What you need to know to perform biofeedback effectively Article 4: Why you need to know how biofeedback devices work By Richard A. Sherman, Ph.D. Research has shown that practitioners who possess an adequate technical understanding of how their biofeedback instrumentation works have more successful outcomes. Fundamental to effective biofeedback is accurate and valid electrophysiological measurement that produces signals that are as free of artifact as possible. When instruments produce feedback that is an accurate reflection of that which is taking place within the body, learning, which is the goal of biofeedback and neurofeedback training, is made possible. The topics below are essential to a sufficiently inclusive course in biofeedback instrumentation: 1. Ohm’s law (the relationship between resistance to the flow of current, amount of current, and voltage / signal magnitude) – how it effects SEMG, EEG, and especially, SCL / GSR signals. 2. Use of the (a) offset, (b) averaging / integration, (c) gain / amplification, and (d) sweep speed (time it takes the signal to cross the screen) to produce an optimal feedback signal from which patients can learn. For example, having your device set these automatically may make learning difficult or impossible? 3. The overwhelming effect of bandwidth / filters on whether the signal has been recorded relative to its power spectrum. 4. How to properly attach the sensors for each physiological parameter so movement artifacts are minimized and the signal rather than noise is recorded. Placement of reference sensors is included. 5. How to easily test each signal before beginning a session to insure that it is recording properly and that the signal changes with changes in the subject’s physiology. 6. How to recognize noise and interference from other signals during a recording. 1 7. How to reset the equipment as a session progresses so you can use shaping techniques to teach patients physiological control by properly setting the feedback display. 8. How to protect your patient from electrical shock when using typical biofeedback devices. 9. Understanding the display when you record: a. Muscle Tension b. Heart Rate / heart rate variability c. Sweating in response to stress / GSR / SCL d. Respiration e. Temperature (for blood flow) f. Electroencephalogram (EEG)? 10. Understanding how these signals are actually generated to avoid common mistakes such as recording across rather than along a muscle? 11. Understanding the relationships between changes in sweating and changes in GSR signals such as conductance and resistance or the relationships between conductance and resistance? 12. Understanding the effect of different devices on the magnitudes of the signals you see? The following pages include a more in-depth view of the kind of information biofeedback practitioners need to understand in order to perform biofeedback effectively. Bandwidth The amplifiers on most biofeedback and psychophysiological recording devices can be set to a variety of frequencies. The basic idea is to set the amplifier to receive the relevant frequencies produced by the physiological signal being recorded (such as sEMG from a muscle) while filtering out frequencies produced by unwanted sources such as the lights in the room, other physiological signals, etc. The window of frequencies an amplifier is set to record is referred to as its bandwidth. An example would be 90 – 500 Hz. No significant amount of the signals picked up by the sensors that are produced by frequencies above or below the bandwidth are amplified, so they don’t appear in the recording. Any power from 0 – 89 Hz and above 500 Hz is never seen. 2 Power Spectrum When examining a raw sEMG signal, many different frequencies are represented. The total amount of the muscle’s electricity produced at any particular level of tension is unevenly divided between the various frequencies. A muscle’s power spectrum refers to the amount of electrical potential in each of the frequencies of waves occurring at a particular level of tension. When a muscle is relaxed, most of the power may be in lowerfrequency waves. This may change as the muscle tenses so that more of the power is in higher-frequency waves. A muscle’s power tends to decrease in frequency with fatigue, so the power spectrum changes during recordings in which the subject is trying to control the muscle’s activity. All principles of the power spectrum apply to the EEG as well. When you look at a raw EEG signal, many different frequencies are represented. The total amount of electricity recorded by surface EEG sensors is unevenly divided between the various frequencies. In the case of the EEG, power spectrum refers to the amount of electrical potential in each of the frequencies of waves recorded from a particular part of the brain at any moment in time. Improperly Set Amplifiers In order to record power properly, it is necessary to set the bandwidth to include the frequencies containing most of the muscle’s power and at various levels of tension. This is illustrated on the next page as follows: Part “a” of the following figure illustrates the effect of changing the amplifier’s bandwidth on how much of the signal actually gets recorded. Note that as the bandwidth is narrowed from 90 – 1,000 Hz (cycles per second) through 250 Hz down to 150 Hz, much of the signal’s characteristics disappear as the more rapid deflections are cut out. This reduces the overall amount of power recorded. Part “b” shows you just how much power much power will be lost. 3 reduction. The following illustration shows what you see on biofeedback meter displays as the frequency block containing most of a muscle’s power shifts from low to high frequencies as the muscle’s tension changes. Two meters are shown in each part of the figure. The upper meter shows the reading of relative tension when recording