PTR-TOF-MS: A New Instrument For Real-Time Analysis Of Multi-Component Systems: Applications To Food Analysis Caroline Lamont-Smith, Steve Mullock & Fraser Reich Kore Technology Ltd. Ely, Cambs Rob Linforth, Annie Blisset, Andy Taylor Dept. of Food Science University of Nottingham Talk: Outline First a confession: This talk is heavy on the instrument description, its properties and possibilities of the instrument, because the speaker is one of the team that designed the instrument. • Overview of the Proton Transfer Reactor method • Why a PTR-TOF-MS? • The Instrument • Some Example Data • Summary PTR: Proton Transfer Reactions Essentially, PTR is a subset of Chemical Ionisation. It was defined and refined by Werner Lindinger in the 1990’s, and resulted in the first commercially available instrument (Ionicon) • Aim is to achieve ‘Soft’ ionisation so as not to fragment the molecule(s) of interest: less fragments = less mass spectral ‘noise’. • In positive ion mode, most frequently used reaction is: H 3 O+ + R H2O + RH+ • For this protonation reaction to occur, the proton affinity of R must be greater than that of water • The ionisation probability is almost the same as the collision cross-section, so the aim is to induce sufficient collisions to ensure efficient analyte ionisation PTR: Proton Transfer Reactions With H3O+ Compound PA (kJ/mol) Oxygen 421 Nitrogen 493 Carbon Dioxide 541 Sulphur Dioxide 672 Water 691 Hydrogen Sulphide 709 Benzene 750 Methanol 754 Toluene 784 Naphthalene 803 Ethyl Acetate 836 Tri ethyl phosphate 909 Compounds above water in the proton affinity table will not be ionised, whereas compounds below will The good news is that most VOCs have higher proton affinities than water and will receive a proton from H3O+ Generally, larger molecules have larger proton affinities Linearity of Response Provided that [RH+]<< [H3O+], the H3O+ signal does not change with analyte concentration, and the detected analyte intensity is linear with concentration: [RH+] = [H3O+]0(1-e-k(R)t) [H3O+]0 [R] kt Plot of H2S Peak Yield vs. H2S Concentration (H2s in Air) (k is rate constant for the reaction, and t is time to travel through the drift tube) 500 0-50ppm H2S calibration: 450 400 H2S Peak Yield (x50) 350 Raw data plot 300 Series1 250 Linear (Series1) 200 150 H2S proton affinity= 709kJ/mol 100 50 0 0 10 20 30 Calibrated H2S Concentration(ppm) 40 50 60 Soft Ionisation vs. Electron Impact Ionisation Significant fragmentation to lower masses Molecular Ion appears at mass 116 m/z 70eV EI spectrum of Ethyl Butyrate Soft Ionisation vs. Electron Impact Ionisation Counts 5000 PTR-TOF-MS Spectrum of 5ppm Ethyl Butyrate 4000 Molecular Ion at 117 m/z 3000 (molecular ion + proton) 2000 1000 Very little fragmentation 0 60 70 ce6.1, 15ml.min, ethylbuty, 2.0mbptr 80 90 100 110 Mass (m/z) 120 130 140 Charge Transfer In The PTR Reactor • Need to induce sufficient collisions: H3O+ and analyte mixed in a reactor tube (drift cell) • Voltage gradient across reactor • Pressure ~ 1 mbar, approximately 2000 collisions down 10cm length reactor • Aim is to have dilute analyte, so that analyte molecules do not collide with anything except H3O+ . This keeps the ionisation scheme and calibration simple, unlike in Ion Mobility Spectrometry (complex inter-analyte charge transfer reactions) The PTR Reactor and the E/n Ratio • E is the voltage gradient down the Analyte in reactor in V/cm Reagent gas in GD Cathode GD Anode GD SD PTR • n is the gas density in molecules / cc (1mbar in reactor, typically) • E/n in Townsend; 1 Townsend = 10-17 cm2 / Vs • Lindinger et al. (1998) showed that 120-140 T is ideal for nonfragmentation of organic molecules • If E/n too high, fragmentation of molecule occurs Transfer Optics • If E/n too low, clustering of water molecules occurs and can present a problem for certain species The PTR Reactor and the E/n Ratio Counts Tri-ethyl phosphate, protonated mass 182 60000 130 Townsend 