Experiment No…… ObjectiveTo plot a graph showing the variation of magnetic field with distance along the axis of a Helmholtz galvanometer and calculate the reduction factor ‘k’, horizontal component of earth’s magnetic field ‘H’ and the radius of coil ‘r’ from graph. Apparatus requiredHelmholtz galvanometer, variable power supply, an ammeter, commutator and connecting wires TheoryHelmholtz galvanometer-A Helmholtz galvanometer, like other galvanometers, is an instrument which can be used to measure dc current in the circuit .It is an improved form of the tangent galvanometer. It consists of two identical coils placed coaxially at a distance equal to the radius of the either coil. The two coils are of the same radius on the same direction. The field at any point on the axis of the coil is equal to the sum of the magnetic field due to the individual coil. It consists of two identical circular coils, each having several hundred turns in it, mounted coaxially at a distance equal to the radius. In the side of each coil, four terminals are provided. First of which is internally connected with the starting end of the coil, second is connected with the end of the second turn, third with the end of the 50th turn and fourth with the 500th turns. With the help of these terminals, equal number of turns of each coil can be connected in series to allow the same current to flow in the same direction For a coil consisting of n turns of wire and having a mean radius r, the magnetic field at a point on the axis at a distance x from the center of the coil is given by F= 2πnir2 3 (r2 +x2 ) ⁄2 Where I is the current passing through the coil. The rate of variation is given by dF = -3x [(2πnir 2 )(r 2 + x 2 ) dx d2 F dx2 = -6πni𝑟 2 [(r 2 + x 2 ) −5⁄ 2 d2 F −5⁄ 2] - 5𝑥 2 (r 2 + x 2 ) −7⁄ 2] dF From which x = ±r/2, if dx2 = 0 or dx = constant If there are two identical coils having the same axis and caring the same current in the direction with their centers r cm. apart, the rate of increase of field due to one coil at the midpoint between the coils is equal to the rate of decrease of field due to the other at the same point. Therefore, if one moves away along the axis from the midpoint, any diminution in the intensity of the field due to one coil is compensated by the increase in the field due to the other so that the field between the coils is practically uniform. The field at the midpoint is given by F=2* 2πnir2 3 [r2 +(r⁄2)2 ] ⁄2 = 32𝜋𝑛𝑖 𝑟 √125 If the coils are in magnetic meridian will be perpendicular to H. If the needle shows the deflection θ, 32𝜋𝑛𝑖 𝑟√125 Or I = = H tan θ r H √125 tan θ 32 𝜋 𝑛 = k tan θ , where k (expressed in emu) is a constant for the galvanometer at a particular place called the reduction factor of the galvanometer. Also k = I / tan θ Where k and I are in amperes. Procedure 1. Place the magnetometer compass box on the sliding bench so that its magnetic needle is at the centre of the coil. By rotating the whole apparatus in the horizontal plane, set the coil in the magnetic meridian roughly. In this case the coil, needle and its image all lie in the same vertical plane. Rotate the compass box till the pointer ends read 0 – 0 on the circular scale. 2. To set the coil exactly in the magnetic meridian set up the electrical connections as shown in fig. Send the current in one direction with the help of commutator and note down the deflection of the needle. Now reverse the direction of the current and again note down the deflection. If the deflections are equal then the coil is in magnetic meridian. Otherwise turn the apparatus a little, 3. Using rheostat Rh adjust the current such that the deflection of nearly 450 produced in the compass needle placed at the centre of the two coils. Read both the ends of the pointer. Reverse the direction of the current and again read both the ends of the mean deflection at x =0. 4. The current is noted from the ammeter. 5. The magnetometer box is slowly moved over the bench and kept the distance 1,2,3,4,5,6,7,8,10,12,14,16,18& 20 cm. on east side from the center. The deflection for both the sides of the pointer is noted at these points of direct as well as reverse currents. The procedure is repeated for west arm. 6. Plot graph taking distances along X-axis and tan θ along Y- axis. 7. Mark the points of inflections on the curve. The distance between the two points will be the radius of coil ObservationsNumber of turns in the coil = …. Current in ampere = ….amp S.No. distance along Deflection of needle the axis from East arm West arm center (x) in Direct Reverse Direct Revers cm. e Mean θ tan θ θ1 θ2 θ3 θ4 θ1 θ2 θ3 θ4 1. 1 cm. 2. 2 cm. 3. 3 cm. 4. 4 cm. 5. 5 cm. 6. 6 cm. 7. 7 cm. 8. 8 cm. 9. 10 cm. 10. 12 cm. 11. 14 cm. 12. 16 cm. 13. 18 cm. 14. 20 cm. Calculation- Mean θ tan θ K = I⁄tan 𝜃 amperes H= 32𝜋𝑛𝑘 10 𝑟 √125 (k/10 is expressed in emu) ResultRadius of the coil from the graph is ………………..cm. The reduction factor of Helmholtz galvanometer for ……turns is ……..ampere. Horizontal component of earth’s magnetic field in the lab ........( Standard value = 0.35 Gauss). Precautions1. The coil should be carefully adjusted in the magnetic meridian. 2. All the magnetic materials and current carrying conductors should be at a considerable distance from the apparatus. 3. The current passed in the coil should be of such a value as to produce a deflection of nearly 450. 4. Parallax should be removed while reading the position of the pointer. Both ends of the pointer should be read.