the wide bandwidth of 8 – 500 Hz (cycles per second) while the lower meter shows the reading you see when The illustration above shows the biofeedback meter display as the frequency block containing most of a muscle’s power shifts from low to high frequencies as the muscle tension changes. Two meters are shown in each part of the figure. The upper meter shows the reading of relative tension when recording the wide bandwidth of 8-500 HZ (cycles per second) while the lower meter shows the reading seen when recording only a narrow bandwidth of 100 – 200 Hz within the wider bandwidth. The heavy, dark line represents how much power the muscle being recording is putting out at each frequency. The top part of the figure shows the situation when most of the power is in lower frequencies so the biofeedback displays for both wide and narrow bandwidth show low readings (needle off to the left) as there is little total tension and most of the power is outside the area recorded by the narrow bandwidth – so it misses most of what is going on. The middle part of the figure shows what happens when tension has changed so that most of the power is in the middle of the frequency spectrum centered at about 200 Hz. Thus, both meters go up. The problem with using a narrow bandwidth becomes evident in the lower part of the figure, which shows the situation when most of the power has shifted to higher bandwidths. The total amount of tension in the muscle has gone up 4 so the top meter shows a higher reading. However, the meter recording only the narrow bandwidth goes down because little power is in the 100 to 200 Hz range. 8 5 100 200 500 Power Spectra For Two Muscles The power spectrum put out by muscles is different 1) for different muscles at the same level of tension and 2) the same muscle at different levels of tension. Power tends to decrease with fatigue, so incorrectly set filters are likely to give less and less accurate readings as fatigue sets in. The following figure shows the tremendous differences in patterns of power in two muscles. The three-dimensional graphs show the power spectra of two muscles as their tension changes over time. The one on the left is in blue and the one on the right is in red. Power (amount of electricity) is on the vertical axis, time goes from front to back, and frequency changes from left to right. This is an illustration of differing spectra further explained in the next figure. Muscle 1 Muscle 2 Time Power 0 60 500 0 60 500 Frequencies recorded for each muscle (Hz) 6 This is an example of the power spectra of three muscles recorded at about the same level of tension. If the recording used only a narrow bandwidth of 100 – 200 Hz, the meter displays would look very different as different amounts of power for each muscle happen to be within the 100 – 200 Hz range. 7 Signal Control A signal can be controlled best if it can be displayed properly. The display should be set so the signal is 1) the right size to provide enough information to learn from, 2) entirely in view on the monitor, and 3) present for sufficient time before the screen refreshes itself so you can see changes. To make an optimally effective display, control the (1) gain (amplification), (2) offset, and (3) sweep speed (horizontal deflection speed). A signal’s gain is the amount of amplification (volume). In the following figure, the gain is initially set so low that important changes in the signal can’t be detected. Here’s a situation in which the gain needs to be lowered to use the signal appropriately. The screen’s viewable area is shown by the red box. HIGH GAIN LOWER GAIN 8 Initially, an important part of the signal is off the viewable area of the monitor, so the size of the signal needs to be reduced. The offset controls the vertical position of the signal on the screen. It does not change the gain (size/magnitude) of the signal. The screen’s viewable area is shown as a red box. INCORRECT CORRECT Initially, an important part of the signal is off the top of the monitor’s viewable area, so the vertical position of the signal needs to be lowered in order for the entire signal to be seen. The deflection time or the rate at which the signal moves across the monitor’s screen from left to right is called the sweep speed. This has to be controlled so you can see enough of the signal’s history to make sense of what is happening. When the signal reaches the right edge of the screen, the display resets itself, so everything that was on the screen disappears, and the signal starts over at the left edge. If the signal is set so it moves across the screen too quickly, too little of the history is available on the screen to tell what happened throughout a series of actions. If it is too slow, the changes can’t be seen because they pile up on each other. For example, if you ask a patient to tense a muscle, hold it for 10 seconds and then relax, you want to see the entire sequence. This event is shown by the blue arrow.Setting the speed so the signal moves across the screen in five seconds would mean that you might see the first part of the movement but not the entire sequence. 9 5 sec Setting it at 10 minutes would make the changes pile up, so you couldn’t tell them apart. 600 sec Setting it at 30 seconds is about right to see the entire sequence. 30 sec What’s next? This article was intended to provide you with some valuable information to use in improving your biofeedback recordings and displays. If there are any topics discussed in this paper that you don’t truly understand, it is recommended that you take the instrumentation portion of a BCIA approved biofeedback course. See the provider training sections on BCIA and AAPB’s websites. AAPB frequently offers a biofeedback instrumentation course during its annual conference. These courses teach the concepts but are not tied to one specific type of equipment. 10