50000 40000 30000 20000 10000 Counts 60000 0 100 120 140 Mass (m/z) 160 180 50000 40000 30000 20000 Counts 10000 120000 0 260 Townsend 100000 100 80000 40000 0 100 120 140 Mass (m/z) 160 160 180 ethyl groups (28 mass units) 20000 TEP mass 182 140 Mass (m/z) 195 Townsend – loss of the PO4 + proton 60000 120 180 200 Why Use A TOF-MS? • Parallel detection means no analytical price to pay for monitoring many species, unlike a sequential analyser such as a quadrupole or magnetic sector • TOF Cycle frequency typically 20-50 kHz = high data rate • Full mass spectra to several hundred masses with 4-5 orders dynamic range in one second • Possibility of ‘real time’ data analysis with < 100ms resolution, multiple species and sensible counting statistics • Ability to interrogate a data set ‘retrospectively’: intensity of any species as a function of time, mass spectra for a variety of ‘time slices’ • Parallel detection and full mass spectra more suitable to software data reduction techniques, e.g. principle component analysis Why Use A TOF-MS? Mass spectrum from breath (U. Nijmegan) This mass spectral plot, acquired with a PTR-Quad-MS (a sequential scanning device) took 40 seconds A full mass spectrum to several hundred m/z can be acquired in less than a second with a PTR-TOF-MS TOF-MS: Accurate Mass Capability 30.00 m/z NO+ 31.99 m/z O2+ Counts 33.04 m/z CH3OH2+ 36.04 m/z (H2O)2+ 1500 1000 500 0 29 30 31 32 33 Daltons 34 35 36 Downside Of A TOF-MS? • Main drawback is analysis of a continuous analyte stream: When the TOF cycle has started, no further ions can be injected without resulting in multiple overlapping spectra • ‘Duty cycle’ describes the % of the analyte stream that is sampled • In an orthogonal TOF (analyte stream enters TOF source at 90°), the duty cycle can reach 10%: when the Kore TOF source pulses, it ‘empties out’ ~5-10 microseconds worth of ions. Usage is therefore 5µs in 50µs, I.e. 10% • Currently we are observing up to 500kc/s of Hydronium at the detector, approximately half that reported by the PTR-Quad (single ion mode). • Prospect of improving duty cycle using ‘Hadamard’ methods – high frequency random pulsing of source with mathematical deconvolution of spectra - still not really proven as a workable solution TOF MS: Orthogonal Pulsing TOF Source Continuous Sample TOF Source Ions Extracted 0 Volts Source Off -2kV Extractor pulses to –400v t Source On Detector Light Ions Heavy Ions Ions of different masses within a single ‘cycle’ arrive at the detector at different times according to the relation: K.E. = mv2/2 Variant on ‘Wiley – Maclaren’ source. Compensation for ion position within source Ions with m/z = 1000 has flight time ~20-50s, therefore ‘cycle time’ = 20-50s, so typical pulsing frequency = 20-50 kHz TOF MS: Orthogonal Pulsing TOF Source Overview schematic of the instrument: TOF source to detector TOF MS: Orthogonal Pulsing TOF Source GD Source 2mbar Overview schematic of the instrument PTR Reactor 1mbar PTR Source Transfer Optics Transfer Optics 10-4 mbar Deflectors Reflectron Extract Pulser Field Free Region Blanker Blanking Pulser TOF Source Detector PreAmplifier Mass spectrometer and detector 10-6 - 10-7 mbar Mass Spectral Analysis: Applications to Food Analysis University of Nottingham Department of Food Science Professor Andy Taylor Dr. Rob Linforth Annie Blisset • “Breath-by-breath” Research • Main technique in lab: APCI • Wanted to add PTR-TOF-MS • Instrument recently delivered; has both PTR and APCI functionality Data acquired at Nottingham by one of us (FR) with Rob Linforth and Annie Blisset Breath-by-Breath Analysis Of Juicy Fruit Gum The principle compound released during chewing of Juicy Fruit gum is ethyl butyrate, molecular mass 116 • Soft tube inserted lightly into the nostril of subject • Breathing normally, the mass spectrometer begins acquiring data. • Most of the exhaled breath passes out into open space, but a side-mounted