy may not be
uniform over its thickness
• universal problem of
imaging 3D anatomy
• in CT & MRI this is known
as partial volume effect
Thickness
Constructive Interference
• 2 echoes
received at
same time
• in phase
• Result
+
higher intensity
=
Destructive Interference
• 2 echoes
received at
same time
• Exactly 180o
out of phase
• Result
flat (zero) wave
=
Acoustic Speckle
• texture seen on image
may not correspond to
tissue texture
• Results from
interference effects
between multiple
reflectors received
simultaneously which
can
add together
» constructive interference
subtract from one another
» destructive interference
Mirror Image & Doppler
• Analogous to mirror image artifact
discussed previously
mirrored structures can include mirrored vessel
duplicate image visible on opposite side of strong
reflector
example: bone
• Doppler data also duplicated
flow & spectrum copied from original vessel
Spectral Duplication
• mirror image of Doppler
spectrum appears on
opposite side of baseline
• causes
electronic duplication caused by receiver
gain set too high
» overloads receiver
True sensing caused by too large Doppler
angle
» beam covers flow in both directions
Blood flows
toward
transducer
Blood flows
away from
transducer
Doppler Artifacts
• Doppler spectrum speckle
• Cause
same as acoustic speckle
random constructive & destructive
interference from sound scattered in blood
Aliasing
• Results in detection of improper flow
direction
• occurs because sampling rate too
slow
• Similar to wagon wheels rotating
backwards in movies
Aliasing
Sufficient Sampling
Insufficient Sampling
Aliasing
• Which way is this shape turning?
#1
#2
#3
Aliasing
Did the shape turn 1/4 turn right
or
3/4 turn left?
1 1/4 turn right?
#1
#2
#3
Aliasing
Does it help to sample more often?
#1
#1A
#2
#2A
#3
#3A
Aliasing
• Maximum detectable Doppler shift
equals half the pulse repetition
frequency
• Sampling rate
Same as pulse repetition frequency
Must be at least twice highest frequency to be
sensed
• Aliasing occurs when Doppler shift
exceeds 0.5 * PRF
Aliasing
• Maximum detectable Doppler Shift
not limited for continuous wave
Doppler
• Maximum detectable Doppler Shift is
limited for pulsed instruments
Maximum Detectable Doppler Shift = half pulse repetition frequency
Coping with Aliasing
• decrease transducer frequency
reduces Doppler shift
shift proportional to operating frequency
• increase pulse repetition frequency
decreases maximum imaging depth
increases likelihood of range ambiguity for pulsed
instruments
77 X fD (kHz)
v (cm/s) = -------------------------fo (MHz) X cosq
Coping with Aliasing
• increase Doppler angle
Reduces relative flow rate between blood & transducer
Reduces Doppler shift sensed by scanner
77 X fD (kHz)
v (cm/s) = -------------------------fo (MHz) X cosq
q
Coping with Aliasing:
Baseline Shifting
• operator instructs scanner to assume that
aliasing is occurring
scanner does calculations based on operator’s assumption
• scanner has no way of determining where in
image aliasing occurs
Any Questions?