MuCool Test Facility Instrumentation Update M. Haney, D. Errede, Z. Conway University of Illinois at Urbana-Champaign 12/13 August 2002 Instrumentation • Seeking an intrinsically safe solution for instrumenting the LH2 absorber in the MTF – Prudent to accommodate MICE and European safety considerations – i.e. do it right the first time… 12/13 August 2002 m. haney, et al. 2 Objective - Safety • Intrinsic safety, if possible – Limited power per channel • Good idea for Fermi; required for Europe • Gas-purged enclosure, if necessary – 19” rack, 40U tall 12/13 August 2002 m. haney, et al. 3 Instrumentation (1) • Lakeshore 218S – 8 channel cryo temperature monitor • +/- 21m K at 10 K – 16 readings/s – GPIB interface (to PC) • TG-120PL GaAlAs diodes (8) – -180mV/K at 4.2 K – B < 5 T; rad hard 12/13 August 2002 m. haney, et al. 4 Instrumentation (2) • FISO Fiber-optic strain and temperature – BUS chassis, with 4 (up to 8) modules • 1000 readings/s • RS-232 interface (to PC) • FOS-N strain sensors – +/- 500 me; 0.01% full scale; 0.2mm O.D. • FOT-L temp sensors – 0.1 K resolution; 1.5mm x 32mm (10mm active) – Slow: 1.5 second response… 12/13 August 2002 m. haney, et al. 5 Instrumentation (3) • Gateway E-4000 PC – – – – 1.8 GHz, 1Gbyte RAM, 0.5Gbyte cache 120Gbyte disk, Windows 2000 15” LCD flat panel display 640x480 CCD camera, microphone • PCI-MIO-16E-1 – 16 channel ADC, 1.25 Msample/s, 12 bits, +/-10 V • Tripplite Smart UPS – 3000 VA 12/13 August 2002 m. haney, et al. 6 Instrumentation (4) • IRM – 64 channel multiplexed ADC • 16 bit, 100 Ksamples/s – ACNET network connection • All PC data will be posted to IRM for ACNET access 12/13 August 2002 m. haney, et al. 7 Instrumentation (5) • MTL7055ac barriers – Low level AC • 24 W per line • 3 V max • MTL7060ac barriers – Star-connected AC • 101 W per line • 8.5 V max 12/13 August 2002 m. haney, et al. 8 (basic) Barrier fuse Power + Signal “safe circuit” Return Ground 12/13 August 2002 m. haney, et al. 9 Signals inside (the cryostat) • 8 (cryo) temperatures – readout via GPIB + LabView (PC) • Multiple fibers – FISO strain, temperature (4) • Pressure, flow? – Laser occlusion (2) – Simple spark detectors (2) • (light pipe to photodiode) • 25(ish) 4-wire bolometry strips 12/13 August 2002 m. haney, et al. 10 (more) Signals inside • Piezo vibration sensor – via (16 chan/12 bit) ADC and LabView (PC) • Local (64 chan/16 bit) IRM channels – via “LabView” (see last slide) • MICE channels (?) – placeholder for the future… 12/13 August 2002 m. haney, et al. 11 Signals outside • USB (4 wires, and shield) for CCD camera and microphone – via PC • Many voltages, currents, temperatures, pressures – magnets, beam position monitors, etc. – Most available from ACNET or local IRM • via “LabView” (see last slide) – or Cyro-PLC 12/13 August 2002 m. haney, et al. 12 Two Possibilities (1) • Intrinsically safe Hazard Safe Sealed Conduit(s) Barrier(s) IRM ACNET Cryo(temp) FISO PC w/16 chan ADC Intrinsically safe signal conditioners and transmitters 12/13 August 2002 m. haney, et al. UPS network power 13 Two Possibilities (2) • Gas-purged box Hazard Safe Barrier(s) Sealed Conduit(s) ACNET IRM Cryo(temp) FISO PC w/16 chan ADC power UPS 12/13 August 2002 network m. haney, et al. 14 This Summer – Z. Conway • LabView interface – GPIB to cryo temp monitor - works – PC-internal ADC (e.g. for piezo) – “works” • Simple ADC data read - works • High rate collection and compression – not yet – IRM communications • IRM-ADC Read - works • Write – any moment now… • Integration – remains! 12/13 August 2002 m. haney, et al. 15 Open Issues • Wire+Shielding Concerns – Noise/sensitivity issues due to wire+barrier resistance? • Approx 12 W /ft for 36 AWG manganin wire • 48 W to 202 W added by barrier – And diode capacitance… – Common mode (surges) due to magnets? • Need to protect electronics without burning barriers – Cable plant into solenoid 12/13 August 2002 m. haney, et al. 16 Open Issues (2) • Software Concerns – Posting data to the IRM • For ACNET access – Reading data from ACNET • For PC access – Remote control of PC • Citrix ICA, PCAnyWhere, etc. • LabView web-remote instrumentation • KVM? 12/13 August 2002 m. haney, et al. 17 Open Issues (3) • FISO strain and temp sensors – Not official rated below -40 C (!) – Temp has been used at 35 K (custom) • No experience with strain gages – Company very willing to work with us • To adjust signal conditioner, correction curves, etc. • Will require some form of agreement – Confidentiality, exclusivity for the instrument – Yet freedom to use and publish measurements – Also pressure, flow? 12/13 August 2002 m. haney, et al. 18