capillary pipe samples the breath into the proton transfer reactor. Signal intensity Breath-by-Breath Analysis: Release of Ethyl Butyrate Idealised appearance of breath markers, such as acetone, as exhalation occurs Time Release of flavours during mastication • First data set: take ‘raw data set’ (all ion flight times recorded as a function of elapsed experiment time) • Integrate the ethyl butyrate protonated ion (mass 117) every second Breath-by-Breath Analysis At Different Data ‘Granularity’ Intensity 2500 Reconstruction of Acetone signal with 1 second integration time 2000 1500 1000 Breathing rate ~5/min 500 0 10 20 Signal for Mass 117 30 40 50 Time 10000 Reconstruction of ethyl butyrate signal with 1 second integration time 8000 6000 4000 2000 0 0 1 second integration 10 20 30 Time in Seconds 40 50 Breath-by-Breath Analysis At Different Data ‘Granularity’ Signal for Mass 117 6000 Reconstruction of ethyl butyrate signal with 0.5 second integration timefurther structure emerging 5000 4000 3000 2000 1000 0 0 10 20 0.5 second integration Signal for Mass 117 30 Time in Seconds 40 50 60 Reconstruction of ethyl butyrate signal with 0.25 second integration timefurther structure emerging still 3000 2000 1000 0 0 0.25 second integration 10 20 30 Time in Seconds 40 50 60 Breath-by-Breath Analysis At Different Data ‘Granularity’ Signal for Mass 117 Reconstruction of ethyl butyrate signal with 0.125 second integration time 2000 1500 1000 500 0 0 10 20 0.125 second Signal for integration Mass 117 30 Time in Seconds 40 50 60 1200 Reconstruction of ethyl butyrate signal with 0.0625 second integration time: no further structure emerging 1000 800 600 400 200 0 0 0.0625 second integration 10 20 30 Time in Seconds 40 50 60 Breath-by-Breath Analysis: Different Person Signal 1400 Reconstruction of Acetone signal with 0.125 second integration time 1200 1000 800 600 400 200 0 0 10 20 Mass 59 30 Time (seconds) 40 50 Note different breathing rate ~10/min for different person Signal Overlay of ethyl butyrate signal with acetone 2000 1500 1000 500 0 0 Mass 59 and 117 10 20 30 Time (seconds) 40 50 Full Mass Spectra From 125ms Time Slices Counts/0.125s 4 34.50 –34.625 seconds 3 2 Acetone and Ethyl butyrate 1 0 20 40 60 Counts/0.125s 4 80 Daltons 100 120 140 34.00 –34.125 seconds 3 2 1 0 20 40 60 80 Daltons 100 120 140 Mass Spectrum From Headspace of Freshly Macerated Tomatoes Counts 3 hexenol 4000 3 hexenal 3000 2000 1000 0 60 80 100 Daltons 120 140 3-Hexenol vs. Hexanal, Both C6H12O In previous slide, masses 101 and 83 were identified as 3Hexenol, but in truth there is the possibility that 101 and 83 can be due to Hexanal. Hexanal: protonated masses at 101, 83, 55 3-Hexenol: protonated masses at 101, 83, 59 and 55 3-Hexenol vs. Hexanal, Both C6H12O Is it possible to find an E/n value that will give a mass 101 for one compound but not the other? Counts Counts 1000 1400 101 Hexanal 800 83 83 1200 3 Hexenol 1000 600 800 101 600 400 400 200 200 0 0 80 85 90 95 Daltons 100 105 80 85 90 95 100 Daltons No ‘threshold phenomenon’ observed Also, no E/n found for 3-Hexenol at which intensity of 101>83 Possibility of modulating PTR voltage and using software tools for compound differentiation? More work required, clearly! 105 Summary • PTR-MS instruments based on quadrupole mass spectrometers are now widely used in studies of environmental and atmospheric chemistry as well as food and medical applications • A TOF-MS increases the possibilities of real-time analysis down to <100ms “data granularity”, with full mass spectral acquisition in each time slice • A TOF-MS permits any ion chromatogram to be re-constructed from the data set • A TOF-MS can be operated at higher mass resolutions than a quadrupole mass spectrometer. Even at relatively low mass resolution the mass accuracy is better than 30 millimass units • Greater mass range capability, with no discrimination against high masses up to several hundred Daltons